Transcript of Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis...
![Page 1: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/1.jpg)
Maximizing Efficiency in Analysis through New GC-MS Approaches
Richard Fussell
Vertical Marketing Manager, Food and Beverage, Thermo Fisher Scientific, Hemel Hempstead, UK Dominic Roberts Senior Applications Scientist, GC-MS, Thermo Fisher Scientific, Runcorn, UK PO71686-EN 0615S
![Page 2: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/2.jpg)
2
Overview • The analytical challenge
• User requirements for GC-MS/MS analysis of pesticides • Critical aspects of the method & improving efficiency
• Injector • Column configuration • Instrumental parameters
• Latest GC-MS/MS developments including GC-Orbitrap for pesticide screening
• Summary
![Page 3: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/3.jpg)
3
Typical Pesticides Workflow Register for future webinars and to view recordings of past webinars at www.chromatographyonline.com/LCGCwebseminars
1. Sample Prep: March 24th 2. LC-MS Analysis: April 29th
3. GC-MS Analysis: June 17th 4. Data Processing/Analysis: July 15th
![Page 4: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/4.jpg)
4
• Wide range of matrices • Food • Environment
• Wide analytical scope • Low limits of detection • High sample throughput • Fast turnaround • Low cost of analysis
Analytical Challenges for GC Pesticides Analysis
RT: 4.57 - 37.12
6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36Time (min)
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
Rela
tive
Abun
danc
e
6.32
6.59
4.63
7.00 11.13
10.12
11.9925.20
14.969.50 23.13 29.6117.388.89
12.207.315.77 19.1315.57 15.848.20
26.4712.87 17.9513.77 23.58
28.7125.3620.625.09 21.7023.8319.25
27.7724.12
35.78
34.8134.3932.9230.11 31.00
NL:3.94E8TIC MS 2july2104_011
![Page 5: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/5.jpg)
5
Pesticide Analysis: GC-MS
• Many compounds not amenable to LC separation • Low polarity–poor atmospheric pressure ionization
• GC offers good separation efficiency • Choice of detectors • Easy coupling with MS for increased
selectivity • EI/CI spectra for identification of analytes
![Page 6: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/6.jpg)
6
Pesticide Analysis: Triple Quad GC-MS
Highly Selective Reaction Monitoring (SRM) Improved detection limits
Longer column lifetime Less frequent inlet maintenance
![Page 7: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/7.jpg)
7
Selectivity: Selected Ion Monitoring (SIM) and SRM
DDE-p,p’, 0.05 mg/kg in green tea, 1.0 uL splitless injection
![Page 8: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/8.jpg)
8
Selectivity: SIM and SRM
DDE-p,p’, 0.001 mg/kg in green tea, 1.0 uL splitless injection
![Page 9: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/9.jpg)
9 The world leader in serving science
Proprietary & Confidential
Key Factors in the GC-MS/MS Method
![Page 10: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/10.jpg)
10
Hot split/splitless
Programmed Temperature
PTV
Typical GC Injector Choices for pesticides
• Liquid introduction by syringe
“GC Injection is the Achilles Heel in GC”
Bertsch 1983, Univ. Alabama
• Most commonly used technique • Split/Splitless injection (SSL) • Programmed temperature (PTV)
![Page 11: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/11.jpg)
11
GC Inlets • Splitless
• maximum sensitivity • excellent repeatability for low volumes • simple, probably most wide used
• Split
• reproducible • less discrimination (short residence time) • Shoot and Dilute
• Programmed Temperature Vaporising (PTV) injector • versatile and excellent performance if optimised • reduced discrimination • many liner types (baffled, dimpled, packed, etc) • packed liners (possible discrimination) • large volume injection (solvent removal/exchange in liner)
• Cool on-column (not widely used)
![Page 12: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/12.jpg)
12
GC Liner Selection for Pesticides
• In pesticide analyisis QuEChERS extractions are typical and result in extractions in acetonitrile.
• Many labs use acetonitrile as GC injection solvent • Requires careful method optimisation
• Considerations in liner selection for acetonitrile injections are: • Internal diameter • Type of injection • Packing of liner • Other liner features ie baffles....
