Chapter 1 Introduction to Spectroscopyndl.ethernet.edu.et/bitstream/123456789/78709/7... · 2020....

417
Chapter 1 Introduction to Spectroscopy Ephrem Tilahun (MSc., Ass.Prof) Jimma University 1 E.T

Transcript of Chapter 1 Introduction to Spectroscopyndl.ethernet.edu.et/bitstream/123456789/78709/7... · 2020....

  • Chapter 1

    Introduction to Spectroscopy

    Ephrem Tilahun (MSc., Ass.Prof)

    Jimma University

    1 E.T

  • • Spectroscopy is the study of the interaction between radiation (electromagnetic radiation or light, as well as particle radiation) and matter.

    • Spectrometry is the measurement of these interactions

    • instrument which performs such measurement

    – spectrometer or

    – spectrograph.

    • A plot of the interaction is referred to as a

    – spectrum

    2 E.T

  • Classification of spectroscopic methods

    • Nature of radiation measured

    • The type of spectroscopy depends on the physical quantity measured.

    • Normally, the quantity that is measured is an amount or intensity of something.

    • Electromagnetic spectroscopy involves interactions with electromagnetic radiation, or light. – Ultraviolet-visible spectroscopy is an example.

    • Electronic spectroscopy involves interactions with electron beams. – Auger spectroscopy involves inducing the Auger effect with an electron

    beam.

    • Mechanical spectroscopy involves interactions with macroscopic vibrations, such as phonons. – An example is acoustic spectroscopy, involving sound waves.

    • Mass spectroscopy involves the interaction of charged species with magnetic and/or electric fields, giving rise to a mass spectrum.

    3 E.T

    http://encyclopedia.thefreedictionary.com/Electromagnetic+spectroscopyhttp://encyclopedia.thefreedictionary.com/electron+beamhttp://encyclopedia.thefreedictionary.com/Auger+spectroscopyhttp://encyclopedia.thefreedictionary.com/Auger+effecthttp://encyclopedia.thefreedictionary.com/phononshttp://encyclopedia.thefreedictionary.com/acoustic+spectroscopyhttp://encyclopedia.thefreedictionary.com/sound+waveshttp://encyclopedia.thefreedictionary.com/Mass+spectroscopyhttp://encyclopedia.thefreedictionary.com/Ion

  • • Measurement process

    • Most spectroscopic methods are differentiated as either atomic or molecular

    based on whether or not they apply to atoms or molecules.

    • Along with that distinction, they can be classified on the nature of their

    interaction:

    • Absorption spectroscopy uses the range of the electromagnetic spectra in

    which a substance absorbs.

    – atomic absorption spectroscopy and

    – various molecular techniques, such as infrared spectroscopy in that region a

    – nuclear magnetic resonance (NMR) spectroscopy in the radio region.

    • Emission spectroscopy uses the range of electromagnetic spectra in which a

    substance radiates (emits).

    – luminescence.

    – Molecular luminescence techniques include spectrofluorimetry.

    • Scattering spectroscopy measures the amount of light that a substance

    scatters at certain wavelengths, incident angles, and polarization angles.

    – The scattering process is much faster than the absorption/emission process.

    • Raman spectroscopy

    4 E.T

    http://encyclopedia.thefreedictionary.com/atomichttp://encyclopedia.thefreedictionary.com/molecularhttp://encyclopedia.thefreedictionary.com/Absorption+spectroscopyhttp://encyclopedia.thefreedictionary.com/atomic+absorption+spectroscopyhttp://encyclopedia.thefreedictionary.com/infrared+spectroscopyhttp://encyclopedia.thefreedictionary.com/nuclear+magnetic+resonance+spectroscopyhttp://encyclopedia.thefreedictionary.com/Radio+frequencyhttp://encyclopedia.thefreedictionary.com/Emission+spectroscopyhttp://encyclopedia.thefreedictionary.com/luminescencehttp://encyclopedia.thefreedictionary.com/spectrofluorimetryhttp://encyclopedia.thefreedictionary.com/Raman+spectroscopy

  • Electromagnetic radiation

    • EMR, or light, is a form of energy whose behaviour is described by the properties of both waves and particles.

    • The optical properties of electromagnetic radiation, such as diffraction, are explained best by describing light as a wave.

    • Many of the interactions between electromagnetic radiation and matter, such as absorption and emission,

    – however, are better described by treating light as a particle, or photon.

    • the dual models of wave and particle behaviour provide a useful description for electromagnetic radiation

    5 E.T

  • • Wave Properties of Electromagnetic Radiation

    • Electromagnetic radiation consists of oscillating electric and magnetic fields that propagate through space along a linear path and with a constant velocity

    • In a vacuum, electromagnetic radiation travels at the speed of light, c, which is 2.99792 x108 m/s.

    • Electromagnetic radiation moves through a medium other than a vacuum with a velocity, v, less than that of the speed of light in a vacuum.

    • Oscillations in the electric and magnetic fields are perpendicular to each other, and to the direction of the wave’s propagation.

    6 E.T

  • Figure shows an example of plane-polarized electromagnetic radiation consisting of

    an oscillating electric field and an oscillating magnetic field, each of which is

    constrained to a single plane. 7 E.T

  • • The interaction of electromagnetic radiation with matter can be

    explained using either the electric field or the magnetic field.

    • For this reason, only the electric field component is shown in

    Figure

    8 E.T

  • • The oscillating electric field is described by a sine wave of the form

    E = Ae sin(2πνt + Φ)

    – where E is the magnitude of the electric field at time t, Ae is the electric field’s

    maximum amplitude,

    – v is the frequency, or the number of oscillations in the electric field per unit time, and

    – Φ is a phase angle accounting for the fact that the electric field’s magnitude need not be zero at t = 0.

    • An identical equation can be written for the magnetic field, M

    M = Am sin(2πνt + Φ)

    – where Am is the magnetic field’s maximum amplitude.

    • only' the electrical vector is employed because it is the electrical force that is responsible for such phenomena as transmission, reflection, refraction, and absorption of radiation.

    • Magnetic responsible for the absorption of radio-frequency waves in NMR

    9 E.T

  • Wave Parameters

    • The time required for the passage of successive maxima through a fixed point in space is called the period (p )of the radiation.

    • The frequency (v) is the number of oscillations of the field that occurs per second and is equal to 1/p.

    • Another parameter of interest is the wavelength λ, which is the linear distance between any two equivalent point on successive maxima or minima of a wave.

    • Multiplication of the frequency in reciprocal seconds by the wavelength in metres gives the velocity of propagation

    10 E.T

    V= vλ

  • • It is important to realize that the frequency is determined by the source and remains invariant regardless of the media traversed by the radiation.

    • In contrast, the velocity of propagation v, the rate at' which the wave front moves through a medium, is dependent upon both the medium and the frequency;

    • In a vacuum the velocity of radiation becomes independent of frequency and is at its maximum.

    • This velocity, given the symbol c, has been accurately determined to be 2.99792 X 1010 cm/s. Thus, for a vacuum,

    • c= νλ = 3 x 1010 cm/s

    11 E.T

  • • In any other medium, the rate of propagation is less because of

    interactions between the electromagnetic field of the radiation

    and the bound electrons in the atoms or molecules making up

    the medium.

    • Since the radiant frequency is invariant and fixed by the

    source, the wavelength must decrease as passes from a

    vacuum to a medium containing matter.

    wave number, , which is the reciprocal of wavelength 12 E.T

  • • Particle Properties of Electromagnetic Radiation

    • An understanding of certain interactions between radiation and

    matter requires that the radiation be treated as packets of

    energy called photons or quanta.

    – The energy of a photon depends upon the frequency of the radiation,

    and is given by

    where h is Planck’s constant, which has a value of 6.626 x10–34 J · s.

    Photon : A particle of EMR having zero mass and an energy of hv.

    13 E.T

  • Example 1:- The “D” line in the solar spectrum was due to the

    absorption of solar radiation by sodium atoms. The wavelength of

    the sodium D line is 589 nm. What are the frequency and the wave

    number for this line?

    SOLUTION

    The frequency and wave number of the sodium D line are

    14 E.T

  • Example 2:- What is the energy per photon of the sodium D line

    (λ= 589 nm)?

    SOLUTION

    The energy of the sodium D line is

    15 E.T

  • Electromagnetic (EM) spectrum

    • The electromagnetic (EM) spectrum is just a name that scientists give a bunch of types of radiation when they want to talk about them as a group.

    • Radiation is energy that travels and spreads out as it goes

    – visible light that comes from a lamp in your house or

    – radio waves that come from a radio station

    • Other examples of EM radiation are microwaves, infrared and ultraviolet light, X-rays and gamma-rays.

