CHEM 4461/5461 INSTRUMENTAL ANALYSIS Laboratory Manual … · 2011-01-21 · CHEM 4461/5461 Course...
Embed Size (px)
Transcript of CHEM 4461/5461 INSTRUMENTAL ANALYSIS Laboratory Manual … · 2011-01-21 · CHEM 4461/5461 Course...

CHEM 4461/5461INSTRUMENTAL ANALYSIS
Laboratory Manual&
Course Guide
Spring Semester, 2011
Department of Chemistry and Biochemistry
The University of Texas at Arlington

2
TABLE OF CONTENTS
Section Page
Course Syllabus… 3
Laboratory Schedule… 6
Laboratory Report Format and Pre-Lab Quiz Information… 7
Overview of Spectroscopy… 11
Overview of Chromatography… 18
Experiment 1: UV-Vis Spectroscopy… 23
Experiment 2: Fluorescence Spectroscopy… 27
Experiment 3: Thermogravimetric Analysis… 28
Experiment 4: Flame Atomic Absorption Spectroscopy… 30
Experiment 5: Gas Chromatography… 33
Experiment 6: High Performance Liquid Chromatography… 38
Experiment 7: Nuclear Magnetic Resonance Demonstration… 42
Experiment 8: Mass Spectrometry Demonstration… 43

CHEM 4461/5461 Course Syllabus Instrumental Analysis
3
CHEM 4461/CHEM 5461
INSTRUMENTAL ANALYSIS (Spring 2011)
LABORATORY COURSE SYLLABUS
Instructors: Richard X. Guan
CRB, 103
817-272-6086
- Contact relating to miscellaneous items by email preferred
Office hours: MWF 10 – 11 am
or by appointment
Text: Skoog, Holler, and Crouch, Principles of Instrumental Analysis, 6th Ed.
and the Student Laboratory Manual.
Class Schedule: CPB 215 Section 101: T, Th: 6 – 9:50 pm
Grading: Lecture course: 500 points
Lab course (400 points total):
Experiments: 240 points (8 reports x 30 pts each)
Practicum: 140 points (oral + written portion)
Attendance/Lab Technique: 20 points
Description and Goals of the Course: This course explores the fundamental basis of chemical analysis. It
is designed to give the student a solid conceptual ground to understand how a given analytical technique
works; including its limits and advantages. The emphasis is on solutions analysis and the course is
roughly divided into: (i) Basic measurements and concepts; (ii) spectroscopy; and (iii) chromatography
and mass spectrometry.
Dropping: When dropping the course, YOU are responsible to see that the proper paperwork is filed
with the Chemistry Department. Failure to do so will result in a grade of ‛F‛.
Drop for non-payment of tuition: If you are dropped from this class for non-payment of tuition, you may
secure an Enrollment Loan through the Bursar’s Office. You may not continue to attend class until your
enrollment loan is applied to outstanding tuition fees.

CHEM 4461/5461 Course Syllabus Instrumental Analysis
4
Grade Replacement: Students enrolling in this course with the intention to of replacing a previous grade
earned in the same course must declare their intention to do so at the registrar’s office by Census Date
(Sept. 7 in F2005) of the same semester in which they are enrolled.
Pass/Fail: If P or F is a grade option in this class and you intend to take this class for a pass/fail grade
instead of a letter grade, you MUST inform the instructor(s), through the necessary paperwork, before the
Census Date.
Bomb Threat Policy: In the event of a bomb threat to a specific facility, University Police will evaluate the
threat. If required, exams may be moved to an alternate location, but they will NOT be postponed. UT-
Arlington will prosecute those phoning in bomb threats to the fullest extent of the law.
Americans with Disabilities Act: The University of Texas at Arlington is on record as being committed to
both the spirit and letter of federal equal opportunity legislation; reference Public Law 93112-The
Rehabilitation Act of 1973 as amended. With the passage of new federal legislation entitled American
with Disabilities Act-(ADA), pursuant to section 504 of The Rehabilitation Act, there is renewed focus on
providing this population with the same opportunities enjoyed by all citizens.
As faculty members, we are required by law to provide ‚reasonable accommodation‛ to students with
disabilities, so as not to discriminate on the basis of that disability. Student responsibility primarily rests
with informing faculty at the beginning of the semester and in providing authorized documentation
through designated administrative channels.
Academic Dishonesty: It is the philosophy of the University of Texas at Arlington that academic
dishonesty is a completely unacceptable mode of conduct and will not be tolerated in any form. All
persons involved in academic dishonesty will be disciplined in accordance with University regulations
and procedures. Discipline may include suspension or expulsion from the University.
‚Scholastic dishonesty includes but is not limited to cheating, plagiarism, collusion, the submission for
credit of any work or materials that are attributable in whole or in part to another person, any act
designed to give unfair advantage to a student or the attempt to commit such acts.‛ (Regents’ Rules and
Regulations, Part One, Chapter VI, Section e, Subsection 3.2, Subdivision 3.22)
Mandatory Online Safety Training: Students registered for this course must complete the University’s
required ‚Lab Safety Training‛ prior to entering the lab and undertaking any activities. Students will be
notified via MavMail when their online training is available. Once notified, students should complete the
required module as soon as possible, but no later than their first lab meeting. Until all required Lab Safety
Training is completed, a student will not be given access to lab facilities, will not be able to participate in
any lab activities, and will earn a grade of zero for any uncompleted work.
1. You should have received an email from the UTA Compliance Department. Click on the link in the
email (or navigate to https://training.uta.edu for the login page)
2. Log on using your network log-on ID and password (what you use to access email). If you do not
know your NetID or need to reset your password, visit
http://oit.uta.edu/cs/accounts/student/netid/netid.html.
3. The available courses for completion will be listed. For Chemistry 1441, complete the course entitled
‘Student Lab Safety Training’

CHEM 4461/5461 Course Syllabus Instrumental Analysis
5
4. If you did not receive the training email and you have not already completed the training you will
need to contact the training helpline (817-272-5100) or email [email protected].
5. Students who have not completed the training by census date may be dropped from the lab (and
consequently the lecture).
Once completed, Lab Safety Training is valid for the remainder of the same academic year (i.e. through
next August) for all courses that include a lab. If a student enrolls in a lab course in a subsequent
academic year, he/she must complete the required training again.
All questions/problems with online training should be directed to the University Compliance Services
Training Helpline at 817-272-5100 or by emailing [email protected].

CHEM 4461/5461 Laboratory Schedule Instrumental Analysis
6
LABORATORY SCHEDULE (Spring 2011)
Experiment 1: UV-VIS Spectrophotometry (2 lab periods)
Experiment 2: Fluorescence Spectrophotometry (2 lab periods)
Experiment 3: Thermogravimetric Analysis (2 lab periods)
Experiment 4: Atomic Absorption Spectrophotometry (2 lab periods)
Experiment 5: Gas Chromatography (2 lab periods)
Experiment 6: Liquid Chromatography (2 lab periods)
Experiment 7: NMR Spectroscopy (Demo Lab, 1 lab period)
Meet in CRB B20 at 6:00 pm
Experiment 8: Mass Spectrometry (Demo Lab, 1 lab period)
Meet in CPB 234 at 6:00 pm
Week of: Task
01/18: No Labs; Complete safety requirements (http://compliance.uta.edu/training)
03/15: No Labs ; Spring break.
03/17: No Labs ; Spring break.
Group1 Group2 Group3 Group4 Group5 Group6
02/01: Expt 1 Expt 2 Expt 3 Expt 4 Expt 5 Expt 6
02/08: Expt 2 Expt 3 Expt 4 Expt 5 Expt 6 Expt 1
02/15: Expt 3 Expt 4 Expt 5 Expt 6 Expt 1 Expt 2
02/22: Expt 4 Expt 5 Expt 6 Expt 1 Expt 2 Expt 3
03/01: Expt 5 Expt 6 Expt 1 Expt 2 Expt 3 Expt 4
03/08: Expt 6 Expt 1 Expt 2 Expt 3 Expt 4 Expt 5
Section 101
03/22:
03/22 (Tu): Expt 7
03/24 (Th): No Labs
03/29:
03/29 (Tu): Expt 8
03/31 (Th): No Labs
03/31: Presentations of Practicum Experiments; Location TBA
Practicum experiments distributed in lecture class on Thursday, February 24
04/05 – 05/05: Perform Practicum Experiments
Practicum Lab Reports due Friday, May 6 by 5:00 pm

