Chan Scope and Dew Point Tester Manual

64
PN 13-2270 REV L Chandler ChanScope and Dew Point Tester User Manual Process Instruments 150 Freeport Road Pittsburgh, PA 15238

description

Dew point tester manual

Transcript of Chan Scope and Dew Point Tester Manual

Page 1: Chan Scope and Dew Point Tester Manual

PN 13-2270 REV L

ChandlerChanScope and Dew Point Tester

User Manual

Process Instruments 150 Freeport Road

Pittsburgh, PA 15238

Page 2: Chan Scope and Dew Point Tester Manual

ii | AMETEK ChanScope and Dew Point Tester

USA - HEADQUARTERSAMETEK Process Instruments150 Freeport RoadPittsburgh, PA 15238, USAPhone: 412-828-9040Fax: 412-826-0399

USA - DelawareAMETEK Process Instruments455 Corporate BoulevardNewark, Delaware 19702, USAPhone: 302-456-4400 (Main) 800-537-6044 (Service) 800-222-6789 (Ordering)Fax: 302-456-4444

USA - TexasAMETEK Process Instruments9750 Whithorn DriveHouston, Texas 77095, USAPhone: 281-463-2820Fax: 281-463-2701

USA - OklahomaChandler Engineering2001 North Indianwood AvenueBroken Arrow, OK 74012, USAPhone: 918-250-7200Fax: 918-459-0165

CANADAAMETEK Process Instruments2876 Sunridge Way N.E.Calgary, AlbertaCanada T1Y 7H9Phone: 403-235-8400 Fax: 403-248-3550

GERMANYAMETEK GmbHPostfach 2165D-40644 Meerbusch ORRudolf-Diesel Strasse 16D-40670 MeerbuschGermanyPhone: 49-21-59-9136-0Fax: 49-21-59-9136-39

SINGAPOREAMETEK Singapore PVT. Ltd.10 Ang Mo Kio Street 65#05-12 TechpointSingapore 569059Republic of SingaporePhone: 65-484-2388Fax: 65-481-6588

FRANCEAMETEK - APIFRond Point de l’épine des champsBuroplus Bat D78990 Elancourt, FrancePhone: 33 1 30 68 89 20Fax: 33 1 30 38 89 29

CHINAAMETEK Process InstrumentsRoom 406, Metro TowerNo. 30 Tian Yao Qiao RoadShanghai, 200030Phone: 8621 6426 7049Fax: 8621 6426 7054

Offices

© 2005 AMETEK

This manual is a guide for the use of the AMETEK ChanScope and Dew Point Tester models. Data herein has been verified and validated and is believed adequate for the intended use of this software. If the software or procedures are used for purposes over and above the capabilities specified herein, confirmation of their validity and suitability should be obtained; otherwise, AMETEK does not guarantee results and assumes no obligation or liability. This publication is not a license to operate under, or a recommendation to infringe upon, any process patents.

Page 3: Chan Scope and Dew Point Tester Manual

| iii

Table of ContentsCHAPTER 1 Overview

Dew Point Tester ................................................................................................. 1-1Description ......................................................................................... 1-1

Bureau of Mines-Type Dew Point Tester (Standard) ................... 1-1ChanScope ................................................................................... 1-1

Theory of Operation ........................................................................................... 1-2Additional Equipment ......................................................................................... 1-3

Chiller ................................................................................................ 1-3Illuminated Magnifier ........................................................................ 1-3Tripod ................................................................................................. 1-3Propane Cylinder ............................................................................... 1-3CO2 Bottle .......................................................................................... 1-3

Part Numbering Codes ........................................................................................ 1-4Technical Support ............................................................................................... 1-5

CHAPTER 2 Specifications Standard Bureau of Mines-type Dew Point Tester ............................................. 2-1

Accuracy ............................................................................................ 2-1Dew Point Temperature Range .......................................................... 2-1Operating Pressure ............................................................................. 2-1Connections ........................................................................................ 2-1Weight and Dimensions ..................................................................... 2-1

ChanScope Digital Dew Point Tester ................................................................. 2-2Accuracy ............................................................................................ 2-2Dew Point Temperature Range .......................................................... 2-2Operating Pressure ............................................................................. 2-2Connections ........................................................................................ 2-2Power ................................................................................................. 2-2Weight and Dimensions ..................................................................... 2-2

CHAPTER 3 Installation and Set Up Inspect Shipping Contents .................................................................................. 3-1Safety Requirements ........................................................................................... 3-2Mechanical Installation ....................................................................................... 3-3

Tools/Equipment Required ................................................................ 3-3Tubing and Fittings ............................................................................ 3-3Installing the Dew Point Tester .......................................................... 3-4Connecting the Refrigerant ................................................................ 3-4Installing the Glycol Filter Assembly ................................................ 3-5Selecting Temperature Display (ChanScope only) ............................ 3-5

Start Up ............................................................................................................... 3-6

Page 4: Chan Scope and Dew Point Tester Manual

iv | AMETEK ChanScope and Dew Point Tester

CHAPTER 4 OperationUsing Glycol Filters ............................................................................................ 4-1

Determining Equilibrium ................................................................... 4-1Test Precautions .................................................................................................. 4-2Test Procedure..................................................................................................... 4-4Confirmation of Dew Point ................................................................................. 4-5

Interpretation of Dew Point Temperature .......................................... 4-5Differentiating between Water and Hydrocarbon Dew Points............................ 4-6Special Test Procedure for Propane .................................................................... 4-7

CHAPTER 5 Maintenance and Troubleshooting Maintenance ........................................................................................................ 5-1

Tools Required ................................................................................... 5-1Replacing the Glycol Filter ................................................................ 5-2Cleaning the Window and Mirror ...................................................... 5-2Recharging Batteries (ChanScope Dew Point Tester) ....................... 5-2Bypassing the RTD ............................................................................ 5-3Calibration .......................................................................................... 5-4Selecting Temperature Display (ChanScope only) ............................ 5-4

Troubleshooting Guide ....................................................................................... 5-5Replacement Parts .............................................................................................. 5-9ChanScope Digital Dew Point Tester ............................................................... 5-11

Models 13-175, 13-200, 13-210 (10,000 Series SN) ....................... 5-11ChanScope II Digital Dew Point Tester ............................................................ 5-12

Models 13-1175, 13-1200, 13-1210 (20,000 Series SN) ................. 5-12

CHAPTER 6 Drawings DWG 13-0052 - Glass Window Assy (sour gas) ................................................ 6-1DWG 13-0054 - Glycol Filter Assy .................................................................... 6-2DWG 13-0072 - Immersion Cooling Rod Assy ................................................. 6-3DWG 13-0075 - Model 13-0075 Dew Point Tester ............................................ 6-4DWG 13-0090 - Chillers, Regular and Quick .................................................... 6-5DWG 13-0095 - Saturated Moisture Content Air or Nitrogen (°F) .................... 6-6DWG 13-0095 - Saturated Moisture Content Air or Nitrogen (°C) ................... 6-7DWG 13-0097 - Saturated Moisture Content Natural Gas (°F) ......................... 6-8DWG 13-0097 - Saturated Moisture Content Natural Gas (°C) ......................... 6-9Table - Equilibrium Water Vapor Contents ....................................................... 6-10DWG 13-0100 - Bureau of Mines Dew Point Tester ........................................ 6-15DWG 13-0125 - Conversion Chart, Dew Point to Moisture Content ............... 6-16DWG 13-0150 - Installation, Dew Point Tester ............................................... 6-17DWG 13-0250 - Dew Point Indicator Assy ...................................................... 6-18DWG 13-0303 - Illuminated Magnifier Assy (Model 13-075) ......................... 6-19DWG 13-0304 - Illuminated Magnifier (Models 13-100, 13-110) ................... 6-20DWG 13-0305 - Illuminator Assy .................................................................... 6-21DWG 13-1250 - ChanScope II ......................................................................... 6-22

Page 5: Chan Scope and Dew Point Tester Manual

| v

Safety Notes

wARNiNgs, cAuTioNs, and NoTEs contained in this manual emphasize critical instructions as fol-lows:

An operating procedure which, if not strictly observed, may result in personal injury or envi-ronmental contamination.

An operating procedure which, if not strictly observed, may result in damage to the equip-ment.

Important information that should not be overlooked.

