CHARMVAL PROJECT FINAL REPORT Introduction The selection of chemistries for study in this project is...
Transcript of CHARMVAL PROJECT FINAL REPORT Introduction The selection of chemistries for study in this project is...
CHARMVAL PROJECTFINAL REPORT
September 2001 - October 2003
Simona GagliardiSteve Grigson
October 2003
Contents
Contents 1
I CHEMICAL SELECTION 3
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2 Chemical Usage Offshore . . . . . . . . . . . . . . . . . 6
3 Regulator and Industry Views . . . . . . . . . . . . . . 10
4 Surfactant Fraction Released Default Values in
the CHARM Model . . . . . . . . . . . . . . . . . . . . . . 17
5 Classification of Surfactants . . . . . . . . . . . . . . . 20
6 Chemistries Suggested . . . . . . . . . . . . . . . . . . . . 29
7 Conclusions and Recommendations . . . . . . . . . . . . 30
II EXPERIMENTAL 32
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . 33
2 Imidazoline Synthesis . . . . . . . . . . . . . . . . . . . . 35
3 Quats Synthesis . . . . . . . . . . . . . . . . . . . . . . . . 44
4 Analytical Techniques . . . . . . . . . . . . . . . . . . . 53
1
4.1 DCM extraction . . . . . . . . . . . . . . . . . . . 55
4.2 SPE extraction . . . . . . . . . . . . . . . . . . . . 58
4.3 HPLC-MS . . . . . . . . . . . . . . . . . . . . . . . . 72
5 Produced Water Analysis . . . . . . . . . . . . . . . . . 73
5.1 Calibration Curves . . . . . . . . . . . . . . . . . . 73
5.2 Preparations . . . . . . . . . . . . . . . . . . . . . . 77
5.3 Quats . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
5.4 Imidazolines . . . . . . . . . . . . . . . . . . . . . . . 93
6 Results Summary . . . . . . . . . . . . . . . . . . . . . . . 101
References 104
2
1 Introduction
The selection of chemistries for study in this project is proving more difficult
than anticipated, I believe for two reasons: confidentiality and the complexity
and diverse range of chemistries used. Partially for this reason I have broad-
ened the content of this report to include information on: current chemical
usage in the North Sea; the types of chemicals used; their chemistry; their
mode of action; and the application of the CHARM model. Background
chemical data has been obtained from the literature, the chemical companies
and the Internet.
It is intended that this document will form the basis for the final selection of
chemicals at the next steering group meeting.
Reasons for the Chemical Selection Exercise
The CHARM model uses an estimate of the produced water discharge con-
centrations of oilfield chemical residues to calculate the Hazard Quotient for
an oilfield production chemical. An integral parameter for some of these
calculations is the value estimated for the chemical’s octanol-water partition
coefficient (this is explained in detail in Section 4). For a number of chem-
icals this can be determined adequately using a shake-flask procedure [? ]
or an HPLC method [? ]. Many of the more hazardous oilfield chemicals
are believed to owe at least some of their properties to surfactant chemistries
contained in their formulations. However, surfactant chemicals do not (the-
oretically) give reliable Pow values and, subsequently, a number of default
values for a limited range of generic surfactant chemistries are specified for
4
use in the CHARM model. However, the chemical industry argue that the
limited range of default values do not cover the diverse range of chemistries
used and, further, no field validation data for these values are available. In
this initial stage of the project, all of the sponsors were contacted, and in
most cases visited, to obtain their views on which chemistries should be the
subject of study and field validation. Where possible the views given have
been presented in such a way as to maintain anonymity. However, I felt that
it would be useful to differentiate the views of the different parties.
Visits and Contacts with Sponsors
Following (and indeed prior to) the first steering group meeting, I visited or
contacted the following organisations:
Baker Petrolite 20/09/01 Visit
CEFAS 25/09/01 Visit
Shell 18/10/01 Visit
Statoil Norway/Dynea 31/10/01 Visit
Enterprise Oil 06/11/01 Visit
DTI, Aberdeen 06/11/01 Visit
FRS 09/11/01 Visit
T R Oil Services 09/11/01 Visit
Huntsman Surface Sciences Telephone request for info
Ondeo Nalco Telephone request for info
Texaco Email request for info
5
The objectives of the visits were as follows:
1. To see which chemicals were being used offshore and in what quantities.
2. To see which chemicals fell into the more hazardous OCNS categories.
3. To see which chemistries were of concern to the regulator (DTI) and
their advisers (CEFAS and FRS).
4. To find out the oil company’s views on chemical selection and substi-
tution, from an environmental/operational standpoint.
5. To ask the chemical companies whether their products’ chemistries
could be put into generic groups and, if yes, what these should be,
in particular those products with surfactant chemistries, for example
demulsifiers.
6. To ascertain how discharge concentrations of oilfield chemicals are cur-
rently calculated.
In addition, the MSDS data sheets for chemicals used offshore on platforms
selected by the sponsors were studied to see what information could be gained
for use in the chemical selection process.
2 Chemical Usage Offshore
The production chemical usage on 9 offshore platforms over periods ranging
from 6 months to one year during 2000/2001, were studied (data on two more
platforms have been received). Their chemical usage by category is shown
6
Table 1: Production chemical usage on seven production platforms.
Chemical Typical OCNSCategory
Platform No.
1 2 3 4 5 6 7 8 9
Anti-foaming A-E x x
Biocide B-E x x x x x x x x x
Deoiler C-E x x x
Corrosion inhibitor B/C x x x x x x x x
Defoamer C x x x x
Demulsifier B-D x x x x x x x x x
Gas treatment E x x x x x x x x x
Oxygen scavenger E x x x x x x x
Scale inhibitor E x x x x x x x x
in Table 1. As can be seen from the table, biocides, corrosion inhibitors,
demulsifiers, gas treatment and scale inhibiting chemicals are almost univer-
sally used on these platforms.
Chemicals were selected from the more hazardous B and C groupings and
tabulated (there is a voluntary agreement amongst operators that category A
chemicals, the most hazardous category, will not be used. However, Govern-
ment has always stated that the Risks involved in both the use and non-use
of these should be weighed up before these decisions are made). Chem-
icals that fell within these groups were corrosion inhibitors, demulsifiers,
defoamers, deoilers and combined corrosion/scale inhibitors (also some bio-
cides). In addition, the amounts of scale inhibitors used were also tabulated
for comparison (quantity wise) to the other groups. It is acknowledged that
7
Table 2: Chemical usage and discharge amounts per annum in kg (average amountper platform). Only category B/C chemicals shown plus scale inhibitors.
ChemicalType
OCNSCategory
Used Discharged Fractionreleased
Number ofplatformsconsidered
CorrosionInhibitor
B/C 79939 64354 0.81 7
Demulsifier B/C 66245 10943 0.17 9
Defoamers C 5650 0 0 3
Deoilers C 36360 36360 1.0 3
CI/SI B 65303 62038 0.95 1
ScaleInhibitor
E 303493 275142 0.91 6
scale inhibitors, although regarded as environmentally benign, are among
those used and discharged in the greatest quantities, relative to other oilfield
chemicals. The data is summarised in Table 21.
Table 3 shows a comparison between the tonnage of three types of oilfield
chemicals used and discharged in 1989 and 2000/01. The amounts of demul-
sifiers and scale inhibitors used has increased by an order of magnitude,
corrosion inhibitor usage has doubled. These figures should be treated with
caution, however, as they are estimates only.
The total amount of chemicals discharged in 2000/2001 is estimated at 53,000
tonnes/annum (based on total discharge data provided by the operators of
1A Safety Factor (0.1) is added to the chemical’s calculated produced water fractionreleased value when using the CHARM model. This is to account for chemical associatedwith oil droplets and silt particles present in produced water and the assumption that astate of equilibrium exists between the concentrations in the oil and water phases in thetopside process system.
8
Table 3: Comparison of the total chemical usage by category in the UK Sector ofthe North Sea, between 1989 and 2000/01. The data is based on information inTable 2 and the paper by ? ]. Installations refer to the number of installationshaving permits to discharge produced water.
Chemical Type Used(tonnes/year)
Discharged(tonnes/year)
Installations
1989
Corrosion Inhibitors 2,500 200 36
Demulsifiers 450 10 36
Scale Inhibitors 1,700 1,150 36
2000/01
Corrosion Inhibitors 5,436 4,376 68
Demulsifiers 4,505 744 68
Scale Inhibitors 20,638 18,710 68
all chemicals for three platforms multiplied by 68/3). Roughly 10% are cor-
rosion inhibitors and demulsifiers. Again, these figures are estimates only.
Table 4 shows a comparison between 1989 and 2000 of total oil, water and
chemical discharges. The theoretical concentrations of corrosion inhibitors,
scale inhibitors and demulsifiers in discharged produced water (assuming for
the purposes of the table use equals discharge) is also shown, for interest.
The data shows that although the number of installations permitted to dis-
charge oil has almost doubled, the total oil discharged has only increased by
63%, reflecting a continuing reduction in the level of oil in produced water.
In contrast, the produced water volumes have increased by some 230%, as
fields mature. Oilfield chemical discharges have also significantly increased
in the UK sector. However, the comparison between 1989 and 2000 should
9
Table 4: A comparison of oil, water and chemical discharges in the UK Sector of theNorth Sea in 1989 and 2000. Oilfield chemical levels in produced water estimatedfrom Table 2 (tonnes discharged) and Table 3 (produced water discharge amounts)(*? ]).
1989 2000
Number of installations permitted to discharge oil 36 68
Total oil discharged (tonnes) 3423 5395
Total water discharged (million tonnes) 106 244
Average oil in water content mg/kg (ppm) 32 21.5
Total production chemicals discharged (tonnes) 6,000* 53,000
Average corrosion inhibitors (ppm) in produced water - 18
Average demulsifiers (ppm) in produced water - 3
Average scale inhibitors (ppm) in produced water - 77
be taken as a very rough guide only.
For Statoil operated fields in the Norwegian sector of the North Sea, produc-
tion chemical usage has also increased steadily, from 5,000 tonnes in 1991 to
30,000 tonnes in 2000 (Aas et al, 2002).
3 Regulator and Industry Views
You may wish to read Section 4 first if you are less familiar with the cal-
culation of octanol water partition coefficients and the issues surrounding
surfactants.
• Chemistries of concern to the regulator and their scientific
advisers, and comments on chemical selection
Phylosophy
10
The regulator and their scientific advisers take a pragmatic view
of chemical regulation. For example, in the UK they see no rea-
son to ban aromatic solvents as they will partition primarily into
the crude oil and any contribution to the produced water will be
masked by water-soluble crude oil aromatics. Similarly, although
the Norwegians have concerns about the use of aromatic solvents,
this is primarily on health and safety grounds. Another exam-
ple is phenol formaldehyde polymers. These do not biodegrade
and therefore are identified by OSPARs Harmonised Pre-screening
scheme as being candidates for substitution. However, the UK
view is that this should only be enforced when a suitable alter-
native with the desired efficacy is available. These substances are
banned in Norway. In the UK a risk-based approach is favoured,
it being up to companies to justify why the chemicals they use are
safe.
Chemicals of concern
Generally those on list for temporary permission/substitution. For
these problem chemicals assurances would be needed that they
would not harm the environment. At present in the UK the chem-
ical companies provide formulations and ecotoxicity data in line
with the requirements of the HOCNF, Government is then ap-
plying OSPAR’s prescreening to identify those components which
are candidates for substitution and are informing suppliers of their
findings.
11
Chemistries raised as a potential problem are:
Polymers
Polymers with Mw > 600 are assumed not to pass through bi-
ological membranes (the figure is 700 in Denmark, however no
upper limit in Norway). This can be seen as a positive factor
as they will not bioaccumulate either. However, many polymers
have biodegradation rates of < 20%. OSPAR sees this as a nega-
tive factor, therefore these polymers are immediately marked for
substitution through the pre-screening criteria.
Surfactants
There is general agreement that all chemicals with surface-active
properties are surfactants. However, contrary to CHARM guide-
lines, OECD 117 [? ] Pow values may be acceptable for some
surfactants, providing the data looks sensible. One chemical sup-
plier commented that OECD 107 (shake-flask method) [? ] can
provide sensible Pow data for high molecular weight surfactants.
The solvents can be removed at 100 oC and the residues weighed.
Defoamers
There is an interest in defoamers due to the presence of fluo-
rosilicones, which are non-biodegradable and therefore marked for
substitution.
Suggestions for chemical selection:
1. Select chemicals based on those most widely used.
12
2. Look at substances where their continued use is under pressure
due to weaknesses in the CHARM approach e.g. not those that
are easily dealt with by CHARM but those that do not fit, for
example, surfactants. The ability of CHARM to deal with surfac-
tants could conceivably be altered as the result of work from the
CHARMval project.
3. CHARM copes reasonably with the non-biodegradable substances
and just applies the relevant results as it goes along. However,
non-biodegradable substances are indicated for substitution by
OSPAR pre-screening. This pre-screening is NOT part of the
CHARM model. If evidence was found by CHARMval to support
them NOT being indicated for substitution this is something that
would need to be taken up with OSPAR via OIC.
4. Chemicals or chemistries that are most misrepresented in CHARM
and pose a significant environmental danger.
Points to note:
1. All surfactants identified as being in CHARMs ‘miscellaneous’ sur-
factant type are automatically assigned fraction released values
of 1 (i.e. assumed from the point of regulation that 100% dis-
charged). It is in the chemical industry’s interest not to have
surfactants in this grouping although it may of course still be ap-
propriate for some chemistries when further work is carried out.
