BJP ADVANCED PROTEIN SEPARATIONS - Vivaproducts

2
ADVANCED PROTEIN SEPARATIONS CLINICAL CONCENTRATORS BJP TECHNICAL INFORMATION & OPERATING INSTRUCTIONS Concentration Factors If you are not filling your unit to the full level, Table 1 will help determine the concentration factor you want to achieve. First, locate the row with the sample volume for your BJP model. Then, locate your desired concentration factor in that row. Finally, read the Graduation Mark value at the top of that column. This is where your sample should be removed. For instance, using the BJP 10, if the starting volume of the sample you are concentrating is 2.0 mL and you need to concentrate to 20X, match the 2 mL starting volume at left and look across until you reach the concentration factor you need to reach (20X) and look at the top of that column. Your sample will be concentrated to 20X when it reaches the 100X graduation mark on the unit. BJP 5 unit k r a M n o i t a u d a r G . l o V t r a t S 5X 10X 25X 50X 100X 5 mL 5X 10X 25X 50X 100X 4 mL 4X 8X 20X 40X 80X 3 mL 3X 6X 15X 30X 60X 2.5 mL 2.5X 5X 12.5X 25X 50X 2 mL 2X 4X 10X 20X 40X 1.5 mL 1.5X 3X 7.5X 15X 30X 1 mL -- 2X 5X 10X 20X BJP 10 unit k r a M n o i t a u d a r G . l o V t r a t S 5X 10X 25X 50X 100X 200X 10 mL 5X 10X 25X 50X 100X 200X 5 mL 2.5X 5X 12.5X 25X 50X 100X 2.5 mL 1.3X 2.5X 6.3X 12.5X 25X 50X 2 mL -- 2X 5X 10X 20X 40X 1.5 mL -- 1.5X 3.8X 7.5X 15X 30X 1 mL -- -- 2.5X 5X 10X 20X BJP 20 unit k r a M n o i t a u d a r G . l o V t r a t S 5X 10X 25X 50X 100X 200X 20 mL 10X 20X 50X 100X 200X 400X 15 mL 7.5X 15X 37.5X 75X 150X 300X 10 mL 5X 10X 25X 50X 100X 200X 5 mL 2.5X 5X 12.5X 25X 50X 100X 2 mL 1X 2X 5X 10X 20X 40X TABLE 1: Concentration Factors BJP-5, BJP-10 & BJP-20 BJP concentrators offer fast and convenient means to prepare multiple clinical samples for analysis by electrophoresis or immunoassay. This is done without a centrifuge, pressure or vacuum source. Absorbent pads pull solvent and microsolute through the ultrafilter concentrating the sample. A dead stop pocket at the bottom of each cell prevents filtration to dryness. BJP Applications — Concentration of urine prior to electrophoresis for diagnosis of multiple myeloma and amyloidosis. — Concentrate spinal fluid prior to electrophoresis for diagnosis of meningitis and multiple sclerosis. — Concentrate bacterial antigens in urine (Legionella, Pneumonia, Streptococcus B) prior to immunoassay. — Concentration of proteins and antigens in other clinical or research samples. Features Benefits Integrated dead stop No risk to over concentration. Uses plastic instead of No glass breakage during glass Pasteur pipettes sample removal. Black print for graduations Easy to read. Large volume models hold Concentration 100X in one up to 20 mL cell and still have 100 µl for IFE tests. Available as individual units Individual units may be or as 8 test blocks disposed after use. Blocks may have to be stored with sample residue. Recovering Dry Samples BJP concentrators have an impermeable concentrate pocket (dead stop) which impedes concentration to dryness. However, if the concentrate inadvertently remains too long in the concentrator, the remaining solvent will eventually evaporate and the sample may go to dryness. Should this occur, proteins may be retrieved to solution by pipetting approximately 100 µl of buffer in and out of the concentrate pocket several times. Operation: Concentrating macromolecules A. Measure sample total protein to determine the desired concentration factor. (For urine samples see below “Suggested Concentration Factors”) B. Pipette sample through aperture at the top of the device. Device can be left unattended until desired concentration is achieved. C. Solvent and micromolecules are pulled through the membrane by a high capacity absorbent, a dead stop feature prevents the sample from concentrating to dryness. D. Once the desired volume is achieved, the concentrate is withdrawn using Pasteur, thin plastic or gel loader type pipettes. The sample is now ready for further analysis. B C D B C D - - - A Division of Vivaproducts

Transcript of BJP ADVANCED PROTEIN SEPARATIONS - Vivaproducts

ADVANCED PROTEIN SEPARATIONS

CLINICALCONCENTRATORSBJPBJP

Technical Information & Operating InstructionsBJP 5, BJP 10 & BJP 20

• Use plastic instead of glass pipettes

• Easy to read concentration factors

• Dead stop prevents over concentration

• Volumes up to 20 mL

BJP CLINICALCONCENTRATORS

Operating Tips: Improving speed ofconcentration:

Speed of filtration is affected by severalparameters including temperature, pH andprotein concentration. While BJPconcentrators will provide rapid filtration inmost environments, the followingsuggestions may be helpful in optimizingspeed. Filtration speed will increaseproportionally to ambient temperature.Should you require faster concentration,place the concentrator near a source of heat.An acid sample with a pH of less than 5 willtake longer to concentrate than a neutralsample. Adjustment to a physiological pH willresult in faster filtration.

Suspended particles will tend to foul thefilter element and slow filtration speed.Prefiltration with a syringe filter will clarifythe sample and result in faster filtrationspeed and improved analytical resultsfollowing concentration. Initial proteinconcentration levels will have a significanteffect on concentration speed. While a highlydilute sample will concentrate rapidly, oncemacromolecular concentration exceeds 2 %the speed of filtration will rapidly decrease.Static concentrators are not practical forconcentrations above 5 %.

Recovering dry samples

BJP concentrators have an impermeableconcentrate pocket (dead stop) whichimpedes concentration to dryness. However,if the concentrate inadvertently remains toolong in the concentrator, the remainingsolvent will eventually evaporate and thesample may go to dryness. Should this occur,proteins may be retrieved to solution bypipetting approximately 100 µl of buffer inand out of the concentrate pocket severaltimes.

Following are the concentration factors forcommon biological solutions. Locate thestarting sample volume and the graduationmark for the liquid level; then read theconcentration factor in the box at the rowand column intersection. To calculatevolume, divide the initial volume by theconcentration factor.

OPERATING INSTRUCTIONS

A. Determine approximate samplemicrosolute content (for electrophoresis orimmunoelectrophoresis protein concentrationshould be approximately 25 mg/mL)

B. Pipette sample through aperture at thetop of the device. Device can be leftunattended until desired concentration isachieved.

C. Solvent and micromolecules are pulledthrough the membrane by a high capacityabsorbent, a dead stop feature prevents thesample from concentrating to dryness.

D. Once the desired volume is achieved,the concentrate is withdrawn using Pasteur,thin plastic or gel loader type pipettes. Thesample is now ready for further analysis.

If you are not filling your unit to the full level, these tables will help determine theconcentration factor you want to achieve. For instance, using the BJP 10, if the startingvolume of the sample you are concentrating is 2.0 mL and you need to concentrate to 20X,match the 2mL starting volume at left and look across until you reach the concentrationfactor you need to reach (20X) and look at the top of that column. Your sample will beconcentrated to 20X when it reaches the 100X graduation mark on the unit.

TABLE 1: Concentration Factors

Page 3

B C D

BJP 20 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X20 mL 10X 20X 50X 100X 200X 400X15 mL 7.5X 15X 37.5X 75X 150X 300X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2 mL 1X 2X 5X 10X 20X 40X

BJP 10 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2.5 mL 1.3X 2.5X 6.3X 12.5X 25X 50X2 mL -- 2X 5X 10X 20X 40X1.5 mL -- 1.5X 3.8X 7.5X 15X 30X1 mL -- -- 2.5X 5X 10X 20X

BJP 5 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X5 mL 5X 10X 25X 50X 100X4 mL 4X 8X 20X 40X 80X3 mL 3X 6X 15X 30X 60X2.5 mL 2.5X 5X 12.5X 25X 50X2 mL 2X 4X 10X 20X 40X1.5 mL 1.5X 3X 7.5X 15X 30X1 mL -- 2X 5X 10X 20X

Operating Tips: Improving speed ofconcentration:

Speed of filtration is affected by severalparameters including temperature, pH andprotein concentration. While BJPconcentrators will provide rapid filtration inmost environments, the followingsuggestions may be helpful in optimizingspeed. Filtration speed will increaseproportionally to ambient temperature.Should you require faster concentration,place the concentrator near a source of heat.An acid sample with a pH of less than 5 willtake longer to concentrate than a neutralsample. Adjustment to a physiological pH willresult in faster filtration.

