Biosensor dr manju jha

42
BIOSENSORS Presented by : Dr. MANJU JHA II- year Resident, M.D. 09-10- 2012

description

BIOSENSOR POWERPOINT

Transcript of Biosensor dr manju jha

Page 1: Biosensor dr manju jha

BIOSENSORS

Presented by :

Dr. MANJU JHA II-year Resident, M.D. Biochemistry J.L.N.Medical College,Ajmer.

09-10-2012

Page 2: Biosensor dr manju jha

HISTORY-1962- first description of a biosensor of : an amperometric enzyme electrode (Glucose sensor) by Clark.1969- first potentiometric biosensor : urease immobilized on an ammonia electrode to detect urea by Guilbault & Montalva.1970- ion-selective Field Effect Transistor (ISFET) by Bergveld.1975- fibre-optic sensor with immobilized indicator to measure carbon-di-oxide or oxygen by Lubbers & Optiz.1975- first commercial biosensor (Yellow Spring Instrumental Biosensor) .1975- first microbe based biosensor (first Immunosensor)..1976- first bedside artificial pancrease.1980- first fibre-optic pH sensor for in-vivo blood gases by Peterson.1982- first fibre-optic based biosensor for glucose.1983- first surface plasmon resonance (SPR) immunosensor.1984- first mediated Amperometric biosensor.1987- Blood Glucose biosensor launched by Medi-sense Exac Tech.1990- SPR based biosensor by Pharmacia BIA Core.1992- hand-held blood biosensor by i-STAT..1996- launching of Gluco-card.1998- blood glucose biosensor launched by Life-scan Fast Take.2000- nanotechnology biosensor, chip,quantum dots etc..

Page 3: Biosensor dr manju jha

What is a BIOSENSOR ?

A Biosensor may be defined as :

a compact analytical device having a biological or biologically-derived sensing element either integrated within or intimately associated with a physico-chemical transducer which is then detected by an electronic component and translated into a measurable electronic signal.

Page 4: Biosensor dr manju jha

IUPAC Definition -

Biosensor is a self-contained integrated device which is capable of providing specific quantitative or semi-quantitative analytical information using a biological recognition element (biochemical receptor) which is in direct spatial contact with a electrochemical transducer element.

Page 5: Biosensor dr manju jha

WORKING PRINCIPLE OF A BIOSENSOR

The interaction of the analyte with the bioreceptor is designed to produce an effect measured by the transducer, which converts the information into a measurable effect, such as an electrical / electronic signal.

Page 6: Biosensor dr manju jha

The interaction of the analyte (a) with the bioreceptor, which identifies the stimulus (b) is designed to produce an effect measured by the transducer (c), which converts it to an electrical signal. The output from the transducer is amplified (d), processed (e) and displayed (f).

(a)(b) (c)

(d)

(e)(f)

Page 7: Biosensor dr manju jha

Main components-

A. Biological recognition element-

It is the sensitive biological element or biological material (tissue ,microorganisms , organelles , cell receptors , enzymes, antibodies, nucleic acids, etc.) or bio- mimetic component that interacts (binds or recognises) the analyte under study.

The biologically sensitive elements can also be created by biological engineering.

Page 8: Biosensor dr manju jha

B. Transducer-

The transducer or the detector element transforms the signal resulting from the interaction of the analyte with the biological recognition element into another signal that can be more easily measured and quantified.

C. Amplifier, Microprocessor and Display-

Biosensor reader device with the associated electronics or signal processors that are primarily responsible for the display of the results.

Page 9: Biosensor dr manju jha

Biological elements and Transducers commonly used in BIOSENSORS-

Biological elements

1. Enzymes2. Antibodies3. Hormone receptors4. Cells 5. Cell organelle6. Tissues7. Membranes8. Micro organisms 9. Nucleic acids10. Biomimetic materials

Transducer elements

1. ElectrochemicalAmperometricPotentiometric

2. ElectricalConductometricIon-sensitive

3. PhotochemicalCalorimetricFluorescenceReflectance

4. OpticalFiber opticSPRLuminometricFluorometric

5. PiezoelectricAcousticMCL

UltrasonicQCM

Page 10: Biosensor dr manju jha

A. Catalytic biosensor:

kinetic devices that measure steady state concentration of a ‘tranducer - detectable species’ formed / lost due to a biocatalytic reaction.

