Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT ALBUMIN CATÁLOG BT 10.002.00 METODOLOGY COLORIMETRIC DETERMINATIONS 250
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Albumin UNITS: g/dL CALCULATION MODE Final Point REACTION CURVE Increase READING MODE Monocromatic READING FILTER 620 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 2 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 3 CALIBRATION CALIBRATOR NUMBER OF CALIBRATORS 1 CONCENTRATION @ CALIBRATION REPLICATES 2 LINEARITY LIMIT 6 BLANK LIMIT 1.000 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT ALPHA-AMYLASE CATÁLOG BT 11.001.00 METODOLOGY KINETIC COLORIMETRICDETERMINATIONS 60
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Amylase UNITS: U/L CALCULATION MODE Delta kinetic with Factor REACTION CURVE Increase READING MODE Monocromatic READING FILTER 405 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 INCUBATION TIME 30 INTERVAL TIME 60 NUMBER OF INTERVALS 3 FACTOR 6950 LINEARITY LIMIT 1038 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT ALKALINE PHOSPHATASE
CATÁLOG BT 11.005.00 METODOLOGY KINETIC COLORIMETRICDETERMINATIONS 50
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Alkaline Phosphatase UNITS: U/L CALCULATION MODE Delta kinetic with FactorREACTION CURVE Increase READING MODE Monocromatic READING FILTER 405 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 INCUBATION TIME 30 INTERVAL TIME 60 NUMBER OF INTERVALS 3 FACTOR 2757 LINEARITY LIMIT 700 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE
*
MAXIMUM NORMAL VALUE
*
CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT ALT/TGP CATÁLOG BT 11.008.00 METODOLOGY KINETIC UV DETERMINATIONS 250
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: ALT/TGP UNITS: U/L CALCULATION MODE Delta kinetic with FactorREACTION CURVE Decrease READING MODE Monocromatic READING FILTER 340 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 INCUBATION TIME 30 INTERVAL TIME 60 NUMBER OF INTERVALS 3 FACTOR 3333 LINEARITY LIMIT 350 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE
*
MAXIMUM NORMAL VALUE
*
CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT AST/TGO CATÁLOG BT 11.007.00 METODOLOGY KINETIC UV DETERMINATIONS 250
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: AST/TGO UNITS: U/L CALCULATION MODE Delta kinetic with FactorREACTION CURVE Decrease READING MODE Monocromatic READING FILTER 340 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 INCUBATION TIME 30 INTERVAL TIME 60 NUMBER OF INTERVALS 3 FACTOR 3333 LINEARITY LIMIT 440 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE
*
MAXIMUM NORMAL VALUE
*
CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT BILIRUBIN DIRECT CATÁLOG BT 10.003.00 METODOLOGY COLORIMETRIC DETERMINATIONS 200
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Bili Direct UNITS: mg/dL CALCULATION MODE Differential Mode REACTION CURVE Increase READING MODE Monocromatic READING FILTER 546 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 2 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 3 CALIBRATION CALIBRATOR NUMBER OF CALIBRATORS 1 CONCENTRATION @ CALIBRATION REPLICATES 2 LINEARITY LIMIT 15 BLANK LIMIT 1.000 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT BILIRUBIN TOTAL CATÁLOG BT 10.003.00 METODOLOGY COLORIMETRIC DETERMINATIONS 200
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Bili Total UNITS: mg/dL CALCULATION MODE Differential Mode REACTION CURVE Increase READING MODE Monocromatic READING FILTER 546 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 2 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 3 CALIBRATION CALIBRATOR NUMBER OF CALIBRATORS 1 CONCENTRATION @ CALIBRATION REPLICATES 2 LINEARITY LIMIT 15 BLANK LIMIT 1.000 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT CÁLCIUM ASX CATÁLOG BT 12.002.00 METODOLOGY ARSENAZO III DETERMINATIONS 100
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Calcium ASX UNITS: mg/dL CALCULATION MODE Final Point REACTION CURVE Increase READING MODE Monocromatic READING FILTER 620 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 2 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 3 CALIBRATION CALIBRADOR NUMBER OF CALIBRATORS 1 CONCENTRATION @ CALIBRATION REPLICATES 2 LINEARITY LIMIT 20 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT CÁLCIUM CATÁLOG BT 12.001.00 METODOLOGY CPC DETERMINATIONS 100