CHEM 4461/5461 Lab Report Format and Pre-Lab Quizzes Instrumental Analysis
7
LABORATORY REPORT FORMAT and PRE-LAB QUIZZES Each lab report + pre-lab quiz is worth 40 points
Your laboratory report for each experiment should contain the following information.
It should roughly conform to the format of a standard scientific research article.
I. Introduction (5 points)
a. Background on analytical technique
i. Theory
ii. Advantages and disadvantages relative to other techniques
iii. Common applications (several literature references appropriate here)
b. Aim of the experiment - one or two sentences describing objective of experiment
II. Experimental (4 points)
a. Description of particular instrument used
b. Experimental procedure
III. Results (8 points)
a. Include data, required calculations, figures and tables. Each Figure and Table
should be addressed/mentioned/clarified/explained in the text of the report.
b. Describe only the results of the experiments
c. Subdivide for clarity when multiple experimental goals are present
d. Include statistics to justify non-significance of data points removed
IV. Discussion (8 points)
a. Analyze and discuss the results
b. How do the results of the experiment achieve the objective described in the
introduction?
c. Were the results the same or different than expected? WHY?!
d. If poor results were obtained, explain why. A bad result is still a result. If it is
explained then it is not wrong.
e. Answer assigned questions in the course of discussion.
V. Conclusion (5 points)
a. One paragraph summarizing: a) the objective; b) the experiment; c) the results;
d) the outcome
VI. References (5 points) – minimum three primary literature references needed (not
including websites or textbooks)
Pre-lab quiz (5 points) – Pre-lab quizzes will be administered at the beginning of each
experiment and will cover the experiment that will be performed on that day. A list of possible
(but not all-inclusive) pre-lab questions is given below. Pre-lab quizzes will only be given for
Experiments 1 – 6.

CHEM 4461/5461 Lab Report Format and Pre-Lab Quizzes Instrumental Analysis
8
HELPFUL HINTS/CHECKLIST (These will be very helpful…):
- Use the headings for each section given in bold type above. Keep to the format. Avoid
redundancy.
- Use references/citations (ACS Style Format) throughout to show where you found
your information (books, articles, web pages). Use a consistent format. Here you can
find a good set of guidelines: http://www.lib.berkeley.edu/CHEM/acsstyle.html
- Number figures, tables, structures, equations and refer to them in the text as they are
presented and discussed.
- Figures and tables should be numbered sequentially and should have appropriate
captions. Captions for figures are placed under the figure. Captions for tables are
placed at the top of the table.
- Figures and tables are most easily created in powerpoint. Then they can be easily pasted
into the word document (best is often “Pasted Special” as a “Window Metafile”) and
properly formatted.
- Write in essay-style prose format. Make the paper flow along the outline above. Avoid use
of bullet-point lists
- Do not include personal pronouns or personal feelings
- Write as though this is the first time such a measurement is being performed and you
are trying to communicate the procedure, the merit, the results, and the overall outcome
of the experiment.
- Look at examples of research articles in the scientific literature if you have questions
about how material should be presented. Look at e.g. Analytical Chemistry or the
Journal of the American Chemical Society
SUBMISSION:
All laboratory reports must be submitted in electronic format to your TA by 5 pm on the
Friday of the week after the lab was completed. For example, if you finish your lab experiment
on Tuesday or Thursday of one week, then the lab report is due on Friday of the following
week.
Note: All lab reports received after 5 pm on the due date will be assessed a 10% penalty. For
each additional day that the report is late, an additional 10% penalty will be assessed.

CHEM 4461/5461 Lab Report Format and Pre-Lab Quizzes Instrumental Analysis
9
PRE-LAB QUIZ QUESTIONS:
Expt 1 (UV-Vis)
Why is hexane a good solvent for performing measurements by UV-Vis
spectrophotometry?
Give Beer’s law and define each variable.
What is the approximate wavelength range associated with ultraviolet light?
Is visible light higher in energy or lower in energy than ultraviolet light? Explain.
Experiment 1 is broken down into three parts. Briefly describe the measurement you
will perform in each part.
Draw an example of the calibration plot you will create in your experiment related to
unknown determination. Be sure to label all parts of the plot as detailed as possible.
Expt 2 (Fluorescence)
For a particular system that fluoresces, the energy associated with fluorescence
emission is always ________________ than the energy associated with the associated
absorptive excitation.
With respect to instrumental set-up why is fluorescence spectrophotometry more
sensitive for the trace measurement of a fluorescent molecule compared to UV-Vis
spectrophotometry?
What is the sample that will serve as the unknown in your laboratory experiment?
What is the role of vibrational relaxation in the fluorescence process?
Two calibration curves will be developed in your laboratory experiment for the
fluorescence measurement of quinine. Which experimental variable will you change
to create these two separate plots?
Sketch a fluorescence spectrophotometer and label each part.
Expt 3 (TGA)
What is thermogravimetric analysis?
Why nitrogen gas is used in thermogravimetric analysis?
For what two analytes will you measure thermograms?
The heating of the hydrated salts in this experiment causes the loss of waters in
characteristic increments. How do you expect the “features” displayed in the
thermogram to change as a function of heating rate?
Does the sample need to be volatile in thermogravimetric analysis?
Expt 4 (Flame Atomic Absorption)
Draw a general diagram of a flame atomic absorption instrumental setup and label all
parts.
What metals will you focus on quantifying in this experiment?

CHEM 4461/5461 Lab Report Format and Pre-Lab Quizzes Instrumental Analysis
10
What is a major limitation of flame AA in terms of monitoring the concentration of a
wide range of metals in a sample?
Why do atomic absorptions appear as sharp lines rather than the broad lines that are
observed in molecular absorption spectroscopy?
Draw and label in detail a sample calibration plot for the experiment you will
perform.
Expt 5 (Gas Chromatography)
Draw a general diagram of a gas chromatograph instrument and label all parts.
What is the carrier gas that you will employ to perform your GC experiments?
What type of detector is used for your GC experiments?
What sample will you measure for determination of Kovatt’s Retention Indices?
Define what is meant by the term “dead time” (t0)?
What parameter is calculated to determine whether or not you have achieved baseline
separation of your analytes of interest?
Expt 6 (High Performance Liquid Chromatography)
In reversed phase chromatography, polar analytes are separated based on their
relative degree of _______________________.
What type of stationary phase will you employ to perform your separations?
What are the two mobile phase components that you will use to perform you
separations?
What type of detector is used on the HPLC system you will be using?
What is meant by the term “isocratic separation”?
How many analyte components will you have to separate?
Draw and label a general diagram of an HPLC instrument.

CHEM 4461/5461 Overview of Spectroscopy Instrumental Analysis
11
OVERVIEW OF SPECTROSCOPY
Spectroscopy is one of the most widely used techniques in chemistry and biochemistry.
The most widely used spectrometric methods are based upon interaction of
electromagnetic radiation (EMR) and matter.
Electromagnetic radiation (e.g. light) is a form of energy with both wave and particle
properties. What can happen to the light intensity as it passes through a medium? It
slows down in media other than vacuum because its electric vector interacts with
electric fields in the medium.
Absorption and emission are two most interesting and most useful processes when
EMR interacts with matter. When a molecule absorbs a photon, the molecule gains
energy. On the other hand, when a molecule emits a photon, the molecule loses energy.
Based on quantum mechanics, the energy of a photon depends on its frequency (v):
Ephoton = hv, where
h = Planck’s constant
h = 6.63 x 10-27 erg sec or 6.63 x 10-34 Js
Postulates of Quantum theory include:
1. Atoms, ions, and molecules exist only in discrete energy states
E0 = ground state
E1, E2 , E3 ... = excited states
- Excitation can be electronic, vibrational or rotational.
- Energy levels for atoms, ions or molecules are different.
- Measuring energy levels gives means of identification – spectroscopy.
2. When an atom, ion or molecule changes energy state, it absorbs or emits energy equal
to the energy difference.
ΔEtransition = E1 - E0 = hv = hc/

