Welcome
“An Implant’s Design Dictates its Clinical Capabilities”
The Bicon System was designed not as a research project to study osseointegration, but rather as a means to restore dentition. Bicon’s unique 1.5˚ locking taper implant to abutment connection follows sound bioengineering principles and provides for 360° of universal abutment positioning. Bicon’s implant to abutment connection also has been definitively proven to be bacterially sealed. The implant’s elegant plateaued design, provides for cortical-like bone with central vascular systems around the implant. This cortical-like bone not only grows faster, but also
provides for functionally different capabilities than the appositional bone around non-plateaued implants. The implant’s sloping shoulder provides sufficient space for the interproximal papillae, which are crucial for gingivally aesthetic restorations. Since its introduction in 1985, the Bicon design has benefited from a sensible biological width, which is only now—some 20 years later—being promoted as switching platforms. For these reasons, Bicon clinicians and their patients do not experience the frustrations and limitations inherent in other implant systems.
800-88-BICON B
SHORT IMPLANTSSince 1985 » Bicon Short Implants maximize implant placement possibilities and minimize the need for grafting procedures. With Bicon, longer implant lengths are not necessarily better. For many clinical situations, shorter implants offer a better solution.
EXTRA-ORAL CEMENTATION & THE IAC™Since 1985 » With the elimination of screws, Bicon’s restorative procedures are conventional, requiring only standard impression techniques and allowing for intra-oral or extra-oral cementation techniques. Because of Bicon’s 360˚ of universal abutment positioning, Bicon introduced the revolutionary Integrated Abutment Crown™ (IAC), a screwless and cementless restoration. The IAC affords the dentist a guaranteed aesthetic subgingival crown margin for every restoration, with no extra effort or expense for the dentist or technician.
LOW-SPEED DRILLINGSince 1985 » Low-speed drilling at 50 RPM without irrigation allows a clinician to harvest the patient’s own bone with titanium reamers for autogenous grafting. Slow drilling is forgiving and is unique to Bicon. Additionally, it greatly extends the longevity of the titanium reamers, thus reducing a clinician’s costs.
1mm 10µm 1.5˚ LOCKING TAPERSince 1985 » Bicon’s 1.5 degree locking taper connection provides a proven bacterial seal at the implant to abutment interface, with a microgap of less than 0.5 microns. Bicon’s bacterial seal avoids the microbial leakage issues that can result in inflammation of the soft tissue around an implant, which could lead to not only bone loss around the implant but also to the loss of the implant itself.Image courtesy of Ziedonis Skobe, PhD, Forsyth Institute and Harvard University, Boston, MA and Thomas G.H. Diekwisch, DDS, PhD, UIC College of Dentistry, Chicago, IL
Image courtesy of Paulo G. Coelho, Ph.D., New York University
PLATEAU DESIGNSince 1985 » The plateau or fin design offers at least 30% more surface area than a screw implant of the same dimensions and allows for the callus formation of mature haversian bone between the fins of the implant. This cortical-like bone forms at a faster rate of 10-50 microns per day in comparison to the appositional bone around non-plateaued implants, which forms at a slower rate of 1-3 microns per day.
HAVERSIAN BONE
BICON IMPLANT
THE BICON SYSTEM
PLATFORM SWITCHINGSince 1985 » Bicon features complete interchangeability of abutment diameters, providing sensiblebiological width — now discussed as platform switching. Bicon’s harmonious biological designprovides for impressive bone maintenance.
4.5mmImplant
5.0mmAbutment
RESTORATIVE FLEXIBILITYSince 1985 » Bicon offers a complete selection of abutments: temporary, healing, stealth shouldered, non-shouldered, fixed-detachable, Brevis™ overdenture, and Locator® overdenture — providing for exceptional restorative flexibility and platform switching. All Bicon abutments are completely interchangeable, and all benefit from the unique 360˚ of universal positioning provided by Bicon’s locking taper connection. Once clinicians appreciate what 360˚ of abutment positioning can do for their clinical practice, implant dentistry will never again be the same for them.
SLOPING SHOULDERSince 1985 » Bicon’s sloping shoulder affords more flexibility at the time of implant placement and provides for impressive bone maintenance. It also provides more room for bone over the implant, which provides support for the interdental papillae, enabling aesthetic gingival contours to be easily and consistently achieved.
4.0mmImplant
4.0mm
2.0mm
C www.bicon.com
The success of all dental implants is highly susceptible to surgical technique which must be meticulously performed. Careful patient selection and final restorative prostheses appropriate to individual patient needs and remaining anatomy are essential. Selection of individual or multiple implants of appropriate size and configuration for the existing anatomy is also crucial if success is to be expected.
For the above reasons, Bicon strongly recommends that all dentists attend one or more endosseous dental implant courses or symposia covering the use of the Bicon dental implant system before attempting to incorporate implants into their dental practices. Because Bicon should not and cannot control factors within the scope of services performed by the dental practitioner including patient selection, surgical technique and restorative techniques, Bicon assumes no responsibility beyond replacement of product for failures or other unfavorable reactions or results to or from any of its buyers, any dental practitioners or patients arising out of or through use of a Bicon implant. Bicon implants are manufactured from surgical grade titanium alloy (Ti 6Al 4V) and are supplied as hydroxylapatite coated (HA), or Integra-Ti™, and are considered by biomaterial researchers to be highly biocompatible in numerous animal and human studies which have been conducted over a period of many years. However, as with all such coatings, the possibility of chipped surfaces resulting from insertion of the implant into the bone is always a possibility. Some implants may have some minor chips present in the HA coating when received. These surface imperfections are not thought to affect the overall function of the HA coated implant. Furthermore, while many authorities claim that HA coatings are non-resorbable, the
material has not been used for a long enough period of time to establish the validity of this claim. Bicon implants and titanium abutments are exceptionally resistant to bending and fracture. It is not possible, however, to guarantee that fracture or bending of an implant or abutment cannot or will not occur from traumatic injury or excessive occlusal loads. For the foregoing reasons, except as specifically stated herein, Bicon makes no warranties extending beyond the representation contained in this warranty or beyond the description contained in any invoice. This warranty is in lieu of and excludes all other warranties, expressed or implied including merchantability and fitness for a particular purpose. Bicon sells and warrants to buyers who are licensed dental practitioners and who purchase implants and abutments for purposes of their use and resale as part of their services which they perform for patients. Bicon makes no written warranty to any consumer, patient or end user and authorizes no person to make any such written warranty on its behalf. The remedies of any buyer from Bicon (and its liabilities) are limited solely and exclusively to replacement or repair of a defective Bicon implant or abutment when that product has been demonstrated to Bicon authorities within six (6) months of delivery to the buyer to have been defective at the time of shipment and is returned to Bicon for inspection, replacement, or repair. Bicon is not liable to any person for any incidental, consequential or other damages of any kind whatsoever, whether any claim or potential claim is based upon theories of contract, negligence, or tort.
The Bicon Dental Implant System Warranty and Limitations
D800-88-BICON
Bicon Implant Packaging
12 3
4
1 Product part number.
2 Manufacturing lot number.
3 Sterilization expiration date.
4 Product description.
5
8
6
7
5 European CE Mark.
6 Product is packaged sterile.
7 Caution: read enclosed information.
8 Single use only.
9 U.S. and International Patents:
4,738,623 5,484,285 5,499,918 5,749,731 5,827,062 5,888,066 6,200,345 6,227,857 6,290,500 6,325,628 6,592,370 6,688,887 6,991,462 7,114,952 7,214,062
EP1037569 EP1039840 EP1006927 EP1017327 EP19950928647
AU9074901 AU1811399 AU1811299 AU6039398 AU6246298 AU3232695
JP2001508686T
WO 03-075783 A 1 WO9929256 WO9929255 WO9834562 WO9831300 WO9534249
9
Implants ................................................................................. 1
Implants ............................................................................... 3
Implant Level Impressions ............................................ 5
Implant Level Impression Components ........................ 6
Abutments ............................................................................ 7
Temporary Abutments...................................................... 9
Healing Abutments .........................................................11
Stealth Transitional Abutments ....................................13
Non-Shouldered Abutments .........................................15
Stealth Shouldered Abutments ....................................23
Brevis™ Abutments ..........................................................27
Locator® Abutments ........................................................29
Fixed-Detachable Abutments .......................................31
Bicon Bar Abutments.......................................................32
Abutment Selection Board ............................................33
Abutment Measurement Guide....................................34
Regenerative Products ................................................ 35
SynthoGraft™ ....................................................................37
Resorbable Collagen Membrane ........................................... 40
Resorbable Collagen Plug ........................................................... 40
Bicon Kits ............................................................................ 41
Instruments ....................................................................... 53
Drill Unit and Handpieces ..............................................55
Handpiece Compatibility and Maintenance ..............56
Table of Contents
E www.bicon.com
Table of Contents
Instrumentation ...............................................................57
Restorative Accessories...................................................63
Surgical Accessories.........................................................65
Transitional Implants.................................................... 67
Education.............................................................................71
Demonstration Models and Replicas...........................74
Reference ............................................................................ 77
Single Crown and Bridge Restorations........................79
Non-Shouldered Abutments
and Prosthetic Components ..........................................81
Stealth Shouldered Abutment System........................83
Pilot Drill ............................................................................. 85
Latch Reamers...................................................................86
Return Policies and Procedures ....................................87
Fax Order Form .................................................................88
F800-88-BICON
3 www.bicon.com
Impl
ants
➤ Implant Selection Considerations ■ All implants are packaged with a healing plug.
■ The choice of implant length and width depends upon available bone and expected occlusal loads.
■ The Bicon implant is designed, for an aesthetic emergence profile, to be placed below the crest of the alveolar ridge to the extent that the implant is 5.0mm below the facial crest of the gingiva.
■ The implant may be placed at or slightly above the alveolar crest, if an aesthetic emergence profile is not a concern.
■ In general, choose the widest but not necessarily the longest implant the bone will allow.
■ Place one implant per tooth, posteriorly to the canine, if practical.
■ Use HA coated implants for poor bone quality or grafted bone cases.
■ See reference pages 85-86 for depth indicators on the 2.0mm pilot drill and the latch reamers.
Implants 2.0mm Well
Diameter and Coating Length Well Part Number
3.5 x 8 3.5 x 11
3.5mm HA Coated Implant 8.0mm 2.0mm 260-335-008
3.5mm HA Coated Implant 11.0mm 2.0mm 260-335-011
3.5mm Integra-Ti™ Implant 8.0mm 2.0mm 260-135-008
3.5mm Integra-Ti™ Implant 11.0mm 2.0mm 260-135-011
Diameter and Coating Length Well Part Number
4 x 8 4 x 11
4.0mm HA Coated Implant 8.0mm 2.0mm 260-340-008
4.0mm HA Coated Implant 11.0mm 2.0mm 260-340-011
4.0mm Integra-Ti™ Implant 8.0mm 2.0mm 260-140-008
4.0mm Integra-Ti™ Implant 11.0mm 2.0mm 260-140-011
Color Coding:
6.0mm diameter5.0mm diameter4.5mm diameter4.0mm diameter3.5mm diameter
Healing Plug Inserter2.0mm
Description Length Part Number
2.0mm Healing Plug Inserter* 16.0mm 260-100-002
2 x 16 Example
*Packaged with each 3.5 and 4.0mm diameter implant.
4800-88-BICON
Implants
Implants 3.0mm Well
Diameter and Coating Length Well Part Number
4.5 x 6 4.5 x 8 4.5 x 11
4.5mm HA Coated Implant 6.0mm 3.0mm 260-345-306
4.5mm HA Coated Implant 8.0mm 3.0mm 260-345-308
4.5mm HA Coated Implant 11.0mm 3.0mm 260-345-311
4.5mm Integra-Ti™ Implant 8.0mm 3.0mm 260-145-308
4.5mm Integra-Ti™ Implant 11.0mm 3.0mm 260-145-311
Diameter and Coating Length Well Part Number
5 x 6 5 x 8 5 x 11
5.0mm HA Coated Implant 6.0mm 3.0mm 260-350-306
5.0mm HA Coated Implant 8.0mm 3.0mm 260-350-308
5.0mm HA Coated Implant 11.0mm 3.0mm 260-350-311
5.0mm Integra-Ti™ Implant 8.0mm 3.0mm 260-150-308
5.0mm Integra-Ti™ Implant 11.0mm 3.0mm 260-150-311
Diameter and Coating Length Well Part Number
6 x 5.7 6 x 8
6.0mm HA Coated Implant 5.7mm 3.0mm 260-360-357
6.0mm HA Coated Implant 8.0mm 3.0mm 260-360-108
™™
Healing Plug Inserter3.0mm
Description Length Part Number
3.0mm Healing Plug Inserter* 16.0mm 260-100-001
3 x 16 Example
*Packaged with each 4.5, 5.0 and 6.0mm diameter implant.
6800-88-BICON
Implant Level Impression Components2.0mm
Description Style Part Number
2.0mm Impression Post (2) Plastic 260-100-193
2.0mm Impression PostPlastic
2.0mm Implant AnalogPlastic
2.0mm Implant AnalogTitanium
2.0mm Implant Analog (2) Plastic 260-100-190
2.0mm Implant Level Impression Kit* Plastic 260-100-189
2.0mm Implant Analog (2) Titanium 260-100-188
* Contains one impression post and one plastic implant analog.
Implant Level Impression Components3.0mm
Description Style Part Number
3.0mm Impression Post (2) Plastic 260-100-393
3.0mm Impression PostPlastic
3.0mm Implant Analog Plastic
3.0mm Implant AnalogTitanium
3.0mm Implant Analog (2) Plastic 260-100-390
3.0mm Implant Level Impression Kit* Plastic 260-100-389
3.0mm Implant Analog (2) Titanium 260-100-388
* Contains one impression post and one plastic implant analog.
Implant Level Impression
1 Tap the impression post into the implant well
2 Inject impression material around the impression post
3 Make Impression 4 Remove the impression post from the implant and attach the
implant analog
5 Place assembled unit into the impression
➤ Impression Components and Technique Considerations
■ Making an Implant Level Impression: The use of an Impression Reamer (see page 62) on a guide pin may be necessary to widen the site for the insertion of the impression post. The impression post must be tapped into the well of the implant to ensure that it is fully seated. Once the impression post is seated, an impression is made. After recording the impression, remove the impression post from the implant (the impression post should not come out with the impression). Assemble the impression post and the implant analog, and then seat the unit into the impression. Before pouring impression in dental stone, pour a soft tissue model.
■ The use of a metal or plastic implant analog is at the discretion of the technician. If it is anticipated that the implant analog will be used in conjunction with any soldering techniques, a metal implant analog must be used.
■ When using a metal implant analog, the titanium abutment must be tapped into the analog to ensure proper axial positioning.