Suspended particles will tend to foul thefilter element and slow filtration speed.Prefiltration with a syringe filter will clarifythe sample and result in faster filtrationspeed and improved analytical resultsfollowing concentration. Initial proteinconcentration levels will have a significanteffect on concentration speed. While a highlydilute sample will concentrate rapidly, oncemacromolecular concentration exceeds 2 %the speed of filtration will rapidly decrease.Static concentrators are not practical forconcentrations above 5 %.

Recovering dry samples

BJP concentrators have an impermeableconcentrate pocket (dead stop) whichimpedes concentration to dryness. However,if the concentrate inadvertently remains toolong in the concentrator, the remainingsolvent will eventually evaporate and thesample may go to dryness. Should this occur,proteins may be retrieved to solution bypipetting approximately 100 µl of buffer inand out of the concentrate pocket severaltimes.

Following are the concentration factors forcommon biological solutions. Locate thestarting sample volume and the graduationmark for the liquid level; then read theconcentration factor in the box at the rowand column intersection. To calculatevolume, divide the initial volume by theconcentration factor.

OPERATING INSTRUCTIONS

A. Determine approximate samplemicrosolute content (for electrophoresis orimmunoelectrophoresis protein concentrationshould be approximately 25 mg/mL)

B. Pipette sample through aperture at thetop of the device. Device can be leftunattended until desired concentration isachieved.

C. Solvent and micromolecules are pulledthrough the membrane by a high capacityabsorbent, a dead stop feature prevents thesample from concentrating to dryness.

D. Once the desired volume is achieved,the concentrate is withdrawn using Pasteur,thin plastic or gel loader type pipettes. Thesample is now ready for further analysis.

If you are not filling your unit to the full level, these tables will help determine theconcentration factor you want to achieve. For instance, using the BJP 10, if the startingvolume of the sample you are concentrating is 2.0 mL and you need to concentrate to 20X,match the 2mL starting volume at left and look across until you reach the concentrationfactor you need to reach (20X) and look at the top of that column. Your sample will beconcentrated to 20X when it reaches the 100X graduation mark on the unit.

TABLE 1: Concentration Factors

Page 3

B C D

BJP 20 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X20 mL 10X 20X 50X 100X 200X 400X15 mL 7.5X 15X 37.5X 75X 150X 300X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2 mL 1X 2X 5X 10X 20X 40X

BJP 10 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2.5 mL 1.3X 2.5X 6.3X 12.5X 25X 50X2 mL -- 2X 5X 10X 20X 40X1.5 mL -- 1.5X 3.8X 7.5X 15X 30X1 mL -- -- 2.5X 5X 10X 20X

BJP 5 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X5 mL 5X 10X 25X 50X 100X4 mL 4X 8X 20X 40X 80X3 mL 3X 6X 15X 30X 60X2.5 mL 2.5X 5X 12.5X 25X 50X2 mL 2X 4X 10X 20X 40X1.5 mL 1.5X 3X 7.5X 15X 30X1 mL -- 2X 5X 10X 20X

Operating Tips: Improving speed ofconcentration:

Speed of filtration is affected by severalparameters including temperature, pH andprotein concentration. While BJPconcentrators will provide rapid filtration inmost environments, the followingsuggestions may be helpful in optimizingspeed. Filtration speed will increaseproportionally to ambient temperature.Should you require faster concentration,place the concentrator near a source of heat.An acid sample with a pH of less than 5 willtake longer to concentrate than a neutralsample. Adjustment to a physiological pH willresult in faster filtration.

Suspended particles will tend to foul thefilter element and slow filtration speed.Prefiltration with a syringe filter will clarifythe sample and result in faster filtrationspeed and improved analytical resultsfollowing concentration. Initial proteinconcentration levels will have a significanteffect on concentration speed. While a highlydilute sample will concentrate rapidly, oncemacromolecular concentration exceeds 2 %the speed of filtration will rapidly decrease.Static concentrators are not practical forconcentrations above 5 %.

Recovering dry samples

BJP concentrators have an impermeableconcentrate pocket (dead stop) whichimpedes concentration to dryness. However,if the concentrate inadvertently remains toolong in the concentrator, the remainingsolvent will eventually evaporate and thesample may go to dryness. Should this occur,proteins may be retrieved to solution bypipetting approximately 100 µl of buffer inand out of the concentrate pocket severaltimes.

Following are the concentration factors forcommon biological solutions. Locate thestarting sample volume and the graduationmark for the liquid level; then read theconcentration factor in the box at the rowand column intersection. To calculatevolume, divide the initial volume by theconcentration factor.

OPERATING INSTRUCTIONS

A. Determine approximate samplemicrosolute content (for electrophoresis orimmunoelectrophoresis protein concentrationshould be approximately 25 mg/mL)

B. Pipette sample through aperture at thetop of the device. Device can be leftunattended until desired concentration isachieved.

C. Solvent and micromolecules are pulledthrough the membrane by a high capacityabsorbent, a dead stop feature prevents thesample from concentrating to dryness.

D. Once the desired volume is achieved,the concentrate is withdrawn using Pasteur,thin plastic or gel loader type pipettes. Thesample is now ready for further analysis.

If you are not filling your unit to the full level, these tables will help determine theconcentration factor you want to achieve. For instance, using the BJP 10, if the startingvolume of the sample you are concentrating is 2.0 mL and you need to concentrate to 20X,match the 2mL starting volume at left and look across until you reach the concentrationfactor you need to reach (20X) and look at the top of that column. Your sample will beconcentrated to 20X when it reaches the 100X graduation mark on the unit.

TABLE 1: Concentration Factors

Page 3

B C D

BJP 20 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X20 mL 10X 20X 50X 100X 200X 400X15 mL 7.5X 15X 37.5X 75X 150X 300X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2 mL 1X 2X 5X 10X 20X 40X

BJP 10 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2.5 mL 1.3X 2.5X 6.3X 12.5X 25X 50X2 mL -- 2X 5X 10X 20X 40X1.5 mL -- 1.5X 3.8X 7.5X 15X 30X1 mL -- -- 2.5X 5X 10X 20X

BJP 5 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X5 mL 5X 10X 25X 50X 100X4 mL 4X 8X 20X 40X 80X3 mL 3X 6X 15X 30X 60X2.5 mL 2.5X 5X 12.5X 25X 50X2 mL 2X 4X 10X 20X 40X1.5 mL 1.5X 3X 7.5X 15X 30X1 mL -- 2X 5X 10X 20X

BJP 5, BJP 10 & BJP 20

BJP concentrators offer fast and convenient means toprepare multiple clinical samples for analysis by electrophoresis or immunoassay. This is done withouta centrifuge, pressure or vaccum source. Absorbent pulls solvent and microsolute through the ultrafilter concentrating the sample. An dead stop pocket at the bottom of each cell prevents filtration to dryness.

Improving Speed of Concentration

Speed of filtration is affected by several parameters including temperature pH and protein concentration. While BJP concentrators will provide rapid filtration in most environments, some factors will slow filtration.

A. Filtration speed will increase proportionally to ambient temperature. Should you require faster concentration, place the concentrator near a source of heat.

B. An acidic sample with a pH of less than 5 will take longer to concentrate than a neutral sample.Adjustment to a physiological pH will result in faster filtration.

C. Suspended particles will lend to foul the filter element and slow filtration speed. Prefiltration with a syringe filter will clarify the sample and result in faster filtration speed and improved analytical results following concentration.

D. Initial protein concentration levels will have a significant effect on concentration speed. While a highly dilute sample will concentrate rapidly. Once protein concentration exceeds 2% the speed of filtration will rapidly decrease. Static concentrators are not practical for concentrations above 5%.

Recovering Dry Samples

BJP concentrators have an impermeable concentrate pocket (dead stop) which impedes concentration to dryness. However, if the concentrate inadvertently remains too long in the concentrator, the remaining solvent will eventually evaporate and the sample may go to dryness. Should this occur, proteins may be retrieved to solution by pipetting approximately 100 µl of buffer in and out of the concentrate pocketseveral times.

