    Application Note

    Pressure Calibration

    Pressure instrumentation is oundin virtually every process plant.Periodic calibration o thesepressure, level, and ow instru-ments is required to keep plantsoperating efciently and saely.Fluke provides a broad range opressure calibration tools to helpyou quickly and reliably calibrate

    your pressure instrumentation.

    Fluke 719 Electric Pressure Calibrator Function

    Measure Pressure to 10,000 psi /

    700 bar with internal sensor

    Measure Pressure to 300 psi / 20 bar

    with internal sensor

    Measure Pressure to 10,000 psi / 700 bar

    with Fluke 700Pxx Pressure ModulesPressure switch test Source pressure with accessory pumps Source pressure with built-in pump

    Measure mA Source

    Loop power supply Multiunction source and measure

    Electronic data capture Communication to PC Integrated HART communication

    These pressure calibratorsaccurately measure pressureby using: Internal sensors, or External Pressure Modules

    A pressure source may beprovided by: A sel-contained internal

    pressure pump, or An external source such as an

    accessory pump or apressure bottle / regulator.

    A summary o the pressurecalibration capabilities o FlukeProcess Tools is shown below.

    Pressure Calibrationwith

    Fluke Calibrators

    717PressureCalibrator

    718PressureCalibrator

    719ElectricPressureCalibrator

    725and726MultifunctionPr

    ocessCalibrators

    753DocumentingProcessCalibrator

    754DocumentingProcessCalibrator

    5520ACalibrator

  • 7/31/2019 Calibradores Fluke Presion

    2/9

    2 Fluke Corporation Pressure Calibration

    Pressure switch calibrationVeriy the setpoint and deadband o pressureswitches using the 718 Pressure Calibrators.

    How to use the 718 to calibrate a pressure switch

    1. Depressurize and isolate thepressure switch rom the pro-cess.

    2. Plumb the 718 and make con-

    nections as per theillustration.3. Turn on the 718 and open

    the vent valve. Press the Zerobutton to clear the zero oset.Close the vent.

    4. Press the Switch Test buttonto enter the switch test mode.

    5. Apply pressure slowly with thehand pump until youapproach the setpoint. Usingthe fne adjust vernier adjustthe pressure until the switchopens and OPEN is displayed

    on the 718.

    6. Release the pressure slowlyusing the fne adjust vernieruntil RCL is displayed.

    7. Press the Switch Test but-

    ton once to read the pressurevalues or switch opening andagain to see the pressure atswitch closing.

    8. Press and hold the Switch Testbutton or 3 seconds to clearthe test results and start over.

    9. Adjust the pressure switchsetpoint until the switchcontacts open and close atthe desired pressure.

    Innovative new pump design Pumps can be easily contaminated with

    process uids Oten requires repair

    New design reduces repairs and costo ownership

    Has two clean out ports! Remove uids, clean with a cotton swab Easy access, can be servied in the feld

    COM

    mA

    30VMA

    X0-30psi

    0-206kP

    aRA

    NGE

    90psi6

    20kPa

    MAX

    S I G N A L

    T E S T

    +

    Red

    Black718

    PRESS

    URE

    CALIBR

    ATOR

    HOLD

    UNITS

    ZERO

    DAMP

    MIN

    CLR

    MAX

    HOLD

    LOOP

    POWE

    R

    BAROM

    ETRICA

    DJ.

    Typical Pressure Applications

    How to calibrate a P / I Transmitter

    With a built-in hand pump,precision measurement o bothpressure and current, and a 24volt loop supply, the Fluke 718Pressure Calibrator is a complete,sel-contained tool or the calibra-tion o P/I transmitters. To calibratea 3-15 psi / 4-20 mA transmitterusing a Fluke 718 30G Pressure

    Calibrator:1. Depressurize the transmitter,and then plumb the transmitterto the 1/8 inch NPT pressureport o the 718. Connect thetest leads per the fgure above.

    2. Turn the calibrator on. (I youneed to power the transmitter,hold down the UNITS key whileturning on the calibrator.)

    3. Press the UNITS key until PSIshows in the display.

    4. With the 718s bleed valveopen to atmosphere, press the

    ZERO key.

    Close the bleed valve. Set thepressure / vacuum valve to +,or positive pressure.

    5. Use the hand pump to applyroughly 3 psi to the transmit-ter. Partial pump strokes willapply small increments opressure. Use the fne-adjust

    knob to get reasonably close to3.00 psi.6. Press the HOLD key, and record

    the psi and mA readings. Pressthe HOLD key to resume read-ing.