Electrical Safety

Always shut down power source(s) before performing maintenance or troubleshooting. Only a qualified electrician should make electrical connections and ground checks.

Any use of the equipment in a manner not specified by the manufacturer may impair the safety protection originally provided by the equipment.

Grounding

Instrument grounding is mandatory. Performance specifications and safety protection are void if instru-ment is operated from an improperly grounded power source.

Verify ground continuity of all equipment before applying power.

NOTE

Page 6: Chan Scope and Dew Point Tester Manual

vi | AMETEK ChanScope and Dew Point Tester

PRoTEcTiVE coNducToR TERMiNAL(BoRNiER dE L’EcRAN dE PRoTEcTioN)schutzerde

cAuTioN - Risk of electric shock(ATTENTioN-RisQuE dE dÉcHARgE ÉLEcTRiQuE)Achtung - Hochspannung Lebensgefahr

cAuTioN - (Refer to accompanying documents)(ATTENTioN-sE RÉFERER AuX docuMENTs JoiNTs)Achtung (Beachten sie beiliegende dokumente)

cAuTioN - Hot surface

(ATTENTioN-suRFAcE cHAudE)Achtung - Heiße Oberfläche

Warning Labels

These symbols may appear on the instrument in order to alert you of existing conditions.

Environmental Information (WEEE)

This AMETEK product contains materials that can be reclaimed and recycled. In some cases the product may contain materials known to be hazardous to the environment or human health. In order to prevent the release of harmful substances into the environment and to conserve our natural resources, AMETEK recommends that you arrange to recycle this product when it reaches its “end of life.”

Waste Electrical and Electronic Equipment (WEEE) should never be disposed of in a municipal waste system (residential trash). The Wheelie Bin marking on this product is a reminder to dispose of the prod-uct properly after it has completed its useful life and been removed from service. Metals, plastics and other components are recyclable and you can do your part by one of the following these steps:

• When the equipment is ready to be disposed of, take it to your local or re-gional waste collection administration for recycling.

• In some cases, your “end-of-life” product may be traded in for credit towards the purchase of new AMETEK instruments. Contact your dealer to see if this program is available in your area.

• If you need further assistance in recycling your AMETEK product, contact our office listed in the front of the instruction manual.

Page 7: Chan Scope and Dew Point Tester Manual

| vii

WARRANTY AND CLAIMS

We warrant that any equipment of our own manufacture or manufactured for us pursuant to our specifications which shall not be, at the time of shipment thereof by or for us, free from defects in material or workmanship under normal use and service will be repaired or replaced (at our option) by us free of charge, provided that written notice of such defect is received by us within twelve (12) months from date of shipment of portable analyzers or within eighteen (18) months from date of shipment or twelve (12) months from date of installation of permanent equipment, whichever period is shorter. All equipment requiring repair or replacement under the warranty shall be returned to us at our factory, or at such other location as we may designate, transportation prepaid. Such returned equipment shall be examined by us and if it is found to be defective as a result of defective materials or workmanship, it shall be repaired or replaced as aforesaid. Our obligation does not include the cost of furnishing any labor in connection with the installation of such repaired or replaced equipment or parts thereof, nor does it include the responsibility or cost of transportation. In addition, instead of repairing or replacing the equipment returned to us as aforesaid, we may, at our option, take back the defective equipment, and refund in full settlement the purchase price thereof paid by Buyer.

Process photometric analyzers, process moisture analyzers, and sampling systems are warranted to perform the intended measurement, only in the event that the customer has supplied, and AMETEK has accepted, valid sample stream composition data, process conditions, and electrical area classification prior to order acknowledgment. The photometric light sources are warranted for ninety (90) days from date of shipment. Resale items warranty is limited to the transferable portion of the original equipment manufacturer’s warranty to AMETEK. If you are returning equipment from outside the United States, a statement should appear on the documentation accompanying the equipment being returned declaring that the goods being returned for repair are American goods, the name of the firm who purchased the goods, and the shipment date.

The warranty shall not apply to any equipment (or part thereof) which has been tampered with or altered after leav-ing our control or which has been replaced by anyone except us, or which has been subject to misuse, neglect, abuse or improper use. Misuse or abuse of the equipment, or any part thereof, shall be construed to include, but shall not be limited to, damage by negligence, accident, fire or force of the elements. Improper use or misapplications shall be construed to include improper or inadequate protection against shock, vibration, high or low temperature, overpressure, excess voltage and the like, or operating the equipment with or in a corrosive, explosive or combustible medium, un-less the equipment is specifically designed for such service, or exposure to any other service or environment of greater severity than that for which the equipment was designed.

The warranty does not apply to used or secondhand equipment nor extend to anyone other than the original purchaser from us.

THIS WARRANTY IS GIVEN AND ACCEPTED IN LIEU OF ALL OTHER WARRANTIES, WHETHER EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION AND WARRANTIES OF FITNESS OR OF MERCHANT-ABILITY OTHER THAN AS EXPRESSLY SET FORTH HEREIN, AND OF ALL OTHER OBLIGATIONS OR LI-ABILITIES ON OUR PART. IN NO EVENT SHALL WE BE LIABLE UNDER THIS WARRANTY OR ANY OTHER PROVISION OF THIS AGREEMENT FOR ANY ANTICIPATED OR LOST PROFITS, INCIDENTAL DAMAGES, CONSEQUENTIAL DAMAGES, TIME CHANGES OR ANY OTHER LOSSES INCURRED BY THE ORIGINAL PURCHASER OR ANY THIRD PARTY IN CONNECTION WITH THE PURCHASE, INSTALLATION, REPAIR OR OPERATION OF EQUIPMENT, OR ANY PART THEREOF COVERED BY THIS WARRANTY OR OTHERWISE. WE MAKE NO WARRANTY, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION ANY WARRAN-TIES OF FITNESS OR OF MERCHANTABILITY, AS TO ANY OTHER MANUFACTURER’S EQUIPMENT, WHETHER SOLD SEPARATELY OR IN CONJUNCTION WITH EQUIPMENT OF OUR MANUFACTURE. WE DO NOT AUTHORIZE ANY REPRESENTATIVE OR OTHER PERSON TO ASSUME FOR US ANY LIABILITY IN CONNECTION WITH EQUIPMENT, OR ANY PART THEREOF, COVERED BY THIS WARRANTY.

Page 8: Chan Scope and Dew Point Tester Manual

viii | AMETEK ChanScope and Dew Point Tester

This page intentionally left blank.

Page 9: Chan Scope and Dew Point Tester Manual

Overview | 1-1

OVERVIEW

Dew Point Tester

Description

Bureau of Mines-Type Dew Point Tester (Standard)

When a gas-water vapor is cooled with no liquid water present, the absolute hu-midity of the vapor content remains constant but the relative humidity increases until it reaches 100% saturation, after which moisture begins to condense. The temperature at which condensation is first detected is known as the dew point. The Bureau of Mines-type dew point tester duplicates the conditions required by the definition of dew point. However, the measurement method is modified so that it is suitable for measuring dew point at high pressures. This is done by enclosing a mirror upon which moisture or hydrocarbons condense (when chilled by a refrigerant) in a pressure chamber. A window in the chamber allows viewing of moisture condensation.

ChanScope

The digital ChanScope provides state-of-the-art temperature measurement of the dew point of pressurized gas. The ChanScope retains all of the proven reliabil-ity and quality of the original Bureau of Mines-type dew point tester and gives instantaneous feedback using an LEDreadout during a test.

Both the standard (Bureau of Mines type) dew point tester and the ChanScope include a sample pressure gauge and tripod socket for ease of use in the field.

All AMETEK Chandler dew point testers meet ASTM D1142-GPA 2140 standards.

NOTE

Page 10: Chan Scope and Dew Point Tester Manual

1-2 | AMETEK ChanScope and Dew Point Tester

Theory of Operation

Water contamination in gases can cause corrosion and other equipment problems, especially in regulation and metering stations. The determination of hydrocarbon contamination in pipeline gas is necessary to meet hydrocarbon dew point speci-fications in natural gas transportation contracts. Gases most frequently tested for water content include natural gas, sour gas, heat treating furnace gas, and indus-trial gases such as nitrogen, oxygen, hydrogen, and other bottled gases. The most accurate and widely used instrument available for determining water content is the Bureau of Mines-type chilled mirror dew point tester. When this instrument is used properly, measured dew point accuracy is ± 0.2°F.