![Page 13: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/13.jpg)
13
GC liners – Type of Injection
• Split • Typically open ended at the bottom • Enables split flow to pass across the bottom of the liner removing a portion of the
sample, allowing a split injection to be performed
• Splitless
• Typically tapered at the bottom with the column inserted into the taper • Funnels sample onto the column and minimizes sample contact with reactive metal
components
• PTV • Generally used with very active compounds such as pesticides • Good option for acetonitrile injection solvents • Thermally liable compounds protected
![Page 14: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/14.jpg)
14
PTV Injector: Key Points
• Minimal thermal mass for fast cooling and heating
• Injection volumes from nano liter up to largevolume
• Cold injection technique
• Clean step possibility for keeping the liner inert
• Multiple injection modes
OVEN column
Liner
Cooling by fan Heater element
Inlet Carrier Septum Purge
Split line
Slide courtesy of Thermo Fisher Scientific
![Page 15: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/15.jpg)
15
System Contamination with Heavy Matrices
9.03 min (start of run)
1.0 µg/ml dimethoate in crude extracts of lettuce - 3 µl splitless
9.06 (Injection ~20)
GC Liner
Slide courtesy of Fera, UK
![Page 16: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/16.jpg)
16
Backflush Injection
(Thermo GCQ Quantum)
Pre-column (2 m x 0.53 mm i.d., deactivated)
![Page 17: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/17.jpg)
17
PTV Backflush of Pear Extracted with AcEt
• No BKFL
• BKFL ON 10 min after injection of sample
• BKFL ON 10 min after injection of standard
xc21_estratto_plus40ppbmixpestethaccy... 16/04/2009 19.02.57
RT: 5.07 - 24.68
6 8 10 12 14 16 18 20 22 24Time (min)
0
20
40
60
80
100
Rel
ativ
e A
bund
ance
0
20
40
60
80
100
Rel
ativ
e A
bund
ance
0
20
40
60
80
100
Rel
ativ
e A
bund
ance
Extract
BKF at 20 min (No BKF) Vitamine E
Sitosterol (area)
Octadecanoic Acid 14.68
14.1921.1513.39
19.9017.707.6421.47
18.3011.769.738.15 16.52
12.64 22.037.42 16.868.72 22.1715.38 19.455.81 11.23 22.516.74 23.47
Extract
BKF at 10 min
Vitamine E
Sitosterol (area)
Octadecanoic Acid14.20
13.7813.04
7.65 19.08
11.619.739.38 17.00
12.387.42 15.938.72
5.83 6.88 15.2711.15 17.5919.24 19.98 23.0520.84 23.69
Standard Mix
BKF at 10 min
Pesticides14.88
15.9712.95
13.72 16.7812.32
14.5117.07
17.827.29 7.80 10.41 11.539.08 18.4411.29
18.657.18 8.39 21.30 23.6520.799.266.45 21.44
NL: 5.69E9TIC MS xc21_estratto_plus40ppbmixpestethaccyc_120ul_ptvbkf_02mindelay_85cto260ptv_r3
NL: 5.12E9TIC MS XC21_EstrattoPeraArmandi_120ulEtAcCycl_PTVBKF_Clean10min
NL: 2.65E9TIC MS xc21_400ppbmixpestarmandi_120uletaccycl_ptvbkf_clean10min
Full scan data acquisition – Trace GC w PTV-BKF – 30 m TR-Pesticides, 5 m pre-column 0.53 mm ID
Area of high boiling matrix
![Page 18: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/18.jpg)
18
Effect of Back-Flush on Carryover RT: 0.00 - 35.10 SM: 7G
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34Time (min)
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
Relativ
e Abun
dance
29.45 30.2133.99
32.4129.14
28.70
28.4527.99
27.2526.98
26.76
26.26
26.04
25.30
24.8424.49
14.2510.44 23.7122.359.108.80 10.79 21.62
7.90 19.7112.64 18.9217.60
12.17
20.71
12.64
28.23
29.22 33.7932.99
27.0310.1713.3810.07 11.52
14.219.32
9.058.64
7.90 14.92 24.96
16.62 17.46 24.3418.11 21.46
NL:2.50E5TIC MS 51845023NL:2.50E5TIC MS 51845038
Fresh Podded Peas
RT: 0.00 - 35.10 SM: 7G
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34Time (min)
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
Relative
Abunda
nce
28.8829.5928.66
27.5830.12
27.3430.98
32.1027.0233.34
26.4934.95
26.25
25.73
24.9824.8014.2524.53
24.099.02 23.839.35
7.39 23.1710.3211.84 20.5619.40
12.068.64 19.0117.23
16.11
12.62 23.1122.92
30.79
31.7014.06
32.3427.50 32.79
27.1821.99
12.0832.97
9.63
9.348.75 11.86 25.75
20.30 25.4514.59 20.187.86
15.05 16.89 18.60
NL:2.00E5TIC MS 51865013NL:2.00E5TIC MS 51865028
Molasses
Chromatograms for EtAc solvent after injection (n=10) of QuEChERS extracts with back-flush and without back-flush
Slide courtesy of Fera, UK
![Page 19: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/19.jpg)
19
Guard columns
• Analytical columns with a length of 5-10 m of deactivated fused silica. They can be purchased already integrated or joined by a union.