    16 E.T

    http://imagine.gsfc.nasa.gov/docs/dict_ei.htmlhttp://imagine.gsfc.nasa.gov/docs/dict_ei.htmlhttp://imagine.gsfc.nasa.gov/docs/dict_ei.htmlhttp://imagine.gsfc.nasa.gov/docs/dict_qz.htmlhttp://imagine.gsfc.nasa.gov/docs/dict_qz.htmlhttp://imagine.gsfc.nasa.gov/docs/dict_qz.htmlhttp://imagine.gsfc.nasa.gov/docs/dict_jp.htmlhttp://imagine.gsfc.nasa.gov/docs/dict_ei.htmlhttp://imagine.gsfc.nasa.gov/docs/dict_qz.htmlhttp://imagine.gsfc.nasa.gov/docs/dict_qz.htmlhttp://imagine.gsfc.nasa.gov/docs/dict_qz.htmlhttp://imagine.gsfc.nasa.gov/docs/dict_qz.htmlhttp://imagine.gsfc.nasa.gov/docs/dict_ei.htmlhttp://imagine.gsfc.nasa.gov/docs/dict_ei.htmlhttp://imagine.gsfc.nasa.gov/docs/dict_ei.html

  • 17 E.T

  • • I. Absorption of Radiation

    • When radiation passes through a transparent medium contain

    chemical species, certain frequencies may be selectively

    attenuate by the process of absorption.

    • EMR energy is transferred to the atoms or molecules

    constituting the sample; as a result, these particles are

    promoted from a lower energy state to higher-energy states, or

    excited states.

    • At room temperature, most substances are in their lowest

    energy or ground state.

    • Absorption then ordinarily involves a transition from the

    ground state to higher-energy states.

    • Atoms, molecules, or ions have only at limited number of

    discrete, quantized energy levels;

    18 E.T

  • • for absorption of radiation to occur, the energy

    of the exciting photon must exactly match the

    energy difference between the ground state and

    one of the excited states of the absorbing

    species.

    • consider the three processes that may occur

    when a two-state system is subjected to

    radiation of frequency or wave number ,

    corresponding to the energy separation ∆E

    where

    19 E.T

  • 1. Induced absorption, in which the molecular (or atom) M absorbs

    a quantum of radiation and is excited from m to n:

    Eg. the appearance of an aqueous solution of copper sulphate as blue

    due to the absorption of the complementary colour, red, by the solution.

    2. Spontaneous emission, in which M* (in state n) spontaneously

    emits a quantum of radiation:

    Eg. a sodium vapour or tungsten filament lamp, is of the

    spontaneous type. 20 E.T

  • 3. Induced, or stimulated, emission.

    This is a different type of emission process from that of type 2 in that

    a quantum of radiation of wave number is required to induce,

    or stimulate, M* to go from n to m.

    The process is represented by

    Since the energy d/n are unique for each species, a study of the

    frequencies of absorbed radiation provides a means of characterizing

    the constituents of a sample of matter.

    For this purpose, a plot of absorbance as a function of wavelength or

    frequency is experimentally derived absorption spectra

    21 E.T

  • The general appearance of an absorption spectrum will depend upon

    the complexity,

    the physical state, and

    the environment of the absorbing species.

    It is convenient to recognize two types of spectra, namely, those

    associated with

    atomic absorption and

    those resulting from molecular absorption.

    22 E.T

  • • A. Atomic Absorption

    • When a beam of polychromatic Uv or Vis

    radiation passes through a medium containing

    gaseous atoms, only a few frequencies are

    attenuated by absorption, and

    • The spectrum consists of a number of very

    narrow absorption lines

    23 E.T

  • 200 300 400 500 600 Wavelength, nm

    A

    3s

    3p

    4p

    5p

    E

    59

    0

    33

    0

    28

    5

    0

    1

    2

    3

    4

    (a)Absorption spectrum for Na vapour (b) partial energy level diagram for Na

    Transitions involves excitation of

    the single outer electron of Na

    24 E.T

  • • B. Molecular Absorption

    • Molecules undergo three type of quantized transitions when

    exited by UV, visible and IR radiation.

    • For UV and VIS radiation excitation involves promoting an

    electron from ground state (low energy molecular or atomic

    orbital )to exited state (a higher – energy orbital).

    • The energy hv of the photon must be exactly the same as the

    energy d/n b/n the two orbital energies.

    • The transition known as Electronic transition

    – electronic absorption

    • Molecules also exhibit two other types of radiation- induced

    transitions

    – Vibrational

    – Rotational

    25 E.T

  • • For each electronic energy state of the molecule, there

    normally exist several possible vibrational states, and for

    each of these, in turn, numerous rotational states,

    26 E.T

  • Symmetric Asymmetric

    A)Stretching Vibrations

    In- plane scissoring

    - + + +

    - In-plane rocking Out-plane wagging Out-plane twisting

    B) Bending vibrations

    Vibrational transition come about b/c a molecule has a multitude of

    quantized energy levels (vibrational state) associated with the bond that

    holed the molecule together.

    27 E.T

  • • Absorption by polyatomic molecules, particularly in the

    condensed state, is a considerably more complex process

    because the number of energy states is greatly enhanced.

    • Here, the total energy of a molecule is given by

    • E =Eele + Evib. + Erot.

    – where Eele describes the electronic energy of the molecule, and

    – Evib refers to the energy of the molecule resulting from various

    atomic vibrations.

    – The third term accounts for the energy associated with the

    rotation of the molecule around its center of gravity.

    28 E.T

  • Figure below is a graphical representation of the energies

    associated with few of the electronic and vibrational states of a

    molecule.

    Fig. 29 E.T

  • • The heavy line labelled Eo represents the electronic

    energy of the molecule in its ground state (its state of

    lowest electronic energy);

    • the lines labelled E1 and E2 represent the energies of

    two excited electronic states.

    • Several vibrational energy levels (0,1,…,4) are shown

    for each of these electronic states.

    • the energy difference between the ground state and an

    electronically excited state is large relative to the

    energy differences b/n vibration levels in a given

    electronic state

    – (typically, the two differ by a factor of 10 to 100).

    30 E.T

  • • Visible radiation causes excitation of an electron from

    Eo to any of the vibration levels associated with E 1'

    • Ultraviolet causes excitation of an electron from Eo to

    any of the vibration levels associated with E2

    • Finally, the less energetic near- and mid infrared

    radiation can only bring about transition among the

    vibrational levels of the ground state.

    • Although they are not shown, several rotational

    energy levels are associated with each vibrational

    level.

    • The energy difference between these is small relative

    to the energy difference between vibrational levels;

    31 E.T

  • In contrast to atomic absorption spectra, which consist of

    a series of sharp, well-defined lines,

    Molecular spectra in the ultraviolet and visible regions are

    ordinarily characterized by absorption bands

    that often encompass a substantial wavelength the spectrum for a

    molecule ordinarily consists of a series of closely spaced

    absorption bands, such as that shown for benzene vapor in Figure

    below

    32 E.T

  • Unless a high-resolution instrument is employed, the

    individual bands may not be detected, and the spectra will

    appear as smooth curves.

    Finally, in the condensed state and in the presence of

    solvent molecules, the individual bands tend to broaden to

    give spectra such as those shown in the Figure below

    33 E.T

  • • Pure vibrational absorption can be observed in the

    infrared region.

    – where the energy of radiation is insufficient to cause

    electronic transitions.

    • Variations in rotational levels may give rise to a series

    of peaks for each vibrational state.

    – However, rotation is often hindered or prevented in liquid

    or solid samples;

    – the effects of these small energy differences are

    not ordinarily detected in such samples.

    34 E.T

  • • Pure rotational spectra for gases can be observed in the

    microwave region.

    – Absorption Induced by a Magnetic Field.

    • When electrons or the nuclei of certain elements are

    subjected to a strong magnetic field additional quantized

    energy levels are produced as a consequence of magnetic

    properties of these elementary particles.

    – Absorption by nuclei or by electrons in magnetic fields is

    studied by nuclear magnetic resonance (NMR) and

    – electron spin resonance (ESR) techniques, respectively

    35 E.T

  • Relaxation Processes

    • Ordinarily, the lifetime of an atom or molecule

    excited by absorption of radiation is short because

    several relaxation processes exist which permit its

    return to the ground state.

    • Two of the most important of these mechanisms

    • nonradiative relaxation

    • fluorescent relaxation

    36 E.T

  • E.T 37

    Fig. 5

  • • Two type of nonradiative relaxation are shown in the figure 5b

    Vibrational deactivation or relaxation,

    – denoted by short arrows b/n vibrational energy levels,

    – involves the loss of energy in a series of small steps,

    – the excitation energy being converted to kinetic energy by collision with

    solvent molecules.

    • These energy transferd to solvent molecule and reflacted by a

    tiny increase in the temperature of the medium.

    • Non radiative relaxation b/n the lowest vibrational level of an

    exited electronic state and the upper vibrational level of another

    electronic level can also occur.

    – Represent by long arrow.

    • Less efficient than vibrational relaxation and its life time is b/n

    10-6 and 10-9

    • As shown in Figure 5c, relaxation can also occur by emission of

    fluorescent radiation.