Concentration Factors

if you are filling your unit to the full level, Table 1 will help determine the concentration factor you want to achieve. First, locate the row with the sample volume for your BP model. Then, locate your desired concentration, factor in that row. Finally, read the Graduation Mark value at the top of that column. This is where your sample should be removed. For instance, using the BJP 10, if the starting volume of the sample you are concentrating is 2.0 mL and you need to concentrate to 20%, match the 2 mL starting volume at left and look across until youreach the concentration factor you need to reach (20X)and look at the top of that column. Your sample will be concentrated to 20X when it reaches the 100X graduation mark on the unit.

TECHNICAL INFORMATION & OPERATING INSTRUCTIONS

.

.

B C D

BJP Applications– Concentration of urine prior to electrophoresis for diagnosis of multiple myeloma and amyloidosis.

– Concentrate spinal fluid prior to electrophoresis for diagnosis of meningitis and multiple sclerosis.

– Concentrate bacterial antigens in urine (Legionella, Pneumonia, Streptococcus B) prior to immunoassay.

– Concentration of proteins and antigens in other clinical or research samples.

FeaturesIntegrated dead stop uses plastic instead of glass pasteur pipettes. Black print for graduations. Variety of volumes to 20 ml.

Available as individual units or as 8 test blocks.

BenefitsNo risk of over concentration. No glass breakage during sample removal. Easy to read. Flexible sample volumes with higher concentration factors.

Individual units may be disposed after use. Blocks may have to be stored with sample residue.

TABLE 1: Concentration Factors

ADVANCED PROTEIN SEPARATIONS

CLINICALCONCENTRATORSBJPBJP

Technical Information & Operating InstructionsBJP 5, BJP 10 & BJP 20

• Use plastic instead of glass pipettes

• Easy to read concentration factors

• Dead stop prevents over concentration

• Volumes up to 20 mL

BJP CLINICALCONCENTRATORS

Operating Tips: Improving speed ofconcentration:

Speed of filtration is affected by severalparameters including temperature, pH andprotein concentration. While BJPconcentrators will provide rapid filtration inmost environments, the followingsuggestions may be helpful in optimizingspeed. Filtration speed will increaseproportionally to ambient temperature.Should you require faster concentration,place the concentrator near a source of heat.An acid sample with a pH of less than 5 willtake longer to concentrate than a neutralsample. Adjustment to a physiological pH willresult in faster filtration.

Suspended particles will tend to foul thefilter element and slow filtration speed.Prefiltration with a syringe filter will clarifythe sample and result in faster filtrationspeed and improved analytical resultsfollowing concentration. Initial proteinconcentration levels will have a significanteffect on concentration speed. While a highlydilute sample will concentrate rapidly, oncemacromolecular concentration exceeds 2 %the speed of filtration will rapidly decrease.Static concentrators are not practical forconcentrations above 5 %.

Recovering dry samples

BJP concentrators have an impermeableconcentrate pocket (dead stop) whichimpedes concentration to dryness. However,if the concentrate inadvertently remains toolong in the concentrator, the remainingsolvent will eventually evaporate and thesample may go to dryness. Should this occur,proteins may be retrieved to solution bypipetting approximately 100 µl of buffer inand out of the concentrate pocket severaltimes.

Following are the concentration factors forcommon biological solutions. Locate thestarting sample volume and the graduationmark for the liquid level; then read theconcentration factor in the box at the rowand column intersection. To calculatevolume, divide the initial volume by theconcentration factor.

OPERATING INSTRUCTIONS

A. Determine approximate samplemicrosolute content (for electrophoresis orimmunoelectrophoresis protein concentrationshould be approximately 25 mg/mL)

B. Pipette sample through aperture at thetop of the device. Device can be leftunattended until desired concentration isachieved.

C. Solvent and micromolecules are pulledthrough the membrane by a high capacityabsorbent, a dead stop feature prevents thesample from concentrating to dryness.

D. Once the desired volume is achieved,the concentrate is withdrawn using Pasteur,thin plastic or gel loader type pipettes. Thesample is now ready for further analysis.

If you are not filling your unit to the full level, these tables will help determine theconcentration factor you want to achieve. For instance, using the BJP 10, if the startingvolume of the sample you are concentrating is 2.0 mL and you need to concentrate to 20X,match the 2mL starting volume at left and look across until you reach the concentrationfactor you need to reach (20X) and look at the top of that column. Your sample will beconcentrated to 20X when it reaches the 100X graduation mark on the unit.

TABLE 1: Concentration Factors

Page 3

B C D

BJP 20 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X20 mL 10X 20X 50X 100X 200X 400X15 mL 7.5X 15X 37.5X 75X 150X 300X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2 mL 1X 2X 5X 10X 20X 40X

BJP 10 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2.5 mL 1.3X 2.5X 6.3X 12.5X 25X 50X2 mL -- 2X 5X 10X 20X 40X1.5 mL -- 1.5X 3.8X 7.5X 15X 30X1 mL -- -- 2.5X 5X 10X 20X

BJP 5 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X5 mL 5X 10X 25X 50X 100X4 mL 4X 8X 20X 40X 80X3 mL 3X 6X 15X 30X 60X2.5 mL 2.5X 5X 12.5X 25X 50X2 mL 2X 4X 10X 20X 40X1.5 mL 1.5X 3X 7.5X 15X 30X1 mL -- 2X 5X 10X 20X

Operating Tips: Improving speed ofconcentration:

Speed of filtration is affected by severalparameters including temperature, pH andprotein concentration. While BJPconcentrators will provide rapid filtration inmost environments, the followingsuggestions may be helpful in optimizingspeed. Filtration speed will increaseproportionally to ambient temperature.Should you require faster concentration,place the concentrator near a source of heat.An acid sample with a pH of less than 5 willtake longer to concentrate than a neutralsample. Adjustment to a physiological pH willresult in faster filtration.

Suspended particles will tend to foul thefilter element and slow filtration speed.Prefiltration with a syringe filter will clarifythe sample and result in faster filtrationspeed and improved analytical resultsfollowing concentration. Initial proteinconcentration levels will have a significanteffect on concentration speed. While a highlydilute sample will concentrate rapidly, oncemacromolecular concentration exceeds 2 %the speed of filtration will rapidly decrease.Static concentrators are not practical forconcentrations above 5 %.

Recovering dry samples

BJP concentrators have an impermeableconcentrate pocket (dead stop) whichimpedes concentration to dryness. However,if the concentrate inadvertently remains toolong in the concentrator, the remainingsolvent will eventually evaporate and thesample may go to dryness. Should this occur,proteins may be retrieved to solution bypipetting approximately 100 µl of buffer inand out of the concentrate pocket severaltimes.

Following are the concentration factors forcommon biological solutions. Locate thestarting sample volume and the graduationmark for the liquid level; then read theconcentration factor in the box at the rowand column intersection. To calculatevolume, divide the initial volume by theconcentration factor.

OPERATING INSTRUCTIONS

A. Determine approximate samplemicrosolute content (for electrophoresis orimmunoelectrophoresis protein concentrationshould be approximately 25 mg/mL)

B. Pipette sample through aperture at thetop of the device. Device can be leftunattended until desired concentration isachieved.

C. Solvent and micromolecules are pulledthrough the membrane by a high capacityabsorbent, a dead stop feature prevents thesample from concentrating to dryness.

D. Once the desired volume is achieved,the concentrate is withdrawn using Pasteur,thin plastic or gel loader type pipettes. Thesample is now ready for further analysis.

If you are not filling your unit to the full level, these tables will help determine theconcentration factor you want to achieve. For instance, using the BJP 10, if the startingvolume of the sample you are concentrating is 2.0 mL and you need to concentrate to 20X,match the 2mL starting volume at left and look across until you reach the concentrationfactor you need to reach (20X) and look at the top of that column. Your sample will beconcentrated to 20X when it reaches the 100X graduation mark on the unit.