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Mix the Reagent A + Reagent B in a proportion 1:1 and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Calcium CPC UNITS: mg/dL CALCULATION MODE Final Point REACTION CURVE Increase READING MODE Monocromatic READING FILTER 578 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 2 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 3 CALIBRATION CALIBRADOR NUMBER OF CALIBRATORS 1 CONCENTRATION @ CALIBRATION REPLICATES 2 LINEARITY LIMIT 20 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT CK NAC CATÁLOG BT 11.002.00 METODOLOGY KINETIC UV DETERMINATIONS 25
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: CK NAC UNITS: U/L CALCULATION MODE Delta kinetic with FactorREACTION CURVE Increase READING MODE Monocromatic READING FILTER 340 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 INCUBATION TIME 180 INTERVAL TIME 60 NUMBER OF INTERVALS 3 FACTOR 8095 LINEARITY LIMIT 1000 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE
*
MAXIMUM NORMAL VALUE
*
CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT CK MB CATÁLOG BT 11.003.00 METODOLOGY KINETIC UV DETERMINATIONS 25
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: CK MB UNITS: U/L CALCULATION MODE Delta kinetic with FactorREACTION CURVE Increase READING MODE Monocromatic READING FILTER 340 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 INCUBATION TIME 180 INTERVAL TIME 60 NUMBER OF INTERVALS 3 FACTOR 6750 LINEARITY LIMIT 600 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE
*
MAXIMUM NORMAL VALUE
*
CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT CHLORIDE CATÁLOG BT 12.003.00 METODOLOGY COLORIMÉTRIC DETERMINATIONS 50
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Chloride UNITS: mEq/L CALCULATION MODE Final Point REACTION CURVE Increase READING MODE Monocromatic READING FILTER 505 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 3 CALIBRATION CALIBRADOR NUMBER OF CALIBRATORS 1 CONCENTRATION @ CALIBRATION REPLICATES 2 LINEARITY LIMIT 150 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT CHOLESTEROL CATÁLOG BT 10.004.00 METODOLOGY ENZIMÁTIC
COLORIMÉTRIC DETERMINATIONS 1000
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Cholesterol UNITS: mg/dL CALCULATION MODE Final Point REACTION CURVE Increase READING MODE Monocromatic READING FILTER 505 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 3 CALIBRATION CALIBRADOR NUMBER OF CALIBRATORS 1 CONCENTRATION @ CALIBRATION REPLICATES 2 LINEARITY LIMIT 800 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT CHOLESTEROL HDL CATÁLOG BT 10.005.00 METODOLOGY PRECIPITATION +
ENZIMATIC COLORIMÉTRIC
DETERMINATIONS 400 Precipitações
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: HDL Cholesterol UNITS: mg/dL CALCULATION MODE Final Point REACTION CURVE Increase READING MODE Monocromatic READING FILTER 505 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 3 CALIBRATION CALIBRADOR NUMBER OF CALIBRATORS 1 CONCENTRATION @ CALIBRATION REPLICATES 2 LINEARITY LIMIT 150 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT CHOLESTEROL HDL DIRECT
CATÁLOG BT 10.006.00 METODOLOGY ENZIMÁTIC
COLORIMÉTRIC DETERMINATIONS 120
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: HDL UNITS: mg/dL CALCULATION MODE Fixed Time with STD REACTION CURVE Increase READING MODE Monocromatic READING FILTER 578 ASP. VOL. 400 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 INCUBATION TIME 30 REACTION TIME 180 CALIBRATION CALIBRATOR NUMBER OF CALIBRATORS
1
CONCENTRATION @ CALIBRATION REPLICATES
2
LINEARITY LIMIT 180 BLANK LIMIT 0.500 MÍNIMUM NORMAL VALUE
*
MAXIMUM NORMAL VALUE
*
CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT CHOLESTEROL LDL DIRECT
CATÁLOG BT 10.015.00 METODOLOGY ENZIMÁTIC
COLORIMÉTRIC DETERMINATIONS 40
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: LDL UNITS: mg/dL CALCULATION MODE Fixed Time with STD REACTION CURVE Increase READING MODE Monocromatic READING FILTER 578 ASP. VOL. 400 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 INCUBATION TIME 30 REACTION TIME 180 CALIBRATION CALIBRATOR NUMBER OF CALIBRATORS
1
CONCENTRATION @ CALIBRATION REPLICATES
2
LINEARITY LIMIT 400 BLANK LIMIT 0.500 MÍNIMUM NORMAL VALUE
*
MAXIMUM NORMAL VALUE
