Honest Family Dental - Affordable Dental Implants | Dental Implants Austin
Bio Compatibility of Polymer Implants
-
Upload
cascarudo-fc -
Category
Documents
-
view
229 -
download
0
Transcript of Bio Compatibility of Polymer Implants
-
7/31/2019 Bio Compatibility of Polymer Implants
1/79
BIOCOMPATIBILITY OF POLYMER IMPLANTS
FOR MEDICAL APPLICATIONS
A Thesis
Presented to
The Graduate Faculty of The University of Akron
In Partial Fulfillment
of the Requirements for the Degree
Master of Science
Christopher M. Brendel
August 2009
-
7/31/2019 Bio Compatibility of Polymer Implants
2/79
ii
BIOCOMPATIBILITY OF POLYMER IMPLANTS FOR MEDICAL APPLICATIONS
Christopher M. Brendel
Thesis
Approved: Accepted:
________________________________ ________________________________
Advisor Department ChairDr. Daniel Ely Dr. Monte Turner
________________________________ ________________________________
Faculty Reader Dean of the CollegeDr. Ronald Salisbury Dr. Chand Midha
________________________________ ________________________________
Committee Member Dean of the Graduate SchoolDr. Qin Liu Dr. George Newkome
________________________________
Date
-
7/31/2019 Bio Compatibility of Polymer Implants
3/79
iii
ABSTRACT
Two separate experiments were performed to test the biocompatibility of
amphiphilic conetwork polymers for medical applications. In experiment 1, four
different types of polymers were tested in the liver and brain of 20 rats to determine
which polymer would be best suited for intervertebral disc repair or replacement. The
samples used for the brain and liver injections were Monomer: 1-cyanoacryl-2,4,4-
trimethylpentane(TMP-CA), Polymer: cyanoacrylate terminated tri-arm star PIB [(PIB-
CA)3, Initiator: N,N-diethyl telechelic [(PIB-NEt3)3] PIB, and Monomer + Polymer
Combination: cyanoacrylate-telechelic 3-arm star polyisobutylene [(PIB-CA)3] + 2,2,4-
trimethylpent-1-cyanoacrylate (pTMP-CA). The solid samples implanted into the liver
were Monomer: poly(1-cyanoacryl-2,4,4-trimethylpentane)poly(TMP-CA), Polymer:
cross linked [(PIB-CA)3] polymer, and Monomer + Polymer Combination: copolymer
of [(PIB-CA)3] + 2,2,4-trimethylpent-1-cyanoacrylate (pTMP-CA).
The rats were divided into three groups. Group number one had four different
honey-consistency gel polymers injected into the liver and a saline injection for
control. Group number two was subjected to the same injections, but this time the site of
injection was the brain. Group number three had three types of solid polymers surgically
-
7/31/2019 Bio Compatibility of Polymer Implants
4/79
iv
implanted into the liver. Tissues were then examined histologically to determine if any
damage had occurred. A polymer sample that resulted in little or no significant tissue
damage would be a good candidate for intervertebral disc replacement and/or repair.
In experiment 2, two types of polymers were used; a polyisobutylene polymer as
well as a Bionate sample for a control, (to determine the biocompatibility of the polymer
for a potential pacemaker coating). Both were implanted on the peritoneal wall of eight
rats. The rats were divided into two groups of four. Group number one had four
polyisobutylene polymer strips sutured to the peritoneal wall. Group number two had
four Bionate strips also implanted on the peritoneal wall. Tissue at the implant site was
removed and examined histologically to determine if any damage had occurred. A
polymer sample that resulted in little or no significant tissue damage was considered a
good candidate for a potential pacemaker coating.
-
7/31/2019 Bio Compatibility of Polymer Implants
5/79
v
ACKNOWLEDGEMENTS
I would like to thank everyone who assisted me with this experimental
investigation over the course of the past two years. Without their help, my research
efforts which have culminated into this thesis would not have been possible. The
following individuals trust and support guided me throughout the laboratory work,
research, and production of this paper, of which I am most appreciative and thankful.
The first individual that I would like to thank is Dr. Kennedy. His expertise,
innovation, and enthusiasm in the field of polymer science have afforded me the
opportunity to participate in biomedical research since the final semester of my
undergraduate career. It has been through this work, that I have realized my passion
towards working in biomedical research. I am excited about the prospect of pursuing my
Ph.D. in Polymer Science in the Fall of 2009, and look forward to any future
collaboration with Dr. Kennedy. I would also like to thank Suresh Jewrajka, Dr.
Kennedys research assistant, for synthesizing the polymer samples that were used in this
study.
My gratitude is also extended to Gail Dunphy and Shannon Boehme for their
assistance in learning the proper technical and surgical protocols in the lab. Without their
insight and guidance this would have been an incredibly cumbersome task. I have gained
-
7/31/2019 Bio Compatibility of Polymer Implants
6/79
vi
a great deal of knowledge and techniques from them, and am truly grateful for all of their
efforts throughout this process.
I would also like to acknowledge Amy Kladar for her work as an undergraduate
assistant, which was invaluable to my research.
Finally, without the quality of time devoted, counseling, consideration, and
mentoring of Dr. Ely from the last semester of my undergraduate career throughout my
graduate studies, I would not be where I am today. Its hard to find the right words to
adequately express my gratitude to Dr. Ely who has played such an instrumental role in
my education and career exploration. It is thanks to Dr. Ely that I will graduate with a
clear career direction. Dr. Elys support and faith in me has been paramount to my
success and I am very fortunate to have had such a distinguished and caring advisor to
guide me through not only my research and graduate program, but also to help me realize
the full potential of the work I can accomplish in the future.
I would also like to thank my family and friends for their unwavering belief in my
ability, which means so very much to me.
As I complete my graduate coursework and embark on the new journey of earning
my Ph.D. in Polymer Science, I am thankful to have so many highly esteemed
professionals as colleagues. It is my sincere hope that I will be able to work with them in
some capacity in the future.
-
7/31/2019 Bio Compatibility of Polymer Implants
7/79
vii
TABLE OF CONTENTS
Page
LIST OF TABLES ...............................................................................................................x
LIST OF FIGURES ........................................................................................................... xi
CHAPTER
I. INTRODUCTION ............................................................................................................1
Hypotheses and Objectives ......................................................................................3
II. LITERATURE REVIEW ................................................................................................5
Medical Applications for Polymer Biomaterials .....................................................6
Biocompatibility ....................................................................................................10
Biofilms..................................................................................................................12
Biofilm Prevention .................................................................................................13
Amphiphilic Conetworks .......................................................................................14
Polymer Biodegradation ........................................................................................15
III. MATERIALS AND METHODS .................................................................................17
Experiment 1: Polymer Biocompatibility for Intervertebral Disc
Repair/Replacement ...............................................................................................17
Surgical Procedure .................................................................................................17
Polymers used for Brain and Liver Injections .......................................................18
-
7/31/2019 Bio Compatibility of Polymer Implants
8/79
viii
Solid Polymers Implanted in the Liver ..................................................................19
Tissue Removal and Histology ..............................................................................19
Experiment 2: Polymer Biocompatibility for Pacemaker Coating ........................20
Surgical Procedure .................................................................................................20
Polymer and Bionate Control Implants..................................................................21
Tissue Removal and Histology ..............................................................................21
Statistical Analysis for Experiments 1 and 2 .........................................................22
IV. RESULTS ....................................................................................................................24
Experiment 1: Polymer Biocompatibility for Intervertebral DiscRepair/Replacement ...............................................................................................24
Injection/Implant Scoring Averages for All Categories ........................................24
Wall Thickness for Liver Injections ......................................................................25
Irregular Cells for Liver Injections ........................................................................25
Lymphocytes for Liver Injections ..........................................................................25
Wall Thickness for Brain Injections ......................................................................26
Irregular Cells for Brain Injections ........................................................................26
Lymphocytes for Brain Injections .........................................................................27
Wall Thickness for Liver Implants ........................................................................27
Irregular Cells for Liver Implants ..........................................................................27
Lymphocytes for Liver Implants ...........................................................................28
Survival Rate for Injections/Implants ....................................................................28
Total Scores for Liver Injections ...........................................................................28
Total Scores for Brain Injections ...........................................................................29
-
7/31/2019 Bio Compatibility of Polymer Implants
9/79
ix
Total Scores for Liver Implants .............................................................................29
Histology for Liver Injections ................................................................................29
Histology for Brain Injections ...............................................................................30
Histology for Liver Implants .................................................................................30
Experiment 2: Polymer Biocompatibility for Pacemaker Coating ........................30
Scoring Averages for Peritoneal Wall Implants and Surgical Notes .....................31