2. Toxicity data is based on preparations, not at substance level. An
expert judgement is therefore made by CEFAS toxicologists as to
13
the toxicity of an oilfield chemical, based on the chemical com-
position of the preparation supplied by the chemical company.
Chemical composition consists of the CAS No, molecular weight
(or average molecular weight), weight percent and chemical de-
scription (which is sufficient for a structure to be drawn) of each
substance in a preparation. This data is supplied by chemical
companies to CEFAS, along with the Pow of each substance in a
preparation.
3. Chemicals could have a high OCNS category (A/B) but a low
RQ. This is mainly due to the different ways in which the OCNS
scheme and CHARM use Log Pow values.
4. There has been voluntary ban in the North Sea for the past 3-4
years on the use of alkyl phenol ethoxylates due to their oestro-
genic properties.
• Views of the Chemical Suppliers
Chemicals of concern
1. Anything with a low Pow (< 2 i.e. water-soluble) and a high
toxicity (for example benzalkonium QUATs and imidazolines,
which generally have EC50s in the range 0-10ppm). These two
parameters most affect the hazard quotient.
2. Ethoxylated amines/acids. These attach to fish membranes
and they drown.
3. Resins. These have low mammalian toxicity (? ]). They are
particularly effective. If banned, difficult to replace.
14
4. The Chemical Regulations do not give a definition for sur-
factants but also say that Log Pows cannot be calculated for
them. However, there may be exceptions.
5. Chemical companies are always looking for alternatives to aro-
matic solvents.
6. Water treatment chemicals used in hydrocyclones. They have
poor degradability, are cationic and toxic.
Chemicals that should be studied
1. General consensus that corrosion inhibitors and demulsifiers
are two groups that should be studied. Also flocculants and
H2Sscavengers (reaction products are not measured but could
be important environmentally).
2. General consensus that five or six demulsifier chemistries widely
used plus unique chemistries.
3. Only nonyl phenol and nonyl phenol ethoxylates are banned/not
used. Does not apply to resins, which are still used in the
North Sea (Jacques et al, 2002).
4. The selected mixture of demulsifier chemistries, which have
a high viscosity, are normally blended with suitable solvents,
such as heavy aromatic naphtha and isopropyl alcohol, to pro-
vide a liquid that pours at the lowest expected temperature.
5. The least complex chemistry to study is probably a defoamer
(solvent + defoaming agent).
6. Corrosion inhibitor chemistries are generic, simpler and better
15
understood (compared to demulsifiers). Therefore should be
studied first.
• View of the Oil Companies
Chemical selection
The oil companies use chemicals that are in OCNS categories B-E
(there is a voluntary agreement not to use chemicals in category
A). The view was expressed that expertise to make environmental
decisions on chemical selection rests with, and should come from,
the government laboratories. Their key priority is performance,
although greener chemicals will be used if equally effective. In the
UK, companies see it as up to the chemical supplier to offer alter-
native (greener) chemicals. In Norway the chemical suppliers may
actually be contractually obliged by the oil companies to develop
and offer greener alternatives. Defoamers and the reaction prod-
ucts of H2S scavengers were both highlighted for study. In Norway
there are ongoing studies on radiolabelled emulsion breakers, but
these are all laboratory based.
Chemical substitution
When asked the question: “have you had any difficulties when
substituting more hazardous chemicals in terms of performance?”
the companies responded that attempts to substitute chemicals
for those in lower OCNS categories have not worked particularly
successfully, i.e. they have not been as effective. However, al-
ternative products are tried on an ongoing basis. One company
16
had rarely substituted chemicals. When it did it was for the same
category and tonnage usage. Another company reported that the
solvents were the most frequently changed substances in a prepa-
ration. Platforms may change chemicals fairly often or have man-
aged longer term (e.g. five-year contracts) evolving around similar
chemistries.
4 Surfactant Fraction Released Default
Values in the CHARM Model
Background Information on Pow Determinations
A critical parameter used to determine the water borne component concen-
tration of an oilfield chemical is the octanol-water partition coefficient (Pow).
Pow values, or more specifically log Pow, are supplied by oilfield chemical
manufacturers/suppliers to the regulatory authorities as part of the HOCNF
process. These values were originally used to assess the theoretical potential
for a chemical to bioaccumulate in marine organisms. Chemicals with log
Pow values of > 3 and a molecular weight of < 600 are deemed likely to
bioaccumulate. These values are now often used to estimate the partitioning
of oilfield chemicals between oil and water and their ready availability has
resulted in their adoption for fraction released determinations both by chem-
ical suppliers and in the CHARM model.
Octanol-water partition coefficients are presently calculated using two dif-
ferent methods. Log Pow values in the range -2 to 4 may be directly mea-
17
sured by the OECD 107 shake-flask method [? ]. However many oilfield
chemicals contain components with Pow values greater than 4. The use of
a high-pressure liquid chromatography (HPLC)-based method (OECD 117)
extends the range of log Pow estimations from 0 to 6 [? ]. This method
exploits the relationship between an intrinsic HPLC property of a substance
and its directly measured octanol-water partition coefficient, as determined
by the shake-flask OECD 107 method. Log Pow of an analyte can be derived
from the correlation between its retention time on an HPLC column and the
retention times of standards with known Pow values.
Due to the system’s capability of separating multi-component mixtures into
a series of separated peaks, HPLC has been used to assign an average Pow
value to materials comprising a mixture of several components, including oil-
field chemicals.
For HOCNF submissions to CEFAS, the log Pow of each individual substance
in a preparation is determined. If more than one peak appears in the chro-
matogram a range of Pow values will be given which encompass the first and
last peaks. If the substance is a series of homologues, a weighted-average
may be calculated and this information provided to CEFAS.
The fraction released equation used in CHARM not only requires information
on the Pow of the individual components in an oilfield chemical formulation
but also their relative abundance. To apportion quantitative information to
the composition of the original analyte requires rigorous criteria to be met.
These are: i) that the various components in the original mixture will have
equal responses - peak area / unit mass injected - for the detector in use,
and; ii) that the detectors have a linear response for each peak over the range
18
of injection quantities being encountered.
The number of component peaks detected in a mixture and their relative
abundance will depend upon the type of HPLC detector used. The most
commonly used detectors to approximate the criteria above are the differ-
ential refractive index (DRI) detector and the ultraviolet absorption (UV)
detector. The RI detector compares the refractivity of the HPLC column
eluant containing the sample with the eluant prior to the injector. Only in
the highly unlikely event of an analyte solution having an identical refractive
index to the eluant would it not be detected. However, the magnitude of the
response is dependent on the difference between the RI of the eluant with
and without analyte, with equal concentrations of different analytes altering
the RI of the solution to differing extents. Additionally, the response of the
DRI to increasing concentrations of analytes is not normally linear. Quan-
tification of individual peaks without proper standards is therefore highly
problematical.
The response of the UV detector to a particular compound is dependent
on the material’s extinction coefficient at the detector’s wavelength and for
many compounds lacking a UV chromophore this is zero. Other analytes
with the requisite uv-absorbing molecular structure will exhibit a very wide
range of responses. These properties indicate that extreme care must be ex-
ercised in data interpretation if the UV detector, which is more sensitive and
easy-to-use than the RI detector, is being employed for the analysis.
19
The Application of CHARM to Surfactant Chemistries
The current OECD 117 method is not suitable for most surfactant chemistries.
Unfortunately, many of the oilfield chemicals in the higher (i.e. potentially
more hazardous to the environment) OCNS categories contain surfactant
chemistries. For these a number of default values have been incorporated
into the CHARM manual. However, they were derived from a very limited
data set of production chemicals and, as such, have been the subject of major
disquiet amongst the chemical suppliers, who argue that they do not cover
the diverse range of chemistries actually used. Further, no field validation
data for these values are available.
The surfactant chemistries and their default fraction released values for use
in the CHARM model is given in Table 5. The values are those given in
the CHARM Manual, Version 1.2 (updated on the 12 July 2001). All other
surfactant chemistries are automatically assigned a fraction released values
of 1, i.e. assumed 100% discharged in produced water.
5 Classification of Surfactants
The chemical suppliers were asked whether they could suggest a classification
scheme for commonly used surfactant chemistries, to enable field validation
data obtained from the project to be applied to a range of products from
different suppliers. Initially, no suggestions were received. In light of this,
information was retrieved from the Internet. In addition, the chemical con-
stituents given on MSDS sheets for a range of demulsifiers and corrosion
inhibitors were put onto a spreadsheet, along with information on their solu-
20
Table 5: CHARM Default Chemistries and Fraction Released Values for Surfac-tants.
Type of surfactant Fraction released
Quaternary amines 1.0
EO-PO Block polymer demulsifier(Ethoxylate - Propoxylate)
0.4
Imidazolines 0.1
Fatty amines 0.1
Primary amines (cationic type, C≥ 12) 0.1
Phosphate esters (anionic type, C≥ 13) 0.1
Others 1.0
bility and any available environmental information, to assist in the chemical
selection process.
General Description of Corrosion Inhibitors and Demul-
sifiers - Their Chemistry and Application
• Corrosion Inhibitors [? ]
Generally, deep oil-and gas-containing formations represent an extremely
corrosive system as a result of the simultaneous presence of mineralised
water, carbon dioxide and hydrogen sulphide, with hydrocarbons. Cor-
rosion damage can be intensified by factors such as elevated temper-
ature and pressure, sulphur present either in the elementary state or
as polysulphide, organic acids, oxygen introduced with injection water,
microbiological activity at certain stages of production, and mineral
acids that may have been used to treat strata of low permeability. Flow
21
rate is also a critical factor in the corrosion process. Corrosion occur-
ring in oil and gas systems takes various forms, e.g. general corrosion,
pitting, sulphide stress corrosion cracking, embrittlement, blistering,
corrosion fatigue, erosion-corrosion, cavitation, galvanic corrosion and
microbiological corrosion.
Types of Inhibitors
Most of the inhibitors currently used in producing wells are or-
ganic nitrogenous compounds. The basic types have long chain
hydrocarbons (usually C18) as a part of the structure.
The most frequently used inhibitors in general petroleum produc-
tion can be classified as follows:
? Amides/imidazolines
? Salts of nitrogenous molecules with carboxylic acids
? Nitrogen quaternaries
? Polyoxyalkylated amines, amides and imidazolines, and
? Nitrogen heterocyclics, and compounds containing P.S.O.
• Demulsifiers
Emulsion formation
Because of the coexistence of the oil and water in the rock strata
of oil reservoirs, the crude oil extracted from the reservoir con-
tains emulsified water. Naturally occurring emulsifying agents i.e.
asphaltenes, paraffins, resins, organic acids, metallic salts and silt,
stabilise these emulsions.
22
Problems caused by emulsions
The presence of water along with the crude oil is undesirable
because of problems directly correlated to foaming, corrosion in
pipelines and tanks, and increased volume and viscosity.
Physical description of emulsions
Macro emulsions are thermodynamically unstable, and given suf-
ficient time, will separate out naturally. However, for financial
and practical reasons the process needs to be speeded up. The
process is termed demulsification. The most widely used process
in the oil industry is chemical demulsification. The difficulty with
using emulsion breakers is that they are typically specific for site
or crude-oil type, which implies that a certain emulsion breaker
that has worked for a crude oil type A will not necessarily work
for a crude oil type B.
Theory of demulsification
Rigorous attempts have been made to correlate between demulsi-
fier performance and physical properties such as molecular struc-
ture, interfacial tension, HLB (for explanation of HLB see later),
interfacial viscosity, partition viscosity, dynamic interfacial ten-
sion, and relative solubility number. However, according to chem-
ical suppliers, it is difficult to give conclusions due to big differ-
ences exhibited by each crude oil.
Demulsifier chemistry
Early demulsifiers included Turkey red oil, sulphuric acid, sul-
23
phated castor oil, polyamines, polyhydric alcohols etc. Since the
1940s, when the technology of alkylene oxide condensation started
to evolve, almost all demulsifier products have been made up of
condensation products of ethylene, propylene and butylene oxide.
Most are alkoxylate polymers that are mainly ethoxylated and
propoxylated or both. They are macromolecules held in chains,
industrially synthesised from petroleum chemicals. Polymers used
as demulsifiers are surfactants that counteract the effect of crude
oil asphaltenes. They contain both hydrophilic and hydrophobic
groups. When added to the petroleum emulsion the polymer sur-
factant locates itself in the interface between the water and oil
molecules. The hydrophilic groups orientate themselves towards
the water whilst the hydrophobic groups orientate themselves to-
wards the oil. The best polymeric surfactants are alkoxylated
material derivatives. Because they are alkoxylated, they are con-
sidered as nonionic polymers. Sometimes a mixture of nonionic,
cationic, or anionic materials are used together, depending on the
oil characteristics. Ethoxylated nonionic surfactants are effective
multi-purpose and versatile substances. Commercial products are
obtained by reaction of ethylene oxide with a hydrophobe having
an active hydrogen group (e.g, fatty acids, alkylphenols, or fatty
alcohols) in the presence of a suitable catalyst. Commercial poly-
mer formulations are accompanied by solvents. These solvents do
not induce a chemical change in the polymer, rather, they affect
the physical properties of the polymer formulation.
24
Synthesis of Surfactant Chemistries used in
Oilfield Chemicals
Long chain fatty acids derived from natural products such as coco oil (C12-
C16), tall oil (pine oil, C18-C20) etc are often used as starting materials for
synthesising actives such as imidazolines [? ]. This results in actives contain-
ing a range of molecular weights reflecting, primarily, the alkyl chain lengths
and degree of unsaturation of the raw materials. Quaternary ammonium for-
mulations derived from tertiary Coco (and also synthetic) amines which have
been quaternerised with a reactive alkylation reagent such as benzyl chloride
or dimethyl sulphate, are used in corrosion inhibitors, biocides and as cationic
surfactants [? ]. Fatty and ether amines with alkyl chain lengths ranging from
C10-C18 may be reacted with ethylene oxide to produce a range of ethoxy-
lated fatty and ether amines containing from 2 to 15 moles of ethylene oxide
producing a range of different properties (www.tomah3.com/Documents/P-
03.pdf). Water solubility increases with addition of ethylene oxide.