    7. Calculate and record the error,using: Error = ([ (i-4)/16] [(P-3)/12]) * 100 whereError is in % o span, i is yourmeasured current in mA andP is your measured pressurein psi.

    8. Repeat steps 5 through 7 atmid-range, around 9 psi, tocheck linearity at mid-span.

    9. Repeat steps 5 through 7, nowat 15 psi, or a check at 100 %o span.

    I your calculated errors arewithin tolerance, the transmitterhas passed your As-ound test,and you are done. I necessary,perorm your zero and spanadjustments, then repeat steps5 through 9 or an As-let test.Depressurize the line, anddisconnect the 718.

  • 7/31/2019 Calibradores Fluke Presion

    3/9

    3 Fluke Corporation Pressure Calibration

    Measuring less than nine inches in lengthand weighing just over two pounds, the

    rugged 718 is easy to carry into the feld.The 718 is oered in 1 psi, 30 psi, 100 psiand 300 psi models. Media compatibility isdry air and non-corrosive gasses. A built-inpump generates pressure or vacuum. Min,Max, Hold and error % calculator unctions

    are available. The 718 can also measurepressure using any o the 29

    Fluke 700Pxx Pressure Modules, to coverapplications up to 10,000 psi. The 718comes complete with protective holster, testleads, test clips, Users Manual, andtwo 9-volt batteries (installed).

    718, 719Models

    Pressure mA

    Range Resolution Accuracy

    Max non-destructivepressure Range Resolution Accuracy

    718-1G -1 psi to 1 psi(-68.9 mbar to 68.8 mbar)

    .001 psi(0.001 mbar)

    0.05 % FS 5X ull scale 0-24 mAMeasure

    0.001 mA 0.015 % + 2 counts

    718-30G -12 psi to 30 psi(-850 mbar to 2 bar)

    0.001 psi(0.01 mbar)

    0.05 % FS 2X ull scale 0-24 mAMeasure

    0.001 mA 0.015 % + 2 counts

    718-100G -12 psi to 100 psi(-850 mbar to 7 bar)

    0.01 psi

    (0.01 mbar)

    0.05 % FS 2X ull scale 0-24 mAMeasure

    0.001 mA 0.015 % + 2 counts

    718-300G -12 to 300 psi(-850 mbar to 20 bar)

    0.01 psi(1 mbar)

    0.05 % FS 375 psi or25 bar

    0-24 mAMeasure

    0.001 mA 0.015 % + 2 counts

    719-30G -12 psi to 30 psi(-850 mbar to 2 bar)

    0.001 psi

    (0.01 mbar)

    0.025 % FS 2X ull scale 0-24 mAMeasureSource

    0.001 mA 0.015 % + 2 counts

    719-100G -12 psi to 100 psi(-850 mbar to 7 bar)

    0.01 psi

    (0.01 mbar)

    0.025 % FS 2X ull scale 0-24 mAMeasureSource

    0.001 mA 0.015 % + 2 counts

    Specifcations are based on a one year calibration cycle and apply or ambient temperature rom +18 C to +28 C. Counts are the number o increments or decrements

    o the least signifcant digit. 719 accuracy specifcation is or 6 months.

    Functional Pressure Specifcations

    Model Range Resolution Over Pressure Functions

    717 1G -1 PSI to 1 PSI, (-68.9 mbar to68.9 mbar, -6.89 to 6.89 kPa)

    0.001 psi, 0.001 mbar Over Pressure 5xFS Zero, Min, Max, Hold, Damp

    717 15G -12 PSI to 15 PSI, (-850 mbar to1.03 bar, -85 to 103 kPa)

    0.001 psi, 0.1 mbar Over Pressure 2xFS Zero, Min, Max, Hold, Damp

    717 30G -12 PSI to 30 PSI, (-850 mbar to 2bar, -85 to 206.84 kPa)

    0.001 psi, 0.1 mbar Over Pressure 2xFS Zero, Min, Max, Hold, Damp

    717 100G -12 PSI to 100 PSI, (-850 mbar to6.895 bar, -85 to 689.48 kPa)

    0.01 psi, 1 mbar Over Pressure 2xFS Zero, Min, Max, Hold, Damp

    717 300G -12 PSI to 300 PSI, (-850 mbar to20.68 bar, -85 to 2068.4 kPa)

    0.01 psi, 1 mbar Over Pressure 375 PSI,25 bar

    Zero, Min, Max, Hold, Damp

    717 500G 0 PSI to 500 PSI, (0 mbar to34.47 bar, 0 to 3447.4 kPa )