Each dew point tester incorporates a pressure-tight chamber to contain the gas or vapor test sample and valves for controlling gas flow. There is a clear plastic win-dow at one end of the chamber to observe the interior and the mirror. The mirror itself is highly polished stainless steel and is attached at its center to a copper thermowell. A chiller attached to the thermowell controls expansions of refriger-ant and the temperature of the mirror. The greater the flow rate of refrigerant, the lower the temperature. The mirror’s temperature is indicated by a thermometer and uses a bulb that is located close to the back surface of the mirror. An elec-tronic Resistance Temperature Detection (RTD) is used on some model dew point testers.

Page 11: Chan Scope and Dew Point Tester Manual

Overview | 1-3

Additional Equipment

Chiller

There are three types of chiller devices available for use with the standard dew point tester or the ChanScope.

• Standard Chiller• Quick Chiller• Liquid Nitrogen Chiller

The Quick Chiller and the Liquid Nitrogen Chiller allow much faster dew point determination but there is an increased risk of error and temperature overshoot if components are not allowed to reach thermal equilibrium.

• The Quick Chiller can be used for dew point measurements down to -20°F (-30°C) with propane, and down to -90°F (-70°C) with liquid CO2 refriger-ant.

• The Nitrogen Chiller is a dipstick type and can be used with liquid N2 for dew points down to and below -200°F (-130°C). The Nitrogen Chiller can also be used with dry ice/acetone for dew points down to -100°F (-75°C).

Certified Thermometers

Certified thermometers that have been tested for accuracy are also available. They are supplied with calibration traceable to NIST.

Illuminated Magnifier

A four-power, prefocused magnifier is available as an option, and can be attached to the instrument to illuminate and magnify the mirror. The battery-illuminated magnifier ensures optimum accuracy of dew point determinations regardless of field lighting. These illuminated magnifiers are available for the following dew point tester models: Model 13-075 (large window, 1000 PSIG) and Models 13-100 (small window, 3000 PSIG), and 13-110 (small window, 5000 PSIG). The magnifier is a simple design and is easy to operate. It is held on the dew point tester using an O-ring.

Page 12: Chan Scope and Dew Point Tester Manual

1-4 | AMETEK ChanScope and Dew Point Tester

Glycol Filter

Glycol vapor contained in a gas that has passed through a glycol dehydrator will condense on the tester’s mirror complicating observation of water dew point. A glycol filter assembly can be installed in the tester to remove the glycol vapor before the contaminant enters the instrument. The filter is especially useful when the glycol condensation temperature is above the water dew point.

Tripod

The tripod is lightweight, rigid aluminum. It has two section legs that are extend-able to 54 1/2” for support in the field.

Propane Cylinder

The propane cylinder is a size 20 DOT-approved cylinder with a delivery control valve and a 1/4” male NPT thread. The capacity of the cylinder is 3.75 gallons (14.2 liters) of propane. The cylinder is shipped empty and can be filled at a local gas supplier.

CO2 Bottle

The CO2 bottle is a size 20 DOT-approved cylinder with a dip tube, delivery con-trol valve, and a 1/4” male NPT thread. The capacity of the cylinder is 5 lb (2.3 kg) of CO2. The cylinder can be purchased empty, or full of welding grade liquid CO2.

Page 13: Chan Scope and Dew Point Tester Manual

Overview | 1-5

Pressure Range075 0-1000 PSIG Model B100 0-3000 PSIG Model A2110 0-5000 PSIG Model A2

Chiller DeviceC - Standard ChillerQ - Quick ChillerL - Liquid Nitrogen Chiller

ApplicationN - Normal ServiceR - Off-Shore ServiceS - Sour Gas ServiceX - Cleaned for O2 Service

Standard Thermometers025 -30 to 110°F027 -90 to +40°F038 -100 to +10°C060 -35 to +50°C096 -200 to 0°F099 -0 to +220°F

Certified Thermometers146 -90 to +40°F148 -200 to 0°F149 -30 to 110°F152 -100 to +10°C157 -35 to +50°C

13-

13- 100- C N 02513-100-C-N-025: Typical Part Number

Part Numbering Code for Standard Dew Point Tester

Pressure Range1175 0-1000 PSIG1200 0-3000 PSIG1210 0-5000 PSIG

Chiller DeviceC - Standard ChillerQ - Quick ChillerL - Liquid Nitrogen Chiller

ApplicationN - Normal ServiceR - Off-Shore ServiceS - Sour Gas ServiceX - Cleaned for O2 Service

Battery Charger Power1 - 115 VAC 50/60 Hz2 - 230 VAC 50/60 Hz

13-

13- 1200- C N 113-1200-C-N-1: Typical Part Number

Part Numbering Code for Digital ChanScope Dew Point Tester

All dew point testers come with spare O-rings, ten feet of stainless steel sample tubing with stainless fittings, and ten feet of stainless tubing with stainless fittings for refrigerant gas, carrying case, and manual.

Part Numbering Codes

Page 14: Chan Scope and Dew Point Tester Manual

1-6 | AMETEK ChanScope and Dew Point Tester

Technical Support

AMETEK is committed to providing you with the best technical support in the industry. If you need service or application assistance, please call Chandler at 918-250-7200, or call your local AMETEK/Chandler representative.

If you need to return equipment, you will be asked to provide the following infor-mation before obtaining a Return Material Authorization (RMA) number.

• Billing and shipping address

• Model number

• Serial number

• Purchase order number

• Telephone number

Before returning material, you must get an RMA number from the factory.NOTE

Page 15: Chan Scope and Dew Point Tester Manual

Specifications | 2-1

SPECIFICATIONS

Standard Bureau of Mines-type Dew Point Tester

Accuracy:

± 0.5°F (± 0.25°C)

Dew Point Temperature Range:

Dependent on which thermometer is chosen.

Operating Pressure:

Model 13-075 Standard Tester: 0 to 1000 PSIG (6.9 mPa)Model 13-100 Standard Tester: 0 to 3000 PSIG (20.7 mPa)Model 13-110 Standard Tester: 0 to 5000 PSIG (34.6 mPa)

Connections:

Sample inlet and outlet: 1/8” NPTRefrigerant: 1/8” tube

Weight and Dimensions

Weight: 25 lb (11 kg)Carrying Case: 14” x 18” x 7” (35.5 cm x 20 cm x 17.8 cm)

Page 16: Chan Scope and Dew Point Tester Manual

2-2 | AMETEK ChanScope and Dew Point Tester

ChanScope Digital Dew Point Tester

Accuracy:

± 0.2°C (± 0.4°F) at 40°C (104°F) down to -90°C (-130°F)

Dew Point Temperature Range:

-129°C (-200°F) to ambient

Ambient Operating Temperature Range:

-20°C (-4°F) to 40°C (104°F)

Operating Pressure:

Model 13-1175 ChanScope Tester: 0 to 1000 PSIG (6.9 mPa)Model 13-1200 ChanScope Tester: 0 to 3000 PSIG (20.7 mPa)Model 13-1210 ChanScope Tester: 0 to 5000 PSIG (34.6 mPa)

Connections:

Sample inlet and outlet: 1/8” NPTRefrigerant: 1/8” tube

Power

Battery-powered (NiMH), 100-240 VAC, 50/60 Hz Smart Charger with three interchangeable blades

Weight and Dimensions

Weight: 25 lb (11 kg)Carrying Case: 14” x 18” x 7” (35.5 cm x 20 cm x 17.8 cm)

Page 17: Chan Scope and Dew Point Tester Manual

Installation and Set Up | 3-1

INSTALLATION AND SET UP

This chapter shows you how to set up your dew point tester. It includes the fol-lowing:

• Inspecting Shipping Contents

• Safety Requirements

• Mechanical Installation- Equipment Required- Tubing and Fittings- Refrigerant Connections- Glycol Filter Assembly- Temperature Display Units (ChanScope only)

• Start-Up- Observing the dew point temperature

Inspect Shipping Contents

The dew point tester and its accessories were inspected before being shipped from the factory. Remove any packing material from the units. Check for dam-age and verify all parts on the packing list were shipped. If any damage is found, notify AMETEK as soon as possible and notify the shipper.