    38 E.T

  • II. Emission Radiation

    • Excited particles (ions, atoms, or molecules)can be excited to

    one or more higher energy levels by a variety of means,

    including

    – bombardment with electrons or other elementary particles,

    – exposure to a high-potential alternating current spark,

    – heat treatment in an arc or a flame, or

    – absorption of electromagnetic radiation.

    • The lifetime of an excited species is generally transitory (10-6

    – 10-9 s) and

    • a relaxation to a lower-energy levels or to their ground states

    takes place with the release of the excess energy in the form of

    Electromagnetic radiation, heat or perhaps both.

    39 E.T

  • • Emission spectra

    • A plot of the relative power of the emitted radiation

    as a function of wavelength or frequency.

    40 E.T

    Figure 6 Emission spectrum of a brine sample obtained with an oxyhydrogen flame. The spectrum consists of the superimposed line. band. and continuum spectra of the constituents of the sample. The character- istic wavelengths of the species contributing to the spectrum are listed beside each feature. (R. Hermann and

    C. T. J. Alkemade, Chemical Analysis by Flame Photometry, 2nd ed.. p. 484. New York: Interscience, 1979.)

  • • There are three type of spectra:- line, band and

    continues.

    • Excitation can be brought about Radiating particles

    that are well separated from one another, as in the

    gaseous state, behave as independent bodies and often

    produce radiation containing relatively few specific

    wavelengths.

    • The resulting spectrum is then discontinuous and is

    termed a line spectrum.

    • Made up of a series of sharp, well-defined peaks

    caused by an excitation of individual atom.

    • In fig. 6 lines for Na, K, Sr and Ca

    41 E.T

  • • The band spectrum consists of several groups of lines so

    closely spaced that they are not completely resolved.

    – The source of the band is small molecules or radicals

    – In fig.6 bands of OH,MgOH and MgO

    • Bands arise from the numerous quantized vibrational levels

    that are superimposed on the ground state electronic energy

    level of a molecule.

    • A continuous spectrum, on the other hand, is one in which all

    wavelengths are represented over an appreciable range, or one

    in which the individual wavelengths are so closely spaced that

    resolution is not feasible by ordinary means.

    – The line and band spectra are superimposed

    • Continuous spectra result from excitation of:

    • (1) solids or liquids, in which the atoms are so closely packed

    as to be incapable of independent behaviour; or

    • (2) complicated molecules possessing many closely related

    energy states.

    42 E.T

  • • When solids are heated to incandescence, the continuous

    radiation that is emitted is more characteristic of the

    temperature of the emitting surface than of the material of

    which that surface is composed.

    • Radiation of this kind (called black-body radiation)

    • is produced by the innumerable atomic and molecular

    oscillations excited in the condensed solid by the thermal

    energy.

    43 E.T

    Figure 7 Blackbody radiation curves for various light sources. Note the shift in the peaks as the temperature of the sources changes.

  • • Both continuous spectra and line spectra are of

    importance in analytical chemistry.

    • The former are frequently employed in

    methods based on the interaction of radiation

    with matter, such as spectrophotometry.

    • Line spectra, on the other hand, are important

    because • they permit the identification and determination of the

    emitting species.

    44 E.T

  • Absorption Law

    Chapter 2

    E.T 45

  • The absorption law

    • also known as the Beer-Lambert law or just Beer's law,

    • Tells us quantitatively how the amount of attenuation depends on the concentration of the absorbing molecules and the path length over which absorption occurs.

    • As light traverses a medium containing an absorbing analyte, decreases in intensity occur as� the analyte becomes excited.

    E.T 46

  • Transmittance and Absorbance

    • The common terms and alternative names/symbols

    employed in spectroscopy are listed in the table

    below. The recommended terms and symbols are

    listed under the column labeled Term and Symbol.

    E.T 47

  • Terms and Symbols

    E.T 48

  • Transmittance

    E.T 49

    Figure 1 depicts a beam of parallel radiation before and after it has passed

    through a layer of solution with a thickness of b cm and a concentration of c of

    an absorbing species.

    As a consequence of interactions between the photons and absorbing particles,

    the power of the beam is attenuated from Po to P.

    The transmittance T of the solution is the fraction

    Equ.1

  • Absorbance

    • Absorbance, like the previous table shows, can

    be defined as the base-ten logarithm of the

    reciprocal of the transmittance :

    A = log 1/T = -log I/Io = -log P/Po

    E.T 50

    Equ.2

  • • Absorptivity and Molar Absorptivity

    • Absorbance is directly proportional to the path length through the solution and the concentration of the absorbing species.

    – That is, A = abc

    • where a is a proportionality constant called – the absorptivity.

    • The magnitude will clearly depend upon the units used for b(cm) and c (g/L).

    • When the concentration is expressed in moles per litter and the cell length is in centimetres,

    – the absorptivity is called the molar absorptivity and given the special symbol ԑ

    – A = ԑbc

    E.T 51

  • Experimental Measurements of Absorbance and

    transmittance

    • The relationship given by Equation 1 and 2 is not directly

    applicable to chemical analysis.

    • Neither Absorbance nor Transmittance, as defined, can be

    conveniently measured in the laboratory

    – because the solution to be studied must be held in some sort of

    container.

    • Interaction between the radiation and the walls is inevitable

    leading to a loss by reflection at each interface;

    • significant absorption may occur within the walls themselves.

    • the beam may suffer a diminution in power during its passage

    through the solution as a result of scattering by large

    molecules or inhomogeneities. E.T 52

  • Reflection and Scattering Losses

    E.T 53

  • Reflection and Scattering Losses (cont)

    • Reflection and scattering losses are significant and to compensate for these effects, the power of the beam transmitted by the analyte solution is ordinarily compared with the power of the beam transmitted by an identical cell containing only the solvent.

    • An experimental absorbance that closely approximates the true absorbance is then obtained with the equation:

    A = log Psolvent/Psolution ≈ log Po/P

    E.T 54

  • • Example:

    • A sample in a 1.0 cm cell is determined with a

    spectrometer to transmit 80 % light at a certain

    wavelength. If the absorptivity of this substance at

    this wavelength is 2.0, what is the concentration of

    the substance?

    • Solution:

    • The percent transmittance is 80 % and so T = 0.80:

    E.T 55

    -1 11log 2.0 cm 1.0 cm0.80

    g L c

    -1log1.25 2.0 g L c

    0.100.050 /

    2.0c g L

  • E.T 56

    0.70T 1

    log log1.43 0.1550.70

    A

    0.155 (0.0100 / )ab g L

    15.5 /ab L g15.5 L/g (4 0.0100 g/L) = 0.620A

    1log 0.620

    T

    0.240T 1 1 1

    2 2 2

    A abc c

    A abc c

    22 1

    1

    40.155 0.620

    1

    cA A

    c

    Solution:

    Therefor

    e,

    The absorbance of the new solution could have

    been calculated more directly:

    Example:

    A solution containing 1.00 mg ion (as the thiocyanate complex) in 100 mL

    was

    observed to transmit 70.0% of incident light compared to an appropriate

    blank.

    (a) What is the absorbance of the solution at this wavelength?

    (b) What fraction of light would be transmitted by a solution of iron four

    times as concentrated?

  • Beer’s Law

    Consider This:

    • A parallel beam of monochromatic radiation with power Po strikes the block perpendicular to a

    surface after passing through a length b of the material , which contains n absorbing particles , the beam’s power is decreased to P as a result of absorption.

    E.T 57

  • an

    area S and an infinitesimal thickness dx.

    Within this section there are dn absorbing

    particles;

    associated with each particle, we can imagine a

    surface

    at which photon capture will occur.

    That is, if a photon reaches one of these areas by

    chance, absorption will follow immediately.

    The total projected area of these capture

    surfaces

    within the section is designated as dS;

    the ratio of the capture area to the total area,

    then, is

    dS/S.

    On a statistical average, this ratio represents the

    probability for the capture of photons within the

    E.T 58

  • The power of the beam entering the section, Px,

    is proportional to the number of photons per

    square centimeter per second, and

    dPx represents the quantity removed per second

    within the section;

    the fraction absorbed is then - dPx /Px, and this

    ratio also equals the average probability for

    capture.

    The term is given a minus sign to indicate that P

    undergoes a decrease. Thus,

    eq.3

    E.T 59

  • Recall, now, that dS is the sum of the capture areas for particles within the

    section; it must therefore be proportional to the number of particles, or

    where dn is the number of particles and a is a proportionality

    constant, which can be called the capture cross section

    Combining Equations 3 and 4 and summing over the interval

    between zero and n, we obtain

    Integration gives

    eq 6

    eq. 4

    eq 5

    E.T 60

  • converting to base 10 logarithms and inverting the fraction to

    change the sign, we obtain

    eq 7

    where n is the total number of particles within the block shown in Figure 2. The

    cross sectional area S can be expressed in terms of the volume of the block V

    and its length b.