    0.01 psi, 1 mbar Over Pressure 2xFS Zero, Min, Max, Hold, Damp

    717 1000G 0 PSI to 1000 PSI, (0 mbar to68.95 bar, 0 to 6894.8 kPa )

    0.1 psi, 1 mbar Over Pressure 2xFS Zero, Min, Max, Hold, Damp

    717 1500G 0 PSI to 1500 PSI, (0 mbar to103.42 bar, 0 to 10342 kPa)

    0.1 psi, 0.01 bar Over Pressure 2xFS Zero, Min, Max, Hold, Damp

    717 3000G 0 PSI to 3000 PSI, (0 mbar to206.84 bar, 0 to 20684 kPa) 0.1 psi, 0.01 bar Over Pressure 2xFS Zero, Min, Max, Hold, Damp

    717 5000G 0 PSI to 5000 PSI, (0 mbar to344.74 bar, 0 to 34474 kPa)

    0.1 psi, 0.01 bar Over Pressure 2xFS Zero, Min, Max, Hold, Damp

    717 10000G 0 PSI to 10,000 PSI, (0 mbar to700 bar, 0 to 70,000 kPa)

    0.1 psi, 0.01 bar Over Pressure 2xFS Zero, Min, Max, Hold, Damp

    Intrinsic Safe Version Available

  • 7/31/2019 Calibradores Fluke Presion

    4/9

    4 Fluke Corporation Pressure Calibration

    HandPump

    PressureModule

    754 DOCUMENTINGPROCESS CALIBRATOR

    HL

    DifferentialModule

    +IsolationValve

    754DOCUMENTING PROCESS CALIBRATOR

    754DOCUMENTINGPROCESS CALIBRATOR

    Hand Pump

    +

    Pressure Module

    Red

    Black

    IsolationValve

    PressureModule

    V mA

    LOOP

    100%

    25%

    25%

    RECALL

    C /F

    MEAS

    SOURCE

    STORESETUP

    TC RTD

    V mA

    PROCESS CALIBRATOR

    MULTIFUNCTION725

    0%

    754DOCUMENTINGPROCESS CALIBRATOR

    +

    Pressure

    Module

    Red

    Black

    Dierential measurementsDierential pressure modules are useul in a wide varietyo applications, e.g., measuring the uid level in a tankor calibrating a dierential pressure transmitter. A Fluke754 Documenting Process Calibrator is shown.

    I to P device calibrationThe I to P device is used to convert 4 mA to 20 mAelectrical analog loop control to pneumatic analog loopcontrol, generally 3 psi to 15 psi. Here, a typical confgu-ration or using a pressure module with a 750 Series DPCis demonstrated.

    Sourcing pressureTo calibrate an instrument with pressure input, pressurerom an external source (such as a hand-held pump) isapplied. Prompts on the 750 Series Calibrator displayindicate when to increase or decrease the input pressure,and when the specifed test points are achieved. Here, aFluke 754 Documenting Process Calibrator is shown.

    Measuring pressureTo measure pressure, the appropriate pressure moduleor the pressure to be tested is attached to the calibrator.The measured pressure can be displayed in a variety oengineering units. A Fluke 725 or 726 multiunction pro-cess calibrator could be used here.

    P to I device calibrationThe P to I device is used to convert pneumatic analogloop control signals o 3 psi to 15 psi to electrical loopanalog control signals o 4 mA to 20 mA. Here, a

    Fluke 717 Pressure Calibrator is used.

    Pressure switch calibrationVeriy and document the setpoint and deadband o pressureswitches using the 750 Series Documenting Process Calibrators.

  • 7/31/2019 Calibradores Fluke Presion

    5/9

    5 Fluke Corporation Pressure Calibration

    A complete amily opressure modules

    A amily o 29 pressure modulescovers the most common pressurecalibrations rom 0-1 H

    2O

    (0-.25 kPa) to 0-10,000 psi(0-70,000 kPa).

    Gage pressure modules haveone pressure ftting and mea-sure the process pressure withrespect to atmospheric pressure.Dierential pressure moduleshave two pressure fttings andmeasure the dierence betweenthe applied pressure on the highftting versus the low ftting.

    Each module is clearly labeled orrange, overpressure specifcation,and media compatibility. A metricadapter is included with all butthe P29 through P31 high pres-sure modules.

    Quick and easymeasurementsFluke 700 Series pressuremodules are easy to operate. Tomeasure pressure, the technicianplumbs the pressure module toa pressure source, and connectsthe pressure module cable tothe calibrator. Pressure is ap-plied, measured by the pressuremodule, and displayed digitallyon the calibrator. At the touch oa button, the pressure may bedisplayed in up to 11 dierentengineering units. When usedwith the 750 Series Document-ing Process Calibrators, pres-sure readings can be date/timestamped and stored electronicallyor later retrieval. This saves time,eliminates errors, and supportscompliance with quality stan-

    dards and regulations.

