BS-200 Parameter Setting
(Fixed-time)
WendyApr, 2008
-
7/27/2019 BS 200 Fixed Time
2/42
Main Content
Preface
Basic parameter setting
Reagent setting
Calibration and QC setting
-
7/27/2019 BS 200 Fixed Time
3/42
Preface
Here we set UREA as the example forall Fixed-time items on
Generally, Fixed-time item includes: UREA,Cre,
-
7/27/2019 BS 200 Fixed Time
4/42
Basic parameter for UREA Read the reagent insert first, eg as below.
-
7/27/2019 BS 200 Fixed Time
5/42
Input the item name forTest.
UREA
-
7/27/2019 BS 200 Fixed Time
6/42
No. will be generated automatically by the system, eg 1 for the first one.
UREA
1
-
7/27/2019 BS 200 Fixed Time
7/42
Full Nameof the item, can be blank if the user dont want to input anything.
UREA
1
-
7/27/2019 BS 200 Fixed Time
8/42
Standard No. of the item, can be blank or involved with the bar code.
1
UREA
-
7/27/2019 BS 200 Fixed Time
9/42
Reac. Type of the item, can be referred from the reagent insert as ...
1
UREA
-
7/27/2019 BS 200 Fixed Time
10/42
So we can set it as Fixed-timetype.
Its another saying of Fixed-time.
Basic parameter for UREA
-
7/27/2019 BS 200 Fixed Time
11/42
Reac. Type is set as Fixed-time as the reagent insert indicates.
1
Fixed-time
UREA
-
7/27/2019 BS 200 Fixed Time
12/42
Pri. Wave&Sec. Waveshould be selected as reagent insert indicates
UREA
1
Fixed-time
-
7/27/2019 BS 200 Fixed Time
13/42
So we can set it as 340nm for Pri. Wave; since nosecondary wavelength is mentioned, Sec. Wave could beblank.
Basic parameter for UREA
-
7/27/2019 BS 200 Fixed Time
14/42
Pri. Waveis selected as 340 nm, Sec. Wave is blank.
UREA
1
Fixed-time
340
-
7/27/2019 BS 200 Fixed Time
15/42
Directionof the absorbance changing during the reaction, refer to ...
1
UREA
Fixed-time
340
-
7/27/2019 BS 200 Fixed Time
16/42
So we can find its
Descending
Moreover, From
Reaction Principle,
we can see NADH(routine indicator
under 340nm) is a
to-be-consumed
reactant, which
will be less and
less during testing,
thus, indicating
the absorbance is
descending in the
measurement.
Basic parameter for UREA
-
7/27/2019 BS 200 Fixed Time
17/42
Directionis selected as Descending.
1
Descending
UREA
Fixed-time
340
-
7/27/2019 BS 200 Fixed Time
18/42
Reac. Time& Incuba. Time for the reaction, refer to ...
1
Descending
UREA
Fixed-time
340
-
7/27/2019 BS 200 Fixed Time
19/42
Mono reagent items' reaction time begins when Sample is dispensed.
Dual reagent items reaction time begins when R2 is dispensed.
The time between S and R2 dispensing is called Incubation time,only dual reagent measurements have incubation time.
R1 Sample
Preheating time Reaction time
To be set from
the insert Fixed 14
cycles
Basic parameter for UREA
Incubation
time
0 cycle of
reaction time
-
7/27/2019 BS 200 Fixed Time
20/42
Time is all calculated into cycles on BS-200, and 1 cycleequals to 16 seconds.
Roughly speaking, we change minutes into cycles forReac. Time and Incuba. Time as below.
Basic parameter for UREA
minute cycle
0.5 2
1 4
2 8
3 12
4 16
5 20
-
7/27/2019 BS 200 Fixed Time
21/42
According to the instruction and the minute-cycle conversion table:
The incubation time is be 1 minute, namely 4 cycles;
The reaction time should be 30seconds and 30+60seconds,
namely 2 and 6 cycles.
Basic parameter for UREA
-
7/27/2019 BS 200 Fixed Time
22/42
Reac. Timeis set as 2 and 6, Incuba. Time as 4.
1
2 6
Descending
UREA
Fixed-time
340
4
-
7/27/2019 BS 200 Fixed Time
23/42
Unitof the item, should be set according to the reagent insert
1
2 6
Descending
UREA
Fixed-time
340
4
-
7/27/2019 BS 200 Fixed Time
24/42
According to the insert, we can set the Unit as mg/dl ormmol/l for serum testing, and g/24h or mmol/24h forurine samples.
