Circuits i Lab 1 Post Report

download Circuits i Lab 1 Post Report

of 14

Transcript of Circuits i Lab 1 Post Report

  • 8/11/2019 Circuits i Lab 1 Post Report

    1/14

    EEL-3115L.001F13

    Laboratory Experiment 1: Demonstration of

    Circuit Laws/Theorems

    Gregory Keener

    U-51971841

    9/11/2013

    This lab report was written and assembled solely by myself, Gregory Keener. In-lab measurements were

    taken in conjunction with my lab partner, Erin Ritter. Use of any portion of this lab report by any other

    students (with the exception of measurements for Erin Ritter) constitutes cheating. In addition to being

    reported to appropriate USF faculty, violators will be publicly ridiculed, berated, and/or cursed out in

    front of large groups of people by me. This document is provided on Scribd.com for the sole purposes of

    demonstrating proper lab report formatting and experimental methods. Copyright 2013. All rights

    reserved. If you are an instructor and are receiving this document, please slap your student for me.

    They're cheating in your course.

    ____________________________________________________ ________________

  • 8/11/2019 Circuits i Lab 1 Post Report

    2/14

    Signature Date

    Table of contents

    Introduction and Objectives..2

    Pre-lab Calculations..2

    Materials and Equipment..4

    Experimental Procedure5

    Experimental Data.....8

    Data Analysis..11

    Experimental Concerns, Issues, and Future Improvements12

    Conclusion...13

    Works Cited15

    List of Tables and Figures

    Figure 2.1: Circuit Diagram of Circuit A3

    Figure 2.2: Circuit Diagram of Circuit B3

    Table 2.1: Circuit A Theoretical Values.3

    Table 2.2: Circuit B Theoretical Values.4

    Figure 4.1: Live of Circuit A...6

    Figure 4.2 Live of Circuit B6

  • 8/11/2019 Circuits i Lab 1 Post Report

    3/14

    Table 5.1: Circuit A Experimental Values.8

    Table 5.2: Circuit B Experimental Values.10

    1. Introduction and Objectives

    This purpose of this lab was to learn the proper use of laboratory equipment, including

    digital multimeters, power supplies, breadboards, and basic circuit components. To accomplish

    this, two circuits were to be created as shown in the lab manual (refer to section 3. Experimental

    Procedures for detailed layout). These circuits were to be used to verify that KVL, KCL, and the

    Thevenin Theorem hold in a laboratory environment. In addition to reinforcing fundamental

    circuit laws and theorems, this experiment provided the opportunity to gain hands-on

    experience with professional lab equipment. Also, this experiment demonstrated how to measure

    experimental values for node voltages, branch currents, Thevenin voltages, Norton currents,

    resistor voltages, and Thevenin resistances.

    2. Pre-lab Preparations and Calculations

    In preparation for this experiment, circuits A and B (see fig. 2.1 and 2.2, respectively)

    were analyzed using KVL, KCL, Thevenin Theorem, and basic circuit laws such as Ohms law.

    For circuit A, the node voltages, resistor voltages, and branch currents were calculated for the

    entire circuit. For circuit B, the Thevenin voltage, Norton current, and Thevenin resistance were

    calculated by hand using basic circuit laws and theorems.

  • 8/11/2019 Circuits i Lab 1 Post Report

    4/14

    Figure 2.1: Circuit A diagram Figure 2.2: Circuit B diagram

    The theoretical values for Circuit A and Circuit B were then recorded and are listed in Tables 2.1

    and 2.2, respectively.

    Circuit A Theoretical Values

    Node Voltages

    V1 12V

    V2 4.062V

    V3 Not calculated, mistakenly read node 3 as

    ground

    Branch Currents

    I1 3.777mA

    I2 2.898mA

    I3

    Not calculated, mistakenly read node 3 as

    ground

    Resistor Voltages

  • 8/11/2019 Circuits i Lab 1 Post Report

    5/14

    V1.5k 5.666V

    V3.3k 2.901V

    V910k 3.437V

    V1k -2.901V

    Table 2.1

    Circuit B Theoretical Values

    Thevenin and Norton Equivalents

    Vth 6.974V

    IN 11.9355mA

    Rth 582.08

    Table 2.2

    3. Materials and Equipment

    The following materials and equipment were used to obtain experimental values:

    1x Agilent U8030A Triple Output DC Power Supply

    1x HP3441A Digital Multimeter

    1x 910 , W resistor

    1x 3.3 k, W resistor

    1x 1.5 k, W resistor

    1x 1 k, W resistor

  • 8/11/2019 Circuits i Lab 1 Post Report

    6/14

    1x Heavy duty, banana plug-to-alligator clip wiring set

    1x Heavy duty, probe test lead set.