*
CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT CREATININE CATÁLOG BT 10.007.00 METODOLOGY COLORIMÉTRIC DETERMINATIONS 500
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Creatinina UNITS: mg/dL CALCULATION MODE Fixed Time with STD REACTION CURVE Increase READING MODE Monocromatic READING FILTER 505 ASP. VOL. 400 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 INCUBATION TIME 30 REACTION TIME 60 CALIBRATION CALIBRATOR NUMBER OF CALIBRATORS
1
CONCENTRATION @ CALIBRATION REPLICATES
2
LINEARITY LIMIT 12 BLANK LIMIT 1.000 MÍNIMUM NORMAL VALUE
*
MAXIMUM NORMAL VALUE
*
CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT GAMMA GT CATÁLOG BT 11.006.00 METODOLOGY KINÉTIC COLORIMÉTRICDETERMINATIONS 50
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Gama GT UNITS: U/L CALCULATION MODE Delta kinetic with FactorREACTION CURVE Increase READING MODE Monocromatic READING FILTER 405 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 INCUBATION TIME 30 INTERVAL TIME 60 NUMBER OF INTERVALS 3 FACTOR 1158 LINEARITY LIMIT 250 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE
*
MAXIMUM NORMAL VALUE
*
CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT GLUCOSE CATÁLOG BT 10.008.00 METODOLOGY ENZIMÁTIC
COLORIMÉTRIC DETERMINATIONS 1000
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Glucose UNITS: mg/dL CALCULATION MODE Final Point REACTION CURVE Increase READING MODE Monocromatic READING FILTER 505 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 3 CALIBRATION CALIBRADOR NUMBER OF CALIBRATORS 1 CONCENTRATION @ CALIBRATION REPLICATES 2 LINEARITY LIMIT 400 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT IRON FEZ CATÁLOG BT 12.005.00 METODOLOGY FERROZINE DETERMINATIONS 50
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagent 1: Reagent A Reagent 2: Reagent B (Ferrozine) CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Iron FEZ UNITS: µg/dL CALCULATION MODE Differential Mode REACTION CURVE Increase READING MODE Monocromatic READING FILTER 546 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 3 CALIBRATION CALIBRATOR NUMBER OF CALIBRATORS 1 CONCENTRATION @ CALIBRATION REPLICATES 2 LINEARITY LIMIT 500 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT IRON CRX CATÁLOG BT 12.004.00 METODOLOGY COLORIMETRIC DETERMINATIONS 100
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Iron CRX UNITS: µg/dL CALCULATION MODE Final Point REACTION CURVE Increase READING MODE Monocromatic READING FILTER 620 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 3 CALIBRATION CALIBRADOR NUMBER OF CALIBRATORS 1 CONCENTRATION @ CALIBRATION REPLICATES 2 LINEARITY LIMIT 500 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT LACTATE DEHYDROGENASE
CATÁLOG BT 11.004.00 METODOLOGY KINÉTICO UV DETERMINATIONS 50
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: LDH UNITS: U/L CALCULATION MODE Delta kinetic with FactorREACTION CURVE Decrease READING MODE Monocromatic READING FILTER 340 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 INCUBATION TIME 30 INTERVAL TIME 60 NUMBER OF INTERVALS 3 FACTOR 8095 LINEARITY LIMIT 2000 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE
*
MAXIMUM NORMAL VALUE
*
CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT MAGNÉSIUM CATÁLOG BT 12.007.00 METODOLOGY MAGON DETERMINATIONS 50
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Magnesium UNITS: mg/dL CALCULATION MODE Final Point REACTION CURVE Increase READING MODE Monocromatic READING FILTER 505 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 2 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 3 CALIBRATION CALIBRADOR NUMBER OF CALIBRATORS 1 CONCENTRATION @ CALIBRATION REPLICATES 2 LINEARITY LIMIT 4.5 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT PHÓSPHORUS UV CATÁLOG BT 12.006.00 METODOLOGY Molibidate UV DETERMINATIONS 50
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Phosphorus UNITS: mg/dL CALCULATION MODE Final Point REACTION CURVE Increase READING MODE Monocromatic READING FILTER 340 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 1 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 3 CALIBRATION CALIBRADOR NUMBER OF CALIBRATORS 1 CONCENTRATION @ CALIBRATION REPLICATES 2 LINEARITY LIMIT 15 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT PROTEÍN TOTAL CATÁLOG BT 10.009.00 METODOLOGY COLORIMÉTRIC DETERMINATIONS 250