Imp
lants
Implant Level Impression Components
9 www.bicon.com
Temporary Abutments2.0mm Post
Description Diameter Height Post Style Part Number
3.5 x 4.5mm Temporary Abutment 2.0mm Post 3.5mm 4.5mm 2.0mm Titanium 260-135-245
3.5 x 4.5 3.5 x 6.53.5 x 6.5mm Temporary Abutment 2.0mm Post 3.5mm 6.5mm 2.0mm Titanium 260-135-265
4.0 x 4.5mm Temporary Abutment 2.0mm Post 4.0mm 4.5mm 2.0mm Titanium 260-140-245
4.0 x 4.5 4.0 x 6.54.0 x 6.5mm Temporary Abutment 2.0mm Post 4.0mm 6.5mm 2.0mm Titanium 260-140-265
5.0 x 4.5mm Temporary Abutment 2.0mm Post 5.0mm 4.5mm 2.0mm Titanium 260-150-245
5.0 x 4.5 5.0 x 6.55.0 x 6.5mm Temporary Abutment 2.0mm Post 5.0mm 6.5mm 2.0mm Titanium 260-150-265
➤ Temporary AbutmentDiameter
Hei
ght
The actual diameter of the temporary abutment accounts for the diameter of the final abutment with spacing for an emergence cuff. The height is measured from the top of the implant to the top of the temporary abutment.
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
➤ Temporary Abutment Considerations
■ Temporary abutments are made of surgical grade titanium alloy Ti6Al4V.
■ Temporary abutments are designed to form a gingival sulcus and to preserve papillae during the one stage surgical procedure or after a two stage uncovering procedure.
■ Choose an abutment height that is sufficient to laterally support the interproximal papillae without its being too tall such that it would receive undesirable forces from the tongue.
■ The actual diameter of the temporary abutment accounts for the final abutment’s diameter with spacing for an emergence cuff. The height is measured from the top of the implant to the top of the temporary abutment.
■ Actual diameters of the temporary abutments are 4.0, 5.5, 6.5, and 7.2mm for the 3.5, 4.0, 5.0, and 6.5mm abutments respectively.
■ The abutment height is measured from the top of the implant to the top of the abutment when the abutment is fully seated.
■ Temporary abutments may be modified with a #1557 carbide bur or other carbide burs.
10800-88-BICON
Temporary Abutments3.0mm Post
Description Diameter Height Post Style Part Number
4.0 x 4.5mm Temporary Abutment 3.0mm Post 4.0mm 4.5mm 3.0mm Titanium 260-340-345
4.0 x 4.5 4.0 x 6.54.0 x 6.5mm Temporary Abutment 3.0mm Post 4.0mm 6.5mm 3.0mm Titanium 260-340-365
5.0 x 4.5mm Temporary Abutment 3.0mm Post 5.0mm 4.5mm 3.0mm Titanium 260-350-345
5.0 x 4.5 5.0 x 6.55.0 x 6.5mm Temporary Abutment 3.0mm Post 5.0mm 6.5mm 3.0mm Titanium 260-350-365
6.5 x 4.5mm Temporary Abutment 3.0mm Post 6.5mm 4.5mm 3.0mm Titanium 260-365-345
6.5 x 4.5 6.5 x 6.56.5 x 6.5mm Temporary Abutment 3.0mm Post 6.5mm 6.5mm 3.0mm Titanium 260-365-365
Sinus Lift Temporary Abutment3.0mm Post
Description Diameter Height Post Style Part Number
6.5 x 2.5mm Sinus Lift Temporary Abutment 3.0mm Post 6.5mm 2.5mm 3.0mm Titanium 260-365-503
6.5 x 2.5
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
11 www.bicon.com
Healing Abutments2.0mm Post
Description Diameter Height Post Angle Style Part Number
4.0 x 5.0mm 0˚ Healing Abutment 2.0mm Post 4.0mm 5.0mm 2.0mm 0˚ Plastic 260-240-500
4.0 x 5.0 4.0 x 8.0
4.0 x 8.0mm 0˚ Healing Abutment 2.0mm Post 4.0mm 8.0mm 2.0mm 0˚ Plastic 260-240-800
5.0 x 5.0mm 0˚ Healing Abutment 2.0mm Post 5.0mm 5.0mm 2.0mm 0˚ Plastic 260-250-500
5.0 x 5.0 5.0 x 8.0 5.0 x 8.0 15˚
5.0 x 8.0mm 0˚ Healing Abutment 2.0mm Post 5.0mm 8.0mm 2.0mm 0˚ Plastic 260-250-800
5.0 x 8.0mm 15˚ Healing Abutment 2.0mm Post 5.0mm 8.0mm 2.0mm 15˚ Plastic 260-250-815
➤ Healing AbutmentDiameter
Hei
ght
The diameter of the healing abutment is measured at the widest part of the healing abutment. The height is measured from the top of the post to the highest portion of the incisal aspect of the healing abutment.
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
➤ Healing Abutment Considerations
■ Healing abutments are polycarbonate and are indicated for cases in which an abutment is needed to preserve the papillae for 180 days or less.
■ Healing abutments may be modified to meet a desired clinical shape.
■ The indicated diameter of a healing abutment is measured at its widest diameter. Its height is measured from the occlusal end of the post to the highest occlusal portion of the healing abutment.
■ If a healing abutment post were to fracture, removal is facilitated with the use of a round carbide bur.
12800-88-BICON
Healing Abutments3.0mm Post
Description Diameter Height Post Angle Style Part Number
5.0 x 5.0mm 0˚ Healing Abutment 3.0mm Post 5.0mm 5.0mm 3.0mm 0˚ Plastic 260-350-500
5.0 x 5.0 5.0 x 8.0
5.0 x 8.0mm 0˚ Healing Abutment 3.0mm Post 5.0mm 8.0mm 3.0mm 0˚ Plastic 260-350-800
6.5 x 5.0mm 0˚ Healing Abutment 3.0mm Post 6.5mm 5.0mm 3.0mm 0˚ Plastic 260-365-500
6.5 x 5.0 6.5 x 8.0 6.5 x 8.0 15˚
6.5 x 8.0mm 0˚ Healing Abutment 3.0mm Post 6.5mm 8.0mm 3.0mm 0˚ Plastic 260-365-800
6.5 x 8.0mm 15˚ Healing Abutment 3.0mm Post 6.5mm 8.0mm 3.0mm 15˚ Plastic 260-365-815
7.5 x 5.0mm 0˚ Healing Abutment 3.0mm Post 7.5mm 5.0mm 3.0mm 0˚ Plastic 260-375-500
7.5 x 5.0 7.5 x 8.0
7.5 x 8.0mm 0˚ Healing Abutment 3.0mm Post 7.5mm 8.0mm 3.0mm 0˚ Plastic 260-375-800
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
13 www.bicon.com
Stealth Transitional Abutments*
2.0mm PostDescription Diameter Height Angle Post Style Part Number 0˚ 10˚
3.5 x 1.5mm 0˚ Stealth Transitional Abutment 2.0mm Post 3.5mm 1.5mm 0˚ 2.0mm Plastic 260-235-150
4.0 x 1.5mm 0˚ Stealth Transitional Abutment 2.0mm Post 4.0mm 1.5mm 0˚ 2.0mm Plastic 260-240-150
4.0 x 3.5mm 0˚ Stealth Transitional Abutment 2.0mm Post 4.0mm 3.5mm 0˚ 2.0mm Plastic 260-240-350
5.0 x 2.0mm 0˚ Stealth Transitional Abutment 2.0mm Post 5.0mm 2.0mm 0˚ 2.0mm Plastic 260-250-201
5.0 x 2.0mm 10˚ Stealth Transitional Abutment 2.0mm Post 5.0mm 2.0mm 10˚ 2.0mm Plastic 260-250-202
5.0 x 4.0mm 0˚ Stealth Transitional Abutment 2.0mm Post 5.0mm 4.0mm 0˚ 2.0mm Plastic 260-250-221
5.0 x 4.0mm 10˚ Stealth Transitional Abutment 2.0mm Post 5.0mm 4.0mm 10˚ 2.0mm Plastic 260-250-222
5.0 x 6.0mm 0˚ Stealth Transitional Abutment 2.0mm Post 5.0mm 6.0mm 0˚ 2.0mm Plastic 260-250-241
*Each 4.0 and 5.0mm abutment is packaged with two corresponding acrylic sleeves. See page 25.
➤ Stealth Transitional AbutmentDiameter
Hei
ght
Example
The diameter of the abutment is measured at the widest part of the abutment. The height of the stealth transitional abutment is measured from the top of the implant to the shoulder of the abutment. The heights are 1.5, 2.0, 3.5, 4.0, or 6.0mm, depending upon the abutment diameter.
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
14800-88-BICON
➤ Stealth Transitional Abutment Considerations
■ Designed to be used in conjunction with a corresponding stealth acrylic sleeve.
■ Designed for cases in which a transitional prosthesis is required for 180 days or less.
Stealth Transitional Abutments*
3.0mm PostDescription Diameter Height Angle Post Style Part Number 0˚ 10˚
4.0 x 1.5mm 0˚ Stealth Transitional Abutment 3.0mm Post 4.0mm 1.5mm 0˚ 3.0mm Plastic 260-340-150
4.0 x 3.5mm 0˚ Stealth Transitional Abutment 3.0mm Post 4.0mm 3.5mm 0˚ 3.0mm Plastic 260-340-350
5.0 x 2.0mm 0˚ Stealth Transitional Abutment 3.0mm Post 5.0mm 2.0mm 0˚ 3.0mm Plastic 260-350-301
5.0 x 2.0mm 10˚ Stealth Transitional Abutment 3.0mm Post 5.0mm 2.0mm 10˚ 3.0mm Plastic 260-350-303
5.0 x 4.0mm 0˚ Stealth Transitional Abutment 3.0mm Post 5.0mm 4.0mm 0˚ 3.0mm Plastic 260-350-321
5.0 x 4.0mm 10˚ Stealth Transitional Abutment 3.0mm Post 5.0mm 4.0mm 10˚ 3.0mm Plastic 260-350-322
5.0 x 6.0mm 0˚ Stealth Transitional Abutment 3.0mm Post 5.0mm 6.0mm 0˚ 3.0mm Plastic 260-350-341
*Each 4.0 and 5.0mm abutment is packaged with two corresponding acrylic sleeves. See page 25.
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
15 www.bicon.com
Non-Shouldered Abutments 2.0mm Post
Description Diameter Height Angle Post Part Number 0˚ 15˚ 25˚
3.5 x 6.5mm 0˚ Abutment 2.0mm Post 3.5mm 6.5mm 0˚ 2.0mm 260-135-001
3.5 x 6.5mm 15˚ Abutment 2.0mm Post 3.5mm 6.5mm 15˚ 2.0mm 260-135-015
3.5 x 6.5mm 25˚ Abutment 2.0mm Post 3.5mm 6.5mm 25˚ 2.0mm 260-135-025
4.0 x 6.5mm 0˚ Abutment 2.0mm Post 4.0mm 6.5mm 0˚ 2.0mm 260-140-002
4.0 x 6.5mm 15˚ Abutment 2.0mm Post 4.0mm 6.5mm 15˚ 2.0mm 260-140-015
4.0 x 6.5mm 25˚ Abutment 2.0mm Post 4.0mm 6.5mm 25˚ 2.0mm 260-140-025
4.0 x 10mm 0˚ Abutment 2.0mm Post 4.0mm 10mm 0˚ 2.0mm 260-140-101
4.0 x 10mm 15˚ Abutment 2.0mm Post 4.0mm 10mm 15˚ 2.0mm 260-140-115
➤ 2.0mm Non-Shouldered AbutmentDiameter
Hei
ght
The diameter of the abutment is measured at the widest part of the abutment. The height of the non-shouldered abutment is measured from the widest portion to the incisal aspect of the abutment.
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
➤ Non-Shouldered Abutment Considerations
■ The abutment can be rotated 360° to reach a desired position or to achieve parallelism prior to being seated.
■ The 3.5mm abutments are recommended for mandibular incisors only; 4.0mm abutments are primarily used for maxillary laterals and bicuspids; 5.0mm abutments are more universal in their use; 6.5mm and 7.5mm abutments are ideally suited for molars.
■ The diameter of each Emergence Cuff (see page 22) corresponds to the diameter of a non-shouldered abutment.
■ Do not make an impression with the emergence cuff.
■ Choose the widest abutment and sleeve (see page 21) that will support the interproximal papillae without encroaching upon them.
16800-88-BICON
Non-Souldered Abutments 2.0mm PostDescription Diameter Height Angle Post Part Number 0˚ 15˚ 25˚
5.0 x 5.0mm 0˚ Abutment 2.0mm Post 5.0mm 5.0mm 0˚ 2.0mm 260-150-050
5.0 x 5.0mm 15˚ Abutment 2.0mm Post 5.0mm 5.0mm 15˚ 2.0mm 260-150-055
5.0 x 6.5mm 0˚ Abutment 2.0mm Post 5.0mm 6.5mm 0˚ 2.0mm 260-150-001
5.0 x 6.5mm 15˚ Abutment 2.0mm Post 5.0mm 6.5mm 15˚ 2.0mm 260-150-015
5.0 x 6.5mm 25˚ Abutment 2.0mm Post 5.0mm 6.5mm 25˚ 2.0mm 260-150-025
5.0 x 10mm 0˚ Abutment 2.0mm Post 5.0mm 10mm 0˚ 2.0mm 260-150-101
5.0 x 10mm 15˚ Abutment 2.0mm Post 5.0mm 10mm 15˚ 2.0mm 260-150-115
5.0 x 12mm 0˚ Abutment 2.0mm Post 5.0mm 12mm 0˚ 2.0mm 260-150-201
5.0 x 12mm 15˚ Abutment 2.0mm Post 5.0mm 12mm 15˚ 2.0mm 260-150-215
6.5 x 5.0mm 0˚ Abutment 2.0mm Post 6.5mm 5.0mm 0˚ 2.0mm 260-165-050
6.5 x 5.0mm 15˚ Abutment 2.0mm Post 6.5mm 5.0mm 15˚ 2.0mm 260-165-055
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
17 www.bicon.com
Non-Shouldered Abutments 3.0mm Post
Description Diameter Height Angle Post Part Number 0˚ 15˚ 25˚
4.0 x 6.5mm 0˚ Abutment 3.0mm Post 4.0mm 6.5mm 0˚ 3.0mm 260-340-001
4.0 x 6.5mm 15˚ Abutment 3.0mm Post 4.0mm 6.5mm 15˚ 3.0mm 260-340-015
4.0 x 10mm 0˚ Abutment 3.0mm Post 4.0mm 10mm 0˚ 3.0mm 260-340-101
4.0 x 10mm 15˚ Abutment 3.0mm Post 4.0mm 10mm 15˚ 3.0mm 260-340-115
5.0 x 5.0mm 0˚ Abutment 3.0mm Post 5.0mm 5.0mm 0˚ 3.0mm 260-350-050
5.0 x 5.0mm 15˚ Abutment 3.0mm Post 5.0mm 5.0mm 15˚ 3.0mm 260-350-055
5.0 x 6.5mm 0˚ Abutment 3.0mm Post 5.0mm 6.5mm 0˚ 3.0mm 260-350-001
5.0 x 6.5mm 15˚ Abutment 3.0mm Post 5.0mm 6.5mm 15˚ 3.0mm 260-350-015
5.0 x 6.5mm 25˚ Abutment 3.0mm Post 5.0mm 6.5mm 25˚ 3.0mm 260-350-025
➤ 3.0mm Non-Shouldered AbutmentDiameter
Hei
ght
The diameter of the abutment is measured at the widest part of the abutment. The height of the non-shouldered abutment is measured from the widest portion to the incisal aspect of the abutment.
