BS 200 Fixed Time

download BS 200 Fixed Time

of 42

Transcript of BS 200 Fixed Time

  • 7/27/2019 BS 200 Fixed Time

    1/42

    BS-200 Parameter Setting

    (Fixed-time)

    WendyApr, 2008

  • 7/27/2019 BS 200 Fixed Time

    2/42

    Main Content

    Preface

    Basic parameter setting

    Reagent setting

    Calibration and QC setting

  • 7/27/2019 BS 200 Fixed Time

    3/42

    Preface

    Here we set UREA as the example forall Fixed-time items on

    Generally, Fixed-time item includes: UREA,Cre,

  • 7/27/2019 BS 200 Fixed Time

    4/42

    Basic parameter for UREA Read the reagent insert first, eg as below.

  • 7/27/2019 BS 200 Fixed Time

    5/42

    Input the item name forTest.

    UREA

  • 7/27/2019 BS 200 Fixed Time

    6/42

    No. will be generated automatically by the system, eg 1 for the first one.

    UREA

    1

  • 7/27/2019 BS 200 Fixed Time

    7/42

    Full Nameof the item, can be blank if the user dont want to input anything.

    UREA

    1

  • 7/27/2019 BS 200 Fixed Time

    8/42

    Standard No. of the item, can be blank or involved with the bar code.

    1

    UREA

  • 7/27/2019 BS 200 Fixed Time

    9/42

    Reac. Type of the item, can be referred from the reagent insert as ...

    1

    UREA

  • 7/27/2019 BS 200 Fixed Time

    10/42

    So we can set it as Fixed-timetype.

    Its another saying of Fixed-time.

    Basic parameter for UREA

  • 7/27/2019 BS 200 Fixed Time

    11/42

    Reac. Type is set as Fixed-time as the reagent insert indicates.

    1

    Fixed-time

    UREA

  • 7/27/2019 BS 200 Fixed Time

    12/42

    Pri. Wave&Sec. Waveshould be selected as reagent insert indicates

    UREA

    1

    Fixed-time

  • 7/27/2019 BS 200 Fixed Time

    13/42

    So we can set it as 340nm for Pri. Wave; since nosecondary wavelength is mentioned, Sec. Wave could beblank.

    Basic parameter for UREA

  • 7/27/2019 BS 200 Fixed Time

    14/42

    Pri. Waveis selected as 340 nm, Sec. Wave is blank.

    UREA

    1

    Fixed-time

    340

  • 7/27/2019 BS 200 Fixed Time

    15/42

    Directionof the absorbance changing during the reaction, refer to ...

    1

    UREA

    Fixed-time

    340

  • 7/27/2019 BS 200 Fixed Time

    16/42

    So we can find its

    Descending

    Moreover, From

    Reaction Principle,

    we can see NADH(routine indicator

    under 340nm) is a

    to-be-consumed

    reactant, which

    will be less and

    less during testing,

    thus, indicating

    the absorbance is

    descending in the

    measurement.

    Basic parameter for UREA

  • 7/27/2019 BS 200 Fixed Time

    17/42

    Directionis selected as Descending.

    1

    Descending

    UREA

    Fixed-time

    340

  • 7/27/2019 BS 200 Fixed Time

    18/42

    Reac. Time& Incuba. Time for the reaction, refer to ...

    1

    Descending

    UREA

    Fixed-time

    340

  • 7/27/2019 BS 200 Fixed Time

    19/42

    Mono reagent items' reaction time begins when Sample is dispensed.

    Dual reagent items reaction time begins when R2 is dispensed.

    The time between S and R2 dispensing is called Incubation time,only dual reagent measurements have incubation time.

    R1 Sample

    Preheating time Reaction time

    To be set from

    the insert Fixed 14

    cycles

    Basic parameter for UREA

    Incubation

    time

    0 cycle of

    reaction time

  • 7/27/2019 BS 200 Fixed Time

    20/42

    Time is all calculated into cycles on BS-200, and 1 cycleequals to 16 seconds.

    Roughly speaking, we change minutes into cycles forReac. Time and Incuba. Time as below.

    Basic parameter for UREA

    minute cycle

    0.5 2

    1 4

    2 8

    3 12

    4 16

    5 20

  • 7/27/2019 BS 200 Fixed Time

    21/42

    According to the instruction and the minute-cycle conversion table:

    The incubation time is be 1 minute, namely 4 cycles;

    The reaction time should be 30seconds and 30+60seconds,

    namely 2 and 6 cycles.

    Basic parameter for UREA

  • 7/27/2019 BS 200 Fixed Time

    22/42

    Reac. Timeis set as 2 and 6, Incuba. Time as 4.

    1

    2 6

    Descending

    UREA

    Fixed-time

    340

    4

  • 7/27/2019 BS 200 Fixed Time

    23/42

    Unitof the item, should be set according to the reagent insert

    1

    2 6

    Descending

    UREA

    Fixed-time

    340

    4

  • 7/27/2019 BS 200 Fixed Time

    24/42

    According to the insert, we can set the Unit as mg/dl ormmol/l for serum testing, and g/24h or mmol/24h forurine samples.