    Pressure moduleperormanceFluke 700 Series pressure mod-ules are highly accurate, with

    total specifcations that applyrom 0 % to 100 % o ull spanand rom 0 C to 50 C (32 F to122 F)a eature that sets themapart rom other pressure cali-brators. Many ranges have totaluncertainties o 0.05 % o ullscale and reerence uncertaintieso 0.025 % o scale (see Table,page 5).

    This perormance is possiblethrough the innovative applica-tion o mathematics and micro-processor power. Fluke pressure

    modules have silicon piezo-resistor sensors which consisto a resistive bridge abricatedin a silicon diaphragm. Pressureapplied to the diaphragm causesa change in the balance o thebridge which is proportional tothe applied pressure. The bridgebalance change is not linear andis very sensitive to temperature.However, since these eectsare quite stable with time andthrough repetitive changes ocondition, the sensors can be

    very accurate in measuring pres-sure provided they are careullycharacterized.

    During manuacture, Flukepressure modules are character-ized by reading temperature andpressure at more than 100 points.A least-squares regression is usedto calculate the coefcients o apolynomial expression or pres-sure. The coefcients, unique tothat pressure module, are storedin the modules memory.

    Each module has its own

    microprocessor, allowing it to runthe measurement circuitry andto communicate digitally with acalibrator. When connected tothe calibrator, the modules coe-fcients are uploaded rom thepressure module to the calibrator.Then, as pressure measurementsare made, raw sensor values orpressure and temperature aredigitally loaded to the calibrator,where the raw sensor values andcoefcients are manipulated toderive and display the pressure

    reading.

    This innovative technique pro-vides several benefts:1. Digital communication elimi-

    nates errors due to poor

    connections and electricalintererence.

    2. The modules are inherentlytemperature-compensated rom0 C to 50 C (32 F to 122 F).

    3. The modules are ully inter-changeable because allmeasurements are completedin the pressure module itseland then communicated tothe calibrator in digitizedorm. Modules are calibratedindependently o the calibrator,and can be used with any 700

    Series calibrator. Each modulehas its own serial number toacilitate traceability.

    Sensor protection inisolated modulesMany o these modules (seeTable) incorporate a stainlesssteel diaphragm to isolate thesensor. With these modules, anymedium that is compatible withstainless steel can be used on thehigh side o the module.

    Rugged constructionA urethane overmolding protectsagainst shock i a module is ac-cidentally dropped and also sealsagainst dirt, dust, and moisture.Pressure connections are 1/4NPT. A BSP/ISO adapter is alsoprovided on all but the P29, P30and P31.

    Convenient setupA one-meter cable between thepressure module and calibratorreduces the length o connect-ing tubing to the pressure source.

    The remote pressure head alsoprovides an extra margin o saetyand convenience by removing thecalibrator and operator rom thepressure source.

    Pressure Modules

  • 7/31/2019 Calibradores Fluke Presion

    6/9

    6 Fluke Corporation Pressure Calibration

    Pressure module specifcations (all specifcations in % o ull span. Specifcations reect a confdence interval o 95 %.)

    Pressure Performance

    Summary calibrator specifcations: (one year, 18 C to 28 C)

    1 Supported Pressure Units on 701 and 702 include: psi, kPa, bar, inches Hg, mm Hg, inches H2O (@ 4 C), and eet H

    2O (@4C) Supported Pressure Units on 753, and 754 include all o the above,

    plus kg/cm2, inches H2O (@ 60F), mm H

    2O (@ 4 C). Supported Pressure Units on 717, 718, and 725 include psi, kPa, bar, mbar, kg/cm2, inches Hg, mm Hg, inches H

    2O (@ 4C), inches H

    2O

    (@ 20 C), cm H2O (@ 4C), and cm H

    2O (@ 20C).

    2 For Pressure Module specifcations, see table below.3 Gasses only or 718 and 719.

    1 Total uncertainty, oneyear or temperaturerange 0 C to +50 C.Total uncertainty, 1.0% o ull span or temp-erature range -10 Cto 0C. For P00 moduleonly, compensated

    temperature range is15 C to 35 C.

    2 Dry indicates dryair or non-corrosivegas as compatiblemedia. 316 SS in-dicates media compat-ible with Type 316Stainless Steel.C276 indicatesmedia compatiblewith Hastelloy C276.

