CLINICAL AREA - TECcare · Test Conditions / Description Contact slide test results from inside...

1
Test No. Test Conditions / Description Contact slide test results from inside macerator [number of microbes recovered in colony forming units (CFU)] Bacteria* added to 2 pulp bedpans (2.5ml of inoculum to each) Doses of TECcare CONTROL added Contact slides taken post cycle Upper drum Lower drum Impeller disc (base of drum) Inside lid 1 - - Y 10 6 10 5 10 6 10 5 2 Y - Y 10 3 10 3 10 10 3 3 Y Y (0.7ml) Y 0 0 0 <10 TEST LOCATION CLINICAL AREA Urology Stroke Unit Ante-Natal CCU General Ward Macerator drum 10 3 10 4 10 4 10 3 10 3 Macerator inside lid 10 3 0 10 5 10 5 10

Transcript of CLINICAL AREA - TECcare · Test Conditions / Description Contact slide test results from inside...

Page 1: CLINICAL AREA - TECcare · Test Conditions / Description Contact slide test results from inside macerator [number of microbes recovered in colony forming units Bacteria* (CFU)] added

Test No.

Test Conditions / Description Contact slide test results from inside macerator [number of microbes

recovered in colony forming units (CFU)]Bacteria*

added to 2 pulp bedpans (2.5ml of inoculum to

each)

Doses of TECcare

CONTROL added

Contact slides taken post

cycle Upper drum

Lower drum

Impeller disc (base of drum)

Inside lid

1 - - Y 106 105 106 105

2 Y - Y 103 103 10 103

3 Y Y(0.7ml) Y 0 0 0 <10

TEST LOCATIONCLINICAL AREA

Urology Stroke Unit Ante-Natal CCU General Ward

Macerator drum 103 104 104 103 103

Macerator inside lid 103 0 105 105 10