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
➤ Non-Shouldered Abutment Considerations
■ The abutment can be rotated 360° to reach a desired position or to achieve parallelism prior to being seated.
■ The 3.5mm abutments are recommended for mandibular incisors only; 4.0mm abutments are primarily used for maxillary laterals and bicuspids; 5.0mm abutments are more universal in their use; 6.5mm and 7.5mm abutments are ideally suited for molars.
■ The diameter of each Emergence Cuff (see page 22) corresponds to the diameter of a non-shouldered abutment.
■ Do not make an impression with the emergence cuff.
■ Choose the widest abutment and sleeve (see page 21) that will support the interproximal papillae without encroaching upon them.
18800-88-BICON
Non-Souldered Abutments 3.0mm PostDescription Diameter Height Angle Post Part Number 0˚ 15˚ 25˚
5.0 x 10mm 0˚ Abutment 3.0mm Post 5.0mm 10mm 0˚ 3.0mm 260-350-101
5.0 x 10mm 15˚ Abutment 3.0mm Post 5.0mm 10mm 15˚ 3.0mm 260-350-115
5.0 x 12mm 0˚ Abutment 3.0mm Post 5.0mm 12mm 0˚ 3.0mm 260-350-201
5.0 x 12mm 15˚ Abutment 3.0mm Post 5.0mm 12mm 15˚ 3.0mm 260-350-215
6.5 x 5.0mm 0˚ Abutment 3.0mm Post 6.5mm 5.0mm 0˚ 3.0mm 260-365-050
6.5 x 5.0mm 15˚ Abutment 3.0mm Post 6.5mm 5.0mm 15˚ 3.0mm 260-365-055
6.5 x 6.5mm 0˚ Abutment 3.0mm Post 6.5mm 6.5mm 0˚ 3.0mm 260-365-001
6.5 x 6.5mm 15˚ Abutment 3.0mm Post 6.5mm 6.5mm 15˚ 3.0mm 260-365-015
7.5 x 8.0mm 0˚ Abutment 3.0mm Post 7.5mm 8.0mm 0˚ 3.0mm 260-375-801
7.5 x 8.0mm 15˚ Abutment 3.0mm Post 7.5mm 8.0mm 15˚ 3.0mm 260-375-815
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
19 www.bicon.com
Restorative/Laboratory Kits for Non-Shouldered Abutments*
Description Diameter Height Part Number
3.5 x 6.5mm Non-Shouldered Restorative/Laboratory Kit 3.5mm 6.5mm 260-135-465
3.5 x 6.5mm
4.0 x 6.5mm Non-Shouldered Restorative/Laboratory Kit 4.0mm 6.5mm 260-140-465
4.0 x 6.5mm
4.0 x 10mm Non-Shouldered Restorative/Laboratory Kit 4.0mm 10mm 260-140-410
4.0 x 10mm
5.0 x 5.0mm Non-Shouldered Restorative/Laboratory Kit 5.0mm 5.0mm 260-150-450
5.0 x 5.0mm
5.0 x 6.5mm Non-Shouldered Restorative/Laboratory Kit 5.0mm 6.5mm 260-150-465
5.0 x 6.5mm
5.0 x 10mm Non-Shouldered Restorative/Laboratory Kit 5.0mm 10mm 260-150-410
5.0 x 10mm
5.0 x 12mm Non-Shouldered Restorative/Laboratory Kit 5.0mm 12mm 260-150-412
5.0 x 12mm
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
➤ Non-Shouldered Prosthetic Components Considerations
■ The impression sleeves and transfer dies are color coded according to the corresponding diameter of the non-shouldered abutment: 3.5mm = blue, 4.0mm = red, 5.0mm = yellow, 6.5mm = green, and 7.5mm = purple.
■ The sleeves may be used for impression making, temporization, corrective copings, and/or as waxing sleeves.
■ The temporization sleeves may be used in lieu of the emergence cuffs.
■ Always confirm the complete seating of snap-on acrylic sleeves with either the non-shouldered abutment or colored abutment transfer die.
■ It is paramount that the proper abutment transfer die height be chosen for pouring of a stone model, since all transfer dies of the same color will snap into an impression sleeve of that color.
■ The diameter and height of the transfer dies are sized to match the diameter and height of a corresponding non-shouldered abutment.
■ Although possible, avoid making an impression using the tooth colored temporization sleeves, since they are not color coded with their corresponding transfer dies and they are more retentive.
■ Sleeves may be easily modified to suit a particular clinical need.
■ Current versions of the non-shouldered abutment has a groove approximately 3.0mm above its height of contour providing retention for the new snap-on impression and temporization sleeves.
■ The snap-on impression and temporization sleeves will not necessarily fit precisely on a non-shouldered abutment without the groove.
20800-88-BICON
Description Diameter Height Part Number
6.5 x 5.0mm Non-Shouldered Restorative/Laboratory Kit 6.5mm 5.0mm 260-165-450
6.5 x 5.0mm
6.5 x 6.5mm Non-Shouldered Restorative/Laboratory Kit 6.5mm 6.5mm 260-165-465
6.5 x 6.5mm
7.5 x 8.0mm Non-Shouldered Restorative/Laboratory Kit 7.5mm 8.0mm 260-175-480
7.5 x 8.0mm
*Each kit contains (1) Transfer Die, (1) Impression/Waxing Sleeve, and (1) Temporization Sleeve.
Impression
Snap on impression sleeve Make abutment transfer impression Insert abutment transfer die
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
21 www.bicon.com
Temporization Sleeves for Non-Shouldered AbutmentsDescription Diameter Part Number
3.5mm Non-Shouldered Temporization Sleeve (2) 3.5mm 260-135-165
3.5mm
Example
4.0mm Non-Shouldered Temporization Sleeve (2) 4.0mm 260-140-165
4.0mm
5.0mm Non-Shouldered Temporization Sleeve (2) 5.0mm 260-150-165
5.0mm
6.5mm Non-Shouldered Temporization Sleeve (2) 6.5mm 260-165-165
6.5mm
7.5mm Non-Shouldered Temporization Sleeve (2) 7.5mm 260-175-165
7.5mm
Temporization
Snap on temporization sleeve Inject transitional material Snap on transitional prosthesis
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
➤ Emergence Cuff Considerations
■ The diameter of each emergence cuff corresponds to the diameter of a non-shouldered abutment.
■ Do not make an impression with an emergence cuff.
■ If you prefer to make an impression of a non-shouldered abutment with a plastic sleeve, please see pages 19-20.
■ The temporization sleeves may be used in lieu of the emergence cuffs.
22800-88-BICON
Emergence Cuffs for Non-Shouldered AbutmentsDescription Diameter Part Number
4.0mm Emergence Cuff (4) 4.0mm 260-140-010
4.0mm
Example
5.0mm Emergence Cuff (4) 5.0mm 260-150-010
5.0mm
Example
6.5mm Emergence Cuff (4) 6.5mm 260-165-065
6.5mm
Example
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
23 www.bicon.com
Stealth Shouldered Abutments* 2.0mm Post
Description Diameter Height Angle Post Part Number 0˚ 10˚
3.5 x 1.5mm 0° Stealth Abutment 2.0mm Post 3.5mm 1.5mm 0˚ 2.0mm 260-200-150
4.0 x 1.5mm 0° Stealth Abutment 2.0mm Post 4.0mm 1.5mm 0˚ 2.0mm 260-200-200
4.0 x 3.5mm 0° Stealth Abutment 2.0mm Post 4.0mm 3.5mm 0˚ 2.0mm 260-200-400
5.0 x 2.0mm 0° Stealth Abutment 2.0mm Post 5.0mm 2.0mm 0˚ 2.0mm 260-200-201
5.0 x 2.0mm 10° Stealth Abutment 2.0mm Post 5.0mm 2.0mm 10˚ 2.0mm 260-200-202
5.0 x 4.0mm 0° Stealth Abutment 2.0mm Post 5.0mm 4.0mm 0˚ 2.0mm 260-200-221
5.0 x 4.0mm 10° Stealth Abutment 2.0mm Post 5.0mm 4.0mm 10˚ 2.0mm 260-200-222
5.0 x 6.0mm 0° Stealth Abutment 2.0mm Post 5.0mm 6.0mm 0˚ 2.0mm 260-200-241
5.0 x 6.0mm 10° Stealth Abutment 2.0mm Post 5.0mm 6.0mm 10˚ 2.0mm 260-200-242
5.0 x 8.0mm 0° Stealth Abutment 2.0mm Post 5.0mm 8.0mm 0˚ 2.0mm 260-200-261
*Each 4.0 and 5.0mm abutment is packaged with two corresponding stealth acrylic sleeves.
➤ Stealth Shouldered AbutmentDiameter
Hei
ght
Example
The diameter of the abutment is measured at the widest part of the abutment. The height of the stealth shouldered abutment is measured from the top of the implant to the shoulder of the abutment. The heights are 1.5, 2.0, 3.5, 4.0, 6.0 or 8.0mm, depending upon the abutment diameter.
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
24800-88-BICON
Stealth Shouldered Abutments*
3.0mm PostDescription Diameter Height Angle Post Part Number 0˚ 10˚
4.0 x 1.5mm 0° Stealth Abutment 3.0mm Post 4.0mm 1.5mm 0˚ 3.0mm 260-300-300
4.0 x 3.5mm 0° Stealth Abutment 3.0mm Post 4.0mm 3.5mm 0˚ 3.0mm 260-300-400
5.0 x 2.0mm 0° Stealth Abutment 3.0mm Post 5.0mm 2.0mm 0˚ 3.0mm 260-300-301
5.0 x 2.0mm 10° Stealth Abutment 3.0mm Post 5.0mm 2.0mm 10˚ 3.0mm 260-300-302
5.0 x 3.0mm 0° Stealth Abutment 3.0mm Post 5.0mm 3.0mm 0˚ 3.0mm 260-300-311
5.0 x 3.0mm 10° Stealth Abutment 3.0mm Post 5.0mm 3.0mm 10˚ 3.0mm 260-300-312
5.0 x 4.0mm 0° Stealth Abutment 3.0mm Post 5.0mm 4.0mm 0˚ 3.0mm 260-300-321
5.0 x 4.0mm 10° Stealth Abutment 3.0mm Post 5.0mm 4.0mm 10˚ 3.0mm 260-300-322
5.0 x 6.0mm 0° Stealth Abutment 3.0mm Post 5.0mm 6.0mm 0˚ 3.0mm 260-300-341
5.0 x 6.0mm 10° Stealth Abutment 3.0mm Post 5.0mm 6.0mm 10˚ 3.0mm 260-300-342
5.0 x 8.0mm 0° Stealth Abutment 3.0mm Post 5.0mm 8.0mm 0˚ 3.0mm 260-300-361
*Each 4.0 and 5.0mm abutment is packaged with two corresponding stealth acrylic sleeves.
➤ Stealth Shouldered AbutmentDiameter
Hei
ght
Example
The diameter of the abutment is measured at the widest part of the abutment. The height of the stealth shouldered abutment is measured from the top of the implant to the shoulder of the abutment. The heights are 1.5, 2.0, 3.0, 3.5, 4.0, 6.0 or 8.0mm, depending upon the abutment diameter.
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
25 www.bicon.com
Stealth Acrylic SleevesDescription Diameter Height Part Number
Thin Wall Stealth Sleeve for 4.0mm Stealth Abutment (4) 4.0mm 5.5mm 260-250-204
Example
Thin Wall Stealth Sleeve for 5.0mm Stealth Abutment (4) 5.0mm 6.5mm 260-250-205
Stealth Acrylic Sleeve for 3.5mm Stealth Abutment (4) 3.5mm 10.0mm 260-250-103
Short Stealth Acrylic Sleeve for 4.0mm Stealth Abutment (4) 4.0mm 6.0mm 260-250-004
Tall Stealth Acrylic Sleeve for 4.0mm Stealth Abutment (4) 4.0mm 10.0mm 260-250-104
Short Stealth Acrylic Sleeve for 5.0mm Stealth Abutment (4) 5.0mm 6.0mm 260-250-005
Tall Stealth Acrylic Sleeve for 5.0mm Stealth Abutment (4) 5.0mm 10.0mm 260-250-105 5.0mmShort
5.0mmTall
4.0mmShort
4.0mmTall
3.5mm
4.0mm 5.0mm
Stealth Aluminum Oxide Sleeves*
Description Diameter Height Part Number
3.5 x 7.0mm Aluminum Oxide Sleeve 3.5mm 7.0mm 260-250-371
3.5 x 7.0mm
Example
4.0 x 7.0mm Aluminum Oxide Sleeve 4.0mm 7.0mm 260-250-471
4.0 x 7.0mm
5.0 x 7.0mm Aluminum Oxide Sleeve 5.0mm 7.0mm 260-250-571
5.0 x 7.0mm
5.0 x 10mm Aluminum Oxide Sleeve 5.0mm 10.0mm 260-250-101
5.0 x 10.0mm
*Each sleeve is packaged with one corresponding abutment transfer die. Choose the diameter of the sleeve based upon the diameter of the abutment. Expansion factor = AL
2 O
3 25°C to 500°C 7 x 10-6/C 25°C to 800°C 8 x 10-6/C
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
➤ Stealth Shouldered Abutment Considerations
■ Although the stealth shouldered abutment is designed for all-ceramic restorations, it may be used in conjunction with metal castings.
■ Use the shoulder depth gauge to determine an appropriate abutment shoulder height, relative to the gingival tissues. The first mark on the shoulder depth gauge corresponds to a 2.0mm shoulder height of a 5.0mm stealth shouldered abutment. The second mark corresponds to a 4.0mm shoulder height, etc. If the soft tissue falls in between two marks, select the shorter of the two abutment heights for an aesthetic result.
■ The diameter of an abutment is dictated by the interproximal papillae. The abutment should support the papillae without encroaching upon them.
■ The stealth acrylic sleeve may be used for making an impression, for creating a temporary crown, for recording an occlusal registration, and also as a waxing sleeve or a corrective coping.
■ When using the one-piece sleeve as an impression transfer, it is essential to confirm that an internal flat of the sleeve has been aligned with an external flat on the abutment.
■ Prior to the intraorally cementing of a crown on a shouldered abutment, a radiograph must be taken to ensure that the crown is fully seated.
