Bio Monitoring 01
description
Transcript of Bio Monitoring 01
1
Overview of system
ECG sensor
Integrated SpO2/Temperature sensor
Central processing unit
Wireless Body Area Network (WBAN) Bluetooth
Sentinel events / Request to view real-
time vital signs
GSM
Wearer’s PDA phoneDoctor’s PDA phone
1. Four physiological signs, e.g., Electrocardiogram (ECG), SPO2, body temperature and blood pressure will be continuously acquired or derived from two wireless sensor node - ECG sensor and integrated SpO2/Temperature sensor node. 2. At the wearer’s location, the PDA-based monitor can be used to acquire real-time and continuous waveform.3. Upon detection of sentinel events, the abnormal vital signs would be sent wirelessly through GSM to doctor/caregiver’s PDA phone.4. Doctor/caregiver can promptly view and analyze the received abnormal vital signs forfurther treatment strategy decision
2
Electrocardiogram (ECG)
3
Wireless ECG sensor
As the heart undergoes depolarization and repolarization, electrical currents spread throughout the body because the body acts as a volume conductor. The electrical currents generated by the heart are commonly measured by electrodes placed on the body surface and the resulting tracing is called an electrocardiogram (ECG).
What is ECG signal?
volume conductor
Entire process of depolarization and repolarization
4
Wireless ECG sensor What is ECG signal?
P wave: the sequential activation (depolarization) of the right and left atria
QRS complex: right and left ventricular depolarization
ST-T wave: ventricular repolarization
5
Wireless ECG sensorConventional ECG sensor1. Standard ECG: 12 leads, specially for hospital use.
Advantage: Standard and comprehensive ECG signalsDisadvantage: Short sampling and can not detect irregular or intermittent arrhythmia
2. ECG holter: 3 or 5 leads for long-term monitoringDisadvantage: Receive power from an electrical outlet, outstanding heavy and not
essentially portable or wearable
Objective:1. A medical grade and battery-supplied
Lead I ECG sensor 2. Can be unobtrusively worn over a
period of several days3. Continually log heart rate data, provide
detection of life threatening events, such as arrhythmia
6
Wireless ECG sensorInput noisy ECG signal
Low-Pass FIR FilterRemove high frequency
noise, such as power line interference
High-Pass FIR Filter
Remove low frequency noise, such as EMG
noise (muscle), electrode contact noise
U3 algorithm Remove base line drift
QRS peak detection
Characterization of QRS width, height, R-R interval
Arrhythmia detection
0 500 1000 1500 2000
Samples
ECG Signal
U3 Response
ECG Electrodes
Battery
Bluetooth module
MCU board
General processing
Fabricated ECG sensors based on TI MSP430FG439
Two square pads used as electrodes for testing
7
Detectable ECG abnormalitiesRate (bpm) and R-R Interval (sec)
Normal ~60-90 bpm corresponds to ~ 0.66 – 1sec Tachycardia > 90 bpm corresponds to < 0.65 sec Bradycardia < 60 bpm corresponds to > 1 sec
Rhythm Difference between the Longest and Shortest R-R
interval detected within 3 sec is Irregular if > 0.12 sec,which indicates
AV block Atrial Fibrillation
QRS Width Normal = 0.06 – 0.10 sec 0.1 – 0.12 sec indicates
Wolff-Parkinson-White syndrome (WPW) Non-specific intraventricular conduction delay (IVCD) Incomplete right or left bundle branch block (RBBB or LBBB)
> 0.12 sec indicates Complete LBBB or RBBB Ventricular tachycardia
Q Wave Width > 0.04 sec or/and height > 25% of R’s height
indicates Myocardial Infarction (MI)
QRS Complex
0.16 s
RBBB
Normal Regular
TachycardiaIrregular
0.59 s0.89 s
0.65 s
BradycardiaIrregular
0.91 s0.91 s
1.08 s 1.08 s
Typical QRS
Q
R
S
Abnormal QRS Appear in MI and Hyperkalemia
Q
R
S
Q
R
Q is 37.5% of R
SpO2
8
AV (Atrio-Ventricular) block
ECG abnormalities
Atrial fibrillation
Exit
9
Wolff-Parkinson-White syndrome (WPW)
ECG abnormalities
Complete left bundle branch block (LBBB) Complete right bundle branch block (RBBB)
Exit
10
Saturation of Arterial Oxygen- SpO2
11
Blood and Hemoglobin (Hb)Circulation
Common carotid artery
Superior vena cava
Pulmonary vein
Inferior vena cava
Pulmonary artery
1 red blood cell: ~ 265 millionmolecules of hemoglobin
1 hemoglobin molecule: 4 heme and 4 globin units.
