Chemical lab report:Analysis of anions(in toothpaste Berocca and an unknown solution) by...
-
Upload
awad-albalwi -
Category
Science
-
view
491 -
download
2
Transcript of Chemical lab report:Analysis of anions(in toothpaste Berocca and an unknown solution) by...
Awad Albalwi
Analysis of Anions by Chromatography (IC) and ion selective electrodes (ISE)
Prac Report
Aim-
The aims of the experiment are to calculate the amount of fluoride and chloride in a
range of samples including toothpaste Berocca and an unknown solution using Ion
Chromatography (IC) and ion selective electrodes (ISE). The experiment also aims to
estimate to limits of detection of each method.
Method-
Method is as per method described in lab manual.
Results-
A tables summarising the results obtained is shown below. The calculations are provided
with the attached sheets. The results will be discussed further in the results section.
Table 1:
Sample ISE ISE IC IC
[Cl-]
(ppm) [F
-] (ppm) [Cl
-]
(ppm) [F
-] (ppm)
Unknown 45±0.6 28.2±0.05 28±1 26±3
Toothpaste A 5±0.5 21.9±0.05 9±1 21±3
Berocca n/a n/a 151±1 n/a
Tap Water 14±0.5 1.2±0.05 20+/-3 0.4±3
Table.1 : Analysis of standard solution & sample for F&CL by using Ionic Chromatography.
Sample No Sample conce
(ppm)
F Ret.Time
(Min)
F Peak area (µS/min)
CL (peak area)
(µS/min
CL Ret.Time
(min) CL)
1 0 0.0491 2.62 0.097 3.75
2 0.1 0.3327 2.62 0.343 3.76
Awad Albalwi
3 10 3.906 2.62 3.070 3.75
4 25 9.720 2.63 8.117 3.74
5 30 11.110 2.63 9.574 3.73
6 40 14.521 2.64 13.038 3.72
7 50 17.563 2.64 16.308 3.71
8 Unknown
9.408 2.62 9.212 3.71
9 Tap Water 0.373 2.61 6.512 3.71
10 Toothpaste 7.647 2.61 3.004 3.73
11 Borroca
0.020 2.62 0.492 3.74
Table.1: Analysis of standard solution & sample for F&CL by using ISE.
Sample No Sample (ppm)
Log[Conce]
F (mV)
CL (Rmv)
1 0.1 -1 137.8 263.2
2 1 0 80.2 242.5
3 10 1 19.8 199.5
4 25 1.397940009 -4.3 171.3
5 30 1.477121255 -8.6 171.0
6 40 1.602059991 -17.4 161.9
7 50 1.698970004 -24.1 158.4
8 unknown ------ -7.8 164.0
9 Tap Water ----- 74.5 184.1
10 Toothpaste ------ -1.4 204.0
11 Borroca ------- a/n 105.8
Graph1: The calibration curve for fluoride (Concentration (ppm) vis Peak area (µS/Min))
Awad Albalwi
Graph2: The calibration curve for chloride (Concentration (ppm) vis Peak area (µS/Min))
Graph3: The calibration curve for Flouride (Log (ppm) vs ( Mv)
Graph4: The calibration curve for chloride ((Log (ppm) vs ( Mv))
Awad Albalwi
Discussion/Questions-
The results obtained with ISE and IC methods are shown in the tables and graphs above.
The calculated errors for the ISE method are significant larger for the chloride
analysis then for fluoride. The main reason for this can be attributed to the calibration
curve. The calibration curve for fluoride is very linear with an R2 value of
0.9997(graph.3). The calibration curve for chloride is not as linear with an R2 value of
0.9692(graph.4). The main error in the chloride calibration curve lies in the
measurements made at lower concentrations, the measurements at 0.1 ppm was below
the linear while the 1 & 10 ppm were above of the linear relationship (graph.4) . When a
trend line ( chloride) was fitted to all these points the line has an R2 value less than
would be obtained if these three points were not involved. The deviation from linearity
results in higher sx values obtained and hence a higher estimate for the precision is
obtained. The most likely reasons for the errors in the measurements at lower
concentrations of chloride was that the electrode was not given sufficient time to
equilibrate. Although the electrodes were allowed significant time (about 10 minets) to
Awad Albalwi
equilibrate before the measurements were taken the output never reached a steady
value as it was always changing slowly. This meant accurate measurements could not be
made as the instrument never reached a steady output. The other likely reason for the
errors in measurements of chloride concentration that the amount of solution (25.5 ml)
in electrode beaker was nearly enough only to cover the electrode heads in comparing
with the nearly double amount of solution (45ml) in fluoride electrode case.
