Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

18
MMT -530M Revision 7 Jan. 2000 Page 1 Site Plannin Guide GE 1.0T Signa Horizon Mobile MR System (CZ-70 and up) @ 1999 Calumet Coach Company All Rights Reserved This guide contains confidential information of Calumet Coach Company. You may not copy it or any part of it without the written permission of Calumet Coach Company. This manual may be used only by you, and only for the purposes for which it was intended. You may not disclose this manual or the confidential information it contains outside your company. If you wish to copy part of this manual or to use it other than as described above, you must write to Calumet Coach Company seeking permission to do so: Keith Moore, President Calumet Coach Company 2150 E. Dolton Road Calumet City, IL 60409-1411 USA The infonnation contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and may not be used. either partially or wholly. for any purpose inconsistent with the purpose for which it was produced. The contents may Dot be reproduced or disclosed without prior consent.

Transcript of Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

Page 1: Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

MMT -530MRevision 7Jan. 2000Page 1

Site Plannin GuideGE 1.0T Signa Horizon Mobile MR System

(CZ-70 and up)

@ 1999 Calumet Coach CompanyAll Rights Reserved

This guide contains confidential information of Calumet Coach Company. You may not copy it or any part of itwithout the written permission of Calumet Coach Company. This manual may be used only by you, and onlyfor the purposes for which it was intended. You may not disclose this manual or the confidential information itcontains outside your company.

If you wish to copy part of this manual or to use it other than as described above, you must write to CalumetCoach Company seeking permission to do so:

Keith Moore, PresidentCalumet Coach Company2150 E. Dolton RoadCalumet City, IL 60409-1411 USA

The infonnation contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and maynot be used. either partially or wholly. for any purpose inconsistent with the purpose for which it was produced. The contents may Dot bereproduced or disclosed without prior consent.

Page 2: Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

MMT-530MRevision 7Jan. 2000Page 2

LIST OF REVISIONS

In accordance with our policy of continued product improvement, Calumet Coach Company reservesthe right to make changes in the equipment, design, specifications and materials of the productdescribed herein.

Any problems or questions related to the components or systems covered in this booklet may bedirected to:

Calumet Coach CompanyService Department

2150 E. Dolton RoadCalumet City, Illinois60409-1411 USA

Telephone: 800-839-0630 (24/7 Service)Fax: 708-868-5101

http://www.calumetcoach.com

The information contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and maynot be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not bereproduced or disclosed without prior consent.

Revisions Date

01 Initial Release 04-1996

02 Revised width to 98"; changedmagnet 07-1996centerline to 128" from rear of unit

03 Revised width to 100" 02-1997

04 Revised magnet centerline for units 47A and up 06-1997

05 Added wireless LAN antenna location 05-1998

06 Added grounding requirements; added phone 05-1999line for data link

07 Revised generator housing 01-2000

Notice

Page 3: Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

Table of Contents

MMT-530MRevision 7Jan. 2000Page 3

The information contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and maynot be used. either partially or wholly. for any purpose inconsistent with the purpose for which it was produced. The contents may not bereproduced or disclosed without prior consent.

Section Page

1. Support Pad Requirements 4

2. Customer Power Requirements 6

3. Mobile Grounding Requirements 8

4. Telephone Service 9

5. Water Requirements 10

6. Figure 1 - Medical System 11

7. Figure 2 - Pad Layout 12

8. Figure 3 - Curb Side Elevation 13

9. Figure 4 - RussellStoll Service Outlet 14

10. Figure 5 - RussellStoll Chart 15

11. Figure 6 - Mobile Grounding Requirements Chart 16

12. Figure 7 - Turning Requirements 17

13. Figure 8 - Stair and Platform Detail 18

14. Appendix: Photographic References Al

Page 4: Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

MMT -530MRevision 7Jan. 2000Page 4

Support Pad RequirementsGE 1.0T Signa Horizon Mobile MR System

The following is a list of recommendations and requirements for a concrete support pad. However, due tovarying site conditions, the actual pad design should be prepared by an appropriately licensed structural orarchitectural engineer.

Recommended Support PadA full padmeasuring10'-0" x 40'-0", locatedas shownonFigure2 andFigure3 as the crosshatching,is therecommended support pad.

Recommended Service PadA full pad measuring 21' -0" x 55'-6", located as shown on Figure 2 and Figure 3 is recommended to provide aservice access.