EO/PO Block Copolymers used in Demulsifier Formulations
The strict definition of an EO/PO block copolymers is just EO & PO i.e.
HO − (CH2CH2)m − (CH2CH(Me)O)n − (CH2CH2O)mH
EO is hydrophilic, PO lipophilic (you need n > 5 PO groups to display
lipophilic properties). m and n can be varied to give a wide range of proper-
ties. Molecular weight goes from ≈ 1, 000 to > 10, 000 but for the oil industry
usually ≈ 4, 000.
25
The EO/PO polymers can be based on any initiator with one or more active
hydrogens. For example:
1. One active hydrogen - monofunctional alcohols, i.e. R-OH (R typically
C12-C14)
R−O − (PO)n(EO)mH
2. Two active hydrogens - glycols in water.
H − (EO)m(PO)nR− (PO)n(EO)mH
R = CH2CH2 or CH(Me)CH2
3. Three active hydrogens - e.g. glycerol
RCH2 − CHR− CH2R
R = (PO)n(EO)mH
4. Four active hydrogens e.g. ethylene diamine.
R2NCH2CH2NR2
R = (PO)n(EO)mH
etc.
With such a wide range of possible substrates it is very difficult to come
up with one generic structure. Also, although the surfactants are generally
26
capped with EO, they can sometimes be capped with PO. The PO chain has
to be > 5 PO groups to show hydrophobic properties, and it is the PO chain
that does not biodegrade (personal communication).
As this class of surfactants can have such a huge range of properties, for
example, depending on the number of EO and PO groups attached they can
range from completely water soluble to completely oil soluble (see below), it
is not sensible to have one discharge factor (0.4) as per the CHARM model.
HLB (Hydrophile-Lipophile Balance)
(www.paddocklabs.com/publications/secundum/secart41.htm)
Surfactants are often described in terms of their HLB values. A brief de-
scription of HLB follows.
The HLB (Hydrophile-Lipophile Balance) system is used for describing char-
acteristics of a surface-active agent. It consists of an arbitrary scale to which
HLB values are determined and assigned. If the HLB value is low, there is a
low number of hydrophilic groups on the surfactant and it is more lipophilic
(oil soluble). If the HLB value is high, there is a large number of hydrophilic
groups on the surfactant and it is more hydrophilic (water soluble) than oil
soluble. Some general applications of materials with various HLB values are
shown below:
1-3 Antifoaming agents
3-6 Emulsifying agents (w/o emulsions)
7-9 Wetting agents
27
8-18 Emulsifying agents (o/w emulsions)
13-16 Detergents
16-18 Solubilising agents
Surfactants with HLB values of 1-10 are oil soluble and 11-18, water soluble.
For example, to prepare an oil in water emulsion using stearic acid, an HLB
of 15 is required.
Information from MSDS Sheets
Some of the chemistries taken from MSDS sheets of currently used corrosion
inhibitors and demulsifiers are summarised below.
• Demulsifiers
Active components
1. Oxyalkylated amines
2. Alkyl benzene sulphonic acid
3. Polyoxyalkylene glycols
4. Alkylene oxide
Solvents
1. Light aromatic naphtha
2. 2-butoxyethanol
3. 2-ethyl hexanol
4. Aliphatic alcohols
28
5. Alkyl benzenes
6. Xylene
7. Aromatic solvent
8. Aromatic hydrocarbons
9. Aromatic hydrocarbons and alcohols
• Corrosion Inhibitors
Active components
1. Amine ethoxylate
2. Amine alkoxylate
3. Alkylamino propionic acid
4. Quaternary ammonium compounds
Solvents
1. Butanediol
2. Butoxyethanol
In addition, amides, imidazolines and phosphate esters are common corrosion
inhibitor ”actives”.
6 Chemistries Suggested
In terms of chemicals there was general agreement that corrosion inhibitors
and demulsifiers should be studied. One chemical company was more specific,
suggesting the project look at the following: quats, imidazolines and phos-
phate esters as corrosion inhibitors, and phosphonates as scale inhibitors.
29
They commented that most emulsion breaker and wax chemistry is poly-
meric (by implication, need not be investigated2), although some of the lower
molecular weight surfactant chemistry is of interest, e.g. alkoxylated alcohols
and amines.
One chemical supplier suggested the following generic demulsifier chem-
istry descriptors:
? Alkoxylated phenols and alcohols
? Alkoxylated alkyl phenol formaldehyde resin
? Alkoxylkated glycol
? Alkoylated amine
? Alkoxylated acids
? Alkoxylated sorbitol
7 Conclusions and Recommendations
I was hoping that the meetings and subsequent communications that I have
had with the sponsors would have enabled me to finalise the chemistries for
investigation. However, this has not been the case. There was a general
consensus that corrosion inhibitors and demulsifiers should be investigated,
but there was a lack of progress on generic classification. One company was
specific on chemistries for investigation. These are given in the previous Sec-
tion. In light of this I would suggest that these form the basis for discussion
2High molecular weight polymers will, theoretically, not pass through cell membranesand, therefore not bioaccumulate although they may be poorly degradable.
30
at the next steering group meeting, along with the existing CHARM default
surfactant chemistries.
31
1 Introduction
The present report summarises experiments and results obtained in the at-
tempt to quantify a series of oilfield chemicals in discharged produced waters
and compare the experimentally determined fractions released to the pre-
dicted CHARM values. The investigation has focused on corrosion inhibitors,
in particular on two types of surfactants:
? quaternary ammonium salts (QUATs)
? imidazolines
In the framework of this project, the instrument of choice is an electrospray
ionisation tandem mass spectrometer (ESI-MS/MS). This technique has the
sensitivity required for this type of analysis, i.e. down to 0.01 - 1 mg/l
depending on the substance. In addition the parent - daughter ion analysis
(MS/MS) is very specific and makes mistaken identification of compounds
very unlikely. In contrast, conventional wet chemical techniques do not have
the sensitivity or specificity to identify and measure specific oilfield chemical
substances in produced water.
In order to measure those compounds by ESI-MS/MS; three major prob-
lems need to be overcome:
1. the presence of salt in produced water which interferes with the mass
spectrometer measurements;
2. the matrix effects: produced water contains a whole range of organic
compounds, some of which may interfere with the analysis of the chem-
ical compounds of interest;
33
3. the adsorption of surfactants to the sampling containers, analytical
equipment and chromatography columns used in the analysis.
The use of deuterated internal standards should overcome these problems.
Deuterated standards have the same properties of their hydrogenous coun-
terparts, therefore the same behaviour but they are clearly distinguished by
the mass spectrometer since they have a different mass due to the substi-
tution of some of the hydrogen atoms with deuterium (see Figures 8 and
15). Further, as already pointed out, the use of tandem mass spectrometry
(MS/MS) avoids incorrect identification of peaks in the produced water (see
Figure 24) so that in no way hydrogenous and deuterated counterparts of the
same molecule can be mistaken for one another.
In this report (period May 2002 - October 2003) results are reported for:
• synthesis of imidazolines and quaternary ammonium salts, in their hy-
drogenous and deuterated forms;
• solid phase extraction (SPE) techniques applied to salty aqueous solu-
tions and produced water samples to remove the salt and extract the
oilfield chemicals to be analysed via mass spectroscopy;
• a new, fast on-line system that couples a high-pressure liquid chro-
matography (HPLC) column with the mass spectrometer;
• the use of the deuterated standards available, i.e. D-C12, D-C14, D-
C16 QUATs and D-(2:1) imidazoline, on produced water samples to
quantify the corresponding chemical residues.
34
N
N
NH2
PENDANT GROUP
HYDROCARBONTAI L
IM IDAZOLINEHEAD GROUP
Figure 1: Molecular structure of imidazolines.
2 Imidazoline Synthesis
Imidazolines are widely used in the oil industry as corrosion inhibitors, how-
ever their properties and behaviour in such complex environment is far from
being well known from a scientific point of view. For example, the mechanism
by which these compounds prevent corrosion is not entirely understood and
their obviously good performance has been only recently supported by ex-
perimental evidence [1–8] and investigated by molecular modeling techniques
[3, 6, 9–12]. This is because they are used in low concentrations and operate
in complex environment. For the very same reasons their quantification in
produced water is particularly difficult and only recently attempted [13–16].
In general, this class of compounds present a pendant group, an imi-
dazoline head group and a hydrocarbon tail, as shown in Figure 1. Recent
molecular modeling [3, 6, 9–12] studies suggest that the head and pendant
35
group promote bonding of the molecule to the surface, while the hydrocar-
bon tail forms a protective monolayer. This hydrophobic barrier protects the
surface from the water and corrosive ions, i.e. makes imidazolines effective
as corrosion inhibitor. However not every imidazoline is as effective: the hy-
drocarbon tail must be long enough (at least C8 [5]) to form the protective
film and cover the surface and the length of both the tail and the pendant
group must be such that the layer forms rapidly [9]. In addition, the pen-
dant group seems to play a role in locking the hydrocarbon tail into position,
reducing free motion of water to the surface [10]. In other words depending
on the specific molecular structure of the imidazoline, it may be more or less
effective as corrosion inhibitor.
Palmitic acid (CH3(CH2)14COOH) and diethylenetriamine (DETA)
(NH(CH2CH2NH2)2) were mixed under vacuum at 210 oC to obtain a mix-
ture of (2:1) and (1:1) palmitic imidazoline (see Figures 2 and 3). The
ratio between the two imidazolines synthesised varied depending on the molar
ratio between the reagents and the reaction time.
However, palmitic imidazolines were soon abandoned in favour of oleic
imidazolines, perhaps the most common class of corrosion inhibitors used
by the oil and gas industry. Oleic imidazolines are widely used because
of their good performances, which agrees with experimental observation that
these molecules absorb rapidly [9] and is the reason why they are often chosen
as a model molecule, including the present project.
Oleic acid (cis CH3(CH2)7CH=CH(CH2)7COOH) and DETA were mixed
under vacuum at 210 oC to obtain a mixture of (2:1) and (1:1) oleic imida-
zolines (see Figures 4 and 5). Note that both molecules carry a polar group,
36
N
N
(CH2)14CH3
NH2
Figure 2: Molecular structure of (1:1) palmitic imidazoline.
N
N
(CH2)14CH3
NH
O
(CH2)14CH3
Figure 3: Molecular structure of (2:1) palmitic imidazoline.
37
N
N
(CH2)7CH=CH(CH2)7CH3
NH2
Figure 4: Molecular structure of (1:1) oleic imidazoline.
i.e. an amine group NH2 for (1:1) oleic imidazolines and an amide group
NHCO for (2:1) oleic imidazolines, although the latter is followed by a rel-
atively long hydrocarbon chain that makes the molecule less water soluble
compared to (1:1) oleic imidazolines. This extra hydrocarbon chain not only
changes the nature of the molecule itself, but also the delicate equilibrium
between the pendant group and the tail that controls the effectivess of the
imidazolines as corrosion inhibitors. Different properties and behaviours are
therefore expected for these two imidazolines, along with a probable different
efficiency as inhibitors.
The reaction proceeds as shown in Figure 6. The reagents are mixed at
room temperature, however the salt formation may not be complete because
of the high viscosity of the mixture itself. The mixture is then transferred
to the heating apparatus, which is shown schematically in Figure 7. The
part of reagents that has not formed the salt is therefore heated up above
the boiling point of both oleic acid (Bp ≈ 195oC at 1.2 mm Hg) and DETA
(Bp ≈ 208oC). In theory they should cool down in the apparatus and reflux
38
N
N
(CH2)7CH=CH(CH2)7CH3
NH
(CH2)7CH=CH(CH2)7CH3O
Figure 5: Molecular structure of (2:1) oleic imidazoline.
back into the reaction flask. In practise because of the vacuum, some, in
particular DETA, are lost from the system. Even an initial molar ratio of
oleic acid to DETA of 1 : 5 does not lead to the exclusive formation of
(1:1) imidazoline as would be expected, but to a ratio of (1:1) : (2:1) oleic
imidazolines of 4 : 1 (see Table 6).
Unfortunately it turned out to be difficult to obtain a pure (1:1) oleic
imidazoline as reaction product when using very low amount of starting ma-
terials, which is the case with deuterated precursors, as explained in the
following paragraphs. Different reaction paths are possible and were briefly
investigated to improve the control over the synthesis products and minimise
the loss of reagents. However, the limited time available for this project did
not allow us further investigation of alternative imidazoline synthesis.
The deuterated imidazoline was synthesised using D2-9,10-oleic acid (Cam-
bridge Isotope Laboratories) instead of oleic acid (see Figure 8). In Figures
39
R
O
OH+ NH
NH2 NH2
OH2-
R
O
NHNH
NH2
R
O
NHNH
NH R
O
+
OH2-
N
N
R
NH2
N
N
R
NH
RO
+
R
O
O- NH
NH3+
NH2
Figure 6: Imidazoline synthesis reaction path.
40
H2O
H2O
T = 210 C
T ~ 20 C
vacuum
Figure 7: Apparatus used for imidazolines synthesis.
CH3
OHO
D
D
Figure 8: Molecular structure of D2-9,10-oleic acid.
41
Tab
le6:
olei
cim
idaz
olin
esy
nthe
sis
(tes
t13
refe
rsto
deut
erat
edim
idaz
olin
esy
nthe
sis)
.
