CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... ·...
Transcript of CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... ·...
![Page 1: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/1.jpg)
USP <232>/<233>
Method Note Compendium
Updated April 6, 2018
![Page 2: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/2.jpg)
ContentsMARSTM 6
Consumer Products
Suppository (Capsule) 2
Vaseline 3
Wax (Candle) 4
Nutraceuticals
Biotin 6
Ginko (Ground) 7
Ginko (Whole Pill) 8
Kelp 9
Whey Powder 10
Pharmaceutical & Biotech
(4-Fluorophenyl)diphenylsulfonium Triflate 12
Allergy Pill (Ground) 13
Allergy Pill (Whole Pill) 14
Antioxidant (Ground) 15
Antioxidant (Whole Pill) 16
APIs with Aromatic Ring Structures 17
Ascorbic Acid 18
Aspirin (Ground) 19
Aspirin (Whole) 20
Beta Carotene 21
Biotin 22
Calcium Carbonate 23
Calpan Pantothenic Acid 24
Cephalexin 25
Chromium Chelate 26
![Page 3: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/3.jpg)
Cold and Flu Medicine (Liquid) 27
Cupric Sulfate 28
Diclofenac K 29
Empty Capsule (gel) 30
Fish Oil (No Capsule) 31
Folic Acid 32
Ginko (Ground) 33
Ginko (Whole Pill) 34
Kelp 35
Losataran K 36
Magnesium Carbonate 37
Magnesium Oxide 38
Manganese Carbonate 39
Metaformin (ground) 40
Multivitamin (Ground) 41
Multivitamin (Whole) 42
Niacinamide 43
Pantothenic Acid 44
Potassium Chelate 45
Pyridoxine 46
Riboflavin 47
Selenium Chelate 48
Stearic Acid 49
Thiamine 50
Trixie Phosphate in iPrep 51
USP 232/233 (Pharmaceuticals) 52
Vitamin B-12 53
Vitamin C (ground) 54
Vitamin C (whole) 55
Vitamin D 56
Vitamin D3 57
Vitamin E 58
Vitamins 59
Zinc Sulfate 60
![Page 4: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/4.jpg)
Plastics, Polymers, & Oils
Mineral Oil 62
Discover® SP-D 80
Consumer Products
Suppository 64
Vaseline 65
Wax (candle) 66
Nutraceuticals
Ginko (Ground) 68
Ginko (Whole Pill) 69
Kelp 70
Pharmaceutical & Biotech
Allergy Pill (ground) 72
Allergy Pill (whole pill) 73
Antioxidant (Ground) 74
Antioxidant (whole pill) 75
Aspirin (Ground) 76
Aspirin (Whole) 77
Calpan Pantothenic Acid 78
Cephalexin 79
Chromium Chelate 80
Cold and Flu Medicine (liquid) 81
Diclofenac K 82
Empty Capsule 83
Fish Oil (No Capsule) 84
Folic Acid 85
Ginko (Ground) 86
Ginko (Whole Pill) 87
![Page 5: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/5.jpg)
Kelp 88
Losataran K 89
Magnesium Oxide 90
Manganese Carbonate 91
Metaformin 92
Multivitamin (ground) 93
Multivitamin (whole) 94
Stearic Acid 95
Thiamine 96
USP 232/233 97
Vitamin C (ground) 98
Vitamin C (whole) 99
Vitamin D3 100
Zinc Sulfate 101
![Page 6: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/6.jpg)
Consumer Products
MARS™ 6
![Page 7: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/7.jpg)
2
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Suppository (Capsule)
Weigh 1 Capsule (approx. 1.0 g) into the digestion vessel. Add 12 mL of HNO3. Gently swirl the mixture and wait approximately 15 minutesbefore closing the vessel.
Ensure that the capsule is completely covered with reagent before sealing the vessel.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
1 Capsule (approx 1 g) Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofSuppository (Capsule)
![Page 8: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/8.jpg)
3
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Vaseline
Weigh 0.5 g of the sample into the digestion vessel. Add 10 mL of HNO3. Gently swirl the mixture and wait approximately 15 minutes beforeclosing the vessel.