    Use o pressure zero isrequired prior to mea-surement or source.Maximum overpressurespecifcation includescommon mode pres-sure. Modules arerated. Metric adapter(s):

    1/4 NPTemale to male BSP/ISO 1/4-19, taperedthread, included withall modules exceptP29, P30, and P31.Eective October 1996,all modules include aNIST traceable certif-cate and test data.

    *Intrinsically saeversion available oruse with the 718Exand 725Ex.

    Reerence High2 Low2 Max over-Range/ Range (approx)/ uncertainty Stability Temperature Total1 side side Fitting pressure

    Model Resolution Resolution (23 3 C) (1 year) (0 to 50 C) uncertainty media media material (x nominal)Dierential

    FLUKE-700P00 1 in. H2O/0.001 0.25 kPa/0.0002 0.300 0.025 0.025 0.350 Dry Dry 316 SS 30xFLUKE-700P01* 10 in. H2O/0.01 2.5 kPa/0.002 0.200 0.050 0.050 0.300 Dry Dry 316 SS 3xFLUKE-700P02 1 psi/0.0001 6900 Pa/0.7 0.150 0.070 0.080 0.300 Dry Dry 316 SS 3xFLUKE-700P22 1 psi/0.0001 6900 Pa/0.7 0.100 0.020 0.030 0.150 316 SS Dry 316 SS 3xFLUKE-700P03 5 psi/0.0001 34 kPa/0.001 0.050 0.020 0.030 0.100 Dry Dry 316 SS 3xFLUKE-700P23 5 psi/0.0001 34 kPa/0.001 0.025 0.010 0.015 0.050 316 SS Dry 316 SS 3xFLUKE-700P04 15 psi/0.001 103 kPa/0.01 0.025 0.010 0.015 0.050 Dry Dry 316 SS 3xFLUKE-700P24* 15 psi/0.001 103 kPa/0.01 0.025 0.010 0.015 0.050 316 SS Dry 316 SS 3xGage

    FLUKE-700P05* 30 psi/0.001 207 kPa/0.01 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3xFLUKE-700P06* 100 psi/0.01 690 kPa/0.07 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3xFLUKE-700P27* 300 psi / 0.01 2070 kPa / 0.1 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3xFLUKE-700P07 500 psi/0.01 3400 kPa/0.1 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3xFLUKE-700P08 1000 psi/0.1 6900 kPa/0.7 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3xFLUKE-700P09* 1500 psi/0.1 10 MPa/0.001 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 2xAbsolute (not compatible with Fluke 701 or 702)

    FLUKE-700PA3 5 psi/0.0001 34 kPa/0.001 0.050 0.010 0.010 0.070 316 SS N/A 316 SS 3xFLUKE-700PA4* 15 psi/0.001 103 kPa/0.01 0.050 0.010 0.010 0.070 316 SS N/A 316 SS 3x

    FLUKE-700PA5 30 psi/0.001 207 kPa/0.01 0.050 0.010 0.010 0.070 316 SS N/A 316 SS 3xFLUKE-700PA6 100 psi/0.01 690 kPa/0.07 0.050 0.010 0.010 0.070 316 SS N/A 316 SS 3xVacuum (not compatible with Fluke 701 or 702)

    FLUKE-700PV3 -5 psi/0.0001 -34 kPa/0.001 0.040 0.015 0.015 0.070 316 SS Dry 316 SS 3xFLUKE-700PV4 -15 psi/0.001 -103 kPa/0.01 0.040 0.015 0.015 0.070 316 SS Dry 316 SS 3xDual

    FLUKE-700PD2 1 psi/0.0001 6900 Pa/0.7 0.150 0.025 0.025 0.200 316 SS Dry 316 SS 3xFLUKE-700PD3 5 psi/0.0001 34 kPa/0.001 0.040 0.015 0.015 0.070 316 SS Dry 316 SS 3xFLUKE-700PD4 15 psi/0.001 103 kPa/0.01 0.025 0.010 0.015 0.050 316 SS Dry 316 SS 3xFLUKE-700PD5 -15/30 psi/0.001 -100/207 kPa/0.01 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3xFLUKE-700PD6 -15/100 psi/0.01 -100/690 kPa/0.07 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3xFLUKE-700PD7 -15/200 psi/0.01 -100/1380 kPa/0.1 0.040 0.015 0.015 0.070 316 SS N/A 316 SS 3xHigh