CHEM 4461/5461 Overview of Spectroscopy Instrumental Analysis
12
Absorption and Emission
Absorption involves transitions from ground
state to excited states. For absorption to
occur, the energy of the photon must exactly
match the energy difference between the
ground state and one of the excited states of
the absorbing species. On the other hand, emission of EMR occurs when excited
particles (atoms or molecules) return to ground state by giving up excess energy.
Atomic Absorption and Molecular Absorption
In atomic absorption, only a few well-defined frequencies are observed in the spectrum,
since for absorption of radiation to occur, the energy of the exciting photon must exactly
match the energy difference between the ground state and one of the excited states of
the absorbing species. Compared with atomic absorption, molecular absorption is more
complex because many more potential transitions exist, i.e., electronic energy level,
vibrational energy level, and rotational energy level transitions. The energy of a
molecule is made up of three components: Emolecule = Eelectronic + Evibrational + Erotational (note
that Eelectronic > Evibrational > Erotational). Eelectronic represents the electric energy of the molecule
that arises from the energy states of its several bonding electrons; Evibrational describes the
energy associated with the interatomic vibrations that are present in molecular species;
while Erotational is the energy caused by various rotational motions within a molecule. It
should be mentioned that the number of rotational states is much larger that that of
vibrational states, and the number of vibrational states is much larger than that of
electric states.
E0
E1
En
……
.
E0
E1
En
……
.
E0
E1
En
……
.
E0
E1
vibrational level
rotational level
E2
E0
E1
vibrational level
rotational level
E2
Energy Levels - Molecular Absorption Energy Levels - Atomic Absorption

CHEM 4461/5461 Overview of Spectroscopy Instrumental Analysis
13
Beer-Lambert Law (Beer’s Law)
Beer’s Law is a fundamental law governing molecular and atomic absorption
spectroscopy. Let’s consider a beam of light with an initial radiant intensity P0 pass
through a layer of solution (with a thickness of b and concentration of c). The intensity
of the light after passage through the solution is P.
According to Beer’s law, there is a linear relationship between absorbance and
concentration of the solution:
A = -log (T) = log (P0/P) = εbc, where
Transmittance (T) = P/P0, and Absorbance (A) = -log (T)
The linear relationship between absorbance and concentration of the solution is the
basis of the quantitative detection employed by molecular and atomic absorption
spectroscopy.
Concentration (c)
Abso
rban
ce (
a)
Incident light P0
hv Emerging light P
b
Absorbing Soln. (C)

CHEM 4461/5461 Overview of Spectroscopy Instrumental Analysis
14
Electromagnetic spectrum encompasses an enormous range of wavelengths and
frequencies (or energies). And thus, various spectroscopic methods are developed by
employing part of the spectrum.

CHEM 4461/5461 Overview of Spectroscopy Instrumental Analysis
15
The following figure shows different portions of the EMR spectrum and different types
of spectroscopy involve quantum states of the atom.
UV-Visible Molecular Absorption Spectrometry is the most common analytical
technique in the analytical laboratory, which involves absorption of ultraviolet and
visible radiation for quantitative purposes. Absorption commonly occurs with many
organic molecules, metals, and metal-organic complexes. In UV-Vis, absorption
involves bonding (outer valence) electrons.
Infrared Spectrometry. Energy of IR photon is insufficient to cause electronic excitation
but can cause vibrational or rotational excitation.
Fluorescence is emission of light from any excited state of a molecule. Fluorescence
and phosphorescence are alike in that excitation is brought by absorption of photons.
The difference between them is that the electronic energy transitions responsible for the
fluorescence do not involve a change in electron spin.
Atomic Spectroscopy is based on excitation of a valence electron to an excited state. In
atomic spectroscopy, a substance is decomposed into atoms by a process called

CHEM 4461/5461 Overview of Spectroscopy Instrumental Analysis
16
“atomization”. The quantity of each element is measured by absorption or emission of
UV or visible radiation by the gaseous atoms. Atomic spectroscopy methods are
categorized based on the type of atomization process, e.g., flame, electrical discharge,
inductively coupled plasma, etc. The type of atomizer used in atomization process will
affect the atomization temperature.
The following diagram is a good schematic representation of underlying energy
chances involved in UV-vis, IR, and fluorescence spectrophotometric experiments.
IR
E0
E1
vis
E2
v1
v2
v3
v4
v1
v2
v3
v4
UV
IR
E0
E1
vis
E2
v1
v2
v3
v4
v1
v2
v3
v4
v1
v2
v3
v4
v1
v2
v3
v4
v1
v2
v3
v4
v1
v2
v3
v4
UV
Spectroscopy and Energy Level Diagram
Relationship between UV, vis, and IR

CHEM 4461/5461 Overview of Spectroscopy Instrumental Analysis
17
Instrumentation Available in the Laboratory
FT/IR
Bruker Vector 22 FT/IR Spectrometer
UV-vis
Jasco V-550 UV-vis Spectrophotometer
Fluorescence
Horiba Jobin Yvon FluoroMax-3 spectrofluorometer
AA
Thermo Scientific - SOLAAR M Series AA Spectrometer

CHEM 4461/5461 Overview of Chromatography Instrumental Analysis
18
OVERVIEW OF CHROMATOGRAPHY
PLANAR
SUPERCRITICAL FLUID
TLC PC
GAS
GSC GLC
BPC-NP GPC GFC BPC-RP
BPCLSC IEC SEC
COLUMN
LIQUID
CHROMATOGRAPHY
PLANAR
SUPERCRITICAL FLUID
TLC PC
GAS
GSC GLC
BPC-NP GPC GFC BPC-RP
BPCLSC IEC SEC
COLUMN
LIQUID
CHROMATOGRAPHY
The chromatographic process:
1. Mobile phase (m.p.) – used for transport of the sample across the stationary phase
a. GC – gas (He)
i. Inert: does not interact with sample
b. HPLC – liquid
i. Reversed phase HPLC (RP-HPLC) – water/organic mixture
ii. Provides a differential affinity from the stationary phase for the analyte
by which a dynamic partitioning effect takes place
2. Stationary phase (s.p.) – present in the column (the “heart” of the chromatograph). A
medium which provides a differential affinity for a mixture of analytes so that each
analyte interacts to a different degree and provides separation.
a. GC – polydimethylsiloxane or other nonpolar/semipolar phase
b. RP-HPLC – C-18 phase
i. Hydrophobic chains which separate analytes based on their degree of
partitioning between the polar m.p. and the hydrophobic s.p.

CHEM 4461/5461 Overview of Chromatography Instrumental Analysis
19
A Typical Chromatogram:
Retention time (tR)
Peak areaPeak
Height
Injection
Baseline
Time
Sig
na
l
t′Rt0
Dead time marker
Retention time (tR)
Peak areaPeak
Height
Injection
Baseline
Time
Sig
na
l
Time
Sig
na
l
t′Rt0
Dead time marker
Components of chromatograph:
1. Mobile phase delivery unit
a. GC – gas cylinder
b. HPLC – pump
2. Injector
a. GC – heated injection port to vaporize sample and deliver it onto column
b. HPLC – manual injection valve with a sample loop
3. Column – stationary phase regime; where the separation takes place
a. GC – polyimide coated fused silica coated internally with s.p.
b. HPLC – 25 cm stainless steel tube packed with silica gel particles (5 µm
diameter); the s.p. (C-18) is bonded onto the silica gel particles
4. Detector – where separated “peaks” are detected
a. GC – flame ionization detector (FID) – detects separated sample components by
“burning” the analyte
i. Others: mass spectrometer, electron capture, thermal conductivity
b. HPLC – UV spectrophotometer; absorbance detection based on analyte
chromophore (reference to UV/Vis fundamentals; Expt. 1)
i. 254 nm: highest absorbance for aromatic structural units

CHEM 4461/5461 Overview of Chromatography Instrumental Analysis
20
ii. Others: mass spectrometer, fluorescence, electrochemical
5. Data collection – computer; converts the detector signal to chromatogram output
Qualitative Analysis
- Determination of the identity of a component in a mixture is done by analyzing
standards under the same experimental conditions as the mixture of interest. The
retention time of an analyte standard will be the same as it is in the sample mixture.
1. Standard
2. Unknown
tR(theophylline) = tR(“X”), therefore, “X” is suggested to be theophylline.
tR(theophylline) = tR(“X”)
Quantitative Analysis
- The response of an analyte (the magnitude of peak height or peak area) will be directly
proportional to the concentration of the analyte under linear operating conditions. A
calibration curve (response vs. concentration) can be made in the same manner as that
performed for previous experiments. This external standard procedure can be used to
extrapolate the unknown concentration of a component in a sample mixture.
Concentration
Re
spo
nse
Standard data points
Unknown concentration
Concentration
Re
spo
nse
Standard data points
Unknown concentration