Support Pad DepthRecommendations for the width and length of the pad are given above. Based upon the weight distributioninformation given in Figure 3 and existing site conditions, the depth should be determined by a local contractor.It is recommended that non-ferrous reinforcement materials be used for pad reinforcement. GE must approveplans for pad construction.

Support Pad LevelnessThe support pad must be level to ensure proper operation of the MRI system. The pad must not exceed .125"deviation in 10'-0". If the minimum support pads are selected, rather then the recommended single pad, theymust also meet this specification.

Vehicle AccessA firm, level surface is required around the mobile unit to provide access to the tractor/trailer, aid with patienthandling, servicing the unit, and delivering of cryogens.

Steel Reinforced Concrete PadA maximum of 4 lbs per square foot of steel is allowed in the concrete pads beneath the mobile unit.

The infonnation contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and maynot be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not bereproduced or disclosed without prior consent.---

Page 5: Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

MMT-S30MRevision 7Jan. 2000Page 5

Support Pad RequirementsGE 1.0T Signa Horizon Mobile MR System

Minimum Support Pad ReauirementsA front pad measuring 10'-0" x 4'-0" and a rear pad measuring 10'-0" x 15'-0" located as shown on Figure 2will provide the minimum support requirements.

Recommended Attachment to FacilitvAn inflatable air bag or soft seal is recommended at the point of connection from the unit to the facility. Fixedor solid connections may hinder imaging quality. Contact Calumet Coach Company or the local GErepresentative prior to construction if the proposed connection varies from the recommended.

Exclusion ZoneAn area of 5'-0" x 5'-0", located directly below the magnet vent should be fenced off to prevent injury in theevent of magnet quenches.

Vehicle MovementThe MRI is very sensitive to vibration and moving metal. Consequently, all vehicle traffic shall be kept as faraway as possible from the pad. Moving ferrous materials having the listed masses should be limited to areas asdescribed in the GE site planning publication. Contact GE to obtain the latest version.

Vibration I Foundation DesiQnIt is essential to minimize vibration to ensure high resolution Image quality. To achieve this, GE has produceda scope of work, specifications, and test methods for new and existing sites. GE in their Signa Horizon siteplanning publication provides this documentation. Contact GE to obtain the latest version.

SwinQ Clearance NotePlease verify the actual dimensions of the rearmost projections on the cab of your tractor to the centerline oftandem suspension or centerline of the fifth wheel plate on your tractor. Refer to Figure 7 for proper tractorsizing information.

The information contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and maynot be used, either partially or wholly. for any purpose inconsistent with the purpose for which it was produced. The contents may not bereproduced or disclosed without prior consent.

Page 6: Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

MMT-530MRevision 7Jan. 2000Page 6

Customer Power RequirementsGE 1.0T Signa Horizon Mobile MR System

Electrical ServiceA single electrical power source is required for operation of the MRI system. 480 Volt A.C., 3 Phase, 125KVA, fused at 150Amps

Load Reaulation at Line FreauencvWiresare to be sizedsuchthat the line voltagedropsfromthe powersourceto themobileunit is less then2%of thenominalvoltageforratedloadof themobileunit.

ConfiaurationThree phase, five wire, wye connection, with ground and neutral (not less than AWG #1/0).

Freauencv. 60 Hz j: 0.5%. Phase Balance: + 2% maximumphase-to-phase line voltage difference lowest phase.. Maximum voltage variation: j: 2% from nominal steady state (under the worst case conditions of line

voltage). Maximum voltage range: 420 volts - 500 volts. Source impedance: 480 volt line to line impedance <= 0.1 ohms

Connector TvpeThe unit is supplied with a 35-foot (usable) cable and male connector. The connector is a RussellSto1l200 Ampplug #DS2504MP.

Customer FacilitvThe facility must have the matching receptacle as specified in Figure 4 and Figure 5. The receptacle is aRussellStoll #DF2504FRAB.

Voltaae SuraesTransient voltage variations caused by external loads must not:. Exceed + 10%. exceed five cycles duration. Occur more than ten times an hour.

The infonnation contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and mayDot be used, either partiaIly or whoIly, for any purpose inconsistent with the purpose for which it was produced. The contents may not bereproduced or disclosed without prior consent.