    Thus,

    Substitution of this quantity into Equation 7 yields

    eq. 8

    E.T 61

  • Note that n/V has the units of concentration (that is, number of particles per

    cubic centimetre); we can readily convert n/V to moles per litter. Thus, number

    of moles given by

    Combining eq.8 and 9

    eq.9

    E.T 62

    and c in mol/L is given by

  • E.T 63

    Finally, the constants in this equation can be collected into a

    single term ԑ to give

    which is Beer's law.

  • Application of Beer`s Law to

    Mixture

    • Beer's law also applies to a solution containing more than one

    kind of absorbing substance.

    • Provided there is no interaction among the various species, the

    total absorbance for a multicomponent system is given by

    Atotal = A1 + A2 + ... + An (6-11) = ԑ1bcl + ԑ2bc2 + ... + ԑnbcn

    • where the subscripts refer to absorbing components 1, 2, ..., n.

    E.T 64

  • Deviations from Beer’s Law

    • Beer’s law states that a plot of absorbance versus

    concentration should give a straight line passing

    through the origin with a slope equal to ab. However,

    deviations from direct proportionality between

    absorbance and concentration are sometimes

    encountered.

    • These deviations are a result of one or more of the

    following three things ; real limitations, instrumental

    factors or chemical factors.

    E.T 65

  • Real Limitations

    • Beer’s law is successful in describing the absorption behavior of dilute solutions only ; in this sense it is a limiting law. At high concentrations ( > 0.01M ),the average distance between the species responsible for absorption is diminished to the point where each affects the charge distribution of its neighbors.

    • This interaction, in turn, can alter the species’ ability to absorb at a given wavelength of radiation – Because the extent of interaction depends upon concentration,

    • the occurrence of this phenomenon thus leading to a deviation from Beer’s law.

    E.T 66

  • Limitations (cont)

    • A similar effect is sometimes encountered in solutions

    containing low absorber concentrations and high

    concentrations of other species, particularly electrolytes.

    • The close proximity of ions to the absorber alters the molar

    absorptivity of the latter by electrostatic interactions;

    – the effect is lessened by dilution

    • Deviations also arise because ԑ is dependent upon the

    refractive index of the solution. Thus, if concentration

    changes cause significant alterations in the refractive index µ

    of a solution, departures from Beer’s law are observed.

    E.T 67

  • Chemical Deviations

    • Apparent deviations from Beer's law are frequently

    encountered as a consequence of association,

    dissociation, or reaction of the absorbing species with

    the solvent to generate a product that has a different

    absorption spectrum with the analyte.

    • A common example of this behavior is found with

    acid/base indicators.

    • Deviations arising from chemical factors can only be

    observed when concentrations are changed.

    E.T 68

  • Instrumental Deviations

    • Unsatisfactory performance of an instrument

    may be caused by fluctuations in the power-

    supply voltage, an unstable light source, or a

    non-linear response of the detector-amplifier

    system.

    E.T 69

  • Polychromatic Radiation

    • Strict adherence to Beer’s law is observed only with truly

    monochromatic radiation. This sort of radiation is only

    approached in specialized line emission sources.

    • All monochromators, regardless of quality and size, have a

    finite resolving power and therefore minimum instrumental

    bandwidth.

    • the use of radiation that is restricted to a single wavelength

    is seldom practical; devices that isolate portions of the

    output from a continuous source produce a more or less

    symmetric band of wavelengths around the desired one

    E.T 70

  • Polychromatic Radiation (cont)

    • A good picture of the effect of polychromatic

    radiation can be presented as follows. When

    radiation consists of two wavelengths, λ’ and λ’’, and

    assuming that Beer’s law applies at each of these

    individually

    • the absorbance at λ’ is given by:

    log ( P’o/P’ ) = A’ = ԑ’bc

    P’o /P’ = 10ԑ’bc

    P’ = p’0 10-ԑ’bc

    E.T 71

    P’’ = p0’’ 10-ԑ’’bc Similarly at λ’’ is given by

  • When an absorbance measurement is made with radiation composed of both

    wavelengths, the power of the beam emerging from the solution is given by (P' +

    P") and that of the beam from the solvent by (P’O + P’’O).

    Therefore, the measured absorbance is

    E.T 72

  • E.T 73

  • • If the two molar absorpitivities differ, however, the

    relationship between AM and concentration will no longer be

    linear; moreover, greater departures from linearity can be

    expected with increasing differences between ԑ' and ԑ`'.

    E.T 74

    FIGURE the effect of polychromatic radiation on beer’s law. In the spectrum at the left, the

    absorptivity of the analyte is nearly constant over band A from the source. Note in the beer’s law

    plot at the right that using band A gives a linear relationship. In the spectrum, band B corresponds

    to a region where the absorptivity shows substantial changes. In the right plot, note the dramatic

    deviation from beer’s law that results.

  • Chapter Three

    Instruments for optical spectroscopy

    75 E.T

  • • The basic components of analytical instruments

    – for emission,

    – absorption and

    – fluorescence spectroscopy

    • Are remarkably alike in function and general performance requirements regardless of whether the instrument are designed for Uv, Vis or IR radiation.

    • Called Optical instruments

    76 E.T

  • • Spectroscopic instruments contain five

    components, including:

    • a stable source of radiant energy;

    • a wavelength selector that permits the isolation

    of a restricted wavelength region;

    • a transparent container for holding the sample;

    • a radiation detector or transducer, that converts

    radiant energy to a usable signal (usually

    electrical); and

    • a signal processor and readout

    77 E.T

  • FIGURE 1 Components for various types of instruments for

    optical spectroscopy: (a) absorption spectroscopy,

    (b), fluorescence and scattering Spectroscopy and (c) emission spectroscopy 78 E.T

  • • As can be seen in the figure, the configuration of components

    (4) and (5) is the same for each type of instruments.

    • Emission spectroscopy differ from the other two types in that

    components 1 & 3are combined,

    – that no external radiation source is required;

    • the sample itself is the emitter.

    – the sample is usually fed into a plasma or a flame, which provides enough

    thermal energy to cause the analyte to emit characteristic radiation..

    • Absorption and fluorescence spectroscopy require an external source of

    radiant energy and a cell to hold the sample.

    • For the absorption measurements

    • The beam from the source passes through the sample after leaving the

    wavelength selector.

    • In fluorescence Here, two wavelength selectors are needed to select the

    excitation and emission wavelengths. The selected source radiation is

    incident on the sample and the radiation emitted is measured. usually at

    right angles to avoid scattering.

    79 E.T

  • Optical Material

    The cell, windows, lenses and wavelength – selecting elements

    in a optical spectroscopy instrument must transmit radiation in the

    wavelength region being investigated.

    80 E.T Figure 2a Transmittance ranges for various construction materials.

  • E.T 81

    Figure 2b Transmittance ranges for wavelength selectors

  • • Ordinary silicate glass is completely adequate for use

    in the visible region and has the considerable

    advantage of low cost.

    • In the UV region at wavelengths shorter than about

    380 nm glass begins to absorb and fused silica or

    quartz must be substituted.

    • Also, in the IR region, glass, quartz, and fused silica

    all absorb at wavelengths longer than about 2500nm.

    • Hence, optical elements for IR spectrometry are

    typically made from halide salts or In some cases

    polymeric materials.

    E.T 82

  • Spectroscopic Sources • To be suitable for spectroscopic studies,

    – a source must generate a beam of radiation that is sufficiently powerful

    to allow easy detection and measurement.

    – its output power should be stable for reasonable periods of time.

    • Typically, for good stability. a well-regulated power supply

    must provide electrical power for the source.

    • Spectroscopic sources are of two types:

    • continuum sources, which emit radiation that changes in

    intensity only slowly as a function of wavelength, and

    • line sources, which emit a limited number of spectral lines,

    each of which spans a very limited wavelength range.

    • The distinction between these sources is illustrated in Figure 3.

    E.T 83

  • 84 E.T

    Fig. 3

  • • Continuous Sources of Ultraviolet -Visible, and Near-Infrared Radiation

    • Tungsten Filament Lamps

    • The most common source of visible and near-infrared radiation is the tungsten filament lamp. – The energy distribution of this source approximates that of a black body and

    is thus temperature dependent.

    • Figure below illustrates the behaviour of the tungsten filament lamp at 3000oK.In most absorption instruments, the operating filament temperature is about 2900oK; the bulk of the energy is thus emitted in the infrared region.

    85 E.T

    Figure 4 (a) A tungsten lamp of the type used in spectroscopy and its spectmm

    (b). Intensity of the tungsten source is usually quite low at wavelengths shorter than about 350 nm.

    Note that the intensity reaches a maximum in the near-IR region of the spectrum ( -1200 nm in this case).

  • • A tungsten filament lamp is useful for the wavelength region between 320 and 2500 nm.

    • In the visible region the energy output of a tungsten lamp varies.

    – As a consequence close voltage control is required for a stable radiation source.

    • Constant voltage transformers or electronic voltage regulators are often employed for this purpose

    86 E.T

  • Hydrogen or Deuterium Lamps

    • A truly continuous spectrum in the ultraviolet region is conveniently

    produced by the electrical excitation of hydrogen or deuterium

    at low pressure.