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Protein Total UNITS: g/dL CALCULATION MODE Final Point REACTION CURVE Increase READING MODE Monocromatic READING FILTER 546 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 2 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 3 CALIBRATION CALIBRADOR NUMBER OF CALIBRATORS 1 CONCENTRATION @ CALIBRATION REPLICATES 2 LINEARITY LIMIT 12 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT PROTEÍN URINÁRY CATÁLOG BT 10.016.00 METODOLOGY COLORIMÉTRIC DETERMINATIONS 100
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Protein Urinary UNITS: mg/L CALCULATION MODE Final Point REACTION CURVE Increase READING MODE Monocromatic READING FILTER 578 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 3 CALIBRATION CALIBRADOR NUMBER OF CALIBRATORS 1 CONCENTRATION @ CALIBRATION REPLICATES 2 LINEARITY LIMIT 4000 BLANK LIMIT 2.00 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT TRIGLYCÉRIDES CATÁLOG BT 10.010.00 METODOLOGY ENZYMÁTIC
COLORIMÉTRIC DETERMINATIONS 500
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Triglycerides UNITS: mg/dL CALCULATION MODE Final Point REACTION CURVE Increase READING MODE Monocromatic READING FILTER 505 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 3 CALIBRATION CALIBRADOR NUMBER OF CALIBRATORS 1 CONCENTRATION @ CALIBRATION REPLICATES 2 LINEARITY LIMIT 1000 BLANK LIMIT 1.500 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT URÉA ENZYMÁTIC CATÁLOG BT 10.013.00 METODOLOGY ENZYMATIC
COLORIMÉTRIC DETERMINATIONS 250
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Please follow the instructions CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Urea UNITS: mg/dL CALCULATION MODE Final Point REACTION CURVE Increase READING MODE Monocromatic READING FILTER 578 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 3 CALIBRATION CALIBRADOR NUMBER OF CALIBRATORS 1 CONCENTRATION @ CALIBRATION REPLICATES 2 LINEARITY LIMIT 200 BLANK LIMIT 1.500 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT URÉA UV CATÁLOG BT 10.012.00 METODOLOGY KINÉTICO UV DETERMINATIONS 250
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Urea UNITS: mg/dL CALCULATION MODE Fixed Time with STD REACTION CURVE Decrease READING MODE Monocromatic READING FILTER 340 ASP. VOL. 400 µL TEMPERATURE 37 ºC DECIMALS 0 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 30 CALIBRATION 60 NUMBER OF CALIBRATORS
CALIBRATOR
CONCENTRATION 1 CALIBRATION REPLICATES
@
LINEARITY LIMIT 2 BLANK LIMIT 250 MÍNIMUM NORMAL VALUE
1.000
MAXIMUM NORMAL VALUE
*
CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT URIC ACID - MONOREAGENT
CATÁLOG BT 10.001.00 METODOLOGY ENZYMATIC
COLORIMETRIC DETERMINATIONS 250
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Uric Acid UNITS: mg/dL CALCULATION MODE Final Point REACTION CURVE Increase READING MODE Monocromatic READING FILTER 500 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 2 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 3 CALIBRATION CALIBRADOR NUMBER OF CALIBRATORS 1 CONCENTRATION @ CALIBRATION REPLICATES 2 LINEARITY LIMIT 20 BLANK LIMIT 1.000 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *
Equipment Applications
BTS 330®
® TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT URIC ACID - BIREAGENTCATÁLOG BT 10.001.00 METODOLOGY ENZYMATIC
COLORIMETRIC DETERMINATIONS 250
- The number of determinations was calculated NOT taking into account the “dead volume” of the equipment flask. - It´s recommended to read the Instructions before using this product.
Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test it´s sugested the use of the BioTécnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
NAME: Uric Acid UNITS: mg/dL CALCULATION MODE Final Point REACTION CURVE Increase READING MODE Monocromatic READING FILTER 500 ASP. VOL. 500 µL TEMPERATURE 37 ºC DECIMALS 2 BLANK REPLICATES 1 SAMPLE REPLICATES 1 STABILIZATION TIME 3 CALIBRATION CALIBRADOR NUMBER OF CALIBRATORS 1 CONCENTRATION @ CALIBRATION REPLICATES 2 LINEARITY LIMIT 20 BLANK LIMIT 1.000 MÍNIMUM NORMAL VALUE * MAXIMUM NORMAL VALUE * CONTROL 1 * CONTROL 1 NAME * CONTROL 1 LOT * CONTROL 1 MÍNIMUM * CONTROL 1 MÁXIMUM * CONTROL 2 * CONTROL 2 NAME * CONTROL 2 LOT * CONTROL 2 MÍNIMUM * CONTROL 2 MÁXIMUM *