CHEM 4461/5461 Overview of Chromatography Instrumental Analysis
21
Instrumentation Available in the Laboratory
Gas Chromatography (GC)
Shimadzu GC-17A
He carrier gas
Split/Splitless injector
Autosampler with 10 µL syringe (Manual injection capability as well)
RTX-5 column (95% methyl, 5% phenyl polysiloxane) (15 m L, 250 µm i.d, 0.25 µm df)
Flame ionization detector
Shimadzu Class VP software (Ver. 7.4)
High Performance Liquid Chromatography (HPLC)
Varian Prostar (set up for Reverse Phase mode separations)
Binary, high pressure gradient, pumping system
MeOH and Water mobile phase reservoirs
ODS Hypersil C18 column (100 mm L x 4.6 mm i.d., 5 µm dp)
Variable wavelength UV detector
Varian Star Chrom Workstation software (Ver. 6.4)
Chromatography Terms and Definitions
1. Retention time (tR) – the time an analyte spends in the column. It is measured from the
time of injection to the peak maximum (when the analyte is eluted). It is characteristic of
an analyte for a given chromatographic method.
2. Dead time (tM or t0) – the time required to elute an unretained peak. It can used to
measure the void volume or open space in the column.
3. Adjusted retention time (t′R) – retention time minus dead time. This measures the time
an analyte spends “sorbed” in the column packing. 0ttt RR
4. Retention factor (k′) – the ratio of time the analyte spends in the stationary phase (t′R) to
the time the analyte spends in the mobile phase (t0). Also known as the capacity factor.
0t
tk R

CHEM 4461/5461 Overview of Chromatography Instrumental Analysis
22
5. Selectivity (α) – separation factor for quantifying the difference in retention for two
analytes of interest. α should be reported as a value greater than 1.
1
2
1
2
1
2
K
K
k
k
t
t
,R
,R
6. Width at base (W) – the portion of the baseline intersected by tangents drawn to the
peak. It is equal to 4σ of a Gaussian-shaped peak and is needed to calculate theoretical
plates (N) and resolution (R).
7. Width at half height (Wh) – the peak width at half the peak height. It is used to
calculate N and R.
8. Number of theoretical plates (N) – the number of theoretical solute equilibria for an
analyte between the mobile and stationary phases during separation in a column. N,
also called the plate number, is a measure of efficiency and the “goodness” of a column.
A column with high N produces narrow peaks. 22
54516
h
R
b
R
W
t.
W
tN
9. Height equivalent to a theoretical plate (H or HETP) – the length of column required
for one equilibrium by a solute between the stationary and mobile phases. It is the
length of one theoretical plate. H = L/N. The smaller the value of H, the more efficient
the column. H is also used to measure the rate of band broadening.
10. Average linear flow velocity (u) – linear m.p. flow usually given in cm/s. It is equal to
the length of the column (L in cm) divided by the dead time (t0 in seconds). This is used
to construct van Deemter plots for column efficiency studies.
11. Partition coefficient (K) – the ratio of concentration of an analyte in the stationary phase
to the concentration of the analyte in the mobile phase. A partition coefficient is a
fundamental physical property and is characteristic of the solute and the solvent. K is
constant for a given chromatographic method.
k
mL/g
mL/g
A
AK
.p.m
.p.s
.p.m
.p.s
12. Phase ratio (β) – the volume of the mobile phase divided by the volume of the stationary
phase in a chromatographic column.
13. Resolution (R) – a quantitative measure of the degree of separation of two peaks. R =
1.5 is baseline separation. A more comprehensive measured of resolution is given by the
Master Resolution Equation (Rs). Rs shows that resolution is a product of three
independent factors (efficiency, selectivity, and capacity).
21
12
21
12 1812
,h,h
,R,R,R,R
WW
)tt(.
WW
)tt(R
2
2
1
1
4 k
kNRs

CHEM 4461/5461 Expt. 1: UV-VIS Instrumental Analysis
23
EXPERIMENT 1: UV-VIS SPECTROPHOTOMETRY
PART 1: BEER’S LAW EVALUATION
Sample Preparation
External Standard Calibration: Create the following solutions using standard variable-
volume micropipettors and plastic tips. Create a stock solution by dissolving accurately
approximately 15 mg of quinine sulfate in 15 mL of 0.5% (v/v) sulfuric acid in a glass
vial. From this stock solution, create a series (at least 12) of diluted standard solutions
ranging from 1 to 250 mg/L. Prepare 5 mL of each solution, and again dilute with 0.5%
(v/v) sulfuric acid. Obtain 5 mL of 0.5% (v/v) sulfuric acid to use as a reference blank
solution.
Obtain ~50 mL of degassed tonic water as your unknown. This sample will be
quantitatively analyzed for quinine content using the external calibration standards
prepared above, as well as by standard addition, as follows:
Standard Addition: Deposit exactly 5.000 mL of tonic water into six 15-mL centrifuge
tubes. Using your stock solution of quinine sulfate (~1000 mg/L), aliquot 0, 100, 200,
500, 600, and 800 μL of this standard into the six separate tubes. Dilute to exactly 6.000
mL total with the appropriate volume of deionized water.
Analysis
First, be sure the instrument is on and the computer is responding. The computer
should open the Jasco software automatically. Use the instructions by the machine to
start Spectrum Measurement and create a baseline. Check with the TA if these items
are unclear. Now, in the Spectrum Measurement window, select Measurement from
the menu bar. Select parameters. Set the measurement range to 220 -450 nm, the data
pitch to 0.2 nm, and the scanning speed to 100 nm/min.
Effect of Changes in Scanning Parameters
Fill a cuvette with 0.5% (v/v) sulfuric acid and place it in the reference cell holder in the
UV-VIS spectrophotometer. Using the solution containing ~50 mg/L of stock solution,
record and print a UV spectrum from 220 – 450 nm using the parameters given above.
Next, increase the scanning speed to 1000 nm/min, change the data pitch to 0.5 nm, and
record and print a second UV spectrum. Set the scanning speed back to 100 nm/min
and adjust the data pitch to 2.0 nm. Record and print a third UV spectrum from 220 –
450 nm using these parameters.

CHEM 4461/5461 Expt. 1: UV-VIS Instrumental Analysis
24
Calibration Plot Linearity
Record two UV spectra from 220 – 450 nm for each of the dilute standard solutions,
which were made from the stock solution. Start with the lowest concentration. First
record a set of spectra with scanning rate set to 100 nm/min and data pitch set to 0.2 nm.
For these UV spectra record the absorbance at the peak maximum and just off the peak
maximum (Make sure to note the specific wavelengths monitored). Second, record a set
of spectra with scanning rate set to 1000 nm/min and data pitch set to 0.5 nm. Record
the absorbance at the peak maximum and just off the peak maximum for each spectrum
recorded. These numbers are found using the Spectra Analysis tweezer tool. Click on
and move the red vertical line to your peaks of interest and record the wavelength and
absorbance reported at the bottom of the Analysis window. Note, only the numbers
need to be recorded. It is not necessary to print out each of the UV spectra, but be
consistent in your choice of on-peak-maximum and off-peak-maximum wavelength
selection.
PART 2: UNKNOWN DETERMINATION
Set the scanning rate to 100 nm/min and the data pitch to 0.2 nm and record and print a
UV spectrum of your unknown degassed tonic water from 220 – 450 nm. Record
specifically the absorbance of the unknown at the same wavelengths used above (on
and off peak maximum). Place a cuvette containing 0.5% (v/v) sulfuric acid in the
sample cell and record and print a UV spectrum from 220 – 450 nm. Expand the y-axis
of the recorded spectrum so that the average noise and baseline can be estimated at the
peak maximum wavelength. Be sure to record the brand name associated with your
tonic water unknown and include this information in your lab report.
Set the scanning rate to 100 nm/min and the data pitch to 0.2 nm. Next, measure in
sequence each of your six standard addition solutions, starting with your lowest
concentration of added standard, first. Record the absorbance measured on- and off-
peak-maximum, as before, for each solution.
PART 3: Ka DETERMINATION
Sample Preparation: Standard variable volume plastic-tip micropipettors can be used to
prepare these aqueous solutions. Place 120 L of the bromcresol purple stock indicator
solution in each of 6 glass vials. Add 3840 L of pH 4.0, 5.0, 6.0, 7.0, 8.0, and 9.0 buffer
solutions to each tube, respectively. Next, place 120 L of the phenol red stock indicator