TABLE 1: Concentration Factors

Page 3

B C D

BJP 20 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X20 mL 10X 20X 50X 100X 200X 400X15 mL 7.5X 15X 37.5X 75X 150X 300X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2 mL 1X 2X 5X 10X 20X 40X

BJP 10 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2.5 mL 1.3X 2.5X 6.3X 12.5X 25X 50X2 mL -- 2X 5X 10X 20X 40X1.5 mL -- 1.5X 3.8X 7.5X 15X 30X1 mL -- -- 2.5X 5X 10X 20X

BJP 5 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X5 mL 5X 10X 25X 50X 100X4 mL 4X 8X 20X 40X 80X3 mL 3X 6X 15X 30X 60X2.5 mL 2.5X 5X 12.5X 25X 50X2 mL 2X 4X 10X 20X 40X1.5 mL 1.5X 3X 7.5X 15X 30X1 mL -- 2X 5X 10X 20X

Operating Tips: Improving speed ofconcentration:

Speed of filtration is affected by severalparameters including temperature, pH andprotein concentration. While BJPconcentrators will provide rapid filtration inmost environments, the followingsuggestions may be helpful in optimizingspeed. Filtration speed will increaseproportionally to ambient temperature.Should you require faster concentration,place the concentrator near a source of heat.An acid sample with a pH of less than 5 willtake longer to concentrate than a neutralsample. Adjustment to a physiological pH willresult in faster filtration.

Suspended particles will tend to foul thefilter element and slow filtration speed.Prefiltration with a syringe filter will clarifythe sample and result in faster filtrationspeed and improved analytical resultsfollowing concentration. Initial proteinconcentration levels will have a significanteffect on concentration speed. While a highlydilute sample will concentrate rapidly, oncemacromolecular concentration exceeds 2 %the speed of filtration will rapidly decrease.Static concentrators are not practical forconcentrations above 5 %.

Recovering dry samples

BJP concentrators have an impermeableconcentrate pocket (dead stop) whichimpedes concentration to dryness. However,if the concentrate inadvertently remains toolong in the concentrator, the remainingsolvent will eventually evaporate and thesample may go to dryness. Should this occur,proteins may be retrieved to solution bypipetting approximately 100 µl of buffer inand out of the concentrate pocket severaltimes.

Following are the concentration factors forcommon biological solutions. Locate thestarting sample volume and the graduationmark for the liquid level; then read theconcentration factor in the box at the rowand column intersection. To calculatevolume, divide the initial volume by theconcentration factor.

OPERATING INSTRUCTIONS

A. Determine approximate samplemicrosolute content (for electrophoresis orimmunoelectrophoresis protein concentrationshould be approximately 25 mg/mL)

B. Pipette sample through aperture at thetop of the device. Device can be leftunattended until desired concentration isachieved.

C. Solvent and micromolecules are pulledthrough the membrane by a high capacityabsorbent, a dead stop feature prevents thesample from concentrating to dryness.

D. Once the desired volume is achieved,the concentrate is withdrawn using Pasteur,thin plastic or gel loader type pipettes. Thesample is now ready for further analysis.

If you are not filling your unit to the full level, these tables will help determine theconcentration factor you want to achieve. For instance, using the BJP 10, if the startingvolume of the sample you are concentrating is 2.0 mL and you need to concentrate to 20X,match the 2mL starting volume at left and look across until you reach the concentrationfactor you need to reach (20X) and look at the top of that column. Your sample will beconcentrated to 20X when it reaches the 100X graduation mark on the unit.

TABLE 1: Concentration Factors

Page 3

B C D

BJP 20 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X20 mL 10X 20X 50X 100X 200X 400X15 mL 7.5X 15X 37.5X 75X 150X 300X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2 mL 1X 2X 5X 10X 20X 40X

BJP 10 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2.5 mL 1.3X 2.5X 6.3X 12.5X 25X 50X2 mL -- 2X 5X 10X 20X 40X1.5 mL -- 1.5X 3.8X 7.5X 15X 30X1 mL -- -- 2.5X 5X 10X 20X

BJP 5 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X5 mL 5X 10X 25X 50X 100X4 mL 4X 8X 20X 40X 80X3 mL 3X 6X 15X 30X 60X2.5 mL 2.5X 5X 12.5X 25X 50X2 mL 2X 4X 10X 20X 40X1.5 mL 1.5X 3X 7.5X 15X 30X1 mL -- 2X 5X 10X 20X

BJP 5, BJP 10 & BJP 20

BJP concentrators offer fast and convenient means toprepare multiple clinical samples for analysis by electrophoresis or immunoassay. This is done withouta centrifuge, pressure or vaccum source. Absorbent pulls solvent and microsolute through the ultrafilter concentrating the sample. An dead stop pocket at the bottom of each cell prevents filtration to dryness.

Improving Speed of Concentration

Speed of filtration is affected by several parameters including temperature pH and protein concentration. While BJP concentrators will provide rapid filtration in most environments, some factors will slow filtration.

A. Filtration speed will increase proportionally to ambient temperature. Should you require faster concentration, place the concentrator near a source of heat.

B. An acidic sample with a pH of less than 5 will take longer to concentrate than a neutral sample.Adjustment to a physiological pH will result in faster filtration.

C. Suspended particles will lend to foul the filter element and slow filtration speed. Prefiltration with a syringe filter will clarify the sample and result in faster filtration speed and improved analytical results following concentration.

D. Initial protein concentration levels will have a significant effect on concentration speed. While a highly dilute sample will concentrate rapidly. Once protein concentration exceeds 2% the speed of filtration will rapidly decrease. Static concentrators are not practical for concentrations above 5%.

Recovering Dry Samples

BJP concentrators have an impermeable concentrate pocket (dead stop) which impedes concentration to dryness. However, if the concentrate inadvertently remains too long in the concentrator, the remaining solvent will eventually evaporate and the sample may go to dryness. Should this occur, proteins may be retrieved to solution by pipetting approximately 100 µl of buffer in and out of the concentrate pocketseveral times.

Concentration Factors

if you are filling your unit to the full level, Table 1 will help determine the concentration factor you want to achieve. First, locate the row with the sample volume for your BP model. Then, locate your desired concentration, factor in that row. Finally, read the Graduation Mark value at the top of that column. This is where your sample should be removed. For instance, using the BJP 10, if the starting volume of the sample you are concentrating is 2.0 mL and you need to concentrate to 20%, match the 2 mL starting volume at left and look across until youreach the concentration factor you need to reach (20X)and look at the top of that column. Your sample will be concentrated to 20X when it reaches the 100X graduation mark on the unit.

TECHNICAL INFORMATION & OPERATING INSTRUCTIONS

.

.

B C D

BJP Applications– Concentration of urine prior to electrophoresis for diagnosis of multiple myeloma and amyloidosis.

– Concentrate spinal fluid prior to electrophoresis for diagnosis of meningitis and multiple sclerosis.

– Concentrate bacterial antigens in urine (Legionella, Pneumonia, Streptococcus B) prior to immunoassay.

– Concentration of proteins and antigens in other clinical or research samples.

FeaturesIntegrated dead stop uses plastic instead of glass pasteur pipettes. Black print for graduations. Variety of volumes to 20 ml.

Available as individual units or as 8 test blocks.

BenefitsNo risk of over concentration. No glass breakage during sample removal. Easy to read. Flexible sample volumes with higher concentration factors.

Individual units may be disposed after use. Blocks may have to be stored with sample residue.

TABLE 1: Concentration Factors

ADVANCED PROTEIN SEPARATIONS

CLINICALCONCENTRATORSBJPBJP

Technical Information & Operating InstructionsBJP 5, BJP 10 & BJP 20

• Use plastic instead of glass pipettes

• Easy to read concentration factors

• Dead stop prevents over concentration

• Volumes up to 20 mL

BJP CLINICALCONCENTRATORS

Operating Tips: Improving speed ofconcentration:

Speed of filtration is affected by severalparameters including temperature, pH andprotein concentration. While BJPconcentrators will provide rapid filtration inmost environments, the followingsuggestions may be helpful in optimizingspeed. Filtration speed will increaseproportionally to ambient temperature.Should you require faster concentration,place the concentrator near a source of heat.An acid sample with a pH of less than 5 willtake longer to concentrate than a neutralsample. Adjustment to a physiological pH willresult in faster filtration.

Suspended particles will tend to foul thefilter element and slow filtration speed.Prefiltration with a syringe filter will clarifythe sample and result in faster filtrationspeed and improved analytical resultsfollowing concentration. Initial proteinconcentration levels will have a significanteffect on concentration speed. While a highlydilute sample will concentrate rapidly, oncemacromolecular concentration exceeds 2 %the speed of filtration will rapidly decrease.Static concentrators are not practical forconcentrations above 5 %.

Recovering dry samples

BJP concentrators have an impermeableconcentrate pocket (dead stop) whichimpedes concentration to dryness. However,if the concentrate inadvertently remains toolong in the concentrator, the remainingsolvent will eventually evaporate and thesample may go to dryness. Should this occur,proteins may be retrieved to solution bypipetting approximately 100 µl of buffer inand out of the concentrate pocket severaltimes.

