By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

36
Anesthetic Gas Modules By Pioneer Biomedical

Transcript of By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Page 1: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Anesthetic Gas ModulesBy Pioneer Biomedical

Page 2: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

HP &Philips

Page 3: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

GE

Page 4: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Other monitor equipment

Page 5: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Introduction

Thank you KAMI for letting us host your meeting!

Mike CorumDan ChapmanBrendan Healy

Page 6: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

www.AGM-parts.comDedicate website for Anesthetic Gas Monitoring, AGM-Parts.com

Page 7: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

M1026A support

Page 8: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

M1026A & M1026B

Page 9: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

M1026A & M1026B

Page 10: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Inside look

Page 11: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

M1026A Insides

Page 12: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.
Page 13: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.
Page 14: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Flow Diagram

Page 15: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Major Problems

Page 16: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Problem Hierarchy

Page 17: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Unit without errors

Page 18: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Common Problem #1

Page 19: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Agent ID replacement

Page 20: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Common Problem # 2

Page 21: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Common Problem # 3

Page 22: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Other IR Bench errors

Measurement Optic Path shows failure and Main board failure

Noisy Baseline on CO2, N2O, or AgentGas Analyzer Equipment Malfunction or

communication problem

Page 23: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Replacing the IR Bench

Page 24: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Transferring Data

Page 25: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Check out procedure

1 Perform Mechanical/Visual Checks  2 Check Error Contents 3 Check System Ventilator Fan Operation 4 Perform Pump Test 5 Replace Internal Nafion Tubing 6 Replace Manifold Seals 7 Replace Room Air Filter 8 Replace Pump Filter 9 Pump Hours: 10 Check Pneumatic System Leakage 11 Check/Calibrate Pneumatic System Flow rates and set (150 and 310 ml/min) 12 Perform Zero Calibration 13 Perform Barometric Pressure Check/Calibration 14 Perform another Zero Calibration 15 Perform Span Calibration Check on all gas channels and agents 16 Repeat Zero Calibration, check displayed values 17 Perform Electrical Safety Check

Page 26: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Quality assurance- what customers get from us

Page 27: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Checking Nafion Tubing

Page 28: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Leak Check

Page 29: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Checking flows

Need:Flow meterTubing

Page 30: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Adjusting flows

Normal Flow Adjustment

Purge Flow Adjustment

Page 31: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Barometric Pressure Check

Page 32: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

SPAN calibration of all gases

Page 33: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

GE

Page 34: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Other monitor equipment

Page 35: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

HP &Philips

Page 36: By Pioneer Biomedical. HP &Philips GE Other monitor equipment.

Conclusion

For parts, tech help, PM kits, Pioneer Biomedical 1-888-644-3274

Air Liquide, 1-215-766-7407, x1235, for combined M1026A & M1026B cal gas

Part # 47068SR 3% isoflurane, 5% C02, 40% Nitrous Oxide, Bal O2 works for both M1026A and M1026B