Bio Compatibility of Polymer Implants

download Bio Compatibility of Polymer Implants

of 79

Transcript of Bio Compatibility of Polymer Implants

  • 7/31/2019 Bio Compatibility of Polymer Implants

    1/79

    BIOCOMPATIBILITY OF POLYMER IMPLANTS

    FOR MEDICAL APPLICATIONS

    A Thesis

    Presented to

    The Graduate Faculty of The University of Akron

    In Partial Fulfillment

    of the Requirements for the Degree

    Master of Science

    Christopher M. Brendel

    August 2009

  • 7/31/2019 Bio Compatibility of Polymer Implants

    2/79

    ii

    BIOCOMPATIBILITY OF POLYMER IMPLANTS FOR MEDICAL APPLICATIONS

    Christopher M. Brendel

    Thesis

    Approved: Accepted:

    ________________________________ ________________________________

    Advisor Department ChairDr. Daniel Ely Dr. Monte Turner

    ________________________________ ________________________________

    Faculty Reader Dean of the CollegeDr. Ronald Salisbury Dr. Chand Midha

    ________________________________ ________________________________

    Committee Member Dean of the Graduate SchoolDr. Qin Liu Dr. George Newkome

    ________________________________

    Date

  • 7/31/2019 Bio Compatibility of Polymer Implants

    3/79

    iii

    ABSTRACT

    Two separate experiments were performed to test the biocompatibility of

    amphiphilic conetwork polymers for medical applications. In experiment 1, four

    different types of polymers were tested in the liver and brain of 20 rats to determine

    which polymer would be best suited for intervertebral disc repair or replacement. The

    samples used for the brain and liver injections were Monomer: 1-cyanoacryl-2,4,4-

    trimethylpentane(TMP-CA), Polymer: cyanoacrylate terminated tri-arm star PIB [(PIB-

    CA)3, Initiator: N,N-diethyl telechelic [(PIB-NEt3)3] PIB, and Monomer + Polymer

    Combination: cyanoacrylate-telechelic 3-arm star polyisobutylene [(PIB-CA)3] + 2,2,4-

    trimethylpent-1-cyanoacrylate (pTMP-CA). The solid samples implanted into the liver

    were Monomer: poly(1-cyanoacryl-2,4,4-trimethylpentane)poly(TMP-CA), Polymer:

    cross linked [(PIB-CA)3] polymer, and Monomer + Polymer Combination: copolymer

    of [(PIB-CA)3] + 2,2,4-trimethylpent-1-cyanoacrylate (pTMP-CA).

    The rats were divided into three groups. Group number one had four different

    honey-consistency gel polymers injected into the liver and a saline injection for

    control. Group number two was subjected to the same injections, but this time the site of

    injection was the brain. Group number three had three types of solid polymers surgically

  • 7/31/2019 Bio Compatibility of Polymer Implants

    4/79

    iv

    implanted into the liver. Tissues were then examined histologically to determine if any

    damage had occurred. A polymer sample that resulted in little or no significant tissue

    damage would be a good candidate for intervertebral disc replacement and/or repair.

    In experiment 2, two types of polymers were used; a polyisobutylene polymer as

    well as a Bionate sample for a control, (to determine the biocompatibility of the polymer

    for a potential pacemaker coating). Both were implanted on the peritoneal wall of eight

    rats. The rats were divided into two groups of four. Group number one had four

    polyisobutylene polymer strips sutured to the peritoneal wall. Group number two had

    four Bionate strips also implanted on the peritoneal wall. Tissue at the implant site was

    removed and examined histologically to determine if any damage had occurred. A

    polymer sample that resulted in little or no significant tissue damage was considered a

    good candidate for a potential pacemaker coating.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    5/79

    v

    ACKNOWLEDGEMENTS

    I would like to thank everyone who assisted me with this experimental

    investigation over the course of the past two years. Without their help, my research

    efforts which have culminated into this thesis would not have been possible. The

    following individuals trust and support guided me throughout the laboratory work,

    research, and production of this paper, of which I am most appreciative and thankful.

    The first individual that I would like to thank is Dr. Kennedy. His expertise,

    innovation, and enthusiasm in the field of polymer science have afforded me the

    opportunity to participate in biomedical research since the final semester of my

    undergraduate career. It has been through this work, that I have realized my passion

    towards working in biomedical research. I am excited about the prospect of pursuing my

    Ph.D. in Polymer Science in the Fall of 2009, and look forward to any future

    collaboration with Dr. Kennedy. I would also like to thank Suresh Jewrajka, Dr.

    Kennedys research assistant, for synthesizing the polymer samples that were used in this

    study.

    My gratitude is also extended to Gail Dunphy and Shannon Boehme for their

    assistance in learning the proper technical and surgical protocols in the lab. Without their

    insight and guidance this would have been an incredibly cumbersome task. I have gained

  • 7/31/2019 Bio Compatibility of Polymer Implants

    6/79

    vi

    a great deal of knowledge and techniques from them, and am truly grateful for all of their

    efforts throughout this process.

    I would also like to acknowledge Amy Kladar for her work as an undergraduate

    assistant, which was invaluable to my research.

    Finally, without the quality of time devoted, counseling, consideration, and

    mentoring of Dr. Ely from the last semester of my undergraduate career throughout my

    graduate studies, I would not be where I am today. Its hard to find the right words to

    adequately express my gratitude to Dr. Ely who has played such an instrumental role in

    my education and career exploration. It is thanks to Dr. Ely that I will graduate with a

    clear career direction. Dr. Elys support and faith in me has been paramount to my

    success and I am very fortunate to have had such a distinguished and caring advisor to

    guide me through not only my research and graduate program, but also to help me realize

    the full potential of the work I can accomplish in the future.

    I would also like to thank my family and friends for their unwavering belief in my

    ability, which means so very much to me.

    As I complete my graduate coursework and embark on the new journey of earning

    my Ph.D. in Polymer Science, I am thankful to have so many highly esteemed

    professionals as colleagues. It is my sincere hope that I will be able to work with them in

    some capacity in the future.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    7/79

    vii

    TABLE OF CONTENTS

    Page

    LIST OF TABLES ...............................................................................................................x

    LIST OF FIGURES ........................................................................................................... xi

    CHAPTER

    I. INTRODUCTION ............................................................................................................1

    Hypotheses and Objectives ......................................................................................3

    II. LITERATURE REVIEW ................................................................................................5

    Medical Applications for Polymer Biomaterials .....................................................6

    Biocompatibility ....................................................................................................10

    Biofilms..................................................................................................................12

    Biofilm Prevention .................................................................................................13

    Amphiphilic Conetworks .......................................................................................14

    Polymer Biodegradation ........................................................................................15

    III. MATERIALS AND METHODS .................................................................................17

    Experiment 1: Polymer Biocompatibility for Intervertebral Disc

    Repair/Replacement ...............................................................................................17

    Surgical Procedure .................................................................................................17

    Polymers used for Brain and Liver Injections .......................................................18

  • 7/31/2019 Bio Compatibility of Polymer Implants

    8/79

    viii

    Solid Polymers Implanted in the Liver ..................................................................19

    Tissue Removal and Histology ..............................................................................19

    Experiment 2: Polymer Biocompatibility for Pacemaker Coating ........................20

    Surgical Procedure .................................................................................................20

    Polymer and Bionate Control Implants..................................................................21

    Tissue Removal and Histology ..............................................................................21

    Statistical Analysis for Experiments 1 and 2 .........................................................22

    IV. RESULTS ....................................................................................................................24

    Experiment 1: Polymer Biocompatibility for Intervertebral DiscRepair/Replacement ...............................................................................................24

    Injection/Implant Scoring Averages for All Categories ........................................24

    Wall Thickness for Liver Injections ......................................................................25

    Irregular Cells for Liver Injections ........................................................................25

    Lymphocytes for Liver Injections ..........................................................................25

    Wall Thickness for Brain Injections ......................................................................26

    Irregular Cells for Brain Injections ........................................................................26

    Lymphocytes for Brain Injections .........................................................................27

    Wall Thickness for Liver Implants ........................................................................27

    Irregular Cells for Liver Implants ..........................................................................27

    Lymphocytes for Liver Implants ...........................................................................28

    Survival Rate for Injections/Implants ....................................................................28

    Total Scores for Liver Injections ...........................................................................28

    Total Scores for Brain Injections ...........................................................................29

  • 7/31/2019 Bio Compatibility of Polymer Implants

    9/79

    ix

    Total Scores for Liver Implants .............................................................................29

    Histology for Liver Injections ................................................................................29

    Histology for Brain Injections ...............................................................................30

    Histology for Liver Implants .................................................................................30

    Experiment 2: Polymer Biocompatibility for Pacemaker Coating ........................30

    Scoring Averages for Peritoneal Wall Implants and Surgical Notes .....................31

    Histology for Polymer Implants.............................................................................32

    Histology for Bionate control Implants .................................................................32

    V. DISCUSSION ...............................................................................................................50

    Experiment 1: Polymer Biocompatibility for Intervertebral DiscRepair/Replacement ...............................................................................................50

    Liver and Brain Injections .....................................................................................50

    Liver Implants ........................................................................................................53