    1x Elenco Precision Model 9433 breadboard circuit design station

    Note: Jumper wires in graphic diagrams are for illustrative purposes only and were not used for

    the experimental circuit. As such, they are not included in this section

    4. Experimental Procedure

    Two identical circuits were prepared on a breadboard in accordance with the

    specifications listed for circuit A and circuit B in figures 2.1 and 2.2, respectively. Although the

    two circuits extremely similar, and could therefore be interchanged via a small change in wiring,

    separate circuits were prepared in advance of the experiment in order to save time. A live

    illustration of these circuits are shown below (see figures 4.1 and 4.2).

  • 8/11/2019 Circuits i Lab 1 Post Report

    7/14

    Figure 4.1: Live model of circuit A

    Figure 4.2: Live model of circuit B

    In this experiment, circuit A was tested first. For the first part of this experiment, the

    power supply was set to provide a constant 12VDC to the circuit, with a maximum current of

    10mA (the lowest current setting possible on the power supply used). In order to obtain the node

    voltages, the voltage was taken from each node, using node zero as the ground point for the

    probes. The circuit was not broken or modified during this series of measurements, thereby

    causing the probes to be in parallel to the selected node during measurement. Next, the voltages

    were measured across each resistor. Once again, the circuit was measured as-is. Then, the

    current was measured through each resistor. For this series of measurements, one side of the

    resistor was removed from the circuit. The circuit was then reconnected using the test lead

    probes to short the previously created open circuit.

  • 8/11/2019 Circuits i Lab 1 Post Report

    8/14

    After this was completed, the power supply was set to 0V and the resistances were

    measured across each resistor while still in the circuit. Next, the power supply was turned off and

    resistances were taken across each node. Once again, all resistors were left connected to the

    circuit. Finally, each resistor was removed from the circuit and individually measured for

    resistance using the digital multimeter.

    Circuit B was the next system to be tested. For this circuit, the power supply was set to

    provide two separate voltages to the circuit. As such, 8VDC and 12VDC signals were connected

    to the circuit at the points shown in figure 4.2. The node voltages and voltages across each

    resistor were measured using the same methods used for circuit A. Next, the open circuit voltage

    was measured across node 3 and 4. Then, a short circuit was created along nodes 3 and 4 using

    the probe kit. The resulting value was recorded as the Norton current. Finally, the power supply

    connections were removed and the circuit was shorted along these points. The resistance between

    nodes 3 and 4 was then recorded as the Thevenin voltage, thereby concluding the experiment.

    5. Experimental Data

    Circuit A Experimental Values

    Node Voltages

    Measurement By-hand Experimental % Error

    V1 12V 12.001V .00833%

    V2

    4.062V

    6.335V

    35.88%

    V3 Not calculated,

    mistakenly read node 3

    as ground

    3.442V N/A

  • 8/11/2019 Circuits i Lab 1 Post Report

    9/14

    Branch Currents

    Measurement

    By-hand

    Experimental

    %Error

    IR1 3.777mA 3.825mA 1.25%

    IR2 2.898mA .896mA 223.44%

    IR3 Not calculated,

    mistakenly read node 3

    as ground

    2.942mA N/A

    IR4

    3.777mA

    3.839mA

    1.62%

    Resistor Voltages

    Measurement By-hand Experimental % Error

    V1.5k 5.666V 5.664V .035%

    V3.3k 2.901V 2.912V 0.38%

    V910 3.437V 3.422V 0.44%

    V1k -2.901V -2.912V 0.38%

    Node Resistances

    Measurement By-hand Experimental % Error

    R(0-1)