    • The mechanism by which a continuum is produced involves the

    formation of excited molecule (D2*or H2*) by absorption of

    electrical energy.

    • The species then dissociates to give two hydrogen or deuterium

    plus ultraviolet photon.

    H2 + Ee H2* H’ + H’’ + hʋ

    The energy of the overall process is

    Ee = E H2* = EH’ + EH’’ + hʋ

    E H2* is fixed quantized energy of H2* , EH’ and EH’’

    kinetic energy of H’ and H’’

    87 E.T

  • • Most modern lamps for generating ultraviolet radiation contain deuterium and are of a low-voltage lamps. – an arc is formed between a heated, oxide-coated filament and a

    metal electrode.

    • The heated filament provides 'electrons to maintain a dc current when a voltage of about 40 V is applied; a regulated power supply is required for constant intensities.

    • Both hydrogen and deuterium lamps produce a continuous spectrum in the region of 160 to 375 nm.

    • Quartz windows must be employed in the tubes, since glass absorbs strongly in this wavelength region.

    • Xenon Arc Lamps.

    • This lamp produces intense radiation by the passage of current through an atmosphere of xenon.

    • The spectrum is continuous over the range between about 250 and 600 nm.

    88 E.T

  • • Continuous Sources of Infrared radiation

    • The common infrared source is an inert solid heated electrically to temperatures between – 1500 and 2000oK.

    • The Nernst glower is composed of Zirconium and yttrium oxides formed into a cylinder having a diameter of 2 mm and length of perhaps 20 mm,

    • It emits infrared radiation when heated to a high temperature by an electric current.

    • The Globar Source

    • A Globar is a silicon carbide rod, usually about 50 mm in length and 5 mm in diameter.

    • Radiation in the region 1 to 40 µm is emitted when the Globar heated to 15000C by passage of electricity

    89 E.T

  • • Line Sources

    • Sources that emit a few discrete lines find use in atomic

    absorption spectroscopy, Raman spectroscopy, refractometry, and

    polarimetry.

    • Hollow Cathode Lamps

    • The most common source for atomic absorption

    • which consists of a tungsten anode and a cylindrical cathode

    sealed in a glass tube that is filled with neon or argon at a

    pressure of 1 to 5 torr

    • The cathode is constructed of the metal whose spectrum is

    desired or serves As a support for a coating of that metal.

    90 E.T

  • CELLS

    UV Spectrophotometer

    Quartz (crystalline silica)

    Visible Spectrophotometer

    Glass

    IR Spectrophotometer

    NaCl

    91 E.T

  • E.T 92

    Figure Typical examples of commercially available cells for the

    UV/visible region.

  • • Wavelength Selection

    • Spectroscopic instruments generally required a device that

    restricts the wavelength that is to be used to narrow band that

    is absorbed or emitted by the analyte.

    – Such device enhance selectivity and sensitivity of an instrument

    • For absorption measurement, narrow bandwidths increase the

    likelihood of adherence to Beer’s law.

    • For this reason we usually try to select a single wavelength

    where the analyte is the only absorbing species.

    – Unfortunately, we cannot isolate a single wavelength of radiation from

    a continuum source.

    • Instead, a wavelength selector passes a narrow band of

    radiation characterized by

    – a nominal wavelength, an effective bandwidth, and a maximum

    throughput of radiation.

    93 E.T

  • • The effective bandwidth is defined as the width of the radiation at

    half the maximum throughput.

    • The ideal wavelength selector has a high throughput of radiation

    and a narrow effective bandwidth.

    – A high throughput is desirable because more photons pass through the

    wavelength selector, giving a stronger signal with less background noise.

    • A narrow effective bandwidth provides a higher resolution, with

    spectral features separated by more than twice the effective

    bandwidth being resolved.

    94 E.T

  • • Generally these two features of a wavelength

    selector are in opposition

    • Conditions favouring a higher throughput of

    radiation usually provide less resolution.

    • Decreasing the effective bandwidth improves

    resolution, but at the cost of a noisier signal.

    • For a qualitative analysis, resolution is

    generally more important than the throughput

    of radiation; thus,

    – smaller effective bandwidths are desirable.

    • In a quantitative analysis a higher throughput

    of radiation is usually desirable

    95 E.T

  • • Wavelength Selection Using Filters

    • The simplest method for isolating a narrow band of radiation is

    to use an absorption or interference filter.

    • Absorption filters

    • work by selectively absorbing radiation from a narrow region of

    the electromagnetic spectrum .

    – Limited application to visible region.

    • A simple example of an absorption filter usually consist of a

    colored glass plate that removes part of incident radiation by

    absorption.

    – A purple filter, for example, removes the complementary color green

    from 500–560 nm.

    – Commercially available absorption filters provide effective bandwidths

    from 30–250 nm.

    • Interference filters are more expensive than absorption filters,

    but have narrower effective bandwidths, typically 10–20 nm,

    with maximum throughputs of at least 40%.

    96 E.T

  • 97 E.T

  • • Wavelength Selection Using Monochromators

    • One limitation of an absorption or interference filter is that

    they do not allow for a continuous selection of wavelength.

    • If measurements need to be made at two wavelengths, then the

    filter must be changed in between measurements.

    • A further limitation is that filters are available for only

    selected nominal ranges of wavelengths.

    • An alternative approach to wavelength selection, which

    provides for a continuous variation of wavelength, is the

    – monochromator

    • Monochromators for ultraviolet, visible, and infrared radiation

    are all similar in mechanical construction in the sense that they

    employ slits, lenses, mirrors, windows, and prisms or gratings.

    98 E.T

  • • Components of Monochromator

    • All monochromators contain an entrance slit, a collimating lens

    or mirror to produce a parallel beam of radiation, a prism or

    grating as a dispersing element and a focusing element,

    • which projects a series of rectangular images of the entrance slit

    upon a plane surface (the focal plane).

    Polychromatic

    Ray

    Infrared

    Red

    Orange

    Yellow

    Green

    Blue

    Violet

    Ultraviolet

    monochromatic

    Ray

    SLIT

    PRISM

    Polychromatic Ray Monochromatic Ray

    99 E.T

  • 100 E.T

    Figure Types of monochromators: (a) grating monochromator: (b) prism monochromator.

    The monochromator design in (a) is a Czerny- Turner design. while the prism

    monochromator in (b) is a Bunsen design. In both cases, , λ1 >λ2

  • • A source of radiation containing two wavelengths, λ1 and λ2, is

    shown for purposes of illustration.

    • This radiation enters the monochromators via a narrow

    rectangular opening or slit, is collimated, and then strikes the

    surface of the dispersing element at an angle.

    • In the prism monochromator, refraction at the two faces results

    in angular dispersal of the radiation, as shown;

    • for the grating, angular dispersion results from diffraction,

    which occurs at the reflective surface.

    • In both designs, the dispersed radiation is focused on the focal

    plane AB where it appears as two "images of the entrance slit

    (one for each wavelength).

    101 E.T

  • • Radiation detectors and Transducers

    • A detector is a device that indicates the existence of some

    physical phenomenon.

    • Photographic film indicating presence of EMR ,pointer of

    balance indicating mass d/n, mercury level in thermometer

    indicating temperature change.

    – Early instruments for the measurement of emission and absorption of

    radiation required visual or photographic methods for detection.

    • A transducer is a special type of detector that converts signals,

    such as light intensity, pH, mass and temperature into electrical

    signals that can be subsequently amplified,

    • manipulated and finally converted into numbers that are

    related to the magnitude of the original signal.

    102 E.T

  • • Properties of Transducers

    • To be useful the detector must respond rapidly to low levels of

    radiant energy over a broad wavelength range.

    • produce an electrical signal that can be readily amplified, and

    have a relatively low noise level (for stability).

    • Finally, it is essential that the signal produced be directly

    proportional to the beam power P; that is,

    G = KP + K'

    – where G is the electrical response of the detector in units of current.

    resistance, or potential. The constant K measures the sensitivity of the

    detector in terms of electrical response per unit of radiant power.

    – Many detectors exhibit a small constant response known as a dark

    currenl K'. even when no radiation impinges on their surfaces.

    103 E.T

  • Types of Transducers

    As indicated in Table,10.4 two general types of radiation transducers

    are employed; one responds to photons, the other to heat.

    104 E.T

  • • All photon detectors are based upon interaction of

    radiation with a reactive surface to produce electrons

    (photoemission) or

    • to promote electrons to energy states in which they

    can conduct electricity (photoconduction).

    • Only ultraviolet, visible, and near-infrared radiations

    have sufficient energy to cause these processes to

    occur.

    • Photoelectric detectors also differ from heat detectors

    in that their electrical signal results from a series of

    individual events (absorption of one photon).

    105 E.T

  • • Photon Detectors

    • Several types of photon detectors are available. including:

    – photovoltaic cells, phototubes, photomultiplier tubes, semiconductor

    detectors, silicon diodes and charge –transfer device

    • Phototubes

    • which consists of a semi cylindrical cathode and a wire

    anode sealed inside an evacuated transparent envelope.