    Total Scores for Injections and Implants in Liver and Brian .................................54

    Summary ................................................................................................................55

    Experiment 2: Polymer Biocompatibility for Pacemaker Coating ........................56

    VI. CONCLUSION............................................................................................................59

    Experiment 1: Polymer Biocompatibility for Intervertebral DiscRepair/Replacement ...............................................................................................59

    Experiment 2: Polymer Biocompatibility for Pacemaker Coating ........................60

    REFERENCES ..................................................................................................................62

    APPENDIX ........................................................................................................................67

  • 7/31/2019 Bio Compatibility of Polymer Implants

    10/79

    x

    LIST OF TABLES

    Tables Page

    1. Important Components to Consider in Material Selection ............................................11

    2. Scoring Averages and Surgical Notes for Each Injection/Implant ................................33

    3. Animal Survivors Associated with Polymer Type.........................................................39

    4. Scoring Averages and Surgical Notes for Each Implant ...............................................45

    5. Scoring Averages for Polymer and Bionate control Implants .......................................46

    6. Total Mean Scores for Peritoneal Wall Implants ...........................................................46

    7. Two-Tail T-Test for Comparison between Polymer and Bionate acrossAll Categories Scored ...................................................................................................47

  • 7/31/2019 Bio Compatibility of Polymer Implants

    11/79

    xi

    LIST OF FIGURES

    Figures Page

    1. Various Applications of Different Polymer Composite Biomaterials .............................7

    2. Diagram of Intervertebral Disc ........................................................................................8

    3. Schematic Representation of Stages 1-3 for Biofilm Formation ...................................13

    4. Experiment 1: Wall Thickness for Liver Injections .......................................................34

    5. Experiment 1: Irregular Cells for Liver Injections ........................................................34

    6. Experiment 1: Lymphocytes for Liver Injections ..........................................................35

    7. Experiment 1: Wall Thickness for Brain Injection ........................................................35

    8. Experiment 1: Irregular Cells for Brain Injections ........................................................36

    9. Experiment 1: Lymphocytes for Brain Injections ..........................................................36

    10. Experiment 1: Wall Thickness for Liver Implants ......................................................37

    11. Experiment 1: Irregular Cells for Liver Implants ........................................................37

    12. Experiment 1: Lymphocytes for Liver Implants ..........................................................38

    13. Experiment 1: Total Mean Scores for Each Group with Liver Injections. ..................40

    14. Experiment 1: Total Mean Scores for Each Group with Brain Injections ...................40

    15. Experiment 1: Total Mean Scores for Each Group with Liver Implants .....................41

    16. Experiment 1: Representative Section for Tissue Analysis for Liver Injections ..............42

  • 7/31/2019 Bio Compatibility of Polymer Implants

    12/79

    xii

    17. Experiment 1: Representative Section for Tissue Analysis for BrainInjections ......................................................................................................................43

    18. Experiment 1: Representative Section for Tissue Analysis for Liver Implants...........44

    19. Experiment 2: Representative Sections for Tissue Analysis for PeritonealWall Polymer Implants ................................................................................................48

    20. Experiment 2: Representative Sections for Tissue Analysis for PeritonealWall Bionate Implants .................................................................................................49

  • 7/31/2019 Bio Compatibility of Polymer Implants

    13/79

    1

    CHAPTER I

    INTRODUCTION

    A large number of polymer associated implantable devices are used in medicine

    today. Polymer based biomaterials such as bone plates, ligaments, intervertebral discs,

    heart valves and pacemakers are used to replace or reestablish function of failing tissues

    or organs. These biomaterials help to heal, increase function, repair abnormalities, and

    thus improve the patients quality of life (Ramakrishna et al., 2000).

    Polymers are utilized in numerous medical applications. This is mainly due to

    their versatility as their composition, properties, and forms can be manipulated to readily

    produce shapes and structures in the form of gels, films, fibers and solids (Deluca, 1988;

    Ramakrishna et al., 2000).

    A cardiac pacemaker is a small battery powered devise which produces electronic

    impulses to help stimulate the heart muscle. It is used to treat heart arrythmias by

    regulating an individuals heart beat. A pacemaker is similar in size to a matchbox and

    weighs only between 20-27 grams. It is composed of a corresponding electronic circuit,

    lithion ion battery and two tubes with electrodes to transmit the electric impulses to the

    heart muscle. These components are enclosed in titanium and covered in a biocompatible

  • 7/31/2019 Bio Compatibility of Polymer Implants

    14/79

    2

    material to isolate the devise from exposure to the bodys tissues (Lechleitner et al.,

    2005).

    Any and all medical implants must be properly sealed to defend against the

    aggregation of surrounding body fluids, tissues and cells. Prevention of internal body

    cavity substances from penetrating the coating is absolutely necessary to thwart the

    destruction of the electrical components. The material(s) used to coat a surgically

    implanted medical devise such as a cardiac pacemaker must be biocompatible and

    biostable to reduce the possibility of tissue rejection reactions (Lechleitner et al., 2005).

    Pacemaker coatings have been known to degrade prematurely due to the build up

    of biofilms, which are collections/populations of bacteria, fungi and protozoa. The

    microorganisms can deteriorate the protective coatings quite rapidly thus resulting in

    damage and often removal and replacement of the devise itself. The function and

    structure of these biomaterials can be spoiled by biofilms in a variety of ways. One way

    this can occur is by masking surface properties of the protective coating and

    contamination of water by microorganisms. Another way is the ability of

    microorganisms to extract monomers and additives out of the polymers via microbial

    degradation. Also, enzymes and/or radicals can cause brittleness and decrease of

    mechanical stability to the polymer coating and its additives. There is also the possibility

    of water accumulation that could potentially penetrate the polymer matrix via microbial

    filaments. This could lead to swelling and an increase in conductivity. Damage and

    biodeterioration can occur to the polymer coating material(s) by any one or combination

    of these mechanisms (Flemming, 1998).

  • 7/31/2019 Bio Compatibility of Polymer Implants

    15/79

  • 7/31/2019 Bio Compatibility of Polymer Implants

    16/79

    4

    that of an implanted pacemaker subject to various body tissues. It is hypothesized that

    the polymer implant will elicit little or no significant tissue damage, nor will it degrade

    itself, and will therefore be suitable for a cardiac pacemaker coating.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    17/79

    5

    CHAPTER II

    LITERATURE REVIEW

    The main objective of this research project was to provide insight to the potential

    uses of polymers for medical applications by exploring their biocompatibility in vivo.

    Various polymers were investigated in two separate experiments. Experiment 1

    examined the possibility of discovering a polymer suitable for intervertebral disc repair

    and/or replacement. Experiment 2 pursued the potential use of a polymer as an improved

    pacemaker coating. This chapter will provide background information from scientific

    articles to justify the reasoning for research in this interdisciplinary academic arena.

    Topics relevant to this project will be discussed such as previous related work in polymer

    biomaterials and their respective biocompatibility. Biofilms and cell adhesion growth, as

    well as prevention, will also be reviewed concerning their roles related to polymer

    implants. Mechanisms regarding biodegradation and erosion of polymers in vivo will

    also be mentioned. Finally, histology of tissue sections will be discussed to better

    determine the biocompatibility of the polymers tested in this project.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    18/79

    6

    Medical Applications for Polymer Biomaterials

    As seen in Figure 1, polymer composite materials have been produced for many

    biomedical applications (Ramakrishna et al., 2000, 2001).

  • 7/31/2019 Bio Compatibility of Polymer Implants

    19/79

    7

    Figure 1. Various Applications of Different Polymer Composite Biomaterials. (modifiedfrom Ramakrishna et al., 2001)

    Dental Implant

    Vascular

    Teritoneal Wall

    Intramedullary

    Li ament/Tendo

    Cartilage

    Bone plate &

    Screws

    External Fixation

    Total Knee Replacement

    Bone Cement

    Total Hip

    Replacement

    Fin er Joint

    Spine Cage, Plate, Rods,

    Screws and Disc

    Bone Replacement

    Material

  • 7/31/2019 Bio Compatibility of Polymer Implants

    20/79

    8

    Fibrocartilaginous intervertebral discs (IVD) separate vertebrae. They are

    composed of a core, nucleus pulposus, and an outer coating known as the annulus fibrosis

    which is composed of 90 concentric layers of fibers as seen in Figure 2. The disc is

    coated on the lower and upper surfaces by a thin layer of cartaliginous endplates. These

    endplates are perforated thus allowing the transfer of nutrients, water, and metabolism

    products. The main role of the intervertebral disc is to cushion adjacent vertebral

    segments that act as a shock absorber for the spine (Anderson et al., 1993; Ramakrishna

    et al., 2001).

    Figure 2. Diagram of Intervertebral Disc. (modified fromhttp://www.aafp.org/afp/990201ap/575.html)

    Over the years many spinal disorders have been discovered. Many include disc

    herniation, facet degeneration, and structural abnormalities such as scoliosis and

    kyphosis. It has been reported that one disorder may have a negative cascading impact

    on another (Ramakrishna et al., 2001).