The use of HF may be required to digest any silicates found in the sample.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofVaseline
![Page 9: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/9.jpg)
4
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Wax (Candle)
Weigh 0.5 g of the sample into the digestion vessel. Add 10 mL of HNO3. Gently swirl the mixture and wait approximately 15 minutes beforeclosing the vessel.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofWax (Candle)
![Page 10: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/10.jpg)
Nutraceuticals
MARS™ 6
![Page 11: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/11.jpg)
6
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Biotin
Weigh 0.5 g of the sample into the digestion vessel. Add 9 ml of HNO3, and 1 ml of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4mL) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
EasyPrepEasyPrep PlusXpressXpress Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
210 20:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 ml.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofBiotin
![Page 12: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/12.jpg)
7
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Ginko (Ground)
Weigh 0.5 g of the sample into the digestion vessel. Add 10 mL of HNO3. Gently swirl the mixture and wait approximately 15 minutes beforeclosing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl willvary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofGinko (Ground)
![Page 13: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/13.jpg)
8
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Ginko (Whole Pill)
Weigh 1 pill (approx 1.0 g) into the digestion vessel. Add 10 mL of HNO3 and 2 mL of H2O2. Gently swirl the mixture and wait approximately15 minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3H2O2
1 pill (Approx 1.0 g) Organic Ramp to Temperature One Touch
210 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofGinko (Whole Pill)
![Page 14: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/14.jpg)
9
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Kelp
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL of HNO3 and 1 mL of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofKelp
![Page 15: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/15.jpg)
10
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Whey Powder
Weigh 0.5 g of the sample into the digestion vessel. Add 10 ml of HNO3. Gently swirl the mixture and wait approximately 15 minutes beforeclosing the vessel.
The addition of Conc. HCl (0-4mL) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
0.5 g Organic Ramp to Temperature One Touch
210 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 ml.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofWhey Powder
![Page 16: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/16.jpg)
Pharmaceutical & Biotech
MARS™ 6
![Page 17: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/17.jpg)
12
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
(4-Fluorophenyl)diphenylsulfonium Triflate
Add 0.25 g of the sample into the digestion vessel. Add 9 mL of HNO3, and 1 mL of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
This application can only be run in the iPrep vessel.
This chemical compound is also know as Sudan Orange.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl willvary depending on the matrix and the concentration of the analytes.
MARS 6 iWave iPrep HNO3HCl
0.25 g Organic Ramp to Temperature One Touch
250 30:00 25:00 N/A 1800 Off
Sample was clear, gold colored, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.d) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion of(4-Fluorophenyl)diphenylsulfonium Triflate
![Page 18: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/18.jpg)
13
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Allergy Pill (Ground)
Weigh 0.25 g of the sample into the digestion vessel. Add 9 mL of HNO3 and 1 mL of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl willvary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
HCl
0.25 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofAllergy Pill (Ground)
![Page 19: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/19.jpg)
14
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Allergy Pill (Whole Pill)
Weigh 1 pill (Approx 1.0 g) into the digestion vessel. Add 10 mL of HNO3 and 2 mL of H2O2 (30%). Gently swirl the mixture and waitapproximately 15 minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl willvary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
H2 O2 (30%)
1 pill (Approx 1.0 g) Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofAllergy Pill (Whole Pill)
![Page 20: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/20.jpg)
15
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Antioxidant (Ground)
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL of HNO3 and 1 mL of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
HCl
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofAntioxidant (Ground)
![Page 21: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/21.jpg)
16
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Antioxidant (Whole Pill)
Weigh 1 pill (approx 1.4 g) into the digestion vessel. Add 10 mL of HNO3 and 2 mL of H2O2 (30%). Gently swirl the mixture and waitapproximately 15 minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl willvary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
H2O2 (30%)
1 pill (Approx 1.4 g) Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofAntioxidant (Whole Pill)
![Page 22: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/22.jpg)
17
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
APIs with Aromatic Ring Structures
Weigh 0.1 g - 0.5 g of the sample into the digestion vessel. Add 9 ml of HNO3 and 1 ml of HCl. Gently swirl the mixture and waitapproximately 15 minutes before closing the vessel.
This application can only be run in the iPrep vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6 iWave iPrep HNO3HCl
0.1 g - 0.5 g Organic Ramp to Temperature One Touch
250 30:00 25:00 N/A 1800 N/A
Sample was clear, colorless, and particle free upon dilution to 50 ml. If the sample contains precious metals the diluted sample may be the color associatedwith the metals.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.d) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofAPIs with Aromatic Ring Structures
![Page 23: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/23.jpg)
18
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Ascorbic Acid
Weigh 0.5 g of the sample into the digestion vessel. Add 9 ml of HNO3, and 1 ml of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4mL) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
EasyPrepEasyPrep PlusXpress PlusXpress
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
210 20:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 ml.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofAscorbic Acid
![Page 24: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/24.jpg)
19
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Aspirin (Ground)
Weigh 0.25 g of the sample into the digestion vessel. Add 9 mL of HNO3 and 1 mL of HCl. Gently swirl the mixture and wait approximately15 minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl willvary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
HCl
0.25 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofAspirin (Ground)
![Page 25: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/25.jpg)
20
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Aspirin (Whole)
Weigh 1 pill (Approx 1.0 g) into the digestion vessel. Add 10 mL of HNO3. Gently swirl the mixture and wait approximately 15 minutes beforeclosing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl willvary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
1 pill (Approx 1.0 g) Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofAspirin (Whole)
![Page 26: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/26.jpg)
21
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Beta Carotene
Weigh 0.5 g of the sample into the digestion vessel. Add 9 ml of HNO3, and 1 ml of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4mL) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6 iWaveMARS 6
EasyPrepEasyPrep PlusXpressXpress Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
210 20:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 ml.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofBeta Carotene
![Page 27: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/27.jpg)
22
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Biotin
Weigh 0.5 g of the sample into the digestion vessel. Add 9 ml of HNO3, and 1 ml of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4mL) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
EasyPrepEasyPrep PlusXpressXpress Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
210 20:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 ml.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofBiotin
![Page 28: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/28.jpg)
23
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Calcium Carbonate
Weigh 0.5 g of the sample into the digestion vessel. Slowly add 9 ml of HNO3, and 1 ml of HCl. Gently swirl the mixture and waitapproximately 15 minutes before closing the vessel.