Monitored quantities are- i. rate of product / acid ( ↑ or ↓pH) formation,ii disappearance / consumption of reactant,iii. inhibition of reaction ,etc.

Biological recognition element— (Classification acc to Biological Signalling method)

Page 11: Biosensor dr manju jha

Biocatalysts used are----

1.Enzymes- Enzyme-based biosensors use their catalytic activity and binding capabilities for specific detection . The catalytic activity of the enzymes

provides these types of biosensors with the ability to detect much lower limits than with normal binding techniques.

This catalytic activity is related to the integrity of the native protein structure.

Page 12: Biosensor dr manju jha

Most common – Glucose oxidase, Urease, Alcohol oxidase etc.

Commercial example is GLUCOSE SENSOR using oxidase (GOD) .

Clarks (1962)

Glucose + O2 -- glusoce oxidase Gluconic acid + H2O2 (1) 

H2O2 O2 + 2H+ + 2e- (3) at anode

O2 + 4H+ + 4 e- 2H2O (2) at cathode-and current flows.   There are three measurement routes-

-pH change (acid production) - O2 consumption (fluorophore monitor)- H2O2 production (electrochemical)

Page 13: Biosensor dr manju jha

2. Micro-organisms – Microorganisms such as bacteria and fungi can be used as biosensors to detect specific molecules or the overall state of the

surrounding environment. Cell behaviour such as cell metabolism,

cell viability, cell respiration, and bio-luminescence can be used as indicators for the detection. Furthermore, proteins that are present in cells can also be used as bio-receptors for the detection of specific analytes.

Rhodococcus erythropolis (in collagen) used to measure BOD.

Page 14: Biosensor dr manju jha

3. Cell / Tissue samples.

Use of cell as a biosensor occurred in 1977 by Rechnitz .

Rechnitz coupled Streptococcus faecium on the surface of an ammonia gas sensing membrane electrode .

This Rechnitz electrode was capable of detecting the amino acid ARGININE.

4. Organelles – Mitochondra , Cell walls etc.

5. Membranes .

Page 15: Biosensor dr manju jha

6. Bio-mimetic materials -

A bio mimetic biosensor is an artificial or synthetic sensor that mimics the function of a natural biosensor . These can include aptasensors , where apta

-sensors use aptamers as the biocomponent.

Aptamers are synthetic single stranded nucleic acid that can be designed to

identify or recognize amino-acids, oligosaccharides , peptides , and proteins . Aptamers have high affinity , high selectivity,

cheaper & easy to synthesize .

Page 16: Biosensor dr manju jha

B. Affinity biosensors: device in which receptor molecules bind analyte molecules

causing physicochemical change that is detected by a transducer.Receptors used are-

1. Antibody / Antigen (Immunosensor) -The high specificity between an antibody and an antigen can be utilized in this type of sensor technology. Biosensors utilizing this specificity must ensure that binding occurs under conditions where nonspecific interactions are minimized . Binding can be detected either through fluorescent labelling or by observing a refractive index or reflectivity changes.

2. Hormone receptor/Antagonist-

Page 17: Biosensor dr manju jha

3. Nucleic acid --

The complementary relationships between nucleic acid ‘bases’ in the DNA form the basis of specificity in nucleic acid based biosensors.

These sensors are capable of detecting trace amounts of microorganism DNA by comparing it to a complementary strand of known DNA.

By unwinding the target DNA strand, adding the DNA probe, and annealing the two strands , the probe

will hydrolyze to the complementary sequence on the adjacent strand . If the probe is tagged with a fluorescent compound , then this annealing can be visualized under a microscope.

eg. DNA Chip.

Page 18: Biosensor dr manju jha

General DNA biosensor scheme--Target DNA is captured at the recognition layer (A),and the resulting hybridization is transduced into a measurable electronic signal (B).

It is used for genome mutation detection & clinical diagnosis.

(Capture probe)

(Target probe )

Page 19: Biosensor dr manju jha

I. Recognition element immobilization (Immobilization of biological receptor)-

Biological receptors, i.e. enzymes, antibodies, cells or tissues with high biological activity, can be immobilized in a thin layer at the transducer surface by using different procedures.

(a) Entrapment behind a membrane - viscous aqueous solution trapped by membrane permeable to analyte.

(b) Entrapment of biological receptors within a polymeric matrix- Membranes— Cellophane, Cellulose acetate, Polyurethane membranes . Gel entrapment– Agarose, Gelatin, Agar gel , Polyacrylamide gel. Microencapsulation- Encapsulation inside Liposomes, or absorbed in fine carbon particles that are incorporated in a gel or membrane

Biosensor construction--

.