TE
ST
Ole
icA
cid
DE
TA
Mol
arT
ime
MA
SS
SP
EC
INT
EN
SIT
IES
NR
(mol
)(m
ol)
Rat
io(h
rs)
AM
IDE
DIA
MID
E(1
:1)
IMID
AZO
LIN
E(2
:1)
IMID
AZO
LIN
ER
AT
IO(1
:1)/
(2:1
)
50.
010
0.05
01:
53
13<
510
025
4:1
150.
020
0.06
01:
33
4010
01:
2.5
80.
0003
50.
0003
31:
13
710
100
(2:1
)
90.
0003
60.
0003
51:
13
53
100
(2:1
)
100.
0003
70.
0003
51:
10.
5<
66
100
(2:1
)
110.
0003
50.
0007
2:1
0.5
510
0(2
:1)
120.
0003
80.
0003
51:
11
1410
0(2
:1)
140.
0003
50.
0035
1:10
0.5
10<
776
100
1:1.
3
110
<10
3010
01:
3.3
210
052
1:0.
5
130.
0003
50.
0003
81:
10.
53-
1010
0(2
:1)
42
0.0
0.1
0.2
300 310 320 330 340 350 360 370 380 390 400
m/z
Inte
nsit
y (n
orm
alis
ed to
1)
(1:1) IMIDAZOLINE
m = 350 AMIDE m = 368
Figure 9: Normalised mass spectra for synthesis 10 (red) and 13 (black) in therange m/z = 300 to 400.
9 and 10 synthesis 10 is compared to synthesis 13, i.e. the deuterated imi-
dazoline. The signal due to the amide is clearly absent in both cases while a
weak signal from the (1:1) imidazoline can be recognised only for synthesis
10 (Figure 9), i.e. the hydrogenous imidazoline synthesis. In Figure 9 the
hydrogenous and deuterated (2:1) imidazolines are clearly visible and distin-
guished from one another by the MS. A weak signal for a diamide ias also
present for synthesis 13.
The purity of the deuterated imidazoline has been assessed by comparing
it to an imidazoline provided by one of the chemical companies involved in
the project. From the chromatogram shown in Figure 11, the ratio between
the (2:1) and (1:1) imidazolines in the substance can be evaluated. Using
selected ion recording (SIR) analysis, and by comparing the peak areas of
the ions at m/z 348 and 612, 1ppm of the substance gives an area of 21.9
million ions (or 0.67ppm) for the (1:1) imidazoline and 10.8 million ions
(or 0.33ppm) for the (2:1) imidazoline. Assuming the relationship between
43
0.0
0.5
1.0
500 525 550 575 600 625 650 675 700
m/z
Inte
nsit
y (n
orm
alis
ed to
1)
(2:1) IMIDAZOLINE
m = 614.6 (2:1) d4 IMIDAZOLINE m = 618.7
DIAMIDE m = 634.6
Figure 10: Normalised mass spectra for synthesis 10 (red) and 13 (black) in therange m/z = 500 to 700.
concentration and signal intensity is linear, 1ppm of pure (2:1) imidazoline
should therefore give a peak area of approximately 32.6 million ions, which
is the area obtained from synthesised deuterated imidazoline. Figures 12
and 13 show the MS-MS spectra from oleic imidazoline, hydrogenous and
deuterated, respectively. The main daughter ion in Figure 12 corresponds to
the part of the molecule highlighted in red (m = 305). The corresponding
deuterated daughter is clearly visible in Figure 13 (m = 307), confirming the
product of the reaction is a deuterated imidazoline.
3 Quats Synthesis
Quaternary ammonium compounds, or QUATs, are cationic surfactants.
Although cationic surfactants account for only 5-6% of the total surfactant
production, they are extremely usefull for some specific uses, because of
their peculiar properties. They are not good detergents nor foaming agents,
44
2 3 4 5 6 7 8 9 10
time (min)
inte
nsit
y provided imidazoline_m=612.50
d4-imidazoline_m=618.70
provided imidazoline_m=348.50
area 32,536,394
area 10,751,146
area 21,936,192
Figure 11: Comparison of the peak areas in a SIR chromatogram between thedeuterated imidazoline and an imidazoline (substance) provided by one of thecompanies (1ppm solutions in MeOH:H2O=90:10).
dau of m = 612.5
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
0 100 200 300 400 500 600 700
m/z
Inte
nsit
y
306.1
NH
(CH2)7CH=CH(CH2)7CH3O
N
N
(CH2)7CH=CH(CH2)7CH3
Figure 12: MS-MS spectrum: daughter ions of m/z 612.50, (2:1) imidazoline.
45
dau of 618.70
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
0 100 200 300 400 500 600 700
Mass
Inte
nsit
y
308.9
NH
(CH2)7CH=CH(CH2)7CH3O
N
N
(CH2)7CD=CD(CH2)7CH3
Figure 13: MS-MS spectrum: daughter ions of m/z 618.70, d(2:1)imidazoline.
and they cannot be mixed in formulations which contain anionic surfactants,
with the exception of non quaternary nitrogenated compounds, or when a
catanionic complex synergetic action is sought. Nevertheless, they exhibit
two very important features:
1. their positive charge allows them to adsorb on negatively charged sub-
strates, as most solid surfaces are at neutral pH. The positive charge
enable them to operate as floatation collectors, hydrophobating agents,
corrosion inhibitors as well as solid particle dispersant
2. many cationic surfactants are bactericides. They are used to formulate
desinfectants for domestic and hospital use, and to sterilize food bottle
or containers.
Among cationic surfactants, QUATs are the dominant commercial ex-
ample. However, QUATs is a term used to describe a very wide range of
molecules. These molecules are in fact only associated through a quaternary
nitrogen in the structure as a common feature. Their use is as broad, span-
46
ning from detergents and biocidals to surfactants and pharmaceuticals. The
individual members of this broad family have very different properties. For
example monoalkyl QUATs are biocidal ingredients but have little surfac-
tant activity, on the other end twin chain (long) QUATs are mainly used as
surfactants and have no significant biocidal activity. The four arms of the
quaternary compound can look very different, and the molecules can be com-
pletely different functionally speaking. On the other hand, the counter-ion,
the negative chloride or methylsulphate ion included in the compound, seems
in most cases to make little difference to the properties of the compound. In
general, the positively charged nitrogen part of the molecule should attach
itself to surfaces (which for the most part are negatively charged) and there,
depending on the side chains, act as plasticiser, antistat agent, biocide, cor-
rosion protection in relation to a surrounding medium or otherwise change
the properties of the surface.
In the framework of this project, the interest is focussed on benzalkonium
QUATs or alkyl dimethyl benzyl-ammonium salts. Therefore in this report
the term QUATs refer to benzalkonium salts, where the alkyl side chain is a
long hydrocarbon chain, typically from C12 to C20 (see Figure 14).
As with the imidazolines, the QUATs synthesis was performed using non-
deuterated (cheaper) precursors, to evaluate the suitability and yield of the
different methods available. Once a suitable method had been established,
deuterated precursors were used. The synthesis is summarised in Figure 14.
Benzyl chloride (C6H5CH2Cl or C6D5CD2Cl) and the appropriate triamine
were mixed under vacuum at T ≈ 100oC for ≈ 1hr. The reaction was carried
out in MeOH:H2O, H2O or no solvent, best results being achieved in the
47
CH2Cl + N R
CH3
CH3
N+
CH3
CH3 R
Cl- R = (CH2)13 CH3
R = (CH2)15 CH3
R = (CH2)11 CH3
Figure 14: QUATs synthesis reaction path.
aqueous environment. The molecular structure of the deuterated ammonium
salts synthesised is shown in Figure 15. Figure 16 shows the mass spectrum
from the product obtained in Synthesis 1, i.e. C16 QUAT. The spectrum was
the same as that of a C16 QUAT standard bought from Aldrich. Further,
MS/MS showed the synthesised QUAT produced the same daughter ion spec-
trum as the Aldrich standard, confirming the synthesis had been successful.
The spectra are shown in Figures 17 and 18. The purity of the reaction
product can be checked with the MS, since any trace of triamine left would
be detected, as shown in Figure 19 for the synthesised C14 QUAT (Synthesis
3), where the product is clearly rich in unreacted triamine (m=242). The
spectrum shown in Figure 20 shows the same quaternary ammonium salt
purchased from Aldrich, no triamine can be detected in this case.
Deuterated QUATs were synthesised using fully deuterated benzyl chlo-
ride instead of its hydrogenous counterpart. Figure 21 shows the mass spec-
trum for D7 C16 QUAT, where traces of triamine and hydrogenous C16 QUAT
48
CD2 N+
CH3
CH3
R
DD
D
D D
Figure 15: Molecular structure of the deuterated quaternary ammonium saltssynthesised where R is either (CH2)11CH3 (C12 QUAT), (CH2)13CH3 (C14 QUAT)or (CH2)15CH3 (C16 QUAT).
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
0 100 200 300 400 500 600
Mass
Inte
nsit
y
QUAT m = 360
Figure 16: Mass spectrum from Synthesis 1 product (C16 QUAT).
49
0.0
0.5
1.0
0 50 100 150 200 250 300 350 400
m/z
Inte
nsit
y(n
orm
alis
ed to
1)
N+
CH3
CH3 C16H33
Figure 17: Daughter ions from C16 QUAT, m/z = 360 (synthesis 1). Note themain daughter ion corresponding to the part of the molecule highlighted in red(m/z = 91).
0.0
0.5
1.0
0 50 100 150 200 250 300 350 400
m/z
Inte
nsit
y(n
orm
alis
ed to
1)
Figure 18: Daughter ions from C16 Quat, m/z = 360 (Aldrich).
50
0.0
0.5
1.0
0 50 100 150 200 250 300 350 400 450 500
m/z
Inte
nsit
y TRIAMINE m = 242
QUAT m = 332
Figure 19: Mass spectra of Synthesis 3 product, C14 QUAT.
0.0
0.5
1.0
0 50 100 150 200 250 300 350 400 450 500
m/z
Inte
nsit
y
QUAT m = 332
Figure 20: Mass spectra of C14 QUAT, Aldrich.
51
0.0
0.5
1.0
100 150 200 250 300 350 400 450 500
m/z
Inte
nsit
y(n
orm
alis
ed to
1) d7-C16 QUAT
m = 367.4
C16 TRIAMINEm = 270.4 C16 QUAT
m = 360.4
Figure 21: Mass spectrum for deuterated C16 QUAT.
can be detected. Figure 22 shows the corresponding MS-MS spectrum for
m/z = 367.40. Note that the main daughter ion corresponds to the one
observed in Figure 17.
In Figure 23 a comparison between the deuterated C16 QUAT synthe-
sised (red arrow) and the hydrogenous one bought from Aldrich (black ar-
row) is presented. The MRM chromatograms clearly show that the Aldrich
compound contains traces of C18, C14 and C12 QUATs. Although they are
present in very small concentrations, they are above background level, as
shown by the comparison with the blank response (blue arrow), especially
the C18 QUAT. On the contrary, D-C16 QUAT signals show no traces of C18,
C14 or C12 above background level, C16 QUAT being only slightly above this
treshold. This is consistent with the H- NMR also carried out on this sample:
deuteration level was found to be 98%, in agreement with the purity assays
of the reagents.
Finally, Figure 24 shows the spectrum of an oilfield preparation spiked
52
0.0
0.5
1.0
0 50 100 150 200 250 300 350 400
m/z
Inte
nsit
y(n
orm
alis
ed to
1) CD2
DD
D
D D
N+
CH3
CH3 C16H33
Figure 22: Daughter ions from D7 C16 QUAT, m/z = 367.40. Note the maindaughter ion corresponding to the part of the molecule highlighted in red (m/z =98).
with deuterated QUATs. The signals from the deuterated standards are
clearly distinguishable from their hydrogenous counterparts, as expected.
4 Analytical Techniques
Different techniques were used to analyse produced water samples containing
QUATs and imidazolines: dichloromethane (DCM) extraction, solid phase
extraction (SPE) and high pressure liquid chromatography (HPLC), all fol-
lowed by ESI MS/MS quantitative analysis.
These different techniques have intrinsic advantages and disadvantages;
for example the DCM and SPE extraction, followed by MS analysis, are time
consuming compared to direct injection onto the HPLC-MS system. On
the other hand the amount of other organic compounds present in produced
water samples varies widely from platform to platform and cleaning steps
53
Figure 23: MRM chromatograms for various QUATs ions from a MeOH:H2O solu-tion (blue arrow), D-C16 QUAT synthesised (red arrow) and C16 QUAT purchasedfrom Aldrich (black arrow) in MeOH:H2O directly injected into the MS.
54
0
20
40
60
80
100
0 50 100 150 200 250 300 350 400
m/z
Rel
ativ
e in
ten
sity C12 QUAT; FW=304 C18 QUAT; FW=388
C16 QUAT; FW=360
d7 C16 QUAT; FW=367
d7 C14 QUAT; FW=339
C14 QUAT; FW=332
Figure 24: Spectrum of a chemical preparation (containing C12, C14, C16 andC18 QUATs) that was spiked with D7 C14 and D7 C16 QUAT.
prior to injection may be necessary. A summary of pros and cons related to
each technique is given in Table 7.
4.1 DCM extraction
DCM extraction was carried out on at least 50ml of produced water, which
was extracted with 2x10ml of DCM. The extract was dried under N2 and re-
dissolved in MeOH:HCOOH (25mM aqueous solution) (50:50). The solution
was then injected into the MS (10ul loop, carrier solvent MeOH:HCOOH
(25mM aqueous solution) (90:10), flow 0.1 ml/min) where selected masses
were detected (MRM mode, i.e. parent/daughter ion transitions).