CHEM 4461/5461 Expt. 1: UV-VIS Instrumental Analysis
25
solution in each of 6 glass vials and similarly dilute each with 3840 L of pH 4.0, 5.0, 6.0,
7.0, 8.0, and 9.0 buffer solutions, respectively.
Spectrophotometry
Set the scanning rate to 100 nm/min and the data pitch to 0.2 nm and record a set of
spectra in the visible absorption range from 400 – 700 nm for the 12 solutions. Use a
solution made from mixing 480 L of 0.01 M NaOH with 3520 L of H2O as the
reference solution. For each set of 6 solutions containing a given indicator, overlay the
spectra and print. Also, record specifically the absorbance of the solutions at the peak
maxima below and above the wavelength of the isosbestic point. Also, record the
observed color changes for the indicator solutions as a function of pH.
DATA ANALYSIS
Make plots of absorbance versus quinine concentration (mg/L) for the four sets of data
(i.e. (a) peak maximum, scan rate 100 nm/min.; (b) off-peak maximum, scan rate 100
nm/min.; (c) peak maximum, scan rate 1000 nm/min.; (d) off-peak maximum, scan rate
1000 nm/min.) acquired in the calibration plot linearity study. Examine these plots and
identify the linear ranges of the plots. Perform a linear regression on the linear data
showing a linear response and determine the slopes and intercepts for the four plots.
Also, determine the standard deviations of the slopes and intercepts, and the relative
standard deviations of the slopes.
Using the slope and intercept of the appropriate external standard calibration plots (on-
and off-peak maximum with 100nm/min scate rate and 0.2 nm data pitch), determine
the concentration of your unknown. Also, estimate the error in this value based on the
errors in the individual values used in the calculation. From the average noise and the
standard deviation of the noise, estimate the limit of detection (LOD) of quinine in 0.5%
(v/v) sulfuric acid using the slope of the calibration plot. Using this value, estimate the
dynamic range for UV detection of quinine in 0.5% (v/v) sulfuric acid.
From your standard addition data, calculate, based on your dilutions, the final standard
concentration in each of the six samples you analyzed. Construct a standard addition
plot and determine the concentration of your unknown.
Determine the isosbestic points for the two indicators using the absorption spectra
taken at various pH’s. Also, calculate the Ka’s for the two indicators.

CHEM 4461/5461 Expt. 1: UV-VIS Instrumental Analysis
26
DISCUSSION POINTS
Discuss the changes in the UV spectra of quinine with changes in the scanning speed
and data pitch. What explanation for the observed changes can you offer?
Discuss the four Beer’s law plots obtained. Are deviations from linearity observed and,
if so, what explanations for these deviations can you offer? Which Beer’s law plot is
most appropriate for determining the concentration of the unknown, and why?
Evaluate the method used to determine the LOD and dynamic range for quinine in 0.5%
(v/v) sulfuric acid. Is the approach described appropriate? If not, what better
approaches might be used? If you have performed the fluorescence lab experiment,
how do those method parameters (LOD, dynamic range) compare with those obtained
in this experiment? If you have not yet performed the fluorescence experiment, how do
you think the data will compare using the different spectroscopic techniques (UV vs.
fluorescence)?
Compare the results for your quantitative analysis of quinine in tonic water by external
standard and standard addition. How well do the results agree? Explain any
discrepancies. In order to have obtained a more valid comparison of the quantitative
determination methods, what should have been done? (i.e., What procedure would
need to be followed in order to determine if the results of the two methods are the same
or different?)
Why is an isosbestic point observed in the pH dependent spectra of the two indicators?
How do the observed color changes correlate with the VIS absorption spectra of the
indicators as a function of pH? How do your calculated Ka values for the two indicators
compare with the literature values?

CHEM 4461/5461 Expt. 2: FLUORESCENCE Instrumental Analysis
27
EXPERIMENT 2: FLUORESCENCE SPECTROPHOTOMETRY
First, be sure the instrument is on and the computer is responding. Check with the TA
if either of these items is unclear. Next, using the instructions provided, set the
measurement increasement to 3 nm, the slit bandpass to 1 nm and the integration time
to 0.5 s. Finally, calibrate excitation and emission monochromators before starting your
measurements (p. 31-38 of fluorescence software manual).
PROCEDURE Prepare a stock solution of quinine in 0.5% sulfuric acid at a concentration of ~100
mg/L. Prepare ~10 mL of this solution in a 15-mL centrifuge tube.
Prepare 5 mL of a 10-fold dilution of your stock solution (~10 mg/L).
With the excitation wavelength set at 250 nm, acquire a fluorescence spectrum of the
10 mg/L solution from 350 nm to 600 nm.
Repeat the above experiment with the excitation wavelength set at 350 nm.
Prepare a set of 8 calibration solutions, spanning a range of 0.1 to 10 mg/L in regular
intervals. Prepare 2 – 3 mL of each calibration standard. Develop two calibration
curves (Intensity vs. Concentration), monitoring the emission at 450 nm using two
separate excitation wavelengths: 250 nm and 350 nm, respectively. Record the
response of each solution under each condition three times.
Take 200 L of the tonic water (unknown) and dilute up to 2.0 mL with 0.5% H2SO4.
Monitor the emission intensities for the two excitation wavelengths, as above.
Repeat two more times.
Take the 10 mg/L quinine solution and add varying amounts (0.02 g to 0.10 g) of
NaCl, monitoring the fluorescence intensity at 450 nm (use 350 nm excitation here).
Make at least 5 solutions. Plot I vs. [Cl-].
REPORT Report the structure of quinine.
Discuss the influence of the excitation wavelength on the emission spectrum.
Discuss the relative sensitivity of analysis at the two excitation wavelengths.
Discuss the quenching effect of Cl-.
Report the concentration of quinine in the tonic water and estimate the uncertainty
in this value.
If you have performed the UV-Vis experiment, compare your results from this
experiment, including the various method paramters (linear range, LOD, accuracy,
precision, etc.), with those parameters obtained by UV-Vis. If you have not yet
performed the UV-Vis experiment, conjecture as to how the methods might compare.

CHEM 4461/5461 Expt. 3: TGA Instrumental Analysis
28
EXPERIMENT 3: Thermogravimetric Analysis Procedure TURN ON NITROGEN GAS! NOT DOING THIS CAN DAMAGE THE INSTRUMENT ELECTRONICS! Log onto the computer and open the TA advantage control software. Make sure the TGA icon in the bottom right of the screen is highlighted.
1. Under summary, input the appropriate sample name and comments (e.g. Copper sulfate pentahydrate @ 10 degrees per minute). Specify a data file in the instrumental folder that includes your group #, the sample name and the date of the run.
2. Under procedure, create a name for the current method e.g. 400C 10dpm. Click the editor button and delete any current segments. Double click the ramp segment type and choose a ramp speed and maximum temperature. You will not exceed 500°C on any run.
3. Save your sequence with your group name and the ramp parameters You will need to perform the following sequences to complete the experiment. Prepare and load a fresh sample for each experiment.
1. Ramp to 300°C @ 5dpm using copper sulfate pentahydrate 2. Ramp to 300°C @ 25dpm using copper sulfate pentahydrate 3. Ramp to 500°C @ 10dpm using cobalt chloride hexahydrate
BEFORE LOADING A SAMPLE, CHECK THE INTERNAL TEMPERATURE READOUT IN THE TOP-LEFT OF THE SCREEN AND ENSURE IT IS BELOW 40°C Always exercise care when handling platinum pans as they are very expensive. Using forceps, pick up the pan by the hook and place it on the loading plate (ensure the groove on the bottom of the pan is aligned with the loading tray). Press the tare button on the front of the TGA and wait for the instrument to complete the tare cycle. Remove the pan from the loading tray and place a small amount of analyte across the surface of the pan (try to make a single uniform layer). Return the pan to the loading tray making sure to align the pan with the tray.