    • The concave surface of the electrode supports a layer of

    photoemissive material that tends to emit electrons upon

    being irradiated.

    • When a potential is applied across the electrodes, the

    emitted electrons flow to the wire anode, generating a

    photocurrent.

    • The currents produced are generally about one-tenth as great as

    those from a photovoltaic cell for a given radiant intensity.

    106 E.T

  • E.T 107

    Figure A phototube and accompanying circuit. The photocurrent induced by

    the radiation causes a voltage across the measuring resistor:

    this voltage is then amplified and measured.

  • • The number of electrons ejected from a

    photoemissive surface is directly proportional

    to the radiant power of the beam striking that

    surface.

    • As the potential applied across the two

    electrodes of the tube is about 90 V, the

    fraction of the emitted electrons reaching the

    anode to give current that is proportional to the

    radiant power.

    • Assignment Discuss in detail for the remaining

    four type of detectors

    108 E.T

  • • Heat detector

    • Infrared radiation generally does not have sufficient energy to

    produce a measurable current when using a photon transducer.

    – A heat detector, therefore, is used for infrared spectroscopy.

    • The absorption of infrared photons by a heat detector increases

    its temperature, changing one or more of its characteristic

    properties.

    • The pneumatic transducer, for example consists of a small tube

    filled with xenon gas equipped with an IR-transparent window

    at one end, and a flexible membrane at the other end.

    • A blackened surface in the tube absorbs photons, increasing

    the temperature and, therefore, the pressure of the gas.

    • The greater pressure in the tube causes the flexible membrane

    to move in and out, and this displacement is monitored to

    produce an electrical signal.

    109 E.T

  • • SIGNAL PROCESSORS AND READOUTS

    • The signal processor is ordinarily an electronic

    device that amplifies the electrical signal from

    the detector; in addition, it may alter the signal

    from dc to ac (or the reverse)

    – change the phase of the signal, and filter it to

    remove unwanted components.

    • Furthermore, the signal processor may be

    called upon to perform such mathematical

    operations on the signal as differentiation

    integration, or conversion to a logarithm.

    110 E.T

  • Chapter Four

    Atomic Absorption and Emission Spectroscopy

    111 E.T

  • ATOMIC SPECTROSCOPY

    • Elemental analysis at the trace or ultratrace level can be

    performed by a number of analytical techniques; however, atomic spectroscopy remains the most popular approach.

    • Atomic spectroscopy can be subdivided into three fields:

    • Atomic emission spectroscopy (AES),

    • Atomic absorption spectroscopy (AAS), and

    • Atomic fluorescence spectroscopy (AFS)

    • that differ by the mode of excitation and the method of measurement of the atom concentrations.

    112 E.T

  • • The selection of the atomic spectroscopic technique to be used for a particular application should be based on the desired result, since each technique involves different measurement approaches.

    • AES excites ground-state atoms and then quantifies the concentrations of excited-state atoms by monitoring their special deactivation.

    • AAS measures the concentrations of ground-state atoms by quantifying the absorption of spectral radiation that corresponds to allowed transitions from the ground to excited states.

    • AFS determines the concentrations of ground-state atoms by quantifying the radiative deactivation of atoms that have been excited by the absorption of discrete spectral radiation.

    113 E.T

  • 114 E.T

  • Flame Atomic Emission

    Spectroscopy • Introduction Flame photometry, now more properly called

    flame atomic emission spectrometry, is a relatively old instrumental analysis method.

    • Atomic emission is a fast, simple, and sensitive method for the determination of trace metal ions in solution.

    • Because of the very narrow (ca. 0.01 nm) and characteristic emission lines from the gas-phase atoms in the flame plasma, the method is relatively free of interferences from other elements.

    • The method is suitable for many metallic elements, especially for those metals that are easily excited to higher energy levels at the relatively cool temperatures of typical flames – Na, K, Ca, Rb, Cs, Cu, and Ba.

    115 E.T

  • Theory

    • Sample solution sprayed or aspirated as fine mist into flame. • Conversion of sample solution into an aerosol by atomizer

    (scent spray) principle. – in this stage no chemical change in the sample

    • Heat of the flame vaporizes sample constituents – Solvent vaporizes leaves solid partial of salt

    • By heat of the flame + action of the reducing gas (fuel), molecules & ions of the sample species are decomposed and reduced to give ATOMS. – eg Na+ + e- --> Na

    • Heat of the flame causes excitation of some atoms into higher electronic states.

    • Excited atoms revert to ground state by emission of light energy, of characteristic wavelength; measured by detector.

    116 E.T

  • E.T 117

  • • Flame Emission -> it measures the radiation

    emitted by the excited atoms that is related to

    concentration.

    • Atomic Absorption -> it measures the radiation

    absorbed by the unexcited atoms that are

    determined.

    • Atomic absorption depends only upon the number

    of unexcited atoms, the absorption intensity is not

    directly affected by the temperature of the flame.

    • The flame emission intensity in contrast, being

    dependent upon the number of excited atoms, is

    greatly influenced by temperature variations.

    118 E.T

  • Instrumentation:

    • Flame photometer consists of following components:

    • Pressure regulators and flow meters for fuel gases

    • Flame source

    • The atomizing device

    • Wavelength selectors: May be a filter or Monochromator

    • Photosensitive detectors

    • Read out unit.

    119 E.T

  • 120 E.T

  • 121 E.T

  • • Requirements of flame:

    • It should have proper temperature

    • Temp. should remain constant throughout operation.

    • There should not be any fluctuation during burning.

    • Function flame includes:

    – To convert the constituent of the liquid sample in to vapor state.

    – To decompose the constituent into atoms or simple molecules.

    – To electronically excite a fraction of the resulting atomic or molecular spectra.

    122 E.T

  • • Transformation of sample into vapour

    • With a pneumatic nebulizer operated by a compressed

    gas, the solution is aspirated from the sample container

    and nebulized into a mist or aerosol of fine droplets.

    • By desolvation, i.e., evaporation of the solvent from the

    droplets, this mist is converted into a dry aerosol which

    is volatilized in the flame.

    • The atomization, i.e., the conversion of volatilized

    analyte into free atoms is performed by the flame or

    other atomizer.

    123 E.T

  • • Nebulizers used in flame photometer

    • According to the source of energy used for

    nebulization as, for example, Pneumatic or

    ultrasonic nebulizers.

    • According to the relative position of the

    capillaries for the nebulizing gas and the

    aspirated liquid, e.g., angular and concentric

    nebulizers

    124 E.T

  • 125 E.T

  • • Burners

    • Flames are produced by means of a burner to which fuel and oxidant are supplied in the form of gases.

    • With the premix burner, fuel and oxidant are thoroughly mixed inside the burner housing before they leave the burner ports and enter the primary combustion or inner zone of the flame.

    • This type of burner usually produces an approximately laminar flame, and is commonly combined with a separate unit for nebulizing the sample.

    • The direct-injection burner combines the function of nebulizer and burner.

    126 E.T

  • 127 E.T

  • 128 E.T

  • • Thermal energy in flame atomization is provided by the combustion of a fuel–oxidant mixture.

    • Common fuels and oxidants and their normal temperature ranges are listed in Table 28.2. Of these, the air–acetylene and nitrous oxide-acetylene flames are used most frequently.

    129 E.T

  • • Flame temperature:

    • The optimum flame temp. depends upon several factors as excitation energy of the element, sensitivity of the measurement, Presence of other elements etc.

    • The temp. of flame lies between 1000OC and 3000OC.

    • Mixture of coal gas and air do not give very hot flames, because of presence of nitrogen.

    • The cyanogens gas produces excellent spectra but it is toxic.

    • Acetylene and hydrogen are most frequent choice

    130 E.T

  • Atomic Absorption Spectroscopy (AAS)

    • Atomic absorption spectroscopy is a quantitative method of

    analysis that is applicable to many metals and a few

    nonmetals.

    • The technique was introduced in 1955 by Walsh in Australia

    (A.Walsh, Spectrochim. Acta, 1955, 7, 108)

    • A much larger number of the gaseous metal atoms will

    normally remain in the ground state.

    • These ground state atoms are capable of absorbing radiant

    energy of their own specific resonance wavelength.

    • If light of the resonance wavelength is passed through a flame

    containing the atoms in question, then part of the light will be

    absorbed.

    • The extent of absorption will be proportional to the number of

    ground state atoms present in the flame.