The addition of Conc. HCl (0-4mL) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
Allow initial reaction to subside before sealing vessel.
MARS 6MARS 6 iWave
EasyPrepEasyPrep PlusXpressXpress Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
210 20:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 ml.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofCalcium Carbonate
![Page 29: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/29.jpg)
24
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Calpan Pantothenic Acid
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL of HNO3 and 1 mL of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
HCl
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofCalpan Pantothenic Acid
![Page 30: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/30.jpg)
25
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Cephalexin
Weigh 0.5 g of the sample into the digestion vessel. Add 10 mL of HNO3. Gently swirl the mixture and wait approximately 15 minutes beforeclosing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl willvary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofCephalexin
![Page 31: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/31.jpg)
26
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Chromium Chelate
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL of HNO3 and 1 mL of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
HCl
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofChromium Chelate
![Page 32: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/32.jpg)
27
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Cold and Flu Medicine (Liquid)
Add 0.75 mL of the sample into the digestion vessel. Add 9 mL of HNO3 and 1 mL of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl willvary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
HCl
0.75 mL Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofCold and Flu Medicine (Liquid)
![Page 33: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/33.jpg)
28
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Cupric Sulfate
Weigh 0.5 g of the sample into the digestion vessel. Add 9 ml of HNO3, and 1 ml of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4mL) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
EasyPrepEasyPrep PlusXpressXpress Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
210 20:00 15:00 800 900-1050 Off
Sample was clear, blue in color, and particle free upon dilution to 50 ml.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofCupric Sulfate
![Page 34: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/34.jpg)
29
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Diclofenac K
Weigh 0.5 g of the sample into the digestion vessel. Add 10 mL of HNO3 and 2 mL of H2O2 (30%). Gently swirl the mixture and waitapproximately 15 minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl willvary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
H2O2 (30%)
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofDiclofenac K
![Page 35: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/35.jpg)
30
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Empty Capsule (gel)
Weigh 1 capsule (Approx 0.1 g) into the digestion vessel. Add 10 mL of HNO3. Gently swirl the mixture and wait approximately 15 minutesbefore closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl willvary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
1 capsule (Approx 0.1 g) Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofEmpty Capsule (gel)
![Page 36: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/36.jpg)
31
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Fish Oil (No Capsule)
Weigh 0.25 g of the sample into the digestion vessel. Add 10 mL of HNO3. Gently swirl the mixture and wait approximately 15 minutes beforeclosing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl willvary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
0.25 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofFish Oil (No Capsule)
![Page 37: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/37.jpg)
32
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Folic Acid
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL of HNO3 and 1 mL of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
HCl
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofFolic Acid
![Page 38: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/38.jpg)
33
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Ginko (Ground)
Weigh 0.5 g of the sample into the digestion vessel. Add 10 mL of HNO3. Gently swirl the mixture and wait approximately 15 minutes beforeclosing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl willvary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofGinko (Ground)
![Page 39: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/39.jpg)
34
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Ginko (Whole Pill)
Weigh 1 pill (approx 1.0 g) into the digestion vessel. Add 10 mL of HNO3 and 2 mL of H2O2. Gently swirl the mixture and wait approximately15 minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3H2O2
1 pill (Approx 1.0 g) Organic Ramp to Temperature One Touch
210 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofGinko (Whole Pill)
![Page 40: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/40.jpg)
35
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Kelp
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL of HNO3 and 1 mL of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofKelp
![Page 41: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/41.jpg)
36
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Losataran K
Weigh 0.5 g of the sample into the digestion vessel. Add 10 mL of HNO3 and 2 mL of H2O2 (30%). Gently swirl the mixture and waitapproximately 15 minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3H2O2 (30%)
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofLosataran K
![Page 42: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/42.jpg)
37
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Magnesium Carbonate
Weigh 0.5 g of the sample into the digestion vessel. Add 9 ml of HNO3, and 1 ml of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
Allow any reaction to subside before sealing the vessel.