• Provides the benefit of protecting the analytical column from
contamination of non-volatile residues. Very important when working with dirty sample extracts eg. QuEChERS.
• Can also act as a retention gap to improve analyte focussing. • Maintain retention time of analytes and SRM segments in the method. • Can be a source of leaks if using a connection. • Added maintenance
![Page 20: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/20.jpg)
Thermo Scientific TSQ 8000 Evo GC-MS/MS
![Page 21: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/21.jpg)
21
Increasing Laboratory Productivity
• Decrease analysis time by shortening the GC run times. • More samples in less time.
• Increase the number of pesticides in a run. • More SRMs to accommodate within an analytical run.
• Improve selectivity for various matrices. • Increased number of SRMs per compound.
• See beyond the targets. • Full Scan and SRM data acquisition in the same experiment.
Expect More Performance
![Page 22: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/22.jpg)
22
Fast GC-MS Pesticide Residue Analysis
Challenges:
• Complexity of elution when using fast GC
• Large number of compounds (SRMs) in short time
• Many SRM transitions can result in sensitivity loss
Solution:
• High speed analyzer • Fast collision cell • Short SRM dwell times with very
short inter-scan delays
![Page 23: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/23.jpg)
23
TSQ 8000 Evo GC-MS/MS
Expect More Capacity
• Analytical instrumentation:
Thermo Scientific™ TSQ ™ 8000 Evo GC-MS/MS Thermo Scientific™ TRACE™ 1310 GC Thermo Scientific™ TriPlus RSH™ autosampler (liquid injection set-up)
• EvoCell
• Rapid, innovative collision cell technology • Increased method capacity • More compounds • More SRM transitions • Up to 4x more transitions whilst maintaining method sensitivity low analyte concentrations
![Page 24: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/24.jpg)
24
Increasing Laboratory Productivity
• Decrease the analysis time by shortening the GC run times. • More samples in less time. • More SRM Increase the number of pesticides in a run. • s to accommodate within an analytical run.
• Improved selectivity for various matrices • Increase the number of SRMs per compound.
• Seeing beyond the targets • Full Scan and SRM data acquisition in the same experiment.
![Page 25: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/25.jpg)
25
Decrease the Analysis Time
RT: 4.57 - 37.12
6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36Time (min)
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
Re
lativ
e A
bu
nd
an
ce
6.32
6.59
4.63
7.00 11.13
10.12
11.9925.20
14.969.50 23.13 29.6117.388.89
12.207.315.77 19.1315.57 15.848.20
26.4712.87 17.9513.77 23.58
28.7125.3620.625.09 21.7023.8319.25
27.7724.12
35.78
34.8134.3932.9230.11 31.00
NL:3.94E8TIC MS 2july2104_011
Full scan 144 pesticides in baby food @ 0.2 mg/kg TG-5 SILMS, 30m x 0.25 mm x 0.25 µm GC run time: ~37 min
![Page 26: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/26.jpg)
26
Decrease the Analysis Time RT: 4.04 - 10.89
4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5 10.0 10.5Time (min)
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
Re
lativ
e A
bu
nd
an
ce
8.73
8.26
8.95
7.687.65 8.45
8.00
8.788.23
8.507.80
8.677.40
6.03 7.467.036.63 7.25
7.026.619.306.826.39 7.20
9.026.17 8.02 10.349.216.35 8.036.57
5.884.65 9.609.944.89 9.44
5.424.72 9.61
9.985.245.21 10.5710.04
4.09 9.73 10.175.85 10.675.504.20 5.684.27 5.11
NL:1.39E9TIC MS 2july2104_048
Full Scan 144 pesticides in baby food @ 0.2 mg/kg TG-5 SILMS, 20m x 0.18 mm x 0.18 µm GC run time: <11 min
![Page 27: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/27.jpg)
27
Pesticide MRM Database
The problem:
• Growing list of target compounds require continuous adjustment to an existing SRM database.