    131 E.T

  • 132 E.T

  • 133 E.T

  • Instrumentation

    134 E.T

  • 135 E.T

  • 136 E.T

  • 137 E.T

  • 138 E.T

  • 139 E.T

  • 140 E.T

  • 141 E.T

  • 142 E.T

  • 143 E.T

  • 144 E.T

  • 145 E.T

  • 146 E.T

  • 147 E.T

  • 148 E.T

  • 149 E.T

  • 150 E.T

  • 151 E.T

  • Application of Flame photometer

    • Determination of Na+, K+, Ca++,Mg++ in

    biological fluids (Serum, plasma & Urine),

    Analysis of Industrial waste for pollutants

    Hardness of water

    • Determination of Heavy metal at trace and

    ulteratrace level

    E.T 152

  • Chapter Five

    UV-Visible spectroscopy

    153 E.T

  • 154 E.T

  • UV-Visible spectrum

    • The part of the electromagnetic radiation

    spectrum that you are most familiar with is

    "visible light" but

    • this is just a small portion of all the possible

    types – Violet :400 - 420 nm

    – Indigo : 420 - 440 nm

    – Blue : 440 - 490 nm

    – Green : 490 - 570 nm

    – Yellow : 570 - 585 nm

    – Orange :585 - 620 nm

    – Red : 620 - 780 nm 155 E.T

  • Ultraviolet and visible spectroscopy

    • It is used to measure the multiple bonds or atomic

    conjugation within the molecule.

    • The UV-Visible region is subdivided as below

    – Vacuum UV: 100-200 nm

    – Near UV: 200 to 400 nm

    – Visible region: 400 to 750 nm

    • Vacuum UV is so named because molecule of air

    absorb radiation in these region.

    • The radiation is assessable only in special

    vacuum equipments. 156 E.T

  • Basics of UV Light Absorption

    • Ultraviolet/visible spectroscopy involves the absorption of ultraviolet/visible light by a molecule

    • causing the promotion of an electron from a ground electronic state to an excited electronic state

    • Absorption of this relatively high-energy light causes electronic excitation.

    • The easily accessible part of this region (wavelengths of 200 to 800 nm)

    • absorption only if conjugated pi-electron systems are present.

    157 E.T

  • • The visible region of the spectrum comprises photon energies of 36 to 72 kcal/mole, and

    • the near ultraviolet region, out to 200 nm, extends this energy range to 143 kcal/mole.

    • Ultraviolet radiation having wavelengths less than 200 nm is difficult to handle, and is seldom used as a routine tool for structural analysis

    • The energies are sufficient to promote or excite a molecular electron to a higher energy orbital.

    • The absorption spectroscopy carried out in this region is sometimes called "electronic spectroscopy".

    158 E.T

  • • Electronic transitions

    – d and f electrons

    – Charge transfer reactions

    • p, s, and n (non-bonding) electrons

    159 E.T

  • Absorption Involving d and f Electrons

    • Most transition-metal ions absorb in the

    ultraviolet or visible region of the spectrum.

    • For the lanthanide and actinide series, the

    absorption process results from electronic

    transitions of 4f and 5f electrons;

    • for elements of the first and second transition-

    metal series, the 3d and 4d electrons are

    responsible for absorption.

    160 E.T

  • 161 E.T

  • D transitions

    • Partially occupied d orbitals

    – Transitions from lower to higher energy levels

    • Splitting of levels due to spatial distribution

    similar

    Axial direction

    162 E.T

  • D transitions

    • Binding ligands on axis have greater effect

    on axial orbitals

    163 E.T

  • D transitions

    • ∆ value dependent upon ligand field

    strength

  • Actinide transitions

    400 500 600 700 8000

    1

    2

    3

    4

    5

    Ab

    so

    rba

    nce

    Wavelength (nm)

    Normal

    Heavy

    Light

    Pu4+

    (489 nm)

    Pu6+

    (835 nm)

    Figure 2: UV-vis spectra of organic phases for

    13M HNO3 system 165 E.T

  • Charge-transfer Transitions

    • Electron donor and acceptor characteristics

    – Absorption involves e- transitions from donor to

    acceptor

    • SCN to Fe(III)

    – Product is Fe(II) and neutral SCN

    – Metal is acceptor

    • An exception is the 1,10--phenanthroline complex of iron(II) or copper(I),

    where the ligand is the acceptor and the metal ion is the donor.

    166 E.T

  • • p, s, and n (non-bonding) electrons

    167 E.T

  • Sigma and Pi orbitals

    168 E.T

  • Electronic transactions involved

    169 E.T

  • 170 E.T

  • Electronic spectroscopy

    • Out of the six transitions outlined, only the two

    lowest energy ones (left-most, colored blue)

    are achieved by the energies available in the

    200 to 800 nm spectrum.

    • As a rule, energetically favored electron

    promotion will be from the highest occupied

    molecular orbital (HOMO) to the lowest

    unoccupied molecular orbital (LUMO), and

    the resulting species is called an excited state

    171 E.T

  • Types of Transitions

    • There are several types of electronic transitions

    available to a molecule including:

    • σ to σ * (alkanes)

    • σ to π * (carbonyl compounds)

    • π to π * (alkenes, carbonyl compounds,

    alkynes, azo compounds)

    • n to σ * (oxygen, nitrogen, sulfur, and halogen

    compounds)

    • n to π * (carbonyl compounds)

    172 E.T

  • σ σ * Transitions

    • An electron in a bonding σ orbital is excited to the corresponding antibonding orbital. • The energy required is large.

    • For example, alkanes as methane (which has only C-H bonds, and can only undergo σ σ * transitions)

    • shows an absorbance maximum at 125 nm.

    • Absorption maxima due to σ σ * transitions are not seen in typical UV-Visible spectra (200 - 700 nm).

    • σ bonds are very strong and requires higher energy of vacuum UV.

    173 E.T

  • s smax = 135 nm (a high energy transition)

    Absorptions having max < 200 nm are difficult to observe because everything (including quartz glass and air) absorbs in this spectral region.

    Ethane

    174 E.T

  • nσ * Transitions

    • Saturated compounds containing atoms with lone pairs (non-bonding electrons) are capable of n σ * as transitions.

    • These transitions usually need less energy than σ σ * transitions.

    • They can be initiated by light whose wavelength is in the range 150 - 250 nm.

    • The number of organic functional groups with n σ * peaks in the UV region is small.

    • These transitions are involved in saturated compound with one hetero atom with unshared pair of electron i.e. saturated halides, ethers, aldehydes, ketones, amines etc.

    • These transitions are sensitive to hydrogen bonding eg alcohol and ethers which absorbs at wavelength shorter than 185 nm therefore used as a solvent in UV

    175 E.T

  • π π * Transitions

    • For molecules that possess π bonding as in alkenes, alkynes, aromatics, acyl compounds or nitriles,

    • energy that is available can promote electrons from a π Bonding molecular orbital to a π Antibonding molecular orbital (π * ).

    • This is called a π π * transition.

    • The absorption peaks for these transitions fall in an experimentally convenient region of the spectrum (200 - 700 nm).

    • These transitions need an unsaturated group in the molecule to provide the π electrons.

    176 E.T

  • C C

    s

    s

    hv

    p

    p

    s

    s

    p

    p

    p pExample: ethylene absorbs at longer wavelengths:

    = hv =hc/

    177 E.T

  • n π * Transitions

    • Lone pairs that exist on Oxygen atoms and Nitrogen atoms may be promoted from their non-bonding molecular orbital to a π antibonding molecular orbital within the molecule.

    • This is called an n π * transition and

    • requires less energy (longer wavelength) compared to a π π * transition within the same chromophore.

    • These are available in compounds with unsaturated centers. eg. Alkenes. They requires lowest energy as compare to others

    178 E.T

  • s

    s

    hv

    p

    p

    n

    s

    s

    p

    p

    n

    C O

    pn

    179 E.T

  • C C

    HOMO

    LUMO

    Conjugated systems:

    Preferred transition is between Highest Occupied Molecular Orbital

    (HOMO) and Lowest Unoccupied Molecular Orbital (LUMO).

    Note: Additional conjugation (double bonds) lowers the HOMO-LUMO energy gap: Example: 1,3 butadiene: max = 217 nm ; = 21,000 1,3,5-hexatriene max = 258 nm ; = 35,000

    180 E.T

  • O

    O

    O

    Similar structures have similar UV spectra:

    max = 238, 305 nm max = 240, 311 nm max = 173, 192 nm

    181 E.T

  • E.T 182

    Table: absorption characteristics of some common chromophores

  • E.T 183

    Activity one:

    Transitions n-> σ *, π -> π*, n->π* , σ -> σ*, σ ->π*

    Choose possible transitions from the above list for each

    chromophores.

    R-OH n-> σ *

    R-O-R n-> σ *

    R-CH2-R σ -> σ*

    R-C≡C-R π -> π*

    R-C≡N n->π*, π -> π*

    R-CHO π -> π*, n->π*

    RCOOH n->π*

  • Instrumentation: :

    184 E.T

  • Figure: Instrumental designs for UV-visible photometers or spectrophotometers. In (a), a single-beam instrument is shown. Radiation from the filter or monochromator passes through either the reference cell or the sample cell before striking the photodetector. In (b), a double-beam-in-space instrument is shown. Here, radiation from the filter or monochromator is split into two beams that simultaneously pass through the reference and sample cells before striking two matched photodetectors. In the double-beam-in-time instrument (c), the beam is alternately sent through reference and sample cells before striking a single photodetector. only a matter of milliseconds separate the beams as they pass through the two cells.