All equipment is calibrated and tested prior to shipment.NOTE

Page 18: Chan Scope and Dew Point Tester Manual

3-2 | AMETEK ChanScope and Dew Point Tester

Safety Requirements

• Vent the sample gas to a safe area to prevent the possibility of creating an explosive atmosphere in an enclosed area.

• Always wear safety glasses when working with refrigerant and high pressure gases.

• Do not exceed the pressure rating of the instrument. Refer to Chapter 2 “Specifications”forinstrumentoperatingpressures.

If operating pressure is exceeded, component “blow out” could result in serious injury to the operator.

• Clean the instrument prior to using it with oxygen.

• FormodelsequippedwithChanScopedigitaldewpointindicators:

Do not expose the battery or internal electronics to temperatures higher than the range -20°C to 70°C (-4°F to 148°F). Battery packs should be recycled at an appropriate facility for Ni-MH batteries. Use only Chandler part #13-1274 batteries as replacements. Using other battery packs will cause damage to the electronics and void your warranty.

Use extreme caution if you are using the instrument with oxygen or any other explosive gas reactive to oil. The entire instrument and all sample tubing must be thoroughly cleaned.

No utilities are required for the standard dew point tester. The Chan-Scope requires 100-265 VAC, 50/60 Hz for the battery charger.

NOTE

Page 19: Chan Scope and Dew Point Tester Manual

Installation and Set Up | 3-3

Mechanical Installation

Tools/Equipment Required

• 7/16”,1/2”open-endedwrenches,10”adjustablewrench

• Isopropyl alcohol (99.5% pure)

• Cotton balls or swabs

• Air duster

• Teflontape

• Refrigerant: propane, liquid CO2 or liquid nitrogen- Liquid CO2 requires using a siphon dip tube cylinder so that low-speci-

fiedtemperaturescanbeachieved.

- Liquid propane requires inverting the propane cylinder.

Do not use regulators on CO2 or propane cylinder tanks.

Tubing and Fittings

Theinstrumentcomeswith1/8”metaltubingandfittingsformakingtheconnec-tion to the gas pipeline or tank from which you will be obtaining the sample. A nylonhoseshouldnotbeusedforthesamplelines(unlessitislinedwithTeflonor equal) because unlined hose may add or remove water from the gas causing an error in the dew point determination. Nylon hose can be used for the refrigerant connection.However,stainlesssteeltubing,1/8”indiameter,isrecommendedifthe tubing length is greater than ten feet.

Page 20: Chan Scope and Dew Point Tester Manual

3-4 | AMETEK ChanScope and Dew Point Tester

Installing the Dew Point Tester

Anadjustabletripodisavailableformountingthetesterwhenusedinthefield.

Observe the following precautions when setting up the instrument and doing test-inginthefield.

• Takearepresentativesample.Werecommendthatyouuseasampleprobeinserted into the center 1/3 of the pipeline. The probe should be inserted in a horizontal, straight run of pipe. It should also be located at least ten pipe di-ameters upstream and downstream from any turbulence causing devices such asanelbow,valve,orificemeter,etc.

• Ensure that the temperature of the sampling line and the dew point tester are above the dew point temperature of the gas in the pipeline. Heat-traced sample tubing may be required in cold ambient conditions.

• Ensure that the sampling line does not contain any condensed liquids such as water, hydrocarbons, glycol, etc.

• There must be no appreciable reduction in pressure between the gas source and the dew point tester.

Connecting the Refrigerant

Once the dew point tester sample connection to the gas supply is properly set up, you must connect the refrigerant valve to a cylinder of propane, liquid CO2, or otherliquefiedgasthatwillbeusedasarefrigerant.

Open the valve on the cylinder and invert it so that the liquid refrigerant is available at the chiller. If you are using a siphon dip tube, you do not have to invert the cylinder.

• Propane is a satisfactory refrigerant for dew point temperatures down to about -20°F (-28°C).

• Liquid carbon dioxide (CO2) may be used for temperatures down to -80°F (-62°C).

• For lower dew points you must use liquid nitro-gen (Figure 3-1).

Installing the Glycol Filter Assem-bly

Flask

ChainClamp

ClampHolder

Inversion Cooling Rod

Figure 3-1.Liquid nitrogen refrigerant connections.

Page 21: Chan Scope and Dew Point Tester Manual

Installation and Set Up | 3-5

Theglycolfilterassemblyisattachedin-linetotheinletofthedewpointtester(Figure3-2).Onceinstalled,thefilterassemblywillnotinterferewiththetester’sattachment to the tripod.

Selecting Temperature Display (ChanScope only)

Remove the rear cover of the ChanScope enclosure to change the temperature display units. Selections are degrees Celsius (°C) or degrees Fahrenheit (°F). There is a circuit assembly inside the enclosure. Set the miniature switch setting #1 to ON for Fahrenheit or to OFF for Celsius. Switches 2, 3 and 4 are not used.

Glycol FilterCartridge

O-RingSeal

Filter Housing O-Ring Seal

Installthis end todew pointtester

Cartridge

Figure 3-2.Glycol filter installation.

Page 22: Chan Scope and Dew Point Tester Manual

3-6 | AMETEK ChanScope and Dew Point Tester

Start Up

1. Connect the sample tubing from the sample source to the dew point tester sample inlet valve.

2. Turn the sample inlet valve until it is fully open.

3. Partiallyopenthesampleoutletvalvejustenoughtoallowflowthroughthetester body but not enough to cause a drop in the sample pressure.

4. Allowthesampletoflowthroughthetesteruntilsufficienttimehaspassedtoallowsystemequilibrium.Thisshouldbeapproximatelyfivetotenminutes.

5. Connect the refrigerant tubing from the refrigerant cylinder to the refrigerant inlet connection valve on the chiller device.

6. Slowlyopentherefrigerantvalvetostarttheflowofrefrigerantthroughthe chiller device. Once the temperature starts to drop, close the refrigerant valve.

7. Slowly open and close the refrigerant valve so that the temperature of the sample gas drops slowly.

8. Observe the mirror and thermometer/digital readout. At the same time, make noteofthetemperatureatwhichaspotofmoisturefirstappearsatthecenterof the mirror.

9. Allow the mirror to warm up and make note of the temperature at which the moisture spot disappears.

10. Repeat Steps 7 and 8 until the temperatures observed are within 2°F (1°C).

11. Record the average or mean temperature that you observed as the dew point temperature.

Page 23: Chan Scope and Dew Point Tester Manual

Operation | 4-1

OPERATION

To prevent serious injury or incorrect measurements, adhere to all test procedures when setting up the tester and performing tests in the field.

Using Glycol Filters

When using the standard dew point tester, there may be times when glycol vapors are present as you are measuring the dew point in the gas. These vapors can lead to difficulty in determining the water dew point because the glycol vapors also condense on the mirror, making observation of the water dew point difficult or uncertain. A glycol filter can be used with the dew point tester under these circumstances to remove the glycol vapor to make it easier to determine the true water dew point.

The filtering material in the glycol filter absorbs both water vapor and glycol va-por. However, though it quickly becomes saturated with water vapor, it continues to absorb glycol vapor for a somewhat longer time. It is during this latter period that water dew points can be determined satisfactorily since once the filter is water saturated, no more water vapor is removed from the gas and the dew points will be the same if read without the filter.

Determining Equilibrium

When the sample gas is introduced at the side of the pressure chamber, it flows through the glycol filter before entering the tester. The filter material should not cause a drop in pressure at normal flow rates. However, you must allow sufficient time for the filter material to attain equilibrium with the sample gas pressure. The time required to reach equilibrium depends on the concentration of water and glycol in the gas and the density of the filter material in the cartridge. Equilib-rium is determined by running preliminary dew point tests. Two consecutive test results without glycol masking of the mirror indicate that the tester is operating normally.

Page 24: Chan Scope and Dew Point Tester Manual

4-2 | AMETEK ChanScope and Dew Point Tester

Test Precautions

These precautions must be observed to prevent serious injury and to ensure ac-curate measurements.

1. Vent the sample gas to a safe area to prevent the possibility of creating an explosive atmosphere in an enclosed area.

2. Always wear safety goggles when working with refrigerants and high -pres-sure gases.

3. Do not exceed the pressure rating of the tester.

- Models 13-100/13-1200 tester maximum working pressure is 3000 PSIG (20.7 mPa).