CHEM 4461/5461 Expt. 3: TGA Instrumental Analysis
29
Before beginning a run, check the nitrogen supply and ensure gas is flowing; if you are not sure, ask your TA; the furnace generates extreme heat and without cooling gas, the instrument will be completely destroyed. When you are ready to begin your run, click the green play button in the top-left corner of the TGA display screen. As soon as a ramp segment completes for each experimental run, you may begin to your data analysis (i.e., begin the data analysis while the instrument cools).
DO NOT BEGIN ANOTHER SEQUENCE UNTIL THE FURNACE HAS COOLED COMPLETELY. DATA ANALYSIS Using the formula mass for each analyte, determine the weight percent of one molecule of H2O. Using the universal analysis software, open your data file and begin your analysis.
1. For each transition in the thermogram, determine the total and individual changes in the weight percent of your analyte using the signal change function.
2. Determine the onset temperature for each transition using the onset point function.
DISCUSSION POINTS Discuss the difference between waters of hydration and waters of crystallization. Propose a coordination sphere for each cation. Based upon your results, propose an order for the dehydration of your analyte assuming each compound contains both waters of hydration and crystallization. Discuss any variation in the thermograms taken at different ramp speeds, and with different analytes. Your lab report should include the onset point and type of water for each molecule of H2O present in the analyte crystal.

CHEM 4461/5461 Expt. 4: AAS Instrumental Analysis
30
EXPERIMENT 4: FLAME ATOMIC ABSORPTION
SPECTROPHOTOMETRY
PART 1: SAMPLE PREPARATION
All samples must be prepared in ultra-pure water. Make 8 standards solutions (~15 mL each) of
Fe in 0.10 M nitric acid (HNO3) ranging in concentration from 1.0 ppm to 100 ppm (w/w).
Assume the density of the HNO3 solution is the same as that for pure water. Make 8 standards
solutions (~15 mL each) of Cu in 0.10 M nitric acid (HNO3) ranging in concentration from 1.0
ppm to 100 ppm (w/w). These samples will be used for calibration in flame atomic absorption
(FAA) experiments. Prepare two blank solutions (no Fe or Cu) of 0.1 M HNO3. Additionally,
obtain a substantial amount (~ 100 mL) of ultra-pure water for rinsing the flow line between
FAA measurements.
Obtain unknowns for FAA determination. Be sure to note the sample codes in your lab notebook
and in your laboratory report.
PART 2: INSTRUMENT OPTIMIZATION
Refer to the software instruction manual for setting up the method. Cookbook methods are
available for both flame and graphite furnace atomic absorption measurement of different metals
in the instrument software. Make sure to use flame methods only.
PART 3: STANDARD PLOTS
Flame Atomic Absorption
Prepare a method and sequence for analyzing your standards by FAA. Each sample will need to
be sequentially placed into the aspirator as prompted by the software. For the first
determination, use a slit width of 0.1 nm. Absorbance associated with the 8 standard Fe and
solutions and the unknown will be recorded. Be sure to print the appropriate report and record
the data in your lab notebook. Next, adjust the method so that the measurements are made at a
secondary absorption wavelength for Fe. Again measure and record the absorbance associated
with the 8 standard Fe solutions. Repeat the two standard plot determinations, using primary and
secondary wavelengths, for a slit width setting of 1.0 nm. Repeat the above for your Cu samples.
REPORT
Construct calibration curves, perform linear regressions and determine the concentration of the
unknowns supplied for FAA measurements. Perform the analysis for primary and secondary

CHEM 4461/5461 Expt. 4: AAS Instrumental Analysis
31
wavelengths and for different slit widths, as appropriate. Explain your results in terms of LOD
and sensitivity. Do these results suggest an optimal method for determining Fe and Cu
concentrations?
Draw and label a diagram for a flame atomic absorption spectrometer. Discuss the parts of the
flame and the importance of proper alignment of the flame between the lamp and detector. Why
is a hollow cathode lamp (HCL) for each element required in AAS? How does the lamp current
affect the width of the spectral lines from the HCL and why? What are other methods for
measuring trace metals other than FAA?

CHEM 4461/5461 Expt. 4: AAS Instrumental Analysis
32
TABLE
Elements Wave length (nm) Special band pass
(nm)
Optimum working
range (g/mL)
Ag 328.1
338.3
0.5
0.5
1-5
3-12
Al 309.3
396.3
237.3
236.7
257.4
256.8
0.5
0.5
0.5
0.5
0.5
0.5
40-200
50-250
200-800
250-1000
400-1600
650-2600
Cd 228.8
326.1
0.5
0.5
0.5-2.0
250-1000
Fe 248.3
372.0
386.0
392.0
0.2
0.2
0.2
0.2
2.5-10
25-100
50-200
800-3200
Hg 253.7 0.5 100-400
K 766.5
769.5
404.4
1.0
1.0
0.5
0.5-20
0.5-60
200-800
Mg 285.2
202.6
0.5
1.0
0.1-0.4
5-20
Pb 217.0
283.3
261.4
202.2
205.3
1.0
0.5
0.5
0.5
0.5
5-20
10-40
200-800
250-1000
2000-8000

CHEM 4461/5461 Expt. 5: GC Instrumental Analysis
33
EXPERIMENT 5: Gas Chromatography
PART 1: OPTIMIZATION OF CHROMATOGRAPHY
First, make sure the instrument in on, the carrier gas is flowing, the flame ionization
detector (FID) is lit, and the computer is responding. Record the dimensions (L, i.d.), the
make and model, and the type and thickness of the stationary phase for the column
installed in the chromatograph. Check with the TA if any of these items are unclear.
Next, using the instructions provided, set the flow rate of the column to 0.80 mL/min,
the split ratio to 30:1, and the injector and detector temperatures to 250 °C.
Sample Preparation
In an autosampler vial, deposit 20 μL of carbon tetrachloride (CCl4) using the
micropipette. In a second autosampler vial, deposit 20 μL of chloroform (CHCl3). In a
third autosampler vial, deposit 20 μL of CCl4 and 20 μL of CHCl3. Finally add sufficient
volume of dichloromethane (CH2Cl2) to each tube to bring the total volumes to 1.000 mL
and cap each vial securely.
Determining Dead Time (t0)
Each student will make five manual injections of butane to determine the dead time and
to determine the reproducibility of each individual’s injection technique. Enter a
temperature program for the column, in which the oven temperature is held constant at
50 °C. This is called an isothermal separation method. Insert the syringe needle into the
outlet of a butane lighter and press the gas release button on the lighter. Do not strike
the flint. Withdraw the plunger on the syringe to between 2 and 4 μL. Prior to
injection, keep the syringe needle pointing down to avoid loss of butane from syringe.
Inject the solution and record a chromatogram. A typical manual injection in split
injection mode should take approximately 2.5 seconds (1 second to insert the needle, 0.5
seconds to inject, and 1 second to withdraw the needle). Be sure to be consistent in
timing the start of the run with the manual injection (insert syringe inject
withdraw syringe start run). Record the retention time of the peak for each repetition
and calculate the average, the standard deviation, and the relative standard deviation
for each individual’s five replicates, as well as for the entire group. This is the dead
time for the column under this method.
Effect of Changes in Temperature Program
Enter a temperature program for the column, in which the oven temperature is held
constant at 30 °C. Inject a 0.5 μL sample of the mixed solution containing both CCl4
and CHCl3 and record the chromatogram. Print the chromatogram and the peak