Histology for Polymer Implants.............................................................................32
Histology for Bionate control Implants .................................................................32
V. DISCUSSION ...............................................................................................................50
Experiment 1: Polymer Biocompatibility for Intervertebral DiscRepair/Replacement ...............................................................................................50
Liver and Brain Injections .....................................................................................50
Liver Implants ........................................................................................................53
Total Scores for Injections and Implants in Liver and Brian .................................54
Summary ................................................................................................................55
Experiment 2: Polymer Biocompatibility for Pacemaker Coating ........................56
VI. CONCLUSION............................................................................................................59
Experiment 1: Polymer Biocompatibility for Intervertebral DiscRepair/Replacement ...............................................................................................59
Experiment 2: Polymer Biocompatibility for Pacemaker Coating ........................60
REFERENCES ..................................................................................................................62
APPENDIX ........................................................................................................................67
-
7/31/2019 Bio Compatibility of Polymer Implants
10/79
x
LIST OF TABLES
Tables Page
1. Important Components to Consider in Material Selection ............................................11
2. Scoring Averages and Surgical Notes for Each Injection/Implant ................................33
3. Animal Survivors Associated with Polymer Type.........................................................39
4. Scoring Averages and Surgical Notes for Each Implant ...............................................45
5. Scoring Averages for Polymer and Bionate control Implants .......................................46
6. Total Mean Scores for Peritoneal Wall Implants ...........................................................46
7. Two-Tail T-Test for Comparison between Polymer and Bionate acrossAll Categories Scored ...................................................................................................47
-
7/31/2019 Bio Compatibility of Polymer Implants
11/79
xi
LIST OF FIGURES
Figures Page
1. Various Applications of Different Polymer Composite Biomaterials .............................7
2. Diagram of Intervertebral Disc ........................................................................................8
3. Schematic Representation of Stages 1-3 for Biofilm Formation ...................................13
4. Experiment 1: Wall Thickness for Liver Injections .......................................................34
5. Experiment 1: Irregular Cells for Liver Injections ........................................................34
6. Experiment 1: Lymphocytes for Liver Injections ..........................................................35
7. Experiment 1: Wall Thickness for Brain Injection ........................................................35
8. Experiment 1: Irregular Cells for Brain Injections ........................................................36
9. Experiment 1: Lymphocytes for Brain Injections ..........................................................36
10. Experiment 1: Wall Thickness for Liver Implants ......................................................37
11. Experiment 1: Irregular Cells for Liver Implants ........................................................37
12. Experiment 1: Lymphocytes for Liver Implants ..........................................................38
13. Experiment 1: Total Mean Scores for Each Group with Liver Injections. ..................40
14. Experiment 1: Total Mean Scores for Each Group with Brain Injections ...................40
15. Experiment 1: Total Mean Scores for Each Group with Liver Implants .....................41
16. Experiment 1: Representative Section for Tissue Analysis for Liver Injections ..............42
-
7/31/2019 Bio Compatibility of Polymer Implants
12/79
xii
17. Experiment 1: Representative Section for Tissue Analysis for BrainInjections ......................................................................................................................43
18. Experiment 1: Representative Section for Tissue Analysis for Liver Implants...........44
19. Experiment 2: Representative Sections for Tissue Analysis for PeritonealWall Polymer Implants ................................................................................................48
20. Experiment 2: Representative Sections for Tissue Analysis for PeritonealWall Bionate Implants .................................................................................................49
-
7/31/2019 Bio Compatibility of Polymer Implants
13/79
1
CHAPTER I
INTRODUCTION
A large number of polymer associated implantable devices are used in medicine
today. Polymer based biomaterials such as bone plates, ligaments, intervertebral discs,
heart valves and pacemakers are used to replace or reestablish function of failing tissues
or organs. These biomaterials help to heal, increase function, repair abnormalities, and
thus improve the patients quality of life (Ramakrishna et al., 2000).
Polymers are utilized in numerous medical applications. This is mainly due to
their versatility as their composition, properties, and forms can be manipulated to readily
produce shapes and structures in the form of gels, films, fibers and solids (Deluca, 1988;
Ramakrishna et al., 2000).
A cardiac pacemaker is a small battery powered devise which produces electronic
impulses to help stimulate the heart muscle. It is used to treat heart arrythmias by
regulating an individuals heart beat. A pacemaker is similar in size to a matchbox and
weighs only between 20-27 grams. It is composed of a corresponding electronic circuit,
lithion ion battery and two tubes with electrodes to transmit the electric impulses to the
heart muscle. These components are enclosed in titanium and covered in a biocompatible
-
7/31/2019 Bio Compatibility of Polymer Implants
14/79
2
material to isolate the devise from exposure to the bodys tissues (Lechleitner et al.,
2005).
Any and all medical implants must be properly sealed to defend against the
aggregation of surrounding body fluids, tissues and cells. Prevention of internal body
cavity substances from penetrating the coating is absolutely necessary to thwart the
destruction of the electrical components. The material(s) used to coat a surgically
implanted medical devise such as a cardiac pacemaker must be biocompatible and
biostable to reduce the possibility of tissue rejection reactions (Lechleitner et al., 2005).
Pacemaker coatings have been known to degrade prematurely due to the build up
of biofilms, which are collections/populations of bacteria, fungi and protozoa. The
microorganisms can deteriorate the protective coatings quite rapidly thus resulting in
damage and often removal and replacement of the devise itself. The function and
structure of these biomaterials can be spoiled by biofilms in a variety of ways. One way
this can occur is by masking surface properties of the protective coating and
contamination of water by microorganisms. Another way is the ability of
microorganisms to extract monomers and additives out of the polymers via microbial
degradation. Also, enzymes and/or radicals can cause brittleness and decrease of
mechanical stability to the polymer coating and its additives. There is also the possibility
of water accumulation that could potentially penetrate the polymer matrix via microbial
filaments. This could lead to swelling and an increase in conductivity. Damage and
biodeterioration can occur to the polymer coating material(s) by any one or combination
of these mechanisms (Flemming, 1998).
-
7/31/2019 Bio Compatibility of Polymer Implants
15/79
-
7/31/2019 Bio Compatibility of Polymer Implants
16/79
4
that of an implanted pacemaker subject to various body tissues. It is hypothesized that
the polymer implant will elicit little or no significant tissue damage, nor will it degrade
itself, and will therefore be suitable for a cardiac pacemaker coating.
-
7/31/2019 Bio Compatibility of Polymer Implants
17/79
5
CHAPTER II
LITERATURE REVIEW
The main objective of this research project was to provide insight to the potential
uses of polymers for medical applications by exploring their biocompatibility in vivo.
Various polymers were investigated in two separate experiments. Experiment 1
examined the possibility of discovering a polymer suitable for intervertebral disc repair
and/or replacement. Experiment 2 pursued the potential use of a polymer as an improved
pacemaker coating. This chapter will provide background information from scientific
articles to justify the reasoning for research in this interdisciplinary academic arena.
Topics relevant to this project will be discussed such as previous related work in polymer
biomaterials and their respective biocompatibility. Biofilms and cell adhesion growth, as
well as prevention, will also be reviewed concerning their roles related to polymer
implants. Mechanisms regarding biodegradation and erosion of polymers in vivo will
also be mentioned. Finally, histology of tissue sections will be discussed to better
determine the biocompatibility of the polymers tested in this project.
-
7/31/2019 Bio Compatibility of Polymer Implants
18/79
6
Medical Applications for Polymer Biomaterials
As seen in Figure 1, polymer composite materials have been produced for many
biomedical applications (Ramakrishna et al., 2000, 2001).
-
7/31/2019 Bio Compatibility of Polymer Implants
19/79
7
Figure 1. Various Applications of Different Polymer Composite Biomaterials. (modifiedfrom Ramakrishna et al., 2001)
Dental Implant
Vascular
Teritoneal Wall
Intramedullary
Li ament/Tendo
Cartilage
Bone plate &
Screws
External Fixation
Total Knee Replacement
Bone Cement
Total Hip
Replacement
Fin er Joint
Spine Cage, Plate, Rods,
Screws and Disc
Bone Replacement
Material
-
7/31/2019 Bio Compatibility of Polymer Implants
20/79
8
Fibrocartilaginous intervertebral discs (IVD) separate vertebrae. They are
composed of a core, nucleus pulposus, and an outer coating known as the annulus fibrosis
which is composed of 90 concentric layers of fibers as seen in Figure 2. The disc is
coated on the lower and upper surfaces by a thin layer of cartaliginous endplates. These
endplates are perforated thus allowing the transfer of nutrients, water, and metabolism
products. The main role of the intervertebral disc is to cushion adjacent vertebral
segments that act as a shock absorber for the spine (Anderson et al., 1993; Ramakrishna
et al., 2001).
Figure 2. Diagram of Intervertebral Disc. (modified fromhttp://www.aafp.org/afp/990201ap/575.html)
Over the years many spinal disorders have been discovered. Many include disc
herniation, facet degeneration, and structural abnormalities such as scoliosis and
kyphosis. It has been reported that one disorder may have a negative cascading impact
on another (Ramakrishna et al., 2001).
