CIF MPMS Procedures: Magnetic Property Measurement System...recommended for removing the sample rod...
Transcript of CIF MPMS Procedures: Magnetic Property Measurement System...recommended for removing the sample rod...
Research SOP: Magnetic Property Measurement System
Central Instrument Facility, Department of Chemistry
© 2018 Colorado State University Proprietary and Confidential Page 1 of 8
CIF MPMS Procedures:
Magnetic Property Measurement System
Control Authority Central Instrument Facility, Department of Chemistry, Colorado
State University
Title Magnetic Property Measurement System – Research SOP
Last Updated 25 April 2018
Author(s) Erin Stuckert
Contributors Brian Newell
Instrument Manager(s) Brian Newell (Materials Analysis Lab Manager –
[email protected], 1-6209)
Research SOP: Magnetic Property Measurement System
Central Instrument Facility, Department of Chemistry
© 2018 Colorado State University Proprietary and Confidential Page 2 of 8
I. Introduction
Additional resources can be found at http://cif.colostate.edu/materials-analysis/.
II. Safety Information
a. Personal Protective Equipment
i. Eye Protection must always be worn in the Central Instrument Facility(CIF).
ii. Hand Protection, specifically gloves appropriate for low temperature are
recommended for removing the sample rod from the MPMS chamber.
b. Burn Hazards
i. Due to the low temperatures of the MPMS chamber, the sample rod can be quite
cold and burn exposed skin.
c. Training
i. Unauthorized operation of the MPMS instrument is strictly prohibited. All users
must be trained by approved personnel within the protocols of the Central
Instrument Facility prior to operation of the instrument.
ii. This document alone does not serve as training, or a replacement in lieu of
training.
III. Operational Procedures
IMPORTANT NOTES:
• Before using the instrument:
o Check that the instrument’s He levels are within usable levels. If liquid helium level
becomes too low, the superconducting magnet may quench. To prevent this from
happening, the liquid helium should be maintained above the top of the magnet. This
corresponds to a liquid helium level of approximately 50%. If the helium level drops
below the top of the magnet, the field used should be limited.
▪ 40-50% He maximum field should be limited to 10,000 gauss
▪ 30-40% He maximum field should be limited to 1,000 gauss
▪ Below 30% He the magnet should not be charged with a magnetic field
o If the He levels are not within usable range please wait until the CIF has filled the He
to usable levels again or contact Brian Newell to request a fill as soon as possible.
o Check that the pump outside of the MPMS housing is functioning by placing your
hand near the fan on the pump. If you can’t feel air movement from the fan and the
pump appears to not be functioning please contact Brian Newell.
• Be sure to copy all of your data to another location for your records as the CIF
cannot guarantee all data will be backed up.
• Fill out the required information on the log sheet by the MPMS computer.
o Be sure to record the initial and final He levels and include any comments about your
type of analysis or problems that arise with the system.
a. Instrument Start-Up
i. Take the instrument out of Standby mode by clicking OK. (When not in
use the MPMS should always be in standby mode to minimize He waste
from the 44 L He dewer in the instrument.)
Research SOP: Magnetic Property Measurement System
Central Instrument Facility, Department of Chemistry
© 2018 Colorado State University Proprietary and Confidential Page 3 of 8
b. Energize the field and set temperature
i. Using the bar on the bottom of the window,
1. double click on Target field and in the Magnetic Field window,
change set field to 1000 Oersted (Oe)
2. double click on System and in the Temperature Parameters
window, under control change set point to 100 K. (If sample is
paramagnetic will likely go above 100 K and if diamagnetic will
likely go below 100 K. Your sample needs to have some
paramagnetic character for centering the sample.)