    Nucleus Pulposus

    Transition Zone

    Anulus Fibrosus

    (Inner)

    Anulus

    Fibrosus

    (Outer)

  • 7/31/2019 Bio Compatibility of Polymer Implants

    21/79

    9

    Intervertebral disc repair is treated by the replacement of the damaged area of the

    nucleus pulposus with a synthetic polymer material or by complete replacement of the

    affected disc with an artificial one. Together, these methods both require the replication

    of the natural substance and significant durability. Current research has proposed the

    injection of silicone elastomers or hydrogels to replace the nucleus pulposus (Bao et al.,

    1996; Ramakrishna et al., 2001). A polyisobutylene (PIB) based intervertebral disc

    replacement has been suggested. The synthesized PIBs which carry cyanoacrylate end

    groups will polymerize when exposed to moisture, proteins, and/or other weak

    nucleophiles. They have the potential to be injected into the space that remains after the

    nucleus pulpous of the herniated disc is excised from the spine, and will polymerize in

    vivo under the influence of moisture and proteins. Adherence and space filling to the

    tissue by the injectable PIB is a perfect method as leakage of the polymer into

    surrounding tissues is prevented. Also, the cyanoacrylate groups of the newly formed

    disc are protected from enzyme degradation by the high molecular flexible weight of the

    hydrophobic PIB coils (Kennedy, 2001; Puskas, 2004).

    Polymer surface coatings for implantable medical devices are of great

    importance. The success of an implant is dependent on the surface chemistry developed,

    which actuates interactions at the implant material-tissue interface. The human body is

    selective in regard to implants as rejection of foreign material is almost inevitable thus

    requiring polymer coatings that are biocompatible and biologically stable (Ramakrishna

    et al., 2001).

    The failure of previous pacemaker polymer coatings is notable. Polyether

    polyurethane (PEU) elastomers were biocompatible and thought to be biostable. These

  • 7/31/2019 Bio Compatibility of Polymer Implants

    22/79

    10

    attributes helped them win their campaign to replace silicone rubber as pacemaker

    surface coatings. Although initial results were positive, the biostability of this polymer

    did not meet long term standards (Lambda et al., 1998; Wiggins et al., 2001).

    There are two mechanisms for the failure of polyther polyurethane elastomers:

    metal ion oxidation (MIO) and environmental stress cracking (ESC). The MIO

    mechanism explained the degradation of the PEU that was in contact with the metal ions

    of the conductor coils. The ESC mechanism depicted the degradation of the PEU that

    was in direct contact with tissue. The outer surface of the PEU in drastic cases has been

    known to crack and completely rupture the outer insulation. (Stokes et al., 1987; Wiggins

    et al., 2001).

    Biocompatibility

    There are many important factors that need to be taken into consideration

    regarding polymer composite materials and compatibility in vivo. Table 1 lists these

    components (Ramakrishna et al., 2001).

  • 7/31/2019 Bio Compatibility of Polymer Implants

    23/79

    11

    Table 1. Important Components to Consider in Material Selection. (modified fromRamakrishna et al., 2001)

  • 7/31/2019 Bio Compatibility of Polymer Implants

    24/79

    12

    Biofilms

    Biofilms are communities of micro-organisms anchored to a surface

    (substratum) and each other by EPS (exrtra-cellular polymeric substance) (Chen et al.,

    1994 as cited in Cogan & Keener, 2004). Cellular adherence and formation of biofilms

    to synthetic surfaces like polymer based medical materials is very well known, but may

    have undesirable consequences. This colonization and bacterial adherence to medical

    implants have the ability to produce infections related to the implant (Jansen et al., 1995).

    In many situations, the formation of biofilms, help to protect embedded

    microorganisms from antibiotics and host defense mechanisms (Evans et al., 1987; Hoyle

    et al., 1990; Hoyle et al., 1991; Jansen et al., 1995). This can potentiate to their

    persistence and ultimate survival on polymer devices. Biofilms may be responsible for

    the complexities in treating foreign body infections and their infamous durability (Jansen

    et al., 1995; Kohnen et al., 1995).

    As there is a large range of applicable biomaterials performing various functions,

    all can be a potential location for microbial colonization and infection. Microbial

    contamination of medical devices is caused by a variety of factors. When a device is

    implanted, it immediately elicits local antimicrobial immune responses that can result in a

    congregation of microbials. Most patients who experience microbial infections due to

    implanted medical devices are more susceptible to them as they may be more likely to be

    immunocompromised (Francolini et al., 2003).

    The mechanism for biofilm formation can be characterized as a three phase

    process. Phase 1 consists of initial and irreversible cellular adherence to the polymer

  • 7/31/2019 Bio Compatibility of Polymer Implants

    25/79

    13

    surface. Phase 2 is the expansion and maturation of the biofilm scaffold. Lastly, phase 3

    is the adhesion of individual cells or cellular colonies from the biofilm. These stages are

    depicted in Figure 3 (Francolini et al., 2003).

    Figure 3. Schematic Representation of Stages 1-3 for Biofilm Formation. (modified fromFrancolini et al., 2003)

    Biofilm Prevention

    Upon implantation of a medical device, the human body responds to the device by

    coating it with a protein layer. This layer consists mainly of albumin, laminin, fibrin, and

    fibronectin (Francolini et al., 2004; Jeng et al., 2003). There are several approaches to

    counteract the formation of this protein layer, or biofilm growth, on medical devices.

    One in particular refers to polymers which are chemically modified to prevent primary

    microbial adherence and are able to release antimicrobials to prevent surface growth

    (Francolini et al., 2004).

    Phase 1 Phase 2 Phase 3

  • 7/31/2019 Bio Compatibility of Polymer Implants

    26/79

    14

    A decrease in bacterial adherence occurs by changing the surface charge of the

    polymer. Since most bacteria are negatively charged, a polymer which has a negative

    charge repels bacteria better than an uncharged polymer (Jansen et al., 1995).

    Another way to deter antimicrobial growth and adhesion is by the addition of a

    silver coating. By binding to microbial DNA and sulphydryl groups, silver ions act to

    restrict bacterial replication and deactivate metabolic enzymes (Francolini et al., 2003;

    Petering et al., 1976).

    Amphiphilic Conetworks

    The polymers used in this study for experiments 1 and 2 for the potential

    applications of intervertebral disc repair and/or replacement and pacemaker coatings are

    amphiphilic conetworks (ACPN). Amphiphilic conetworks are two-component

    networks of covalently interconnected hydrophilic/hydrophobic (HI/HO) phases of

    cocontinuous morphology; as such they swell both in water and hydrocarbons, and

    respond to changes in the medium by morphological isomerization (smart networks)

    (Erdodi et al., 2006). ACPNs are responsive to stimuli as they have the ability to

    rearrange their morphology very rapidly. This allows the conetwork to produce favorable

    conformations in reaction to a medium of any polarity as well as amphiphilic protein

    contact (Erdodi et al., 2006).

    Amphiphilic conetworks have the potential to be used in a variety of ways for

    biomedical applications. They are currently used as controlled implantable drug delivery

    devices. In previous work, they exhibit delayed drug delivery profiles, and in some cases

  • 7/31/2019 Bio Compatibility of Polymer Implants

    27/79

    15

    yield desirable zero-order drug delivery kinetics (Chen et al., 1989; Erdodi et al., 2006;

    Ivan et al., 1989, 1990; Kennedy et al., 1990; Keszler, 1993). Most recently, ACPNs

    have been used to support cell proliferation (Erdodi et al., 2006; Haigh et al., 2000,

    2002). Another application was for a thin film, or antimicrobial coating, which showed

    that the biocide was released into the environment over a long time period (Erdodi et al.,

    2006; Tiller, 2005a; Tiller, 2005b). Peroxides that become trapped in the hydrophilic

    domains of the ACPNs have displayed increased activity and stability. The increased

    activity has been attributed to the hydrophilic/hydrophobic interphase of the ACPN

    (Bruns, 2005a; Bruns, 2005b; Erdodi et al., 2006).

    Polymer Biodegradation

    Investigations of polymeric applications in medicine have elucidated serious

    questions about the suitability and degradability of polymers in certain circumstances.

    The stability of sensitive compounds such as protein and peptide drugs as well as the

    ability of living cells to survive in close proximity to an eroding polymer are of concern.

    Research regarding the loss of mechanical stability of polymers during erosion can occur

    too quickly and cause high concentrations of toxicity due to the degradated products. An

    understanding of polymer erosion and degradation is key to the solution of these

    undesirable effects (Gopferich, 1996).

    In this study for both experiments 1 and 2, polymer toxicity was measured by

    lymphocyte infiltration (used as a marker in this thesis). Lymphocytes are inflammatory

    cells which migrate to the wound site to assist in the healing process and rid the body of

  • 7/31/2019 Bio Compatibility of Polymer Implants

    28/79

    16

    any foreign invaders. So the level of polymer toxicity can be directly correlated to the

    amount of lymphocytes found at the implant site. An increased number of lymphocytes

    discovered at the site is indicative of an undesirable chronic inflammatory response

    (Jewrajka et al., 2007).

    The process of degradation refers to the chain scission process in which polymer

    chains are cleaved resulting in the formation of oligomers, then monomers. Erosion is

    described as the loss of material owing to monomers and oligomers leaving the polymer

    (Gopferich, 1996; Tamada et al., 1993).