Each heme and globin unit can carry 1 molecule of oxygen
Hemoglobin changes color: Oxygenated (HbO2): bright red Deoxygenated (Hb): dark red
This color change is used to measure hemoglobin oxygen saturation.
12
Blood and HemoglobinDiffusion of oxygen
Feeder arteriole
Precapillary sphincter
Drainage venule
Tissue cells
Shunt True capillary
Once blood is oxygenated, although it may pass through oxygen depleted tissue, oxygen does not diffuse until it reaches the capillaries with one cell thickness in the wall.
Oxygen diffuses into the interstitial fluid and into the cells.
True capillary
Lymph capillary
Arteriole end
Venule end
Φ6-8μm
13
Definition of SO2
2
2
22
2
100% 100%HbO
HbO Hb
CHbOSO
HbO Hb C C
SO2: Saturation of Oxygen
Percent of oxygen present in the hemoglobin present in blood
SaO2: Arterial oxygen saturation (in arterial blood)
Normal range for a health adult: 95 - 100 %
SpO2: Oxygen saturation derived from pulse oximetry.
Non-invasive method
SvO2: Venous oxygen saturation (in venous blood)
The normal SvO2 is 75%, which indicates that under normal conditions, tissues extract 25% of the oxygen delivered.
14
Principle of SpO2
( )0
ClI I e
Beer’s law (Beer-Lambert’s law or Bouguer’s law)
lCε(λ): absorption coefficient of the substance at a specific wavelength λ.
C: concentration
l: optical path length
I0
I
1 1 2 2( ) ( )0
C l C lI I e lC1
I0
I
C2
15
Principle of SpO2
2 2 2 2 20
Hb aHb HbO aHbO t HbO vHb HbO vHbOC l C l A m C n C nI I e e e
Arterial Tissue Venous
Beer’s law for measurement of oxygen saturationεHb: absorption coefficient of HbεHbO2: absorption coefficient of HbO2
In arterial blood:
CaHb: concentration of Hb
CaHbO2: concentration of HbO2
In venous blood:CvHb: concentration of HbCvHbO2: concentration of HbO2
l
n
I0
I
Arterial blood
Tissue and capillaries
Venous blood
m
( l is variable in pulsed arterial blood)
16
2 2( ) ( )( ) ( )0
Hb Hb HbO HbODC DC DC C C lC lI I e e
timeOne cardiac cycle
Other tissue
Venous bloodNon-pulsating arterial blood
Pulsating arterial blood
I0
I
DC
AC
I0
IDC
Δl
( ) ( )0
DC DC DCC lDCI I e
2 2( ) ( )Hb Hb HbO HbOC C l
DC
IN e
I Normalization:
Principle of SpO2
Pulsation of the blood
Photodiode
LED
17
2 2 1
1
( 1) ( 1)Hb Hb HbO HbOC C lN e
Principle of SpO2
Ratio of normalized signals
Wavelength λ1,
2 2 2
2
( 2) ( 2)Hb Hb HbO HbOC C lN e
Wavelength λ2,
1
2
1 2 1 2 1
2 2 2 2 2
ln ( ) ( )
( ) ( )lnHb Hb HbO HbO
Hb Hb HbO HbO
N C C lR
C C lN
Ratio
When the optical path lengths for the two wavelengths (λ1 λ2) are equal, Δlλ1 = Δlλ2
2 1 2
22 1 2 1 2 2 2
( ) ( )100%
( ) ( ) ( ) ( )HbO Hb Hb
Hb HbO Hb HbO Hb HbO
C RSpO
C C R
LED2
Photodiode
LED1
18
Principle of SpO2Criteria for the choice of wavelengths
500 600 700 800 900 1000
16
12
8
4
0
Absorptivity
Wavelength (nm)
εHbO2
εHb
660nm 940nm
The red skin pigmentation absorbs a great amount of light at wavelengths shorter than 600 nm.