- Models 13-075/13-1175 tester working pressure is limited to 1000 PSIG (6.9 mPa).

- Models 13-110/13-1210 tester working pressure is 5000 PSIG (34.5 mPa).

Working pressure must not be exceeded as this can result in compo-nent “blow out” causing serious injury to the operator.

4. Clean the tester properly before using oxygen.

5. Take a representative sample using a sample probe inserted into the center 1/3 of the pipeline.

6. Ensure that the sampling line does not contain condensed water, hydrocar-bons, glycol, etc.

7. Ensure that the temperature of the sampling line and the dew point tester are above the dew point temperature of the sample gas. When the atmospheric temperature is below the dew point temperature of the tester or the gas in the pipeline or tank, the tester and sample line should be heated with a flexible heat tape to avoid condensation between the gas source and the instrument.

8. Ensure that there is no appreciable reduction in pressure between the gas source and the dew point tester.

9. If you have installed a filter to remove glycol from the gas, the material in the filter absorbs both glycol and water. The filter usually becomes saturated with water within five minutes but continues to absorb glycol. Only after the filter is saturated with water can you make accurate dew point readings.

Page 25: Chan Scope and Dew Point Tester Manual

Operation | 4-3

Use extreme caution if you are using the tester on oxygen or other explosive gases reactive to oil. The entire instrument, including the sample line, must be thoroughly cleaned. Testers that have been ordered “cleaned for oxygen” from the factory will have an oxygen pressure gauge installed and will be tagged with the date they were cleaned.

Page 26: Chan Scope and Dew Point Tester Manual

4-4 | AMETEK ChanScope and Dew Point Tester

Test Procedure

Follow this procedure in the ordered sequence to ensure accurate readings.

1. Open the sample inlet valve and purge the tester by flowing the test gas through the sample chamber to expel all the air.

2. Regulate the flow of the test sample by partially closing the outlet valve and then adjust to ensure that the sample flows continuously at full sample pres-sure through the chamber.

3. Adjust the outlet valve so that the gas sample can just be detected on your fingertip. An aspirator bulb or vacuum pump can be used to bring atmospher-ic or low-pressure samples past the mirror.

A low flow of approximately 5 SCFH is recommended.

4. ChanScope For models with a digital dew point indicator, verify that the decimal point at

the far right side of the display is not blinking.

A blinking decimal point indicates that the tester is in pause or hold mode. When in this mode, the displayed temperature is “frozen” and will not change until you press the HOLD button. The HOLD button is a pushbut-ton or mementary switch; press HOLD at any time to ENTER or EXIT this mode.

5. Open the refrigerant valve periodically for a few seconds; allow sufficient time between valve openings to lower the mirror temperature. You can make a quick observation to determine an approximate dew point. However, for the final determination, you should lower the temperature at a rate of no more than one or two degrees per minute.

6. Observe the mirror and thermometer/digital readout at the same time and note the temperature at which a spot of moisture first appears at the center of the mirror.

7. Allow the mirror to warm up and note the temperature at which the moisture spot disappears.

8. Repeat Steps 5 and 6 until the temperatures observed are within 2°F (1°C).

9. Record the mean or average temperature observed as the dew point.

NOTE

Page 27: Chan Scope and Dew Point Tester Manual

Operation | 4-5

Confirmation of Dew Point

In extreme cases, water dew point determination can be difficult, even after the gas has passed through a glycol filter. An alternate method for determining if the dew point is water is to operate at a different pressure.

Example:

1. A dew point determination is done in the normal manner at 1000 PSIG and a dew point of 30°F is obtained. Plotting on the “Saturated Moisture Content of Natural Gas as a Function of Temperature and Pressure” chart (Drawing 13-0097) obtains a water vapor content of 6.3 lb/million cubic feet.

2. Pressure in the tester is reduced to 300 PSIG by partially closing the inlet valve and adjusting the outlet valve to obtain the approximate same flow rate.

3. A dew point determination is now run at this reduced pressure.

If a dew point of 9°F is obtained, you can be sure that the original determina-tion at 1000 PSIG was correct since the water content remained the same (6.3 lb/MMscf) with the change is pressure and dew point. The water dew point is lowered as the pressure is lowered, but the water content of the gas remains the same. Had the water content been different at the second pressure and dew point, then one of the dew points might have been a hydrocarbon, glycol or alcohol dew point.

Interpretation of Dew Point Temperature

Dew points can be converted to water content using Drawings 13-0095, 13-0097, and 13-0125 found in this manual. The ASTM tables are also included.

Page 28: Chan Scope and Dew Point Tester Manual

4-6 | AMETEK ChanScope and Dew Point Tester

Differentiating between Water and Hydrocarbon Dew Points

Sometimes, the hydrocarbon dew point of the gas will be the same or higher temperature than the water dew point. In this case, the two dew points must be differentiated.

• The hydrocarbon wets the mirror and the film expands in an iridescent ring to cover the mirror’s surface, or appears as a waterfall of droplets that appear to be pouring from the mirror’s center.

• The water condensate remains concentrated in the center of the mirror.

You can also verify the substance on the mirror by determining the dew point at a lower pressure.

NOTE

Page 29: Chan Scope and Dew Point Tester Manual

Operation | 4-7

Special Test Procedure for Propane

Dew point tests of commercial propane are usually made at atmospheric pres-sure instead of at full sample flow as with other gases. Therefore, when testing propane, the procedure for opening and closing the valves should be reversed to ensure atmospheric pressure inside the tester.

Commercial propane is usually considered “dry” if the dew point is -15°F or lower when measured at atmospheric pressure.

For detailed information refer to GPA Publication 2140-80 “Liq-uid Petroleum Gas Specification and Test Methods,” 1980 revision, Tulsa, Oklahoma.NOTE

Page 30: Chan Scope and Dew Point Tester Manual

4-8 | AMETEK ChanScope and Dew Point Tester

This page intentionally left blank.

Page 31: Chan Scope and Dew Point Tester Manual

Maintenance and Troubleshooting | 5-1

Maintenance and troubleshooting

Maintenance

tools required

standard dew Point tester

• 7/16”,1/2”,1/4”open-endwrenches• 10”adjustablewrench• 1/8”Allen-headdriverforlooseningandtighteningthewindownut• Slottedscrewdriver• Benchvise• Isopropylalcohol(99.5%)• Airduster• Teflontape• Cottonballsorswabs

chanscope dew Point tester

Sameasaboveplus:

• Phillipsscrewdriver• 1/8”hexwrench

Page 32: Chan Scope and Dew Point Tester Manual

5-2 | AMETEK ChanScope and Dew Point Tester

replacing the glycol Filter

Ifaglycolfilteritbeingused,itwilleventuallybecomesaturatedwithglycolandwillneedtobereplaced.

1. Disconnectthesampletubingfromtheglycolfilterhousing.

2. Openthefilterhousingandremovetheoldfiltercartridge.RemovetheO-ringfromtheoldfiltercartridgeandplaceitonthenewfiltercartridge.

3. Placethenewfiltercartridgeinthefilterhousingandreassemble.

4. Reconnectthesampleinlettubing.

cleaning the Window and Mirror

Boththewindowandmirrorofthetesterareeasilyremovedforcleaning.

• Useacottonball,cottonswab,orsoftclothtocleanthewindowandmirrorsurfaces.Becarefultoavoidscratchingthesurfaces.

• Ifahydrocarbonorothercontaminantremainsaftercleansing,useasoftclothwithsoapandwaterfollowedbyarinseusingisopropylalcohol.

• Tocleanthehigh-pressurewindow,removethescrewandpulltheunitoutofthewindownut.Becarefultoavoidscratchingthewindow.

A scratched window should be replaced. A fracture could develop from the scratch when the tester is subjected to high pressure.

recharging batteries (chanscope dew Point tester)

IftheLEDdisplayontheChanScopedewpointtestergoesblank,itisbecausethebatterypackneedstoberecharged.Thebatterypackisanickelmetalhy-dride(NiMH)1600mAhandwilllastapproximately12hoursiffullycharged.Itrequiresapproximately3hourstofullyrecharge.

• Plugthebatterychargerintothewallreceptaclefrom100-265VAC,50/60Hzusingtheappropriatesuppliedbladeadapter,andplugtheotherendintothejackontheChanScope.

The charger is not for outdoor use.