Nucleus Pulposus
Transition Zone
Anulus Fibrosus
(Inner)
Anulus
Fibrosus
(Outer)
-
7/31/2019 Bio Compatibility of Polymer Implants
21/79
9
Intervertebral disc repair is treated by the replacement of the damaged area of the
nucleus pulposus with a synthetic polymer material or by complete replacement of the
affected disc with an artificial one. Together, these methods both require the replication
of the natural substance and significant durability. Current research has proposed the
injection of silicone elastomers or hydrogels to replace the nucleus pulposus (Bao et al.,
1996; Ramakrishna et al., 2001). A polyisobutylene (PIB) based intervertebral disc
replacement has been suggested. The synthesized PIBs which carry cyanoacrylate end
groups will polymerize when exposed to moisture, proteins, and/or other weak
nucleophiles. They have the potential to be injected into the space that remains after the
nucleus pulpous of the herniated disc is excised from the spine, and will polymerize in
vivo under the influence of moisture and proteins. Adherence and space filling to the
tissue by the injectable PIB is a perfect method as leakage of the polymer into
surrounding tissues is prevented. Also, the cyanoacrylate groups of the newly formed
disc are protected from enzyme degradation by the high molecular flexible weight of the
hydrophobic PIB coils (Kennedy, 2001; Puskas, 2004).
Polymer surface coatings for implantable medical devices are of great
importance. The success of an implant is dependent on the surface chemistry developed,
which actuates interactions at the implant material-tissue interface. The human body is
selective in regard to implants as rejection of foreign material is almost inevitable thus
requiring polymer coatings that are biocompatible and biologically stable (Ramakrishna
et al., 2001).
The failure of previous pacemaker polymer coatings is notable. Polyether
polyurethane (PEU) elastomers were biocompatible and thought to be biostable. These
-
7/31/2019 Bio Compatibility of Polymer Implants
22/79
10
attributes helped them win their campaign to replace silicone rubber as pacemaker
surface coatings. Although initial results were positive, the biostability of this polymer
did not meet long term standards (Lambda et al., 1998; Wiggins et al., 2001).
There are two mechanisms for the failure of polyther polyurethane elastomers:
metal ion oxidation (MIO) and environmental stress cracking (ESC). The MIO
mechanism explained the degradation of the PEU that was in contact with the metal ions
of the conductor coils. The ESC mechanism depicted the degradation of the PEU that
was in direct contact with tissue. The outer surface of the PEU in drastic cases has been
known to crack and completely rupture the outer insulation. (Stokes et al., 1987; Wiggins
et al., 2001).
Biocompatibility
There are many important factors that need to be taken into consideration
regarding polymer composite materials and compatibility in vivo. Table 1 lists these
components (Ramakrishna et al., 2001).
-
7/31/2019 Bio Compatibility of Polymer Implants
23/79
11
Table 1. Important Components to Consider in Material Selection. (modified fromRamakrishna et al., 2001)
-
7/31/2019 Bio Compatibility of Polymer Implants
24/79
12
Biofilms
Biofilms are communities of micro-organisms anchored to a surface
(substratum) and each other by EPS (exrtra-cellular polymeric substance) (Chen et al.,
1994 as cited in Cogan & Keener, 2004). Cellular adherence and formation of biofilms
to synthetic surfaces like polymer based medical materials is very well known, but may
have undesirable consequences. This colonization and bacterial adherence to medical
implants have the ability to produce infections related to the implant (Jansen et al., 1995).
In many situations, the formation of biofilms, help to protect embedded
microorganisms from antibiotics and host defense mechanisms (Evans et al., 1987; Hoyle
et al., 1990; Hoyle et al., 1991; Jansen et al., 1995). This can potentiate to their
persistence and ultimate survival on polymer devices. Biofilms may be responsible for
the complexities in treating foreign body infections and their infamous durability (Jansen
et al., 1995; Kohnen et al., 1995).
As there is a large range of applicable biomaterials performing various functions,
all can be a potential location for microbial colonization and infection. Microbial
contamination of medical devices is caused by a variety of factors. When a device is
implanted, it immediately elicits local antimicrobial immune responses that can result in a
congregation of microbials. Most patients who experience microbial infections due to
implanted medical devices are more susceptible to them as they may be more likely to be
immunocompromised (Francolini et al., 2003).
The mechanism for biofilm formation can be characterized as a three phase
process. Phase 1 consists of initial and irreversible cellular adherence to the polymer
-
7/31/2019 Bio Compatibility of Polymer Implants
25/79
13
surface. Phase 2 is the expansion and maturation of the biofilm scaffold. Lastly, phase 3
is the adhesion of individual cells or cellular colonies from the biofilm. These stages are
depicted in Figure 3 (Francolini et al., 2003).
Figure 3. Schematic Representation of Stages 1-3 for Biofilm Formation. (modified fromFrancolini et al., 2003)
Biofilm Prevention
Upon implantation of a medical device, the human body responds to the device by
coating it with a protein layer. This layer consists mainly of albumin, laminin, fibrin, and
fibronectin (Francolini et al., 2004; Jeng et al., 2003). There are several approaches to
counteract the formation of this protein layer, or biofilm growth, on medical devices.
One in particular refers to polymers which are chemically modified to prevent primary
microbial adherence and are able to release antimicrobials to prevent surface growth
(Francolini et al., 2004).
Phase 1 Phase 2 Phase 3
-
7/31/2019 Bio Compatibility of Polymer Implants
26/79
14
A decrease in bacterial adherence occurs by changing the surface charge of the
polymer. Since most bacteria are negatively charged, a polymer which has a negative
charge repels bacteria better than an uncharged polymer (Jansen et al., 1995).
Another way to deter antimicrobial growth and adhesion is by the addition of a
silver coating. By binding to microbial DNA and sulphydryl groups, silver ions act to
restrict bacterial replication and deactivate metabolic enzymes (Francolini et al., 2003;
Petering et al., 1976).
Amphiphilic Conetworks
The polymers used in this study for experiments 1 and 2 for the potential
applications of intervertebral disc repair and/or replacement and pacemaker coatings are
amphiphilic conetworks (ACPN). Amphiphilic conetworks are two-component
networks of covalently interconnected hydrophilic/hydrophobic (HI/HO) phases of
cocontinuous morphology; as such they swell both in water and hydrocarbons, and
respond to changes in the medium by morphological isomerization (smart networks)
(Erdodi et al., 2006). ACPNs are responsive to stimuli as they have the ability to
rearrange their morphology very rapidly. This allows the conetwork to produce favorable
conformations in reaction to a medium of any polarity as well as amphiphilic protein
contact (Erdodi et al., 2006).
Amphiphilic conetworks have the potential to be used in a variety of ways for
biomedical applications. They are currently used as controlled implantable drug delivery
devices. In previous work, they exhibit delayed drug delivery profiles, and in some cases
-
7/31/2019 Bio Compatibility of Polymer Implants
27/79
15
yield desirable zero-order drug delivery kinetics (Chen et al., 1989; Erdodi et al., 2006;
Ivan et al., 1989, 1990; Kennedy et al., 1990; Keszler, 1993). Most recently, ACPNs
have been used to support cell proliferation (Erdodi et al., 2006; Haigh et al., 2000,
2002). Another application was for a thin film, or antimicrobial coating, which showed
that the biocide was released into the environment over a long time period (Erdodi et al.,
2006; Tiller, 2005a; Tiller, 2005b). Peroxides that become trapped in the hydrophilic
domains of the ACPNs have displayed increased activity and stability. The increased
activity has been attributed to the hydrophilic/hydrophobic interphase of the ACPN
(Bruns, 2005a; Bruns, 2005b; Erdodi et al., 2006).
Polymer Biodegradation
Investigations of polymeric applications in medicine have elucidated serious
questions about the suitability and degradability of polymers in certain circumstances.
The stability of sensitive compounds such as protein and peptide drugs as well as the
ability of living cells to survive in close proximity to an eroding polymer are of concern.
Research regarding the loss of mechanical stability of polymers during erosion can occur
too quickly and cause high concentrations of toxicity due to the degradated products. An
understanding of polymer erosion and degradation is key to the solution of these
undesirable effects (Gopferich, 1996).
In this study for both experiments 1 and 2, polymer toxicity was measured by
lymphocyte infiltration (used as a marker in this thesis). Lymphocytes are inflammatory
cells which migrate to the wound site to assist in the healing process and rid the body of
-
7/31/2019 Bio Compatibility of Polymer Implants
28/79
16
any foreign invaders. So the level of polymer toxicity can be directly correlated to the
amount of lymphocytes found at the implant site. An increased number of lymphocytes
discovered at the site is indicative of an undesirable chronic inflammatory response
(Jewrajka et al., 2007).
The process of degradation refers to the chain scission process in which polymer
chains are cleaved resulting in the formation of oligomers, then monomers. Erosion is
described as the loss of material owing to monomers and oligomers leaving the polymer
(Gopferich, 1996; Tamada et al., 1993).
There are two principal ways by which polymer bonds can be cleaved: passively
by hydrolysis or actively by enzymatic reaction (Lenz et al., 1993; Gopferich, 1996).
The most important criterion for monitoring polymer degradation is an observed decrease
in molecular weight. Other parameters include loss of mechanical strength and absolute
chemical degradation into monomers, or monomer release by scission of the polymer
backbone (Gopferich, 1996).