    FLUKE-700P29* 3000 psi/0.1 20.7 M Pa/0.001 0.050 0.010 0.020 0.080 C276 N/A C276 2xFLUKE-700P30 5000 psi/0.1 34 M Pa/0.001 0.050 0.010 0.020 0.080 C276 N/A C276 2xFLUKE-700P31 10000 psi/1 69 M Pa/0.007 0.050 0.010 0.020 0.080 C276 N/A C276 1.5x

    71

    7

    71

    8/719

    72

    5/726

    75

    3

    75

    4

    55

    20A

    Function Range Resolution Accuracy Notes

    Measure Pressure1 1 psi to 10000 psi 0.0001 psi to 0.05 % ull scale Gasses/liquids 3 (non(internal sensor) see table 0.01 psi corrosive) Zero, Min,

    see table Max, Hold, Damp

    Measure Pressure1 1 psi to 300 psi 0.0001 psi to 0.05 % ull scale (718) Gasses (noncorrosive)

    (internal sensor) see table 0.01 psi 0.025 % ull scale (719) Zero, Min, Max, Hold,100 psi (719) see table Damp

    Measure Pressure1 29 Pressure Modules, To 0.0001 psi, per To 0.025 % o ull span, Media compatibility per(with Pressure Modules) 1 in. H

    2O / 0.25 kPa Pressure Module per Pressure Module Pressure Module specs2

    Over pressureper pressure to specs2 specs2 Zero, Min, Max, Hold,module specs2 10,000 psi / 69 MPa Damp

    Source pressure -12 psig to ull scale N/A N/A Pressure or vacuum,built-in pump overpressure protected

    Measure mA 0 to 24 mA 0.001 mA 0.05 % reading + 1 count 719 Source mA

    Measure mA 0 to 24 mA 0.001 mA 0.020 % reading + 2 count

    Measure mA 0 to 24 mA 0.001 mA 0.010 % reading + 0.015 %o ull scale

    Loop power supply 24 V dc N/A + 10 %

  • 7/31/2019 Calibradores Fluke Presion

    7/9

    7 Fluke Corporation Pressure Calibration

    Fluke 700PTP-1Pneumatic Test Pump

    Pressure Accessories

    Fluke 700HTP-1Hydraulic Test Pump

    Fluke 700LTP-1Low-Pressure Test Pump

    For use with: Fluke 700 SeriesPressure Modules and the Fluke710 Series Pressure Calibrators.

    Description: The Fluke 700PTP-1is a handheld pressure pumpdesigned to generate eithervacuum to -13 psi/-0.9 bar orpressure to 600 psi/40 bar.

    The Fluke 700PTP-1 has twopressure ports: 3/8-BSP (ISO228) emale

    parallel thread ftting or thereerence gauge or pressuremodule

    1/8-BSP (ISO228) emaleparallel thread ftting or the

    unit under testApplication: The Fluke 700PTP-1 eatures an integral pressureadjustment vernier which variesthe pressurized volume by 2.0cc over approximately eleventurns o the vernier knob. Thepressure variation achievablewith the vernier will depend onthe nominal pressure and totalpressurized volume, but with aminimum volume and maximumpressure, the vernier provided600 20 psi adjustment range.

    With a minimum volume and nopressure applied, the vernier canalso be used to provide a 0 to70 H2O range. Larger volumeswill provide a smaller range oadjustment, but greater resolu-tion. The length o the stroke canbe adjusted to limit the maximumoutput pressure. Maximum outputpressure is adjustable rom2.5 psi to 600 psi.

    For use with: Fluke 700 SeriesPressure Modules and the Fluke710 Series Pressure Calibrators.

    Description: The Fluke 700HTP-1 is designed to generate pres-sures up to 10,000 psi/700 bar.The Fluke 700HTP-1 has twopressure ports: 3/8-BSP (ISO228) emale

    parallel thread ftting or thereerence gauge or pressuremodule

    1/8-BSP (ISO228) emale par-allel thread ftting or the unitunder test

    Note: The user must provide a hose withappropriate end fttings rom this port to

    the unit under test.Application: This pump canprovide up to 10,000 psi usingdistilled water or mineral-basedhydraulic oil. The pump is operat-ed by pumping several strokes toprime the system, then switchingto high pressure mode when theresistance increases. An integralpressure adjustment vernier knobvaries the pressurized volume by0.6 cc. The pressure variationachievable with the vernier willdepend on the nominal pressure

    and total pressurized volume, butwith a minimum volume, thevernier provided 150-3,000 psi(at 150 psi nominal) and 3,000-10,000 psi (at 3,000 psi nomi-nal) adjustment ranges. With aminimum volume and no pressureapplied, the vernier can also beused to provide a 0 to 1.7 psirange. Larger volumes will pro-vide a smaller range o adjust-ment, but greater resolution.