    There are two principal ways by which polymer bonds can be cleaved: passively

    by hydrolysis or actively by enzymatic reaction (Lenz et al., 1993; Gopferich, 1996).

    The most important criterion for monitoring polymer degradation is an observed decrease

    in molecular weight. Other parameters include loss of mechanical strength and absolute

    chemical degradation into monomers, or monomer release by scission of the polymer

    backbone (Gopferich, 1996).

  • 7/31/2019 Bio Compatibility of Polymer Implants

    29/79

    17

    CHAPTER III

    MATERIALS AND METHODS

    Experiment 1: Polymer Biocompatibility for Intervertebral Disc Repair/Replacement

    It is the goal of experiment 1 to determine if any of the four polymers (polymer,

    monomer, initiator, polymer + monomer combination) could be used for intervertebral

    disc replacement, or to repair any degenerated discs or disc space that has occurred in the

    spine.

    Surgical Procedure

    A group (n=27) of adult male, SHR/y rats were injected with the anesthesia,

    sodium pentathol (50mg/kg,ip) before surgery. For the rats designated for liver

    injections, an incision approximately 8 cm long was made on the ventral side of the

    animal. The liver was exposed and injected with the respective polymer, or saline for

    control.

    For the rats designated for brain injections, an incision approximately 5 cm in

    length was made on dorsal side of the cranium. With the skull exposed, a dremel with

  • 7/31/2019 Bio Compatibility of Polymer Implants

    30/79

    18

    drill bit attachment was used to manufacture a hole adequate for a 10 gauge needle to be

    inserted. The brain was then injected with the respective polymer, or saline for control.

    Polymers used for Brain and Liver Injections

    Monomer: 1-cyanoacryl-2,4,4-trimethylpentane(TMP-CA)

    Polymer: cyanoacrylate terminated tri-arm star PIB [(PIB-CA)3

    Initiator: N,N-diethyl telechelic [(PIB-NEt3)3] PIB

    Monomer + Polymer Combination: cyanoacrylate-telechelic 3-arm star polyisobutylene

    [(PIB-CA)3] + 2,2,4-trimethylpent-1-cyanoacrylate (pTMP-CA)

    The incision made on the ventral side of the animal for the liver injections was

    closed with sutures.

    The incision made on the dorsal side of the cranium for this procedure was closed

    with staples.

    For the group of rats designated for solid polymer liver implants, an incision was

    generated approximately 8 cm long on the ventral side of the animal. A 2 cm incision

    was made on a lobe of the exposed liver, and the respective solid polymer was implanted.

    The incision site was then sutured to prevent the solid implant from leaving the

    appropriated tissue site.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    31/79

    19

    Solid Polymers Implanted in the Liver

    Monomer: poly(1-cyanoacryl-2,4,4-trimethylpentane)poly(TMP-CA)

    Polymer: cross linked [(PIB-CA)3] polymer

    Monomer + Polymer Combination: copolymer of [(PIB-CA)3] + 2,2,4-trimethylpent-1-

    cyanoacrylate (pTMP-CA)

    The incision made on the ventral side of the animal for the liver implants was

    closed with sutures.

    Upon completion of each surgical procedure, suturing and/or stapling, the post-

    operative antibioticadministered to the animals was penicillin (5,000 units, im). Theanimals were housed one to a cage with Sani-Chips bedding (Murphy Forest Products,

    Montville, NJ) changed once per week and food (PMI Nutrition Intl., Brentwood, MO)

    and water were provided ad libitum. The animals were regularly examined for signs of

    pain, infection, or implant/injection rejection.

    Tissue Removal and Histology

    After two weeks the rats were euthanized via a sodium pentothal overdose, and

    the tissues containing the polymer injections and/or implants were removed. The tissue

    samples were then placed in a 10% buffered formalin phosphate solution for

    preservation.

    Tissues were dehydrated and embedded with paraffin in a tissue processor

    (Tissue-Tek, Miles Scientific). The tissue samples were then embedded into paraffin wax

  • 7/31/2019 Bio Compatibility of Polymer Implants

    32/79

    20

    blocks. The tissue blocks were sectioned at 7 microns thick by a Reichert-Jung 820-II

    microtome. Sections were taken at the site of the polymer implantation or injection and

    placed in a 40 degree Celsius water bath. These sections were transferred to gelatin

    coated slides where they were de-waxed, rehydrated, and stained with hematoxylin and

    eosin stain (H&E) for microscopic analysis. The slides were observed through a light

    microscope and micrographs were taken using an Olympus BX 60 microscope, an MTI

    digital camera and the SCION image program. To attain an accurate quantitative

    evaluation of the tissue site where the polymer had been implanted or injected, wall

    thickness, irregular cells, and lymphocyte infiltration were scored to measure tissue

    response. These catergories were blindly scored on a scheme of 0-3 (0 = none, 1 =

    negligible, 2 = moderate, 3 = excessive) (Jewrajka et al. 2007).

    Experiment 2: Polymer Biocompatibility for Pacemaker Coating

    It is the goal of experiment 2 to determine if the polymer tested here, can be used

    as an improved coating for cardiac pacemakers.

    Surgical Procedure

    During experiment 2, each rat was injected with the anesthesia, sodium pentathol

    (50mg/kg,ip) before surgery. For the first group of animals (n=4) designated for polymer

    implants, an incision approximately 8 cm long was made on the ventral side of the

    animal. The peritoneal wall was exposed and the 3 cm long polymer strip was sutured to

  • 7/31/2019 Bio Compatibility of Polymer Implants

    33/79

    21

    the wall at both ends. For the second group of animals (n=4) designated for the control

    Bionate implants, an incision approximately 8 cm long was made on the ventral side of

    the animal. The peritoneal wall was exposed and the 3 cm long polymer or Bionate strip

    was sutured to the wall at both ends.

    Polymer and Bionate Control Implants

    Polymer: Polyisobutylene polytetramethylene oxide

    Bionate: Thermoplastic polycarbonate urethane

    The incision made on the abdominal muscle wall of the animal for the polymer

    and Bionate implants were closed with sutures, and the skin was sealed with staples.

    Upon completion of each surgical procedure, the post-operative antibioticadministered tothe animals was penicillin (5,000 units, im). The animals were housed one to a cage with

    Sani-Chips bedding (Murphy Forest Products, Montville, NJ) changed once per week and

    food (PMI Nutrition Intl., Brentwood, MO) and water were provided ad libitum. The

    animals were regularly examined for signs of pain, infection, or implant rejection.

    Tissue Removal and Histology

    After an interval of four weeks, the rats were euthanized via a sodium pentothal

    overdose, and the tissue areas containing the polymer and Bionate implants were

    removed. The tissue samples were then placed in a 10% buffered formalin phosphate

    solution for preservation.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    34/79

    22

    Tissues were dehydrated and embedded with paraffin in a tissue processor

    (Tissue-Tek, Miles Scientific). The tissue samples were then embedded into paraffin wax

    blocks. The tissue blocks were sectioned at 7 microns thick by a Reichert-Jung 820-II

    microtome. Sections were taken at the site of polymer or Bionate implantation and

    situated in a 40 degree Celsius water bath. These sections were transferred to gelatin

    coated microscope slides and were allowed 24 hours to dry. The slides were then de-

    waxed, rehydrated, and stained with hematoxylin and eosin stain (H&E). The slides were

    observed through a light microscope and micrographs were taken using an Olympus BX

    60 microscope, an MTI digital camera and the SCION image program. To attain an

    accurate quantitative evaluation of the tissue site where the polymer/Bionate had been

    implanted, lymphocyte infiltration, irregular tissue, irregular cells and fat infiltration were

    scored to measure tissue response. These catergories were blindly scored on a scheme of

    0-3 (0 = none, 1 = negligible, 2 = moderate, 3 = excessive) (Jewrajka et al., 2007).

    Statistical Analysis for Experiments 1 and 2

    Statistical tests performed in this study were 1 way ANOVA and t-test. One way

    ANOVA tests were used in each graph to test for significance between injections and

    implants. A two-sample t-test was used by way of SPSS (Statsistal Package for the

    Social Sciences) to compare the four categories scored between the polymer and Bionate

    implants to explore the possibility of any statistically significant differences. Total scores

    and averages for each category respective to experiments 1 and 2 are presented. A

  • 7/31/2019 Bio Compatibility of Polymer Implants

    35/79

    23

    survival table is also shown to focus on the toxicity of the polymers tested. Graphs were

    created using Microsoft Excel. Tables were created using Microsoft Word.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    36/79

    24

    CHAPTER IV

    RESULTS

    Experiment 1: Polymer Biocompatibility for Intervertebral Disc Repair/Replacement

    In experiment 1, three categories were blindly scored on a scale of 0-3, by four

    individuals, to determine wall thickness around the injection/implant site, number of

    irregular cells, and number of lymphocytes present. A score of zero represents a thin

    wall thickness and little or no cells present. A score of three is indicative of a thick wall

    and a considerably high number of cells present at the site.

