Evidence in focus Clinical case series: Helping to prevent infection with intravenous (IV) therapy using IV3000 Moisture Responsive Intravenous Catheter Dressing

Transcript of Clinical case series - smith-nephew.com

Page 1: Clinical case series - smith-nephew.com

Evidence in focus

Clinical case series: Helping to prevent infection with intravenous (IV) therapy using IV3000◊ Moisture Responsive Intravenous Catheter Dressing

Page 2: Clinical case series - smith-nephew.com

Contents

IV3000 Dressing Ported IV3000 Dressing Non-Ported

Page 3: Clinical case series - smith-nephew.com

• Peripheral venous catheters (PVCs) For short-term use, usually in the hand or forearm

• Midline catheters For mid-term use (up to 4 weeks), inserted in the antecubital fossae with the tip in the axillary vein of the upper arm

• Central venous catheters (CVCs; non tunnelled)Temporary access in emergency and critical care situations via the subclavian, jugular or femoral veins; require close monitoring and have the highest risk of catheter-related bloodstream infections (CRBSIs)

• Tunnelled CVCs Can be indwelling for years; inserted into a jugular or subclavian vein

• Peripherally inserted central catheters (PICCs)Long-term use, particularly in oncology settings; inserted into a peripheral vein until the tip lies within a central vein

• Implanted ports Long-term intermittent therapy; inserted into a surgically created subcutaneous pocket and the port is connected to a central vein via a catheter

Types of vascular access devices1

Major risks with all vascular access devices are:2

• Increased risk of CRBSIs

• Dislodgement

• Accidental device loss

• Phlebitis

• Leakage of infusate

• Extravasion

Page 4: Clinical case series - smith-nephew.com

IV dressings should help to:3,4 • Prevent device dislodgement

• Facilitate assessment of the device and surrounding skin

• Protect the device entry site and thereby reduce the risk of infection

IV3000◊ Dressing with REACTIC◊ Film* incorporates the features of an ideal IV dressing4 • Specifically designed for IV sites,5 the dressing film

has a low-allergy adhesive with a unique grid pattern,6 is transparent,4,7 and is easy to remove7–9

• IV3000 Dressing helps to keep the skin dry and minimise the risk of skin maceration due to its high moisture vapour transmission rate (MVTR) compared with other film dressings5,10,11

• IV3000 Dressing helps to prevent moisture build-up and provide a bacterial barrier,12–15 which helps to minimise the risk of infection and external contamination

Reducing the risks of venous access devices using IV dressings1

Grid pattern adhesive:Low allergy 6

High MVTR: 5,10,11

Helps keep skin dry and reduces risk of skin maceration

*This resource reviews use of IV3000 Moisture Responsive Intravenous Catheter Dressing for frame delivery (not use of the infusion set).

Page 5: Clinical case series - smith-nephew.com

Case studies: real-world use of IV3000◊ Dressing

To learn more about the use of IV3000 Dressing with various vascular access devices in both hospital and ambulatory care settings, please select one of the following eight studies:4

Page 6: Clinical case series - smith-nephew.com

Case 1: Non-tunnelled CVC following surgery4

IV3000 Dressing in situ over a dual-lumen non-tunnelled CVC

Source: Steve Hill, Procedure Team Manager, Christie NHS Foundation Trust, Manchester, UK

Patient profile

• 64-year-old female undergoing elective surgery to remove a cancerous pelvic mass

• Preoperatively, a quadruple lumen non-tunnelled CVC was inserted in the patient’s internal jugular vein for postoperative management (in place for 7 days), which was secured with a ported IV3000 Dressing

Results with use of IV3000◊ Dressing

• Easy to apply and provided good securement while the CVC was in place

• Transparency of the dressing meant that the insertion site and skin could be easily assessed

• No signs of moisture build up, infection, erythema, tenderness or oozing at the entry site

• The patient did not find the CVC troublesome and found the dressing comfortable

• Remained intact throughout the postoperative period and was considered to be cost effective