    Description: The Fluke 700LTP-1is a hand operated pressurepump designed to generate either

    vacuum to -12 psi/-.85 bar orpressures to 100 psi/6.9 bar. TheFluke 700LTP-1 has twopressure ports with push ftconnectors. These push ft con-nectors, one or the reerenceport or connection to a Fluke700 series pressure module andone to connect to a unit undertest, connect to the suppliedtest hoses. These test hoses areterminated with 1/4-BSP (ISO228)emale parallel thread fttings thatcan be adapted using the fttings

    included.Application: The Fluke 700LTP-1is primarily intended or lowpressure applications. It eatures afne adjust vernier with.00145 / PSI resolution at lowpressures. The pressure variationachievable with the vernier willdepend on the nominal pressureand total pressurized volume,but with minimum volume andmaximum pressure the vernierprovides 30 psi 6 psi. Theadjustable pressure relie valve

    eatures a slow-bleed capabilitythat allows the user to slowly re-lease pressure at a controlled rateto achieve a desired pressure.

  • 7/31/2019 Calibradores Fluke Presion

    8/9

    8 Fluke Corporation Pressure Calibration

    Fluke 700 PCKCalibration KitThe Fluke 700PCK PressureCalibration Kit makes it pos-sible to calibrate your pressuremodules at your acility usingyour own precision pressurestandards. The kit consists oa power supply, an interaceadapter, appropriate cables,and Fluke 700PC Pressure

    Module Calibration sotware.When installed on your PC,the Windows-based sotwareeasily steps you through anas-ound verifcation, a calibra-tion adjustment, and an as-let verifcation. Calibrationdata is captured or import toyour database. A 386 or betterPC, running Windows 3.1, orlater is required, along with aprecision pressure standardwith an uncertainty o lessthan 1/4 that o the pressure

    module being verifed.

    For use with: Fluke 717, 718,and 719 Pressure Calibrators tohelp isolate the calibrator romliquid contamination.

    Description: Use to isolate cali-brator rom contact with liquids.Input is 1/8 NPT emale. Output is1/8 NPT male.

    Application: The Fluke 700ILFcan be used to isolate the calibra-tor rom incidental contact withuids present in the gas mediumbeing measured. Particularlyuseul with the 718/719 calibra-tor to help keep moisture or oilsrom contaminating the on-boardpump. The 1/8 NPT male thread-ed output threads directly intothe 717, 718, and 719 calibra-tors. Maximum pressure is 100psi. Burst pressure is 375 psi.

    Maximum ow is 10 SCFM andfltration to 1 micron.

    Fluke 700HTH-1Hydraulic Test Hose

    Fluke 700ILFIn-line Filter

    12345678

    Pressure

    Calibrator

    Interface

    Unit

    PC Serial Port

    Line Power

    Pressure

    Module

    Note:

    With a Fluke 700PCK and any Fluke Pressure Module, aFluke 5520A Calibrator becomes a precision pressure standard.

    Description: The Fluke700HTH-1 hydraulic test hose isa 10,000 psi, 700 bar workingpressure test hose. The hose usessel-sealing fttings with easyfnger tight connections.

    Application:The Fluke700HTH-1 allows connectionsto a unit under test rom a Fluke700HTP-1 hydraulic test pumpin use with the Fluke 700 seriespressure modules. The 700HTH-1is compatible with water andnon-corrosive oil.

    Fluke 700PRV-1 PressureRelie Valve KitFor use with: Fluke 700HTP-1Hydraulic Test Pump.

    Description: The Fluke 700PRV-1consists o two relie valves (1360

    and 5450 psi) to be used with the700HTP-1 Hydraulic Test Pump.These relie valves will protectthe Fluke pressure modules romdamage due to over-pressuriza-tion. 1/4 BSP male parallel threadto ft Fluke 700HTP-1.

    Application: Repeatability 10 % o nominal setting. Multi-turn adjustment screw to setpreload on internal disc springs.

    Fluke 71X Hose KitFor use with: Fluke 718 and 719Pressure Calibrators.

    Description: The Fluke 71X hosekit includes (2) quick disconnectfttings to connect to the 718 or719, three 1 meter translucenthoses and one BSP adapter.

    Fluke 700PMPPressure Pump

    For use with: Fluke 700 SeriesPressure Modules and the Fluke710 Series Pressure Calibrators.

    Description: The Fluke 700PMPis a hand-operated pressurepump to provide pressures up to150 psi/1000 kPa. Output fttingis 1/8 FNPT.