CHEM 4461/5461 Expt. 5: GC Instrumental Analysis
34
analysis report for your records. Make sure all printed chromatograms are adequately
labeled with sample and method information for future reference. Also, expand the
chromatogram on the computer and record the width at half-height (Wh) for each peak
(Note: width at base (W) may also be recorded, if necessary and if peaks are adequately
resolved).
Enter an isothermal temperature program where the oven is set to 50 °C. Inject a 0.5 μL
sample of the mixed solution containing both CCl4 and CHCl3 and record the
chromatogram. Print the chromatogram and the peak analysis report, and record Wh
for each peak.
Enter an isothermal temperature program where the oven is set to 90 °C. Inject a 0.5 μL
sample of the mixed solution containing both CCl4 and CHCl3 and record the
chromatogram. Print the chromatogram and the peak analysis report, and record Wh
for each peak.
Enter a temperature program in which the oven temperature begins at 50 °C and
increases to 90 °C at a rate of 5 °C/min. This is called a temperature gradient separation
program. Inject a 0.5 μL sample of the mixed solution containing both CCl4 and CHCl3
and record the chromatogram. Print the chromatogram and the peak analysis report,
and record Wh for each peak.
Finally, record chromatograms of 0.5 μL injections of each of the two pure compounds
dissolved in CH2Cl2 using the temperature program above. Print the chromatogram and
the peak analysis report for each run, and record Wh for each peak.
Effect of Changes in Flow Rate
Adjust the column flow of the carrier gas to 1.50 mL/min. Set the oven temperature for
an isothermal run at 50 °C, make a 0.5 μL injection of the mixed solution, and record the
chromatogram. Print the chromatogram and the peak analysis report, and record Wh
for each peak. Adjust the column flow of the carrier gas to 3.00 mL/min and repeat the
experiment. Print the chromatogram and the peak analysis report, and record Wh for
each peak.
PART 2: STANDARD CURVE
Sample Preparations
In four (4) autosampler vials, deposit 30, 40, 50, and 60 μL of CCl4, respectively, using
the micropipette. Next, add 20 μL of CHCl3 to each tube. Finally, add sufficient volume

CHEM 4461/5461 Expt. 5: GC Instrumental Analysis
35
of CH2Cl2 to each vial to bring the total volume to 1.000 mL and cap each vial securely.
These are your calibration standard samples. Prepare a sample “blank”, by filling an
autosampler vial with 1 mL of CH2Cl2. In an autosampler vial, add 20 μL of CHCl3 to
500 μL of your unknown. Add sufficient CH2Cl2 to bring the total volume to 1.000 mL
and cap securely. This is your unknown sample for quantitative determination.
Chromatography
Place the sample vials in the autosampler and create a sequence in the software for the
analysis of your samples, using the same temperature program as previously and a
carrier gas flow rate of 0.8 mL/min. Program the autosampler to sequentially inject 0.5
μL of each of the solutions (calibration standards, blank, unknown). Print the
chromatograms and the peak analysis reports for your records.
PART 3: KOVAT’S RETENTION INDICES
Using the isothermal method with column temperature of 50 °C and flow rate of 0.8
mL/min, record the chromatogram for 0.1 μL injection of a neat mixture of the C5 – C8
hydrocarbons. Print the chromatogram and the peak analysis report for your records.
DATA ANALYSIS
Part 1
Calculate the resolution (R) for the separation of CHCl3 and CCl4 in the 4 chromatograms
recorded using the various oven temperature settings (isothermal at 30, 50, and 90 °C;
and gradient temperature programming). Also, calculate the capacity factors (k’) for
CHCl3 and CCl4 and the selectivity factor (α) resulting from each method. Assume t0 for
each of the methods is equivalent to that which was determined initially from the
butane injections at 50 °C. How do the calculated values (R, k’, α ) change for
separations using the different oven temperature methods?
Calculate R between CHCl3 and CCl4 at the 1.5 mL/min and the 3.0 mL/min flow rates.
Calculate the number of theoretical plates (N) and the height equivalent to a theoretical plate
(HETP) for CHCl3 and CCl4 at each of the 3 flow rates (0.8, 1.5, and 3.0 mL/min). Using
the HETP values, calculate the A, B, and C constants in the Van Deemter equation for
each of the two compounds (Note: You will need to convert the flow rates to linear
velocities; divide by the area of the cross section of the open portion of the column).
Finally, take the first derivative of the Van Deemter equation and estimate the optimum
column flow rate.

CHEM 4461/5461 Expt. 5: GC Instrumental Analysis
36
Part 2
Using the chromatograms obtained from your standard solutions perform a linear
regression of the plot of area of the CCl4 peak versus percent CCl4 (external standard
method). Determine the slope and the intercept of the best-fit line to the data. Also,
determine the standard deviation of the slope and the intercept. Next, for each of the
five chromatograms of the standard samples, divide the area of the CCl4 peak by the
area of the CHCl3 peak and make a second plot of the these values versus percent CCl4
(internal standard method). Determine the slope and the intercept of the best-fit line to
the data and determine the standard deviation of the slope and the intercept. How does
the relative standard deviation of the slope of the second plot compare with the relative
standard deviation of the slope of the first plot?
Determine the concentration of the unknown by the external standard method (using
the regression equation from the first plot, determine the concentration of unknown
from the measured area) and the internal standard method (divide the area of the CCl4
peak by the area of CHCl3 from your unknown chromatogram and, using the regression
equation from the second plot, determine the concentration of unknown from the
normalized area). Estimate the error in these values from the errors in the calibration
plot. How do the absolute values determined and their uncertainty compare? From the
blank injection, estimate the average background signal and standard deviation of this
value. From the slope of the first calibration plot, estimate the limit of detection (LOD)
and the limit of quantification (LOQ) for the CCl4.
Part 3
Make a plot of the log of the adjusted retention times for the C5 to C8 hydrocarbons
versus the number of parafinnic carbon atoms x 100 (e.g. for C5, this equals 500 and so
on). Perform a linear regression and determine the slope and the intercept of the best-fit
line. Use this plot and the adjusted retention times for CHCl3 and CCl4, calculate the
Kovat’s retention index for each compound.
DISCUSSION POINTS
In your laboratory report, discuss the reasons for using temperature programming in
gas chromatography rather than simple isothermal separations. Relate this discussion
to the specific examples in this experiment.

CHEM 4461/5461 Expt. 5: GC Instrumental Analysis
37
How reproducible is your injection technique versus that of the group? Why is it
important to make the injection quickly and to be consistent in your technique? Would
expect the dead time to increase or decrease with an increase in oven temperature?
Discuss the Van Deemter equation and the implications of the three characteristic
constants A, B, and C. Which of these constants are particularly relevant to gas
chromatography and why?
Compare and contrast the two calibration plots. Why is the second method of
preparing a calibration plot using an internal standard considered superior in gas
chromatography? Is this borne out in your results? If you had performed the unknown
determination using manual injection, instead of an autosampler, would you expect any
substantial differences in your determinations? In determining the LOD and LOQ,
what assumptions have you made? Are these reasonable for gas chromatography with
an FID detector?
What is the purpose of determining the Kovat’s retention indices of the 2 compounds?
How can this information and method by useful for designing additional gas
chromatographic separation methods and predicting the outcome?

CHEM 4461/5461 Expt. 6: HPLC Instrumental Analysis
38
EXPERIMENT 6: High Performance Liquid Chromatography
PART 1: OPTIMIZATION OF CHROMATOGRAPHY
First, make sure the instrument in on, the UV lamp is on, and the computer is
responding. Make sure the 2 solvent reservoirs are filled with HPLC-grade methanol
and HPLC-grade water, respectively, and that each bottle is being gently sparged by
helium to remove dissolved gases in the solvent. Record the dimensions (L, i.d.), the
make and model, the type of the stationary phase, and the type of packing for the
column installed in the chromatograph. Check with the TA if any of these items are
unclear.
Sample Preparation
The stock solution contains methyl paraben, ethyl paraben, propyl paraben, and butyl
paraben, each at a concentration of 200 ppm (v/v), in 50/50 methanol:water. Make serial
dilutions of this mixture using 50/50 methanol:water to create a series of calibration
standard mixtures at 150 ppm, 100 ppm, 75 ppm, 50 ppm, and 10 ppm concentrations.
Sample volumes of 1.000 mL are sufficient. All solutions should be filtered using a
syringe filter prior to use.
Effect of Mobile Phase Composition
Using the instructions provided, set the flow rate of the column to 1.00 mL/min, the
solvent ratio of the mobile phase to 99% methanol / 1% water, and turn the pump on.
Set the analysis time to 8 minutes and the detector wavelength to 254 nm. Check all
tubing fittings for leaks. Allow at least 5 column volumes of solvent to pass through the
column before beginning the analysis. Prior to analysis, note the pressure of the system
and make sure it is stable prior to beginning the run.
Fill your syringe with the 200 ppm mixture to make a full-loop injection (fill the
injection needle with approximately 1 – 2 μL more sample than the volume of the loop).
Make sure the injector is in the “load” position, insert the syringe and fill the loop.
Leave the empty syringe in the injector as you switch the injector to the “inject” position
and start the run. You can remove the syringe after the injection is made and the run is
started. Print the chromatogram and the peak analysis report for your records. Make
sure all printed chromatograms are adequately labeled with sample and method
information for future reference.
Repeat this analysis (using the 200 ppm mixture), sequentially, for solvent compositions
of 90/10, 80/20, 70/30, and 60/40 methanol:water. The separation under each of these