Page 7: Clinical case series - smith-nephew.com

Case 2: CVC in an outpatient haematology unit4

Source: Steve Hill, Procedure Team Manager, Christie NHS Foundation Trust, Manchester, UK

IV3000 Dressing in situ over a dual-lumen tunnelled CVC

Patient profile

• 25-year-old female with acute lymphocytic leukaemia preparing for bone marrow transplant required a dual-lumen cuffed tunnelled CVC (for chemotherapy and other IV support)

– The patient was prone to infection

• At time of evaluation, the CVC, secured with an IV3000 Dressing (changed weekly), had been in place for 68 days

• Catheter site was examined daily over a 7-day evaluation period when the patient attended the outpatient haematology unit for treatment

Results with use of IV3000◊ Dressing

• Successful application as part of a care and maintenance strategy to help prevent infection in a vulnerable patient

• Over 6 days, there was slight redness (~0.5cm around exit site), which the patient reported had occurred previously; however, the site was not painful or uncomfortable (swabs and blood samples did not indicate bacterial growth)

• The patient found the dressing comfortable and showered daily without a protective cover over the dressing, which only started to peel slightly at the edges on day 7

– Peeling did not affect integrity over the CVC site

Page 8: Clinical case series - smith-nephew.com

Case 3: PICC in an ICU setting4

Source: Vicki Shawyer, Senior Vascular Access Nurse Specialist, Royal Devon and Exeter Hospital, UK

Patient skin before first IV3000 Dressing was applied

Patient skin after dressing removal

Patient skin intact despite 2-hourly patient repositioning

Patient profile

• 70-year-old male who developed sepsis after bowel cancer surgery and required a PICC for parenteral nutrition and IV medication (up to 14 days)

• The patient was febrile with clammy skin when the PICC was inserted, which was left in place for 6 weeks secured with a ported IV3000 Dressing (changed weekly)

Results with use of IV3000◊ Dressing

• The skin beneath the ported dressing was dry and intact at each dressing change with no signs of irritation; there was no need to switch to chlorhexidine gluconate-containing IV dressings to help prevent infection

• IV3000 Dressing secured the PICC line and its stabilisation device, reducing the risk of migration and contamination; the dressing design helped to secure the PICC line

• It was considered to be cost effective as it remained intact until the scheduled dressing changes

• IV3000 Dressing was easy to remove using the clear guidance on the order of removal; no adhesive was left on the skin and there was no skin irritation

• It remained intact despite the patient being repositioned every 2 hours, who did not complain of dressing-related discomfort

Page 9: Clinical case series - smith-nephew.com

Case 4: PICC in an acute oncology setting4

Source: Janice Gabriel, Former Nurse Director, Central South Coast Cancer Network, Portsmouth, UK

PICC immediately after insertion into the upper arm with IV3000 Dressing in situ

Patient arm after removal of PICC and IV3000 Dressing

Patient profile

• 77-year-old female receiving chemotherapy after breast cancer surgery

• Developed pyrexia due to neutropenic sepsis after 10 days into the third cycle of chemotherapy, which was administered via peripheral cannula

• Cannulation for IV antibiotics and fluids was difficult; therefore, a single-lumen, 4 French, open-ended PICC was required

• The insertion site was covered with sterile gauze and secured with a ported IV3000 Dressing

• Dressings were changed after 24 hours, the insertion site was cleansed with chlorhexidine gluconate and allowed to dry; a new IV3000 Dressing was applied (without gauze) and IV antibiotics were continued for 5 days

Results with use of IV3000◊ Dressing

• Transparency allowed easy observation of the PICC site; there were no signs of exudate or phlebitis

• The waterproof barrier protected the PICC site during showers and maintained its integrity; the skin was dry at each dressing change

• The patient found the PICC and IV3000 Dressing comfortable

• There was no redness or irritation upon removal of the dressing at patient discharge

Page 10: Clinical case series - smith-nephew.com

Results with use of IV3000◊ Dressing

• IV3000 Dressing was comfortable for the patient when assessed after 7 days

• It was easy to apply and transparency facilitated inspection of the PICC site

– It protected the PICC entry site, reducing the risk of contamination and accidental removal, and there was no build up of moisture under the dressing