Page 20: Biosensor dr manju jha

(c) Entrapment of biological receptors within self-assembled monolayers (SAMs)

or bilayer lipid membranes (BLMs).

(d) Adsorption: direct adsorption onto membrane or transducer;

can also be adsorbed onto pre -adsorbed proteins .

(e) Covalent binding (via –COOH, -NH2, -OH ) , or cross linking (via glutaraldehyde ) to transducer or membrane surface.

Receptors are immobilized either alone or are mixed with other proteins, such as bovine serum albumin (BSA), either directly on the transducer surface, or on a polymer membrane covering it .

Page 21: Biosensor dr manju jha

II. Inner & Outer membranes--

These serve three important functions -

(a) Protective barrier --

Membrane prevents large molecules, such as proteins or cells of biological samples, from entering and interfering with the reaction layer. It also reduces leakage of the reacting layer components into the sample solution. This function of the outer membrane is important, for example, for implanted glucose sensors, since its glucose oxidase is of non-human origin and may cause immunological reactions.

Page 22: Biosensor dr manju jha

(b) Diffusional outer barrier for the substrate – The thinner the membrane, the shorter the

biosensor response time so such a diffusional barrier also makes the sensor response independent of the amount of active enzyme

present and improves the sensor response stability.

(c) Biocompatible and biostable surfaces –Biosensors are subject to modifications when they are in direct contact with biological tissues or fluid , i.e. implanted in vivo or, more generally, in biologically active matrices, such as cell cultures.

Page 23: Biosensor dr manju jha

BIOCOMPATIBILITY-

If the implantation of the biosensor does not materially affect the normal functioning of the host medium and if the medium does not materially affect the normal operation of the biosensor , then the biosensor is considered to be biocompatible.

Page 24: Biosensor dr manju jha

Detection or Measurement mode (Transducer)—

1. Electrochemical- translate a chemical event to an electrical event by measuring current passed.

a. Amperometric- most common : movement of electron produced in a redox reaction.

In 2000 a wearable non-invasive Glucose monitor has been introduced.SONY has developed a biofuel cell using sugar as the fuel and enzymes as catalysts to power a Walkman.

 b. Potentiometric - electrical potential producing phenomenon.

Page 25: Biosensor dr manju jha

2. Electrical –

a. Conductometric

b. Ion-sensitive – The use of ion channels has been shown to offer

highly sensitive detection of target biological molecules.

By imbedding the ion channels in supported or tethered bilayer membranes (t-BLM) attached to a gold electrode, an electrical circuit is created.Capture molecules such as antibodies can be bound to the ion channel so that the binding of the target molecule controls the ion flow through the channel . This results in a measurable change in the electrical conduction which is proportional to the concentration of the target.

Page 26: Biosensor dr manju jha
Page 27: Biosensor dr manju jha
Page 28: Biosensor dr manju jha

3. Photochemical – translate chemical event to a photochemical event,

measure light intensity & wave length.

a. Calorimetric – measure heat changes of a reaction.

Isothermal , Heat conduction , orIsoperibol .

b. Fluorescence - property of fluorescence is used.eg. DNA Microarray

c. Reflectance – use the property of reflection which is detected by photo detector.

Page 29: Biosensor dr manju jha

4. Optical –

There is light output during the reaction or a light absorbance difference between the reactants &

products. It uses property of fluorescence , phosphorescence , refraction, and dispersion spectrometry etc.

a. Fiber optic (optrode/optode)

b. Surface plasmon resonance (SPR)

c. Luminometric

d. Fluorometric

Page 30: Biosensor dr manju jha

5. Piezoelectric–

These devices are used to detect the specific angle at which electron waves are emitted when the

substance is exposed to laser light or crystals, such as quartz ,which vibrate under the influence of an electric field .

It translates a mass change from a chemical adsorption event to electrical signal. The change in frequency is proportional to the mass of absorbed material.

a. Acoustic (Surface acoustic wave)b. Microcantileverc. Ultrasonicd. Quartz crystal microbalance (QCM)

Page 31: Biosensor dr manju jha

Examples—

Quantum dots (QD's)

QCMMicrocantileverNanoparticlesNanotubes(CNTs)Nanowires

Nanotechnology will enable us to design sensors that are much smaller, less power hungry, and more sensitive than current micro- or macro sensors.