Research SOP: Magnetic Property Measurement System
Central Instrument Facility, Department of Chemistry
© 2018 Colorado State University Proprietary and Confidential Page 4 of 8
c. Preparing your sample for the instrument:
i. Sample mass should be 50-100 mg. (Need greater mass if looking at high
MW samples to get more free electrons/molecule.)
ii. Use a polycarbonate (usually) straw with PEG capsules, gelatin or PE bags
to contain sample. (*all samples should be in containment for ease of
retrieval should the sample fall into the chamber*)
iii. Bend straw in half,
1. use push pin to create 4 holes evenly spaced just below the halfway
point, insert sample into straw, then create 4 more holes evenly
spaced just above the sample. These holes help keep the sample
contained in the straw.
2. alternatively, you can use a scissors to cut an ~1mm section of the
straw and push it into the center of the straw creating a ledge for
your sample, insert your sample, then use the push pin to create 4
holes evenly spaced just above the sample.
iv. Get the sample rod, being careful of the Au/Au coated junction as it is
delicate and can be easily
damaged.
1. DC rod typically
used, use carbon fiber rod for less thermal expansion.
v. Insert the straw onto the end of the sample rod into the ferrule. Gently
twist the straw to secure it to the rod. Lower the blue cap to meet the
ferrule
d. To insert the sample rod into the instrument: (See Figure d below)
i. On the instrument, barely push down CW on the black lever to activate the
sensor to vent, and push the lever back up (orange venting light should
flash).
ii. Be sure one of the white dots on the blue cap on the rod is facing you and
the other is directed to the left. Watch that the rod does not hit the ceiling
and the straw does not pop off of the rod while slowly inserting the rod
straight down into the instrument until the straw is fully observed in the
viewing window and the blue cap sits in place of the plug. Close the black
side tabs to ‘lock’ the blue cap in place, and press black button to purge
airlock once, then once again after Ready lights up green.
iii. When instrument Ready for second time, push the black lever clockwise to
be vertical and in the open position. Slowly push the rod carefully near the
end going into the system, allowing the rod to go in as straight as possible.
(be careful of the joint between the two different metals, if the rod
Research SOP: Magnetic Property Measurement System
Central Instrument Facility, Department of Chemistry
© 2018 Colorado State University Proprietary and Confidential Page 5 of 8
catches on this joint or near the rod/straw joint, carefully twist the
rod as you push) Be sure the black bar on top of the rod is flush with the
instrument (twist the rod so the black plate fits under the loosened black
beads. Tighten the black beads to hold the rod in place.
e. Centering sample:
*The field should be 1000 Oe or 0.1 T before centering. Depending on your
experiment needs you may want to center your sample under conditions similar to
your experiment settings.
i. Click the center tab, then DC.
1. A window should pop up. Select initialize transport, then full DC
scan. (The full range the instrument can scan
through is 12 cm before taking any
measurements.) Click view to see a graph
showing you the centering progress. The
sample should be centered at 2 ± 0.02 cm
6b.
6c.
6a.
Research SOP: Magnetic Property Measurement System
Central Instrument Facility, Department of Chemistry
© 2018 Colorado State University Proprietary and Confidential Page 6 of 8
2. Always start with Full DC Scan (i.e. auto center) first. If the
sample won’t center or if sample loses magnetism, you will need to
Adjust Position (i.e. manual center).
a. Try to Auto Adjust first using 0.5 to 1 cm increments to
approach 2 ± 0.02 cm. *Larger increments are likely to
cause the adjustment to fail.
b. If the adjustment needed is several cm or Auto Adjust
continues to fail, use Manual Adjust and follow the
instructions in the pop up windows.
c. Loosen the metal ferrule on top of rod, click Ok, then
slightly (barely a few mm) pull up or push down on the rod,
finger tighten the ferrule, then click ok again. The
instrument will redo a 4 cm short center scan to check the
new center. *You may need to repeat the Manual Adjust
process or a combination of Manual Adjust and Auto
Adjust.
ii. Sample head options and their centering
3. DC transport sample head option
a. Scans 32 pts across the entire position range
then goes back through the scan a second time
from the beginning.