The addition of Conc. HCl (0-4mL) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
EasyPrepEasyPrep PlusXpressXpress Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
210 20:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 ml.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofMagnesium Carbonate
![Page 43: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/43.jpg)
38
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Magnesium Oxide
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL of HNO3 and 1 mL of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofMagnesium Oxide
![Page 44: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/44.jpg)
39
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Manganese Carbonate
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL of HNO3 and 1 mL of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofManganese Carbonate
![Page 45: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/45.jpg)
40
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Metaformin (ground)
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL of HNO3 and 1 mL of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofMetaformin (ground)
![Page 46: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/46.jpg)
41
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Multivitamin (Ground)
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL of HNO3 and 1 mL HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
HCl
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofMultivitamin (Ground)
![Page 47: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/47.jpg)
42
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Multivitamin (Whole)
Weigh 1 pill (Approx 2.5 g) into the digestion vessel. Add 10 mL of HNO3 and 2 mL of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
EasyPrepEasyPrep Plus
HNO3
HCl
1 pill (Approx 2.5 g) Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofMultivitamin (Whole)
![Page 48: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/48.jpg)
43
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Niacinamide
Weigh 0.5 g of the sample into the digestion vessel. Add 9 ml of HNO3, and 1 ml of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4mL) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
EasyPrepEasyPrep PlusXpressXpress Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
210 20:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 ml.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofNiacinamide
![Page 49: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/49.jpg)
44
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Pantothenic Acid
Weigh 0.5 g of the sample into the digestion vessel. Add 9 ml of HNO3, and 1 ml of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4mL) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
EasyPrepEasyPrep PlusXpressXpress Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
210 20:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 ml.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofPantothenic Acid
![Page 50: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/50.jpg)
45
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Potassium Chelate
Weigh 0.5 g of the sample into the digestion vessel. Add 9 ml of HNO3, and 1 ml of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4mL) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
EasyPrepEasyPrep PlusXpressXpress Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
210 20:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 ml.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofPotassium Chelate
![Page 51: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/51.jpg)
46
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Pyridoxine
Weigh 0.5 g of the sample into the digestion vessel. Add 9 ml of HNO3, and 1 ml of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4mL) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
EasyPrepEasyPrep PlusXpressXpress Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
210 20:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 ml.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofPyridoxine
![Page 52: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/52.jpg)
47
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Riboflavin
Weigh 0.25 g of the sample into the digestion vessel. Add 9 ml of HNO3, and 1 ml of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4mL) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
EasyPrepEasyPrep Plus
HNO3HCl
0.25 g Organic Ramp to Temperature One Touch
230 20:00 30:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 ml.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofRiboflavin
![Page 53: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/53.jpg)
48
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Selenium Chelate
Weigh 0.5 g of the sample into the digestion vessel. Add 9 ml of HNO3, and 1 ml of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4mL) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
EasyPrepEasyPrep PlusXpressXpress Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
210 20:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 ml.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofSelenium Chelate
![Page 54: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/54.jpg)
49
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Stearic Acid
Weigh 0.25 g of the sample into the digestion vessel. Add 10 mL of HNO3. Gently swirl the mixture and wait approximately 15 minutes beforeclosing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
EasyPrepEasyPrep Plus
HNO3
0.25 g Organic Ramp to Temperature One Touch
210 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofStearic Acid
![Page 55: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/55.jpg)
50
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Thiamine
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL of HNO3 and 1 mL of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
HCl
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofThiamine
![Page 56: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/56.jpg)
51
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Trixie Phosphate in iPrep
Weigh 0.1 g of the sample into the digestion vessel. Add 4 mL of HNO3 and 1 mL of HCl. Gently swirl the mixture and wait approximately 15minutes to pre-digest before closing the vessel.
This application can only be run in the iPrep vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl willvary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
MARS 6 iWave iPrep HNO3HCl
0.1 - 0.5 g Organic Ramp to Temperature Classic
260 30:00 25:00 N/A 1800 Off
Sample was clear, colorless and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the required results on yoursample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or below roomtemperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.d) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofTrixie Phosphate in iPrep
![Page 57: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/57.jpg)
52
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
USP 232/233 (Pharmaceuticals)
Weigh 0.2 g of the sample into the digestion vessel. Add 9 ml of HNO3, and 1 ml HCl. Gently swirl the mixture before closing the vessel.