    185 E.T

  • Instrumentation: • A beam of light from a visible and/or UV light source

    is separated into its component wavelengths by a

    prism or diffraction grating.

    • Each monochromatic (single wavelength) beam in

    turn is split into two equal intensity beams by a half-

    mirrored device.

    • One beam, the sample beam , passes through a small

    transparent container (cuvette) containing a solution

    of the compound being studied in a transparent

    solvent.

    • The other beam, the reference , passes through an

    identical cuvette containing only the solvent. 186 E.T

  • • The intensities of these light beams are then measured

    by electronic detectors and compared.

    • The intensity of the reference beam, which should

    have suffered little or no light absorption, is defined

    as I0.

    • The intensity of the sample beam is defined as I.

    • Over a short period of time, the spectrometer

    automatically scans all the component wavelengths in

    the manner described.

    • The ultraviolet (UV) region scanned is normally from

    200 to 400 nm, and the visible portion is from 400 to

    800 nm.

    187 E.T

  • Solvent effect:

    • Different compounds may have very different

    absorption maxima and absorbances.

    • Intensely absorbing compounds must be examined in

    dilute solution, so that significant light energy is

    received by the detector, and this requires the use of

    completely transparent (non-absorbing) solvents.

    • The most commonly used solvents are water,

    ethanol, hexane and cyclohexane.

    • Solvents having double or triple bonds, or heavy

    atoms (e.g. S, Br & I) are generally avoided.

    188 E.T

  • • Solvents must be transparent in the region to

    be observed; the wavelength where a solvent

    is no longer transparent is referred to as the

    cutoff

    189 E.T

    Table: solvents for the UV and Visible regions

  • 190 E.T

    • H-bonding further complicates the effect of vibrational and

    rotational energy levels on electronic transitions, dipole-

    dipole interacts less so

    • The more non-polar the solvent, the better (this is not

    always possible)

    Figure: effect of solvent on the absorption spectrum of acetaldehy

  • 191

    The Spectrum

    1. The x-axis of the spectrum is in wavelength; 200-350 nm for UV,

    200-700 for UV-VIS determinations

    2. Due to the lack of any fine structure, spectra are rarely shown in their

    raw form, rather, the peak maxima are simply reported as a numerical

    list of “lambda max” values or λmax

    λmax = 206 nm

    252

    317

    376

    O

    NH2

    O

    E.T

  • 192

    The Spectrum

    1. The y-axis of the spectrum is in absorbance, A

    2. From the spectrometers point of view, absorbance is the inverse of

    transmittance: A = log10 (p0/p)

    3. From an experimental point of view, three other considerations must

    be made:

    i. a longer path length, l through the sample will cause

    more UV light to be absorbed – linear effect

    ii. the greater the concentration, c of the sample, the more

    UV light will be absorbed – linear effect

    iii. some electronic transitions are more effective at the

    absorption of photon than others – molar absorptivity, ε

    this may vary by orders of magnitude…

    192 E.T

  • 193

    4. These effects are combined into the Beer-Lambert Law: A = ε c l

    i. for most UV spectrometers, l would remain constant (standard

    cells are typically 1 cm in path length)

    ii. concentration is typically varied depending on the strength of

    absorption observed or expected – typically dilute – sub .001

    M

    iii. molar absorptivities vary by orders of magnitude:

    • values of 104-106 are termed high intensity absorptions

    • values of 103-104 are termed low intensity absorptions

    • values of 0 to 103 are the absorptions of forbidden

    transitions

    A is unit less, so the units for ε are cm-1 · M-1

    5. Since path length and concentration effects can be easily factored

    out, absorbance simply becomes proportional to ε, and the y-axis is

    expressed as e directly or as the logarithm of ε

    E.T

  • E.T 194

    The Spectrum

  • 195

    Terminology Chromophore: • Chromophore: A covalently unsaturated group

    responsible for electronic absorption.

    • Or Any group of atoms that absorbs light whether or not

    a color is thereby produced. Eg C=C, C=O, NO2 etc

    • A compound containing chromophore is called

    chromogen.

    • There are two types of chromophore

    – Independent chromophore: single chromophore is sufficient to

    import color to the compound eg. Azo group

    – Dependent chromophore: When more then one chromophore is

    required to produce color.

    • Eg acetone having 1 kentone group is colorless where as diacetyl

    having two kentone group is yellow

    E.T

  • Auxochrome:

    • A saturated group with non bonding electron when

    attached to chromophore alters both wavelengths as

    well as intensity of absorption. Eg OH, NH2, NHR

    etc.

    OR

    • A group which extends the conjugation of a

    chromophore by sharing of nonbonding electrons.

    196 E.T

  • Substituent Effects

    • General – Substituents may have any of four

    effects on a chromophore

    • Bathochromic shift (red shift) – a shift to

    longer λ; lower energy

    • Hypsochromic shift (blue shift) – shift to

    shorter λ; higher energy

    • Hyperchromic effect – an increase in intensity

    • Hypochromic effect - a decrease in intensity

    197 E.T

  • 200 nm 700 nm

    H

    ypochro

    mic

    Hypsochromic

    Hyperc

    hro

    mic

    Bathochromic

    E.T

  • 199

    • Substituent Effects

    1. Conjugation – most efficient means of bringing about a

    bathochromic and hyperchromic shift of an unsaturated

    chromophore:

    H2CCH2

    -carotene

    O

    O

    max nm 175 15,000

    217 21,000

    258 35,000

    n p* 280 27 p p* 213 7,100

    465 125,000

    n p* 280 12 p p* 189 900

    E.T

  • 200

    • Substituent Effects

    2. Conjugation – Alkenes

    Extending this effect out to longer conjugated systems the

    energy gap becomes progressively smaller:

    Energy

    ethylene

    butadiene

    hexatriene

    octatetraene

    Lower energy = Longer wavelengths

    E.T

  • Activity 1

    What type of shift observed in 1&2,3&5 and

    4&6

    • max ᵋmax 1 .CH3-COH 293 nm 12

    2. CH3-COCl 235 nm 53

    E.T 201

    5. n p* 280 27 6. p p* 213 7,100

    3. n p* 280 12 4. p p* 189 900

    Hypsoc

    hromic

    &

    Hyperc

    hromic

    Bathoc

    hromic

    &

    Hyperc

    hromic

    Hyperc

    hromic 3&5 4&6

  • 202

    • We will find that the effect of substituent groups can be reliably

    quantified from empirical observation of known conjugated

    structures and applied to new systems

    • This quantification is referred to as the Woodward-Fieser Rules

    which we will apply to two specific chromophores:

    1. Conjugated dienes

    2. Aromatic systems

    E.T

    A. Dienes

    1. General Features

    For acyclic butadiene, two conformers are possible – s-cis and s-trans

    s-trans s-cis

    The s-cis conformer is at an overall higher potential energy than the s-trans;

    therefore the HOMO electrons of the conjugated system have less of a jump to

    the LUMO – lower energy, longer wavelength

  • 203

    A. Dienes 1. General Features Two possible p p* transitions can occur for butadiene Y2 Y3

    and Y2 Y4

    *

    The Y2 Y4

    * transition is not typically observed:

    • The energy of this transition places it outside the region typically observed – 175 nm

    • For the more favorable s-trans conformation, this transition is forbidden

    The Y2 Y3* transition is observed as an intense absorption

    s-trans s-cis

    175 nm –forb.

    217 nm 253 nm

    175 nm

    Y4

    Y2

    Y1

    Y3

    E.T

  • 204

    A. Dienes

    1. General Features The Y2 Y3

    * transition is observed as an intense absorption ( =

    20,000+) based at 217 nm within the observed region of the UV While this band is insensitive to solvent (as would be expected) it is

    subject to the bathochromic and hyperchromic effects of alkyl substituents as well as further conjugation

    Consider:

    max = 217 253 220 227 227 256 263 nm

    E.T

  • 205

    A. Dienes 2. Woodward-Fieser Rules - Dienes The rules begin with a base value for max of the chromophore being

    observed: acyclic butadiene = 217 nm The incremental contribution of substituents is added to this base value

    from the group tables:

    Group Increment

    Extended conjugation +30

    Each exo-cyclic C=C +5

    Alkyl +5

    -OCOCH3 +0

    -OR +6

    -SR +30

    -Cl, -Br +5

    -NR2 +60 E.T

  • 206

    A. Dienes

    2. Woodward-Fieser Rules - Dienes For example: Isoprene - acyclic butadiene = 217 nm one alkyl subs. + 5 nm 222 nm Experimental value 220 nm Allylidenecyclohexane - acyclic butadiene = 217 nm one exocyclic C=C + 5 nm 2 alkyl subs. +10 nm 232 nm Experimental value 237 nm

    E.T

  • E.T 207

    Calculate wavelengths for the fo