Basic parameter for UREA
-
7/27/2019 BS 200 Fixed Time
25/42
2 6
4
We can set the Unitas mg/dl.
1
mg/dl
Descending
UREA
Fixed-time
340
1
-
7/27/2019 BS 200 Fixed Time
26/42
2 6
4
Precision of the items results, should be referred to the reagent insert
1
Descending
UREA
Fixed-time
340
mg/dl
-
7/27/2019 BS 200 Fixed Time
27/42
If the unit is set as mg/dl, its normal to report results
without decimals. We can set the Precision as 0.
Basic parameter for UREA
-
7/27/2019 BS 200 Fixed Time
28/42
2 6
4
Precision of the item is set as 0.
1
0
Descending
UREA
Fixed-time
340
mg/dl
-
7/27/2019 BS 200 Fixed Time
29/42
2 6
4
R1, R2 and Sample Volume of the item, should be referred to
1
Descending
UREA
Fixed-time
340
0
mg/dl
-
7/27/2019 BS 200 Fixed Time
30/42
To process with the sample testing, we need 10ul sample, 1000ulR1 and 250ul R2 (namely, the ratio of S/R1/R2 is 1/100/25);
According to the volume range of BS-200 (sample: 3-45ul, R1:
180-450ul; R2: 30-450ul), and the indicated S/R1/R2 ratio;
We can set Sample volume as 3ul, R1 as 300ul, and R2 as 75ul.
Basic parameter for UREA
-
7/27/2019 BS 200 Fixed Time
31/42
2 6
4
R1, R2 and Sample Volume are set as 300, 75 and 3.
3
300
Descending
UREA
Fixed-time
340
0
mg/dl
751
-
7/27/2019 BS 200 Fixed Time
32/42
2 6
4
Absorbance ofR1 Blank (mono rgt.) and Mixed Rgt. Blank (dual rgt.),
set according to experience or reagent insert; if no idea, can be blank.
1
3
300
Descending
UREA
Fixed-time
340
0
mg/dl
75
-
7/27/2019 BS 200 Fixed Time
33/42
2 6
4
1
3
300
Linearity Range means the reagents measuring range, refer to
Descending
UREA
Fixed-time
340
0
mg/dl
75
-
7/27/2019 BS 200 Fixed Time
34/42
Basic parameter for UREA
From the part of Linearity, we can get aware of that, theupper limit of Linearity Range is 300 mg/dl, or 50 mmol/l.
-
7/27/2019 BS 200 Fixed Time
35/42
2 6
4
3
300
We can set the upper limit ofLinearity Range as 300 mg/dl.
300
Descending
UREA
Fixed-time
340
0
mg/dl
75
-
7/27/2019 BS 200 Fixed Time
36/42
2 6
4
1
3
300
Linearity Limit, Substrate Limit and Factorshould be set for the
Kinetics tests, so here we can leave them as blank.
Descending
UREA
Fixed-time
340
0
mg/dl
75
300
-
7/27/2019 BS 200 Fixed Time
37/42
2 6
4
1
3
300
Prozone check is set for the turbidimetric tests, which can also be blank.
Descending
UREA
Fixed-time
340
0
mg/dl
75
300
-
7/27/2019 BS 200 Fixed Time
38/42
The basic parameters for UREA havebeen finished
Next, we should set the reagentinformation as following
Reagent parameter for UREA
-
7/27/2019 BS 200 Fixed Time
39/42
-
7/27/2019 BS 200 Fixed Time
40/42
-
7/27/2019 BS 200 Fixed Time
41/42
Basic reagent information (Position andBottle Type) has been set up
We should set relevant calibration and QC
information for the item
(Please refer to the video of Calibration & QC setting)
Calibration and QC for UREA
http://images.google.cn/imgres?imgurl=http://news.xinhuanet.com/auto/2006-09/01/xinsrc_06209030108183282472421.jpg&imgrefurl=http://news.xinhuanet.com/auto/2006-09/01/content_5033813.htm&h=295&w=400&sz=18&hl=zh-CN&start=15&um=1&tbnid=qOsPXSyElbEfgM:&tbnh=91&tbnw=124&prev=/images%3Fq%3D%25E7%259C%258B%25E7%2594%25B5%25E5%25BD%25B1%26um%3D1%26complete%3D1%26hl%3Dzh-CN%26newwindow%3D1%26rls%3Dcom.microsoft:zh-cn:IE-SearchBox%26rlz%3D1I7SUNA -
7/27/2019 BS 200 Fixed Time
42/42
Thank you!
NOTE:Fixed-time items should be set in the same way as UREA.