- --

Page 7: Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

MMT-530MRevision 7Jan. 2000Page 7

Customer Power RequirementsGE 1.0T Signa Horizon Mobile MRSystem

Ground ConductorAn insulated ground conductor sized in accordance with National, State, and local codes shall be installedbetween the facility vault and the MRI System ground bus location in the power distribution unit.

Exclusion ZoneThe MRI unit is equipped with magnetic shielding. The exclusion zone for cardiac pacemakers,neurostimulators, and other biostimulation devices is recommended at 5 gauss (0.5mT). Signs, provided by GE,must be posted to alert all who approach the unit of this requirement.

The l.OT magnet systems exclusion zone does not extend into normally accessible locations that are exterior tothe mobile unit. The appropriate warning signs are permanently attached to the scan room doors. The 3 gaussline (0.3mT) is located approximately 3 feet from the unit walls. The 1 gauss line (O.lmT) is locatedapproximately 8 feet from the van walls. The 0.5 gauss line (0.05mT) is located approximately 12 feet from thevan walls.

R.F. ShieldinaTheR.F.shieldis includedwiththeMRIsystemandwillprovideminimumlevelof attenuation:10Mhz- 100Mhz,85db (planewave,electricfield,magneticwave).

Note:The standard connector for the unit is a RussellStoll #DS2504MP. If an existing site currently implements adifferent connector, connector configuration, or the available power supply varies from above specifications,please contact Calumet Coach Company to arrange for a compatible power connector before the unit leaves thefacility.

WARNING:It is the operator's responsibility to verify that the shore power receptacle is electrically compatible with themobile system's umbilical connection prior to plugging in. Plugging into a receptacle not electricallycompatible could cause injury or damage.

It is also recommended to inspect the power cord, connector, and fasteners prior to connection. If duringinspection you suspect internal or external damage has occurred, have a qualified electrician inspect and repairthe damage before further use. Follow the maintenance schedule in the Operator and Maintenance Manual forsafe operation of the mobile unit.

The information contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and maynot be used. either panially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not bereproduced or disclosed without prior consent.

--- - -

Page 8: Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

MMT -530MRevision 7Jan. 2000Page 8

Mobile Grounding RequirementsGE 1.0T Signa Horizon Mobile MR System

NOTE: For the associated drawinQplease refer to FiQure6. All work is to be done in accordance with local and national electrical codes.. Information shown here is only a recommendation and must be verified local site codes and national

site codes.. Ground wires inside enclosures are to be taped green for the entire visual length for identification.. If a separately derived secondary system transformer is used, a bonding jumper between the grounded

conductor (neutral) and the equipment - grounding conductor must be used.

GroundinQTheground for our system shall originate at the system power source (transformer) or first access point ofpower into the facility, and be continuous to our system power disconnect in the room. This ground can bespliced with "High Compression Fittings" and should be terminated at each distribution panel it passes through.When it is broken for a connection to a panel, it shall be connected into an approved grounding block with theincoming and outgoing ground in the same grounding block. This grounding block is then connected to thesteel panel. Never use the steel or any other material of the panel as the grounding block.

The connection at the power source shall be at the grounding point of the "Neutral - Ground" if a "WYE"transformer is being used, or at the typical grounding points of a separately derived system. In the case of anexternal facility, it shall be bonded to the facility ground point at the service entrance.

Ground WireThe ground wire shall be copper wire and the same size as the disconnect feeders. The ground wire impedancefrom our system disconnect, including the ground rod, shall not have an impedance greater than 2 ohms to earthas measured by one of the applicable techniques described in Section 4 of ANSIlIEEE Standard 142-1982.

Special GroundinQ Note:The unit must have an earth driven groundrod withinfive (5) feet of the hospitablepower receptacle. Agrounding cable of a minimum 4 GA must be connected between the grounding rod and the grounding pin ofthe hospital power receptacle, and another cable to be kept as short as possible, and must not exceed 8 feet inlength. A separate grounding conductor must still be run with the phase conductors to the source of power fromthe grounding pin of the hospital power receptacle in accordance with NEe. Article 250-23.

The infonnation contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and maynot be used, either partially or wholly. for any purpose inconsistent with the purpose for which it was produced. The contents may not bereproduced or disclosed without prior consent.- -

Page 9: Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

MMT-530MRevision 7Jan. 2000Page 9

Telephone Service RequirementsGE 1.0T Signa Horizon Mobile MR System

Telephone Service. The unit is supplied with three telephone connections and three data line connections.