NOTE

NOTE

Page 33: Chan Scope and Dew Point Tester Manual

Maintenance and Troubleshooting | 5-3

charger Functionality

Chargingbeginswhenthebatterypackisconnectedtothecharger.TheLEDisorangebeforethefastchargestarts,andthentheLEDchangestored.Whenthecellsarefullychargedandthevoltagedrops(becauseofthe-dVsignalfromthecells),thechargerwillgointoatop-offchargemodebeforeswitchingtotrickle charge mode.

Duringthetop-offchargesession,theLEDwillbegreenwithshort,intermit-tentorangeflashes.Whenthetop-offchargeiscomplete,thechargerwillgointotrickle charge mode,changingtheLEDtogreen.Thechargecurrentisnowreducedtoasafelevel,permittingthechargertostayconnectedtoaNiCdbatterypackwithoutdamagingthecells.DependingonthespecificmanufacturerofNiMHcells,tricklecharginginexcessof24hoursmaynotbepermitted.

Ifthesafetytimerdisconnectsbefore-dV,thetop-offchargewillnotbeengaged.Thechargerwillthengodirectlytotrickle charge modeandtheLEDwillbegreen.Ifthebatteryvoltageisfarbelownormal,thechargerwillstopthefast-chargecurrentandswitchtotrickle charge mode.TheLEDwillthenindicate“error”byflickeringgreenandred.

Ifthemainsisturnedoff,thechargerwillresetandstartanewchargecyclewhenthemainsisturnedonagain.

You can operate the tester during charging. The tester cannot be operated without a battery.

• Turnthetesteroffwhennotinusetoconserveenergy.

Use a ChanScope II battery charger only (PN 13-2019). Other char-gers will cause internal damage.

RecyclethebatterypackatanappropriatefacilityforNiMHbatteries.Visitwww.rbrc.orgorcall1-800-8-BATTERYforlocations.

NOTE

NOTE

Page 34: Chan Scope and Dew Point Tester Manual

5-4 | AMETEK ChanScope and Dew Point Tester

bypassing the rtd

YoucanbypasstheRTDatanytimebyoperatingthetesterwiththethermometerinstead.

• LoosentheglandnutontheRTDandgentlypulltheRTDout.Replacewiththeappropriatethermometer.

• Turnthepowerswitchontoilluminatethemirrorandproceedasusual.

Remember that the LED displays now indicates ambient temperature when operating in this mode.

Replacing the batteries and changing the temperature display unit selection are the only operations allowed that require removing the back cover of the dew point indicator. All optical and electronic adjustments are made only at the factory. Failure to observe this may void the warranty.

NOTE

Page 35: Chan Scope and Dew Point Tester Manual

Maintenance and Troubleshooting | 5-5

calibration

Exceptforthethermometer,calibrationisnotpossibleornecessaryontheStan-dardDewPointTesters(Models13-075,13-100,and13-110.Thethermometercanberemovedandcheckedforaccuracyatanyreputablecalibrationserviceequippedtodoso.

CalibrationoftheChanScopeRTDandtheelectronicsisdoneatthefactory.ReturnthetesterafterobtaininganRMA.

selecting temperature display (chanscope only)

RemovetherearcoveroftheChanScopeenclosuretochangethetemperaturedisplayunits.SelectionsaredegreesCelsius(°C)ordegreesFahrenheit(°F).Thereisacircuitassemblyinsidetheenclosure.Settheminiatureswitchsetting#1toONforFahrenheitortoOFFforCelsius.Switches2,3and4arenotused.

Page 36: Chan Scope and Dew Point Tester Manual

5-6 | AMETEK ChanScope and Dew Point Tester

troubleshooting the chanscoPe

ProbleM cause solution

Intermittent display RTD, battery connections, or switch

Check RTD cable, battery connections, and switch

Display is segmented, or no display.

Low batteries, voltage less than minimum

Check charger output, replace battery pack; recharge.

With Chanscope OFF, char-ger turns OFF prematurely. Battery not fully charged.

Battery may have developed a “memory”.

Remove charger from outlet. Turn on Chanscope and discharge battery completely. Plug charger into outlet to charge.

Excessive glare or light on mirror.

Lamp too deep in holder. Remove cover, adjust lamp depth.

Temperature not reading correctly.

RTD not making proper contact.

Calibration.

Check for corrosion in well and depth of RTD in well to bottom. Use thrmometer to verify. Check RTD at 32°F (0°C) in insulated ice water bath, stirring the crushed ice.

Contact factory for calibration.

Temperature not changing, but decimal point is flashing in lower right portion of display.

Device is in “HOLD” mode.

Press HOLD button on the front panel to resume normal operation.

Page 37: Chan Scope and Dew Point Tester Manual

Maintenance and Troubleshooting | 5-7

troubleshooting deW Point readings

ProbleM cause solution

Haze on entire monitor obscures dew point.

Possible glycol contamination. Install glycol filter.

Condensed water in sample line.

Temperature of sample line and tester below the temperature of sample gas. Not using a probe to sample gas, slugs of water entering the sample line from the pipe wall.

Heat trace sample line. Purge line with dry nitrogen. Avoid excessive fittings, gauges, etc., in sample line that may be difficult to dry down after condensation. Use sample probe.

Low dew point using glycol filter.

Moisture equilibrium in glycol filter not yet reached.

Glycol filter requires time to saturate with water. Allow to saturate for 5 to 10 minutes.

CO2 and propane refrigerant not chilling to specified temperature.

Using gas instead of liquid. Using a regulator on the cylinder.

CO2 cylinder must be a siphon dip tube cylinder. Propane cylinders must be inverted. Do not use a regulator.

Unable to distinguish water dew point from hydrocarbon dew point.

Both dew points can occur at a lower, same, or higher temperature depending on hydrocarbon components present and the amount of water.

Hydrocarbon dew point wets the mirror with an iridescent ring followed by condensing droplets. This is your dew point. Viewing the hydrocarbon dew point at an angle to the mirror can be helpful. The water condensate begins at the mirror’s center and moves outward across the mirror.

Page 38: Chan Scope and Dew Point Tester Manual

5-8 | AMETEK ChanScope and Dew Point Tester

rePlaceMent PartsModels 13-075, 13-100, 13-110, 13-210, 13-1175, 13-1200 and 13-1210

Part # descriPtion13-0005 Window, Models 13-100 & 13-200 and 13-120013-0006 Window retaining nut, Model 13-100 & 13-11013-0007 Mirror assembly, regular chiller13-0009 Mirror insulator, Models 13-153 & 13-154 kits13-0018 Chiller body, regular13-0020 Thermoguard assembly13-0021 Lock nut, thermoguard13-0022 Bulb shield, thermoguard13-0024 Shield, thermoguard13-0025 Thermometer, range -30 to +110°F13-0027 Thermometer, range -90 to +40°F13-0038 Thermometer, range -100 to +10°C13-0046 Retaining nut, quick chiller13-0047 Block, with tube, quick chiller13-0048 Quick chiller assembly

13-0052 Glass window assembly, Models 13-100, 13-110, 13-200, 13-210, 13-1200, 13-1210 (sour gas and 5,000 PSIG)

13-0053 Cooling rod, nitrogen chiller13-0054 Glycol filter assembly13-0060 Thermometer, range -35 to +50°C

13-0064 Window, glass, Models 13-100, 13-110, 13-200, 13-210, 13-1200, 13-1210 (sour gas and 5,000 PSIG)

13-0065 Adapter, glass window, Models 13-100, 13-200, 13-210, 13-1200, 13-1210 (sour gas and 5,000 PSIG)

13-0068 Regular chiller assembly13-0071 Extension, tripod adapter, nitrogen chiller13-0072 Liquid nitrogen chiller assembly13-0077 Nut, window (Model 13-075)13-0078 Nut, mirror, retaining13-0079 Window, plastic, Models 13-075, 13-175, & 13-117513-0084 Adapter, gauge13-0085 Body, Models 13-075, 13-175, & 13-117513-0086 Adapter, tripod13-0101M Body, Model 13-100, 13-110, 13-200, 13-210, 13-1200, 13-121013-0102 Mirror, swivel13-0111 Body assembly, magnifier13-0116 Window, magnifier, eye guard13-0117 Lens retainer, Model 13-07513-0119 Lens, magnifier, objective13-0122 Eye guard, plastic, magnifier