The addition of Conc. HCl (0-4mL) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpress PlusEasyPrepEasyPrep Plus
HNO3HCl
0.2 g Organic Ramp to Temperature One Touch
210 20:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 ml.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofUSP 232/233 (Pharmaceuticals)
![Page 58: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/58.jpg)
53
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Vitamin B-12
Weigh 0.5 g of the sample into the digestion vessel. Add 9 ml of HNO3, and 1 ml of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4mL) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
EasyPrepEasyPrep PlusXpressXpress Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
210 20:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 ml.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofVitamin B-12
![Page 59: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/59.jpg)
54
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Vitamin C (ground)
Weigh 0.5 g of the sample into the digestion vessel. Add 10 mL of HNO3. Gently swirl the mixture and wait approximately 15 minutes beforeclosing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofVitamin C (ground)
![Page 60: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/60.jpg)
55
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Vitamin C (whole)
Weigh 1 pill (Approx 1.4 g) into the digestion vessel. Add 10 mL of HNO3. Gently swirl the mixture and wait approximately 15 minutes beforeclosing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
1 pill (Approx 1.4 g) Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofVitamin C (whole)
![Page 61: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/61.jpg)
56
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Vitamin D
Weigh 0.5 g of the sample into the digestion vessel. Add 9 ml of HNO3, and 1 ml of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4mL) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
EasyPrepEasyPrep PlusXpressXpress Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
210 20:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 ml.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofVitamin D
![Page 62: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/62.jpg)
57
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Vitamin D3
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL of HNO3 and 1 mL of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofVitamin D3
![Page 63: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/63.jpg)
58
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Vitamin E
Weigh 0.25 g of the sample into the digestion vessel. Add 9 ml of HNO3, and 1 ml of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4mL) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
EasyPrepEasyPrep Plus
HNO3HCl
0.25 g Organic Ramp to Temperature One Touch
230 20:00 30:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 ml.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofVitamin E
![Page 64: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/64.jpg)
59
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Vitamins
Weigh 0.5 g (0.25 g w/Xpress Vessels) of the sample into the digestion vessel. Add 9 ml of HNO3, 1 ml HCl, and 1 ml of HF. Gently swirl themixture and wait approximately 15 minutes before closing the vessel.
This procedure uses hydrofluoric acid. If it is necessary to complex the residual hydrofluoric acid or redissolve insoluble fluorides formed, an additionalcomplexation step with boric acid should be used. This procedure can be found in the One Touch Method note entitled “Boric Acid HF Neutralization”.
The addition of Conc. HCl (0-4mL) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl willvary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
MARS 6MARS 6 iWave
MARSXpressMARSXpress PlusEasyPrepEasyPrep Plus
HNO3
HFHCl
0.5 g Organic Ramp to Temperature One Touch
210 15:00 30:00 800 900-1050 Off
Sample was clear, colorless and particle free upon dilution to 50 ml.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the required results onyour sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or below roomtemperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofVitamins
![Page 65: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/65.jpg)
60
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Zinc Sulfate
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL of HNO3 and 1 mL of HCl. Gently swirl the mixture and wait approximately 15minutes before closing the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. Theamount of HCl will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit thetechniques or increase the difficulties of analysis.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3HCl
0.5 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofZinc Sulfate
![Page 66: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/66.jpg)
Plastics, Polymers, & Oils
MARS™ 6
![Page 67: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/67.jpg)
62
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type Method Type
Reagents
Max Sample Weight Sample Type
www.cem.com Form prepared by:
MARS 6TM
Method Note
Results
General Precautions
* Ramp times and power may vary depending on type and number of vessels.
Mineral Oil
Weigh 0.25 g of the sample into the digestion vessel. Add 10 mL of HNO3. Gently swirl the mixture and wait approximately 15 minutes beforeclosing the vessel.
MARS 6MARS 6 iWave
XpressXpressPlusEasyPrepEasyPrep Plus
HNO3
0.25 g Organic Ramp to Temperature One Touch
200 15:00 15:00 800 900-1050 Off
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) The control / reference vessel must contain the largest and most reactive sample.d) Manual venting of CEM vessels should be performed when wearing hand/eye/body protection and when the vessel contents are at or belowroom temperature to avoid the potential for chemical burns. Always point the vent hole away from the operator.e) If programming as One Touch, the ramp time and power will be automatically determined based on the number and type of vessels detected.
Microwave Digestion ofMineral Oil
![Page 68: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/68.jpg)
Discover® SP-D 80
Consumer Products
![Page 69: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/69.jpg)
64
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Suppository
Weigh 1 Capsule (approx. 1 g) of the sample into the digestion vessel. Add 12 mL of HNO3. Gently swirl the mixture and wait approximately15 minutes before capping the vessel.
The use of HF may be required to digest any silicates found in the sample.
Discover SP-D 80 mL QuartzQuartz with Teflon liner
HNO3
1 capsule Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) If using HF follow restictions listed in HF addendum.
Microwave Digestion ofSuppository
![Page 70: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/70.jpg)
65
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Vaseline
Weigh 0.5 g of the sample into the digestion vessel. Add 8 mL of HNO3 and 2 mL of HF. Gently swirl the mixture and wait approximately 15minutes before capping the vessel.
This procedure uses hydrofluoric acid. If it is necessary to complex the residual hydrofluoric acid or redissolve insoluble fluorides formed, an additionalcomplexation step with boric acid should be used. This procedure can be found in the method note entitled “Boric Acid HF Neutralization”.
Discover SP-D 80 mL Quartz with Teflon liner HNO3HF
0.5 g Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) If using HF follow restictions listed in HF addendum.