DCM extracts organic compounds from the aqueous solution, including
the ones under investigation, however most of the other compounds that may
cause interference, i.e. signal suppression, with the MS analysis, are also
extracted. Therefore, depending on the actual concentrations of analytes to
55
Table 7: Summary of advantages and disadvantages of the different techniquestested on produced water (PW) samples.
PROs CONs
SPE
1. SPE removes salts and pos-sibly other interfering or-ganic compounds from thePW
2. Samples can be concen-trated or diluted
3. QUATs can be eluted withMeOH and therefore di-rectly injected into the MS
1. The procedure is time con-suming
2. Imidazolines can be elutedwith CHCl3, therefore theextract needs to be driedand redissolved in a moresuitable solvent for the MS
DCM
1. Samples can be concen-trated or diluted depend-ing on the levels of analytespresent
1. The procedure is time con-suming
2. Other organic compoundsare extracted which may in-terfere with the analysis
HPLC
1. PW samples are injected di-rectly into the HPLC-MSsystem, the analysis is veryfast
2. There may be separationthrough the column, effec-tively cleaning the samplefrom interference
1. PW samples need to be di-luted (50:50) with MeOH-HCOOH (25mM)
2. If there is no separationthrough the HPLC column,the signal may be weak dueto interference
56
0 1 2 3 4 5
time (min)
Inte
nsit
yC18 (m/z=388.5)
dC16 (m/z=367.4)
C16 (m/z=360.4)
dC14 (m/z=339.3)
C14 (m/z=332.4)
C12 (m/z=304.5)
area = 730,000
area = 267,000
area = 53,000
Figure 25: DCM extract chromatogram (sample L) using a direct injection intothe MS.
be investigated and eventual matrix effects, the extract can be diluted or
concentrated simply by changing the amount of MeOH:HCOOH used in the
attempt to optimise the MS response.
An example of the matrix effects on DCM extracts is shown in Figures 25
and 26: the same sample (sample L - DCM extract) is directly injected into
the MS, giving the chromatogram shown in Figure 25, or passed through
the HPLC system, Figure 26. The areas obtained from the D-standards
are much lower than expected in both cases but clearly increasing when the
sample goes through the HPLC column. The increase could be due to some of
the contaminants being flushed off during the washing procedure (see Section
4.3) and/or some separation between the contaminants and the analytes in
the column. The signals from the hydrogenous QUATs are also very low but
increasing with the aid of the HPLC column, although the areas are too low
to consider the output reliable.
57
0 2 4 6 8 10
time (min)
Inte
nsit
y
C18 (m/z=388.5)
dC16 (m/z=367.4)
C16 (m/z=360.4)
dC14 (m/z=339.3)
C14 (m/z=332.4)
C12 (m/z=304.5)
area = 1,814,000
area = 603,000area = 72,000
area = 48,000
Figure 26: DCM extract chromatogram (sample L) using the HPLC - MS onlinesystem.
Figure 27 shows another DCM extract (sample Z) where the only signal
clearly detected, is the one for C12 QUAT, followed by a weak signal for
C14 QUAT, indicating a greater amount of the former in the produced water
sample. The fact that the D-standards are not visible at all, although present
at approximately 0.5 ppm each, points to a very high suppression of the signal
due to interference. In this case the HPLC column did not have the cleaning
effect clearly observed for sample L.
4.2 SPE extraction
SPE columns are commonly used to clean-up complex samples and/or con-
centrate low levels of analytes, prior to analysis. A variety of different SPE
columns are available depending on the chemical nature of the analyte re-
quiring analysis, and a number of methods well established. Unfortunately
protocols are not available for the chemistries under investigation.
58
0 2 4 6 8 10
time (min)
Inte
nsit
y
C18 (m/z=388.5)
dC16 (m/z=367.4)
C16 (m/z=360.4)
dC14 (m/z=339.3)
C14 (m/z=332.4)
C12 (m/z=304.5)
area = 447,000
area = 27,000
Figure 27: DCM extract chromatogram (sample Z) using the HPLC - MS onlinesystem.
The solid phase extraction method adopted is known as reversed phase,
tipically used to extract hydrophobic or even polar organic analytes from
aqueous samples. Reversed phase extractions are in fact relatively non-
specific, therefore a wide range of organic compounds is typically retained.
Figure 28 schematically shows how it works:
• First of all the sorbent mass is conditioned, which means that it is
prepared for effective interaction with the analytes by solvation, i.e
unwrapping of the side chains (see for example Figure 29 to 31).
• Then the SPE is equilibrated, i.e. prepared for the specific sample
matrix with a similar solvent. These first two steps are not shown in
Figure 28.
• At this point the SPE is ready to be used and the loading step follows:
the hydrocarbon chains on both the analyte and sorbent are attracted
59
to one another by low energy Van der Waals dispersion forces; the
analytes, and likely some of the contaminants as well, are retained.
• In order to remove the impurities that are bound to the sorbent less
strongly than the analytes and to rinse residual, unretained sample
components that may remain from the sample loading step (for example
salt), this is followed by a washing step. In other words, this step should
wash the contaminants off the column.
• Finally, the elution step should selectively desorb and recover the ana-
lytes by disrupting the analyte - sorbent interactions so that the target
analyte is eluted and collected.
The recovery and purity of the target analytes can be optimised by adjusting
the composition of the solvents used during each of the steps, i.e. loading,
washing and elution. In reversed phase, a combination of MeOH, MeCN and
a strong acid (such as HCl) is often required to obtain quantitative recoveries
for polar analytes. However, it is desirable to choose an elution solvent
that is compatible with the final analytical method. In the case of mass
spectrometry, HCl suppresses the signal, even at very low concentrations,
therefore should be avoided. In conclusion, very often a compromise between
optimum recoveries and compatible solvents needs to be reached.
In a previous project (MIME, Grigson et al, 2000), benzalkonium quater-
nary ammonium salts were successfully measured in marine sediments using
SPE for sample preparation. However, the method was not entirely satisfac-
tory, as acidified methanol was required as a solvent, reducing the sensitivity
of the ESI-MS analysis. A synthesised (1:1) imidazoline could be eluted from
60
LOADING RETENTION
WASHING
ELUTIONANALYTES of INTERESTDISPLACED BY THEAPPROPRIATE SOLVENT
SALT and CONTAMINANTS WASHEDOFF BY THE DISTILLED WATER
WATER andCONTAMINANTS NOTRETAINED BY THE SPECOLUMN
Figure 28: Schematic representation of reversed phase extraction through SPEcartridge.
61
-Si
-Si
-Si
-Si
-Si
-Si
-Si
-Si
-Si
-Si
-Si
-Si
OH
OH
OH
CH3
CH3
CH3
CH3
OH
OH
CH3
CH3
CH3
Figure 29: Schematic representation of a silica particle’s surface carrying C18 hy-drocarbon chains (C18 SPE column).
the columns but a synthesised (2:1) imidazoline could not. It should be noted
that the synthesised imidazoline was a palmitic imidazoline with a terminal
hydroxyl in place of the amine in the pendant group (see Figures 2 and 3).
In this project different SPE columns were tested using solutions contain-
ing imidazolines or QUATs in either H2O, brine or produced water (known
not to contain the substances under investigation), prior to analysis of pro-
duced water samples from platforms selected for the project. The two types of
columns selected were C18 (Figure 29), C8 (Figure 30) and C18 end-capped
(Figure 31) silica based columns, that is silica particles with hydrocarbon
side chains of different length and different amount of silanol groups, along
with polystyrene (PS) based (Figure 32) polymer columns. The best results
were obtained on C18 and PS columns for QUATs and PS for imidazolines.
62
-Si
-Si
-Si
-Si
-Si
-Si
-Si
-Si
-Si
-Si
-Si
-Si
OH
OH
OH
C CH3
C CH3
CH3
CH3
C CH3
C CH3
C CH3
OH
OH
Figure 30: Schematic representation of a silica particle’s surface carrying C8 hy-drocarbon chains (C8 SPE column).
-Si
-Si
-Si
-Si
-Si
-Si
-Si
-Si
-Si
-Si
-Si
-Si
CH3
CH3
CH3
CH3
CH3
CH3
CH3
C O CH3
C O CH3
C O CH3
C O CH3
C O CH3
Figure 31: Schematic representation of a silica particle’s surface carrying C18 hy-drocarbon chains where hydroxyl are substituted with metoxy groups(C18 end-capped SPE column).
63
C
Figure 32: Polystyrene chain (PS SPE column).
Some of those results are reported below3:
spe C18 - 1ml (MIME project)
sample C16 QUAT (0.001 mg in MeOH:Brine)
conditioning MeOH, H2O
washing 10 ml H2O
elution 6 ml Acidified MeOH
recovery 100%
spe PS - 1ml (trial X31)
sample C16 QUAT (0.001 mg in MeOH:Brine)
conditioning MeOH, MeOH:H2O
washing* 10 ml H2O
elution 6 ml Acidified MeOH
recovery 66%
3washing* means that the column has been dried before being eluted; MeOH:H2O isMeOH:H2O=90:10; acidified MeOH is MeOH with HCl (2% vol); wsi is a water solubleimidazoline; pw is produced water known not to contain the compounds of interest.)
64
spe PS - 6ml (trial X54)
sample C16 QUAT (0.005 mg in Brine)
conditioning MeOH, MeOH:H2O, H2O
washing 5 ml H2O
elution 10 ml MeOH
recovery 100%
spe PS - 1ml (trial X55)
sample C16 QUAT (0.005 mg in Brine)
conditioning MeOH, MeOH:H2O, H2O
washing 1 ml H2O
elution 5 ml MeOH
recovery 100%
spe PS - 1ml (trial X12)
sample WSI (0.01 mg in MeOH:H2O) followed by 5 ml Brine
conditioning MeOH, MeOH:H2O
washing* 5 ml H2O
elution 5 ml CHCl3
recovery (2:1) WSI = 100% (1:1) WSI = 30%
65
spe PS - 1ml (trial X25)
sample WSI (0.01 mg in MeOH:H2O)
conditioning MeOH, MeOH:H2O
washing 5 ml H2O
elution 3 ml MeOH(ext 1) then 5 ml CHCl3 (ext 2)
recovery ext 1: (2:1) WSI = 82% ; (1:1) WSI = 0% ;
ext 2: (2:1) WSI = 18% ; (1:1) WSI = 6%
spe PS - 1ml (trial X26)
sample WSI (0.01 mg in MeOH:H2O) followed by 5ml Brine
conditioning MeOH, MeOH:H2O
washing 5 ml H2O
elution 3 ml MeOH(ext 1) then 5 ml CHCl3 (ext 2)
recovery ext 1: (2:1) WSI = 72% ; (1:1) WSI = 0% ;
ext 2: (2:1) WSI = 14% ; (1:1) WSI = 8%
spe C18E - 6ml (trial E21)
sample WSI (0.01 mg in MeOH:H2O) mixed to 5ml Brine
just before loading
66
conditioning MeOH, MeOH:H2O
washing* 20 ml H2O
elution 20 ml CHCl3
recovery (2:1) WSI = 12% ; (1:1) WSI = 65%
spe PS - 6ml (trial X30)
sample WSI (0.01 mg in MeOH:H2O) mixed to 5ml Brine
just before loading
conditioning MeOH, MeOH:H2O
washing 4 ml H2O
elution 10 ml MeOH:H2O(ext 1) then 10 ml CHCl3 (ext 2)
recovery ext 1: (2:1) WSI = 7% ; (1:1) WSI = 0% ;
ext 2: (2:1) WSI = 75% ; (1:1) WSI = 40%
spe PS - 1ml (trial X32)
sample WSI (0.01 mg in MeOH:H2O) mixed to 5ml pw
just before loading
conditioning MeOH, MeOH:H2O
washing 5 ml H2O
elution 5 ml MeOH
recovery (2:1) WSI = 65% ; (1:1) WSI = 7%
67
spe PS - 6ml (trial X34)
sample WSI (0.01 mg in MeOH:H2O) mixed to 5ml pw
just before loading
conditioning MeOH, MeOH:H2O
washing 10 ml H2O
elution 2.5 ml MeOH:H2O(ext 1) then 10 ml CHCl3 (ext 2)
recovery ext 1: (2:1) WSI = 11% ; (1:1) WSI = 0% ;
ext 2: (2:1) WSI = 83% ; (1:1) WSI = 33%
Overall the best results were obtained using polystyrene (PS) based poly-
mer columns (Strata X, Phenomenex), using MeOH to elute QUATs and
CHCl3 to displace imidazolines after washing off the salt with distilled H2O.
The extract was directly injected into the MS when in MeOH. The chloro-
form extract was dried under N2 and redissolved in MeOH:HCOOH (25mM
aqueous solution) (90:10).
Figures 33 and 34 show the chromatograms obtained from sample L and
Z extracts, respectively. In the examples shown, the methanol solutions were
directly injected into the MS. It should be noticed that the two produced wa-
ter samples were spiked with different amounts of D-standards and that they
were concentrated 3 and 2 times, respectively, during the SPE extraction.
In order to prove that the analytes of interest are indeed retained by the
SPE column, the produced water “waste” from sample Z was collected during
68
0 1 2 3 4 5
time (min)
Inte
nsity
C18 m/z=388.5
dC16 m/z=367.4
C16 m/z=360.4
dC14 m/z=339.3
C14 m/z=332.4
C12 m/z=304.5
area = 1,622,000
area = 777,000
area = 735,000
area = 289,000
area = 77,000
area = 62,000
Figure 33: Chromatogram obtained from sample L SPE extract (concentrated x3) via direct injection into the MS.