Page 39: Chan Scope and Dew Point Tester Manual

Maintenance and Troubleshooting | 5-9

Part # descriPtion13-0124 Lens retainer, Model 13-100, 13-110 magnifier13-0146 Thermometer, certified range -90 to +40°F13-0147 Expansion tube only, quick chiller13-0148 Thermometer, certified range -200 to +0°F13-0149 Thermometer, certified range -30 to +110°F

13-0153

Replacement Seal Kit for 13-100 includes:(6) P-0063 O-ring(1) P-0065 O-ring(3) P-0067 O-ring(1) P-0071 O-ring(1) P-0221 O-ring(1) 13-0009 Insulator (2 parts)

13-0154 Same as 13-153, plus P-70 plug O-ring used for Model 13-07513-0157 Thermometer, certified range -5 to +50°C13-0158 Thermometer, certified range -130 to -20°C13-0161 Extension, tripod adapter13-0210 L-bracket, battery holder13-0277 CO2 bottle, 20 cu ft, with siphon tube, nut and tailpiece (full)13-0303 Illuminated magnifier assembly, Model 13-07513-0304 Illuminated magnifier assembly, Model 13-100 & 13-11013-2277 CO2 bottle, 20 cu.ft., with siphen tube, nut and tailpiece (empty)36-0028 Hose, refrigeration, 1/4” flexible, 1250 PSIG x 10 feet36-0045 Tube, assembly, 1/8” x 10 feet SST, brass fittings, welded36-0050 Refrigerant tube assembly, nylon (sour gas service)36-0087 Tube, assembly, 1/8” x 10 feet, SST, with brass fittings36-0000-10 Tubing, coiled, SST, 1/8” x 10 feet187-20403 Connector, 1/4 MP x 1/8 T, SST188-13044 Connector, 1/8 MP x 1/8 T, SSTC03322 Connector, 1/8 FP x 1/8 T, SSTC05013 LubricantC08161 Valve, inlet and outletC08435 Cartridge, mega-glysorb filter (package of 10)C08517 Connector, 1/4 FP x 1/8 T, SSTC08884 O-ring for gauge adapter (sour gas service)C08885 O-ring for window (sour gas service)C08886 O-ring, mirror (sour gas service)C09490 Battery holder, magnifierC09509 Valve handleC09549 Batteries, AAAC09550 Light bulb, magnifierC09608 Pressure gauge, oxygen service, 1000 PSIGC10499* Gauge, digital pressure, 1000 PSIC10500* Gauge, digital pressure, 3000 PSIC10501* Gauge, digital pressure, 5000 PSI* Consult factory for optional measurement units.

Page 40: Chan Scope and Dew Point Tester Manual

5-10 | AMETEK ChanScope and Dew Point Tester

P-0018 Clamp, magnifierP-0020 Clamp holder, nitrogen chillerP-0063 O-ring, gauge adapter, quick chillerP-0067 O-ring, window mirror (included in 13-153 & 13-154 kits)P-0070 O-ring, 13-0303 magnifier (included in 13-154 kit)

P-0071 O-ring, window for Model 13-075, 13-0304 magnifier (included in 13-153 and 13-154 kits)

P-0182 Hose nipple, chillersP-0301 Valve, chillerP-0535 TripodP-0536 20# propane cylinder with valve & 1/4” MNPT connectorP-0583 Pressure gauge, brass, 0-15 PSIG, Model 13-075P-0591 Pressure gauge, brass, 0-1000 PSIG, Model 13-075P-0594 Pressure gauge, brass, 0-3000 PSIG, Model 13-100, 13-110P-0711 Quick disconnect coupling, chillerP-0720 Pressure gauge, brass, 0-200 PSIG, Model 13-075P-0986 Flask, N2, chillerP-0987 Chain clampo, nitrogen chillerP-0988 Plastic handle for carrying caseP-1037 Pressure gauge, brass, oxygen service, 3000 PSIGP-1360 Nut, brass (for CO2 bottle)P-1367 Tailpiece, brass, 1/4 MP (for CO2 bottle)P-1392 Pressure gauge, 0-3000 PSIG, Model 13-100 sour gas serviceP-1818 Pressure gauge, brass, 0-5000 PSIG, Model 13-110 & 13-210 P-2300 Valve, SST, 6000 PSI (sour gas service)P-2346 Pressure gauge, 0-5000 PSIG, Model 13-0100 & 13-210, SST

Page 41: Chan Scope and Dew Point Tester Manual

Maintenance and Troubleshooting | 5-11

chanscope digital dew Point testerModels 13-175, 13-200, 13-210 (10,000 series sn)

Part # descriPtion13-0204 Window nut, Model 13-17513-0205 Window nut, Model 13-0200, 13-021013-0210 L-bracket, battery13-0214 Optical assembly13-0215 Beam splitter assembly13-0217 Prism assembly13-0219* Logic board assembly13-0220 LED board assembly13-0227 Eyepiece assembly13-0230 Gland nut13-0234 Switch13-0236* RTD assembly with strain relief (spring-loaded)13-0241 Battery holder, two cell13-0250 Dewpoint indicator (ChanScope) assembly13-0253 Wiring harness assembly13-0254 Cable assembly13-1238 Black case with foam inserts187-20535 Adapter, 220 VAC to 115 VAC, 50/60 Hz, use with P-3116 chargerC08846 Kit, 230 VAC plug adapterP-0221 O-ring, RTD gland nutP-1044 Lamp, screw baseP-2770 Lamp, bayonet baseP-2820 O-ring, tube supportP-2822*** Eye CupP-2826 Screw, 2-56 x .250, rnd Head, slotP-2828 Screw, 6-32 x .250, rnd head, PhillipsP-2830 Screw, 6-32 x 1.0, skt head, SSTP-3113 Battery, C cell, NiCad rechargeP-3115 Holder, (1) C cell, #2173P-3116 Battery charger, NiCad

* RTD and circuit assembly should be purchased together and be factory-calibrated. Consult the factory for price and availability.*** OBSOLETE. Replaced by C11468 Eye Cup and 13-0227 Eyepiece Assembly. Consult the factory for price and availability.

Page 42: Chan Scope and Dew Point Tester Manual

5-12 | AMETEK ChanScope and Dew Point Tester

chanscope ii digital dew Point testerModels 13-1175, 13-1200, 13-1210 (20,000 series sn)

Part # descriPtion13-0204 Window nut, Model 13-117513-0205 Window nut, Model 13-1200, 13-121013-0215 Beam splitter assembly13-0227 Eyepiece assembly13-0236* RTD assembly with strain relief (spring-loaded)13-0307* Circuit assembly13-1214 Optical assembly13-1217 Prism assembly13-1238 Case with foam inserts13-1250 ChanScope II assembly13-1252 ChanScope LED/lamp wiring harness assembly13-1274 Battery pack, rechargeable13-1276 Switch/battery/harness assembly13-1277 Assembly, battery charger connector wiring harness13-2019** Battery charger, NiMH, 1600 mAh, 100-265 VAC, 50/60 Hz, with 3 bladesC09867 Pushbutton switchC11003 Charger connection cap assemblyC11468 Eye cupC11921 Adapter, charger, US bladeC11922 Adapter, charger, Euro bladeC11923 Adapter, charger, UK bladeC12024 Charger, Z22 PDA, 110-240VP-2820 O-ring, tube supportP-2821 Rocker switchP-2826 Screw, 2-56 x .250 rnd head, slottedP-2828 Screw, 6-32 x .250, rnd head, PhillipsH-6033 Screw, 6-32 x 1.125”, skt head, SST (4 required for eyepiece)P-2836 Strain relief, RTD, cable

* RTD and circuit assembly should be purchased together and be factory-calibrated. Consult the factory for price and availability.

** Replaces 13-1275 charger, 187-20535, 220 VAC to 115 VAC adapter, and C08846 plug adapter kit.

Page 43: Chan Scope and Dew Point Tester Manual

Drawings | 6-1

DRAWINGS

Figure 6-1.Glass window assembly for sour gas/5000 PSIG DPTDWG 13-0052.