Microwave Digestion ofVaseline
![Page 71: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/71.jpg)
66
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Wax (candle)
Weigh 0.5 g of the sample into the digestion vessel. Add 8 mL of HNO3 and 2 mL of HF. Gently swirl the mixture and wait approximately 15minutes before capping the vessel.
This procedure uses hydrofluoric acid. If it is necessary to complex the residual hydrofluoric acid or redissolve insoluble fluorides formed, an additionalcomplexation step with boric acid should be used. This procedure can be found in the method note entitled “Boric Acid HF Neutralization”.
Discover SP-D 80 mL Quartz with Teflon liner HNO3HF
0.5 g Organic Ramp to Temperature
200 5:00 5:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.c) If using HF follow restictions listed in HF addendum.
Microwave Digestion ofWax (candle)
![Page 72: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/72.jpg)
Nutraceuticals
Discover® SP-D 80
![Page 73: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/73.jpg)
68
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Ginko (Ground)
Weigh 0.5 g of the sample into the digestion vessel. Add 10 mL HNO3. Gently swirl the mixture and wait approximately 15 minutes beforecapping the vessel.
Discover SP-D 80 mL Quartz HNO3
0.5 g Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofGinko (Ground)
![Page 74: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/74.jpg)
69
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Ginko (Whole Pill)
Weigh 1 pill (approx 1 g) of the sample into the digestion vessel. Add 10 mL HNO3 and 2 mL H2O2 (30%). Gently swirl the mixture and waitapproximately 15 minutes before capping the vessel.
Discover SP-D 80 mL Quartz HNO3
H2O2 (30%)
1 pill Organic Ramp to Temperature
210 5:00 5:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofGinko (Whole Pill)
![Page 75: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/75.jpg)
70
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Kelp
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL HNO3 and 1 mL HCl. Gently swirl the mixture and wait approximately 15minutes before capping the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl \will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
Discover SP-D 80 mL Quartz HNO3
HCl
0.5 g Organic Ramp to Temperature
200 5:00 5:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofKelp
![Page 76: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/76.jpg)
Pharmaceutical & Biotech
Discover® SP-D 80
![Page 77: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/77.jpg)
72
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Allergy Pill (ground)
Weigh 0.25 g of the sample into the digestion vessel. Add 9 mL HNO3 and 1 mL H2O2 (30%). Gently swirl the mixture and wait approximately15 minutes before capping the vessel.
Discover SP-D 80 mL Quartz HNO3
H2O2 (30%)
0.25 g Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofAllergy Pill (ground)
![Page 78: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/78.jpg)
73
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Allergy Pill (whole pill)
Weigh 1 pill (approx 1 g) into the digestion vessel. Add 10 mL HNO3 and 2 mL H2O2 (30%). Gently swirl the mixture and wait approximately15 minutes before capping the vessel.
Discover SP-D 80 mL Quartz HNO3
H2O2 (30%)
1 pill Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofAllergy Pill (whole pill)
![Page 79: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/79.jpg)
74
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Antioxidant (Ground)
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL HNO3 and 1 mL HCl. Gently swirl the mixture and wait approximately 15minutes before capping the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HClwill vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
Discover SP-D 80 mL Quartz HNO3 HCl
0.5 g Organic Ramp to Temperature
200 5:00 5:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofAntioxidant (Ground)
![Page 80: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/80.jpg)
75
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Antioxidant (whole pill)
Weigh 1 pill (approx1.4 g) of the sample into the digestion vessel. Add 10 mL HNO3 and 2 mL H2O2 (30%). Gently swirl the mixture and waitapproximately 15 minutes before capping the vessel.
Discover SP-D 80 mL Quartz HNO3
H2O2 (30%)
1 pill Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofAntioxidant (whole pill)
![Page 81: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/81.jpg)
76
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Aspirin (Ground)
Weigh 0.25 g of the sample into the digestion vessel. Add 9 mL HNO3 and 1 mL H2O2 (30%). Gently swirl the mixture and wait approximately15 minutes before capping the vessel.
Discover SP-D 80 mL Quartz HNO3
H2O2 (30%)
0.25 g Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofAspirin (Ground)
![Page 82: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/82.jpg)
77
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Aspirin (Whole)
Weigh 1 pill (approx 1 g) of the sample into the digestion vessel. Add 10 mL HNO3. Gently swirl the mixture and wait approximately 15minutes before capping the vessel.
Discover SP-D 80 mL Quartz HNO3
1 pill Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofAspirin (Whole)
![Page 83: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/83.jpg)
78
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Calpan Pantothenic Acid
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL HNO3 and 1 mL HCl. Gently swirl the mixture and wait approximately 15minutes before capping the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl \will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
Discover SP-D 80 mL Quartz HNO3
HCl
0.5 g Organic Ramp to Temperature
200 5:00 5:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofCalpan Pantothenic Acid
![Page 84: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/84.jpg)
79
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Cephalexin
Add1 mL of the sample into the digestion vessel. Add 10 mL HNO3. Gently swirl the mixture and wait approximately 15 minutes beforecapping the vessel.