• Some SRM transitions are not suitable for all matrices. Addition of new SRM transitions can be time consuming.
The solution:
• Automated SRM development with AutoSRM.
![Page 28: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/28.jpg)
28
AutoSRM: Fast, Simple Route to Optimized SRM
1) Precursor ion selection
2) Product ion selection
3) Collision energy optimization
AutoSRM automates the development of SRM methodology
![Page 29: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/29.jpg)
29
Highlights of AutoSRM
• Automates the following: • Creation of full scan, product ion scan, and SRM methods • Creation of sample sequences • Creation of data layouts for analyzing results • Selection of precursor, product, and collision energies
End result showing optimized transition
![Page 30: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/30.jpg)
30
Timed-SRM: Using Dwell Times Efficiently
Classical segmented SRM
TSQ 8000 EVO timed SRM
Classical segmented SRM:
• Complex to set up • Wasted dwell time • Reduced sensitivity • Reduced tolerance to RT shifts
TSQ 8000 Evo timed-SRM:
• Automated set-up • Full optimized dwell time • Optimal sensitivity • Increased resistance to RT shifts
![Page 31: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/31.jpg)
31
Increase Laboratory Productivity
• Decrease the analysis time by shortening the GC run times. • More samples in less time.
• Increase the number of pesticides in a run. • More SRMs to accommodate within an analytical run.
• Improved selectivity for various matrices • Increase the number of SRMs per compound.
• Seeing beyond the targets • Full Scan and SRM data acquisition in the same experiment.
![Page 32: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/32.jpg)
32
More SRM Transitions/Compound for More Confidence
Tecnazene in baby food at 0.01 mg/kg level
![Page 33: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/33.jpg)
33
RT: 14.95 - 16.71
15.0 15.2 15.4 15.6 15.8 16.0 16.2 16.4 16.6Time (min)
0
10
20
30
40
50
60
70
80
90
100
0
10
20
30
40
50
60
70
80
90
100
Rel
ativ
e A
bund
ance
RT: 15.79AA: 4093580SN: INF
RT: 15.79AA: 1901556SN: INF
NL: 2.60E6TIC F: + c EI SRM ms2 [email protected] [176.095-176.105] MS Genesis 19May2014_03
NL: 1.21E6TIC F: + c EI SRM ms2 [email protected] [245.995-246.005] MS Genesis 19May2014_03
RT: 14.84 - 16.62
15.0 15.2 15.4 15.6 15.8 16.0 16.2 16.4 16.6Time (min)
0
10
20
30
40
50
60
70
80
90
100
0
10
20
30
40
50
60
70
80
90
100
Rel
ativ
e A
bund
ance
RT: 15.79AA: 3837740
RT: 15.79AA: 1690756
NL: 2.35E6TIC F: + c EI SRM ms2 [email protected] [176.095-176.105] MS Genesis 19May2014_05
NL: 1.03E6TIC F: + c EI SRM ms2 [email protected] [245.995-246.005] MS Genesis 19May2014_05
More Speed Maintaining the Sensitivity
DDE-p,p’ in green tea, 1917 SRMs Inj. 0.01 mg/kg on column
DDE-p,p’ in green tea, 44 SRMs Inj. 0.01 mg/kg on column
0.7 ms 27 ms
![Page 34: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/34.jpg)
34
Linearity: Dichlorvos
• Dichlorvos peak area response over 0.5–10 ppb (mg/kg), matrix-matched standard (baby food).