wavelength > 600 nm
Large differences in the absorption coefficients of Hb and HbO2
To get high sensitivity
Flatness of absorption spectra
To get high stability
Hemoglobin absorbance spectra
19
Principle of SpO2SpO2 and R
0 0.5 1 1.5 20
0.2
0.4
0.6
0.8
1
1.2
R
SpO
2660 nm / 940 nm
20
Measurement of R
1
2
lnln ln
lnRmin IRmin
Rmax IRmax
N I IR
I IN
Red transmittance Infrared transmittance
Lig
ht
inte
nsi
ty
Lig
ht
inte
nsi
ty
IRmin
IRmax IIRmax
IIRmin
Logarithmic method
Photo-plethysmogram (PPG) waveforms
21
Measurement of R
1 1 ( )0 0
l L tI I e I e
Lig
ht
inte
nsi
ty
Imin
Imax
Derivative method
( )l L t
1
1
( )0 ( )
0 1 1
( ) ( )L t
L td I edI dL t dL t
I e Idt dt dt dt
1
( )dI dt dL t
I dt
1 2 1 2 1
2 2 2 2 2
( ) ( )
( ) ( )Hb Hb HbO HbO Rmax Rmin IRmax IRminR IR
R IR Rmax Rmin IRmax IRminHb Hb HbO HbO
C C I I I IdI dt dI dtR
I I I I I IC C
1 2 1 2( ) ( )0
Hb Hb HbO HbOC C lI I e
1 1 2 1 2( ) ( )Hb Hb HbO HbOC C Let
We have
Its derivative is
Δt
At point M, 2 2min min max min maxI I I I IdI I
dt t t t
one cardiac cycle
(Imax+Imin)/2M
2max minI I I
For λ1,
22
Measurement of R
Logarithmic method Derivative method
Complicated calculation
Requires powerful computation capability
Simple calculation
Suitable for real-time computation in microcontroller applications
ln lnRmin IRmin
Rmax IRmax
I IR
I I
Comparison
Rmax Rmin IRmax IRmin
Rmax Rmin IRmax IRmin
I I I IR
I I I I
23
History of SpO2 1982, pulse oximetry was developed by Nellcor (William NEw,
Jack LLoyd and Jim CORenman).
1983, pulse oximetry was introduced into the US operating room market.
By 1987, pulse oximetry was included in the standard of care for the administration of a general anesthetic in the US.
Now, pulse oximetry is widely used in hospitals.
The current researches are focused on:
Noise reduction and signal processing reduction of motion artifact noise
Venous pulsation
Low perfusion
Portable and long-time wearable SpO2
Sensor fusion
24
SpO2 Products
25
Limitation of SpO2• Motion artifact• Pulsed venous blood• Low perfusion states• Abnormal hemoglobins (primarily carboxyhemoglobin [COHb]
and met-hemoglobin [metHb])• Intravascular dyes• Exposure of measuring probe to ambient light during
measurement• Skin pigmentation• Inability to detect saturations below 83% with the same
degree of accuracy and precision seen at higher saturations• Inability to quantitate the degree of hyperoxemia present
26
Limitation of SpO2Motion artifact
27
Limitation of SpO2• The amplitude of the plethysmographic wave form is directly proportional to the
vascular distensibility, which is significantly greater in the arterial system.• However, the venous signal can have significant impact on the calculation of the
SpO2 if it reaches the threshold for a pulsation.
Venous pulsation
1
1
( 1) lN e
2
2
( 2) lN e
1
2
1
2
ln ( )
( )ln
NR
N
1 1
1
( 1) ( 1)' l nN e
2 2
2
( 2) ( 2)' l nN e
1
2
1 1 1 1
2 2 1 1
ln ' ( ) ( )'
( ) ( )ln '
N n lR
n lN
1 2l l 1 2l l 1 2n n
No venous pulsation: With venous pulsation:
'R R 2 2 'SpO SpO
1 2( ) ( ) Assuming CHbO2= 75%, CHb= 25% in venous blood, for 660nm/940nm LEDs,
28
Limitation of SpO2
• The venous pulsation cause the measured SpO2 value lower than the real value.
Venous pulsation
SpO2 measured from fingertip
0 20 40 60 80 100 1200.5
0.6
0.7
0.8
0.9
1
1.1
Time (sec)
SpO
2
Venous blood is accumulated in fingertip
Venous blood is minimized
29
Limitation of SpO2
• Received light intensity is determined by position of LED emitter.
Low perfusion
Cross section of fingertip
LED position 1(correct position)
Bone
LED position 2 (inappropriate position)
Artery
Vein
Position 1: normal signal Position 2: weak and noisy signal (low perfusion)
Red
Infrared
Red
Infrared
• Signal from photodiode (Photo-plethysmogram waveform,PPG):
30
Measurement Position
Fingertip Earlobe
Severe motion artifact noise
Problem of venous pulsation
Problem of low perfusion (inappropriate sensor position -> weak signals)
Long-time wearable: it may influence human activity
31
Prototype of SpO2
MCU board
Thermopile
SpO2 Probe
Bluetooth
32
Prototype of SpO2Hardware
LED(660nm) LED(940nm)
Photodiode
Mic
roco
ntro
ller
3.2mm
Red/IR bicolor LED:
Photodiode (light converter):Light-to-Frequency
Light-to-Voltage
Light intensity
Vo
lta
ge
(digital output)
(analog output)
light intensity: low high
PDA/Phone
Blu
etoo
th
Fingertip/earlobe
33
Signal processing
0 20-20-10
0
Frequency (Hz)
Magnitude Response (dB)
General processing
IR
red
MAX
MIN
Interval
U3
IR
RED
Band-pass
Segmentation
IR
Band-pass
Max Min extraction
SpO2
0 20-20-10
0
Frequency (Hz)
Magnitude Response (dB)
Signal processing
One cardiac cycle
FIR filter
Raw data with high-frequency noises (ambient light, circuit, etc)
Smooth data without high-frequency noises Motion
artifact
Venous pulsation
Low perfusion
34
Blood Pressure
35
Blood PressureDefinition of blood pressure
Blood pressure is the pressure exerted by the blood at the normal direction to the walls of the blood vessels.