. The connector type that is used is a model Hubbell PH-6595 (inlet), supplied by Calumet Coach.

. Two Hubbell PH-6599 50 foot telephone-connectingcables are included with the unit.

. The data connections utilize a 50-foot UTP CAT 5 Cable with an RJ-45 connector.. The customer is also required to supply three Hubbell phone connectors, model PH-6597

(weatherproof phone outlets) at the site.

The infonnation contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and maynot be used, either partially or wholly. for any purpose inconsistent with the purpose for which it was produced. The contents may not bereproduced or disclosed without prior consent.

Page 10: Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

MMT-530MRevision 7Jan. 2000Page 10

WaterRequirementsGE 1.0T Signa Horizon Mobile MR System

During winter conditions provisions must be made to ensure that any and all water lines do not freeze a result ofweather conditions.

Humidifier Water FillThe unit contains a water storage tank for the humidifier. This tank is located in the equipment room and mustalways contain water to insure the specified humidity level remains constant. There are two options for fillingthe tank:. A %" I.P.S. male threaded hose connection is located under the equipment room.

. If a water supply hose is not available, the humidifier storage tank may be filled manually via a portlocated on the top of the tank.

Potable Water Supplv Reauirements (option)A cold water supply line is required, with a flow rate of 5 gallons per minute, 45-60 psi and a maximumtemperature of 70 degrees Fahrenheit. The unit will be supplied with a %" diameter, 20' long hose terminatedwith a %" I.P.S. male threaded hose connector. The facility is to provide a W' female connector to connect tothe units 20' long hose.

Waste Water Connections (sink option)A 1Y2"IPS male connection is required to sink wastewater.

Both the water supply and the wastewater drain connections are to be contained in a recessed water connectionbox located as shown in Figure 1.

Any pipes and drains within 20'-0" of the magnet isocenter should be of non-ferrous materials such as PVC,copper, or brass. All plumbing must comply with applicable codes.

The unit is supplied with a 20' long 1W' diameter drain hose terminated with a 1W' male threaded connectorfor sanitary wastewater drainage. The facility must provide means of sanitary wastewater drainage from thesystem that comply with locally applicable codes.

The information contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and maynot be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not bereproduced or disclosed without prior consent.

----

Page 11: Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

~ ~~~!!

128.

EXHAUSTrAN

'00'

L L COL H£ATEXCHANCERPATIENTBLOWER

\VENT DEFUCTOR

~ rPAcI\PM~CER'AN '~

3';"&.

~~

U~I~I,.'I

108.

128.

f-2&'

(iWR :tiH~ENTE~:.~l)

270.

~e'"OPTIONAL

MY "'tN~OR4' 18....

1&8'LA. 3"

I v ~ ~G.\CYMEDIAOF'TION:--2 4' '~2','-..

12;-

SEIMCEDOOR

,"01:;~~z~d

.. I~

h~3 208volTPANElHVAC UNIT72,000 BTU

MMT-530MRevision 7Jan. 2000Page 11

36.-

I ~ nREDETEcnoNPANELr--- IANDCJ 208 VOlTPANELGENERATOR

13&' ~ 32'---I1-4'EST.WEIGHT . 2215801 48

(roR KINGPIN) I

,&1/2't---

II1UI;:::::j:

~gl:;~~u~

'---

VARE DETEC~ONPANEL

H I rWATERy TANKHVAC UNIT60,000 BTU

~UIi4IDIF'IE

HVAC UNIT

/,,20 TMK

~ f--tiUUIOIFlEI

I

.r_.l-~hmT=

,2.....

7~ d'/4' u

4'-

123-

~5","~~~

i~§!

The information contained in this booklet is property of Calumet Coach Company, The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent withthe purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.

~2'r ," 2"~ 7" 1 AI. Q!,!!;L..

96 1

8.1-- §"'RI o.ERGENCY STOF

~ Of IUGNI

.,.