• Chromatograms (quantification and confirmation ions) at 10 ppb level.
8 SRMs/compound
2 SRMs/compound R2= 0.998
R2= 0.997
![Page 35: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/35.jpg)
35
Examples of Peak Area Repeatability: Fast GC Method
Data analysis method maintenance
GC Performance Maintenance
MS Performance Maintenance
Acquisition Method Maintenance
Routine results
%RSD pesticides (0.001 mg/kg on column) baby food matrix ~144 compounds in <10 minutes
Compound % RSD (n = 10) BHC, Alpha 7.0 BHC, Beta 8.8 BHC, gamma 9.2 Chlorobenzilate 12.5 Chlorothalonil 12.6 Clomazone 6.3 Cyfluthrin peaks 1-4 9.3 DDE p, p 8.2 Dichlobenil 5.3 Dichlorvos 8.1 EPTC 5.3 Etridiazole (Terrazole) 4.6 Hexachlorobenzene 8.9 Methacrifos 7.6 Propachlor 11.0 Propham 11.5 Simazine 9.1 Tecnazene 5.6 Tefluthrin 7.0 Triallate 11.0
![Page 36: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/36.jpg)
36
Thermo Scientific TSQ 8000 EVO Pesticide Analyzer
A complete pesticide method implementation, management and maintenance solution to drive unstoppable result productivity
TSQ 8000 EVO PA designed to create powerful pesticide methods that are:
1. Self-customized 2. Auto-optimized
Pesticide Analyzer
![Page 37: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/37.jpg)
37
Powering the TSQ 8000 Pesticide Analyzer
• Preconfigured performance leading TSQ 8000 EVO GC-MS/MS system featuring the award winning TRACE1310 GC
• Pre-loaded acquisition methods • Thermo Scientific TraceGOLD GC Column
and consumable technology • Tracefinder 3.2 EFS Data Processing
software • 600+ Pesticide compound database (CDB)
with 1500 + SRM transitions • AutoSRM & timed SRM (t-SRM) • Pesticide Analyzer installation guide
![Page 38: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/38.jpg)
GC High Resolution Mass Spectrometry for Pesticide Analysis
![Page 39: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/39.jpg)
39
Q Exactive GC for Pesticide Analysis
• Launched at ASMS June 2015. • Screen (qualitative and quantitative) samples for pesticides within a single
analysis, fast and at a competitive cost.
• To increase the scope of the analysis, by using high resolution full scan mass spectrometry.
• Untargeted analysis where a generic full scan acquisition is run, followed by targeted data processing of a list of compounds.
• Retrospective data analysis is possible to identify new compounds that were not screened for at the time of acquisition.
![Page 40: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/40.jpg)
40
The Power of Accurate Mass
![Page 41: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/41.jpg)
41
An Example Study
• To evaluate the performance of Thermo Scientific Q Exactive GC hybrid quadrupole-Orbitrap mass spectrometer for the reliable screening of GC amenable pesticides.
• To screen for a wide range of pesticides in different sample matrices with the highest level of confidence possible.
• To determine if a pesticide is present in a sample above the MRL which is typically 10 ng/g (ppb).
![Page 42: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/42.jpg)
42
Experimental • Sample introduction was performed using a Thermo Scientific™ TriPlus™ RSH
autosampler, and chromatographic separation was obtained with a Thermo Scientific™ TRACE™ 1310 GC. Thermo Scientific™ TraceGOLD TG-5SilMS 15 m x 0.25 mm I.D. x 0.25 µm film capillary column.
• Q Exactive GC hybrid quadrupole-orbitrap mass spectrometer was used. The system was
operated in EI using full scan and 15k, 30k, 60K and 120k resolution (FWHM, m/z 200). Data was acquired with a minimum of 10 points/peak.
• Data was acquired and processed using the TraceFinder version 3.3 software.
