    Measure low-level total organic carbon (TOC) anywhere,anytime with the new CheckPoint* Pharma andCheckPoint e On-Line/Portable Total Organic Carbon(TOC) Sensors from GE Analytical Instruments.Affordable, versatile, and ultra-portable, the CheckPointSensors can be used online for continuous monitoringor hand-carried to any point in a water system for rapiddiagnostic sampling and troubleshooting.The two CheckPoint models have the same specificationsexcept as follows:

    The CheckPoint Pharma has a dynamic range of 0.21to 1,000 ppb and can measure TOC in hot or ozonat-ed water. It meets US, European, Chinese, and IndianPharmacopeia specifications.

    The CheckPointe, which is designed to meet thegreater sensitivity requirements of the semiconductor,power, and other markets, has a dynamic range of 0.05 to 1,000 ppb.

    Key Benefits

    Cost Effective The CheckPoint Sensors use the simpleTOC methodology of measuring initial sample conductiv-ity, UV oxidation of organics, and a final, post-UV con-ductivity measurement (Direct Conductometric). Theyprovide a reliable low-cost TOC approach useful forpharmaceutical Purified Water (PW) and Water forInjection (WFI) and semiconductor ultrapure water (UPW)monitoring applications .

    Rapid Analysis The two Sensors provide the rapidresults and fast rinsedown required for time-critical

    diagnostics and troubleshooting. Featuring a defaultmeasurement every 15 seconds, users can adjust out-put intervals from 15 seconds up to eight hours.

    Versatility and Convenience The Sensors can be usedfor continuous on-line monitoring, rapid on-line monitor-ing, or grab sampling. They also feature advanced digitalcommunication capabilities, including Ethernet (ModbusTCP/IP), and a USB port for easy data download.

    Easy to Use with Low Maintenance The reagentlessCheckPoints are easy to operate and need minimal main-tenance. Calibration is typically stable for six months.

    Sensor-to-Sensor Matching The Sensors can becalibrated to a reference TOC instrument, allowing

    excellent low-level TOC sensor-to-sensor matching.

    Ultra Portability The CheckPoint TOC Sensors haveredefined portability and flexibility. The Sensors typical-ly operate four to f ive hours using the optional battery, or

    continuously with AC power. Weighing just 3.6 kg (7.9 lb)with battery, the CheckPoint Sensors give you the free-dom to sample anywhere.

    ApplicationsPharmaceutical The CheckPoint Pharma provides

    pharmaceutical manufacturers with a cost-effectiveand flexible process monitor for process trending,screening potential trouble areas, and diagnosingproblems in real time. The CheckPoint Pharma can beplaced at a dedicated point of use or convenientlymoved throughout the pharmaceutical manufacturingfacility for multi-point water monitoring, diagnostics,and troubleshooting. The CheckPoint Pharma comeswith a simple and convenient Standard OperatingProcedure (SOP) to meet installation, operation, andperformance qualification requirements.Hard Disk, Thin Display, and Semiconductor TheCheckPointes low cost and portability gives microelec-tronics manufacturers new problem-solving anddiagnostic capabilities. In addition to providing contin-uous on-line TOC monitoring to detect critical ultrapurewater changes, the CheckPoint e Sensor makes itpossible to quickly check TOC samples from pressurizedor non-pressurized sources. CheckPoint e can alsomonitor UP W distribution points or fab tools forpotential contamination sources.

    Power (UPW, Cation Conductivity Control, Makeup, orCogeneration Condensate Polish Control) TheCheckPointe provides sensitive detection of traceorganic contamination in power feed and loop UPW.Controlling cation conductivity can be difficult if its

    CheckPoint Highlights Enables on-line and off-line testing in one instrument

    Display screen and documentation available inEnglish, Chinese, or Japanese

    Self-contained pump for easy system suitability andcalibration testing

    Ethernet Modbus transmission to facilitate data system

    IQ/OQ/PQ validation protocols provided

    User-friendly TOC graph that indicates trends

    Stores 28,800 records with TOC display data accessi-ble using USB memory stick

    Five-hour battery available for maximum portability aswell as 100240 volt operation for installed use

  • 8/10/2019 Check Point Pharma and Check Point-e (Br 3 169 F Cp en)

    3/4

    System Specifications 1Total Organic Carbon

    Linear Range (CheckPoint Pharma) 0.211,000 ppb CLinear Range (CheckPointe) 0.051,000 ppb CAccuracy 5% at 500 ppb C of sucrose 2,3Precision (CheckPoint Pharma) The greater of

  • 8/10/2019 Check Point Pharma and Check Point-e (Br 3 169 F Cp en)

    4/4

    CheckPoint Installation Diagrams

    Wall Mounting Bracket Dimensions

    Vial Sampling Kit

    CheckPoint Pharma and Vial Sampling Kit Mounting Dimensions

    Ordering InformationPharmaceutical Industry Part Number

    Checkpoint TOC Sensor basic model CPRD 97150Checkpoint with I/O board CPRD 97160

    Electronics/Semiconductor IndustryCheckpointe TOC Sensor basic model CPRD 97000Checkpointe with I/O board CPRD 97100

    OptionsVial Sampling Kit CAPK 97900Checkpoint In-line Filter Kit (60 m) CAPK 97300Printer CHMI 9705040-mL Certified Sample Vials (Case of 72) CHMI 90606

    ConsumablesOne-year kit (2 UV lamps and 2 pump heads) CAPK 97010One UV lamp (6-month continuous use) CARK 35001One pump head (6-month continuous use) CAPK 97000

    MountingBracket

    Options and AccessoriesBattery Pack Kit The battery and charger system typicallyenable four to five hours of operation. Extra batteries are alsoavailable.

    I/O Board The optional I/O board has three programmableanalog outputs (select from TOC, raw conductivity, temperature-compensated conductivity or resistivity, error, warning, orstandby options), one binary input (remote start/stop), and twoalarms with 24 V supply.

    Vial Sampling Kit The CheckPoint Vial Sampling Kit, whichattaches easily to the Sensor, allows users to measure stan-dards and grab samples.

    Low-Pressure Sampling Kit This includes Teflon and stainlesssteel sampling tubes and a waste bag for collecting zero-pressure samples in the laboratory or fab.

    Sample Inlet Filter The 60- m sample inlet filter is recom-mended for on-line monitoring.

    300 00169 Rev F

    MC12-060

    I S O 9001:2008

    QUALITYMANAGEMENT

    SYSTEM

    The AmericasGE Analytical Instruments6060 Spine RoadBoulder, CO 80301-3687 USAT +1 800 255 6964T +1 303 444 2009F +1 303 527 [email protected]

    Europe/Middle East/AfricaGE Analytical InstrumentsUnit 3, Mercury WayUrmston, ManchesterUK M41 7LYT +44 (0) 161 864 6800F +44 (0) 161 864 [email protected]

    [email protected]@[email protected]

    For specific country contacts, visitwww.geinstruments.com/sales

    www.geinstruments.com 2012 , General Electric Company. All rights reserved.