0 1 2 3 4 5
time (min)
Inte
nsity
C18 m/z=388.5
dC16 m/z=367.4
C16 m/z=360.4
dC14 m/z=339.3
C14 m/z=332.4
C12 m/z=304.5
area = 2,280,000
area = 998,000
area = 864,000
area = 572,000
area = 93,000
Figure 34: Chromatogram obtained from sample Z SPE extract (concentrated x2) via direct injection into the MS.
69
0 2 4 6 8 10 12
time (min)
Inte
nsit
y
0 2 4 6 8 10 12
enlarged 100 times
Figure 35: Chromatogram from the produced water sample Z eluted from the SPEcartridge during the loading process. The inset shows the same graph enlarged (yaxis) 100 times.
sample loading on the SPE cartridge. The collected waste was diluted with
MeOH(HCOOH 25mM) (50:50) before being injected into the MS.
Figure 35 shows that there is no signal from the QUATs or from the
standards, as expected (note that the scale for the y-axis is the same as in
Figure 34; only when the scale is enlarged 100 times the flat signals can be
seen on the chromatogram, as shown in the inset, confirming the absence of
peaks).
As for the DCM extracts, when the SPE extracts are passed through the
online HPLC system, a further cleaning and/or separation of the analytes
from the contaminants may take place giving greater areas. This is shown in
Figures 36 and 37 for sample L and Z, respectively.
70
0 1 2 3 4 5 6
time (min)
Inte
nsit
y
C18 m/z=388.5
dC16 m/z=367.4
C16 m/z=360.4
dC14 m/z=339.3
C14 m/z=332.4
C12 m/z=304.5
area = 2,889,000
area = 1,170,000
area = 914,000
area = 390,000
area = 199,000
area = 119,000
Figure 36: Chromatogram from sample L SPE extract via HPLC-MS analysis (seealso Figure 33).
0 50 100 150 200 250 300 350 400 450 500
time (min)
Inte
nsit
y
C18 m/z=388.5
dC16 m/z=367.4
C16 m/z=360.4
dC14 m/z=339.3
C14 m/z=332.4
C12 m/z=304.5
area = 4,890,000
area = 1,603,000
area = 1,602,000
area = 884,000
area = 137,000
Figure 37: Chromatogram from sample Z SPE extract via HPLC-MS analysis (seealso Figure 34).
71
A B
10µµµµlLOOP
MSP M
W
SA B
10µµµµlLOOP10µµµµlLOOP
MSMSP M
W
S
Figure 38: Schematic representation of the HPLC - MS configuration: A andB are the carrier solvent reservoirs, P and M the pre and main HPLC column,respectively, S is the switching valve that directs the flow either to waste (W) orto the MS.
4.3 HPLC-MS
For direct LC-MS/MS analysis, the HPLC cartridge was set up online to the
MS using a unique configuration developed for the project. Figure 38 shows
a schematic representation of the system. The switching valve S is turned to
waste (W) when the sample is injected in the 10 µl loop (flow 0.2 ml/min),
once it reaches the pre-column (P) the analytes and some other organic
compounds are retained while the salt is washed off with distilled water.
After 2 min from injection, the carrier solvent is changed to MeOH:HCOOH
(25mM aqueous solution) (90:10) and the valve S is set towards the MS. The
sample thus travels to the main column (M) and then to the MS cleaned of
salt and possibly other organics that may interfere with the MS detection of
QUATs and imidazolines.
As pointed out in the previous sections, the produced water samples ex-
tracted with DCM or cleaned up using SPE may be either injected directly
into the MS via a loop or through the online HPLC column.
72
0 1 2 3 4 5 6
time (min)
Inte
nsit
yC18 m/z=388.5
dC16 m/z=367.4
C16 m/z=360.4
dC14 m/z=339.3
C14 m/z=332.4
C12 m/z=304.5
area = 1,493,000
area = 878,000
area = 730,000
area = 620,000
area = 97,000
Figure 39: Chromatogram obtained from sample Z via the online HPLC-MS sys-tem.
The best results were obtained using a cyano column (Luna CN, Phe-
nomenex) for QUATs and a titanium oxide column for imidazolines (Titan-
sphere TiO, GL Sciences Inc., Japan - through Hichrom Ltd, UK), both
coupled with a cyano pre-column (Security Guard, Phenomenex).
An example of chromatogram obtained from produced water sample Z is
shown in Figure 39, others are presented in the following sections.
5 Produced Water Analysis
5.1 Calibration Curves
Prior to the analysis of the produced water samples, tests were carried
out on solutions containing known amount of actives in order to check the
extent of a possible linear relationship between peak area and concentration.
QUATs solutions were made up using benzalkonium QUATs purchased from
73
Aldrich in either brine, i.e. NaCl (35g/l) in distilled water, or produced
water from an offshore platform known not to contain any of the analytes
under investigation and believed to be relatively clean. For imidazolines
however, a suitable method was only recently developed via TiO HPLC
column, therefore the calibration curve is built on MeOH:H2O solutions,
directly injected in the MS. Also, as the commercial imidazolines provided
are always a mixture of (2:1), (1:1) imidazolines, amides and diamides, the
use of D(2:1) was preferred.
In order to take into account any fluctuation in the MS response and/or
any interference effects, the QUATs solutions were spiked with the deuterated
standards and the peak area was reported as relative intensity, i.e. the area
was divided by the area corresponding to a fixed amount of D-standard.
Unfortunately this was not possible at the time when the D(2:1) imidazoline
calibration curve was run, since D-QUATs were synthetised later during the
project.
Small fluctuations could be observed in the areas of the D-QUATs from
sample to sample but in general the response remained constant, as shown
in Figure 40. Note that the various volumetric flasks (numbered from 1 to
9, see x-axis in Figure 40) contain variable amount of hydrogenous QUATs
(from zero to ≈ 600 ppb) and ≈ 100 ppb of deuterated standards.
An example of the calibration curves obtained is shown in Figure 41: the
red and black lines correspond to the best linear fit for brine and produced
water solutions, respectively.
The solutions were passed through the HPLC column set up on-line to the
MS, including the washing step. The fact that the two lines almost coincide
74
0
1000000
2000000
3000000
4000000
5000000
6000000
0 1 2 3 4 5 6 7 8 9 10
Volumetr ic #
Are
a
d C16 Brine d C14 Brine
d C16 Produced Water d C14 Produced Water
dC14 104 ppbdC16 102 ppbH-Quats ~600 ppb
dC14 104 ppbdC16 102 ppbNo H-Quats
Figure 40: D-QUATs areas for volumetric flasks 1 to 9 containing different amountof QUATs and a fixed amount of d-QUATs.
C16 PW y = 0.0085x
R2 = 0.9939
C16 Briney = 0.0084x
R2 = 0.9872
0
1
2
3
4
5
6
0 100 200 300 400 500 600 700 800
Conc (ppb)
Are
a no
rmal
ised
to d
-std
C16 Brine
C16 Produced Water
Figure 41: C16 QUAT HPLC analysis from brine and produced water solutions.Peak areas are normalised to D-C16 QUAT area (102 ppb) and reported as afunction of concentration, see text for details.
75
0 200 400 600 800 1000 1200
Concentration (ppb)
Are
aMRM 618.7 > 308.9
Figure 42: Calibration curve for deuterated (2:1) imidazoline up to 1 ppm. Peakareas were measured for various solutions (from ≈ 7 ppb to ≈ 1 ppm) of D(2:1)imidazoline in MeOH:H2O= 90:10
indicates that either the produced water used is fairly clean from interfering
compounds or the HPLC column has cleaned up the sample, either through
the washing step or because of separation along the column. Nevertheless,
the relationship between area and concentration is linear up to 600 ppb.
No further measurements at higher concentrations were carried out simply
because the system would be overloaded with QUATs.
As for imidazolines, as already pointed out, the procedure was slightly dif-
ferent: D(2:1) imidazoline solutions were prepared up to≈ 1ppm in MeOH:H2O=
90:10 and directly injected in the MS. The resulting calibration curve is shown
in Figure 42, the relationship is believed to be linear up to around 500ppb.
Note that in both cases conventional ±5% error bars were added to the
experimental data. In order to take into account a possible change in the
slope of the normalised area - concentration linear relationship when using
76
other produced waters containing the chemicals of interest but with a possible
very different matrix, a direct comparison to the deuterated standards was
preferred to the use of the calibration curves for the calculations.
5.2 Preparations
Deuterated (D4) (2:1) oleic imidazoline and deuterated (D7) C12, C14 and
C16 quaternary ammonium salts (QUATs) are available for use as internal
standards.
The platforms, preparations and produced water (PW) samples will be
referred to as F, G, L, M, S, T, W and Z. The five ones (F, G, L, M and
Z) where QUATs are the main focus are described in Section 5.3, the other
three (S, T and W) where imidazolines are the component of interest are
outlined in Section 5.4. Note that preparations F and Z contains both QUATs
and imidazolines, however, when more than one component is relevant for
CHARM calculations, only the one with the highest fraction release default
value is used in the calculations. Therefore, QUATs were the focus of the
analysis of produced water samples F and Z.
The preparations used on the platforms were spiked with deuterated stan-
dards in order to relate the amount of substances detected in the produced
water sample to the dosing concentration of the preparation used on the same
platform. Note that to calculate C18 - QUAT in the preparations, the C18
peak was computed using D7 C16 QUAT, since its deuterated counterpart is
not available. For the same reason, (2:1) and (1:1) imidazolines, amides and
diamides were all compared to D4 - (2:1) oleic imidazoline. Table 8 shows the
77
percentage of QUATs and/or imidazolines contained in the relevant prepa-
rations, along with the amounts of the main contribution to each class of
actives.
As it will be shown in the following sections, sample F has been par-
ticularly difficult to analyse and the same problems seem to apply to the
preparation used, which is why is reported as the percentage corresponding
to the information obtained by the chemical supplier. Preparation F is rich
in C12 QUAT, ≈ 5% according to MS measurements, so that the signals
from the other QUATs seem to be influenced by it. Various solutions of
preparation F, from ≈ 5 to ≈ 50 ppm, were analysed in order to quantify
all the QUATs present, however, the results obtained were not consistent so
that the nominal concentration given by the chemical supplier is reported in
Table 8. It should also be noted that whenever we received the same prepa-
ration at different times, we found the ingredients to be present in slightly
different quantities, probably due to batch to batch variation, which is why
whenever possible we preferred to use the results of the MS analysis rather
than the composition supplied. However, that implies that the preparation
sample received comes from the same batch as the one used on the platform
when the PW samples were taken. Two examples are shown in Figure 43:
the vials contain two different preparations, supposedly the same for the two
vials next to each other but even their colour is clearly different.
78
Table 8: Composition of preparations from MS analysis, dosage rate and relevantnotes.
SAMPLE
PREPARATION
DOSAGE NOTESMASS - SPEC
QUATs IMIDAZOLINEs
F 6.5 % 8.1% 75 ppmpreparationcomposition givenby chemical supplier
G4.0 %
10 ppm →≈ 4.7 ppm
water stream dilutedwith PW beforesampling point
C12 = 2.8 % -
C14 = 1.2 %
L5.2 %
10 ppm
C18 compared toD − C16; PW storedin a tank for a fewdays
C18 = 3.5 % -
C16 = 1.5 %
M5.1 %
200 ppm →≈ 40 ppm
PW diluted withwater from anotherplatform; storagetank as for sample L
C12 = 3.4 % -
C14 = 1.5 %
S20.9 %
≈ 6.0 ppmpreparationcomposition givenby chemical supplier
- (2:1) IMIDAZ
DIAMIDES
T20.9 %
≈ 4.5 ppmpreparationcomposition givenby chemical supplier
- (2:1) IMIDAZ
DIAMIDES
W -
20.1 %
N.A.samples taken beforeand after cleaningprocess
(2:1) IMIDAZ = 12.7 %
(1:1) IMIDAZ = 6.0 %
DIAMIDES = 1.3 %
Z3.8 %
30 ppm PW is reinjectedC12 = 2.8 % IMIDAZ 1 - 5%
C14 = 0.9 % according to MSDS
79
Figure 43: Preparation samples: the two vials on the right contain preparation Fand the two on the left preparation J
5.3 Quats
Depending on the specific sample, i.e. on the specific platform in terms of
chemicals used, oil extracted etc., the so-called matrix effect can be totally
different. Consequently the response to the different techniques may vary
greatly from sample to sample.
sample F
Before spiking the PW samples, preliminary tests are usually run on the
PW in order to eventually add the right amount of deuterated standards to
the containers. In the case of sample F that stage was never reached for two
reason:
1. the two glass bottles broke when in the freezer; they were put into
beakers when defrosted, therefore giving extra surface for the corrosion
inhibitors to stick on
80
0
20000000
40000000
60000000
80000000
100000000
120000000
140000000
0 2 4 6 8 10 12 14 16
time (min)
Inte
nsit
y
C12 overloaded
C14 overloaded
PW diluted 1:2 + 205 ppb dC16 + 208 ppb dC14
Figure 44: MS/MS chromatogram for sample F produced water diluted 1:2 throughHPLC on-line system for salt removal.