■ The aluminium oxide sleeve must be ground wet either using a sintered diamond disk or a diamond wheel on a model trimmer.
■ Only aluminous porcelains may be used in conjunction with the aluminum oxide sleeve.* (n.b., metal-ceramic type porcelains and pressable ceramics are NOT compatible with Bicon’s aluminum oxide sleeve.)
■ A 3.5mm stealth shouldered abutment crown MUST be extra-orally cemented with DiamondLink™ by DRM or another comparable cement.
26800-88-BICON
Stealth Abutment Transfer DiesDescription Diameter Part Number
3.5mm Stealth Abutment Transfer Die (4) 3.5mm 260-250-193
3.5mmPlastic
4.0mm Brass
5.0mm Brass
4.0mm Brass Stealth Abutment Transfer Die (2) 4.0mm 260-250-184
5.0mm Brass Stealth Abutment Transfer Die (2) 5.0mm 260-250-185
Abutment Shoulder Depth GaugeDescription Part Number
Abutment Shoulder Depth Gauge 260-101-380
8mm6mm4mm2mm
Seated gauge Seated abutment
*Cerabien™ by Noritake, Vitadur Alpha by Vident and Allceram™ by Degussa-Ney are recommended for use with the Aluminum Oxide Sleeve. You may check with your porcelain manufacturer about the use of a particular porcelain with the Aluminum Oxide Sleeve.
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
27 www.bicon.com
Brevis™ Abutments*
2.0mm PostDescription Height Angle Post Part Number 0˚ 15˚
2.0mm 0˚ Brevis Abutment 2.0mm Post 2.0mm 0˚ 2.0mm 260-100-404
2.0mm 15˚ Brevis Abutment 2.0mm Post 2.0mm 15˚ 2.0mm 260-100-405
4.0mm 0˚ Brevis Abutment 2.0mm Post 4.0mm 0˚ 2.0mm 260-100-406
4.0mm 15˚ Brevis Abutment 2.0mm Post 4.0mm 15˚ 2.0mm 260-100-407
6.0mm 0˚ Brevis Abutment 2.0mm Post 6.0mm 0˚ 2.0mm 260-100-408
6.0mm 15˚ Brevis Abutment 2.0mm Post 6.0mm 15˚ 2.0mm 260-100-409
*Each abutment is packaged with one titanium Brevis housing.
Brevis™ Abutments* 3.0mm Post
Description Height Angle Post Part Number 0˚ 15˚
2.0mm 0˚ Brevis Abutment 3.0mm Post 2.0mm 0˚ 3.0mm 260-300-434
2.0mm 15˚ Brevis Abutment 3.0mm Post 2.0mm 15˚ 3.0mm 260-300-435
4.0mm 0˚ Brevis Abutment 3.0mm Post 4.0mm 0˚ 3.0mm 260-300-436
4.0mm 15˚ Brevis Abutment 3.0mm Post 4.0mm 15˚ 3.0mm 260-300-437
6.0mm 0˚ Brevis Abutment 3.0mm Post 6.0mm 0˚ 3.0mm 260-300-438
6.0mm 15˚ Brevis Abutment 3.0mm Post 6.0mm 15˚ 3.0mm 260-300-439
*Each abutment is packaged with one titanium Brevis housing.
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
28800-88-BICON
Brevis™ Abutment Restorative Components
Description Part Number
Brevis Abutment Chairside Kit 260-100-212
Rubber O-Ring Brevis Housing
Brevis Impression Kit without Housing 260-100-218
Impression Cap Aluminum Transfer Die
Brevis Abutment Impression Kit 260-100-217
Rubber O-Ring Brevis Housing Impression Cap Aluminum Transfer Die
Brevis Rubber O-Rings (10) 260-100-014 Rubber O-Ring
➤ Brevis™ Abutment
2.0mm
3.0mm
3.0mm
4.0mm
3.0mm
6.0mm
Measurement Guide Example
The height of the Brevis abutment is measured from the top of the implant to the shoulder of the Brevis abutment. The available heights are 2.0, 4.0 or 6.0mm. The total height from the top of the implant to the top of the abutment is approximately 5.0, 7.0 or 9.0mm, respectively.
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
29 www.bicon.com
Locator® Abutments*
2.0mm PostDescription Diameter Height Angle Post Part Number
1.0mm Bicon Locator Abutment 2.0mm Post 4.0mm 1.0mm 0˚ 2.0mm 260-200-501
1.0mm 2.0mm 3.0mm 4.0mm 5.0mm
2.0mm Bicon Locator Abutment 2.0mm Post 4.0mm 2.0mm 0˚ 2.0mm 260-200-502
3.0mm Bicon Locator Abutment 2.0mm Post 4.0mm 3.0mm 0˚ 2.0mm 260-200-503
4.0mm Bicon Locator Abutment 2.0mm Post 4.0mm 4.0mm 0˚ 2.0mm 260-200-504
5.0mm Bicon Locator Abutment 2.0mm Post 4.0mm 5.0mm 0˚ 2.0mm 260-200-505
*Each Locator abutment is packaged with a male processing kit.
Locator® Abutments*
3.0mm PostDescription Diameter Height Angle Post Part Number
1.0mm Bicon Locator Abutment 3.0mm Post 4.0mm 1.0mm 0˚ 3.0mm 260-300-501
1.0mm 2.0mm 3.0mm 4.0mm 5.0mm
2.0mm Bicon Locator Abutment 3.0mm Post 4.0mm 2.0mm 0˚ 3.0mm 260-300-502
3.0mm Bicon Locator Abutment 3.0mm Post 4.0mm 3.0mm 0˚ 3.0mm 260-300-503
4.0mm Bicon Locator Abutment 3.0mm Post 4.0mm 4.0mm 0˚ 3.0mm 260-300-504
5.0mm Bicon Locator Abutment 3.0mm Post 4.0mm 5.0mm 0˚ 3.0mm 260-300-505
*Each Locator abutment is packaged with a male processing kit.
➤ Locator® Abutment with housing
Example
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
➤ Locator® Abutment Considerations
■ The Locator® Attachment features a denture component with a skirt that easily locates the mating implant abutment.
■ The self-aligning ability of the attachment aids the patient in positioning their prosthesis in similar manner as a guide plane created by a milled bar.
■ The implant retained overdenture can be properly seated without damage to the attachment components. This is especially important for a patient lacking anatomical structures necessary to orient their denture due to a severely resorbed mandibular ridge.
30800-88-BICON
Locator® AccessoriesDescription Part Number
Locator Male Processing Kit (2) 260-100-519
2.3 kg (5.0 lb.) Retention Male
1.4 kg (3.0 lb.) Retention Male
0.7 kg (1.5 lb.) Retention Male
ProcessingMale
Metal Housing Cap
BlockoutSpacer
Locator Male Processing Kit (10) 260-100-520
Locator Extended Range Green Cap (4)
260-100-521 1.4 kg (3.0 lb.) Retention Cap
Locator Core Tool 260-101-839
Locator Core Tool
➤ Locator® Abutment
2.0mm 3.0mm 4.0mm 5.0mm
0.56mm
1.0mm
Measurement Guide
The height of the Locator® abutment is measured from the bottom of the gold portion of the abutment to the shoulder of the Locator® abutment. The available heights are 1.0, 2.0, 3.0, 4.0, or 5.0mm. The total height from the top of the implant to the top of the abutment is approximately 4.0mm.
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
31 www.bicon.com
➤ Fixed-Detachable Abutment
Hei
ght
Diameter The diameter of the abutment at its widest point is 4.0mm. The height of the fixed-detachable abutment is measured from the widest portion of the abutment to the top of the implant. The two available heights are 3.5mm or 5.0mm. This abutment has a 20˚ tapered head.
Fixed-Detachable Abutments2.0mm Post
Description Diameter Height Angle Post Part Number 0˚ 15˚
4.0 x 3.0mm 0˚ Abutment 2.0mm Post 4.0mm 3.0mm 0˚ 2.0mm 260-100-003
Occlusal view
4.0 x 3.0mm 15˚ Abutment 2.0mm Post 4.0mm 3.0mm 15˚ 2.0mm 260-100-015
4.0 x 5.0mm 0˚ Abutment 2.0mm Post 4.0mm 5.0mm 0˚ 2.0mm 260-100-004
4.0 x 5.0mm 15˚ Abutment 2.0mm Post 4.0mm 5.0mm 15˚ 2.0mm 260-100-115
Fixed-Detachable Abutments3.0mm Post
Description Diameter Height Angle Post Part Number 0˚ 15˚
4.0 x 3.0mm 0˚ Abutment 3.0mm Post 4.0mm 3.0mm 0˚ 3.0mm 260-300-003
Occlusal view
4.0 x 3.0mm 15˚ Abutment 3.0mm Post 4.0mm 3.0mm 15˚ 3.0mm 260-300-015
4.0 x 5.0mm 0˚ Abutment 3.0mm Post 4.0mm 5.0mm 0˚ 3.0mm 260-300-004
4.0 x 5.0mm 15˚ Abutment 3.0mm Post 4.0mm 5.0mm 15˚ 3.0mm 260-300-115
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
32800-88-BICON
Bicon Bar Abutments2.0mm Post (package of 2)
Description Diameter Height Angle Post Part Number
3.0mm Bicon Bar Abutment 2.0mm Post 4.0mm 3.0mm 0˚ 2.0mm 260-100-303
3.0mm 4.0mm 5.0mm 6.0mm
4.0mm Bicon Bar Abutment 2.0mm Post 4.0mm 4.0mm 0˚ 2.0mm 260-100-304
5.0mm Bicon Bar Abutment 2.0mm Post 4.0mm 5.0mm 0˚ 2.0mm 260-100-305
6.0mm Bicon Bar Abutment 2.0mm Post 4.0mm 6.0mm 0˚ 2.0mm 260-100-306
Bicon Bar AbutmentRestorative Components
Description Length Part Number
Bicon Bar (2) 22.0mm 260-100-300
Brass Bicon Bar Abutment Transfer Die (2) 260-100-309
Fixed-Detachable AbutmentsRestorative Components
Description Height Part Number
Acrylic Transfer and Castable Copings (4) 260-100-005
Titanium Transfer Copings (4) 260-100-019
Brass Abutment Analogs (4) 260-100-006
5.0mm Hex Coping Screws (4) 5.0mm 260-100-016
5.0mm 10.0mm
10.0mm Hex Coping Screws (4) 10.0mm 260-100-017
Hex Retention Screws (4) 260-100-020
Hex Cover Screws (4) 260-100-021
Hex Screwdriver 260-101-024
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
33
Abutment Selection Board
Abutment Selection Board (Plastic Abutments Included)
Description Part Number
Abutment Selection Board 260-101-037
Non-Shouldered Abutment Overlay
Non-Shouldered Abutment Overlay
Description Part Number
Non-Shouldered Abutment Overlay 260-103-024
Stealth Abutment Overlay
Stealth Abutment Overlay
Description Part Number
Stealth Abutment Overlay 260-103-025
Implant Overlay
Implant Overlay
Description Part Number
Implant Overlay 260-103-004
Implant Ruler4.5 x 8mm 5 x 8mm 5 x 8mm 4 x 11mm 3.5 x 11mm 4.5 x 8mm 5 x 8mm 5 x 8mm 4 x 11mm 3.5 x 11mm 6 x 5.7mm 6 x 5.7mm
0 1 2 3 4 5 6 7 8
1/8 inch
1mm
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
(100% SCALE) (125% SCALE)
260-103-005 R0904
Implant Ruler
Description Part Number
Implant Ruler 260-103-005
www.bicon.com
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
34800-88-BICON
Abutment Measurement GuideThe design of the Bicon abutment system is such that the hemispherical base of the abutment does not sit flush with the neck of the implant. By design, there is a radiolucency below the abutment post when the abutment is fully seated. Please note that the post of any abutment may not be modified. Altering the abutment post in any manner or
using a cementing medium will affect the frictional fit of the locking taper resulting in a potentially non-retentive abutment. The following diagrams depict the final seating of each abutment style as well as the method for measuring each abutment.
Temporary Abutment Healing Abutment Fixed-Detachable Abutment
Diameter
Hei
ght
Diameter
Hei
ght
Hei
ght
Diameter
The diameter of the temporary abutment accounts for the diameter of the final abutment with spacing for an emergence cuff. The height is measured from the top of the implant to the top of the temporary abutment.
The diameter of the healing abutment is measured at the widest part of the healing abutment. The height is measured from the top of the post to the highest portion of the incisal aspect of the healing abutment.
The diameter of the abutment at its widest point is 4.0mm. The height of the fixed-detachable abutment is measured from the widest portion of the abutment to the top of the implant. The two available heights are 3.5mm or 5.0mm. This abutment has a 20˚ tapered head.
Non-Shouldered Abutment —2.0mm Post Non-Shouldered Abutment —3.0mm Post Brevis™ Abutment
Diameter
Hei
ght
Diameter
Hei
ght
Hei
ght
The diameter of the abutment is measured at the widest part of the abutment. The height of the non-shouldered abutment is measured from the widest portion to the incisal aspect of the abutment. The abutment post height is constant and is not variable.
The diameter of the abutment is measured at the widest part of the abutment. The height of the non-shouldered abutment is measured from the widest portion to the incisal aspect of the abutment. The abutment post height is constant and is not variable.
The height of the Brevis™ abutment is measured from the top of the implant to the shoulder of the Brevis™ abutment. The available heights are 2.0, 4.0 or 6.0mm. The total height from the top of the implant to the top of the abutment is approximately 5.0, 7.0 or 9.0mm, respectively.
Locator® Abutment Stealth Abutment Stealth Transitional Abutment
Diameter
Hei
ght
Diameter
Hei
ght
Diameter
Hei
ght
The height of the Locator® abutment is measured from the bottom of the gold portion of the abutment to the shoulder of the Locator® abutment. The available heights are 1.0, 2.0, 3.0, 4.0, or 5.0mm. The total height from the top of the implant to the top of the abutment is approximately 4.0mm.
The height of the stealth shouldered abutment is measured from the top of the implant to the shoulder of the abutment. The heights are 1.5, 2.0, 3.0, 3.5, 4.0, 6.0 or 8.0mm, depending upon the abutment diameter.
The height of the stealth transitional abutment is measured from the top of the implant to the shoulder of the abutment. The heights are 1.5, 2.0, 3.5, 4.0 or 6.0mm, depending upon the abutment diameter.
Ab
utmentsA
but
men
ts
Ab
utmentsA
but
men
ts
➤ SynthoGraft™ is a synthetic, biocompatible and resorbable granulate ceramic made of pure phase Beta-Tricalcium Phosphate Ca3(PO4)2 which is used as a matrix for oral and maxillofacial bone regeneration or augmentation.
➤ While several varieties of βTCP are now commercially available, they are not all equal in bone regenerating capability. These differences can profoundly affect the rate and quality of bone regeneration, as well as the rate of resorption of the graft material as it is being replaced by the patients’ own bone during the healing process.