    Application: Linear strokeo 1.6 (4 cm). Multi-turn vernieror fne adjustment o pressure.Includes controlled pressurebleed valve.

  • 7/31/2019 Calibradores Fluke Presion

    9/9

    9 Fluke Corporation Pressure Calibration

    CERTIF

    IEDTOMEET ISO

    900

    1

    QUAL

    ITYMAN

    AGEMENTSYSTEM

    ISO 9001

    PSID pounds per square inchdierential.

    PSIG pounds per square inchgage (same as psi).

    Square root extractor aninstrument or sotware programthat takes the square root o inputand puts the result on its output.Square root extraction is neededto linearize many ow signals.Example: orifce plates, ventu-ris, target ow meters, and pitottubes all require the transmittersoutput signal to be linearized.Mag ow meters, turbine owmeters, Doppler ow meters, andvortex shedding ow meters dont

    require square root extraction.Static pressure the zero-velocity pressure at any arbitrarypoint within a system.

    Wet/dry dierential a di-erential pressure transduceror transmitter that uses a metaldiaphragm at the wet port whereuids can be applied, and no dia-phragm at the dry port. The dryport exposes the sensor materialto the medium, so only clean drygas can be applied to this port.

    Wetted parts the diaphragmand pressure port material thatcomes in direct contact with themedium (gas, liquid).

    Gage pressure transducer a transducer that measurespressure relative to atmosphericpressure.

    Ideal Gas Law combiningBoyles Law and Charles Law,results in the Ideal Gas Law:PV=nRT, where nR is constantor a particular gas analogous tothe number o molecules and therelative size o the molecule.

    I/P (I to P) a current to pres-sure transmitter. A commoninstrument in modern industrialplants. A typical large paper millor refnery could have 5,000 I/Psin use.

    Line pressure the maximumpressure in the pressure vesselor pipe or dierential pressuremeasurement.

    Orifce plate a very low costand common primary sensing el-ement (PSE) or measuring ow. Itmust be used in conjunction witha d/p cell. It creates a venturi anda resulting P is developed acrossthe plate whose square root isproportional to ow.

    P/I (P to I) a pressure to cur-

    rent transducer.

    Pneumatic relay reers to apneumatic instrument that per-orms a unction to its input andprovides the result on its output(Example: square root extractor,adder, etc.).

    PSI pounds per square inch(same as psig).

    PSIA pounds per square inchabsolute.

    Absolute pressure absolutepressure measurements arereerenced to zero pressure,(a perect vacuum.)

    Absolute pressuretransducer a transducer thathas an internal reerencechamber sealed at or close to zeropressure (ull vacuum)when exposed to atmospherea reading o approximately14.7 psi results.

    Boyles Law the volume o agas is inversely proportionalto the pressure o the gas atconstant temperature: V=1/P.

    Charles Law essentially

    states or a fxed volume o gas,i the temperature is raised, thepressure will increase.P = Constant x T.

    Common mode pressure theunderlying common pressure(or static pressure) within asystem rom which a dierentialmeasurement is being made.

    D/P: Dierential pressure,(pronounced DP) other namesused to mean the same thingare d/p cell, d/p transmitter and

    P transmitter (where is deltaor dierential). This is the mostcommon type o transmitter usedin most process industries. It canbe used to measure level, ow,pressure, dierential pressure,and density or specifc gravity.With some modifcations, it canmeasure such things as tempera-ture and oxygen purity. The d/ptransmitter can be pneumatic,electromechanical, or solid state.It can also be a smart transmitter.A typical large process plant can

    have hundreds or thousands od/p transmitters in service.

    Gage pressure the pressurerelative to atmospheric pressure.Gage pressure = absolute pres-sure minus one atmosphere.

    Pressure Terminology

    Fluke Corporation

    PO Box 9090, Everett, WA 98206 U.S.A.Fluke Europe B.V.PO Box 1186, 5602 BDEindhoven, The Netherlands

    For more information call:In the U.S.A. (800) 443-5853 orFax (425) 446-5116In Europe/M-East/Arica +31 (0) 40 2675 200 orFax +31 (0) 40 2675 222In Canada (800)-36-FLUKE orFax (905) 890-6866From other countries +1 (425) 446-5500 orFax +1 (425) 446-5116

    Web access: http://www.luke.com

    2006-2011 Fluke Corporation.Specifcations subject to change without notice.Printed in U.S.A. 7/2011 1281620J D-EN-N

    Modification of this document is not permittedwithout written permission from Fluke Corporation.

    Fluke. Keeping your worldup and running.