Discover SP-D 80 mL Quartz HNO3
1 mL Organic Ramp to Temperature
210 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofCephalexin
![Page 85: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/85.jpg)
80
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Chromium Chelate
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL HNO3 and 1 mL HCl. Gently swirl the mixture and wait approximately 15minutes before capping the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl \will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
Discover SP-D 80 mL Quartz HNO3
HCl
0.5 g Organic Ramp to Temperature
200 5:00 5:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofChromium Chelate
![Page 86: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/86.jpg)
81
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Cold and Flu Medicine (liquid)
Add 0.75 mL of the sample into the digestion vessel. Add 9 mL HNO3 and 1 mL H2O2 (30%) Gently swirl the mixture and wait approximately15 minutes before capping the vessel.
Discover SP-D 80 mL Quartz HNO3
H2O2 (30%)
0.75 mL Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofCold and Flu Medicine (liquid)
![Page 87: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/87.jpg)
82
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Diclofenac K
Weigh 0.5 g of the sample into the digestion vessel. Add 10 mL HNO3 and 2 mL H2O2 (30%) Gently swirl the mixture and wait approximately15 minutes before capping the vessel.
Discover SP-D 80 mL Quartz HNO3
H2O2 (30%)
0.5 g Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofDiclofenac K
![Page 88: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/88.jpg)
83
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Empty Capsule
Weigh 0.1 g of the sample into the digestion vessel. Add 10 mL HNO3. Gently swirl the mixture and wait approximately 15 minutes beforecapping the vessel.
Discover SP-D 80 mL Quartz HNO3
0.1 g Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofEmpty Capsule
![Page 89: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/89.jpg)
84
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Fish Oil (No Capsule)
Weigh 0.25 g of the sample into the digestion vessel. Add 10 mL HNO3. Gently swirl the mixture and wait approximately 15 minutes beforecapping the vessel.
Discover SP-D 80 mL Quartz HNO3
0.25 g Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofFish Oil (No Capsule)
![Page 90: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/90.jpg)
85
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Folic Acid
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL HNO3 and 1 mL HCl. Gently swirl the mixture and wait approximately 15minutes before capping the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HClwill vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
Discover SP-D 80 mL Quartz HNO3
HCl
0.5 g Organic Ramp to Temperature
200 5:00 5:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofFolic Acid
![Page 91: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/91.jpg)
86
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Ginko (Ground)
Weigh 0.5 g of the sample into the digestion vessel. Add 10 mL HNO3. Gently swirl the mixture and wait approximately 15 minutes beforecapping the vessel.
Discover SP-D 80 mL Quartz HNO3
0.5 g Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofGinko (Ground)
![Page 92: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/92.jpg)
87
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Ginko (Whole Pill)
Weigh 1 pill (approx 1 g) of the sample into the digestion vessel. Add 10 mL HNO3 and 2 mL H2O2 (30%). Gently swirl the mixture and waitapproximately 15 minutes before capping the vessel.
Discover SP-D 80 mL Quartz HNO3
H2O2 (30%)
1 pill Organic Ramp to Temperature
210 5:00 5:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofGinko (Whole Pill)
![Page 93: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/93.jpg)
88
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Kelp
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL HNO3 and 1 mL HCl. Gently swirl the mixture and wait approximately 15minutes before capping the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl \will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
Discover SP-D 80 mL Quartz HNO3
HCl
0.5 g Organic Ramp to Temperature
200 5:00 5:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofKelp
![Page 94: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/94.jpg)
89
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Losataran K
Weigh 0.5 g of the sample into the digestion vessel. Add 10 mL HNO3 and 2 mL H2O2 (30%). Gently swirl the mixture and wait approximately15 minutes before capping the vessel.
Discover SP-D 80 mL Quartz HNO3
H2O2 (30%)
0.5 g Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofLosataran K
![Page 95: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/95.jpg)
90
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Magnesium Oxide
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL HNO3 and 1 mL HCl. Gently swirl the mixture and wait approximately 15minutes before capping the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl \will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
Discover SP-D 80 mL Quartz HNO3
HCl
0.5 g Organic Ramp to Temperature
200 5:00 5:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofMagnesium Oxide
![Page 96: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/96.jpg)
91
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Manganese Carbonate
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL HNO3 and 1 mL HCl. Gently swirl the mixture and wait approximately 15minutes before capping the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl \will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
Discover SP-D 80 mL Quartz HNO3
HCl
0.5 g Organic Ramp to Temperature
200 5:00 5:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofManganese Carbonate
![Page 97: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/97.jpg)
92
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Metaformin
Weigh 0.5 g of the sample into the digestion vessel. Add 10 mL HNO3 and 2 mL H2O2 (30%). Gently swirl the mixture and wait approximately15 minutes before capping the vessel.