-
7/31/2019 Bio Compatibility of Polymer Implants
29/79
17
CHAPTER III
MATERIALS AND METHODS
Experiment 1: Polymer Biocompatibility for Intervertebral Disc Repair/Replacement
It is the goal of experiment 1 to determine if any of the four polymers (polymer,
monomer, initiator, polymer + monomer combination) could be used for intervertebral
disc replacement, or to repair any degenerated discs or disc space that has occurred in the
spine.
Surgical Procedure
A group (n=27) of adult male, SHR/y rats were injected with the anesthesia,
sodium pentathol (50mg/kg,ip) before surgery. For the rats designated for liver
injections, an incision approximately 8 cm long was made on the ventral side of the
animal. The liver was exposed and injected with the respective polymer, or saline for
control.
For the rats designated for brain injections, an incision approximately 5 cm in
length was made on dorsal side of the cranium. With the skull exposed, a dremel with
-
7/31/2019 Bio Compatibility of Polymer Implants
30/79
18
drill bit attachment was used to manufacture a hole adequate for a 10 gauge needle to be
inserted. The brain was then injected with the respective polymer, or saline for control.
Polymers used for Brain and Liver Injections
Monomer: 1-cyanoacryl-2,4,4-trimethylpentane(TMP-CA)
Polymer: cyanoacrylate terminated tri-arm star PIB [(PIB-CA)3
Initiator: N,N-diethyl telechelic [(PIB-NEt3)3] PIB
Monomer + Polymer Combination: cyanoacrylate-telechelic 3-arm star polyisobutylene
[(PIB-CA)3] + 2,2,4-trimethylpent-1-cyanoacrylate (pTMP-CA)
The incision made on the ventral side of the animal for the liver injections was
closed with sutures.
The incision made on the dorsal side of the cranium for this procedure was closed
with staples.
For the group of rats designated for solid polymer liver implants, an incision was
generated approximately 8 cm long on the ventral side of the animal. A 2 cm incision
was made on a lobe of the exposed liver, and the respective solid polymer was implanted.
The incision site was then sutured to prevent the solid implant from leaving the
appropriated tissue site.
-
7/31/2019 Bio Compatibility of Polymer Implants
31/79
19
Solid Polymers Implanted in the Liver
Monomer: poly(1-cyanoacryl-2,4,4-trimethylpentane)poly(TMP-CA)
Polymer: cross linked [(PIB-CA)3] polymer
Monomer + Polymer Combination: copolymer of [(PIB-CA)3] + 2,2,4-trimethylpent-1-
cyanoacrylate (pTMP-CA)
The incision made on the ventral side of the animal for the liver implants was
closed with sutures.
Upon completion of each surgical procedure, suturing and/or stapling, the post-
operative antibioticadministered to the animals was penicillin (5,000 units, im). Theanimals were housed one to a cage with Sani-Chips bedding (Murphy Forest Products,
Montville, NJ) changed once per week and food (PMI Nutrition Intl., Brentwood, MO)
and water were provided ad libitum. The animals were regularly examined for signs of
pain, infection, or implant/injection rejection.
Tissue Removal and Histology
After two weeks the rats were euthanized via a sodium pentothal overdose, and
the tissues containing the polymer injections and/or implants were removed. The tissue
samples were then placed in a 10% buffered formalin phosphate solution for
preservation.
Tissues were dehydrated and embedded with paraffin in a tissue processor
(Tissue-Tek, Miles Scientific). The tissue samples were then embedded into paraffin wax
-
7/31/2019 Bio Compatibility of Polymer Implants
32/79
20
blocks. The tissue blocks were sectioned at 7 microns thick by a Reichert-Jung 820-II
microtome. Sections were taken at the site of the polymer implantation or injection and
placed in a 40 degree Celsius water bath. These sections were transferred to gelatin
coated slides where they were de-waxed, rehydrated, and stained with hematoxylin and
eosin stain (H&E) for microscopic analysis. The slides were observed through a light
microscope and micrographs were taken using an Olympus BX 60 microscope, an MTI
digital camera and the SCION image program. To attain an accurate quantitative
evaluation of the tissue site where the polymer had been implanted or injected, wall
thickness, irregular cells, and lymphocyte infiltration were scored to measure tissue
response. These catergories were blindly scored on a scheme of 0-3 (0 = none, 1 =
negligible, 2 = moderate, 3 = excessive) (Jewrajka et al. 2007).
Experiment 2: Polymer Biocompatibility for Pacemaker Coating
It is the goal of experiment 2 to determine if the polymer tested here, can be used
as an improved coating for cardiac pacemakers.
Surgical Procedure
During experiment 2, each rat was injected with the anesthesia, sodium pentathol
(50mg/kg,ip) before surgery. For the first group of animals (n=4) designated for polymer
implants, an incision approximately 8 cm long was made on the ventral side of the
animal. The peritoneal wall was exposed and the 3 cm long polymer strip was sutured to
-
7/31/2019 Bio Compatibility of Polymer Implants
33/79
21
the wall at both ends. For the second group of animals (n=4) designated for the control
Bionate implants, an incision approximately 8 cm long was made on the ventral side of
the animal. The peritoneal wall was exposed and the 3 cm long polymer or Bionate strip
was sutured to the wall at both ends.
Polymer and Bionate Control Implants
Polymer: Polyisobutylene polytetramethylene oxide
Bionate: Thermoplastic polycarbonate urethane
The incision made on the abdominal muscle wall of the animal for the polymer
and Bionate implants were closed with sutures, and the skin was sealed with staples.
Upon completion of each surgical procedure, the post-operative antibioticadministered tothe animals was penicillin (5,000 units, im). The animals were housed one to a cage with
Sani-Chips bedding (Murphy Forest Products, Montville, NJ) changed once per week and
food (PMI Nutrition Intl., Brentwood, MO) and water were provided ad libitum. The
animals were regularly examined for signs of pain, infection, or implant rejection.
Tissue Removal and Histology
After an interval of four weeks, the rats were euthanized via a sodium pentothal
overdose, and the tissue areas containing the polymer and Bionate implants were
removed. The tissue samples were then placed in a 10% buffered formalin phosphate
solution for preservation.
-
7/31/2019 Bio Compatibility of Polymer Implants
34/79
22
Tissues were dehydrated and embedded with paraffin in a tissue processor
(Tissue-Tek, Miles Scientific). The tissue samples were then embedded into paraffin wax
blocks. The tissue blocks were sectioned at 7 microns thick by a Reichert-Jung 820-II
microtome. Sections were taken at the site of polymer or Bionate implantation and
situated in a 40 degree Celsius water bath. These sections were transferred to gelatin
coated microscope slides and were allowed 24 hours to dry. The slides were then de-
waxed, rehydrated, and stained with hematoxylin and eosin stain (H&E). The slides were
observed through a light microscope and micrographs were taken using an Olympus BX
60 microscope, an MTI digital camera and the SCION image program. To attain an
accurate quantitative evaluation of the tissue site where the polymer/Bionate had been
implanted, lymphocyte infiltration, irregular tissue, irregular cells and fat infiltration were
scored to measure tissue response. These catergories were blindly scored on a scheme of
0-3 (0 = none, 1 = negligible, 2 = moderate, 3 = excessive) (Jewrajka et al., 2007).
Statistical Analysis for Experiments 1 and 2
Statistical tests performed in this study were 1 way ANOVA and t-test. One way
ANOVA tests were used in each graph to test for significance between injections and
implants. A two-sample t-test was used by way of SPSS (Statsistal Package for the
Social Sciences) to compare the four categories scored between the polymer and Bionate
implants to explore the possibility of any statistically significant differences. Total scores
and averages for each category respective to experiments 1 and 2 are presented. A
-
7/31/2019 Bio Compatibility of Polymer Implants
35/79
23
survival table is also shown to focus on the toxicity of the polymers tested. Graphs were
created using Microsoft Excel. Tables were created using Microsoft Word.
-
7/31/2019 Bio Compatibility of Polymer Implants
36/79
24
CHAPTER IV
RESULTS
Experiment 1: Polymer Biocompatibility for Intervertebral Disc Repair/Replacement
In experiment 1, three categories were blindly scored on a scale of 0-3, by four
individuals, to determine wall thickness around the injection/implant site, number of
irregular cells, and number of lymphocytes present. A score of zero represents a thin
wall thickness and little or no cells present. A score of three is indicative of a thick wall
and a considerably high number of cells present at the site.
Injection/Implant Scoring Averages for All Categories
Table 2 depicts the scoring averages for each injection/implant to compare each
category (wall thickness, irregular cells, lymphocytes) scoring average. Injection/implant
site is shown, as well as surgical notes during removal from each tissue.