➤ SynthoGraft’s™ unique structure provides increased stability and its micro-porosity and nano-porosity allow for rapid vascularization and subsequent resorption. SynthoGraft™ achieves the greatest efficacy when mixed with the patient’s own blood. Other additives, including autogenous bone particles, are not needed or recommended.
37 www.bicon.com
The Next Generation of Regeneration™
TM
RegenerativeRege
nera
tive
RegenerativeRege
nera
tive
38800-88-BICON
Histological Core: Lateral sinus lift at 12 weeks
OB: Osteoblasts
NB: New bone layer
OC: Newly entrapped osteocytes
BV: Blood vessels
TCP: Remnants of Tricalcium Phosphate surrounded by loose fibrous tissue
CT: Dense connective tissue
RegenerativeRege
nera
tive
RegenerativeRege
nera
tive
➤ SynthoGraft™ Considerations
Indications and Usage:
SynthoGraft™ is designed for the filling and/or reconstruction of either traumatic or degenerative multi-wall bone defects, augmentation of the sinus floor, augmentation of atrophied alveolar ridges, filling of periodontal or other alveolar bone defects, tooth sockets and osteotomies as well as the preservation of the alveolus for the preparation of an implant site.
➤ Warnings:■ SynthoGraft™ should be used only by or under the supervision
of trained personnel with experience in the surgical techniques associated with biomaterials.
■ SynthoGraft™ is packaged and sterilized for single use only.
■ Do not apply SynthoGraft™ dry.
■ Do not saturate SynthoGraft™ with any solution (e.g. physiological saline, NaCl or antibiotics) other than the patient’s blood.
■ Do not mix SynthoGraft™ with any other bone grafting material.
■ Do not resterilize SynthoGraft™.
■ Discard any unused SynthoGraft™ granulate.
■ Not intended for immediate loading.
■ Do not overfill defect site.
■ Secure SynthoGraft™ to prevent migration of materials.
■ Do not use if package has been opened, damaged or if the expiration dated has passed.
■ Do not compromise blood supply to the defect area.
39 www.bicon.com
SynthoGraft™
Description Particle Size Grams per Vial Vials Part Number
SynthoGraft™ Pure Phase Beta-Tricalcium Phosphate 50-500µm 0.5g 5 260-400-150
SynthoGraft™ Pure Phase Beta-Tricalcium Phosphate 50-500µm 1.0g 5 260-400-151
SynthoGraft™ Pure Phase Beta-Tricalcium Phosphate 50-500µm 2.0g 5 260-400-152
SynthoGraft™ Pure Phase Beta-Tricalcium Phosphate 500-1000µm 0.5g 5 260-400-500
SynthoGraft™ Pure Phase Beta-Tricalcium Phosphate 500-1000µm 1.0g 5 260-400-501
SynthoGraft™ Pure Phase Beta-Tricalcium Phosphate 500-1000µm 2.0g 5 260-400-502
Regeneration with SynthoGraft™
1 Remove glass vial 2 Dispense desired quantity 3 Wet with patient’s blood
4 Prepare site 5 Place SynthoGraft mixture 6 Primary closure
RegenerativeRege
nera
tive
RegenerativeRege
nera
tive
40800-88-BICON
Bicon Resorbable Collagen MembraneDescription Size Part Number
Bicon Resorbable Collagen Membrane 15mm x 20mm x 0.3mm 260-509-600
Bicon Resorbable Collagen Membrane 20mm x 30mm x 0.3mm 260-509-300
Bicon Resorbable Collagen Membrane 30mm x 40mm x 0.3mm 260-509-800
➤ Membrane Considerations
■ Intended for use in oral surgical procedures as a resorbable or non-resorbable material for placement in the areas of dental implants, bone defects and ridge augmentation procedures.
■ Ensure that the membrane is stable and securely placed under the tissue.
■ Primary closure is desirable to allow for full coverage of the membrane.
■ Dissecting and suturing the buccal aspect of the flap will allow for the coronal movement of the flap facilitating primary closure.
■ Each membrane can be easily trimmed during surgery for an appropriate clinical adaptation.
In Vivo Rate of Resorption of Membrane
Perc
ent o
f Mat
eria
l Rem
ain
ing
(%)
Implantation Time (weeks)
100
80
60
40
20
00 4 8 12 16 20 24 28 32 36
Bicon Resorbable Collagen Membrane
BioMend®
Resorbable Collagen PlugDescription Size Part Number
Resorbable Collagen Plug (10) 10.0mm x 20.0mm 260-509-400
Resorbable Collagen Plug (10)
➤ Collagen Plug Considerations
■ The Bicon Resorbable Collagen Plug is fully absorbed in 10 to 14 days.
■ Its cylindrical shape adapts easily to surgical sites.
Bicon CPRESORBABLE COLLAGEN PRODUCTS
™
RegenerativeRege
nera
tive
RegenerativeRege
nera
tive
www.bicon.com43
➤ Comprehensive Surgical Kit–Plus Considerations
■ A pilot drill should be replaced after approximately 12 surgeries.
■ Latch reamers may be used for up to 100 surgeries without replacement.
■ It is advisable to have a spare pilot bur, a spare set of latch reamers, and a spare set of handpieces.
■ The tray may be sterilized at temperatures up to 275° F (135° C).
Comprehensive Surgical Kit–Plus
Tray depicted with cover
Description Part Number
Comprehensive Surgical Kit–Plus* 260-101-095
*All items not depicted
KitsKit
s KitsKit
s
44800-88-BICON
Comprehensive Surgical Kit–PlusDescription Part Number
2.0mm Pilot Drill 260-101-001
2.0mm Implant Inserter/Retriever 260-101-003
3.0mm Implant Inserter/Retriever 260-101-004
2.0mm Long Implant Inserter/Retriever 260-101-005
3.0mm Long Implant Inserter/Retriever 260-101-006
Threaded Offset Handle 260-101-009
5.0mm Threaded Knob (4) 260-101-014
Threaded Straight Handle 260-101-016
2.0mm Implant/Angled Abutment Seating Tip 260-101-010
Standard Abutment Seating Tip 260-101-012
3.0mm Implant Seating Tip 260-101-018
Large Abutment Seating Tip 260-101-019
Removal Wrench 260-101-017
Hand Reamer Extension 260-101-020
Latch Reamer Extension 260-101-022
2.5mm Hand Reamer 260-101-025
3.0mm Hand Reamer 260-101-030
3.5mm Hand Reamer 260-101-035
4.0mm Hand Reamer 260-101-040
4.5mm Hand Reamer 260-101-045
5.0mm Hand Reamer 260-101-050
5.5mm Hand Reamer 260-101-055
6.0mm Hand Reamer 260-101-060
Healing Plug Removal Instrument (Manual) 260-101-114
Healing Plug Removal Instrument (Latch) 260-101-115
2.5mm Standard Latch Reamer 260-101-125
3.0mm Standard Latch Reamer 260-101-130
3.5mm Standard Latch Reamer 260-101-135
4.0mm Standard Latch Reamer 260-101-140
4.5mm Standard Latch Reamer 260-101-645
5.0mm Standard Latch Reamer 260-101-650
5.5mm Standard Latch Reamer 260-101-655
6.0mm Standard Latch Reamer 260-101-660
2.5mm Mid- Size Latch Reamer 260-101-325
3.0mm Mid- Size Latch Reamer 260-101-330
3.5mm Mid- Size Latch Reamer 260-101-335
4.0mm Mid- Size Latch Reamer 260-101-340
4.5mm Mid- Size Latch Reamer 260-101-345
5.0mm Mid- Size Latch Reamer 260-101-350
5.5mm Mid- Size Latch Reamer 260-101-355
6.0mm Mid- Size Latch Reamer 260-101-360
Comprehensive Plus KitDescription Part Number
Depth Gauge/Bone Depth Plugger 260-101-370
Abutment Shoulder Gauge 260-101-380
Silicone Dappen Dish 260-103-030
Instrument Storage Tray 260-104-099
Paralleling Pin 0˚ (3) 260-701-001
Paralleling Pin 15˚ (2) 260-701-053
Paralleling Pin 25˚ (2) 260-701-054
2.0mm Standard Guide Pin (3) 260-101-180
3.0mm Standard Guide Pin (3) 260-101-183
3.5mm Sulcus Reamer 260-101-435
4.0mm Sulcus Reamer 260-101-440
5.0mm Sulcus Reamer 260-101-450
6.5mm Sulcus Reamer 260-101-465
7.5mm Sulcus Reamer 260-101-475
Healing Plug Cutter 260-801-001
Abutment Carrying Forceps 260-801-002
Surgical Mallet 260-801-165
3.0mm Bone Expander 260-901-430
3.5mm Bone Expander 260-901-435
4.0mm Bone Expander 260-901-440
4.5mm Bone Expander 260-901-445
5.0mm Bone Expander 260-901-450
5.5mm Bone Expander 260-901-455
6.0mm Bone Expander 260-901-460
3.0mm Osteotome 260-901-530
3.5mm Osteotome 260-901-535
4.0mm Osteotome 260-901-540
5.0mm Osteotome 260-904-550
6.0mm Osteotome 260-901-560
3.0mm Threaded Expanding Chisel 260-901-613
5.0mm Threaded Expanding Chisel 260-901-615
6.0mm Threaded Expanding Chisel 260-901-616
KitsKit
s KitsKit
s
45 www.bicon.com
➤ Comprehensive Surgical Kit Considerations
■ A pilot drill should be replaced after approximately 12 surgeries.
■ Latch reamers may be used for up to 100 surgeries without replacement.
■ It is advisable to have a spare pilot bur, a spare set of latch reamers, and a spare set of handpieces.
■ The tray may be sterilized at temperatures up to 275° F (135° C).
Comprehensive Surgical Kit
Tray depicted with cover
Description Part Number
Comprehensive Surgical Kit* 260-101-098
*All items not depicted
KitsKit
s KitsKit
s
46800-88-BICON
Comprehensive Surgical KitDescription Part Number
2.0mm Pilot Drill 260-101-001
2.0mm Implant Inserter/Retriever 260-101-003
3.0mm Implant Inserter/Retriever 260-101-004
Threaded Offset Handle 260-101-009
5.0mm Threaded Knob 260-101-014
Threaded Straight Handle 260-101-016
2.0mm Implant/Angled Abutment Seating Tip 260-101-010
Standard Abutment Seating Tip 260-101-012
3.0mm Implant Seating Tip 260-101-018
Large Abutment Seating Tip 260-101-019
Removal Wrench 260-101-017
Latch Reamer Extension 260-101-022
2.5mm Hand Reamer 260-101-025
3.0mm Hand Reamer 260-101-030
3.5mm Hand Reamer 260-101-035
4.0mm Hand Reamer 260-101-040
4.5mm Hand Reamer 260-101-045
5.0mm Hand Reamer 260-101-050
5.5mm Hand Reamer 260-101-055
6.0mm Hand Reamer 260-101-060
Healing Plug Removal Tool (Latch) 260-101-115
2.5mm Standard Latch Reamer 260-101-125
3.0mm Standard Latch Reamer 260-101-130
3.5mm Standard Latch Reamer 260-101-135
4.0mm Standard Latch Reamer 260-101-140
4.5mm Standard Latch Reamer 260-101-645
5.0mm Standard Latch Reamer 260-101-650
5.5mm Standard Latch Reamer 260-101-655
6.0mm Standard Latch Reamer 260-101-660
Comprehensive Plus KitDescription Part Number
Depth Gauge/ Bone Depth Plugger 260-101-370
Abutment Shoulder Gauge 260-101-380
Silicone Dappen Dish 260-103-030
Surgical Tray 260-104-994
Paralleling Pin 0˚ (3) 260-701-001
Paralleling Pin 15˚ (2) 260-701-053
Paralleling Pin 25˚ (2) 260-701-054
2.0mm Standard Guide Pin (3) 260-101-180
3.0mm Standard Guide Pin (3) 260-101-183
3.5mm Sulcus Reamer 260-101-435
4.0mm Sulcus Reamer 260-101-440
5.0mm Sulcus Reamer 260-101-450
6.5mm Sulcus Reamer 260-101-465
7.5mm Sulcus Reamer 260-101-475
Healing Plug Cutter 260-801-001
Surgical Mallet 260-801-165
3.0mm Osteotome 260-901-530
3.5mm Osteotome 260-901-535
4.0mm Osteotome 260-901-540
5.0mm Osteotome 260-901-550
6.0mm Osteotome 260-901-560
KitsKit
s KitsKit
s
47 www.bicon.com
➤ Intrductory Surgical Kit Considerations
■ A pilot drill should be replaced after approximately 12 surgeries.
■ Latch reamers may be used for up to 100 surgeries without replacement.
■ It is advisable to have a spare pilot bur, a spare set of latch reamers, and a spare set of handpieces.
■ The tray may be sterilized at temperatures up to 275° F (135° C).
Introductory Surgical Kit
Tray depicted with cover
Description Part Number
Introductory Surgical Kit* 260-101-057
*All items not depicted
KitsKit
s KitsKit
s
48800-88-BICON
Introductory Surgical KitDescription Part Number
2.0mm Pilot Drill 260-101-001
Threaded Offset Handle 260-101-009
5.0mm Threaded Knob 260-101-014
Threaded Straight Handle 260-101-016
2.0mm Implant/Angled Abutment Seating Tip 260-101-010
Standard Abutment Seating Tip 260-101-012
3.0mm Implant Seating Tip 260-101-018
Latch Reamer Extension 260-101-022
2.5mm Standard Latch Reamer 260-101-125
3.0mm Standard Latch Reamer 260-101-130
3.5mm Standard Latch Reamer 260-101-135
4.0mm Standard Latch Reamer 260-101-140
4.5mm Standard Latch Reamer 260-101-645
5.0mm Standard Latch Reamer 260-101-650
5.5mm Standard Latch Reamer 260-101-655
6.0mm Standard Latch Reamer 260-101-660
Comprehensive Plus KitDescription Part Number
Depth Gauge / Bone Depth Plugger 260-101-370
Silicone Dappen Dish 260-103-030
Surgical Tray 260-104-994
Healing Plug Removal Instrument (Latch) 260-101-115
Surgical Mallet 260-801-165
Paralleling Pin 0˚ (1) 260-701-001
Paralleling Pin 15˚ (1) 260-701-053
Paralleling Pin 25˚ (1) 260-701-054
2.0mm Standard Guide Pin (2) 260-101-180
3.0mm Standard Guide Pin (2) 260-101-183
4.0mm Sulcus Reamer 260-101-440
5.0mm Sulcus Reamer 260-101-450
6.5mm Sulcus Reamer 260-101-465
KitsKit
s KitsKit
s
49 www.bicon.com
➤ Restorative Kit Considerations
■ Ideally used for the uncovering of a two stage placement and for other restorative procedures.
■ The tray may be sterilized at temperatures up to 275° F (135° C).