~ ~

I

~- r Duct~ 10: AIR OUCT'11~

u 'TI I~16/1,NET

!-<CY MIt>'"

*

t=

[F, II I I

~~~~

/

It. LANDING LEGS

48'-0"

Fiaure 1: GE 1.0T Siana Horizon Mobile _MRSystem

KING PIN It

117 3/".

I " I ....1'\ I RO<.JoooR

PATIENTun wrTH 1---6'SAF'ETYHANDRAILS66 1/2'62

I

LJ I' 90'

9&'36" 12-

Page 12: Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

~ ~~~!!

I4'-6"

10' I 8'-4'

6'-6"

L

6'

APPRO~. . Of' MAG. VENTING

IS'MINIMUM SUPPORT PAD

10'AXL£ SPACING

10'-0' IC.l.jMAGNET

P"t'RAMID STANDS

CZ 70 AND UP

55'-6'FULL SERvICE PAD

21' 4'MINIMUMSUPPOI

PAD

APPRQXIWATE LOCATION 'tt:a:CAl/TEIAND G.E, EXTERN.t.l. OPTfHS INTERfACE P.

9'-6"

MMT-530MRevision 7Jan. 2000Page 12

1'-141/2"

3'1

lAIC

I

21'IA/C

I~~J,tPPAOXIMATELY e'-o.IN FRONTOF' THE AleUNITS F"QR SERVICENTER THE UNIT IS INPLACE.

OF IHUMIDIFIER FILL

The information contained in this booklet is property of Calumet Coach Company, The contents are considered to be confidential and may not be used, either partially or wholly. for any purpose inconsistent withthe purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.

~~ ~NT-CAAQE TOCEN.UN[OPENING

r m... APPoo"' ,[1.Y "-0"X20'-0"CL£AR rOR NELIUNFlLl ACCESS 0 SERVICE

CLIP-ON STAIRS

NOTE: PLATFORM, WHENREQUIRED, SHALLBE A SITESPECIFIC OPTION.WHEN USED,THE PLATFORMMUST BE MECHANICALLYISOLATED FROM THE TRAILER,

8 7'-6" 10'-7"3'-8"

.25'-4 1/2" 2'-4 1/2'MININUM Cl£AR OPENING REQUIRED

FDA F'ACIUT'I' UNI(

10'-3"

Figure 2: Pad Layout

Page 13: Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

~ ~~~!!

10'-8' jC.L/w.GNET I

7'-3 1/2' I 5'-71JSERVICE HATCh-"""""""

VENT DEFlECTOR

13'-6"

rIIIIIILFLOOR LEVEL

g'

,L./PYRANIOSTANDS10'

AXLE SPACIN'

IEST. REAR WT.34000 LBS,

MMT-530MRevision 7Jan. 2000Page 13

1'-4 1/2-

4S'

CONTACT GE HEN..THNET SVCS FOROPTIONAL WIRELESS LAN COt.lMUNICATION INSTALLATIONOF WAlL MOUNTED COMPONENTS AT HOSPITAL SITE

3'l

2'-6-

..5"RECESS FRO'" FACE/TRAILER II AIC

GEN.

HANDRAILS

7'-6" IPATIENT Uf'T

C.l.KINGPIN

.'

2 1/2"4"

'-S" 10'-3"S'-2"FRAMED OPENING

MINIMUM CLEAR Jp~',.;~ ~6~ACIUTY UNK9'-6 3/".

TOTALWEIGHTAPPROXIMATElY62500 LBS.

II

IEST. FRONT WT.24000 LBS.

C.L/LANDING GEAR

Fiaure 3: Curbside Elevation

The infc)fJnation contained in this booklet is property of Calumet Coach Company, The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent withthe purpose for which it was produced, The contents may not be reproduced or disclosed without prior consent.

Page 14: Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

~ ~~~!!

SAFETY DISCONNECTSWITCH

150 AMPERE FUSINGFRS-R-100 OR EQUAL.

ON 0OFF0

RECEPTACLE:r #DF2504FRABW/SPRING DOOR

IHEIGHT TO BEGOVERNED BYNATIONALANDLOCAL CODES

480 VOLT30, 60HzONLY

CUSTOMER MUSTCLEARLY LABEL

GROUNDROD FINISHEDGRADE

~~~~~~3 PHASES + GROUNDAC 480V 150A SERVICE

WARNING:HOSPITALRECEPTACLEWIRINGAND MOBILE UNIT PLUG WIRINGMUSTBE INITIALLYCHECKEDFOR PROPERCONTINUITYBY A QUALIFIEDELECTRICIAN

.lliITL.THE INFORMATIONCONTAINEDIN THIS DIAGRAMIS RECOMMENDEDTOMEETTHE REQUIREMENTSOF THE EQUIPMENTOPERATEDAND ISSUBJECTTO LOCALCODEAPPROVAL.