    Basic parameter for UREA

  • 7/27/2019 BS 200 Fixed Time

    25/42

    2 6

    4

    We can set the Unitas mg/dl.

    1

    mg/dl

    Descending

    UREA

    Fixed-time

    340

    1

  • 7/27/2019 BS 200 Fixed Time

    26/42

    2 6

    4

    Precision of the items results, should be referred to the reagent insert

    1

    Descending

    UREA

    Fixed-time

    340

    mg/dl

  • 7/27/2019 BS 200 Fixed Time

    27/42

    If the unit is set as mg/dl, its normal to report results

    without decimals. We can set the Precision as 0.

    Basic parameter for UREA

  • 7/27/2019 BS 200 Fixed Time

    28/42

    2 6

    4

    Precision of the item is set as 0.

    1

    0

    Descending

    UREA

    Fixed-time

    340

    mg/dl

  • 7/27/2019 BS 200 Fixed Time

    29/42

    2 6

    4

    R1, R2 and Sample Volume of the item, should be referred to

    1

    Descending

    UREA

    Fixed-time

    340

    0

    mg/dl

  • 7/27/2019 BS 200 Fixed Time

    30/42

    To process with the sample testing, we need 10ul sample, 1000ulR1 and 250ul R2 (namely, the ratio of S/R1/R2 is 1/100/25);

    According to the volume range of BS-200 (sample: 3-45ul, R1:

    180-450ul; R2: 30-450ul), and the indicated S/R1/R2 ratio;

    We can set Sample volume as 3ul, R1 as 300ul, and R2 as 75ul.

    Basic parameter for UREA

  • 7/27/2019 BS 200 Fixed Time

    31/42

    2 6

    4

    R1, R2 and Sample Volume are set as 300, 75 and 3.

    3

    300

    Descending

    UREA

    Fixed-time

    340

    0

    mg/dl

    751

  • 7/27/2019 BS 200 Fixed Time

    32/42

    2 6

    4

    Absorbance ofR1 Blank (mono rgt.) and Mixed Rgt. Blank (dual rgt.),

    set according to experience or reagent insert; if no idea, can be blank.

    1

    3

    300

    Descending

    UREA

    Fixed-time

    340

    0

    mg/dl

    75

  • 7/27/2019 BS 200 Fixed Time

    33/42

    2 6

    4

    1

    3

    300

    Linearity Range means the reagents measuring range, refer to

    Descending

    UREA

    Fixed-time

    340

    0

    mg/dl

    75

  • 7/27/2019 BS 200 Fixed Time

    34/42

    Basic parameter for UREA

    From the part of Linearity, we can get aware of that, theupper limit of Linearity Range is 300 mg/dl, or 50 mmol/l.

  • 7/27/2019 BS 200 Fixed Time

    35/42

    2 6

    4

    3

    300

    We can set the upper limit ofLinearity Range as 300 mg/dl.

    300

    Descending

    UREA

    Fixed-time

    340

    0

    mg/dl

    75

  • 7/27/2019 BS 200 Fixed Time

    36/42

    2 6

    4

    1

    3

    300

    Linearity Limit, Substrate Limit and Factorshould be set for the

    Kinetics tests, so here we can leave them as blank.

    Descending

    UREA

    Fixed-time

    340

    0

    mg/dl

    75

    300

  • 7/27/2019 BS 200 Fixed Time

    37/42

    2 6

    4

    1

    3

    300

    Prozone check is set for the turbidimetric tests, which can also be blank.

    Descending

    UREA

    Fixed-time

    340

    0

    mg/dl

    75

    300

  • 7/27/2019 BS 200 Fixed Time

    38/42

    The basic parameters for UREA havebeen finished

    Next, we should set the reagentinformation as following

    Reagent parameter for UREA

  • 7/27/2019 BS 200 Fixed Time

    39/42

  • 7/27/2019 BS 200 Fixed Time

    40/42

  • 7/27/2019 BS 200 Fixed Time

    41/42

    Basic reagent information (Position andBottle Type) has been set up

    We should set relevant calibration and QC

    information for the item

    (Please refer to the video of Calibration & QC setting)

    Calibration and QC for UREA

    http://images.google.cn/imgres?imgurl=http://news.xinhuanet.com/auto/2006-09/01/xinsrc_06209030108183282472421.jpg&imgrefurl=http://news.xinhuanet.com/auto/2006-09/01/content_5033813.htm&h=295&w=400&sz=18&hl=zh-CN&start=15&um=1&tbnid=qOsPXSyElbEfgM:&tbnh=91&tbnw=124&prev=/images%3Fq%3D%25E7%259C%258B%25E7%2594%25B5%25E5%25BD%25B1%26um%3D1%26complete%3D1%26hl%3Dzh-CN%26newwindow%3D1%26rls%3Dcom.microsoft:zh-cn:IE-SearchBox%26rlz%3D1I7SUNA
  • 7/27/2019 BS 200 Fixed Time

    42/42

    Thank you!

    NOTE:Fixed-time items should be set in the same way as UREA.