Restorative Kit
Tray depicted with cover
Description Part Number
Restorative Kit* 260-101-096
*All items not depicted
KitsKit
s KitsKit
s
50800-88-BICON
Restorative KitDescription Part Number
Threaded Offset Handle 260-101-009
5.0mm Threaded Knob 260-101-014
Threaded Straight Handle 260-101-016
2.0mm Implant/Angled Abutment Seating Tip 260-101-010
Standard Abutment Seating Tip 260-101-012
Crown Seating Tip 260-101-015
Large Abutment Seating Tip 260-101-019
Healing Plug Removal Instrument (Latch) 260-101-115
Healing Plug Removal Instrument (Manual) 260-101-114
Abutment Shoulder Gauge 260-101-380
2.0mm Impression Reamer 260-101-452
3.0mm Impression Reamer 260-101-453
Restorative Instrument Tray 260-104-996
2.0mm Standard Guide Pin (2) 260-101-180
3.0mm Standard Guide Pin (2) 260-101-183
2.0mm Extended Guide Pin (2) 260-101-184
3.0mm Extended Guide Pin (2) 260-101-185
Comprehensive Plus KitDescription Part Number
3.5mm Sulcus Reamer 260-101-435
4.0mm Sulcus Reamer 260-101-440
5.0mm Sulcus Reamer 260-101-450
6.5mm Sulcus Reamer 260-101-465
7.5mm Sulcus Reamer 260-101-475
Abutment Carrying Forceps 260-801-002
Surgical Mallet 260-801-165
KitsKit
s KitsKit
s
51 www.bicon.com
➤ Transitional Implant Kit Considerations
■ Kit can be used for placement of Stealth, O-Ring, or Orthodontic Transitional Implants.
■ A pilot drill should be replaced after approximately 12 surgeries.
■ The tray may be sterilized at temperatures up to 275° F (135° C).
Transitional Implant Kit
Tray depicted with cover
Description Part Number
Transitional Implant Kit 260-101-097
KitsKit
s KitsKit
s
52800-88-BICON
Transitional Implant KitDescription Part Number
2.0mm Pilot Drill (2) 260-101-001
Latch Reamer Extension 260-101-022
Paralleling Pin 0˚ (2) 260-701-001
Paralleling Pin 15˚ (2) 260-701-053
Paralleling Pin 25˚ (2) 260-701-054
Straight Stealth Latch Driver 260-101-905
Angled Stealth Latch Driver 260-101-906
Orthodontic Implant Driver 260-101-907
Straight O-Ring Latch Driver 260-101-902
Angled O-Ring Latch Driver 260-101-903
Transitional Implant Kit Tray 260-104-995
KitsKit
s KitsKit
s
➤ Handpiece Considerations
■ The 400:1 handpiece has sufficient torque to cut through bone at 50 RPM.
■ The 18:1 handpiece is designed to be used with the pilot drill at ~1,100 RPM with external irrigation.
■ Irrigation is not necessary at speeds of 50 RPM or lower.
■ Failure to observe speed limitation may result in burned or overheated bone which could result in substantial bone necrosis.
■ The 20,000 RPM Air Motor attaches to conventional four hole dental tubing and is an alternative to the electric drill unit.
■ Air pressure of at least 80 PSI is recommended to drive the air motor with sufficient torque in dense bone.
■ Proper maintenance of handpieces is crucial for long term success of handpieces.
■ Do not exceed 132°C (270°F) when sterilizing.
■ In order to prevent discoloration and/or damage to the plating of the handpieces from chemicals that are not sufficiently cleaned from other instruments, do not autoclave the handpieces with other instruments.
■ Always check the handpiece for any abnormal vibration, heating, noise, or sluggish operation. If any abnormality is noticed, cease the use of the handpiece.
55 www.bicon.com
Drill Unit and HandpiecesDescription Part Number
30,000 RPM Electric Drill Unit (110 volt) 260-028-110
(Foot pedal not depicted.)
30,000 RPM Electric Drill Unit (220 volt) 260-028-220
18:1 Contra-Angle Handpiece 260-028-181
400:1 Contra-Angle Handpiece 260-028-401
20,000 RPM Air Motor 260-024-120
Contra-Angle Handpiece Head 260-028-405
Pana Spray* 260-028-500
*Available only in the USA via ground shipping.
InstrumentsIn
stru
men
ts
InstrumentsIn
stru
men
ts
56800-88-BICON
Handpiece CompatibilityGear Reduction
1:1 16:1 18:1 64:1 256:1 400:1 1000:1
Motor Speed (RPM)
5,000 5,000 313 278 78 20 13 5 Reamer Speed (RPM)
20,000 20,000 1,250 1,111 313 78 50 20
30,000 30,000 1,875 1,666 469 117 75 30
To determine bur or Reamer Speed, you must know the Motor Speed and Divide that number by the Gear Reduction of the contra-angle.
Motor Speed ÷ Gear Reduction = Reamer Speed (RPM)e.g. 20,000 ÷ 400:1 = 50 RPM
Handpiece Maintenance
Cleaning and Lubrication➤ Clean and lubricate the contra-angle handpiece after each use.
■ Attach the metal spray nozzle into the back of the handpiece and insert the pana spray into the metal spray nozzle. (Figure 1)
■ Spray for approximately 2 seconds
■ Disassemble the head from the handpiece using the supplied wrench. (Figure 2)
■ Spray into the head. (Figure 3)
■ Assemble the head to the handpiece sheath in the reverse order of disassembly. Make sure that the two keys at the union nut align with the slots in the sheath and tighten the union nut.
■ Wipe the assembled handpiece clean.
Figure 1 Figure 2 Figure 3
Sterilization➤ The 400:1 and 18:1 handpieces are autoclavable.
■ Wipe any debris off with an alcohol-soaked cloth.
■ Lubricate the handpiece using spray lubricant. Insert the handpiece into a sterilizing pouch and seal it.
■ Autoclave for 15 minutes at 132° C (270° F). Do not set the sterilizer temperature above 132° C (270° F).
■ To expel excess oil, operate the handpiece before using it intra-orally.
InstrumentsIn
stru
men
ts
InstrumentsIn
stru
men
ts
57 www.bicon.com
Pilot Drill*
Description Type Part Number
2.0mm Pilot Drill SS** 260-101-001
Hand Reamers* ‡
Reamer Removal WrenchDescription Part Number
Removal Wrench 260-101-017
ExtendersDescription Part Number
Hand Reamer Extension 260-101-020
Latch Reamer Extension 260-101-022
Paralleling PinsDescription Part Number
0˚ Paralleling Pin 260-701-001
0˚ 15˚ 25˚
15˚ Paralleling Pin 260-701-053
25˚ Paralleling Pin 260-701-054
*See reference pages 85-86 for measurement guide.**SS denotes stainless steel.
‡ All threaded instrumentation is interchangeable.
All instruments are Ti6Al4V unless otherwise noted.
Diameter Length Part Number
2.5mm 21.0mm 260-101-025
2.5mm 3.0mm 3.5mm 4.0mm 4.5mm 5.0mm 5.5mm 6.0mm
3.0mm 21.0mm 260-101-030
3.5mm 21.0mm 260-101-035
4.0mm 21.0mm 260-101-040
4.5mm 21.0mm 260-101-045
5.0mm 21.0mm 260-101-050
5.5mm 21.0mm 260-101-055
6.0mm 21.0mm 260-101-060
InstrumentsIn
stru
men
ts
InstrumentsIn
stru
men
ts
58800-88-BICON
GaugesDescription Part Number
Depth Gauge/ Bone Depth Plugger 260-101-370
Abutment Shoulder Gauge 260-101-380
Standard Latch Reamers*
Diameter Length Part Number
2.5mm 33.0mm 260-101-125
2.5mm 3.0mm 3.5mm 4.0mm 4.5mm 5.0mm 5.5mm 6.0mm
3.0mm 31.0mm 260-101-130
3.5mm 31.0mm 260-101-135
4.0mm 31.0mm 260-101-140
4.5mm 31.0mm 260-101-645
5.0mm 31.0mm 260-101-650
5.5mm 29.0mm 260-101-655
6.0mm 29.0mm 260-101-660
Mid-Size Latch Reamers*
Diameter Length Part Number
2.5mm 38.0mm 260-101-325
2.5mm 3.0mm 3.5mm 4.0mm 4.5mm 5.0mm 5.5mm 6.0mm
3.0mm 38.0mm 260-101-330
3.5mm 38.0mm 260-101-335
4.0mm 38.0mm 260-101-340
4.5mm 38.0mm 260-101-345
5.0mm 38.0mm 260-101-350
5.5mm 32.0mm 260-101-355
6.0mm 32.0mm 260-101-360
*See reference pages 85-86 for measurement guide. All instruments are Ti6Al4V unless otherwise noted.
InstrumentsIn
stru
men
ts
InstrumentsIn
stru
men
ts
59 www.bicon.com
Threaded Components‡
Description Part Number
Threaded Straight Handle 260-101-016
Threaded Offset Handle 260-101-009
5.0mm Threaded Knob 260-101-014
5.0mm
Seating Tips‡
Description Part Number
2.0mm Implant/Angled Abutment Seating Tip 260-101-010
2.0mm 3.0mm
3.0mm Implant Seating Tip 260-101-018
Standard Seating Tip (4.0mm) 260-101-012
4.0mm 6.5mm
Large Abutment Seating Tip (6.5mm) 260-101-019
Crown Seating Tip 260-101-015
Inserter/Retriever ◊ ‡
Description Part Number
2.0mm Standard Implant Inserter/Retriever 260-101-003
2.0mm Standard 3.0mm Standard
Example
3.0mm Standard Implant Inserter/Retriever 260-101-004
2.0mm Long Implant Inserter/Retriever 260-101-005
2.0mm Long 3.0mm Long
3.0mm Long Implant Inserter/Retriever 260-101-006
◊5.0mm Threaded Knob not included with Inserter/Retriever.‡
All threaded instrumentation is interchangeable.
All instruments are Ti6Al4V unless otherwise noted.
InstrumentsIn
stru
men
ts
InstrumentsIn
stru
men
ts
60800-88-BICON
Bone Expanders‡
Diameter Part Number
3.0mm 260-901-430
3.0mm 3.5mm 4.0mm 4.5mm 5.0mm 5.5mm 6.0mm
3.5mm 260-901-435
4.0mm 260-901-440
4.5mm 260-901-445
5.0mm 260-901-450
5.5mm 260-901-455
6.0mm 260-901-460
Osteotomes‡
Diameter Part Number
3.0mm 260-901-530
3.0mm 3.5mm 4.0mm 5.0mm 6.0mm
3.5mm 260-901-535
4.0mm 260-901-540
5.0mm 260-901-550
6.0mm 260-901-560
‡ All threaded instrumentation is
interchangeable.All instruments are Ti6Al4V unless otherwise noted.
Instrument Storage TraysDescription Part Number Description Part Number
Surgical Instrument Tray 260-104-099 Surgical Instrument Tray 260-104-994
Description Part Number
Restorative Instrument Tray 260-104-996
InstrumentsIn
stru
men
ts
InstrumentsIn
stru
men
ts
61 www.bicon.com
Guide PinsDescription Part Number
2.0mm Standard Guide Pin (4) 260-101-180
2.0mm 3.0mm 2.0mm 3.0mmStandard Extended
Example
3.0mm Standard Guide Pin (4) 260-101-183
2.0mm Extended Guide Pin (4) 260-101-184
3.0mm Extended Guide Pin (4) 260-101-185
Sulcus Reamers† ‡
Diameter Part Number
3.5mm 260-101-435
3.5mm 4.0mm 5.0mm 6.5mm 7.5mm
Example
4.0mm 260-101-440
5.0mm 260-101-450
6.5mm 260-101-465
7.5mm 260-101-475
Tissue Punches† ‡
Diameter Part Number
3.5mm 260-101-535
3.5mm 4.0mm 5.0mm 6.5mm
Example
4.0mm 260-101-540
5.0mm 260-101-550
6.5mm 260-101-565
†Each part is packaged with a 2.0 and 3.0mm standard guide pin.
‡ All threaded instrumentation is interchangeable.
All instruments are Ti6Al4V unless otherwise noted.
InstrumentsIn
stru
men
ts
InstrumentsIn
stru
men
ts
62800-88-BICON
Impression Reamers
Threaded Knob
Guide PinImpression Reamer
2.0mm 3.0mm
Threaded Knob
Guide Pin
Impression Reamer
1 Insert guide pin 2 Slide impression reamer over guide pin 3 Impression reamers
4 Rotate while pressing 5 Tap impression post
Impression Reamers† ‡
Diameter Part Number
2.0mm 260-101-452
2.0mm 3.0mm
Example
3.0mm 260-101-453
†Each part is packaged with a 2.0 and 3.0mm standard guide pin.
‡ All threaded instrumentation is interchangeable.
All instruments are Ti6Al4V unless otherwise noted.
InstrumentsIn
stru
men
ts
InstrumentsIn
stru
men
ts
63 www.bicon.com
Restorative AccessoriesDescription Part Number
Abutment Preparation Holder 260-101-3902.0mm 3.0mm
Crown Alignment Device 260-101-315
Includes Two Crown Seating Tips
Crown Seating Tip 260-101-015
Hydroplastic Resin 260-103-031
IAC Light Cure Booth (110 volt) 260-499-110
IAC Light Cure Booth (multi voltage) 260-499-220 ™
Crown Alignment Device
IAC
Seating Jig (molded plastic)
Crown Seating Tip
IAC or extra-orally cemented crown Crown alignment device Plastic molded over crown
Seating jig fabrication Proper direction for seating force Custom seating jig
InstrumentsIn
stru
men
ts
InstrumentsIn
stru
men
ts
64800-88-BICON
BursDescription Type Part Number
#1557 Regular Length Bur (5) Carbide 260-155-7RL
#1557 Latch Bur (5) Carbide 260-155-7LT
#4 Round Bur Regular Length (5) Carbide 260-2RL-GW4
11.0mm Abutment Cutting Bur Carbide 260-201-011
ForcepsDescription Type Part Number
Abutment Removal Forceps #150 SS** 260-801-050
Abutment Removal Forceps #151 SS 260-801-051
Abutment Removal Forceps for 3.0mm Post SS 260-801-052
Abutment Carrying Forceps SS 260-801-002
Shade Enhancer Plastic 260-103-006
**SS denotes stainless steel. All instruments are Ti6Al4V unless otherwise noted.
Denture AccessoriesDescription Part Number
Denture Duplicator 260-103-98F
Denture Duplicator Kit ¤ 260-103-99K
¤Consists of denture duplicator, silicone spray and Ortho-Jet™ Acrylic. Available only in the USA via ground shipping.