    Injection/Implant Scoring Averages for All Categories

    Table 2 depicts the scoring averages for each injection/implant to compare each

    category (wall thickness, irregular cells, lymphocytes) scoring average. Injection/implant

    site is shown, as well as surgical notes during removal from each tissue.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    37/79

    25

    Wall Thickness for Liver Injections

    Figure 4 is a graph showing the comparison of wall thickness between polymer

    injections and saline control delivered to the liver (mean, s.e.m., 1 way ANOVA). The

    scoring means for each group were respectively: Monomer (1.44), Initiator (.57),

    Polymer (1.88), Monomer + Polymer (2.25), Saline (.38). There was statistical

    significance between Monomer and Saline; p=.016. There was statistical significance

    between Polymer and Saline; p=.004. There was statistical significance between

    Monomer + Polymer and Saline; p=.003.

    Irregular Cells for Liver Injections

    Figure 5 is a graph which depicts the comparison of irregular cells for injections

    across all treatments administered to the liver (mean, s.e.m., 1 way ANOVA). The

    scoring means for each group were: Monomer (2.38), Initiator (.63), Polymer (.25),

    Monomer + Polymer (1.0), Saline (0). There was statistical significance between

    Monomer and Saline; p=9.73E-5. There was statistical significance between Monomer +

    Polymer and Saline; p=.003.

    Lymphocytes for Liver Injections

    Figure 6 is a graph showing the comparison of lymphocytes between liver

    injections for all treatments (mean, s.e.m., 1 way ANOVA). The scoring means for each

  • 7/31/2019 Bio Compatibility of Polymer Implants

    38/79

    26

    treatment and control group were: Monomer (2.63), Initiator (.82), Polymer (1.13),

    Monomer + Polymer (2.25), Saline (.13). There was statistical significance between

    Monomer and Saline; p=.001. There was statistical significance between Initiator and

    Saline; p=.007. There was statistical significance between Polymer and Saline; p=.001.

    There was statistical significance between Monomer + Polymer and Saline; p=0009.

    Wall Thickness for Brain Injections

    Figure 7 is a graph which shows the comparison of wall thickness between

    polymer injections and saline control delivered to the brain (mean, s.e.m., 1 way

    ANOVA). The scoring means for each treatment and control group were respectively:

    Monomer (.75), Initiator (.63), Monomer + Polymer (1.0), Saline (.82). There was no

    statistical significance between these average scores.

    Irregular Cells for Brain Injections

    Figure 8 is a graph depicting the comparison of irregular cells between brain

    injections for all treatments used (mean, s.e.m., 1 way ANOVA). The scoring averages

    for each brain injection were respectively: Monomer (.50), Initiator (.38), Monomer +

    Polymer (.38), Saline (0). There was statistical significance between Monomer and

    Saline; p=.049. There was statistical significance between Monomer + Polymer and

    Saline; p=.002.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    39/79

    27

    Lymphocytes for Brain Injections

    Figure 9 is a graph showing the comparison of lymphocytes between all

    treatments administered to the brain (mean, s.e.m., 1 way ANOVA). The scoring

    averages for each brain injection were: Monomer (1.25), Initiator (1.75), Monomer +

    Polymer (.63), Saline (.57). There was statistical significance between Initiator and

    Saline; p=.001.

    Wall Thickness for Liver Implants

    Figure 10 is a graph comparing wall thickness between all treatments implanted in

    the liver. Means and s.e.m. were shown in this graph. The scoring averages for each

    liver implant were: Monomer (.69), Polymer (2.38), Monomer + Polymer (.88). There

    was statistical significance between Polymer and Monomer; p=.0002. There was

    statistical significance between Polymer and Monomer + Polymer; p=.0002.

    Irregular Cells for Liver Implants

    Figure 11 is a graph which compares irregular cells between all types of liver

    implants used. Means and s.e.m. were shown in this graph. The scoring averages for

    each implant delivered to the liver were: Monomer (.69), Polymer (1.5), Monomer +

    Polymer (.44). There was statistical significance between Polymer and Monomer +

    Polymer; p=.009.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    40/79

    28

    Lymphocytes for Liver Implants

    Figure 12 is a graph comparing lymphocytes between all treatment implants in the

    liver. Means and s.e.m. were shown in this graph. The scoring averages for each implant

    in the liver were: Monomer (1.25), Polymer (2.0), Monomer + Polymer (1.82). There

    was no statistical significance between treatment groups.

    Survival Rate for Injections/Implants

    Table 3 shows the number of rats injected or implanted with each polymer or

    saline control, site at which the injection or implant was placed, and the number that

    survived the procedure. All injections/implants experienced 100% survival rates except:

    Monomer + Polymer injection delivered to the liver (50%), Monomer + Polymer

    delivered to the brain (50%), Polymer implanted in the liver (50%). Polymer injection

    delivered to the brain (0%).

    Total Scores for Liver Injections

    Figure 13 compares the total score averages across all treatment injections in the

    liver (mean, s.e.m., 1 way ANOVA). Total score represents the sum of the averages for

    all three categories (wall thickness, irregular cells, lymphocytes). Total score averages

    for each treatment was: Monomer (6.45), Initiator (2.02), Polymer (3.26), Monomer +

    Polymer (5.50), Saline (.51). There was statistical significance between Monomer and

  • 7/31/2019 Bio Compatibility of Polymer Implants

    41/79

    29

    Saline; p=.006. There was statistical significance between Initiator and Saline; p=.02.

    There was statistical significance between Monomer + Polymer and Saline; p=.018.

    Total Scores for Brain Injections

    Figure 14 compares the total score averages for brain injections between all

    treatments used (mean, s.e.m., 1 way ANOVA). Total score averages for each treatment

    was: Monomer (2.50), Initiator (2.76), Monomer + Polymer (2.01), Saline (1.39). There

    was no statistical significance between treatment injections and Saline.

    Total Scores for Liver Implants

    Figure 15 compares the total score averages for treatment implants delivered to

    the liver. Means and s.e.m. were shown in this graph. Total score averages for each

    treatment was: Monomer (2.63), Polymer (5.88), Monomer + Polymer (3.14). There was

    significance between Polymer and Monomer; p=.027.

    Histology for Liver Injections

    Figure 16 is a representative section of liver tissue used for the analysis and

    scoring of tissue response to treatment injections. Digital pictures of the slides were

    taken at 40x magnification. Notice the area between the pointers for the combo

    (monomer + polymer) slide. The monomer + polymer injection caused an increased wall

  • 7/31/2019 Bio Compatibility of Polymer Implants

    42/79

    30

    thickness formation around the injection site. Also, notice the increased amount of

    irregular cells between the pointers for the monomer slide.

    Histology for Brain Injections

    Figure 17 is a representative section of brain tissue used for the analysis and

    scoring of tissue responses to treatment injections. Digital pictures of the slides were

    taken at 40x magnification. On the initiator injection slide, there is an increased

    lymphocyte infiltration depicted between the pointers. The combo (monomer + polymer)

    slide shows a significant amount of irregular cells, a result from the injection.

    Histology for Liver Implants

    Figure 18 is a representative section of liver tissue used for the analysis and

    scoring of tissue response to treatment implants. Digital pictures of the slides were taken

    at 40x magnification. The set of pointers labeled A on the polymer slide specifies wall

    thickness caused by the implant. The area between the B labeled pointers represents the

    increased amount of irregular cells present from the polymer implant.

    Experiment 2: Polymer Biocompatibility for Pacemaker Coating

    In experiment 2, four categories were blindly scored on a scale of 0-3, by three

    individuals, to determine lymphocyte infiltration, fat infiltration, irregular cells, and

  • 7/31/2019 Bio Compatibility of Polymer Implants

    43/79

    31

    irregular tissue. A score of zero represented little or no cells/tissue present. A score of

    three was indicative of a considerably high number of cells/tissue present at the site. The

    following table depicts the scoring averages for each category listed as well as notes

    taken during the surgical removal of each sample.

    Scoring Averages Peritoneal Wall Implants and Surgical Notes

    Table 4 depicts the scoring averages for each implant delivered to the peritoneal

    wall to compare each category (lymphocyte infiltration, fat infiltration, irregular cells,

    irregular tissue) scoring average. Surgical notes regarding tissue adherence and polymer

    condition during the removal of each respective tissue is provided.

    Table 5 shows the scoring averages for Polymer and Bionate control implants to

    compare each category respectively. Lymphocyte infiltration average scores for Polymer

    implant (1.9) is compared to Bionate (2.3). Fat infiltration average scores for Polymer

    implant (.8) is in comparison to Bionate (1.4). Irregular cells average scores for Polymer

    (1.1) is compared to Bionate (1.6). Irregular tissue average score for Polymer (1.2) is

    compared to Bionate (1.7).

    Table 6 compares the total score averages for Polymer and Bionate implants

    delivered to the peritoneal wall. Total score represents the sum of the averages for all

    four categories (lymphocyte infiltration, fat infiltration, irregular cells, irregular tissue).

    The total scores were: Polymer (4.99) and Bionate (6.99). There was no statisitcal

    significance.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    44/79

    32

    Table 7 shows the results of a two-tailed t-test to measure any significance

    between categories scored in comparison of Polymer and Bionate implants. There was

    statistical significance for fat infiltration at 95% confidence; p=.032.