Fiaure 4: RusseliStoll Service Outlet

MMT-530MRevision 7Jan. 2000Page 14

The information contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and may not be used. either partially or wholly. for any purpose inconsistent withthe purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.

Page 15: Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

~ ~~~!! MMT-530MRevision7Jan,2000Page 15

3C/N/G AC 480V 60 Hz RECEPTACLE

MAIN DISCONNECT3C/N/G AC 480V 150 AMPERE

NEUTRAL-WHITE PHASE 'C'-BLUE(208/240V), BROWN(480V)

o GROUND - GREEN

PHASE 'A' -BLACK(208/240V), YELLOW(480V) PHASE 'B'-RED(208/240V), ORANGE(480V)

VIEW FROM BACK SIDE OF RECEPTACLEINSTALLATIONBY CUSTOMER

ABBREVIATIONS:

3C-PHASE CONDUcroRSN - NEUTRALG-GROUND

Fiaure 5: RusseliStoll Chart

The infonnation contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent withthe purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.

WIRINGCHART

AMPS/WIRE DESCRIPTION RECEPTACLE

WIRES POLES PART NUMBER

5 5 480 VOLT (200 AMP)5 WIRE RUSSELLSTOLL

MAXIMUMWIRE SIZE AS RECEPTACLEREQUIREDBY #DF 2504 FRABLOCAL CODES THIS RECEPTACLEMUST

BE WATERPROOF

3 WYE CONNECTION150 AMP480 VOLT ACPHASETOTAL 3 <YNEUTRAL& GROUND

Page 16: Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

~ ~~~!!: MMT-530MRevision 7Jan. 2000Page 16

GROUND BUS BARI - ul

r lIIII

DISCONNECT LOCATEDAT VAN CONNECTONSEE ELECTRICALREQUIREMENTS FOR AMP.RATINGS

RUSSELLSTOLLDS2S04FP FEMALEr RECEPTACLE

I OREQUIVALENTL1

L2

#1/0 AWG COPPERI I I MIN.

I TO GROUND RODL AT RECEPTACLE

J---------------------

G

SERVICE EARTHMAIN GROUNDINGAT FACILITY SERVICE

Fiaure 6: Mobile Groundina Reauirements Chart

The information contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and may not be used. either partially or wholly. for any purpose inconsistent withthe purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.

ELECTRICALPIPING \II FEEDER

[!> I Ll

fI L2FROM FACILITYSERVICE EQUIPMENT

I L3

INI

IG

I I I IN

Page 17: Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

~ ~~~!! MMT-530MRevision 7Jan. 2000Page 17

PATH OF OVERHANG

---PATH OF RIGHT FRONT I./HEEL./ -

///

/

/1II~

\1\1\\1 \"v I

"/~I ......... -I -------------------------------IIIIr

1

27'-0'

C.LSUSPENS ION

50'-0'OUTER RADIUS

TO CENTER POINT OFRIGHT FRONT I./HEEL

PATH OF LEFT REAR SIDE

-----------------------------ISO' S'-O'

S'-O'

Fiaure 7: Turnina Reauirements

A minimum "A" dimension of 92' is required from rearmost projection to centerline of tandem suspension. This provides swing clearance for generator set which ismounted on the front of the trailer. Hospital is responsible to ensure the access route is clear of obstructionswhen the trailer is scheduled to arrive or depart.

The 50' minimum outside turning radius shown here has been calculated using an international harvester (Navistar) tractor Model COF-9670 with a 161" wheelbase.Turning radius will vary with towing tractor. Customer must confirm the turning radius on their tractor and prepare each site with adequate space to accommodate it.

The information contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent withthe purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.

Page 18: Calumet Coach MMT-530M Site Planning Guide - CZ-70 & Up

~ eAh~~!!

r-- 4'-4i!'_0.I J. 9!' TYP.

V

I3'-O!'

L1'-4i!'

~4'-O~'

STAIR ANO PLATFORM ASSEMBLY

MMT-530MRevision 7Jan,2000Page 18

3'-0a.0i!T10NAL STAIR LOCATlDNr -o...o-.-o-.-o-.-o-o o '

0000000000.0'.0.0.0.0."

TRAILER FLOOR

6i!'

STAIR WITHOUT PLATFORM ASSEMBLY

Fiaure 8: Stair and Platform Detail

The infonnation contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent withthe purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.