Discover SP-D 80 mL Quartz HNO3
H2O2 (30%)
0.5 g Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofMetaformin
![Page 98: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/98.jpg)
93
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Multivitamin (ground)
Weigh 0.5 g of the sample into the digestion vessel. Add 10 mL HNO3 and 2 mL H2O2 (30%). Gently swirl the mixture and wait approximately15 minutes before capping the vessel.
Discover SP-D 80 mL Quartz HNO3
H2O2 (30%)
0.5 g Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofMultivitamin (ground)
![Page 99: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/99.jpg)
94
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Multivitamin (whole)
Weigh 1 pill (approx 2.5 g) of the sample into the digestion vessel. Add 10 mL HNO3 and 2 mL H2O2 (30%). Gently swirl the mixture and waitapproximately 30 minutes before capping the vessel.
Discover SP-D 80 mL Quartz HNO3
H2O2 (30%)
1 pill Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofMultivitamin (whole)
![Page 100: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/100.jpg)
95
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Stearic Acid
Weigh 0.25 g of the sample into the digestion vessel. Add 10 mL HNO3. Gently swirl the mixture and wait approximately 15 minutes beforecapping the vessel.
Discover SP-D 80 mL Quartz HNO3
0.25 g Organic Ramp to Temperature
210 10:00 10:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofStearic Acid
![Page 101: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/101.jpg)
96
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Thiamine
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL HNO3 and 1 mL HCl. Gently swirl the mixture and wait approximately 15minutes before capping the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl \will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
Discover SP-D 80 mL Quartz HNO3
HCl
0.5 g Organic Ramp to Temperature
200 5:00 5:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofThiamine
![Page 102: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/102.jpg)
97
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
USP 232/233
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL of HNO3 and 1 mL of HCl . Gently swirl the mixture and wait approximately 15minutes before capping the vessel.
Discover SP-D 80 mL Quartz HNO3
HCl
0.5 g Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofUSP 232/233
![Page 103: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/103.jpg)
98
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Vitamin C (ground)
Weigh 0.5 g of the sample into the digestion vessel. Add 10 mL HNO3. Gently swirl the mixture and wait approximately 15 minutes beforecapping the vessel.
Discover SP-D 80 mL Quartz HNO3
0.5 g Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofVitamin C (ground)
![Page 104: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/104.jpg)
99
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Vitamin C (whole)
Weigh 1 pill (approx1.4 g) of the sample into the digestion vessel. Add 10 mL HNO3 and 2 mL H2O2 (30%). Gently swirl the mixture and waitapproximately 15 minutes before capping the vessel.
Discover SP-D 80 mL Quartz HNO3
H2O2 (30%)
1 pill Organic Ramp to Temperature
200 5:00 3:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofVitamin C (whole)
![Page 105: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/105.jpg)
100
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Vitamin D3
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL HNO3 and 1 mL HCl. Gently swirl the mixture and wait approximately 15minutes before capping the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl \will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
Discover SP-D 80 mL Quartz HNO3
HCl
0.5 g Organic Ramp to Temperature
200 5:00 5:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofVitamin D3
![Page 106: CEM Pharmaceutical Method Note Compendiumcem.com/media/contenttype/media/literature/MetNote... · 2018-04-09 · d) Manual venting of CEM vessels should be performed when wearing](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e83243bc3f89e686b130120/html5/thumbnails/106.jpg)
101
Procedure
Notes
Recommended Equipment Recommended Vessels
Microwave Digestion of
Heating Program
1
Ramp (mm:ss) Hold (mm:ss) Pressure (psi) Power (W) StirringStage Temp (°C)
Control Type
Reagents
Max Sample Weight Sample Type / Vent Program
www.cem.com Form prepared by:
Discover® SP-D 80TM
Method Note
Results
General Precautions
Zinc Sulfate
Weigh 0.5 g of the sample into the digestion vessel. Add 9 mL HNO3 and 1 mL HCl. Gently swirl the mixture and wait approximately 15minutes before capping the vessel.
The addition of Conc. HCl (0-4 ml) is appropriate for the stabilization of Ag, Ba and Sb, and high concentrations of Fe and Al in solution. The amount of HCl \will vary depending on the matrix and the concentration of the analytes. The addition of HCl may, however, limit the techniques or increase the difficulties ofanalysis.
Discover SP-D 80 mL Quartz HNO3
HCl
0.5 g Organic Ramp to Temperature
200 5:00 5:00 400 300 Med
Sample was clear, colorless, and particle free upon dilution to 50 mL.
a) This procedure is a reference point for sample digestion using a CEM system and may need to be modified or changed to obtain the requiredresults on your sample.b) If using a vessel other than the recommended choice, adjust sample size and pressure limit to values appropriate for the vessel chosen.
Microwave Digestion ofZinc Sulfate