    Histology for Polymer Implants

    Figure 19 is a representative section of peritoneal wall tissue used for the analysis

    and scoring of tissue response to polymer implants. Digital pictures of the slides are

    shown. Polymer 1 is depicted at 40x magnification. Polymer 2 and Polymer 3 are shown

    at 100x magnification. The set of pointers on the polymer 1 slide represents an

    aggregation of lymphocytes. The area between the pointers on the polymer 2 slide

    depicts healthy tissue.

    Histology for Bionate control Implants

    Figure 20 is a representative section of peritoneal wall tissue used for the analysis

    and scoring of tissue response to Bionate implants. Digital pictures of the slides were

    taken at 100x magnification. The set of pointers on the Bionate 1 slide represent an area

    of irregular cells. The set of pointers labeled A on the Bionate 2 slide specifies

    irregular tissue. The set of pointers on the Bionate 2 slide labeled B shows fat

    infiltration. The set of pointers on the Bionate 3 slide is representative of lymphocyte

    infiltration in response to the implant.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    45/79

    33

    Table 2. Scoring Averages and Surgical Notes for Each Injection/Implant

    Injection/Implant

    TissueSite

    WallThickness

    IrregularCells

    Lymphocytes Surgical Notes

    Monomer Liver 1.44 2.38 2.63 Abdominal adhesion

    to liverInitiator Liver .57 .63 .82 Abdominal adhesionto liver

    Polymer Liver 1.88 .25 1.13 Minimal adhesionCombo Liver 2.25 1.0 2.25 Minimal adhesionSaline Liver .38 0 .13 Minimal adhesionMonomer Brain .75 .50 1.25 Injection pushed outInitiator Brain .63 .38 1.75 Injection pushed outCombo Brain 1.0 .38 .63 Injection site goodSaline Brain .82 0 .57 Injection site goodSolid

    Monomer

    Liver .69 .69 1.25 Significant adhesion

    SolidPolymer

    Liver 2.38 1.5 2.0 Mesentery adhesion

    SolidCombo

    Liver .88 .44 1.82 Intestinal adhesion

    Table 2. Experiment 1: Scoring Averages and Surgical Notes for Each Injection/Implant.Wall thickness, irregular cells and lymphocytes represent average tissue response scoresfor each implant/injection. A low score denotes a low tissue response vs. a high scorewhich denotes a high tissue response.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    46/79

    34

    Figure 4. Experiment 1: Wall Thickness for Liver Injections. Comparison of wallthickness across treatment injections (means, s.e.m., 1 way ANOVA) in the liver. There

    was statistical significance between Monomer and Saline; p=.016. There was statisticalsignificance between Polymer and Saline; p=.004. There was statistical significancebetween Monomer + Polymer and Saline; p=.003.

    Figure 5. Experiment 1: Irregular Cells for Liver Injections. Comparison of irregular cells

    across treatment injections (means, s.e.m., 1 way ANOVA) in the liver. There wasstatistical significance between Monomer and Saline; p=9.73E-5. There was statisticalsignificance between Monomer + Polymer and Saline; p=.003.

    *

    **

    **

  • 7/31/2019 Bio Compatibility of Polymer Implants

    47/79

    35

    Figure 6. Experiment 1: Lymphocytes for Liver Injections. Comparison of lymphocytesacross treatment injections (means, s.e.m., 1 way ANOVA) in the liver. There was

    statistical significance between Monomer and Saline; p=.001. There was statisticalsignificance between Initiator and Saline; p=.007. There was statistical significancebetween Polymer and Saline; p=.001. There was statistical significance betweenMonomer + Polymer and Saline; p=0009.

    Figure 7. Experiment 1: Wall Thickness for Brain Injection. Comparison of wallthickness across treatment injections (means, s.e.m., 1 way ANOVA) in the brain. Therewas no statistical significance between treatment injections and Saline.

    ***

    ***

    ***

    **

  • 7/31/2019 Bio Compatibility of Polymer Implants

    48/79

    36

    Figure 8. Experiment 1: Irregular Cells for Brain Injections. Comparison of irregular cellsacross treatment injections (means, s.e.m., 1 way ANOVA) in the brain. There was

    statistical significance between Monomer and Saline; p=.049. There was statisticalsignificance between Monomer + Polymer and Saline; p=.002.

    Figure 9. Experiment 1: Lymphocytes for Brain Injections. Comparison of lymphocytesacross treatment injections (means, s.e.m., 1 way ANOVA) in the brain. There wasstatistical significance between Initiator and Saline; p=.001.

    **

    *

    ***

  • 7/31/2019 Bio Compatibility of Polymer Implants

    49/79

    37

    Figure 10. Experiment 1: Wall Thickness for Liver Implants. Comparison of wallthickness across treatment implants (means, s.e.m., 1 way ANOVA) in the liver. There

    was statistical significance between Polymer and Monomer + Polymer; p=.0002. Therewas statistical significance between Monomer and Polymer; p=.0002.

    Figure 11. Experiment 1: Irregular Cells for Liver Implants. Comparison of irregular cellsacross treatment implants (means, s.e.m., 1 way ANOVA) in the liver. There wasstatistical significance between Polymer and Monomer + Polymer; p=.009.

    ******

    **

  • 7/31/2019 Bio Compatibility of Polymer Implants

    50/79

    38

    Figure 12. Experiment 1: Lymphocytes for Liver Implants. Comparison of lymphocytesacross treatment implants (means, s.e.m., 1 way ANOVA) in the liver. There was no

    statistical significance between implant groups.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    51/79

    39

    Table 3. Animal Survivors Associated with Polymer Type

    Type Site Number Injected Number Survived Survival Rate

    Monomer Liver 2 2 100%

    Initiator Liver 2 2 100%Polymer Liver 2 2 100%Combo Liver 2 1 50%Saline Liver 2 2 100%

    Monomer Brain 3 1 33%

    Initiator Brain 2 2 100%Polymer Brain 2 0 0%Combo Brain 2 1 50%Saline Brain 2 2 100%Solid

    MonomerLiver 2 2 100%

    SolidPolymer

    Liver 2 1 50%

    SolidCombo

    Liver 2 2 100%

    Table 3. Experiment 1: Animal Survivors Associated with Polymer Type. Number ofanimals which received the respective implant and/or injection and the number of thoseanimals that survived the duration of the experiment.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    52/79

    40

    Figure 13. Experiment 1: Total Mean Scores for Each Group with Liver Injections.Comparison of total scores across all treatment injections (means, s.e.m., 1 way

    ANOVA) in the liver. Total score represents the sum of the averages for all threecategories (wall thickness, irregular cells, lymphocytes). There was statisticalsignificance between Monomer and Saline; p=.006. There was statistical significancebetween Initiator and Saline; p=.02. There was statistical significance between Monomer+ Polymer and Saline; p=.018.

    Figure 14. Experiment 1: Total Mean Scores for Each Group with Brain Injections.Comparison of total scores across all treatment injections (means, s.e.m., 1 wayANOVA) in the brain. Total score represents the sum of the averages for all threecategories (wall thickness, irregular cells, lymphocytes). There was no statisticalsignificance between treatment injections and Saline.

    **

    *

    *

  • 7/31/2019 Bio Compatibility of Polymer Implants

    53/79

    41

    Figure 15. Experiment 1: Total Mean Scores for Each Group with Liver Implants.Comparison of total scores across all treatment implants (means, s.e.m., 1 way ANOVA)

    in the liver. Total score represents the sum of the averages for all three categories (wallthickness, irregular cells, lymphocytes). There was statistical significance betweenMonomer and Polymer; p=.027.

    *

  • 7/31/2019 Bio Compatibility of Polymer Implants

    54/79

    42

    Figure 16. Experiment 1: Representative Section for Tissue Analysis for Liver Injections.Slides were used for tissue response scoring and analysis. All five pictures here are shown at40x magnification. Combo refers to Monomer + Polymer injection. Notice the area betweenthe pointers for the combo slide. The combo injection caused an increased wall thicknessformation around the injection site. Also, notice the increased amount of irregular cellsbetween the pointers for the monomer slide.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    55/79

    43

    Figure 17. Experiment 1: Representative Section for Tissue Analysis for Brain Injections.Slides were used for tissue response scoring and analysis. All five pictures here areshown at 40x magnification. Combo refers to monomer + polymer injection. On theinitiator injection slide, there is an increased lymphocyte infiltration depicted between thepointers. The combo slide shows a significant amount of irregular cells, a result from theinjection.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    56/79

    44

    Figure 18. Experiment 1: Representative Section for Tissue Analysis for Liver Implants.Slides were used for tissue response scoring and analysis. All five pictures here areshown at 40x magnification. Combo refers to monomer + polymer injection. The set ofpointers labeled A on the polymer slide specifies wall thickness caused by the implant.The area between the B labeled pointers represents the increased amount of irregularcells present from the polymer implant.