Following are the concentration factors forcommon biological solutions. Locate thestarting sample volume and the graduationmark for the liquid level; then read theconcentration factor in the box at the rowand column intersection. To calculatevolume, divide the initial volume by theconcentration factor.

OPERATING INSTRUCTIONS

A. Determine approximate samplemicrosolute content (for electrophoresis orimmunoelectrophoresis protein concentrationshould be approximately 25 mg/mL)

B. Pipette sample through aperture at thetop of the device. Device can be leftunattended until desired concentration isachieved.

C. Solvent and micromolecules are pulledthrough the membrane by a high capacityabsorbent, a dead stop feature prevents thesample from concentrating to dryness.

D. Once the desired volume is achieved,the concentrate is withdrawn using Pasteur,thin plastic or gel loader type pipettes. Thesample is now ready for further analysis.

If you are not filling your unit to the full level, these tables will help determine theconcentration factor you want to achieve. For instance, using the BJP 10, if the startingvolume of the sample you are concentrating is 2.0 mL and you need to concentrate to 20X,match the 2mL starting volume at left and look across until you reach the concentrationfactor you need to reach (20X) and look at the top of that column. Your sample will beconcentrated to 20X when it reaches the 100X graduation mark on the unit.

TABLE 1: Concentration Factors

Page 3

B C D

BJP 20 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X20 mL 10X 20X 50X 100X 200X 400X15 mL 7.5X 15X 37.5X 75X 150X 300X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2 mL 1X 2X 5X 10X 20X 40X

BJP 10 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2.5 mL 1.3X 2.5X 6.3X 12.5X 25X 50X2 mL -- 2X 5X 10X 20X 40X1.5 mL -- 1.5X 3.8X 7.5X 15X 30X1 mL -- -- 2.5X 5X 10X 20X

BJP 5 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X5 mL 5X 10X 25X 50X 100X4 mL 4X 8X 20X 40X 80X3 mL 3X 6X 15X 30X 60X2.5 mL 2.5X 5X 12.5X 25X 50X2 mL 2X 4X 10X 20X 40X1.5 mL 1.5X 3X 7.5X 15X 30X1 mL -- 2X 5X 10X 20X

Operating Tips: Improving speed ofconcentration:

Speed of filtration is affected by severalparameters including temperature, pH andprotein concentration. While BJPconcentrators will provide rapid filtration inmost environments, the followingsuggestions may be helpful in optimizingspeed. Filtration speed will increaseproportionally to ambient temperature.Should you require faster concentration,place the concentrator near a source of heat.An acid sample with a pH of less than 5 willtake longer to concentrate than a neutralsample. Adjustment to a physiological pH willresult in faster filtration.

Suspended particles will tend to foul thefilter element and slow filtration speed.Prefiltration with a syringe filter will clarifythe sample and result in faster filtrationspeed and improved analytical resultsfollowing concentration. Initial proteinconcentration levels will have a significanteffect on concentration speed. While a highlydilute sample will concentrate rapidly, oncemacromolecular concentration exceeds 2 %the speed of filtration will rapidly decrease.Static concentrators are not practical forconcentrations above 5 %.

Recovering dry samples

BJP concentrators have an impermeableconcentrate pocket (dead stop) whichimpedes concentration to dryness. However,if the concentrate inadvertently remains toolong in the concentrator, the remainingsolvent will eventually evaporate and thesample may go to dryness. Should this occur,proteins may be retrieved to solution bypipetting approximately 100 µl of buffer inand out of the concentrate pocket severaltimes.

Following are the concentration factors forcommon biological solutions. Locate thestarting sample volume and the graduationmark for the liquid level; then read theconcentration factor in the box at the rowand column intersection. To calculatevolume, divide the initial volume by theconcentration factor.

OPERATING INSTRUCTIONS

A. Determine approximate samplemicrosolute content (for electrophoresis orimmunoelectrophoresis protein concentrationshould be approximately 25 mg/mL)

B. Pipette sample through aperture at thetop of the device. Device can be leftunattended until desired concentration isachieved.

C. Solvent and micromolecules are pulledthrough the membrane by a high capacityabsorbent, a dead stop feature prevents thesample from concentrating to dryness.

D. Once the desired volume is achieved,the concentrate is withdrawn using Pasteur,thin plastic or gel loader type pipettes. Thesample is now ready for further analysis.

If you are not filling your unit to the full level, these tables will help determine theconcentration factor you want to achieve. For instance, using the BJP 10, if the startingvolume of the sample you are concentrating is 2.0 mL and you need to concentrate to 20X,match the 2mL starting volume at left and look across until you reach the concentrationfactor you need to reach (20X) and look at the top of that column. Your sample will beconcentrated to 20X when it reaches the 100X graduation mark on the unit.

TABLE 1: Concentration Factors

Page 3

B C D

BJP 20 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X20 mL 10X 20X 50X 100X 200X 400X15 mL 7.5X 15X 37.5X 75X 150X 300X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2 mL 1X 2X 5X 10X 20X 40X

BJP 10 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2.5 mL 1.3X 2.5X 6.3X 12.5X 25X 50X2 mL -- 2X 5X 10X 20X 40X1.5 mL -- 1.5X 3.8X 7.5X 15X 30X1 mL -- -- 2.5X 5X 10X 20X

BJP 5 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X5 mL 5X 10X 25X 50X 100X4 mL 4X 8X 20X 40X 80X3 mL 3X 6X 15X 30X 60X2.5 mL 2.5X 5X 12.5X 25X 50X2 mL 2X 4X 10X 20X 40X1.5 mL 1.5X 3X 7.5X 15X 30X1 mL -- 2X 5X 10X 20X

Operating Tips: Improving speed ofconcentration:

Speed of filtration is affected by severalparameters including temperature, pH andprotein concentration. While BJPconcentrators will provide rapid filtration inmost environments, the followingsuggestions may be helpful in optimizingspeed. Filtration speed will increaseproportionally to ambient temperature.Should you require faster concentration,place the concentrator near a source of heat.An acid sample with a pH of less than 5 willtake longer to concentrate than a neutralsample. Adjustment to a physiological pH willresult in faster filtration.

Suspended particles will tend to foul thefilter element and slow filtration speed.Prefiltration with a syringe filter will clarifythe sample and result in faster filtrationspeed and improved analytical resultsfollowing concentration. Initial proteinconcentration levels will have a significanteffect on concentration speed. While a highlydilute sample will concentrate rapidly, oncemacromolecular concentration exceeds 2 %the speed of filtration will rapidly decrease.Static concentrators are not practical forconcentrations above 5 %.

Recovering dry samples

BJP concentrators have an impermeableconcentrate pocket (dead stop) whichimpedes concentration to dryness. However,if the concentrate inadvertently remains toolong in the concentrator, the remainingsolvent will eventually evaporate and thesample may go to dryness. Should this occur,proteins may be retrieved to solution bypipetting approximately 100 µl of buffer inand out of the concentrate pocket severaltimes.

Following are the concentration factors forcommon biological solutions. Locate thestarting sample volume and the graduationmark for the liquid level; then read theconcentration factor in the box at the rowand column intersection. To calculatevolume, divide the initial volume by theconcentration factor.

OPERATING INSTRUCTIONS

A. Determine approximate samplemicrosolute content (for electrophoresis orimmunoelectrophoresis protein concentrationshould be approximately 25 mg/mL)

B. Pipette sample through aperture at thetop of the device. Device can be leftunattended until desired concentration isachieved.

C. Solvent and micromolecules are pulledthrough the membrane by a high capacityabsorbent, a dead stop feature prevents thesample from concentrating to dryness.

D. Once the desired volume is achieved,the concentrate is withdrawn using Pasteur,thin plastic or gel loader type pipettes. Thesample is now ready for further analysis.

If you are not filling your unit to the full level, these tables will help determine theconcentration factor you want to achieve. For instance, using the BJP 10, if the startingvolume of the sample you are concentrating is 2.0 mL and you need to concentrate to 20X,match the 2mL starting volume at left and look across until you reach the concentrationfactor you need to reach (20X) and look at the top of that column. Your sample will beconcentrated to 20X when it reaches the 100X graduation mark on the unit.