• It did not restrict the patient’s arm movement, was easy to remove and did not mark the skin or affect skin integrity

• The patient could easily shower using a waterproof protector over the dressing and PICC

Case 5: PICC in a community setting4

Source: Nicole Moodley and Lesley Channer, Rapid Response Nurse Practitioners, Your Healthcare, Kingston, Surrey, UK

Skin integrity before application of IV3000 Dressing

Skin integrity upon removal of IV3000 Dressing after 2 weeks

Patient profile

• 65-year-old male with colorectal cancer was referred to a community IV therapy service to have a PICC inserted for chemotherapy

• The ported IV3000 Dressing was changed weekly and the entry site was inspected for phlebitis (used for more than 2 months in total)

Page 11: Clinical case series - smith-nephew.com

Case 6: Midline in a surgical ward4

Source: Vicki Shawyer, Senior Vascular Access Nurse Specialist, Royal Devon and Exeter Hospital, UK

Secure application of IV3000 Dressing

Removal of IV3000 Dressing did not impair skin integrity

Mobility did not affect conformability of IV3000 Dressing

Patient profile

• 41-year-old female with a complex bowel history who was admitted to hospital with abdominal pain and vomiting due to a bowel stricture; the patient could not tolerate oral medication

• An IV midline was inserted for administration of fluids, pain relief and anti-emetics

Results with use of IV3000◊ Dressing

• Skin was sweaty and clammy when the midline was inserted; bleeding at the entry site stopped after it was covered with gauze and a ported IV3000 Dressing for 24 hours

• IV3000 Dressing was then used without gauze for 14 days (replaced every 7 days); its transparency facilitated daily inspection of the insertion site for phlebitis

• The midline was kept secure and free from contamination; the ported top ensured the midline was secured well, reducing the risk of migration and contamination

• IV3000 Dressing remained intact over each 7-day period and was not affected by patient movement

• At dressing change, the skin was dry and skin integrity was similar to elsewhere on the arm

• The dressing was easy to remove, left no residue and there were no signs of skin irritation; the patient commented that IV3000 Dressing felt secure and comfortable compared with other IV dressings she had received and did not pull on her skin

Page 12: Clinical case series - smith-nephew.com

Case 7: Peripheral cannula in a community setting4

Source: Carly Edmed, Lead Intravenous Specialist Nurse, Kent Community NHS Foundation Trust, Kent, UK

Application of IV3000 Dressing with an adhesive strip to secure the cannula

Skin integrity was not impaired upon removal of IV3000 Dressing

Patient profile

• 48-year-old female who required IV antibiotics for an infection at the mastectomy site after breast cancer surgery

• Community nurses administered the antibiotics at the patient’s home, once daily, via a peripheral cannula secured with a ported IV3000 Dressing (4 days in total)

Results with use of IV3000◊ Dressing

• IV3000 Dressing easily covered the cannula and adhered well to the patient’s skin, reducing the risk of contamination

• Despite daily washing, the dressing remained intact and the skin was free of infection

• Transparency of the dressing permitted easy daily inspection for infection, allergic reaction or problems with the line (eg, misplacement or migration)

• Nurses found the dressing easy to apply and remove assisted by its clear markings

• The patient reported that IV3000 Dressing did not limit daily activities

• IV3000 Dressing was removed at the end of treatment using a stretch technique causing minimal distress to the patient and the skin

Page 13: Clinical case series - smith-nephew.com

Case 8: Peripheral cannula in an acute setting4

Skin integrity before application of IV3000 Dressing Skin integrity after using IV3000 Dressing

for 25 hours

Source: Pinky Virhia and Jennifer Pennycock, Practice Development Department, NHS Greater Glasgow and Clyde, Scotland, UK

Patient profile

• 76-year-old female with an ovarian mass requiring chemotherapy administered via peripheral cannula; the patient had vulnerable skin integrity

• Insertion site was cleansed with chlorhexidine gluconate in alcohol and allowed to dry; cannulation was achieved at the second attempt and secured with a ported IV3000 Dressing and two sterile adhesive fixing strips