    A

    B

  • 7/31/2019 Bio Compatibility of Polymer Implants

    57/79

  • 7/31/2019 Bio Compatibility of Polymer Implants

    58/79

    46

    Table 5. Scoring Averages for Polymer and Bionate control Implants

    Implant Lymphocyte

    Infiltration

    Fat

    Infiltration

    Irregular

    Cells

    Irregular

    Tissue

    Polymer 1.9.11 0.8.22 1.1.26 1.2.22

    Bionate

    Control

    2.3.33 1.4.18 1.6.18 1.7.17

    Table 5. Experiment 2: Scoring Averages for Polymer and Bionate Control Implants.Lymphocyte infiltration, fat infiltration, irregular cells and irregular tissue scoresrepresent the average for each respective implant in the peritoneal wall. A high score isindicative of a high tissue response vs. a low score which denotes a low tissue response.

    Table 6. Total Mean Scores for Peritoneal Wall Implants

    Implant Total Score

    Polymer 4.99.23

    Bionate Control 6.991.19

    Table 6. Experiment 2: Total Mean Scores for Peritoneal Wall Implants. Comparison oftotal scores for both treatment implants (means, s.e.m., 1 way ANOVA) in the peritoneal

    wall . Total score represents the sum of the averages for all four categories (lymphocyteinfiltration, fat infiltration, irregular cells, irregular tissue). There was no statisticalsignificance.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    59/79

    47

    Table 7. Two-Tail T-Test for Comparison between Polymer and Bionate across AllCategories Scored

    Levenes Test

    for Equality of

    Variances t-test for Equality of Means

    95% Confidence

    Interval of the

    Difference

    F Df

    Sig. (2-

    tailed)

    Std. Error

    Difference Lower Upper

    Lymphocyte

    Infiltration

    Equal variances assumed 24.123 16 .224 .35136 -1.18930 .30041

    Equal variances not assumed 9.756 .235 .35136 -1.22999 .34110

    Fat

    Infiltration

    Equal variances assumed .038 16 .032 * .28328 -1.26719 -.06614

    Equal variances not assumed 15.191 .032 .28328 -1.26980 -.06353

    Irregular

    Cells

    Equal variances assumed .400 16 .176 .31427 -1.11067 .22178

    Equal variances not assumed 14.027 .179 .31427 -1.11836 .22947

    Irregular

    Tissue

    Equal variances assumed .291 16 .129 .27778 -1.03331 .14442

    Equal variances not assumed 14.837 .131 .27778 -1.03708 .14819

    Note. Comparison of polymer and Bionate control for all four categories (lymphocyteinfiltration, fat infiltration, irregular cells, irregular tissue). F=F variance, df=degrees offreedom. * denotes significance.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    60/79

    48

    Figure 19. Experiment 2: Representative Sections for Tissue Analysis for Peritoneal WallPolymer Implants. Polymer 1 is shown at 40x magnification. Polymer 2 and Polymer 3are shown at 100x magnification. The set of pointers on the polymer 1 slide represents anaggregation of lymphocytes. The area between the pointers on the polymer 2 slide depictshealthy tissue.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    61/79

    49

    Figure 20. Experiment 2: Representative Sections for Tissue Analysis for Peritoneal WallBionate Implants. Slides were used for tissue response scoring and analysis. All fivepictures here are shown at 100x magnification. The set of pointers on the Bionate 1 sliderepresent an area of irregular cells. The set of pointers labeled A on the Bionate 2 slidespecifies irregular tissue. The set of pointers on the Bionate 2 slide labeled B shows fatinfiltration. The set of pointers on the Bionate 3 slide is representative of lymphocyteinfiltration in response to the implant.

    A

    B

  • 7/31/2019 Bio Compatibility of Polymer Implants

    62/79

    50

    CHAPTER V

    DISCUSSION

    Experiment 1: Polymer Biocompatibility for Intervertebral Disc Repair/Replacement

    The goal for Experiment 1 was to determine if any of the four polymers (polymer,

    monomer, initiator, polymer + monomer combination) tested in the brain and liver could

    be used for intervertebral disc replacement, or to repair any degenerated discs or disc

    space that occurred in the spine. It was hypothesized that the polymer sample will elicit

    little or no significant tissue damage, and will therefore be best suited for intervertebral

    disc replacement and/or repair.

    Liver and Brain Injections

    Prior experimentation regarding the polymer (cyanoacrylate terminated tri-arm

    star PIB [(PIB-CA)3) and monomer + polymer combination (cyanoacrylate-telechelic

    3-arm star polyisobutylene [(PIB-CA)3] + 2,2,4-trimethylpent-1-cyanoacrylate (pTMP-

    CA)) tested for swelling, extractables, and oxidative resistance among others.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    63/79

    51

    Under well defined laboratory conditions, these materials exhibited mechanical

    properties promising for medical applications, specifically intervertebral discs (Jewrajka

    et al., 2008).

    The liver was chosen as an alternative tissue site to the brain to evaluate potential

    biocompatibility issues with the polymers. Tissue damage analysis for liver injections

    showed three treatments: monomer, polymer, and monomer + polymer all having a

    higher wall thickness in comparison to the saline control. The initiator may have

    recorded the lowest wall thickness for the liver injections because of its lack of

    cyanoacrylate, found in the above three mentioned treatment injections. The

    cyanoacrylate has induced an increased immune response, in particular wall thickness

    around the injection site. Literature supports this result as the polymerization of CA is

    exothermic, and the heat released may be responsible for the cell damage in cell culture

    (Leggat et al., 2004). Another possible mechanism for CA toxicity refers to the by-

    products of degradation: cyanoacetate and formaldehyde, responsible for evoking an

    intense inflammatory response (Schwade et al., 2008).

    When analyzing irregular cells in the liver it was determined that the monomer

    and monomer + polymer injections elucidated an increased response in comparison to the

    saline control. This result may be due to the 2, 2, 4-trimethylpentane-1-cyanoacrylate

    (TMP-CA) groups that are found in these samples. The polymer and initiator groups did

    not contain TMP-CA, and recorded scores that were not statistically significant from the

    saline control.

    Tissue analysis showed three treatments as having higher scores for lymphocytes

    in the liver when compared to saline: initiator, polymer, and monomer + polymer.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    64/79

    52

    Although the liver responded to the monomer with an increase in wall thickness around

    the injection site, lymphocytes were not statistically significant when compared to the

    saline injection. This result may be due to the absence of polyisobutylene (PIB), which is

    found in the other three treatment injections. The initiator, polymer, and monomer +

    polymer may have activated an increase in lymphocyte infiltration in response to an

    infection caused by the PIB.

    In support of the hypothesis, the irregular cell score of the polymer injection to

    the liver was very similar to that of the saline injection. So although an initial immune

    response was evoked in the form of lymphocyte infiltration and increased wall thickness,

    no tissue damage in the form of irregular cells occurred in comparison to the saline

    control. These results suggest negligible polymer toxicity, and little or no cellular

    damage.

    It is difficult for the chemicals to leak out from the polymer injections and enter

    the central nervous system (CNS) via a systemic route, due to the presence of the blood

    brain barrier (Kou et al., 1997). This makes it difficult to determine biocompatibility

    when referring to a potential polymer application that will be located in the CNS having

    direct contact with cerebral spinal fluid (CSF). Therefore, treatment injections into the

    brain were performed as suggested by Dr. Kennedy. The brain is among the softest of

    biological tissues, but was chosen as a target tissue site to help create an environment

    similar to that of the spinal column, in which the polymer implanted there would be

    bathed in CSF.

    It was found that monomer and monomer + polymer injections into the brain

    resulted in an increase in irregular cells compared to the saline control. This finding is

  • 7/31/2019 Bio Compatibility of Polymer Implants

    65/79

    53

    exactly the same as the results for irregular cell scores recorded for the liver injections in

    this study. This finding in the brain once again suggests that cellular damage in the form

    of irregular cells was due to the TMP-CA in the monomer and monomer + polymer

    combination. Different percentage compositions should be examined in the future for

    better biocompatibility results.

    Tissue damage analysis showed an elevated lymphocyte infiltration in the brain

    for the initiator injection compared to the saline control. To explain this result, the

    initiator was the only sample that contained PIB without TMP-CA. The absence of this

    chemical interaction may have caused this increased immune response in the brain. The

    monomer contained TMP-CA, but not PIB, and resulted in the second highest average

    score for this category. Perhaps the PIB with TMP-CA is more biocompatible at this

    tissue site in regards to lymphocyte infiltration. This suggestion can be supported, as the

    monomer + polymer combination recorded a score (.63) very similar to the saline

    injection (.57).

    Liver Implants

    The polymer implant recorded a higher statistically significant score for wall

    thickness when compared separately to the monomer and monomer + polymer implants.

    This may be the result of the polymer implant not having a TMP-CA group attached.

    The monomer and monomer + polymer implants both contained TMP-CA, and had a

    lesser tissue response in the liver.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    66/79

    54

    The polymer implant also recorded a higher statistically significance score for

    irregular cells when compared to the monomer + polymer combination. The lesser score

    of the monomer + polymer supports the idea that an implant, which is composed of PIB

    and TMP-CA, may be the most biocompatible. So a polymer composed of PIB alone

    may not be as biologically compatible as compared to PIB-TMP-CA.

    Total Scores for Injections and Implants in Liver and Brain

    The monomer injected in the liver recorded the highest total tissue response score

    (6.45) and statistical significance was noted compared to the saline control. The initiator

    and monomer + polymer treatments also recorded a higher statistically significant score

    when compared to the control. The polymer injection recorded a total score that was

    found not to be statistically significant compared to saline.