-
7/31/2019 Bio Compatibility of Polymer Implants
37/79
25
Wall Thickness for Liver Injections
Figure 4 is a graph showing the comparison of wall thickness between polymer
injections and saline control delivered to the liver (mean, s.e.m., 1 way ANOVA). The
scoring means for each group were respectively: Monomer (1.44), Initiator (.57),
Polymer (1.88), Monomer + Polymer (2.25), Saline (.38). There was statistical
significance between Monomer and Saline; p=.016. There was statistical significance
between Polymer and Saline; p=.004. There was statistical significance between
Monomer + Polymer and Saline; p=.003.
Irregular Cells for Liver Injections
Figure 5 is a graph which depicts the comparison of irregular cells for injections
across all treatments administered to the liver (mean, s.e.m., 1 way ANOVA). The
scoring means for each group were: Monomer (2.38), Initiator (.63), Polymer (.25),
Monomer + Polymer (1.0), Saline (0). There was statistical significance between
Monomer and Saline; p=9.73E-5. There was statistical significance between Monomer +
Polymer and Saline; p=.003.
Lymphocytes for Liver Injections
Figure 6 is a graph showing the comparison of lymphocytes between liver
injections for all treatments (mean, s.e.m., 1 way ANOVA). The scoring means for each
-
7/31/2019 Bio Compatibility of Polymer Implants
38/79
26
treatment and control group were: Monomer (2.63), Initiator (.82), Polymer (1.13),
Monomer + Polymer (2.25), Saline (.13). There was statistical significance between
Monomer and Saline; p=.001. There was statistical significance between Initiator and
Saline; p=.007. There was statistical significance between Polymer and Saline; p=.001.
There was statistical significance between Monomer + Polymer and Saline; p=0009.
Wall Thickness for Brain Injections
Figure 7 is a graph which shows the comparison of wall thickness between
polymer injections and saline control delivered to the brain (mean, s.e.m., 1 way
ANOVA). The scoring means for each treatment and control group were respectively:
Monomer (.75), Initiator (.63), Monomer + Polymer (1.0), Saline (.82). There was no
statistical significance between these average scores.
Irregular Cells for Brain Injections
Figure 8 is a graph depicting the comparison of irregular cells between brain
injections for all treatments used (mean, s.e.m., 1 way ANOVA). The scoring averages
for each brain injection were respectively: Monomer (.50), Initiator (.38), Monomer +
Polymer (.38), Saline (0). There was statistical significance between Monomer and
Saline; p=.049. There was statistical significance between Monomer + Polymer and
Saline; p=.002.
-
7/31/2019 Bio Compatibility of Polymer Implants
39/79
27
Lymphocytes for Brain Injections
Figure 9 is a graph showing the comparison of lymphocytes between all
treatments administered to the brain (mean, s.e.m., 1 way ANOVA). The scoring
averages for each brain injection were: Monomer (1.25), Initiator (1.75), Monomer +
Polymer (.63), Saline (.57). There was statistical significance between Initiator and
Saline; p=.001.
Wall Thickness for Liver Implants
Figure 10 is a graph comparing wall thickness between all treatments implanted in
the liver. Means and s.e.m. were shown in this graph. The scoring averages for each
liver implant were: Monomer (.69), Polymer (2.38), Monomer + Polymer (.88). There
was statistical significance between Polymer and Monomer; p=.0002. There was
statistical significance between Polymer and Monomer + Polymer; p=.0002.
Irregular Cells for Liver Implants
Figure 11 is a graph which compares irregular cells between all types of liver
implants used. Means and s.e.m. were shown in this graph. The scoring averages for
each implant delivered to the liver were: Monomer (.69), Polymer (1.5), Monomer +
Polymer (.44). There was statistical significance between Polymer and Monomer +
Polymer; p=.009.
-
7/31/2019 Bio Compatibility of Polymer Implants
40/79
28
Lymphocytes for Liver Implants
Figure 12 is a graph comparing lymphocytes between all treatment implants in the
liver. Means and s.e.m. were shown in this graph. The scoring averages for each implant
in the liver were: Monomer (1.25), Polymer (2.0), Monomer + Polymer (1.82). There
was no statistical significance between treatment groups.
Survival Rate for Injections/Implants
Table 3 shows the number of rats injected or implanted with each polymer or
saline control, site at which the injection or implant was placed, and the number that
survived the procedure. All injections/implants experienced 100% survival rates except:
Monomer + Polymer injection delivered to the liver (50%), Monomer + Polymer
delivered to the brain (50%), Polymer implanted in the liver (50%). Polymer injection
delivered to the brain (0%).
Total Scores for Liver Injections
Figure 13 compares the total score averages across all treatment injections in the
liver (mean, s.e.m., 1 way ANOVA). Total score represents the sum of the averages for
all three categories (wall thickness, irregular cells, lymphocytes). Total score averages
for each treatment was: Monomer (6.45), Initiator (2.02), Polymer (3.26), Monomer +
Polymer (5.50), Saline (.51). There was statistical significance between Monomer and
-
7/31/2019 Bio Compatibility of Polymer Implants
41/79
29
Saline; p=.006. There was statistical significance between Initiator and Saline; p=.02.
There was statistical significance between Monomer + Polymer and Saline; p=.018.
Total Scores for Brain Injections
Figure 14 compares the total score averages for brain injections between all
treatments used (mean, s.e.m., 1 way ANOVA). Total score averages for each treatment
was: Monomer (2.50), Initiator (2.76), Monomer + Polymer (2.01), Saline (1.39). There
was no statistical significance between treatment injections and Saline.
Total Scores for Liver Implants
Figure 15 compares the total score averages for treatment implants delivered to
the liver. Means and s.e.m. were shown in this graph. Total score averages for each
treatment was: Monomer (2.63), Polymer (5.88), Monomer + Polymer (3.14). There was
significance between Polymer and Monomer; p=.027.
Histology for Liver Injections
Figure 16 is a representative section of liver tissue used for the analysis and
scoring of tissue response to treatment injections. Digital pictures of the slides were
taken at 40x magnification. Notice the area between the pointers for the combo
(monomer + polymer) slide. The monomer + polymer injection caused an increased wall
-
7/31/2019 Bio Compatibility of Polymer Implants
42/79
30
thickness formation around the injection site. Also, notice the increased amount of
irregular cells between the pointers for the monomer slide.
Histology for Brain Injections
Figure 17 is a representative section of brain tissue used for the analysis and
scoring of tissue responses to treatment injections. Digital pictures of the slides were
taken at 40x magnification. On the initiator injection slide, there is an increased
lymphocyte infiltration depicted between the pointers. The combo (monomer + polymer)
slide shows a significant amount of irregular cells, a result from the injection.
Histology for Liver Implants
Figure 18 is a representative section of liver tissue used for the analysis and
scoring of tissue response to treatment implants. Digital pictures of the slides were taken
at 40x magnification. The set of pointers labeled A on the polymer slide specifies wall
thickness caused by the implant. The area between the B labeled pointers represents the
increased amount of irregular cells present from the polymer implant.
Experiment 2: Polymer Biocompatibility for Pacemaker Coating
In experiment 2, four categories were blindly scored on a scale of 0-3, by three
individuals, to determine lymphocyte infiltration, fat infiltration, irregular cells, and
-
7/31/2019 Bio Compatibility of Polymer Implants
43/79
31
irregular tissue. A score of zero represented little or no cells/tissue present. A score of
three was indicative of a considerably high number of cells/tissue present at the site. The
following table depicts the scoring averages for each category listed as well as notes
taken during the surgical removal of each sample.
Scoring Averages Peritoneal Wall Implants and Surgical Notes
Table 4 depicts the scoring averages for each implant delivered to the peritoneal
wall to compare each category (lymphocyte infiltration, fat infiltration, irregular cells,
irregular tissue) scoring average. Surgical notes regarding tissue adherence and polymer
condition during the removal of each respective tissue is provided.
Table 5 shows the scoring averages for Polymer and Bionate control implants to
compare each category respectively. Lymphocyte infiltration average scores for Polymer
implant (1.9) is compared to Bionate (2.3). Fat infiltration average scores for Polymer
implant (.8) is in comparison to Bionate (1.4). Irregular cells average scores for Polymer
(1.1) is compared to Bionate (1.6). Irregular tissue average score for Polymer (1.2) is
compared to Bionate (1.7).
Table 6 compares the total score averages for Polymer and Bionate implants
delivered to the peritoneal wall. Total score represents the sum of the averages for all
four categories (lymphocyte infiltration, fat infiltration, irregular cells, irregular tissue).
The total scores were: Polymer (4.99) and Bionate (6.99). There was no statisitcal
significance.
-
7/31/2019 Bio Compatibility of Polymer Implants
44/79
32
Table 7 shows the results of a two-tailed t-test to measure any significance
between categories scored in comparison of Polymer and Bionate implants. There was
statistical significance for fat infiltration at 95% confidence; p=.032.
Histology for Polymer Implants
Figure 19 is a representative section of peritoneal wall tissue used for the analysis
and scoring of tissue response to polymer implants. Digital pictures of the slides are
shown. Polymer 1 is depicted at 40x magnification. Polymer 2 and Polymer 3 are shown
at 100x magnification. The set of pointers on the polymer 1 slide represents an
aggregation of lymphocytes. The area between the pointers on the polymer 2 slide
depicts healthy tissue.