InstrumentsIn
stru
men
ts
InstrumentsIn
stru
men
ts
65 www.bicon.com
Surgical AccessoriesDescription Type Part Number
Surgical Mallet SS** 260-801-165
Healing Plug Removal Instrument Manual 260-101-114
Healing Plug Removal Instrument Latch 260-101-115
Healing Plug Cutter SS 260-801-001
Silicone Dappen Dish 260-103-030
30 x 42mm Radiograph Grid 260-103-002
Implant Spacing Template 260-101-305
2.0mm Cotton Tip Applicator (100) 260-103-027
3.0mm Cotton Tip Applicator (100) 260-103-028
Chisels‡
Description Part Number
3.0mm Threaded Expanding Chisel 260-901-613
5.0mm Threaded Expanding Chisel 260-901-615
6.0mm Threaded Expanding Chisel 260-901-616
3.0mm Expanding Chisel 260-801-330
4.5mm Expanding Chisel 260-801-345
5.5mm Expanding Chisel 260-801-355
**SS denotes stainless steel.‡
All threaded instrumentation is interchangeable.
All instruments are Ti6Al4V unless otherwise noted.
InstrumentsIn
stru
men
ts
InstrumentsIn
stru
men
ts
66800-88-BICON
TrephinesDiameter Outside Diameter Type Part Number
3.75mm 4.6mm SS** 260-801-837
3.75mm 4.25mm 5.25mm 6.0mm 7.0mm 8.0mm 10.0mm
4.25mm 5.2mm SS 260-801-840
5.25mm 6.2mm SS 260-801-850
6.0mm 6.95mm SS 260-801-860
7.0mm 7.95mm SS 260-801-870
8.0mm 8.95mm SS 260-801-880
10.0mm 10.0mm SS 260-801-810
Bone Screw Tack and Seating InstrumentsDescription Part Number
3.0mm Bone Screw Tack (10) 260-201-310
Bone Screw Tack Cassette 260-801-305
Tack Seating Instrument 260-201-700
**SS denotes stainless steel. All instruments are Ti6Al4V unless otherwise noted.
PeriotomesDescription Type Part Number
Double-Ended Straight Periotome SS 260-801-007
Double-Ended Offset Periotome SS 260-801-008
CurettesDescription Type Part Number
Straight Curette SS 260-801-520
Angled Curette SS 260-801-522
Lateral Sinus Lift InstrumentationDescription Type Part Number
Severe Bend Sinus Lift Curette SS 260-801-515
Standard Bend Sinus Lift Curette SS 260-801-517
Antralplasty Elevator SS 260-801-518
Sinus Lift Bur Carbide 260-201-029
InstrumentsIn
stru
men
ts
InstrumentsIn
stru
men
ts
69 www.bicon.com
Stealth Transitional Implants and Components*
Description Angle Diameter Height Part Number
2.5 x 8.0 x 4.0mm 0˚ Stealth Transitional Implant 0˚ 2.5mm 8.0mm 260-425-108
8.0mm 0˚ 8.0mm 10˚
2.5 x 8.0 x 4.0mm 10˚ Stealth Transitional Implant 10˚ 2.5mm 8.0mm 260-425-109
2.5 x 12 x 4.0mm 0˚ Stealth Transitional Implant 0˚ 2.5mm 12.0mm 260-425-112
12.0mm 0˚ 12.0mm 10˚
2.5 x 12 x 4.0mm 10˚ Stealth Transitional Implant 10˚ 2.5mm 12.0mm 260-425-113
4.0mm Thin Wall Stealth Sleeve (4) 4.0mm 5.5mm 260-250-204
4.0mm
Straight Stealth Latch Driver 260-101-905
Straight Driver Angled Driver
Angled Stealth Latch Driver 260-101-906
*Packaged with two 4.0mm stealth acrylic sleeves.
Orthodontic Implant and ComponentsDescription Diameter Length Part Number
8.0mm Orthodontic Implant 2.5mm 8.0mm 260-325-208
8.0mm
Orthodontic Implant Driver 260-101-907
Orthodontic Driver
Transitional Implant Kit Tray
Description Part Number
Transitional Implant Kit Tray 260-104-995
Example
TransitionalTran
siti
onal
TransitionalTran
siti
onal
70800-88-BICON
O-Ring Transitional Implants and ComponentsDescription Angle Diameter Height Part Number
2.5 x 8.0mm 0˚ O-Ring Transitional Implant 0˚ 2.5mm 8.0mm 260-425-208
8.0mm 0˚ 8.0mm 15˚
2.5 x 8.0mm 15˚ O-Ring Transitional Implant 15˚ 2.5mm 8.0mm 260-425-209
2.5 x 12.0mm 0˚ O-Ring Transitional Implant 0˚ 2.5mm 12.0mm 260-425-212
12.0mm 0˚ 12.0mm 15˚
2.5 x 12.0mm 15˚ O-Ring Transitional Implant 15˚ 2.5mm 12.0mm 260-425-213
Brevis™ Abutment Chairside Kit 260-100-212
O-Ring Titanium Housing
Straight O-Ring Latch Driver 260-101-902
Straight Driver Angled DriverAngled O-Ring Latch Driver 260-101-903
➤ Transitional Implant Considerations
Procedure:
■ Drill a 2.0mm osteotomy to appropriate depth with Bicon’s 2.0mmpilot drill with external irrigation.
■ Take care not to overheat the bone while drilling at ~1,100 RPM.
■ Pilot bur is marked at 6.0mm, 8.0mm, 11.0mm and 14.0mm.*
■ Be sure to drill to appropriate depth of intended implant to be placed or slightly beyond if the anatomy permits.
■ Make note of the quality of bone.
■ The Stealth and O-Ring implants are available in 8.0mm and 12.0mmlengths and either straight or angled.
■ Check angulation or trajectory of osteotomy with a paralleling pin.
■ Remove appropriate transitional implant from packaging and insert apical end of implant into osteotomy.
■ Assemble appropriate transitional implant driver to Bicon’s 400:1handpiece or a comparable slow speed latch-type handpiece.
■ Slowly screw implant into the prepared osteotomy.
■* Some pilot burs may have markings only at 8.0mm, 11.0mm, and 14.0mm. Check burs
before drilling. (See page 83)
2.5mm
8.0mm
4.0mm
2.5mm
12.0mm
8.0mm
2.5mm
4.0mm
12.0mm
2.5mm
8.0mm
4.5mm
2.5mm
TransitionalTran
siti
onal
TransitionalTran
siti
onal
73 www.bicon.com
Educ
atio
nal
Education and LiteratureDescription Part Number
Short Implant Brochure LIT-200
History of Bicon LIT-015
HISTORY OF BICON
SynthoGraft™ Brochure LIT-231
Introduction to Bicon LIT-235
Bicon Patient Teasers (25) LIT-240
Patient Education Brochures (25) LIT-400
Description Part Number
Poster for Waiting Room LIT-405
Surgical Manual LIT-005
Porcelain Fused to MetalTechnique ManualNon-Shouldered Abutment
LIT-250
Brevis™ Overdenture AbutmentTechnique Manual
LIT-255
DVDsDescription
Instructional DVD 260-102-111
Patient DVDOn one DVD:
English, Chinese, German,
Italian, Japanese, Korean,
Portuguese, and Spanish
260-102-112
74800-88-BICON
Demonstration Models and Replicas*
Description Style Part Number
Crown Model with two Single Crowns Acrylic 260-102-115
O-Ring Model with Snap-On Denture Acrylic 260-102-120
Practice Model Foam 260-102-104
4.0 x 8.0mm Implant Replica 2.0mm Well Brass 260-540-008
260-540-008 260-540-011(Actual Size)
4.0 x 11.0mm Implant Replica 2.0mm Well Brass 260-540-011
4.5 x 8.0mm Implant Replica 3.0mm Well Brass 260-545-308
260-545-308 260-550-306 260-550-308 260-560-357(Actual Size)
5.0 x 6.0mm Implant Replica 3.0mm Well Brass 260-550-306
5.0 x 8.0mm Implant Replica 3.0mm Well Brass 260-550-308
6.0 x 5.7mm Implant Replica 3.0mm Well Brass 260-560-357
4.0 x 6.5mm 0˚ Non-Shouldered Abutment Replica 2.0mm Post Brass 260-240-002
260-240-002 260-550-301 260-400-301(Actual Size)
5.0 x 6.5mm 0˚ Non-Shouldered Abutment Replica 3.0mm Post Brass 260-550-301
5.0 x 2.0mm 0˚ Stealth Shouldered Abutment Replica 3.0mm Post Brass 260-400-301
2.0mm Brevis™ Abutment Replica 2.0mm Post Brass 260-500-204
260-500-204 260-500-234(Actual Size)
2.0mm Brevis™ Abutment Replica 3.0mm Post Brass 260-500-234
5.0 x 6.0mm Implant Replica 3.0mm Well Aluminum 260-102-306
260-102-306 260-102-308 260-102-311(4x Actual Size)
5.0 x 8.0mm Implant Replica 3.0mm Well Aluminum 260-102-308
5.0 x 11.0mm Implant Replica 3.0mm Well Aluminum 260-102-311
5.0 x 6.5mm 0˚ Non-Shouldered Abutment Replica 3.0mm Well Aluminum 260-102-301
260-102-301 260-102-302(4x Actual Size)
5.0 x 2.0mm 0˚ Stealth Shouldered Abutment Replica 3.0mm Well Aluminum 260-102-302
*Demonstration and teaching models only. Not for clinical use.
EducationalEd
ucat
iona
l EducationalEd
ucat
iona
l
75 www.bicon.com
Titanium Implant Replicas (3x actual size)Description Part Number Description Part Number
3.5 x 11.0mm Implant 260-035-011 5.0 x 6.0mm Implant 260-050-306
4.0 x 11.0mm Implant 260-040-011 5.0 x 8.0mm Implant 260-050-308
4.5 x 6.0mm Implant 260-045-306 6.0 x 5.7mm Implant 260-060-357
4.5 x 8.0mm Implant 260-045-308
3.5 x 11.0mm 4.0 x 11.0mm 4.5 x 6.0mm 4.5 x 8.0mm 5.0 x 6.0mm 5.0 x 8.0mm 6.0 x 5.7mm
260-035-011 260-040-011 260-045-306 260-045-308 260-050-306 260-050-308 260-060-357
Titanium Non-Shouldered Abutment Replicas (3x actual size)Description Part Number Description Part Number
4.0 x 6.5mm Non-Shouldered Abutment 3.0mm Post 260-040-001 5.0 x 6.5mm Non-Shouldered Abutment 3.0mm Post 260-050-001
4.0 x 6.5mm Non-Shouldered Abutment 2.0mm Post 260-040-002 5.0 x 6.5mm Non-Shouldered Abutment 2.0mm Post 260-050-002
4.0 x 6.5mm 4.0 x 6.5mm 5.0 x 6.5mm 5.0 x 6.5mm
260-040-001 260-040-002 260-050-001 260-050-002
EducationalEd
ucat
iona
l EducationalEd
ucat
iona
l
76800-88-BICON
Titanium Stealth Abutment Replicas (3x actual size)Description Part Number Description Part Number
5.0 x 2.0mm Stealth Abutment 2.0mm Post 260-000-201 5.0 x 2.0mm Stealth Abutment 3.0mm Post 260-000-301
5.0 x 4.0mm Stealth Abutment 2.0mm Post 260-000-221 5.0 x 4.0mm Stealth Abutment 3.0mm Post 260-000-321
5.0 x 2.0mm 5.0 x 4.0mm 5.0 x 2.0mm 5.0 x 4.0mm
260-000-201 260-000-221 260-000-301 260-000-321
Titanium Brevis™ Abutment Replicas (3x actual size)Description Part Number Description Part Number
2.0mm Brevis™ Abutment 2.0mm Post 260-000-404 2.0mm Brevis™ Abutment 3.0mm Post 260-000-434
260-000-404 260-000-434
EducationalEd
ucat
iona
l EducationalEd
ucat
iona
l
79 www.bicon.com
Method ofImpression
ImpressionProduct Needed
TemporizationProduct Needed
Laboratory Product Needed
Non-Shouldered Abutment Stealth Shouldered Abutment Non-Shouldered Abutment orStealth Shouldered Abutment
Implant Level Implant Level Implant LevelIndirect
Abutment Level
Implant LevelImpression Kit
Implant LevelImpression Kit
Implant LevelImpression Kit
Temporary or HealingAbutment and/or
Transitional Crown
Temporary or HealingAbutment and/or
Transitional Crown
Temporary or HealingAbutment and/or
Transitional Crown
Final CrownFinal Crown Final Crown
Non-ShoulderedAbutment
Aluminum OxideSleeve &
Stealth Shouldered Abutment
Aluminum OxideSleeve &
AbutmentTransfer Die
Non-ShoulderedAbutment or
Stealth ShoulderedAbutment
DirectAbutment Level
Emergence Cuff or Temporization Sleevefor Non-Shouldered
Abutment
None
None
IndirectAbutment Level
Impression Sleevefor Non-Shouldered
Abutment
Stealth Acrylic Sleevefor Shouldered
Abutment
Stealth Acrylic Sleevefor Shouldered
Abutment
Temporization Sleevefor Non-Shouldered
Abutment
Abutment Transfer Dieand Acrylic Sleeve
for Non-Shouldered Abutment
• For extra-oral cementation, follow the implant level impression technique. • Each implant level impression kit contains one impression post and one implant analog.
Single Crown and Bridge Restorations
PFM
ReferenceRefe
renc
e ReferenceRefe
renc
e
80800-88-BICON
Method ofImpression
ImpressionProduct Needed
TemporizationProduct Needed
Laboratory Product Needed
Non-Shouldered Abutment Stealth Shouldered Abutment Non-Shouldered Abutment orStealth Shouldered Abutment
Implant Level Implant Level Implant LevelIndirect
Abutment Level
Implant LevelImpression Kit
Implant LevelImpression Kit
Implant LevelImpression Kit
Temporary or HealingAbutment and/or
Transitional Crown
Temporary or HealingAbutment and/or
Transitional Crown
Temporary or HealingAbutment and/or
Transitional Crown
Final CrownFinal Crown Final Crown
Non-ShoulderedAbutment
Aluminum OxideSleeve &
Stealth Shouldered Abutment
Aluminum OxideSleeve &
AbutmentTransfer Die
Non-ShoulderedAbutment or
Stealth ShoulderedAbutment
DirectAbutment Level
Emergence Cuff or Temporization Sleevefor Non-Shouldered
Abutment
None
None
IndirectAbutment Level
Impression Sleevefor Non-Shouldered
Abutment
Stealth Acrylic Sleevefor Shouldered
Abutment
Stealth Acrylic Sleevefor Shouldered
Abutment
Temporization Sleevefor Non-Shouldered
Abutment
Abutment Transfer Dieand Acrylic Sleeve
for Non-Shouldered Abutment
• For extra-oral cementation, follow the implant level impression technique. • Each implant level impression kit contains one impression post and one implant analog.
™
Integrated Abutment Crown™All-Ceramic
ReferenceRefe
renc
e ReferenceRefe
renc
e
81 www.bicon.com
Non-Shouldered Abutments and Prosthetic Components
Notes: Snap-on sleeves are only specific for abutment diameter. Abutment height is not a criterion for proper selection of snap-on sleeves.Transfer dies correspond to exact diameter and height of abutment placed.