2. the preliminary analysis was inconclusive, as it will be outlined below
As usual a first set of measurements are carried out on the PW diluted 1:2,
but, as you can see in Figure 44, the sample is clearly overloaded by both
C12 and C14. Only when the chromatogram is enlarged 140 times, then
the deuterated standards’ signals are visible, as shown in Figure 45. The
sample was then diluted 10 times, as shown in Figure 46: although C12 is
still clearly overloaded, the signal from C14 and D7C14 have increased rather
then decresead. That could either indicate an interference from the great
amount of C12 or from some other compounds present in the PW that are
obviously been diluted too. The sample was therefore diluted up to 1:1000,
other examples are shown in Figures 47 and 48 where the PW was diluted
1:25 and 1:370, respectively. Although in Figure 47 the peaks’ areas seem
to indicate that the sample is diluted enough, the signals do not decrease at
all, again areas increase (note that D7C12 area does not change much but it
corresponds to a lower amount of deuterated standard). Figure 49 reports
81
0 2 4 6 8 10 12 14 16
time (min)
Inte
nsit
y
dC16 area = 683,000
d C14 area = 730,000
PW diluted 1:2 + 205 ppb dC16 + 208 ppb dC14
C16 area = 730,000
Figure 45: MS/MS chromatogram for sample F produced water diluted 1:2 throughHPLC on-line system for salt removal, as shown in Figure 44 with y axis enlargedin order to see the deuterated standards signals.
0 2 4 6 8 10 12
time (min)
Inte
nsit
y
C12 overloaded
dC14 area = 1,746,000
C14 area = 2,146,000
PW diluted 1:10 + 208 ppb dC14
Figure 46: MS/MS chromatogram for sample F produced water diluted 1:10through HPLC on-line system for salt removal.
82
0 2 4 6 8 10 12
time (min)
Inte
nsit
y
C12 area = 1,686,000
dC16 area = 1,233,000
C14 area = 175,000
dC12 area = 2,825,000
PW diluted 1:25 + 103ppb dC16 + 312 ppb dC12
Figure 47: MS/MS chromatogram for sample F produced water diluted 1:25through HPLC on-line system for salt removal.
0 1 2 3 4 5 6 7 8 9
time (min)
Inte
nsit
y
C12 area = 1,903,000
C14 area = 730,000
dC12 area = 2,319,000PW diluted 1:370 + 260 ppb dC12
Figure 48: MS/MS chromatogram for sample F produced water diluted 1:370through HPLC on-line system for salt removal.
83
0
3000
6000
0.00 0.05 0.10
1 / DILUTION
C12
ppb 1 : 250 1 : 50 1 : 25
1 : 10
Figure 49: C12 calculated amount in ppb from solutions at variuos dilution asa function of 1/dilution. Note that the line through the data points is not aninterpolation but simply a guide to the eye.
the results from the various measurements at different dilution, as a funtion
of inverse dilution, which means that the C12 concentration increases along
the x-axis from left to right. Three regions can be distinguished:
1. the response seems to be linear for low C12 concentration, corresponding
to PW solutions diluted at least 1:360, as shown in details in Figure
50. Peak areas typical values in this region can be up to a few millions.
2. This is followed by a non linear region for PW samples diluted less than
1:250 where the MS is overloaded.
3. With the C12 concentration further increasing, for example the PW is
diluted only 1:10, the MS is flooded and the peak area value is over 34
millions!
The same kind of behaviour is found for C14 QUAT although on a smaller
scale: as shown in Figure 51 a dilution of at least 1 : 20 is necessary to obtain
84
0
250
500
0 0.001 0.002 0.003 0.004 0.005
1 / DILUTION
C12
ppb 1 : 360
1 : 300
1 : 250
Figure 50: C12 calculated amount in ppb from solutions at variuos dilution as afunction of 1/dilution. In this case only the linear region is shown.
reliable results (see also the C14 signals in the chromatograms from Figure
44 to 48). Since the amount of C12 seems to be so much compared to any
other QUATs, the amount of preparation can be calculated simply using C12
and the fact that C12 equals ≈ 5% of the preparation, according to the MS
measurement of preparation F. If only the measurements carried out on PW
diluted at least 1:360 can be considered reliable, then we obtain the results
reported in Table 9 for sample F analysed via HPLC. It is not surprising that
there is quite a big variation in the results, C12 QUAT quantity goes from 68
to 82 ppm. That is simply a consequence of such a high dilution that implies
multiplying for a high factor in the calculations.
sample G
Two samples of about 250 ml were received in glass containers. Sample
G duplicates were analysed using all the three techniques available. HPLC
analysis gave a very poor response, probably due to low levels of QUATs
85
0
200
400
600
800
0.00 0.05 0.10
1 / DILUTION
C14
ppb
1 : 1001 : 50
1 : 25
1 : 10
1 : 15
Figure 51: C14 calculated amount in ppb from solutions at variuos dilution as afunction of 1/dilution.
Table 9: HPLC analysis results from various trials on diluted PW samples fromSample F.
DILUTION D7C12 (ppb) C12 (ppb meas) C12 (ppm calc) PREP. F (ppm)
357 520 207 74 1473
370 260 213 79 1576
417 520 174 73 1447
454 520 167 76 1513
500 520 136 68 1357
500 250 163 82 1628
1000 250 78 68 1560
86
0 2 4 6 8 10 12
time (min)
Inte
nsit
y
dC14 area = 833,000
C12 area = 176,000
C14 area = 36,000
dC12 area = 747,000 PW diluted 1:1.25 + 104ppb dC14 + 104 ppb dC12
Figure 52: MS/MS chromatogram for sample G produced water diluted 1:1.25through HPLC on-line system for salt removal.
rather than interference. As shown in Figure 52, the deuterated internal
standards response was in fact not too low. Note that the produced water
sample was diluted 80:20 rather than 50:50 in an attempt to improve such low
signals from the hydrogenous QUATs. This early conclusion was supported
by the DCM extraction and SPE results: these techniques allowed us to
clean and concentrate the sample, leading to a more reliable answer. Figure
53 shows the same sample DCM extracted and concentrated 5 times, the
signal clearly improved compared to the HPLC chromatogram. The same
can be said for the SPE analysis, again concentrating the extract 5 times
(Figure 54, see also Table 10). A further “cleaning” of the signals may be
obtained coupling DCM or SPE extraction with HPLC, whenever separation
occurs through the HPLC column, i.e. a further cleaning of the analytes
from possible contaminants may take place via chromatography, improving
the signals as shown in Figure 55, when the SPE extract from sample G is
passed through the online HPLC system.
87
0 0.5 1 1.5 2 2.5 3 3.5 4 4.5
time (min)
Inte
nsit
y
dC14 area = 3,304,000
C12 area = 823,000
C14 area = 402,000
dC12 area = 3,024,000
DCM conc x 5+ 260 ppb dC14 + 260 ppb dC12
Figure 53: MS/MS chromatogram for sample G produced water DCM extract,concentrated 5 times, direct injection into the MS.
0 0.5 1 1.5 2 2.5 3 3.5 4 4.5
time (min)
Inte
nsit
y
dC14 area = 3,425,000
C12 area = 752,000
C14 area = 308,000
dC12 area = 2,907,000SPE conc x 5
260 ppb dC14 260 ppb dC12
Figure 54: MS/MS chromatogram for sample G produced water SPE extract,concentrated 5 times, direct injection into the MS.
88
0 2 4 6 8 10 12 14
time (min)
Inte
nsit
ydC14 area = 4,403,000
C12 area = 1,228,000
C14 area = 381,000
dC12 area = 4,088,000SPE + HPLC
conc x 5260 ppb dC14 260 ppb dC12
Figure 55: MS/MS chromatogram for sample G produced water SPE extract,concentrated 5 times, injected through the HPLC on-line system into the MS.
sample L
Two produced water samples were received, one of the sample was exten-
sively used for preliminary analysis and various trials, while the other one
was kept frozen. These samples in fact were received at an early stage of
the project, when the different techniques were being throughly investigated,
therefore the numeric results reported on Table 10 are based on the second
sample, analysed once the preliminary trials were considered satisfactory.
Samples L gave poor response with both HPLC and DCM extraction. The
HPLC-MS response was so low that the same sample was injected into the
MS using a 50µl loop, that is simply increasing the amount of sample actu-
ally entering the MS. This was done because the signals from the deuterated
standards were greatly reduced, indicating interference effects. Increasing the
loop capacity, it was possible to check that hydrogenous QUATs were indeed
present. Figures 56 and 57 show the chromatograms obtained using a 10 and
50 µl loop, respectively. In Figure 25 the DCM extract chromatogram was
89
0 2 4 6 8 10 12 14 16time (min)
Inte
nsit
yC18 m/z=388.5
dC16 m/z=367.4
C16 m/z=360.4
dC14 m/z=339.3
C14 m/z=332.4
C12 m/z=304.5
area = 367,000
area = 166,000
area = 23,000
Figure 56: MS/MS chromatogram for sample L produced water diluted 1:2 withMeOHthrough HPLC on-line system.
shown; coupling the two techniques improved the response but not enough
to ensure a reliable outcome (Figure 26). Note that with the DCM technique
the sample was concentrated 5 times, however, the signals did not increase ac-
cordingly, indicating that this technique did not help cleaning this particular
sample from other organic compounds that are causing a lot of interference.
In conclusion, for sample L only solid phase extraction gave a good enough
chromatogram, as shown in Figure 33. Obviously then, this is the technique
of choice in this case, where “cleaning” of the sample seems necessary and
achieved using solid phase extraction. The results were confirmed by cou-
pling SPE and HPLC, when the signals improved but the overall answer
remained the same, indicating a further separation through the HPLC col-
umn between contaminants and analytes but also a very efficient cleaning
step for this sample via SPE.
90
0 1 2 3 4 5 6 7time (min)
Inte
nsit
yC18 m/z=388.5
dC16 m/z=367.4
C16 m/z=360.4
dC14 m/z=339.3
C14 m/z=332.4
C12 m/z=304.5
area = 11,388,000
area = 5,075,000
area = 136,000
area = 257,000
area = 418,000
area = 126,000
Figure 57: MS/MS chromatogram for sample L produced water diluted 1:2 withMeOHthrough HPLC on-line system.
sample M
Two duplicates were received in plastic bottles. In this case the HPLC
technique not only gave a good result, but the amount so calculated was
confirmed by SPE. Examples are shown in Figures 58 and 59. The washing
step in the on-line HPLC system and the solid phase extraction both managed
to properly “clean” the sample, or this particular produced water sample is
not particularly affected by contaminants. DCM extraction was not carried
out on these samples.
sample Z
Duplicates were received from platform Z in metal tins. As for sample L, the
first sample was throughly used to test the techniques while the numerical
results reported in Table 10 are based on the analysis of its duplicate. In this
case the HPLC technique gave good results, as shown in Figure 39. However,
the sample gave a poor response when analysed via DCM extraction, in
91
0 2 4 6 8 10 12 14
time (min)
Inte
nsit
y
C12 area = 3,158,000
dC14 area = 1,320,000
C14 area = 292,000
dC12 area = 3,309,000
PW diluted 1:2 + 104ppb dC14 + 312 ppb dC12
Figure 58: MS/MS chromatogram for sample M produced water diluted 1:2 withMeOH:HCOOHinjected through the HPLC into the MS.
0 0.5 1 1.5 2 2.5 3 3.5 4
time (min)
Inte
nsit
y
C12 area = 1,978,000
dC14 area = 570,000
C14 area = 216,000
dC12 area = 1,942,000
SPE conc x 2 + 104ppb dC14 + 520ppb dC12
Figure 59: MS/MS chromatogram for sample M produced water SPE extract,concentrated 2 times, direct injection into the MS.
92
particular, although the sample was concentrated 2.5 times, the deuterated
standards signals where not present, indicating a large interference effect
(see Figure 27). On the other hand, the SPE procedure confirmed the HPLC
results, effectively cleaning the sample from the contaminants, as shown in
Figures 34 and 37.
A summary of the results obtained for all the samples containing QUATs,
is reported in Table 10. Note that calculations are not reported in Table 10
for sample Z analysis by SPE coupled with HPLC, this is due to the fact that
the increase in the peak area of C12 is such that it could be outside the linear
relationship range. Also not reported is DCM coupled HPLC for sample L
simply because the chromatogram did not improve enough compared to DCM
extract analysis and the response was still too low.
5.4 Imidazolines
Compared to QUATs, many more trials were necessary before a suitable
analytical protocol could be established for imidazolines. In particular, SPE
recoveries can be very low, depending on the sample matrix and HPLC re-
tention times very long, up to 30-40 min for (1:1) imidazolines, depending on
the column chosen. Times were eventually shortened using a TiO column
(retention times ≈ 5 min for (1:1) imidazolines). This was coupled with the
cyano pre-column for the salt removal (note that the pre-column has no effect
on the retention times).
93
0
100000
200000
300000
400000
500000
0 2 4 6 8 10 12 14
time (min)
Inte
nsi
ty(1:1)IMIDAZ @ 346.6
(1:1)IMIDAZ @ 348.5
(1:1)IMIDAZ @ 350.6
area = 634,000
area = 297,000
area = 371,000
Figure 60: MS/MS chromatogram for sample W3 produced water, diluted 1:2,HPLC on-line analysis: (1:1) imidazolines.
sample W
13 samples (in glass bottles) were received from platform W, of which 8
have been analysed. The samples are referred to as W1, W2 and so on.
HPLC analysis gave good results for these samples, some of them were also
investigated via DCM and SPE extraction, however these techniques gave
poor results. As an example, sample W3 chromatograms are reported in
Figure 60 to 62. Although the corresponding preparation contains about
12.6% of (2:1) imidazolines, 1.3% of the corresponding diamides and 6.0%
of (1:1) imidazolines, all the samples analysed resulted to contain only (1:1)
imidazolines.
samples S and T
Since samples S and T come from different platforms but the same field and
the same preparation is used, they are described here togheter. They were
received as duplicates in plastic bottles. In this case the preparation con-
94
0
40000
80000
120000
0 2 4 6 8 10 12 14
time (min)
Inte
nsity
AMIDE @ 364.5
AMIDE @ 366.5
AMIDE @ 368.5
DIAMIDE @ 626.6
DIAMIDE @ 628.7
DIAMIDE @ 630.7
DIAMIDE @ 632.5
Figure 61: MS/MS chromatogram for sample W3 produced water, diluted 1:2,HPLC on-line analysis: amides and diamides.