![Page 43: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/43.jpg)
43
RESULTS
![Page 44: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/44.jpg)
44
Screening Criteria Used for Positive Identification
Iden%fica%on Point Tolerance Primary ID Confirmatory ID
Reten%on %me 20 seconds
Accurate Mass 2 ppm
Fragment ions 2 ppm
Isotopic pa8ern >70%
NIST Library match >600
Ion ra%o 30%
![Page 45: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/45.jpg)
45
TraceFinder Screening Browser Positively Identified Pesticides
![Page 46: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/46.jpg)
46
Sensitivity: Wheat
![Page 47: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/47.jpg)
47
Sensitivity: Horse Feed
![Page 48: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/48.jpg)
48
D:\Dom Data\...\21jan_038 01/22/15 03:37:38
21jan_026 #3957 RT: 5.78 AV: 1 NL: 6.28E6T: FTMS + p EI Full ms [50.00-500.00]
127.008 127.010 127.012 127.014 127.016 127.018 127.020 127.022 127.024 127.026 127.028 127.030 127.032 127.034m/z
0
20
40
60
80
100
Rel
ativ
e A
bund
ance
127.02067
21jan_030 #2240 RT: 5.78 AV: 1 NL: 2.39E6T: FTMS + p EI Full ms [50.00-500.00]
127.008 127.010 127.012 127.014 127.016 127.018 127.020 127.022 127.024 127.026 127.028 127.030 127.032 127.034m/z
0
20
40
60
80
100
Rel
ativ
e A
bund
ance
127.02164127.01821
21jan_034 #1192 RT: 5.79 AV: 1 NL: 3.61E6T: FTMS + p EI Full lock ms [50.00-500.00]
127.008 127.010 127.012 127.014 127.016 127.018 127.020 127.022 127.024 127.026 127.028 127.030 127.032 127.034m/z
0
20
40
60
80
100
Rel
ativ
e A
bund
ance
127.02117127.01826
21jan_038 #623 RT: 5.79 AV: 1 NL: 5.49E6T: FTMS + p EI Full lock ms [50.00-500.00]
127.008 127.010 127.012 127.014 127.016 127.018 127.020 127.022 127.024 127.026 127.028 127.030 127.032 127.034m/z
0
20
40
60
80
100
Rel
ativ
e A
bund
ance
127.01833
127.02118
127.02261
30K
60K
120K Chlorpropham Matrix
Mass difference = 18.4 ppm
Mass difference = 0.9 ppm
Mass difference = 0.5 ppm
15K
Mass difference = 0 ppm
Effect of Resolving Power on Mass Accuracy Chlorpropham in Leek (10 ng/g)
![Page 49: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/49.jpg)
49
Scan Speed and Accurate Mass Across Peaks 60K
• XIC of diazinon(m/z 179.11789 ±5 ppm mass window) in wheat at 10 ng/g showing ~11 scans/peak (peak width 1.8 sec).
Average = 0.33 ppm RMS
![Page 50: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/50.jpg)
50
Maintaining Sensitivity with Resolution
![Page 51: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/51.jpg)
51
Linearity
• XIC (quan and confirm ions) and calibration curve for Fenpropimorph in leek.
• Triplicate injections of the calibration series was performed with good linearity across (0.5 – 50 ng/g).
• No internal standard correction.
R2 = 0.9999
![Page 52: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/52.jpg)
52
Conclusions
• Careful method optimisation focussing on the injection parameters.
• Routine pesticides analysis with the EVO offers sensitivity, high analysis speed and easy database management at low cost
• Using the available dwell time wisely: • Timed-SRM ensures minimal loss of time spent to acquire data.
• Q Exactive GC system improves efficiency by increasing the scope of the analysis:
• Full scan non-targeted acquisition. • Provides the required sensitivity and selectivity in complex matrices for routine
pesticide screening and quantification. • Enables the detection and identification of unknown compounds.
Efficient and robust pesticide analysis can be achieved by:
![Page 53: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/53.jpg)
53
Thermo Scientific Food and Environmental Communities: Resources • View application notes, on-demand webinars, product information, and
many more resources on our Pesticides and Food Communities Libraries: www.thermoscientific.com/pesticides www.thermoscientific.com/foodandbeverage
![Page 54: Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maximizing Analysis Efficiency through GC-MS Approaches](https://reader030.fdocuments.in/reader030/viewer/2022032618/55b84ad4bb61eb64028b47fb/html5/thumbnails/54.jpg)
54
Thank You for Listening
Questions?