4. Reciprocating (RSO) sample head option
a. Scans 1000 points in a smaller position range
then extrapolates the rest of the curve.
b. This is a faster scan for most typical
experiments with less sample.
c. If T < ~130-150 K, can’t use the RSO because
the sample chamber is exposed.
f. Creating and running a sequence for data collection
i. Click Sequence tab, new to create
new sequence for data collection.
1. Add Sequence Commands
from the right sidebar by
double clicking on a
command. (These will be
dependent upon your
individual needs.)
scan
Position (cm)
2
Res
po
nse
(en
u)
Position (cm)
2
Res
po
nse
(en
u)
scan
Research SOP: Magnetic Property Measurement System
Central Instrument Facility, Department of Chemistry
© 2018 Colorado State University Proprietary and Confidential Page 7 of 8
*All sequences should include the following*
a. Under Sequence Commands>Measure, add set datafile to
create a folder where data are saved. You may add remarks
to describe the sequence in the sequence window.
i. This folder should be created under C drive in the
QdMPMS >Data folder.
b. Next, add Start Data Logging to your sequence before the
experiment starts (i.e. before set magnetic field) (this allows
us to monitor data collection and parameters to help
diagnose any potential issues that may arise).
i. In the Log Data window that opens, under Standard
Items tab, Select All.
ii. Under the General tab, choose Browse and give
your log file a name specifying log and save it in
your Data folder.
c. Add in the steps you desire to change/measure
temperature/field etc. (see below for more information)
d. Add steps Set temp 4.2 K, temp stable, and set magnetic
field 0 to your sequence to put the instrument into Standby
mode upon completion of your data collection. This will
conserve He until you can put the system in Standy mode in
person or using Team Viewer (click Instrument>Standby).
e. End your sequence with Stop Data Logging.
Things you may include in a sequence
f. You can add in pre-experiment steps to your sequence,
such as exposing the sample to a field or starting at a
specific temperature.
g. You can set sampling rate (specifically with Pd standard).
i. Longitudinal or translational squid
Research SOP: Magnetic Property Measurement System
Central Instrument Facility, Department of Chemistry
© 2018 Colorado State University Proprietary and Confidential Page 8 of 8
h. Set magnetic field to value desired (e.g. 50,000 Oe or 5 T)
i. Select oscillate from the dropdown menu (this is
quickest to set).
ii. If you don’t want to overshoot a particular signal
select no overshoot (this is a slower process)
iii. If select hysteresis, ramps the field positive then
negative alternatively so there is no remnant
magnetic field
i. Always want to have in high resolution mode.
j. Select a DC measurement will have a 4 cm scan length of
24 data pts, with 2 scans per measurement. You shouldn’t
have to change any other settings.
ii. To save a sequence select file save as and be sure to save the file under C
drive in the QdMPMS >Sequence folder.
iii. Click the play/start (top toolbar) or Run (left sidebar) button to start the
sequence data collection when ready.
iv. At any time during a measurement you can observe the sequence
status/step being executed by looking at the Sequence Status bar on the left
sidebar or by clicking Sequence tab>Status.
g. Team Viewer
i. You can look into your data collection and operate the instrument
remotely using Team Viewer. This allows you to check your data
collection during long data collection periods. (To use team viewer you
will need to record the ID and Password provided on the desktop. Keep in
mind if the computer is restarted, the password will be reset.)
h. Removing the sample rod and starting Instrument Standby
i. Remove the sample rod from the instrument by first loosening the black
beads, twist the black bar to disengage from the beads, and pull the rod up
until the straw is fully in the viewing window.
ii. Move the black knob back CCW to close.
iii. Flip the black tabs to ‘unlock’ the blue cap, pull the sample rod completely
out of the instrument and replace the blue plug, flip the black tabs to ‘lock’
the plug.
iv. Push the black Purge button once.
v. If the instrument is not already in Standby, click Instrument>Standby.
vi. Record your final He
consumption,
Instrument time and
any remaining pieces
of the log sheet before
leaving.