Blood pressure usually refers to systemic arterial blood pressure, i.e., the pressure in the large arteries delivering blood to body parts.
Systolic pressure is defined as the peak pressure in the arteries during the cardiac cycle.
Diastolic pressure is the lowest pressure at the resting phase of the cardiac cycle.
Blood pressure
36
Blood PressureKorotkoff method (1905)
No sound when blood is flowing through smooth vessel (normal condition)
Sound is generated when blood is passing through interrupt changed cross section.
1) The artery is completely occluded. No blood flow, no sound.
2) Sound is listened when blood just starts to flow in the artery.
The first sound: cuff pressure = systolic BP
3) Silent when cuff pressure drops below the diastolic blood pressure.
Cuff pressure = diastolic BP
Add pressure to cuff
Release pressure slowly
Release pressure slowly
Flow
37
Blood Pressure• Oscillometric method is functionally the same as for the auscultatory
method. It also requires a cuff.
• An electronic pressure sensor (transducer) fitted in the cuff detects blood flow.
Oscillometric method (1970)
1) Artery is completely occluded: cuff pressure > systolic BP2) Blood flow is unimpeded: cuff pressure < diastolic BP
cuff pressure is constant.
3) Blood flow is present: diastolic BP < cuff pressure < systolic BP
cuff pressure varies periodically in synchrony with the cyclic expansion and contraction of the artery, i.e., it oscillates.
Blood Pressure
Time cuff pressure: constant oscillate constant
Cuff
38
Blood Pressure21
2Fd mv mgh
Cuff-less method using PTT (pulse transit time)
m
F
h
d
v
F BP a
21
2
m mBP v gh
ad ad
m
ad
a: cross section of artery
ΔBP: pressure difference
ρ: density of blood, 1035 kg/m3
g: 9.8 m/s2
dv
PTT PTT: duration while m moves distance d
21 1
0.7 0.7 2
BP dBP gh
PTT
0.6h m
Examples: 0.4h m 0.4d h m (1) Heart - Fingertip
(2) Heart - Fingertip
(3) Heart - Earlobe
Systolic
39
Blood Pressure
0 500 1000 1500 2000 2500 30001
1.5
2
2.5
3
3.5
4x 10
4
PTT
ECG
0 500 1000 1500 2000 25002.3
2.35
2.4
2.45
2.5
2.55
2.6
2.65
2.7x 10
4
PTT extracted from PPG(SpO2) and ECG
(2)
(3)
(1)
PPG graph:Point 1: Greatest light intensity -> minimum blood volume in arterySlope 2: Fresh arterial blood is filling artery due to heart contraction. Point 3: Lowest light intensity -> maximum blood volume
Ligh
t in
tens
ity
Time (ms)
PPG
40
Demonstration
SpO2 & Temperature data packet
GSM
Bluetooth
Bluetooth
Bluetooth
ECG Electrodes
SpO2 ProbeLight-to-
frequency photodetector
IR/R LED driver and control
TI MSP430F1611
Thermopile
ECG Sensor
Integrated SpO2 & Temperature Sensor
Central Processing
Unit
(Deriving blood pressure and saving all vital
signs)
ECG data packet
Physiological signs acquisition module
Real-time Display
Alarm
Local storage
Patient’s PDA phone
Offline data display
Alarm
Local storage
Doctor’s PDA phone Sentinel vital signs
Command for real-time display / Sentinel events
Lead ITI MSP430FG439 Analog front-end
Block diagram of the implemented MEMSWear-Biomonitoring System
ECG data packets (bytes)
E GEight ECG signals Heart beat type
Peak position
SpO2 data packets (bytes)
SIR RED
Peak/non-Peak
T
Temperature
O
Heat beat rate
Sampling rate: ECG (512Hz), SpO2 (64Hz)
41