The precision of the chloride measurements could be improved by allowing the
electrodes more time to stabilise in solution or by applying certain statistical tests to
justify rejecting the points at lower concentrations to give a more linear line and hence a
smaller sx value.
The sensitivity of ISE towards each analyte can be obtained from the slope of the
calibration curve. As the calibration curve plots the instrument response vs the log of
the concentration a more negative slope will mean a higher sensitivity towards the
particular analyte. the sensitivities of ISE towards Cl- and F- in units of mv/ log(ppm)
are -40.71 and –59.828 respectively(graph. 3&4). This indicates the lanthanum fluoride
(LaF3) electrode used to measure fluoride is more responsive to changes in fluoride
concentration than the (AgCl ) chloride electrode is to changes in chloride concentration.
The limits of detection of the chloride and fluoride ISE’s were not calculated.
The chloride ISE uses a reference electrode that consists of a double junction
silver/silver chloride cell. Reference electrodes are used to complete the redox reaction
and to provide a constant potential against which the potential for the electrode that
Awad Albalwi
responds to the analyte can be measured against. The silver/silver chloride reference
electrode consists of a silver wire coated with silver chloride and immersed in a solution
containing concentrated (KCl) saturated with silver chloride. The redox reaction in the
cell occurs as the (AgCl) coated on the wire receiving an electron causing chloride to go
into solution and vice versa. The fluoride concentration in the berocca solution was
found to be very low concentration as such the result is not reported.
During the analysis of fluoride with the ISE it was extremely important that the
pH of the solution was controlled. If the pH is greater than 8 hydroxide ions can
interfere with the measurement. The hydroxide ions can diffuse into the electrode and
cause the electrode to respond as if fluoride was present leading to incorrect
calculations of the fluoride content in the solution. At low pHs undissociated HF may
form which does not cause the a response in the elcetrode. It is for this reason that a
Total Ionic Strength Adjustment Buffer (TISAB) was added to all fluoride solutions to
be analysed by ISE. The TISAB contains some citrate to bind all Fe(III) ions and a pH
buffer to reduce the effect of hydroxide ions or HF on the measurement of fluoride in
the solutions.
The calibration curves for the IC data were very linear with R2 values of 0.9997
and 0.9974 for chloride and fluoride respectively (graph2&1). The precision obtained
for the fluoride measurements were mostly similar between the ISE and IC methods in
unknown and toothpaste samples . The errors for fluoride in IC were larger than those
obtained in ISE, the reason for this is due to the lower R2 value for the IC calibration
Awad Albalwi
curve than the ISE curve resulting in larger sx values being obtained. However the
result obtained was not excepted for Tap water because the floured concentration
from ISE was larger by triple than it obtained from IC.
The precision obtained for the chloride measurements were not similar between the
ISE and IC methods in unknown , toothpaste and tap water samples . The differences
between the measurements might be due to the lower R2 value for the ISE calibration
curve than the IC curve(graph.2&4) . Therefore , It can be said that the IC is more
accuracy method compare to ISE.
The limit of detection (LOD) of the IC method was not measured. In order to obtain
accurate results using a conductivity detector for IOC it is crucial a suppressor column
is used so that the conductivity of the electrolyte solution does not mask the
conductivity of the analytes. In this investigation a Na2CO3 resin is used. As fluoride and
chloride anions pass through then column they exchange with the carbonate anions in the
column. The released carbonate anions then travel through the column. If they did not
pass through a suppressor column as they reach the detector a huge conductivity will be
measured which will mask the signal of the analyte. Instead the carbonate passes
through the suppressor column where it reacts with the H2SO4 to produce the weak acid
H2CO3 which will have a much lower conductivity the carbonate anion thus allowing good
detection of the analytes.