Histology for Bionate control Implants
Figure 20 is a representative section of peritoneal wall tissue used for the analysis
and scoring of tissue response to Bionate implants. Digital pictures of the slides were
taken at 100x magnification. The set of pointers on the Bionate 1 slide represent an area
of irregular cells. The set of pointers labeled A on the Bionate 2 slide specifies
irregular tissue. The set of pointers on the Bionate 2 slide labeled B shows fat
infiltration. The set of pointers on the Bionate 3 slide is representative of lymphocyte
infiltration in response to the implant.
-
7/31/2019 Bio Compatibility of Polymer Implants
45/79
33
Table 2. Scoring Averages and Surgical Notes for Each Injection/Implant
Injection/Implant
TissueSite
WallThickness
IrregularCells
Lymphocytes Surgical Notes
Monomer Liver 1.44 2.38 2.63 Abdominal adhesion
to liverInitiator Liver .57 .63 .82 Abdominal adhesionto liver
Polymer Liver 1.88 .25 1.13 Minimal adhesionCombo Liver 2.25 1.0 2.25 Minimal adhesionSaline Liver .38 0 .13 Minimal adhesionMonomer Brain .75 .50 1.25 Injection pushed outInitiator Brain .63 .38 1.75 Injection pushed outCombo Brain 1.0 .38 .63 Injection site goodSaline Brain .82 0 .57 Injection site goodSolid
Monomer
Liver .69 .69 1.25 Significant adhesion
SolidPolymer
Liver 2.38 1.5 2.0 Mesentery adhesion
SolidCombo
Liver .88 .44 1.82 Intestinal adhesion
Table 2. Experiment 1: Scoring Averages and Surgical Notes for Each Injection/Implant.Wall thickness, irregular cells and lymphocytes represent average tissue response scoresfor each implant/injection. A low score denotes a low tissue response vs. a high scorewhich denotes a high tissue response.
-
7/31/2019 Bio Compatibility of Polymer Implants
46/79
34
Figure 4. Experiment 1: Wall Thickness for Liver Injections. Comparison of wallthickness across treatment injections (means, s.e.m., 1 way ANOVA) in the liver. There
was statistical significance between Monomer and Saline; p=.016. There was statisticalsignificance between Polymer and Saline; p=.004. There was statistical significancebetween Monomer + Polymer and Saline; p=.003.
Figure 5. Experiment 1: Irregular Cells for Liver Injections. Comparison of irregular cells
across treatment injections (means, s.e.m., 1 way ANOVA) in the liver. There wasstatistical significance between Monomer and Saline; p=9.73E-5. There was statisticalsignificance between Monomer + Polymer and Saline; p=.003.
*
**
**
-
7/31/2019 Bio Compatibility of Polymer Implants
47/79
35
Figure 6. Experiment 1: Lymphocytes for Liver Injections. Comparison of lymphocytesacross treatment injections (means, s.e.m., 1 way ANOVA) in the liver. There was
statistical significance between Monomer and Saline; p=.001. There was statisticalsignificance between Initiator and Saline; p=.007. There was statistical significancebetween Polymer and Saline; p=.001. There was statistical significance betweenMonomer + Polymer and Saline; p=0009.
Figure 7. Experiment 1: Wall Thickness for Brain Injection. Comparison of wallthickness across treatment injections (means, s.e.m., 1 way ANOVA) in the brain. Therewas no statistical significance between treatment injections and Saline.
***
***
***
**
-
7/31/2019 Bio Compatibility of Polymer Implants
48/79
36
Figure 8. Experiment 1: Irregular Cells for Brain Injections. Comparison of irregular cellsacross treatment injections (means, s.e.m., 1 way ANOVA) in the brain. There was
statistical significance between Monomer and Saline; p=.049. There was statisticalsignificance between Monomer + Polymer and Saline; p=.002.
Figure 9. Experiment 1: Lymphocytes for Brain Injections. Comparison of lymphocytesacross treatment injections (means, s.e.m., 1 way ANOVA) in the brain. There wasstatistical significance between Initiator and Saline; p=.001.
**
*
***
-
7/31/2019 Bio Compatibility of Polymer Implants
49/79
37
Figure 10. Experiment 1: Wall Thickness for Liver Implants. Comparison of wallthickness across treatment implants (means, s.e.m., 1 way ANOVA) in the liver. There
was statistical significance between Polymer and Monomer + Polymer; p=.0002. Therewas statistical significance between Monomer and Polymer; p=.0002.
Figure 11. Experiment 1: Irregular Cells for Liver Implants. Comparison of irregular cellsacross treatment implants (means, s.e.m., 1 way ANOVA) in the liver. There wasstatistical significance between Polymer and Monomer + Polymer; p=.009.
******
**
-
7/31/2019 Bio Compatibility of Polymer Implants
50/79
38
Figure 12. Experiment 1: Lymphocytes for Liver Implants. Comparison of lymphocytesacross treatment implants (means, s.e.m., 1 way ANOVA) in the liver. There was no
statistical significance between implant groups.
-
7/31/2019 Bio Compatibility of Polymer Implants
51/79
39
Table 3. Animal Survivors Associated with Polymer Type
Type Site Number Injected Number Survived Survival Rate
Monomer Liver 2 2 100%
Initiator Liver 2 2 100%Polymer Liver 2 2 100%Combo Liver 2 1 50%Saline Liver 2 2 100%
Monomer Brain 3 1 33%
Initiator Brain 2 2 100%Polymer Brain 2 0 0%Combo Brain 2 1 50%Saline Brain 2 2 100%Solid
MonomerLiver 2 2 100%
SolidPolymer
Liver 2 1 50%
SolidCombo
Liver 2 2 100%
Table 3. Experiment 1: Animal Survivors Associated with Polymer Type. Number ofanimals which received the respective implant and/or injection and the number of thoseanimals that survived the duration of the experiment.
-
7/31/2019 Bio Compatibility of Polymer Implants
52/79
40
Figure 13. Experiment 1: Total Mean Scores for Each Group with Liver Injections.Comparison of total scores across all treatment injections (means, s.e.m., 1 way
ANOVA) in the liver. Total score represents the sum of the averages for all threecategories (wall thickness, irregular cells, lymphocytes). There was statisticalsignificance between Monomer and Saline; p=.006. There was statistical significancebetween Initiator and Saline; p=.02. There was statistical significance between Monomer+ Polymer and Saline; p=.018.
Figure 14. Experiment 1: Total Mean Scores for Each Group with Brain Injections.Comparison of total scores across all treatment injections (means, s.e.m., 1 wayANOVA) in the brain. Total score represents the sum of the averages for all threecategories (wall thickness, irregular cells, lymphocytes). There was no statisticalsignificance between treatment injections and Saline.
**
*
*
-
7/31/2019 Bio Compatibility of Polymer Implants
53/79
41
Figure 15. Experiment 1: Total Mean Scores for Each Group with Liver Implants.Comparison of total scores across all treatment implants (means, s.e.m., 1 way ANOVA)
in the liver. Total score represents the sum of the averages for all three categories (wallthickness, irregular cells, lymphocytes). There was statistical significance betweenMonomer and Polymer; p=.027.
*
-
7/31/2019 Bio Compatibility of Polymer Implants
54/79
42
Figure 16. Experiment 1: Representative Section for Tissue Analysis for Liver Injections.Slides were used for tissue response scoring and analysis. All five pictures here are shown at40x magnification. Combo refers to Monomer + Polymer injection. Notice the area betweenthe pointers for the combo slide. The combo injection caused an increased wall thicknessformation around the injection site. Also, notice the increased amount of irregular cellsbetween the pointers for the monomer slide.
-
7/31/2019 Bio Compatibility of Polymer Implants
55/79
43
Figure 17. Experiment 1: Representative Section for Tissue Analysis for Brain Injections.Slides were used for tissue response scoring and analysis. All five pictures here areshown at 40x magnification. Combo refers to monomer + polymer injection. On theinitiator injection slide, there is an increased lymphocyte infiltration depicted between thepointers. The combo slide shows a significant amount of irregular cells, a result from theinjection.
-
7/31/2019 Bio Compatibility of Polymer Implants
56/79
44
Figure 18. Experiment 1: Representative Section for Tissue Analysis for Liver Implants.Slides were used for tissue response scoring and analysis. All five pictures here areshown at 40x magnification. Combo refers to monomer + polymer injection. The set ofpointers labeled A on the polymer slide specifies wall thickness caused by the implant.The area between the B labeled pointers represents the increased amount of irregularcells present from the polymer implant.