TABLE 1: Concentration Factors

Page 3

B C D

BJP 20 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X20 mL 10X 20X 50X 100X 200X 400X15 mL 7.5X 15X 37.5X 75X 150X 300X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2 mL 1X 2X 5X 10X 20X 40X

BJP 10 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2.5 mL 1.3X 2.5X 6.3X 12.5X 25X 50X2 mL -- 2X 5X 10X 20X 40X1.5 mL -- 1.5X 3.8X 7.5X 15X 30X1 mL -- -- 2.5X 5X 10X 20X

BJP 5 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X5 mL 5X 10X 25X 50X 100X4 mL 4X 8X 20X 40X 80X3 mL 3X 6X 15X 30X 60X2.5 mL 2.5X 5X 12.5X 25X 50X2 mL 2X 4X 10X 20X 40X1.5 mL 1.5X 3X 7.5X 15X 30X1 mL -- 2X 5X 10X 20X

BJP 5, BJP 10 & BJP 20

BJP concentrators offer fast and convenient means toprepare multiple clinical samples for analysis by electrophoresis or immunoassay. This is done withouta centrifuge, pressure or vaccum source. Absorbent pulls solvent and microsolute through the ultrafilter concentrating the sample. An dead stop pocket at the bottom of each cell prevents filtration to dryness.

Improving Speed of Concentration

Speed of filtration is affected by several parameters including temperature pH and protein concentration. While BJP concentrators will provide rapid filtration in most environments, some factors will slow filtration.

A. Filtration speed will increase proportionally to ambient temperature. Should you require faster concentration, place the concentrator near a source of heat.

B. An acidic sample with a pH of less than 5 will take longer to concentrate than a neutral sample.Adjustment to a physiological pH will result in faster filtration.

C. Suspended particles will lend to foul the filter element and slow filtration speed. Prefiltration with a syringe filter will clarify the sample and result in faster filtration speed and improved analytical results following concentration.

D. Initial protein concentration levels will have a significant effect on concentration speed. While a highly dilute sample will concentrate rapidly. Once protein concentration exceeds 2% the speed of filtration will rapidly decrease. Static concentrators are not practical for concentrations above 5%.

Recovering Dry Samples

BJP concentrators have an impermeable concentrate pocket (dead stop) which impedes concentration to dryness. However, if the concentrate inadvertently remains too long in the concentrator, the remaining solvent will eventually evaporate and the sample may go to dryness. Should this occur, proteins may be retrieved to solution by pipetting approximately 100 µl of buffer in and out of the concentrate pocketseveral times.

Concentration Factors

if you are filling your unit to the full level, Table 1 will help determine the concentration factor you want to achieve. First, locate the row with the sample volume for your BP model. Then, locate your desired concentration, factor in that row. Finally, read the Graduation Mark value at the top of that column. This is where your sample should be removed. For instance, using the BJP 10, if the starting volume of the sample you are concentrating is 2.0 mL and you need to concentrate to 20%, match the 2 mL starting volume at left and look across until youreach the concentration factor you need to reach (20X)and look at the top of that column. Your sample will be concentrated to 20X when it reaches the 100X graduation mark on the unit.

TECHNICAL INFORMATION & OPERATING INSTRUCTIONS

.

.

B C D

BJP Applications– Concentration of urine prior to electrophoresis for diagnosis of multiple myeloma and amyloidosis.

– Concentrate spinal fluid prior to electrophoresis for diagnosis of meningitis and multiple sclerosis.

– Concentrate bacterial antigens in urine (Legionella, Pneumonia, Streptococcus B) prior to immunoassay.

– Concentration of proteins and antigens in other clinical or research samples.

FeaturesIntegrated dead stop uses plastic instead of glass pasteur pipettes. Black print for graduations. Variety of volumes to 20 ml.

Available as individual units or as 8 test blocks.

BenefitsNo risk of over concentration. No glass breakage during sample removal. Easy to read. Flexible sample volumes with higher concentration factors.

Individual units may be disposed after use. Blocks may have to be stored with sample residue.

TABLE 1: Concentration Factors

TECHNICAL INFORMATION& OPERATING INSTRUCTIONS

Concentration Factors

If you are not filling your unit to the full level, Table 1 will help determine the concentration factor you want to achieve. First, locate the row with the sample volume for your BJP model. Then, locate your desired concentration factor in that row. Finally, read the Graduation Mark value at the top of that column. This is where your sample should be removed. For instance, using the BJP 10, if the starting volume of the sample you are concentrating is 2.0 mL and you need to concentrate to 20X, match the 2 mL starting volume at left and look across until you reach the concentration factor you need to reach (20X) and look at the top of that column. Your sample will be concentrated to 20X when it reaches the 100X graduation mark on the unit.

ADVANCED PROTEIN SEPARATIONS

CLINICALCONCENTRATORSBJPBJP

Technical Information & Operating InstructionsBJP 5, BJP 10 & BJP 20

• Use plastic instead of glass pipettes

• Easy to read concentration factors

• Dead stop prevents over concentration

• Volumes up to 20 mL

BJP CLINICALCONCENTRATORS

Operating Tips: Improving speed ofconcentration:

Speed of filtration is affected by severalparameters including temperature, pH andprotein concentration. While BJPconcentrators will provide rapid filtration inmost environments, the followingsuggestions may be helpful in optimizingspeed. Filtration speed will increaseproportionally to ambient temperature.Should you require faster concentration,place the concentrator near a source of heat.An acid sample with a pH of less than 5 willtake longer to concentrate than a neutralsample. Adjustment to a physiological pH willresult in faster filtration.

Suspended particles will tend to foul thefilter element and slow filtration speed.Prefiltration with a syringe filter will clarifythe sample and result in faster filtrationspeed and improved analytical resultsfollowing concentration. Initial proteinconcentration levels will have a significanteffect on concentration speed. While a highlydilute sample will concentrate rapidly, oncemacromolecular concentration exceeds 2 %the speed of filtration will rapidly decrease.Static concentrators are not practical forconcentrations above 5 %.

Recovering dry samples

BJP concentrators have an impermeableconcentrate pocket (dead stop) whichimpedes concentration to dryness. However,if the concentrate inadvertently remains toolong in the concentrator, the remainingsolvent will eventually evaporate and thesample may go to dryness. Should this occur,proteins may be retrieved to solution bypipetting approximately 100 µl of buffer inand out of the concentrate pocket severaltimes.

Following are the concentration factors forcommon biological solutions. Locate thestarting sample volume and the graduationmark for the liquid level; then read theconcentration factor in the box at the rowand column intersection. To calculatevolume, divide the initial volume by theconcentration factor.

OPERATING INSTRUCTIONS

A. Determine approximate samplemicrosolute content (for electrophoresis orimmunoelectrophoresis protein concentrationshould be approximately 25 mg/mL)

B. Pipette sample through aperture at thetop of the device. Device can be leftunattended until desired concentration isachieved.

C. Solvent and micromolecules are pulledthrough the membrane by a high capacityabsorbent, a dead stop feature prevents thesample from concentrating to dryness.

D. Once the desired volume is achieved,the concentrate is withdrawn using Pasteur,thin plastic or gel loader type pipettes. Thesample is now ready for further analysis.

If you are not filling your unit to the full level, these tables will help determine theconcentration factor you want to achieve. For instance, using the BJP 10, if the startingvolume of the sample you are concentrating is 2.0 mL and you need to concentrate to 20X,match the 2mL starting volume at left and look across until you reach the concentrationfactor you need to reach (20X) and look at the top of that column. Your sample will beconcentrated to 20X when it reaches the 100X graduation mark on the unit.

TABLE 1: Concentration Factors

Page 3

B C D

BJP 20 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X20 mL 10X 20X 50X 100X 200X 400X15 mL 7.5X 15X 37.5X 75X 150X 300X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2 mL 1X 2X 5X 10X 20X 40X

BJP 10 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2.5 mL 1.3X 2.5X 6.3X 12.5X 25X 50X2 mL -- 2X 5X 10X 20X 40X1.5 mL -- 1.5X 3.8X 7.5X 15X 30X1 mL -- -- 2.5X 5X 10X 20X

BJP 5 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X5 mL 5X 10X 25X 50X 100X4 mL 4X 8X 20X 40X 80X3 mL 3X 6X 15X 30X 60X2.5 mL 2.5X 5X 12.5X 25X 50X2 mL 2X 4X 10X 20X 40X1.5 mL 1.5X 3X 7.5X 15X 30X1 mL -- 2X 5X 10X 20X

Operating Tips: Improving speed ofconcentration:

Speed of filtration is affected by severalparameters including temperature, pH andprotein concentration. While BJPconcentrators will provide rapid filtration inmost environments, the followingsuggestions may be helpful in optimizingspeed. Filtration speed will increaseproportionally to ambient temperature.Should you require faster concentration,place the concentrator near a source of heat.An acid sample with a pH of less than 5 willtake longer to concentrate than a neutralsample. Adjustment to a physiological pH willresult in faster filtration.