Conclusion:
Both ion chromatography and ion selective electrodes were successfully used to
measure the concentration of chloride and fluoride anions in a variety of solution.
However, IC has many advantages in operation than ISE .
Awad Albalwi
References:
1. Harris ,D C,2007, Quantitative chemical analysis,(7th Ed) W H Freeman.
2- Rundle, C C, 2000, A Beginners Guide toIon-Selective Electrode Measurements, http://www.nico2000.net/Book/Guide1.html 3- USEPA, 1996, Method 9214: Potentiometricdetermination of fluoride in aqueous samples with an
ion-selective electrode 3- USEPA, 1996, Method 9212: Potentiometric determination of chloride in aqueous samples with
an ion-selective electrode
No. of points log[Cl
-]
(ppm) y (mV) xy x^2 d d^2 6
-1 266.1 -266.1 1 -5.63746 31.78097
0 238.2 0 0 5.953512 35.4443
1 201.6 201.6 1 8.844485 78.22491
1.301029996 183.4 238.6089 1.692679 2.532452 6.413314
1.397940009 170.1 237.7896 1.954236 -6.94048 48.17022
1.698970004 160.4 272.5148 2.886499 -4.75251 22.58635
0 0 0 0
0 0 0 0
0 0 0 0
0 0 0 0
0 0 0 0
0 0 0 0
0 0 0 0
0 0 0 0
0 0 0 0
Awad Albalwi
0 0 0 0
0 0 0 0
0 0 0 0
Sums 4.397940009 1219.8 684.4133 8.533414 4.55E-13 222.6201
+/-
S.d. (y) 7.460229 D= 31.85861
m= -39.491 3.237534253
b = 232.2465 3.861003228
Measured y 191.1
Measured 1 times
Calc. X 1.041921
-0.205611627
Sample ISE IC
[Cl
-]
(ppm) [F
-]
(ppm) [Cl
-]
(ppm) [F
-]
(ppm) Unknown 20+/-4 27+/-1 20+/-3 27+/-5 Toothpaste
A 9+/-4 23+/-1 8+/-3 18+/-5 Toothpaste
B 8+/-4 15+/-1 6+/-3 13+/-4 Berocca 177+/-5 <LOQ 131+/-7 <LOD Tap Water 13+/-4 <LOQ 20+/-3 5+/-5
Method ISE IC Cl
- F- Cl
- F-
LOD (ppm) 2 3 0.8 0.3
LOQ (ppm) 6 9 2.54 1
y = -39.491x + 232.25 R² = 0.9738
0
50
100
150
200
250
300
-1.5 -1 -0.5 0 0.5 1 1.5 2
mV
log[Cl] (log ppm)
Cl Caibration Curve
Awad Albalwi
Sensitivity
Recoveries ISE IC
Cl- F- Cl
- F-
Berocca 11132% n/a 8240%
Toothpaste A n/a
55+/-2% n/a
43.6+/-12%
Toothpaste B n/a 41+/-3 n/a
35.9+/-10%
Awad Albalwi
lab manual
INTRODUCTION One measure of the health of a natural water system is the level of dissolved salts (ionic
substances) it contains. Although conductivity measurements can be used, the concentration
of individual ions in natural water systems is of considerable environmental interest. For
example high concentrations of chloride contribute to salinization of the water. High
concentrations of sulfate ion may indicate that the body of water is receiving significant
quantities of runoff in which contains sulfuric acid (such as that from acid sulfate soils).
High concentrations of nitrate and phosphate ions may indicate that the body of water is
receiving run-off from lawns or agricultural land which often are treated with high levels of
fertilizer. These nutrients stimulate the growth of algae which in turn contribute to
eutrophication (oxygen depletion) of the body of water and subsequent death of aquatic
animals. Fluoride is added to drinking water and toothpaste to inhibit dental caries; it is also
present in effluents from many industrial processes, e.g. the manufacture of fluoro-
polymers like Teflon.
There are many different methods of anion analysis. Two of the most widely used in industry
today are Ion Chromatography (IC) and Ion Selective Electrodes (ISEs).