A
B
-
7/31/2019 Bio Compatibility of Polymer Implants
57/79
-
7/31/2019 Bio Compatibility of Polymer Implants
58/79
46
Table 5. Scoring Averages for Polymer and Bionate control Implants
Implant Lymphocyte
Infiltration
Fat
Infiltration
Irregular
Cells
Irregular
Tissue
Polymer 1.9.11 0.8.22 1.1.26 1.2.22
Bionate
Control
2.3.33 1.4.18 1.6.18 1.7.17
Table 5. Experiment 2: Scoring Averages for Polymer and Bionate Control Implants.Lymphocyte infiltration, fat infiltration, irregular cells and irregular tissue scoresrepresent the average for each respective implant in the peritoneal wall. A high score isindicative of a high tissue response vs. a low score which denotes a low tissue response.
Table 6. Total Mean Scores for Peritoneal Wall Implants
Implant Total Score
Polymer 4.99.23
Bionate Control 6.991.19
Table 6. Experiment 2: Total Mean Scores for Peritoneal Wall Implants. Comparison oftotal scores for both treatment implants (means, s.e.m., 1 way ANOVA) in the peritoneal
wall . Total score represents the sum of the averages for all four categories (lymphocyteinfiltration, fat infiltration, irregular cells, irregular tissue). There was no statisticalsignificance.
-
7/31/2019 Bio Compatibility of Polymer Implants
59/79
47
Table 7. Two-Tail T-Test for Comparison between Polymer and Bionate across AllCategories Scored
Levenes Test
for Equality of
Variances t-test for Equality of Means
95% Confidence
Interval of the
Difference
F Df
Sig. (2-
tailed)
Std. Error
Difference Lower Upper
Lymphocyte
Infiltration
Equal variances assumed 24.123 16 .224 .35136 -1.18930 .30041
Equal variances not assumed 9.756 .235 .35136 -1.22999 .34110
Fat
Infiltration
Equal variances assumed .038 16 .032 * .28328 -1.26719 -.06614
Equal variances not assumed 15.191 .032 .28328 -1.26980 -.06353
Irregular
Cells
Equal variances assumed .400 16 .176 .31427 -1.11067 .22178
Equal variances not assumed 14.027 .179 .31427 -1.11836 .22947
Irregular
Tissue
Equal variances assumed .291 16 .129 .27778 -1.03331 .14442
Equal variances not assumed 14.837 .131 .27778 -1.03708 .14819
Note. Comparison of polymer and Bionate control for all four categories (lymphocyteinfiltration, fat infiltration, irregular cells, irregular tissue). F=F variance, df=degrees offreedom. * denotes significance.
-
7/31/2019 Bio Compatibility of Polymer Implants
60/79
48
Figure 19. Experiment 2: Representative Sections for Tissue Analysis for Peritoneal WallPolymer Implants. Polymer 1 is shown at 40x magnification. Polymer 2 and Polymer 3are shown at 100x magnification. The set of pointers on the polymer 1 slide represents anaggregation of lymphocytes. The area between the pointers on the polymer 2 slide depictshealthy tissue.
-
7/31/2019 Bio Compatibility of Polymer Implants
61/79
49
Figure 20. Experiment 2: Representative Sections for Tissue Analysis for Peritoneal WallBionate Implants. Slides were used for tissue response scoring and analysis. All fivepictures here are shown at 100x magnification. The set of pointers on the Bionate 1 sliderepresent an area of irregular cells. The set of pointers labeled A on the Bionate 2 slidespecifies irregular tissue. The set of pointers on the Bionate 2 slide labeled B shows fatinfiltration. The set of pointers on the Bionate 3 slide is representative of lymphocyteinfiltration in response to the implant.
A
B
-
7/31/2019 Bio Compatibility of Polymer Implants
62/79
50
CHAPTER V
DISCUSSION
Experiment 1: Polymer Biocompatibility for Intervertebral Disc Repair/Replacement
The goal for Experiment 1 was to determine if any of the four polymers (polymer,
monomer, initiator, polymer + monomer combination) tested in the brain and liver could
be used for intervertebral disc replacement, or to repair any degenerated discs or disc
space that occurred in the spine. It was hypothesized that the polymer sample will elicit
little or no significant tissue damage, and will therefore be best suited for intervertebral
disc replacement and/or repair.
Liver and Brain Injections
Prior experimentation regarding the polymer (cyanoacrylate terminated tri-arm
star PIB [(PIB-CA)3) and monomer + polymer combination (cyanoacrylate-telechelic
3-arm star polyisobutylene [(PIB-CA)3] + 2,2,4-trimethylpent-1-cyanoacrylate (pTMP-
CA)) tested for swelling, extractables, and oxidative resistance among others.
-
7/31/2019 Bio Compatibility of Polymer Implants
63/79
51
Under well defined laboratory conditions, these materials exhibited mechanical
properties promising for medical applications, specifically intervertebral discs (Jewrajka
et al., 2008).
The liver was chosen as an alternative tissue site to the brain to evaluate potential
biocompatibility issues with the polymers. Tissue damage analysis for liver injections
showed three treatments: monomer, polymer, and monomer + polymer all having a
higher wall thickness in comparison to the saline control. The initiator may have
recorded the lowest wall thickness for the liver injections because of its lack of
cyanoacrylate, found in the above three mentioned treatment injections. The
cyanoacrylate has induced an increased immune response, in particular wall thickness
around the injection site. Literature supports this result as the polymerization of CA is
exothermic, and the heat released may be responsible for the cell damage in cell culture
(Leggat et al., 2004). Another possible mechanism for CA toxicity refers to the by-
products of degradation: cyanoacetate and formaldehyde, responsible for evoking an
intense inflammatory response (Schwade et al., 2008).
When analyzing irregular cells in the liver it was determined that the monomer
and monomer + polymer injections elucidated an increased response in comparison to the
saline control. This result may be due to the 2, 2, 4-trimethylpentane-1-cyanoacrylate
(TMP-CA) groups that are found in these samples. The polymer and initiator groups did
not contain TMP-CA, and recorded scores that were not statistically significant from the
saline control.
Tissue analysis showed three treatments as having higher scores for lymphocytes
in the liver when compared to saline: initiator, polymer, and monomer + polymer.
-
7/31/2019 Bio Compatibility of Polymer Implants
64/79
52
Although the liver responded to the monomer with an increase in wall thickness around
the injection site, lymphocytes were not statistically significant when compared to the
saline injection. This result may be due to the absence of polyisobutylene (PIB), which is
found in the other three treatment injections. The initiator, polymer, and monomer +
polymer may have activated an increase in lymphocyte infiltration in response to an
infection caused by the PIB.
In support of the hypothesis, the irregular cell score of the polymer injection to
the liver was very similar to that of the saline injection. So although an initial immune
response was evoked in the form of lymphocyte infiltration and increased wall thickness,
no tissue damage in the form of irregular cells occurred in comparison to the saline
control. These results suggest negligible polymer toxicity, and little or no cellular
damage.
It is difficult for the chemicals to leak out from the polymer injections and enter
the central nervous system (CNS) via a systemic route, due to the presence of the blood
brain barrier (Kou et al., 1997). This makes it difficult to determine biocompatibility
when referring to a potential polymer application that will be located in the CNS having
direct contact with cerebral spinal fluid (CSF). Therefore, treatment injections into the
brain were performed as suggested by Dr. Kennedy. The brain is among the softest of
biological tissues, but was chosen as a target tissue site to help create an environment
similar to that of the spinal column, in which the polymer implanted there would be
bathed in CSF.
It was found that monomer and monomer + polymer injections into the brain
resulted in an increase in irregular cells compared to the saline control. This finding is
-
7/31/2019 Bio Compatibility of Polymer Implants
65/79
53
exactly the same as the results for irregular cell scores recorded for the liver injections in
this study. This finding in the brain once again suggests that cellular damage in the form
of irregular cells was due to the TMP-CA in the monomer and monomer + polymer
combination. Different percentage compositions should be examined in the future for
better biocompatibility results.
Tissue damage analysis showed an elevated lymphocyte infiltration in the brain
for the initiator injection compared to the saline control. To explain this result, the
initiator was the only sample that contained PIB without TMP-CA. The absence of this
chemical interaction may have caused this increased immune response in the brain. The
monomer contained TMP-CA, but not PIB, and resulted in the second highest average
score for this category. Perhaps the PIB with TMP-CA is more biocompatible at this
tissue site in regards to lymphocyte infiltration. This suggestion can be supported, as the
monomer + polymer combination recorded a score (.63) very similar to the saline
injection (.57).
Liver Implants
The polymer implant recorded a higher statistically significant score for wall
thickness when compared separately to the monomer and monomer + polymer implants.
This may be the result of the polymer implant not having a TMP-CA group attached.
The monomer and monomer + polymer implants both contained TMP-CA, and had a
lesser tissue response in the liver.
-
7/31/2019 Bio Compatibility of Polymer Implants
66/79
54
The polymer implant also recorded a higher statistically significance score for
irregular cells when compared to the monomer + polymer combination. The lesser score
of the monomer + polymer supports the idea that an implant, which is composed of PIB
and TMP-CA, may be the most biocompatible. So a polymer composed of PIB alone
may not be as biologically compatible as compared to PIB-TMP-CA.
