Blood pressure indirect methods

16
Indirect Blood Pressure Measurement - Sphygmomanometer

description

Blood pressure indirect methods sphygmomanometer , ultrasonic method

Transcript of Blood pressure indirect methods

Page 1: Blood pressure indirect methods

Indirect Blood Pressure Measurement - Sphygmomanometer

Page 2: Blood pressure indirect methods

Indirect Blood Pressure Measurement - Sphygmomanometer

Page 3: Blood pressure indirect methods

Indirect Blood Pressure Measurement - Sphygmomanometer

Page 4: Blood pressure indirect methods

Indirect Blood Pressure Measurement - Sphygmomanometer

• The pressure cuff on the upper arm is first inflated to a pressure well above the systolic pressure.

• At this point no sound can be heard through the stethoscope, which is placed over the brachial artery, for that artery has been collapsed by the pressure of the cuff.

• The pressure in the cuff is then gradually reduced.

Page 5: Blood pressure indirect methods

Sphygmomanometery…

• When the systolic peaks are higher than the occlusive pressure, the blood spurts under that cuff and causes a palpable pulse in the wrist (Riva-Rocci Method)

• Audible sounds (Korotkoff (named after Dr. Nikolai Korotkoff) sounds) generated by the flow of blood and vibrations of the vessel under the cuff are heard through the stethoscope.

Page 6: Blood pressure indirect methods

Sphygmomanometery …

• The pressure of the cuff that is indicated on the manometer when the first Korotkoff sound is heard is recorded as the systolic blood pressure.

• As the pressure in the cuff is continues to drop, the Korotkoff sounds continue until the cuff pressure is no longer sufficient to occlude the vessel during any part of the cycle. Below this pressure the Korotkoff sounds disappear, marking the value of the diastolic pressure.

Page 7: Blood pressure indirect methods

Sphygmomanometery …

• Auscultatory (based on the Korotkoff sounds) technique is simpler and requires a minimum of equipment.• Cannot be used in noisy environments.

• Palpation (based on pulse on the blood vessel) technique doesn’t require a noise free environment.

• Normal respiration and vasomotor waves modulate the normal blood-pressure levels.

Page 8: Blood pressure indirect methods

Automated Indirect Blood Pressure measurement techniques

• Involves an automatic sphygmomanometer that inflates and deflates an occlusive cuff at a predetermined rate.

• A sensitive detector is used to measure the distal pulse or cuff pressure.

Page 9: Blood pressure indirect methods

Automated Auscultatory device

• Microphone replaces the stethoscope for sensing the Korotkoff sounds.

• The process begins with a rapid (20-30mm Hg/s) inflation of the occlusive cuff to a preset pressure about 30mm Hg higher that the suspected systolic pressure.

• The flow of blood beneath the cuff is stopped by the collapse of the vessel.

Page 10: Blood pressure indirect methods

Automated Auscultatory device

• Cuff pressure is then reduced slowly (2-3 mm Hg/s).

• The first Korotkoff sound is detected by the microphone, at which time the level of the cuff pressure is stored.

Page 11: Blood pressure indirect methods

Automated Auscultatory device

• The muffling and silent period of the Korotkoff sound is detected, and the value of the diastolic pressure is also stored.

• After a few minutes the instrument displays the systolic and diastolic pressure and recycles the operation

Page 12: Blood pressure indirect methods

Ultrasonic Based Blood Pressure Measurement

• Employs a transcutaneous Doppler sensor that detects the motion of the blood-vessel walls in the various states of occlusion.

• The Doppler ultrasonic transducer is focused on the vessel wall and the blood.

• The reflected signal (shifted in frequency) is detected by the receiving crystal and decoded.

Page 13: Blood pressure indirect methods

Ultrasonic Based Blood Pressure Measurement…

Page 14: Blood pressure indirect methods

Ultrasonic Based Blood Pressure Measurement…

• The difference in frequency, in the range of 40 to 500 Hz, between the transmitted and received signals is proportional to the velocity of the wall motion and the blood velocity.

Page 15: Blood pressure indirect methods

Ultrasonic Based Blood Pressure Measurement…

• As the applied pressure is further increased, the time between the opening and closing decreases until they coincide. The reading at this point is the systolic pressure.

• When the pressure is cuff is reduced, the time between the opening and closing increases until the closing signal from one pulse coincides with opening signal from the next. The reading at this point is the diastolic pressure.

Page 16: Blood pressure indirect methods

Ultrasonic Based Blood Pressure Measurement…

• Advantages:• Doesn’t require a noise free environment.

• Disadvantage:• Movement of the subject‘s body cause

changes in ultrasonic path between the sensor and the blood vessel.