Direct potentiometric measurements using ISEs provide a rapid, convenient and relatively
inexpensive method for determining a number of anions (and cations). ISEs are available for
measuring more than 20 different cations (e.g., Ag+, Na
+ , K
+, Ca
2+, Cu
2+) and anions (e.g., F
-,
Cl-, S
2-, CN
-). These probes are widely used inside and outside of laboratories. They are
readily adapted to automatic and/or continuous recording of analytical measurements within
labs, within industrial plants where monitoring is needed, and in environmental settings. Page
11
IC is particularly useful in the analysis of mixtures of cations or mixtures of anions. By using
an ion exchange medium (either cation or anion) as the stationary phase in what amounts to a
liquid chromatography system, separation of the ions in solution can be achieved. This
permits IC to carry out multi-analyte determinations on a single sample, which has almost
made it the method of choice for anion analysis.
In this practical you will apply both of these methods to the analysis of chloride and fluoride
in several different water samples, as well as in the determination of the fluoride and
chloride contents of toothpaste. Page 12
Awad Albalwi
SAMPLES
Prior to undertaking the analysis with either method, you will need to collect and prepare
the required samples. Once prepared, you should use the same samples in both weeks.
Collect your unknown in a rinsed 250.0 mL volumetric flask, and make up to volume. In
addition, collect a sample of tap water. To prepare your toothpaste sample:
1. The amount of toothpaste required will depend on the manufacturer – refer to the
approximate weights provided for each brand. Accurately weigh the required amount (usually
1.5 – 2.0 g) of toothpaste into each of two 100 mL plastic digestion tubes. Add about 50 mL
of water to each vessel.
2. Heat the mixture for 10 minutes. Cool the solution, quantitatively transfer the liquid to a
100.0 mL volumetric flask and dilute to volume.
3. With the remaining beaker, stir well and quantitatively transfer the liquid to a 100.0 mL
volumetric flask and dilute to volume.
4. Prepare the “Borrocca samples as per “typical” serve described on the side of the
container.
EXPERIMENTS Carry out the ISE and IC experiments in your groups in the order determined by the lab
schedule. Complete the calculations required for each method.
REPORTING Your report should provide all calibration graphs and estimates of uncertainties in the
determinations carried out with both methods.
You should present your results for each sample in a comparison table, and provide a
discussion of the relative merits of each analytical method. Don’t forget to consider the
commonly used analytical figures of merit (such as accuracy, precision, selectivity, limit of
detection) in your discussion. You should also comment on any differences you observed
between the two toothpaste sample preparation methods, as well as the impact of Fe3+
on
your results. In addition, your discussion should consider the following technical questions:
1. The reference electrode required for a chloride ISE is a double junction silver/silver
chloride electrode. What is this electrode and why would it be needed?
2. Why is pH control important in the case of fluoride determination by ISE?
3. Why would a sulfide (Ag2S) ISE have less interference than the other silver halide based
ISEs?
4. Try to assign all of the anion peaks you observe in the ion chromatogram of the
toothpaste and Borocca™ sample.
5. In IC, explain the role of the suppressor used in conjunction with the conductivity
detector.
Awad Albalwi
PART A ION SELECTIVE ELECTRODES The fluoride ISE consists of a single crystal of lanthanum fluoride that acts as an insoluble
membrane. Ionic conduction is the mechanism by which this membrane responds to changes
in the activity of fluoride ion in a solution. Lattice defects in the crystal allow small ions with
a low formal charge (F-) a degree of mobility within the membrane. It is this shift in charge
centres which give rise to the membrane potentials.