    The solid polymer implant in the liver registered the highest total score (5.88).

    This finding contradicts the result for total score regarding the polymer hydrogel injected

    in the liver. The difference seen in the study may be explained by the morphology of the

    polymer implant vs. polymer injection. Biomaterial interactions with tissues are directed

    by surface molecules, specific receptors, atomic geometry and the electronic state of the

    biomaterial surface (Gadkari et al., 1989). It appears that both of these polymer forms are

    not biocompatible.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    67/79

    55

    Summary

    It was very difficult to find literature pertaining to toxic polymer effects in the

    liver and brain for the application of intervertebral disc replacement. Searches were

    carried out on various websites, electronic journal databases, and search engines

    including: Google.com, Googlescholar.com, Dogpile.com, and OhioLink. Experiments

    involving polymer injections and/or implants in the brain were inclusive of successful

    biodegradable polymers associated with controllable drug carriers.

    The brain may be the most important site to focus our attention to when

    examining these results, because as mentioned above, this area has a similar environment

    to that of the spinal column with the presence of cerebral spinal fluid (CSF). There are

    some drawbacks with brain injections which include the risk of infection due to very

    frequent reservoir replenishment and peripheral toxicity due to rapid CSF turnover (Kou

    et al., 1997). The chemical makeup of the monomer, polymer, and monomer + polymer

    combination is the most logical explanation for this increased rate of mortality (~ 55%)

    with these injections. The samples are responsible for delivering toxic effects to the

    CNS.

    The initiator generated the highest total score (2.76) for the brain injections, but

    there was not any significant difference between it and the saline control. The polymer

    injection failed to record a score here, because both animals injected failed to survive.

    Any type of injection into the brain has to be done very carefully as room for error is

    small, but I do not believe technique was to blame for this particular result. Both the

    initiator and saline injections for this group achieved a 100% survival rate. Two of the

  • 7/31/2019 Bio Compatibility of Polymer Implants

    68/79

    56

    three animals injected with the monomer at this site died, as well as one of the two

    animals injected with the monomer + polymer combination. It was discovered that the

    animal with the monomer injection into the brain that did survive had actually pushed the

    injection out of the site. Had the injection stayed in the brain, this animals chance of

    surviving the two week time interval may have significantly decreased. Based on this

    data, it seems as if any injection into the brain other than that of the initiator or saline

    control may illicit negative effects on the animal model, most likely leading to death.

    Only one of the two animals receiving the solid polymer implant survived. So the

    polymer hydrogel and solid form elicited a high degree of mortality (75%) for animals

    receiving these treatments, deeming this particular sample extremely toxic.

    In 16 of 20 animal models, tissue involvement/adherence was noted. This would

    undoubtedly be problematic when pertaining to a prosthetic polymer disc or polymer

    injection into intervertebral disc space. Of the four animals that did not show tissue

    adhesion, two of the injections (initiator and monomer) were displaced from the target

    site (brain), one was a Saline injection into the brain, and the other was a monomer +

    polymer combination injection into the brain.

    Experiment 2: Polymer Biocompatibility for Pacemaker Coating

    The goal for Experiment 2 was to determine if the polymer tested, in the

    peritoneal wall, could be used as an improved coating for cardiac pacemakers. It was

    hypothesized that the polymer implant will elicit little or no significant tissue damage,

  • 7/31/2019 Bio Compatibility of Polymer Implants

    69/79

    57

    nor will it degrade itself, and will therefore be a suitable candidate for a cardiac

    pacemaker coating.

    It should be noted that tissue sections were not obtained from two of eight animals

    in this study. The two tissue samples were extremely brittle, dry, and flaky during the

    sectioning process. I met with Dr. Ely and Dr. Salisbury to discuss the matter and it first

    was suggested to me to soak the wax block samples in 10% ammonia hydroxide for ten

    minutes and then try to section them. After doing this the sections still remained brittle

    and dry, so I soaked the wax blocks for three hours. Again, I tried to section each sample

    to no avail. I then spoke with Dr. Salisbury once more and we decided to reprocess the

    tissue samples. The tissues were removed from their wax blocks and placed in a cassette

    to soak in Hemo-De for five days to help remove the paraffin wax still infiltrating the

    tissue. After this, the tissue samples were removed from the cassettes and placed in a

    cartridge (approximately 2.5x 1 in.) containing Hemo-De, and placed on a rotator for six

    days to remove any wax that may have still been present in the tissue. The tissues were

    then reprocessed and re-embedded in paraffin wax. Sectioning resumed with the

    reprocessed tissue samples and the same result was found as before; the samples

    remained too dry to section properly.

    Lymphocyte infiltration, irregular cells, and irregular tissue are all signs of the

    bodys immune response to a foreign material, in this case, polymer implants onto the

    peritoneal wall. A high score in any of these areas is indicative of tissue damage and

    overall poor biocompatibility.

    The Bionate control implant achieved the highest total score (6.99) compared to

    the polymer implant (4.99). This result suggests that although no statistical significance

  • 7/31/2019 Bio Compatibility of Polymer Implants

    70/79

    58

    between the polymer and Bionate groups was noted, the polymer implant did elicit less of

    a tissue response versus the Bionate control. In fact, the polymer induced less of a tissue

    response in every category scored, including a statistically significant difference in the fat

    infiltration category. The hydrophilic and hydrophobic cross-linked domains that

    compose this amphiphilic polymer allows for the counter-current diffusion of water and

    oxygen and better overall performance compared to the control (Jewrajka et al., 2007).

    In 8 of 8 animals, fibrous encapsulation of the polymer and fat adherence to it

    was noted. There was no evidence of fluid accumulation or a visual infectious response.

    None of the animals demonstrated any signs of implant rejection (implants protruding

    from the skin) (Jewrajka et al., 2007). In support of the fat adherence finding, a similar

    result was discovered in a study conducted for a polymer (polypropylene) implanted in

    the abdominal wall. The tissue site where the polymer was implanted was removed after

    four and eight week time periods. Fat infiltration was found after both durations, but was

    greater after twelve weeks (Kaleya, 2005).

    The adipose tissue adherence is an interesting phenomenon. The reason for the

    fat adhesion to the implants may be a line of defense associated with these animal

    models. The fat may have migrated from the epididymis to the implant site on abdominal

    wall to protect any potential injury to vital organs

    (http://massage.largeheartedboy.com/archive/2009/01/role_of_fat.html).

  • 7/31/2019 Bio Compatibility of Polymer Implants

    71/79

    59

    CHAPTER VI

    CONCLUSION

    Experiment 1: Polymer Biocompatibility for Intervertebral Disc Repair/Replacement

    As the progression of intervertebral disc degeneration fails to cease, research

    involving candidate biomaterials for the spinal column increases. Improving implant

    design and delivery to the target site is of importance, but biocompatibility and durability

    of the polymer is of chief concern.

    It is extremely difficult to support my hypothesis from the data presented in this

    experiment. The solid polymer implant into the liver recorded the highest total score

    compared to the other implants and one out of two animals implanted with the solid

    polymer failed to survive. Also, there were no survivors with the polymer injection into

    the brain.

    The in vivo findings in this study contradict the in vitro results discovered by

    Jewrajka et al. Indeed physiological settings are difficult to reproduce in the laboratory,

  • 7/31/2019 Bio Compatibility of Polymer Implants

    72/79

    60

    but the polymer tested here was extremely toxic and not well tolerated in the brain tissue,

    thus the hypothesis was rejected.

    To understand the mechanisms involved in the immunological response for these

    implants and injections requires a close relationship between multiple disciplines such as

    polymer science, biomedical engineering, and biology.

    Experiment 2: Polymer Biocompatibility for Pacemaker Coating

    As a medical device coating becomes susceptible to chemical degradation it

    becomes brittle and cracking ensues. Biological enzymes attack these weakened areas

    and can promote further damage via electrical malfunction of the pacemaker itself.

    Improved polymer coatings for pacemakers should prevent cracking and biodegradation

    and serve as a successful protective barrier for pacemaker.

    Both implants in this experiment failed to repel tissue adherence completely, but

    lower inflammatory values in each category scored were recorded for the polymer

    compared to the Bionate control. Fat adherence, which was noted in each implant, was

    significantly less in the polymer implant compared to the control. The low tissue

    inflammatory response suggested by the data supports the hypothesis that the polymer

    implant will elicit little or no significant tissue damage.

    All four polymers recovered were visually in tact, meaning they looked the same

    post-surgery as they did pre-surgery without tears or physical damage. It is my belief

    that this sample is both biocompatible and biostable, thus making it suitable as a synthetic

    polymer pacemaker coating.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    73/79

    61

    The polymer coating tested in experiment two may have other roles in future

    medical devices. Applications for this polymer may include catheter and cardiac stent

    coatings, but future research is needed in the area of biological durability. This study

    requires further research to test the long-term biostability of this material in physiological

    conditions as in vivo testing involves much more complex biological effects on polymers

    and does not always support in vitro testing.

  • 7/31/2019 Bio Compatibility of Polymer Implants

    74/7