Nanobiosensors

Page 32: Biosensor dr manju jha

Ideal Biosensor Characteristics

1. Sensitivity: high ΔS/ Δcanalyte (S = signal) 2. Simple calibration (with standards) 3. Linear Response: ΔS/ Δcanalyte constant over large concentration range .4. Accuracy must be there. 5. No hysteresis—signal independent of prior history of measurements . 6. Response time / Recovery time - less . 7. Selectivity—response only to changes in target analyte concentration

Page 33: Biosensor dr manju jha

Ideal Biosensor Characteristics –

8. Long-term Stability—not subject to fouling, poisoning, or oxide formation that interferes with signal;

prolonged stability of biological molecule .9. Dynamic Response-rapid response to variation in analyte concentration . 10. Biocompatibility—minimize clotting, platelet interactions, activation of complement when in direct contact with bloodstream.11. Must be cost effective , smaller or portable in size .

Page 34: Biosensor dr manju jha

Applications1. Medical-a. Biosensors are used in both clinical and laboratory use in medical care. Glucose monitoring in diabetes patients .

Medtronic glucose sensor - implants in major vein of heart.

b. Tumor cells are used as biosensors to monitor the susceptibility of chemotherapeutic drugs.

c. Routine analytical measurement of folic acid, biotin, vitamin B12 and pentothenic acid .

d. Micro- and nanoscale biosensors—Genome mutation detection , cancer detection & clinical diagnosis.Bacterial-UTI , Human Immunodeficiency Virus (HIV) Detection, Hepatitis and Anthrax detection.

Page 35: Biosensor dr manju jha

2. Bio / Pharmaceutical research-

a. Quality assurance

b. Study of biomolecules and their interaction

c. Protein engineering .

d. Drug discovery , evaluation monitor the manufacturing of biological activity of new compounds (research field).

e. Biosensors are used for measuring concentration of various metal ions by specific protein concentration or by using genetically modified organisms.

f. Aptamers are used for the detection of proteins –Thrombin , IgE , HIV – tat protein , Lysozyme , Abrin toxin.

Page 36: Biosensor dr manju jha

2. Bio / Pharmaceutical research-

g. Biosensors are used in monitoring of the glutamate and acetyl choline , which is the main cause in neurodegenerative diseases.

h. Microbiology: bacterial and viral analysis .

i. Biosensors are used in analysing micro dialysis samples.

j. Biosensors are used in biotechnological process such as to determine proteins or peptides.

k. Biosensors are also used in determining intracellular proteins and also plasmids.

l. Detection of cancer biomarkers – CEA , PSA , CA-125 &Tumour necrosis factor .

Page 37: Biosensor dr manju jha

3. Industrial / Agricultural –

a. Biosensors used in process control will be able to measure materials present in the process.

b. Use of biosensors in industry will improve manufacturing techniques, this will allow for usage of wider variety of sensing molecules.

c. Biosensors are used in controlling the industrial processes.

d. Microbial sensor measure Ammonia & Methane.

Page 38: Biosensor dr manju jha

4. Environmental-

a. Biosensors are used in detecting environmental pollutants and monitoring of Mines, Industries and toxic gases.

b. Biosensors are used in the BOD measurement during waste water treatment.

c. Biosensors are used in the detection of poly aromatic hydrocarbons present in water.

d. Environmental applications e.g. the detection of pesticides and river water contaminants such as heavy metal ions.

e. Detection and determination of organophosphates .

Page 39: Biosensor dr manju jha

5. Food industry-

a. Quality assurance in food industries , ex. E. Coli, Salmonella.

b. Food & drink production analysis.

c. Biosensors are used for detection of food freshness marker determining parameters in wine industry.

d. Determination of drug residues in food, such as antibiotics and growth promoters, particularly meat and honey.

e. Detection of toxic metabolites such as mycotoxins.

Page 40: Biosensor dr manju jha

6. Biodefence-

a. Detection of pathogens (SARS in 2003).

b. Remote sensing of airborne bacteria / virus , e.g. In counter bio-terrorist activities.

c. Detection system for biological welfare agent eg. Bacillus anthracis (anthrax) spores.

d. Determining levels of toxic substances before and after bioremediation.

e. Crime detection.

Page 41: Biosensor dr manju jha

?

Page 42: Biosensor dr manju jha

THANK YOU