The fluoride electrode is one of the best ion-selective electrodes available. The electrode
has a Nernstian response to 10-5
M F- (0.2 ppm) and selectivity of at least 1000:1 for F
- over
all other anions except OH-. Free fluoride concentrations are, however, subject to
interference through complexation by metal ions. Page 13
The response of the ISE to the F- ion concentration in solution at 298K is (ideally) given by
the Nernst equation:
pFKvE0592.0+= Ev = electrode potential (V) vs a reference electrode
K = cell constant (V)
pF= -log [F-], where [F
-] is in mol L
-1
The chloride ion electrode works on a similar principle, using a silver halide-silver sulfide
membrane. The membrane actually responds to the silver concentration (or activity) in
solution,
E = Eo + (RT/F) ln a
Ag +
but due to the low solubility of silver chloride, the silver concentration is dependent on the
chloride concentration:
K sp
= aAg
+ aCl-
E = Eo+ (RT/F) ln K
sp/ a
Cl-
E = E o- (RT/F) ln a
Cl-
Hence the measurements can be directly related to the chloride concentration. In its more
usual form, at 298K:
pClKvE0592.0+= where these symbols were defined above.
The chloride ISE is susceptible to interference from other ions. This is due to the
insolubility of other ionic silver compounds, so these anions also govern the activity of silver
ions in solution and thus the response of the membrane.
For ISE measurements, temperature and ionic strength of the samples and standards should
be the same. Recall that the electrode responds to the activity of the analyte, not
specifically the concentration. The activity and concentration are related by the activity
coefficient γ, which is dependent on the ionic strength of the solution. The ionic strength is
made identical in all solutions by addition of a known amount of some form of Ionic Strength
Adjuster (ISA). The Page 14
Awad Albalwi
ionic strength of the analyte solution will also influence the rate and extent of interaction
of the analyte ions with the membrane of the ISE.
Several methods of getting anion concentration can be employed:
o a direct reading is simple and quick, but the instrument is only calibrated against a single
standard. It is inherently error-prone, as it assumes a Nernstian response, although is useful
for the routine analysis of a large number of samples.
o a standard addition procedure, which involves adding a standard of known concentration to
a sample solution. From the change in electrode potential before and after addition, the
original sample concentration is determined. Standard addition is also a useful method for
measuring occasional samples. Any convenient concentration unit can be used for direct
measurement and standard addition.
o the most time-consuming, but perhaps most accurate, involves the preparation of a
calibration curve - a plot of mV vs log([F-]) or log([Cl
-]) for a series of standards.
In this part of the practical exercise the concentration of F- and Cl
- in various samples is
estimated by using a calibration curve.
A potentiometric method is dependant upon ionic adsorption equilibria being established at
the electrode (crystal/solution) interface. This can take a little time so wait for the
response to stabilise before you take a reading. This will normally take about 30 s or so
(What does this mean about the measurement bandwidth?).
EXPERIMENT Each pair or individual should carry out both the chloride and fluoride measurements.
REAGENTS
Total Ionic Strength Adjustment Buffer (TISAB) for F-
Ionic Strength Adjuster (ISA) for Cl- (5M NaNO
3)
Chloride stock solutions (1000 mg/L)
Fluoride stock solutions (1000 mg/L)
Fluoride stock containing Fe3+
APPARATUS
Orion pH/mV meter
Magnetic stirrer
Fluoride ion selective electrode
Chloride ion selective electrode
Reference electrode
Double junction reference
electrode
Awad Albalwi Student No:3343297
1
C A U T I O N: Do not touch the ISE ion sensitive membrane. Rinse it
with deionized water between measurements and then with a small volume
of the new solution. Do not wipe it dry.
PROCEDURE
You will use the Nernstian relationship to draw up a calibration curve to
estimate the concentration of fluoride and chloride your samples. Since the
ISE response is logarithmic you should prepare standards that are 0.1, 1, 10,
20, 30, 40, and 50 ppm F- and Cl-.
CALIBRATION CURVE FOR FLUORIDE ANALYSIS
1. All readings should be done in solutions which are 1:1 TISAB/sample. Pipette
20 mL of your first standard (lowest concentration) and 20 mL TISAB into a
clean, dry beaker.
2. Analyse using the Orion ion selective electrode placing the electrodes in the
stirred solution. Record the mV reading.
3. Repeat for each of the standard solutions and for your unknown sample and
tap water. Rinse electrode with distilled water between samples.
4. Now measure the 25 ppm fluoride standard that contains Fe3+
(check with
your demonstrator about concentrations). Repeat the measurement procedure
using 20 mL of this solution, and compare your result with the iron-free 20
ppm fluoride reading.