Suspended particles will tend to foul thefilter element and slow filtration speed.Prefiltration with a syringe filter will clarifythe sample and result in faster filtrationspeed and improved analytical resultsfollowing concentration. Initial proteinconcentration levels will have a significanteffect on concentration speed. While a highlydilute sample will concentrate rapidly, oncemacromolecular concentration exceeds 2 %the speed of filtration will rapidly decrease.Static concentrators are not practical forconcentrations above 5 %.

Recovering dry samples

BJP concentrators have an impermeableconcentrate pocket (dead stop) whichimpedes concentration to dryness. However,if the concentrate inadvertently remains toolong in the concentrator, the remainingsolvent will eventually evaporate and thesample may go to dryness. Should this occur,proteins may be retrieved to solution bypipetting approximately 100 µl of buffer inand out of the concentrate pocket severaltimes.

Following are the concentration factors forcommon biological solutions. Locate thestarting sample volume and the graduationmark for the liquid level; then read theconcentration factor in the box at the rowand column intersection. To calculatevolume, divide the initial volume by theconcentration factor.

OPERATING INSTRUCTIONS

A. Determine approximate samplemicrosolute content (for electrophoresis orimmunoelectrophoresis protein concentrationshould be approximately 25 mg/mL)

B. Pipette sample through aperture at thetop of the device. Device can be leftunattended until desired concentration isachieved.

C. Solvent and micromolecules are pulledthrough the membrane by a high capacityabsorbent, a dead stop feature prevents thesample from concentrating to dryness.

D. Once the desired volume is achieved,the concentrate is withdrawn using Pasteur,thin plastic or gel loader type pipettes. Thesample is now ready for further analysis.

If you are not filling your unit to the full level, these tables will help determine theconcentration factor you want to achieve. For instance, using the BJP 10, if the startingvolume of the sample you are concentrating is 2.0 mL and you need to concentrate to 20X,match the 2mL starting volume at left and look across until you reach the concentrationfactor you need to reach (20X) and look at the top of that column. Your sample will beconcentrated to 20X when it reaches the 100X graduation mark on the unit.

TABLE 1: Concentration Factors

Page 3

B C D

BJP 20 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X20 mL 10X 20X 50X 100X 200X 400X15 mL 7.5X 15X 37.5X 75X 150X 300X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2 mL 1X 2X 5X 10X 20X 40X

BJP 10 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2.5 mL 1.3X 2.5X 6.3X 12.5X 25X 50X2 mL -- 2X 5X 10X 20X 40X1.5 mL -- 1.5X 3.8X 7.5X 15X 30X1 mL -- -- 2.5X 5X 10X 20X

BJP 5 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X5 mL 5X 10X 25X 50X 100X4 mL 4X 8X 20X 40X 80X3 mL 3X 6X 15X 30X 60X2.5 mL 2.5X 5X 12.5X 25X 50X2 mL 2X 4X 10X 20X 40X1.5 mL 1.5X 3X 7.5X 15X 30X1 mL -- 2X 5X 10X 20X

Operating Tips: Improving speed ofconcentration:

Speed of filtration is affected by severalparameters including temperature, pH andprotein concentration. While BJPconcentrators will provide rapid filtration inmost environments, the followingsuggestions may be helpful in optimizingspeed. Filtration speed will increaseproportionally to ambient temperature.Should you require faster concentration,place the concentrator near a source of heat.An acid sample with a pH of less than 5 willtake longer to concentrate than a neutralsample. Adjustment to a physiological pH willresult in faster filtration.

Suspended particles will tend to foul thefilter element and slow filtration speed.Prefiltration with a syringe filter will clarifythe sample and result in faster filtrationspeed and improved analytical resultsfollowing concentration. Initial proteinconcentration levels will have a significanteffect on concentration speed. While a highlydilute sample will concentrate rapidly, oncemacromolecular concentration exceeds 2 %the speed of filtration will rapidly decrease.Static concentrators are not practical forconcentrations above 5 %.

Recovering dry samples

BJP concentrators have an impermeableconcentrate pocket (dead stop) whichimpedes concentration to dryness. However,if the concentrate inadvertently remains toolong in the concentrator, the remainingsolvent will eventually evaporate and thesample may go to dryness. Should this occur,proteins may be retrieved to solution bypipetting approximately 100 µl of buffer inand out of the concentrate pocket severaltimes.

Following are the concentration factors forcommon biological solutions. Locate thestarting sample volume and the graduationmark for the liquid level; then read theconcentration factor in the box at the rowand column intersection. To calculatevolume, divide the initial volume by theconcentration factor.

OPERATING INSTRUCTIONS

A. Determine approximate samplemicrosolute content (for electrophoresis orimmunoelectrophoresis protein concentrationshould be approximately 25 mg/mL)

B. Pipette sample through aperture at thetop of the device. Device can be leftunattended until desired concentration isachieved.

C. Solvent and micromolecules are pulledthrough the membrane by a high capacityabsorbent, a dead stop feature prevents thesample from concentrating to dryness.

D. Once the desired volume is achieved,the concentrate is withdrawn using Pasteur,thin plastic or gel loader type pipettes. Thesample is now ready for further analysis.

If you are not filling your unit to the full level, these tables will help determine theconcentration factor you want to achieve. For instance, using the BJP 10, if the startingvolume of the sample you are concentrating is 2.0 mL and you need to concentrate to 20X,match the 2mL starting volume at left and look across until you reach the concentrationfactor you need to reach (20X) and look at the top of that column. Your sample will beconcentrated to 20X when it reaches the 100X graduation mark on the unit.

TABLE 1: Concentration Factors

Page 3

B C D

BJP 20 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X20 mL 10X 20X 50X 100X 200X 400X15 mL 7.5X 15X 37.5X 75X 150X 300X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2 mL 1X 2X 5X 10X 20X 40X

BJP 10 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X 200X10 mL 5X 10X 25X 50X 100X 200X5 mL 2.5X 5X 12.5X 25X 50X 100X2.5 mL 1.3X 2.5X 6.3X 12.5X 25X 50X2 mL -- 2X 5X 10X 20X 40X1.5 mL -- 1.5X 3.8X 7.5X 15X 30X1 mL -- -- 2.5X 5X 10X 20X

BJP 5 unitkraM noitaudarG.loV tratS

5X 10X 25X 50X 100X5 mL 5X 10X 25X 50X 100X4 mL 4X 8X 20X 40X 80X3 mL 3X 6X 15X 30X 60X2.5 mL 2.5X 5X 12.5X 25X 50X2 mL 2X 4X 10X 20X 40X1.5 mL 1.5X 3X 7.5X 15X 30X1 mL -- 2X 5X 10X 20X

BJP 5, BJP 10 & BJP 20

BJP concentrators offer fast and convenient means toprepare multiple clinical samples for analysis by electrophoresis or immunoassay. This is done withouta centrifuge, pressure or vaccum source. Absorbent pulls solvent and microsolute through the ultrafilter concentrating the sample. An dead stop pocket at the bottom of each cell prevents filtration to dryness.

Improving Speed of Concentration

Speed of filtration is affected by several parameters including temperature pH and protein concentration. While BJP concentrators will provide rapid filtration in most environments, some factors will slow filtration.

A. Filtration speed will increase proportionally to ambient temperature. Should you require faster concentration, place the concentrator near a source of heat.

B. An acidic sample with a pH of less than 5 will take longer to concentrate than a neutral sample.Adjustment to a physiological pH will result in faster filtration.

C. Suspended particles will lend to foul the filter element and slow filtration speed. Prefiltration with a syringe filter will clarify the sample and result in faster filtration speed and improved analytical results following concentration.

D. Initial protein concentration levels will have a significant effect on concentration speed. While a highly dilute sample will concentrate rapidly. Once protein concentration exceeds 2% the speed of filtration will rapidly decrease. Static concentrators are not practical for concentrations above 5%.

Recovering Dry Samples

BJP concentrators have an impermeable concentrate pocket (dead stop) which impedes concentration to dryness. However, if the concentrate inadvertently remains too long in the concentrator, the remaining solvent will eventually evaporate and the sample may go to dryness. Should this occur, proteins may be retrieved to solution by pipetting approximately 100 µl of buffer in and out of the concentrate pocketseveral times.