    N/A

    3.122k

    N/A

    R(1-2) N/A 1.474k N/A

    R(3-0) N/A 890.1 N/A

    R(2-3) N/A 988.8 N/A

  • 8/11/2019 Circuits i Lab 1 Post Report

    10/14

    Resistances of Resistors in Connected Circuit

    Measurement By-hand Experimental % Error

    R1.5k N/A 1.472k N/A

    R3.3k N/A 3.247k N/A

    R910k N/A 890.1 N/A

    R1k N/A 988.8 N/A

    Resistances of disconnected resistors

    Measurement By-hand Experimental % Error

    R1.5k N/A

    1.740k

    N/A

    R3.3k N/A 757.77 N/A

    R910k N/A 757.74 N/A

    R1k N/A 890.44 N/A

    Table 5.1

    Circuit B Experimental Values

    Node Voltages

    Measurement

    By-hand

    Experimental

    % Error

    V1 N/A 7.99V N/A

    V2 N/A 8.908V N/A

    V3 N/A 3.090V N/A

    V4 N/A -.3415V N/A

    Resistor Voltages

    Measurement By-hand Experimental % Error

  • 8/11/2019 Circuits i Lab 1 Post Report

    11/14

    VR1 N/A 0.567V N/A

    VR2 N/A 8.907V N/A

    VR3 N/A 3.090V N/A

    VR4 N/A -0.3415V N/A

    Thevenin and Norton Equivalents

    Measurement By-hand Experimental % Error

    Vth -6.974V -3.089V 125.77%

    IN

    11.9355mA

    5.337mA

    123.64%

    Rth 582.08 573.96 1.41%

    Table 5.2

    6. Data Analysis

    For circuit A, the node voltage V2and branch current IR2showed extremely high percent errors

    of 35.88% and 223.44%, respectively. After reviewing the experimental process and pre-lab

    calculations, it was determined that the discrepancy was caused by incorrect analysis in the by-

    hand calculations. Specifically, node 3 was mistakenly used as a reference node, in addition to

    the reference node at node 0. This was due to misreading the circuit and overlooking R4(910

    resistor) while performing circuit analysis. Once the circuit is correctly analyzed without node 3

    being treated as an additional reference, the theoretical values of V2and IR2become 6.24V and

    .835mA, respectively. This causes a percent error of 1.4996% for V2and 6.81%, respectively.

    These values can also be determined using voltage and current division, in which they hold the

    same respective values. All other calculations correlated properly with experimental results.

  • 8/11/2019 Circuits i Lab 1 Post Report

    12/14

    The same error occurred during analysis of circuit B. In this case, node 4 was treated as a

    reference node, in addition to the reference node at node 0. As a result, the Thevenin voltage

    values had a percent error of 125.77%, while the Norton current showed a percent error of

    123.64%. Once again, correct analysis of the circuit (by removing the ground from node 4)

    substantially reduced the percent error for both Vthand IN. Proper analysis resulted in a sub-4%

    percentage error for both values. The experimental Thevenin resistance was within an acceptable

    error margin of calculated values.

    7. Experimental Concerns, Issues, and Future Improvements

    While the experiment was ultimately successful, there were a few issues. First, the

    breadboard circuits were configured incorrectly at first. This caused delays in completing the

    experiment due to requiring time to troubleshoot anomalous multimeter readings (0 voltage).

    This could have been avoided by designing the physical circuits in a manner that more closely

    resembled circuit diagrams 1.1 and 1.2. Instead, the circuits were configured in a manner that

    used as little breadboard space as possible, causing them to appear cluttered and causing

    confusion during assembly. An example of this cleaner layout is given in figures 4.1 and 4.2.

    A substantial amount of time was also used for familiarization with both the digital

    multimeter and power supply. Since neither party performing this experiment had any experience

    with the specific equipment models being used, a large amount of laboratory time was wasted

    attempting to figure out the proper use of said equipment. Specifically, setting maximum

    currents on the power supply and ensuring the proper test lead connections on the digital

    multimeter proved to be time-consuming due to lack of experience. This could have been easily

    avoided by reading the instruction manuals provided in the lab manual prior to arriving in the

    laboratory.

  • 8/11/2019 Circuits i Lab 1 Post Report

    13/14

  • 8/11/2019 Circuits i Lab 1 Post Report

    14/14

    theorem, current/voltage division, and resistance combinations to the experiment itself yields

    values that are within 1-2% of expected values. As such, the experiment was able to successfully

    verify the aforementioned circuit laws and theorems.

    Additionally, this experiment provided invaluable experience with both the digital

    multimeters and power supplies used in this course. This experience vastly increased the

    experimenters proficiency and confidence in dealing with such equipment in a professional

    laboratory environment. In the future, this learning curve should become irrelevant, as most basic

    operations of said lab equipment have now been performed and mastered.

    9. Works Cited

    None

    Remainder of page intentionally left blank.