Results with use of IV3000◊ Dressing

• Transparency allowed easy observation of the site without compromising the dressing integrity; visual infusion phlebitis score remained at zero

• IV3000 Dressing acted as a waterproof barrier during washing; the dressing maintained its integrity and protected the cannula

• The patient did not report discomfort upon application or removal of the dressing after 25 hours

– There was no redness or irritation of the skin upon removal of the dressing and the patient’s skin integrity was not compromised

Page 14: Clinical case series - smith-nephew.com

Use of IV3000◊ Dressing was shown to:4

Summary

Facilitate inspection of the insertion site

(transparency)

Not impair skin integrity, irritate the skin

or leave a residue

Maintain its integrity with movement and showering

Keep skin dry, helping to reduce

the risk of infection

Be easy to apply and remove

Remain intact for the duration of use

Be comfortable for patients

Page 15: Clinical case series - smith-nephew.com

1. Cheung E, Baerlocher MO, Asch M, Myers A. Venous access: a practical review for 2009. Can Fam Physician. 2009;55(5):494–496.

2. Mattox EA. Complications of peripheral venous access devices: prevention, detection, and recovery strategies. Crit Care Nurse. 2017;37(2):e1–14.

3. Marsh N, Webster J, Mihala G, Rickard CM. Devices and dressings to secure peripheral venous catheters to prevent complications. Cochrane Database Syst Rev. 2015;Jun 12(6):CD011070.

4. Oliver G. Infection prevention in IV therapy. IV3000 dressing meeting the challenge. Br J Nurs. 2015;24(14 Suppl):S4.

5. Jones A. Dressings for the management of catheter sites: a review. JAVA. 2004;9(1):26–33.

6. Smith+Nephew 2019. Low Allergy Adhesive, Non-Sensitising and Non-Irritant Claims Support. Internal report. PSS 366.

7. Wille JC, Blussé van Oud Albas A, Thewessen EA. A comparison of two transparent film-type dressings in central venous therapy. J Hosp Infect. 1993;23(2):113–121.

8. Maki DG, Stolz SS, Wheeler S, Mermel LA. A prospective, randomised trial of gauze and two polyurethane dressings for site care of pulmonary artery catheters: implications for catheter management. Crit Care Med. 1994;22(11):1729–1737.

9. Keenlyside D. Avoiding an unnecessary outcome. A comparative trial between IV3000 and a conventional film dressing to assess rates of catheter-related sepsis. Prof Nurse. 1993;8(5):288–291.

10. Richardson MC. The research and development of a new transparent film dressing for intravenous catheter care. An in-vivo assessment of the microbial proliferation beneath transparent film dressings. In: Maki DG, ed. Improving Catheter Site Care. Royal Society of Medicine Services International Congress and Symposium, Series No. 179, Harlow, Essex, UK: Royal Society of Medicine Services Limited; 1991:29–33.

11. Smith+Nephew 2001. The Moisture Vapour Transmission Rate (MVTR) of Permeable Film Dressings. Internal report. PS/WR/01/11/003.

12. Smith+Nephew 1990. The Bacterial Barrier Properties of Opsite IV3000. Internal report. SR/TW002/MB89-3.

13. Conyers DA. 1994. IV3000 - Bacterial Barrier Claim. Internal communication.

14. Smith+Nephew 2005. Bacterial Barrier Properties of IV3000 against Methicillin-Resistant Staphylococcus aureas (MRSA). Internal report. DOF 0505005.

15. Smith+Nephew 2013. Bacterial Barrier Testing of IV3000. Internal report. WRP-TW042-281.

References

Page 16: Clinical case series - smith-nephew.com

Advanced Wound Management Smith & Nephew Medical Ltd 101 Hessle Road Hull HU3 2BN, UK

www.smith-nephew.com◊Trademark of Smith+Nephew All Trademarks acknowledged ©August 2021 Smith+Nephew AWM-AWC-30478 | GMC1369

Developed by Evidence Communications, Global Clinical & Medical Affairs

www.smith-nephew.com/education