0
500000
1000000
1500000
2000000
2500000
0 2 4 6 8 10 12 14 16 18
time (min)
Inte
nsit
y
D(2:1) IMIDAZOLINE (500ppb) @ 618.7
(2:1) IMIDAZOLINE @ 608.7
(2:1) IMIDAZOLINE @ 610.6
(2:1) IMIDAZOLINE @ 612.7
(2:1) IMIDAZOLINE @ 614.7
area = 501,000
Figure 62: MS/MS chromatogram for sample W3 produced water, diluted 1:2,HPLC on-line analysis: (2:1) imidazolines.
95
0
50000
100000
0 2 4 6 8 10
time (min)
Inte
nsity
DIAMIDE @ 630.6 DIAMIDE @ 628.7
DIAMIDE @ 626.7 DIAMIDE @ 624.8
Figure 63: MS/MS chromatogram for sample S produced water, diluted 1:2, HPLCon-line analysis: diamides.
tained (2:1) imidazolines and corresponding diamides but no (1:1) imidazo-
lines. However, samples S and T gave no signal for any of the amides/imida-
zolines present in the preparation, as shown in Figure 63 to 65. Note that
on these two samples DCM extraction did not work as well as for other sam-
ples, especially sample S did not separate properly. This may indicate the
presence of other chemicals that enhance water miscibility of hydrophobic
compounds (surfactants) including DCM itself. Also these samples gave an
unidentified white precipitate when mixed with MeOH.
Imidazolines behaviour over time
One of the two duplicates of samples S and T, along with six samples coming
from platform W, were spiked with D(2:1) imidazoline and their behaviour
was followed over time via HPLC analysis. To samples S and T were added
1ppm of D(2:1) imidazoline while samples W were spiked with 500ppb. Since
their behaviour turned out to be quite similar, samples S and T analysis will
96
0
1000000
2000000
3000000
4000000
5000000
0 2 4 6 8 10
time (min)
Inte
nsit
y
IMIDAZ @ 614.6
IMIDAZ @ 612.8
IMIDAZ @ 610.8
IMIDAZ @ 608.6
D(21) IMIDAZ @ 618.7
0
60000
120000
0 2 4 6 8 10
area = 1,580,000
Figure 64: MS/MS chromatogram for sample S produced water, diluted 1:2, HPLCon-line analysis: (2:1) imidazolines.
0
1000000
2000000
3000000
4000000
5000000
0 2 4 6 8 10
time (min)
Inte
nsit
y
IMIDAZ @ 614.6
IMIDAZ @ 612.8
IMIDAZ @ 610.8
IMIDAZ @ 608.6
D(21) IMIDAZ @ 618.7
0
60000
120000
0 2 4 6 8 10
area = 1,104,000
Figure 65: MS/MS chromatogram for sample T produced water, diluted 1:2, HPLCon-line analysis: (2:1) imidazolines.
97
be outlined first:
? the produced water was analysed via HPLC over a 7 days period, each
time a sample was taken out of the original container and diluted 1:2
with MeOH:HCOOH (25mM). As described in previous reports, this
dilution gives the best response from PW analysed via HPLC.
? diamides/imidazolines signals were always at background level, i.e. be-
low ≈ 20ppb/40ppb. At these low levels, the signal is very noisy and
its outcome is therefore affected by a large error. For example the
highest signal recorded (inset in Figure 66) corresponds to just 18ppb
of imidazoline!
? D(2:1) imidazoline signal is constant (see Figure 67 or compare Fig-
ure 64 to Figure 66). This indicates no adsorption and/or exchange
with hydrogenous amides/imidazolines on the container walls (plastic
in these cases). In fact after 11 days the (emptied) containers were
extracted with DCM releasing back only a small amount of deuterated
imidazoline (see for example sample S DCM bottle extraction result-
ing chromatogram in Figure 68 and Figure 64 for comparison). To
compute the D(2:1) signal, the peak area needs to be compared to an
external standard, which leaves us with two options: the area is com-
pared to the area of a known quantity of D(2:1) in MeOH:HCOOH
or to the response of the deuterated standard in the produced water.
Since the extract is redissolved in MeOH:HCOOH and should be rel-
atively “clean” compared to the produced water, the first method is
preferred. The calculations showed that in 11 days only 48ppb and
98
0
1000000
2000000
3000000
4000000
5000000
0 2 4 6 8 10
time (min)
Inte
nsi
tyIMIDAZ @ 614.6
IMIDAZ @ 612.8
IMIDAZ @ 610.8
IMIDAZ @ 608.6
D(21) IMIDAZ @ 618.7
0
60000
120000
0 2 4 6 8 10
area = 1,570,000
Figure 66: MS/MS chromatogram for sample S, HPLC on-line analysis: (2:1)imidazolines after 7 days from spiking, in the inset the highest response from (2:1)imidazoline at m/z = 614.6, see text for details.
58ppb of D(2:1) have deposited on the containers out of ≈ 1ppm.
Six different samples coming from platform W were analysed over time: sam-
ples W1, W2 and W3 behaved in a similar way, different though from samples
W4, W5 and W6, again similar to each others:
? all samples were followed over a period of at least 7 days after been
added the deuterated standard;
? although preparation W contains both (2:1) and (1:1) imidazolines,
only (1:1) imidazolines can be detected in all the produced water sam-
ples;
? in samples W1, W2 and W3 the deuterated standard signal quickly de-
creases (2-3 days) (Figure 69) indicating absorption to glassware rather
than exchange since the hydrogenous diamides and (2:1) imidazolines
signals do not increase and the (1:1) imidazolines signals remain con-
stant (Figure 70). Note that while the exchanging between hydrogenous
99
0.000
0.002
0.004
0.006
0.008
0.010
0 1 2 3 4 5 6 7 8
time (days)
area
per
ppb
S T
Figure 67: Normalised peak area for D(2:1) imidazoline in samples S and T as afunction of time over 7 days from spiking.
0
1000000
2000000
0 2 4 6 8 10
time (min)
Inte
nsit
y
IMIDAZ @ 614.6
IMIDAZ @ 612.8
IMIDAZ @ 610.8
IMIDAZ @ 608.6
D(21) IMIDAZ @ 618.7
DIAMIDE @ 630.6
DIAMIDE @ 628.7
DIAMIDE @ 626.7
DIAMIDE @ 624.8
area = 282,000
Figure 68: MS/MS chromatogram for sample S DCM bottle extraction, HPLCon-line analysis.
100
and deuterated counterparts of the same molecule is expected until
equilibrium is reahced, the same cannot be said about the exchange
between molecules such as (2:1) and (1:1) imidazolines. These are in
fact part of the same family but at the same time different molecules.
? for samples W4, W5 and W6 the behaviour over time is completely dif-
ferent: the peak area of D(2:1) imidazoline is constant, i.e. there is no
absorption and/or exchange (Figure 71). Note that for these three sam-
ples, each measurement was carried out on samples spiked with 205ppb
of D-C16 QUAT just before injection, in order to take into account MS
efficiency. This was done in view of the possible D(2:1) absorption, ob-
served in the first three samples, that makes the deuterated imidazoline
useless in taking the variable MS efficiency into account.
? If we take the very first results as the most reliable for samples W1, W2
and W3, i.e. before the internal standard started to adsorb, then the
preparation present in the samples goes from 6ppm of W4 to 12ppm of
W5.
? Note that SPE and DCM gave poor results in all cases, DCM extraction
on the bottles gave generally poor results, and (1:1) imidazoline seem
to stay in the water layer.
6 Results Summary
A summary of QUATs and imidazolines measurements using the various
techniques on all the samples is reported in Table 10, along with the cal-
101
0
500
1000
1500
2000
2500
3000
0 2 4 6 8 10 12 14 16
time (days)
norm
are
a
Figure 69: Normalised peak area for D(2:1) imidazoline in sample W1 as a functionof time over 12 days from spiking.
0
500000
1000000
1500000
2000000
2500000
0 2 4 6 8 10 12 14
time (days)
area
(11) @ 346 (11) @ 348(11) @ 350
Figure 70: Peak areas for (1:1) imidazolines in sample W2 as a function of timeover 12 days from spiking.
102
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0 1 2 3 4 5 6 7 8
time (days)
nor
m a
rea
Figure 71: Normalised peak area for D(2:1) imidazoline in sample W6 as a functionof time over 7 days from spiking.
culated equivalent of preparations. Whenever possible, all duplicates were
analysed, this was not the case for samples S, T and Z. Apart from those, if
only a single result is reported, the duplicates simply gave the same answer.
Although the single QUATs or amides/imidazolines are quantified by
comparison to their deuterated counterparts (unless stated), the correspond-
ing amount of preparation is calculated from the sum of QUATs or amides/imidazolines
measured. This is compared to their overall percentage in the preparation,
as measured by MS analysis and reported in Table 8.
Finally, Table 11 reports the CHARM predicted concentration of the
preparations in produced water, C ′pw, calculated as follows:
C ′pw = Cpw + 0.1Cpw (1)
C ′pw is the theoretical fraction released Cpw plus a safety factor that equals
10% of its value. Cpw is calculated as follows:
103
Cpw =fr · Ci · Fi
Fpw
(2)
where fr is the fraction released, default value for QUATs is 1, for imidazoline
0.1; Ci is the concentration of surfactants in total fluid, i.e. the dosage
rate (ppm); Fi is the total fluid production (m3/d) and Fpw is the total
water production (m3/d). Measured quantities are reported both as ppm of
preparation (see also Table 10) and as percentage of the predicted CHARM
values.
104
Table 10: Results obtained with the various techniques on the PW samples con-taining QUATs or amides/imidazolines (column 3).
SAMPLE METHODACTIVES PREPAR.
NOTES(sum in ppb) (ppm)
G
HPLC - - poor response
SPE 17 - 18 0.4 - 0.5 } PW concentrated x 5SPE + HPLC 16 - 20 0.4 - 0.5
DCM 16 - 20 0.4 - 0.5 } PW concentrated x 5DCM + HPLC 17 - 21 0.4 - 0.5
LHPLC - - no signals
SPE 118 - 131 2.3 - 2.5 } PW concentrated x 3SPE + HPLC 118 2.3
DCM - - Poor response
M HPLC 602 - 642 12
SPE 557 - 560 11
SHPLC 97 (0.5) background levels
SPE - - PW conc x 5
DCM - - Poor separation
THPLC 72 (0.3) background levels
SPE - - PW conc x 2.5
DCM - -
WHPLC 1126 - 2512 6 - 12 very low signals
SPE - - } poor resultsDCM - -
Z HPLC 517 14
SPE 637 17 PW conc x 2
105
Table 11: Comparison between measured concentration and CHARM predictionsfor samples containing QUATs or imidazolines.
SAMPLE
Ci C ′pw Cm % CHARM
EFFECTIVE CHARMMEASURED
PREDICTED
DOSAGE PREDICTION VALUE
(ppm) (ppm) (ppm) 100 · Cm/Cpw
G 4.7 7 0.4 - 0.5 6 - 8 %
L 10 12 2.3 - 2.5 21 - 23 %
M 40 48 11 - 12 25 - 28 %
S 6 1.8 0.5 28 %
T 4.5 1.1 0.3 27 %
W ? ? 6 - 12 ? %
Z 30 47.9 14 - 17 32 - 39 %
106
References
[1] A. J. McMahon. Colloid Surface, 59, (1991).
[2] M. Joseph, J. Hodge, and D. Klenerman. Faraday Discuss., 94, (1992).
[3] A. Edwards, C. Osborne, S. Webster, D. Klenerman, M. Joseph, P. Os-
tovar, and M. Doyle. Corros. Sci., 36, (1994).
[4] D. A. Raval and V. M. Mannari. Res. and Ind., 39: 94, (1994).
[5] V. Jovancicevic, S. Ramachandran, and P. Prince. Corrosion, 55: 449,
(1999).
[6] D. Wang, S. Li, Y. Ying, M. Wang, H. Xiao, and Z. Chen. Corros. Sci.,
41: 1911, (1999).
[7] Y. Chen, T. Hong, M. Gopal, and W. P. Jepson. Corros. Sci., 42: 979,
(2000).
[8] A. J. Szyprowski. Brit. Corros. J., 35: 155, (2000).
[9] S. Ramachandran, B. L. Tsai, M. Blanco, H. Chen, Y. Tang, and
W. A. Goddard III. Langmuir, 12: 6419, (1996).
107
[10] S. Ramachandran, B. L. Tsai, M. Blanco, H. Chen, Y. Tang, and
W. A. Goddard III. J. Chem. Phys. A, 101: 83, (1997).
[11] S. Ramachandran and V. Jovancicevic. Corrosion, 55: 259, (1999).
[12] M. A. San-Miguel and P. M. Rodger. J. Mol. Struc. Theochem., 506:
263, (2000).
[13] M. A. Gough and G. J. Langley. Rapid Commun. Mass Spectrom., 13:
227, (1999).
[14] S. J. W Grigson, A. Wilkinson, P. Johnson, C. F. Moffat, and A. D.
McIntosh. Rapid Commun. Mass Spectrom., 14: 2210, (2000).
[15] P. McCormack, P. Jones, M. J. Hetheridge, and S. J. Rowland. Wat.
Res., 35: 3567, (2001).
[16] P. McCormack, P. Jones, and S. J. Rowland. Rapid Commun. Mass
Spectrom., 16, (2002).
108