Concentration Factors

if you are filling your unit to the full level, Table 1 will help determine the concentration factor you want to achieve. First, locate the row with the sample volume for your BP model. Then, locate your desired concentration, factor in that row. Finally, read the Graduation Mark value at the top of that column. This is where your sample should be removed. For instance, using the BJP 10, if the starting volume of the sample you are concentrating is 2.0 mL and you need to concentrate to 20%, match the 2 mL starting volume at left and look across until youreach the concentration factor you need to reach (20X)and look at the top of that column. Your sample will be concentrated to 20X when it reaches the 100X graduation mark on the unit.

TECHNICAL INFORMATION & OPERATING INSTRUCTIONS

.

.

B C D

BJP Applications– Concentration of urine prior to electrophoresis for diagnosis of multiple myeloma and amyloidosis.

– Concentrate spinal fluid prior to electrophoresis for diagnosis of meningitis and multiple sclerosis.

– Concentrate bacterial antigens in urine (Legionella, Pneumonia, Streptococcus B) prior to immunoassay.

– Concentration of proteins and antigens in other clinical or research samples.

FeaturesIntegrated dead stop uses plastic instead of glass pasteur pipettes. Black print for graduations. Variety of volumes to 20 ml.

Available as individual units or as 8 test blocks.

BenefitsNo risk of over concentration. No glass breakage during sample removal. Easy to read. Flexible sample volumes with higher concentration factors.

Individual units may be disposed after use. Blocks may have to be stored with sample residue.

TABLE 1: Concentration Factors

BJP-5, BJP-10 & BJP-20

BJP concentrators offer fast and convenient means to prepare multiple clinical samples for analysis by electrophoresis or immunoassay. This is done without a centrifuge, pressure or vacuum source. Absorbent pads pull solvent and microsolute through the ultrafilter concentrating the sample. A dead stop pocket at the bottom of each cell prevents filtration to dryness.

BJP Applications

— Concentration of urine prior to electrophoresis for diagnosis of multiple myeloma and amyloidosis.

— Concentrate spinal fluid prior to electrophoresis for diagnosis of meningitis and multiple sclerosis.

— Concentrate bacterial antigens in urine (Legionella, Pneumonia, Streptococcus B) prior to immunoassay.

— Concentration of proteins and antigens in other clinical or research samples.

Features Benefits

Integrated dead stop No risk to over concentration.

Uses plastic instead of No glass breakage duringglass Pasteur pipettes sample removal.

Black print for graduations Easy to read.

Large volume models hold Concentration 100X in oneup to 20 mL cell and still have 100 µl for IFE tests.

Available as individual units Individual units may be or as 8 test blocks disposed after use. Blocks may have to be stored with sample residue.

Recovering Dry Samples

BJP concentrators have an impermeable concentrate pocket (dead stop) which impedes concentration to dryness. However, if the concentrate inadvertently remains too long in the concentrator, the remaining solvent will eventually evaporate and the sample may go to dryness. Should this occur, proteins may be retrieved to solution by pipetting approximately 100 µl of buffer in and out of the concentrate pocket several times.

Operation: Concentrating macromolecules

A. Measure sample total protein to determine the desired concentration factor.(For urine samples see below “SuggestedConcentration Factors”)

B. Pipette sample through aperture at thetop of the device. Device can be leftunattended until desired concentration isachieved.

C. Solvent and micromolecules are pulledthrough the membrane by a high capacityabsorbent, a dead stop feature prevents thesample from concentrating to dryness.

D. Once the desired volume is achieved,the concentrate is withdrawn using Pasteur,thin plastic or gel loader type pipettes. Thesample is now ready for further analysis.

B C D

Operation: Concentrating macromolecules

A. Measure sample total protein to determine the desired concentration factor.(For urine samples see below “SuggestedConcentration Factors”)

B. Pipette sample through aperture at thetop of the device. Device can be leftunattended until desired concentration isachieved.

C. Solvent and micromolecules are pulledthrough the membrane by a high capacityabsorbent, a dead stop feature prevents thesample from concentrating to dryness.

D. Once the desired volume is achieved,the concentrate is withdrawn using Pasteur,thin plastic or gel loader type pipettes. Thesample is now ready for further analysis.

B C D

-

-

-

A Division of Vivaproducts

Improving Speed of Concentration

Speed of filtration is affected by several parameters including temperature, pH and protein concentration. While BJP concentrators will provide rapid filtration in most environments, some factors will slow filtration.

— Filtration speed will increase proportionally to ambient temperature. Should you require faster concentration, place the concentrator near a source of heat.

— An acidic sample with a pH of less than 5 will take longer to concentrate than a neutral sample. Adjustment to a physiological pH will result in faster filtration.

— Suspended particles will tend to foul the filter element and slow filtration speed. Prefiltration with a syringe filter or centrifugation will clarify the sample and result in faster filtration speed and improved analytical results following concentration.

— Initial protein concentration levels will have a significant effect on concentration speed. A highly dilute sample will concentrate rapidly. Once protein concentration exceeds 2 G/dL the speed of filtration will rapidly decrease. Static concentrators are not practical for concentrations above 5 G/dL.

Suggested Concentration Factors for Urine Samples

For a BJP Concentration Factor Calculator please visit: www.vivaproducts.com/calculator.html

Suggested CAP Validation Procedures

For Recommended Procedures please visit: www.vivaproducts.com/downloads/lab-procedure-performing-the-test.pdfTo download the CAP Calculation Table please visit: www.vivaproducts.com/downloads/cap-recovery-table.xls

TECHNICAL SPECIFICATIONS Concentration Capacity BJP-5 BJP-10 BJP-20 Normal Volume 5 ml 10 ml 10 ml With optional expansion reservoir (BJPA-ER20) NA NA 20 ml Dimensions For BJP-5/30 and 5/100, BJP-10/30 and 10/100, BJP-20/30 and 20/100 Width 38 mm 38 mm 45 mm Height 100 mm 100 mm 100 mm Depth 24 mm 24 mm 27 mm For BJP-5/40, 10/40 Width 147 mm 147 mm NA Height 94 mm 94 mm NA Depth 70 mm 70 mm NA Active membrane area 25 cm2 25 cm2 28 cm2

Dead stop volume 50 µl 50 µl 50 µl Materials of Construction Membrane Polyethersulfone (7,500 MWCO) Reservoir Acrylonitrile Butadiene Styrene Polymer Typical Performance Time to concentrate 10x min. 200 C Concentrate recovery % 7,500 MWCO PES BJP-5 BJP-10 BJP-20 BJP-20 BJP-5 BJP-10 BJP-20 Start volume 5 ml 10 ml 10 ml 20 ml 5 ml 10 ml 10 ml Albumin (66,000 MW) (0.25 mg/ml) 30 60 55 110** 92% 92% 92% lgG (160,000 MW) (0.25 mg/ml) 35 70 65 130** 65% 68% 68% Time to concentrate 50x min. 200 C Concentrate recovery % Albumin (66,000 MW) (0.25 mg/ml) 40 80 75 150** 90% 90% 90% lgG (160,000 MW) (0.25 mg/ml) 45 90 85 170** 56% 60% 60%** All 20 ml samples start at 0.10 mg/ml

ORDERING INFORMATION Type of Device No. Tests BJP-5 BJP-10 BJP-20 Individual device (no pipettes) 30 BJP-5/30 BJP-10/30 BJP-20/30 Individual device (with pipettes) 30 BJP-5/30P BJP-10/30P BJP-20/30P Individual device (no pipettes) 100 BJP-5/100 BJP-10/100 BJP-20/100 Individual device (with pipettes) 100 BJP-5/100P BJP-10/100P BJP-20/100P 8 test block (no pipettes) 40 BJP-5/40 BJP-10/40 8 test block (with pipettes) 40 BJP-5/40P BJP-10/40P

ACCESSORIES Plastic disposable pipettes (qty of 250) BJPA-P250 Plastic disposable pipettes (qty of 100) BJPA-P100 Plastic disposable pipettes (qty of 40) BJPA-P40 Plastic disposable pipettes (qty of 30) BJPA-P30 Expansion reservoirs for use with BJP-20 devices (qty of 10) BJPA-ER20 Acrylic stand for BJP individual units BJPA-AS

Vivaproducts • 521 Great Road • Littleton, MA 01460 • Toll-free phone (U.S. & Canada) 800-456-4633 • Fax 978-952-6143 • Outside U.S. & Canada: 978-952-6868