1. (13-0052) Glass Window Assembly for sour gas/5000 PSIG DPT includes:

- (13-0064) Glass Window- (13-0065) Adapter

2. UseperfluoroelastomerO-ringsforcorrosivegases.

3. LubricateO-ring(C08885)withsiliconO-ringlubricant(C05013).

Page 44: Chan Scope and Dew Point Tester Manual

6-2 | AMETEK ChanScope and Dew Point Tester

Figure 6-2.Glycol filter assembly.DWG 13-0054.

1

2

5

6

4

3

7

8

Installthis end todew pointtester

9 10

11

ITEM PART # DESCRIPTION

1 36-0000-10 1/8” SST Tubing2 C03322 Tube Connector

3 Valve - supplied with dew point tester

4 P-2577 Bushing5 O-Ring Seal6 C08435 Filter Cartridge7 P-3178 Elbow8 C04364 Adapter9 C08434 Filter Housing10 O-Ring Seal11 C08435 Cartridge

Page 45: Chan Scope and Dew Point Tester Manual

Drawings | 6-3

Figure 6-3.Immersion cooling rod assembly.DWG 13-0072.

Approximate Dew Points obtainable with refrigerants:DryIceinsolvent: -60°FLiquidnitrogen:-200°F

Dew point tester and tripod are not included in 13-0072 immersion cooling rod assembly.

NOTE

ITEM PART # DESCRIPTION

1 P-0020 Clamp Holder2 13-0071 Extension

3 Model 13-0100 or Model 13-0075 with 13-0007 Regular Mirror

4 13-0053 Immersion Cooling Rod5 P-0986 Flask6 P-0987 Chain Clamp7 P-0535 Tripod

1

2

3 4

5

6

7

Page 46: Chan Scope and Dew Point Tester Manual

6-4 | AMETEK ChanScope and Dew Point Tester

Figure 6-4.Dew Point Tester, Model 13-0075 with quick chiller assembly.DWG 13-0075.

ITEM PART # DESCRIPTION1 C03322 Connector

2 C08161 Valve

3 P-1190 Elbow

4 Refrigerant Outlet

5 P-0301 Chiller valve

6 13-0086 Adapter, Tripod

7 C08161 Valve

8 13-0079 Window

9 13-0077 Nut

10 13-0085 Body

11 P-0071 Extension

12 13-0084 Adapter

13 P-0591 Pressure Gauge

14 P-0063 O-Ring, Gauge Adapter

15 P-0067 O-Ring, Window Mirror

16 13-0078 Nut

17 13-0048 Quick Chiller Assembly

18 13-0046 Retaining Nut

19 13-0021 Lock Nut, Thermoguard

20 13-0020 Thermoguard Assembly

21 13-0024 Shield

22 13-0023 Adjusting Knob

23 Refrigerant Inlet 1/8 FNPT

1

2

3

4

5

9

8

7

6

23

18

17

1920

2122

1615

14

13

12

11

10

Thermometerspecify temperature range

1/8 NPT

1/8 NPT

14 3/4 approx

Gas Outlet

10 1/2approx

Page 47: Chan Scope and Dew Point Tester Manual

Drawings | 6-5

Figure 6-5.Chillers, regular and quick.DWG 13-0090.

Page 48: Chan Scope and Dew Point Tester Manual

6-6 | AMETEK ChanScope and Dew Point Tester

Figure 6-6.Chart.

Page 49: Chan Scope and Dew Point Tester Manual

Drawings | 6-7

Figure 6-7.Chart.

Page 50: Chan Scope and Dew Point Tester Manual

6-8 | AMETEK ChanScope and Dew Point Tester

Figure 6-8.Chart.

Page 51: Chan Scope and Dew Point Tester Manual

Drawings | 6-9

Figure 6-9.Chart.

Page 52: Chan Scope and Dew Point Tester Manual

6-10 | AMETEK ChanScope and Dew Point Tester

Figure 6-10.Equilibrium water vapor contents in natural gas above the critical temperature, Page 1.

Page 53: Chan Scope and Dew Point Tester Manual

Drawings | 6-11

Figure 6-11.Equilibrium water vapor contents in natural gas above the critical temperature, Page 2.

Page 54: Chan Scope and Dew Point Tester Manual

6-12 | AMETEK ChanScope and Dew Point Tester

Figure 6-12.Equilibrium water vapor contents in natural gas above the critical temperature, Page 3.

Page 55: Chan Scope and Dew Point Tester Manual

Drawings | 6-13

Figure 6-13.Equilibrium water vapor contents in natural gas above the critical temperature, Page 4.

Page 56: Chan Scope and Dew Point Tester Manual

6-14 | AMETEK ChanScope and Dew Point Tester

Figure 6-14.Equilibrium water vapor contents in natural gas above the critical temperature, Page 5.

Page 57: Chan Scope and Dew Point Tester Manual

Drawings | 6-15

Figure 6-15.Bureau of Mines dew point tester with regular chiller.DWG 13-0100.

Page 58: Chan Scope and Dew Point Tester Manual

6-16 | AMETEK ChanScope and Dew Point Tester

Figure 6-16.Dew point to moisture content conversion chart.

Page 59: Chan Scope and Dew Point Tester Manual

Drawings | 6-17

Figure 6-17.Installation of dew point tester.DWG 13-0150.

Page 60: Chan Scope and Dew Point Tester Manual

6-18 | AMETEK ChanScope and Dew Point Tester

Figure 6-17.Dew point indicator assembly.DWG 13-2016.

Page 61: Chan Scope and Dew Point Tester Manual

Drawings | 6-19

Figure 6-18.Illuminated magnifier assembly, Model 13-075.DWG 13-0303.

INST

ALLA

TION

INST

RUCT

IONS

1.Rem

ove13-0102observationmirrorfrom

thedewpointtesterbody.

2.LubricatethemagnifierO

-ringandpushthemagnifierinsideoftheround

windownut.

3.Screw13-0305illuminatorassem

blyinto13-0117lensretainer.

4.Rotate13-0305illum

inatorcounterclockw

isetoturniton.

ITEM

PAR

T #

DES

CR

IPTI

ON

1H

-801

4#8

-32

x .5

LG

, BH

P2

13-0

116

Win

dow

Eye

guar

d3

13-0

122

View

er4

13-0

111

Bod

y A

ssem

bly

51/

4-20

x .2

5 LG

, Alle

n H

D s

et s

crew

613

-011

9Le

ns7

13-0

117

Lens

Ret

aine

r8

P-0

070

O-R

ing

913

-030

5Ill

umin

ator

Ass

embl

y

Page 62: Chan Scope and Dew Point Tester Manual

6-20 | AMETEK ChanScope and Dew Point Tester

Figure 6-19.Illuminated magnifier assembly, Models 13-100 and 13-110.DWG 13-0304.

UsedwithModels13-100& 13-110 Dew Point Tester, with13-0006roundwindownut.

INSTALLATION INSTRUCTIONS

1. Remove13-0102observationmirrorfromthedewpointtesterbody.2. LubricatethemagnifierO-ringandpushthemagnifierovertheoutsideofthe

roundwindownut.3. Screw13-0305illuminatorassemblyinto13-0124lensretainer.4. Rotate13-0305illuminatorcounterclockwisetoturniton.

ITEM PART # DESCRIPTION

1 13-0305 Illuminator2 P-0071 O-Ring3 13-0124 Lens Retainer4 H-8014 #8-32 x .5 LG, BHP5 13-0122 Viewer

6 1/4-20 x .25 LG, Allen HD set screw

7 13-0116 Window Eyeguard8 13-0111 Body Assembly9 13-0119 Lens

Page 63: Chan Scope and Dew Point Tester Manual

Drawings | 6-21

Figure 6-20.Illuminator Assembly.DWG 13-0305.

ITEM PART # DESCRIPTION

1 C09490 Flashlight, Maglight Model M3A016

2 13-0301 Adapter, flashlight x 3/8-32

3 13-0302 Light pipe, 0.25

4 C09549 Batteries, AAA

5 C09550 Light bulb (replacement only), LM-3A-001

Page 64: Chan Scope and Dew Point Tester Manual

6-22 | AMETEK ChanScope and Dew Point Tester

Figure 6-21.ChanScope II.DWG 13-2017.

Not

Sho

wn:

13-2

019

B

atte

ry C

harg

erC

1192

2

Cha

rger

Ada

pter

(Eur

o bl

ades

)C

1192

3

Cha

rger

Ada

pter

(UK

bla

des)