Since the dilution with TISAB applies to all solutions you are to analyse there
is no need to take this dilution into account. For example, 50 mL 100ppm F- +
50 mL TISAB can still called 100ppm F- when recording the result.
CALIBRATION CURVE FOR CHLORIDE ANALYSIS
1. For each of the standards and samples take a 25.00mL aliquot and add
0.5mL of the ISA (for Cl-).
2. Place electrodes into the standard with the lowest concentrations and stir
using the magnetic stirrer. Wait until there is a stable reading. Repeat this
for all the other standards (in the order of increasing concentration), and
then the samples.
CALCULATIONS Based upon results obtained above, and the Nernst equation (as given), draw
calibration curves with the appropriate axes and estimate the concentration
of fluoride and chloride in the samples tested; use these curves to provide an
estimate of the measurement precision of your results.
REFERENCES • Rundle, C C, 2000, A Beginners Guide to Ion-Selective Electrode
Measurements, http://www.nico2000.net/Book/Guide1.html
• USEPA, 1996, Method 9214: Potentiometric determination of fluoride in
aqueous samples with an ion-selective electrode
• USEPA, 1996, Method 9212: Potentiometric determination of chloride in
aqueous samples with an ion-selective electrode
Awad Albalwi Student No:3343297
2
• Thermo-Orion ISE manuals Reproduction rights obtained Feb 2007
Page 15
Awad Albalwi Student No:3343297
3
PART B ION CHROMATOGRAPHY Example ion chromatogram illustrating the separation of anions commonly found in water
samples
BEFORE YOU COME TO THE LABORATORY
Ion chromatography systems have several components (injection valve, guard
column, supressor, etc.) that in our case are integrated into a single unit. You
should be familiar with the function of each of these components prior to the
laboratory session. Refer to your 214 chromatography notes, or a relevant
text.
EXPERIMENT
REAGENTS
Chloride stock solutions (1000 mg/L)
Fluoride stock solutions (1000 mg/L)
Samples and unknowns
APPARATUS
Dionex ICS-90 ion chromatography system
Note down column, eluent and injection loop details
Caution: ALL solutions must be filtered using a 0.45 μm membrane filter
as they are injected into the ion chromatography system.
PROCEDURE
1. Familiarise yourself with the operation of the ion chromatography system.
Ask one of the demonstrators to help you can identify the following
components:
o The eluent reservoirs (record the nature of the eluent solution)
o The injection loop (note its volume) and valve
o The pump
o The chromatography column (make sure you write down the column type)
o The suppressor
o The conductivity detector
2. Using one of your sample solutions, fill the injection loop using a syringe
that has a 0.45 μm filter attached. Start the pre-loaded chromatography
method.
Page 16
Awad Albalwi Student No:3343297
4
3. While the first chromatogram is developing, prepare six combined standard
solutions with 0.1, 1, 10, 20, 30, 40 and 50 mg/L concentrations by making
suitable dilutions of the anion stock standard.
4. Examine your chromatogram and identify the anions responsible for each
peak. You may carry out additional injections (eg. of single component
solutions) if this helps
5. Continue to inject your group’s samples and standards into the
chromatography system. Make sure you wait until the injection loop is full and
the last chromatogram has fully developed before you start the next run.
6. Using the chromatogram for the 0 ppm standard, make an estimate of the
noise in the baseline.
CALCULATIONS Tabulate the areas of each of the peaks in your sample and standard
chromatograms. Use the results for your standards to generate a calibration
curve for each anion, using linear regression to generate an equation for your
response function. Use the regression equations to determine the
concentration of each anion in your samples, along with estimates of the
measurement precision.
Calculate the concentration Limit of Detection for each anion that you can
achieve with this method. Is the sensitivity the same for each anion?
REFERENCES
1. Harris, D C, 2003, Quantitative Chemical Analysis, (6th
Ed), W H Freeman
2. Dionex ISC-90 Manual
3. Dionex IC Application notes: http://www1.dionex.com/en-
us/by_technique/ic/anion/lp5008.html Copyright © 2006 by Sidney Harris