CHEM 4461/5461 Expt. 6: HPLC Instrumental Analysis
39
conditions is called an isocratic separation method, because the solvent composition
does not change throughout the analytical run. Remember to allow sufficient time for
equilibration at each new mobile phase (approximately 5 column volumes) and record
the pressure prior to each run. Print the chromatograms and the peak analysis reports
for each run. Also, expand the chromatogram on the computer and record the width at
half-height (Wh) for each peak, if possible (Note: width at base (W) may also be recorded,
if necessary and if peaks are adequately resolved). (Note: You may need to increase the
run time for methods which contain a higher proportion of water in the mobile phase.)
PART 2: STANDARD CURVE
Chromatography
Pick an isocratic separation program that provides adequate retention (k’ (peak 1) > 1)
and resolution (R > 1.5) of all three sample analytes. This can be, but does not have to
be, one of the methods investigated above. Inject each of your calibration sample
mixtures (full-loop injection) under this method, starting with the lowest concentration
(10 ppm) first. Record the retention times and areas for each peak in each calibration
sample mixture analyzed. Inject your unknowns (unknown 1 and unknown 2) under
the same conditions and record the chromatograms, also noting the retention times and
areas of the peaks in your samples. Note the sample codes for your unknowns and
indicate these in your report.
PART 3: SOLVENT GRADIENT SEPARATION
Set up the instrument to perform a gradient solvent program separation method. Start
with an initial mobile phase composition of 50/50 methanol:water, a hold at that
composition for 1 min, an increase to 99/1 methanol:water in 6 minutes, and a 1 minute
hold at the final composition, as follows (between 1.0 min and 7.0 min, the instrument
will change the solvent composition, in a linear fashion, from 50/50 to 99/1
methanol:water):
Table 1: Initial solvent gradient program
Run time (min) methanol (%) water (%)
0.0 50 50
1.0 50 50
7.0 99 1
8.0 99 1

CHEM 4461/5461 Expt. 6: HPLC Instrumental Analysis
40
Inject the 200 ppm mixture and record a chromatogram. Calculate k’ for each peak, and
the resolution between each component.
Next, modify the solvent gradient to create a method for base line separation (R ≥ 1.5) of
all three components in under 5 minutes. Note, that k’ (peak 1) must be greater than 1
for the optimized method. You may change any parameter in Table 1 above.
Remember to allow sufficient time for equilibration prior to each analytical run.
DATA ANALYSIS
Part 1
Calculate the capacity factor (k’) and number of theoretical plates (N) for each analyte
under each of the different mobile phase compositions tested. Also, calculate resolution
(R) and selectivity (α) for each pair of analyte peaks in each chromatogram. How do the
calculated values change with the different instrumental settings? What is the dead
time (t0) for your experimental set-up and how was this determined?
Part 2
Using the chromatograms obtained from your standard solutions perform a linear
regression of the plot of area of each peak versus concentration (external standard method)
for each analyte. You will have three separate calibration curves. Determine the slope
and the intercept of the best-fit lines to the data for each plot. Also, determine the
standard deviation of the slope and the intercept for each plot. Determine the
concentration of each component in your unknown mixtures and estimate the error in
each of these values from the error in each calibration plot.
Part 3
Record the solvent program for each method you try, as well as k’ (peak 1), k’ (peak 3),
and R (each pair of analyte peaks) for each trial.
DISCUSSION POINTS
In your laboratory report, discuss the reasons for using gradient solvent programming
in HPLC rather than simple isocratic separations. Relate this discussion to the specific
examples in this experiment.
What is the mechanism of a reverse phase HPLC separation? Which mobile phase
component is the “strong” component and which is the “weak” component? Why are

CHEM 4461/5461 Expt. 6: HPLC Instrumental Analysis
41
these designated as such? What other types of solvents are typically employed in
reverse phase HPLC?
Discuss the effect of mobile phase composition on the separation of your analytes in
terms of retention time, capacity factor, selectivity, and resolution. Also, discuss the
change in pressure observed for each isocratic separation method. Why is it important
that the column is “equilibrated” prior to performing each analysis? How is the
pressure an indication of equilibration?
Present the structures of the analytes and logically reason the elution order of the
compounds based on the reverse phase HPLC separation methods you have employed.
What is the identity and concentration of each analyte in your unknown mixtures?
Why is a wavelength of 254 nm set for the detector? What is the “chromophore” in
your analytes that allows them to be detected at this wavelength?
For your gradient separation method, why is it important to have k’ (peak 1) > 1 and R ≥
1.5?
Tabulate all of your data for effective presentation.

CHEM 4461/5461 Expt. 7: NMR Demo Instrumental Analysis
42
EXPERIMENT 7: Nuclear Magnetic Resonance Spectroscopy
Demonstration
Address the following points when preparing your laboratory report for this
demonstration of nuclear magnetic resonance (NMR) spectroscopy. You may need to
independently search for some of this information in your textbook, in the literature,
and in prominent websites.
Briefly describe the basic measurement that is being made in proton NMR
spectroscopy.
What is a superconducting magnet? Why is it necessary/desirable to use a
superconducting magnet in high magnetic field NMR spectroscopy? What are
the dangers associated with these instruments?
Why is it necessary to spin the sample in solution phase proton NMR
spectroscopy?
Using the magnetogyric constant for a proton, calculate the approximate
magnetic field associated with a 60 MHz instrument. Do the same for a 500 MHz
instrument.
How do the energies associated with the splitting of the nuclear spin states in
NMR compare with the energies associated with molecular rotation?
Currently there are efforts underway to create a 24 T magnet. What would be
the frequency of splitting for a proton in an instrument with this magnetic field
strength?
The x-axis of both the 60 MHz and 500 MHz NMR spectra is the same, but the
resolution of the peaks in the 500 MHz instrument is much better. What factors
contribute to the improved resolution in the 500 MHz instrument?
What is the function of the “lock solvent” in high field NMR spectroscopy?

CHEM 4461/5461 Expt. 8: MS Demo Instrumental Analysis
43
EXPERIMENT 8: Mass Spectrometry Demonstration
Address the following points when preparing your laboratory report for this
demonstration of mass spectrometry and the use of specific ion sources. You may need
to independently search for some of this information in your textbook, in the literature,
and in prominent websites (e.g. www.asms.org).
PART 1: MASS SPECTROMETRY, GENERAL ASPECTS
What are the general instrumental components which compose a mass
spectrometer?
Why is a high vacuum necessary in mass spectrometer instruments?
What is the difference between high resolution and low resolution mass
analyzers? Provide examples.
PART 2: ELECTROSPRAY IONIZATION – MASS SPECTROMETRY
Briefly describe the measurement which was made in the demonstration of ESI-
MS. Include aspects of instrumentation, sample components, and the purpose of
the experiment.
Describe the electrospray process which leads to the formation of gas phase ions.
What is the molecular form of ions created in positive mode and in negative
mode ionization?
What kind of mass spectrometer/mass analyzer was employed and what is the
basic theory behind its operation?
Briefly describe the applicability of ESI-MS to different sample types. This
information can also be effectively communicated by citing specific examples
from the literature.
How is ESI-MS capable of analyzing large molecules with masses higher than the
upper mass-to-charge limit of common mass analyzers?
Why is ESI referred to as a soft ionization source?

CHEM 4461/5461 Expt. 8: MS Demo Instrumental Analysis
44
PART 3: MATRIX-ASSISTED LASER DESORPTION/IONIZATION - MS
Briefly describe the measurement which was made in the demonstration of
MALDI-MS. Include aspects of instrumentation, sample components, and the
purpose of the experiment.
What is the purpose of the matrix? What processes occur to result in the
formation of ions from the analytes of interest?
Why is MALDI ideally suited to time-of-flight mass analysis detection? What are
the analytical advantages of a TOF mass analyzer?
Briefly describe the applicability of MALDI-MS to different sample types.
What is post-source decay and how is it useful?
Is MALDI a soft or hard ionization technique and why? In this respect, how does
it compare to laser desorption/ionization with no matrix?