Total Scores for Injections and Implants in Liver and Brain
The monomer injected in the liver recorded the highest total tissue response score
(6.45) and statistical significance was noted compared to the saline control. The initiator
and monomer + polymer treatments also recorded a higher statistically significant score
when compared to the control. The polymer injection recorded a total score that was
found not to be statistically significant compared to saline.
The solid polymer implant in the liver registered the highest total score (5.88).
This finding contradicts the result for total score regarding the polymer hydrogel injected
in the liver. The difference seen in the study may be explained by the morphology of the
polymer implant vs. polymer injection. Biomaterial interactions with tissues are directed
by surface molecules, specific receptors, atomic geometry and the electronic state of the
biomaterial surface (Gadkari et al., 1989). It appears that both of these polymer forms are
not biocompatible.
-
7/31/2019 Bio Compatibility of Polymer Implants
67/79
55
Summary
It was very difficult to find literature pertaining to toxic polymer effects in the
liver and brain for the application of intervertebral disc replacement. Searches were
carried out on various websites, electronic journal databases, and search engines
including: Google.com, Googlescholar.com, Dogpile.com, and OhioLink. Experiments
involving polymer injections and/or implants in the brain were inclusive of successful
biodegradable polymers associated with controllable drug carriers.
The brain may be the most important site to focus our attention to when
examining these results, because as mentioned above, this area has a similar environment
to that of the spinal column with the presence of cerebral spinal fluid (CSF). There are
some drawbacks with brain injections which include the risk of infection due to very
frequent reservoir replenishment and peripheral toxicity due to rapid CSF turnover (Kou
et al., 1997). The chemical makeup of the monomer, polymer, and monomer + polymer
combination is the most logical explanation for this increased rate of mortality (~ 55%)
with these injections. The samples are responsible for delivering toxic effects to the
CNS.
The initiator generated the highest total score (2.76) for the brain injections, but
there was not any significant difference between it and the saline control. The polymer
injection failed to record a score here, because both animals injected failed to survive.
Any type of injection into the brain has to be done very carefully as room for error is
small, but I do not believe technique was to blame for this particular result. Both the
initiator and saline injections for this group achieved a 100% survival rate. Two of the
-
7/31/2019 Bio Compatibility of Polymer Implants
68/79
56
three animals injected with the monomer at this site died, as well as one of the two
animals injected with the monomer + polymer combination. It was discovered that the
animal with the monomer injection into the brain that did survive had actually pushed the
injection out of the site. Had the injection stayed in the brain, this animals chance of
surviving the two week time interval may have significantly decreased. Based on this
data, it seems as if any injection into the brain other than that of the initiator or saline
control may illicit negative effects on the animal model, most likely leading to death.
Only one of the two animals receiving the solid polymer implant survived. So the
polymer hydrogel and solid form elicited a high degree of mortality (75%) for animals
receiving these treatments, deeming this particular sample extremely toxic.
In 16 of 20 animal models, tissue involvement/adherence was noted. This would
undoubtedly be problematic when pertaining to a prosthetic polymer disc or polymer
injection into intervertebral disc space. Of the four animals that did not show tissue
adhesion, two of the injections (initiator and monomer) were displaced from the target
site (brain), one was a Saline injection into the brain, and the other was a monomer +
polymer combination injection into the brain.
Experiment 2: Polymer Biocompatibility for Pacemaker Coating
The goal for Experiment 2 was to determine if the polymer tested, in the
peritoneal wall, could be used as an improved coating for cardiac pacemakers. It was
hypothesized that the polymer implant will elicit little or no significant tissue damage,
-
7/31/2019 Bio Compatibility of Polymer Implants
69/79
57
nor will it degrade itself, and will therefore be a suitable candidate for a cardiac
pacemaker coating.
It should be noted that tissue sections were not obtained from two of eight animals
in this study. The two tissue samples were extremely brittle, dry, and flaky during the
sectioning process. I met with Dr. Ely and Dr. Salisbury to discuss the matter and it first
was suggested to me to soak the wax block samples in 10% ammonia hydroxide for ten
minutes and then try to section them. After doing this the sections still remained brittle
and dry, so I soaked the wax blocks for three hours. Again, I tried to section each sample
to no avail. I then spoke with Dr. Salisbury once more and we decided to reprocess the
tissue samples. The tissues were removed from their wax blocks and placed in a cassette
to soak in Hemo-De for five days to help remove the paraffin wax still infiltrating the
tissue. After this, the tissue samples were removed from the cassettes and placed in a
cartridge (approximately 2.5x 1 in.) containing Hemo-De, and placed on a rotator for six
days to remove any wax that may have still been present in the tissue. The tissues were
then reprocessed and re-embedded in paraffin wax. Sectioning resumed with the
reprocessed tissue samples and the same result was found as before; the samples
remained too dry to section properly.
Lymphocyte infiltration, irregular cells, and irregular tissue are all signs of the
bodys immune response to a foreign material, in this case, polymer implants onto the
peritoneal wall. A high score in any of these areas is indicative of tissue damage and
overall poor biocompatibility.
The Bionate control implant achieved the highest total score (6.99) compared to
the polymer implant (4.99). This result suggests that although no statistical significance
-
7/31/2019 Bio Compatibility of Polymer Implants
70/79
58
between the polymer and Bionate groups was noted, the polymer implant did elicit less of
a tissue response versus the Bionate control. In fact, the polymer induced less of a tissue
response in every category scored, including a statistically significant difference in the fat
infiltration category. The hydrophilic and hydrophobic cross-linked domains that
compose this amphiphilic polymer allows for the counter-current diffusion of water and
oxygen and better overall performance compared to the control (Jewrajka et al., 2007).
In 8 of 8 animals, fibrous encapsulation of the polymer and fat adherence to it
was noted. There was no evidence of fluid accumulation or a visual infectious response.
None of the animals demonstrated any signs of implant rejection (implants protruding
from the skin) (Jewrajka et al., 2007). In support of the fat adherence finding, a similar
result was discovered in a study conducted for a polymer (polypropylene) implanted in
the abdominal wall. The tissue site where the polymer was implanted was removed after
four and eight week time periods. Fat infiltration was found after both durations, but was
greater after twelve weeks (Kaleya, 2005).
The adipose tissue adherence is an interesting phenomenon. The reason for the
fat adhesion to the implants may be a line of defense associated with these animal
models. The fat may have migrated from the epididymis to the implant site on abdominal
wall to protect any potential injury to vital organs
(http://massage.largeheartedboy.com/archive/2009/01/role_of_fat.html).
-
7/31/2019 Bio Compatibility of Polymer Implants
71/79
59
CHAPTER VI
CONCLUSION
Experiment 1: Polymer Biocompatibility for Intervertebral Disc Repair/Replacement
As the progression of intervertebral disc degeneration fails to cease, research
involving candidate biomaterials for the spinal column increases. Improving implant
design and delivery to the target site is of importance, but biocompatibility and durability
of the polymer is of chief concern.
It is extremely difficult to support my hypothesis from the data presented in this
experiment. The solid polymer implant into the liver recorded the highest total score
compared to the other implants and one out of two animals implanted with the solid
polymer failed to survive. Also, there were no survivors with the polymer injection into
the brain.
The in vivo findings in this study contradict the in vitro results discovered by
Jewrajka et al. Indeed physiological settings are difficult to reproduce in the laboratory,
-
7/31/2019 Bio Compatibility of Polymer Implants
72/79
60
but the polymer tested here was extremely toxic and not well tolerated in the brain tissue,
thus the hypothesis was rejected.
To understand the mechanisms involved in the immunological response for these
implants and injections requires a close relationship between multiple disciplines such as
polymer science, biomedical engineering, and biology.
Experiment 2: Polymer Biocompatibility for Pacemaker Coating
As a medical device coating becomes susceptible to chemical degradation it
becomes brittle and cracking ensues. Biological enzymes attack these weakened areas
and can promote further damage via electrical malfunction of the pacemaker itself.
Improved polymer coatings for pacemakers should prevent cracking and biodegradation
and serve as a successful protective barrier for pacemaker.
Both implants in this experiment failed to repel tissue adherence completely, but
lower inflammatory values in each category scored were recorded for the polymer
compared to the Bionate control. Fat adherence, which was noted in each implant, was
significantly less in the polymer implant compared to the control. The low tissue
inflammatory response suggested by the data supports the hypothesis that the polymer
implant will elicit little or no significant tissue damage.
All four polymers recovered were visually in tact, meaning they looked the same
post-surgery as they did pre-surgery without tears or physical damage. It is my belief
that this sample is both biocompatible and biostable, thus making it suitable as a synthetic
polymer pacemaker coating.
-
7/31/2019 Bio Compatibility of Polymer Implants
73/79
61
The polymer coating tested in experiment two may have other roles in future
medical devices. Applications for this polymer may include catheter and cardiac stent
coatings, but future research is needed in the area of biological durability. This study
requires further research to test the long-term biostability of this material in physiological
conditions as in vivo testing involves much more complex biological effects on polymers
and does not always support in vitro testing.
-
7/31/2019 Bio Compatibility of Polymer Implants
74/7