Indirect Abutment Level Impression
Modified and unmodified color-coded impression sleeves are definitively seated on their
corresponding abutments.
Impression material is injected around the impression sleeves for the making of an abutment
level transfer impression.
Acrylic impression sleeves withdrawn in impression material prior to impression being sent to
the laboratory.
Soft tissue material being injected around impression sleeves and
abutment transfer dies.
5.0mm
5.0
x 12
15˚
260-
150-
215
5.0
x 12
0˚
260-
150-
201
12.0mm Height
5.0
x 10
15˚
260-
150-
115
5.0
x 10
0˚
260-
150-
101
10.0mm Height
5.0
x 6.
5 25
˚26
0-15
0-02
5
5.0
x 6.
5 15
˚26
0-15
0-01
5
5.0
x 6.
5 0˚
260-
150-
001
6.5mm Height
5.0
x 5.
0 15
˚26
0-15
0-05
5
5.0
x 5.
0 0˚
260-
150-
050
5.0mm Height
2.0mm
5.0
x 5.
0 0˚
260-
350-
050
5.0
x 5.
0 15
˚26
0-35
0-05
5
5.0mm Height
3.0mm
5.0
x 12
0˚
260-
350-
201
5.0
x 12
15˚
260-
350-
215
12.0mm Height
5.0
x 6.
5 0˚
260-
350-
001
5.0
x 6.
5 15
˚26
0-35
0-01
5
5.0
x 6.
5 25
˚26
0-35
0-02
5
6.5mm Height
5.0 x 12.0mmRestorative/Laboratory Kit
260-150-412
12.0mm Height
5.0mmTemporization Sleeve (2)
260-150-165
5.0 x 6.5mmRestorative/Laboratory Kit
260-150-465
6.5mm Height
5.0 x 5.0mmRestorative/Laboratory Kit
260-150-450
5.0mm Height
5.0 x 10.0mmRestorative/Laboratory Kit
260-150-410
10.0mm Height
6.5
x 5.
0 0˚
260-
365-
050
6.5
x 5.
0 15
˚26
0-36
5-05
5
6.5
x 6.
5 0˚
260-
365-
001
6.5
x 6.
5 15
˚26
0-36
5-01
5
6.5mm Diameter
6.5 x 6.5mmRestorative/Laboratory Kit
260-165-465
6.5 x 5.0mmRestorative/Laboratory Kit
260-165-450
6.5mmTemporization Sleeve (2)
260-165-165
3.0mm Post
6.5mm Height5.0mm Height
6.5
x 5.
0 0˚
260-
165-
050
6.5
x 5.
0 15
˚26
0-16
5-05
5
5.0mm Height
2.0mm Post
5.0mm Height6.5mm Height
7.5 x 8.0mmRestorative/Laboratory Kit
260-175-480
7.5
x 8.
0 0˚
260-
375-
801
7.5
x 8.
0 15
˚26
0-37
5-81
5
7.5mm Diameter
7.5mmTemporization Sleeve (2)
260-175-165
3.0mm Post
8.0mm Height
8.0mm Height
3.5mmTemporization Sleeve (2)
260-135-165
3.5 x 6.5mmRestorative/Laboratory Kit
260-135-465
3.5mm Diameter
2.0mm Post
6.5mm Height
3.5
x 6.
5 0˚
260-
135-
001
3.5
x 6.
5 25
˚26
0-13
5-02
5
3.5
x 6.
5 15
˚26
0-13
5-01
5
6.5mm Height
4.0
x 6.
5 0˚
260-
140-
002
4.0 x 6.5mmRestorative/Laboratory Kit
260-140-465
4.0
x 10
15˚
260-
140-
115
4.0
x 10
0˚
260-
140-
101
4.0
x 6.
5 25
˚26
0-14
0-02
5
4.0
x 6.
5 15
˚26
0-14
0-01
5
4.0
x 6.
5 0˚
260-
340-
001
4.0
x 6.
5 15
˚26
0-34
0-01
5
4.0mm Diameter
4.0 x 10.0mmRestorative/Laboratory Kit
260-140-410
4.0mmTemporization Sleeve (2)
260-140-165
6.5mm Height
10.0mm Height
2.0mm Post
3.0mm Post
6.5mm Height
6.5mm Height
10.0mm Height
4.0
x 10
15˚
260-
340-
115
4.0
x 10
0˚
260-
340-
101
10.0mm Height
5.0
x 10
15˚
260-
350-
115
5.0
x 10
0˚
260-
350-
101
10.0mm Height
Diameter
Post
Post
ReferenceRefe
renc
e ReferenceRefe
renc
e
82800-88-BICON
Because of machining tolerances, acrylic sleeves may not reach the height of contour for some angled abutments.
Direct Abutment Level Impression
Non-shouldered abutment being prepared with a #1557 carbide bur.
Two prepared non-shouldered abutments.
Impression material being injected around non-shouldered abutments.
Full arch impression.
5.0mm
5.0
x 12
15˚
260-
150-
215
5.0
x 12
0˚
260-
150-
201
12.0mm Height
5.0
x 10
15˚
260-
150-
115
5.0
x 10
0˚
260-
150-
101
10.0mm Height
5.0
x 6.
5 25
˚26
0-15
0-02
5
5.0
x 6.
5 15
˚26
0-15
0-01
5
5.0
x 6.
5 0˚
260-
150-
001
6.5mm Height
5.0
x 5.
0 15
˚26
0-15
0-05
5
5.0
x 5.
0 0˚
260-
150-
050
5.0mm Height
2.0mm
5.0
x 5.
0 0˚
260-
350-
050
5.0
x 5.
0 15
˚26
0-35
0-05
5
5.0mm Height
3.0mm
5.0
x 12
0˚
260-
350-
201
5.0
x 12
15˚
260-
350-
215
12.0mm Height
5.0
x 6.
5 0˚
260-
350-
001
5.0
x 6.
5 15
˚26
0-35
0-01
5
5.0
x 6.
5 25
˚26
0-35
0-02
5
6.5mm Height
5.0 x 12.0mmRestorative/Laboratory Kit
260-150-412
12.0mm Height
5.0mmTemporization Sleeve (2)
260-150-165
5.0 x 6.5mmRestorative/Laboratory Kit
260-150-465
6.5mm Height
5.0 x 5.0mmRestorative/Laboratory Kit
260-150-450
5.0mm Height
5.0 x 10.0mmRestorative/Laboratory Kit
260-150-410
10.0mm Height
6.5
x 5.
0 0˚
260-
365-
050
6.5
x 5.
0 15
˚26
0-36
5-05
5
6.5
x 6.
5 0˚
260-
365-
001
6.5
x 6.
5 15
˚26
0-36
5-01
5
6.5mm Diameter
6.5 x 6.5mmRestorative/Laboratory Kit
260-165-465
6.5 x 5.0mmRestorative/Laboratory Kit
260-165-450
6.5mmTemporization Sleeve (2)
260-165-165
3.0mm Post
6.5mm Height5.0mm Height
6.5
x 5.
0 0˚
260-
165-
050
6.5
x 5.
0 15
˚26
0-16
5-05
5
5.0mm Height
2.0mm Post
5.0mm Height6.5mm Height
7.5 x 8.0mmRestorative/Laboratory Kit
260-175-480
7.5
x 8.
0 0˚
260-
375-
801
7.5
x 8.
0 15
˚26
0-37
5-81
5
7.5mm Diameter
7.5mmTemporization Sleeve (2)
260-175-165
3.0mm Post
8.0mm Height
8.0mm Height
3.5mmTemporization Sleeve (2)
260-135-165
3.5 x 6.5mmRestorative/Laboratory Kit
260-135-465
3.5mm Diameter
2.0mm Post
6.5mm Height
3.5
x 6.
5 0˚
260-
135-
001
3.5
x 6.
5 25
˚26
0-13
5-02
5
3.5
x 6.
5 15
˚26
0-13
5-01
5
6.5mm Height
4.0
x 6.
5 0˚
260-
140-
002
4.0 x 6.5mmRestorative/Laboratory Kit
260-140-465
4.0
x 10
15˚
260-
140-
115
4.0
x 10
0˚
260-
140-
101
4.0
x 6.
5 25
˚26
0-14
0-02
5
4.0
x 6.
5 15
˚26
0-14
0-01
5
4.0
x 6.
5 0˚
260-
340-
001
4.0
x 6.
5 15
˚26
0-34
0-01
5
4.0mm Diameter
4.0 x 10.0mmRestorative/Laboratory Kit
260-140-410
4.0mmTemporization Sleeve (2)
260-140-165
6.5mm Height
10.0mm Height
2.0mm Post
3.0mm Post
6.5mm Height
6.5mm Height
10.0mm Height
4.0
x 10
15˚
260-
340-
115
4.0
x 10
0˚
260-
340-
101
10.0mm Height
5.0
x 10
15˚
260-
350-
115
5.0
x 10
0˚
260-
350-
101
10.0mm Height
Diameter
Post
Post
ReferenceRefe
renc
e ReferenceRefe
renc
e
83 www.bicon.com
Stealth Shouldered Abutment System2.0mm Post
Product 3.5mm 4.0mm 5.0mm
Abutment Transfer Dies
3.5mmPlastic
4.0mmBrass
5.0mmBrass
Stealth Acrylic Sleeve
3.5mm
Thin Wall Stealth Sleeves
4.0mm 5.0mm
Abutment Shoulder Gauge
Stealth ShoulderedAbutments 2.0mm Post
3.5mm 4.0mm 5.0mm
Aluminum Oxide Sleeves
3.5 x 7.0mm 4.0 x 7.0mm 5.0 x 7.0mm 5.0 x 10.0mm ReferenceRefe
renc
e ReferenceRefe
renc
e
84800-88-BICON
Stealth Shouldered Abutment System3.0mm Post
Product 4.0mm 5.0mm
Abutment Transfer Dies
4.0mmBrass
5.0mmBrass
Thin Wall Stealth Sleeves
4.0mm 5.0mm
Abutment Shoulder Gauge
Stealth ShoulderedAbutments 3.0mm Post
4.0mm
5.0mm
Aluminum Oxide Sleeves
4.0 x 7.0mm 5.0 x 7.0mm 5.0 x 10.0mm
ReferenceRefe
renc
e ReferenceRefe
renc
e
85 www.bicon.com
Pilot Drill WarningIt is imperative that the depth indicators on the 2.0mm Pilot Drill are identified prior to surgery. No assumptions should be made about the height of the first marking on the pilot drill.
!
Pilot Drill
11.0mm
8.0mm6.0mmNEW!
14.0mm
Old New
ReferenceRefe
renc
e ReferenceRefe
renc
e
86800-88-BICON
Latch Reamer Measurements with Corresponding Implant Examples
Latch Reamers with Measurements
2.5mm 3.0mm 3.5mm 4.0mm 6.0mm5.5mm5.0mm
14.0mm
11.0mm8.0mm
4.5mm
6.0mm
5.0mm 5 x 115 x 84.5mm4.0mm3.5mm3.0mm
14.0mm
11.0mm8.0mm
5 x 65.5mm 6.0mm
6.0mm
2.5mm
Latch Reamer WarningThe reamers are used sequentially beginning with a 2.5mm diameter and ending with the diameter of the intended implant. The latch reamers have horizontal markings at 6.0, 8.0, 11.0 and 14.0mm, whereas older reamers did not have the 6.0mm mark. It is imperative that the depth indicators on the latch reamers are identified prior to surgery. No assumptions should be made about the height of the first marking on any latch reamer.
!
Latch Reamers
ReferenceRefe
renc
e ReferenceRefe
renc
e
87 www.bicon.com
Return Policies and Procedures
Claims for short shipment must be made within 48 hours of receipt. ICC regulations require that claims for damaged merchandise be made with the carrier within 10 days of receipt. Do not accept damaged ship ments unless such damage is noted on the Bill of Lading at the time of receipt. Bicon will not issue credit for shipping damages unless the above procedure is followed.
Claims
Orders may be placed online at www.bicon.com, by fax, or phone. Order forms can be obtained by visiting the Bicon website.Online orders receive a 2% discount.
Bicon501 ArborwayBoston, MA 02130 U.S.A.
Fax: 800.282.4266 or + 1 617.524.0096
Voice: 800.882.4266 or + 1 617.524.4443
E-Mail: [email protected] Web: www.bicon.com
Office hours for customer service are from 8:00am-6:00pm Mon-Fri EST.Orders received by 5:00pm EST Mon-Fri will be processed and shipped the same day.
Placing Orders
Items returned within 30 days from the date of purchase are eligible for full credit.
Items returned between 31 and 90 days from the date of purchase are subject to a 15% restocking fee.
Items will not be accepted for credit or exchange after 90 days from the date of purchase.
Items must be in their original undamaged and unopened packaging with all seals intact.
Implants and abutments being returned for credit or exchange must have at least a two year shelf life remaining.
Shipping charges for any return must be paid by the customer.
Individual items from a kit or previously exchanged items may not be returned for credit.
Return Policy
Return Procedure Call the Customer Service Department at 800.88.BICON to receive a Return Authorization Number (RA#). The granting of an RA# does not guarantee that the product will be accepted for return. All return policy conditions must be met for the return to be processed.
A copy of the corresponding invoice must be enclosed with the shipment.
Send all returns to:
Items missing from package, or damaged due to improper packaging will not be given credit.
Bicon Returns Department501 ArborwayBoston, MA 02130 USA
Refe
renc
e
88800-88-BICON
FAX Order Form » No cover sheet necessary
Page of
Orders for overnight delivery must be received by 5:00 PM EST.Toll-Free fax number available 24 hours a day : 800.282.4266 or + 1 617.524.0096.Confirmation will be made by fax only, if indicated.
Confirmation by fax?(circle one)
YES NO
Customer Number Date
Account Name
Ordered By PO #
Telephone FAX
Shipping AddressCity State ZIP
Quantity Item Number Description
260 - -
260 - -
260 - -
260 - -
260 - -
260 - -
260 - -
260 - -
260 - -
260 - -
Check here if order continues on a second page
Shipping » Exclusively via Fed Ex
Next Day AM10:30am to most ZIP codes
Next Day PM4:30pm to most ZIP codes
2nd Day PM4:30pm on 2nd day
InternationalDepends upon customs clearance. Includes Canada and Puerto Rico
If an option is not selected, shipment is sent 2nd day. Bicon is not responsible for weather or Fed Ex delays.
Payment Information » Please select one
MasterCard Visa American Express CODAdds 2 hours to delivery time + a $5.00 fee
Card Number Expiration Date
Bicon implants, abutments, instrumentation, and components are engineered, manufactured, coated, packaged, and/or sterilized at ISO Certified facilities.
Bicon fulfills the stringent requirements of the European Directive 93/42/EEC for medical devices as well as strict adherence to the ISO 13485 standard.