Bicon Brevis Technique Manual
-
Upload
bicon-implant-ina -
Category
Documents
-
view
238 -
download
0
Transcript of Bicon Brevis Technique Manual
-
7/29/2019 Bicon Brevis Technique Manual
1/16
Technique Manual
Brevis Overdenture Abutment
-
7/29/2019 Bicon Brevis Technique Manual
2/16
-
7/29/2019 Bicon Brevis Technique Manual
3/16
-
7/29/2019 Bicon Brevis Technique Manual
4/16
1
Brevis Chairside Technique
Uncover implants2
A registration of the dentures occlusal
relationship prior to the uncovering ofthe implants will assure that the denture
is not inappropriately displaced by the
abutment or its housing.
occlUsal registration1
2
Use a shoulder depth gauge to facilitate the selection of the abutment height.
Use a shoUlder depth gaUge to determine abUtment height3
Uncover implants
Place guide pins into the implants todetermine their axial inclinations.
place gUide pins4
Uncover implants2
Uncover each implant using a small
crestal incision and use the healing plugremoval instrument to facilitate theremoval of the black healing plug.
Uncover implants2
-
7/29/2019 Bicon Brevis Technique Manual
5/16
2
Brevis Chairside Technique
Rotate a combination of 0 and/or 15 angled abutments to achieve parallelism prior to their being
seated. Brevis abutments are available in heights of 2.0, 4.0 and 6.0mm.
abUtments5a parallel abUtments5b
Place soft wax in the denture to act as a pressure indicator to determine the relative position of theabutments. Alternatively, the top of the abutment may be marked with a felt tip pen to indicate thelocation of the abutment on the denture.
place blUe wax6a mark abUtments6b
Place a Brevis housing on each abutmentintra-orally.
brevis hoUsing7
Liberally relieve denture to accommodatethe Brevis housings. Confirm clearancefor housings by placing the denture overthe housings.
relieve dentUre8
4.0 15 6.0 04.0 02.0 152.0 0 6.0 15
2.0 0 6.0 156.0 04.0 154.0 02.0 153.0mmP
ost
2.0mmP
ost
-
7/29/2019 Bicon Brevis Technique Manual
6/16
3
Brevis Chairside Technique
Remove the Brevis housings andplace a piece of rubber dam over theabutments to act as an apron to preventacrylic from locking onto an abutment.nsure that the ruer a rtetsthe unerut the autent raryl, y lang t ave theshuler the autent.
remove hoUsings9
Inject Vaseline under the rubberdam aprons to serve as an additionalprecaution to prevent acrylic fromlocking under the abutments during thechairside pickup of the Brevis housings.
inject vaseline10
Place the denture into the mouth andinstruct patient to lenh laterally nttn rlls t assure rer seatng the enture.
assUre proper seating12
Inject flowable acrylic around the Brevishousings and into the relieved portionsof the existing denture.
inject acrylic11
-
7/29/2019 Bicon Brevis Technique Manual
7/16
4
Use an occlusal registration jig to prevent inadvertent displacement of the denture during
the chairside technique.
Use the 15 Brevis abutment to help achieve parallelism for non-parallel implants.
The denture should NOT rock or pivot on the abutments or the housings.
Do not make a direct impression of the overdenture abutments.
Use a rubber dam and Vaseline to prevent locking of the denture beneath the undercut
of the abutments.
Place acrylic into a syringe for ease of use and greater control.
A too viscous mix of acrylic may inappropriately displace the orientation of the Brevis housing
on the abutment.
It is essential for the patient to clench bilaterally on cotton rolls to ensure proper seating of the
housings in the denture.
If the denture is too retentive, slightly relieve the inside of the o-ring lumen with a round bur.
A common cause of accelerated o-ring wear is a Brevis housing whose retentive acrylic was
polymerized while the Brevis housing was not appropriately aligned on the abutment.
3keys t uess
Note: If the denture is inadvertently locked onto the Brevis abutment, it is advisable to tap it off rather than
attempting to cut it off.
The denture may be notched to facilitate the placement of a tapping instrument. Either the denture will
be removed from the abutment or the abutment will be removed from the implant.
Brevis Chairside Technique
Polish the excess acrylic around theBrevis housings after the removal of therubber dam.
polish excess acrylic13
Radiograph of two Bicon implants andBrevis abutments. it s araunt thatverentures e letely tssuerne an nly lant retane.i the aryl s t vsus r nlylae n the enture, t ay auseslaeent the husng resultngn a salgnent the husng an
exessve wear the ruer -rngs.
radiograph14
-
7/29/2019 Bicon Brevis Technique Manual
8/16
5
Indirect Transfer Technique
Uncover implants2
A registration of the dentures occlusal
relationship prior to the uncovering ofthe implants will assure that the denture
is not inappropriately displaced by the
abutment or its housing.
occlUsal registration1
2
Use a shoulder depth gauge to facilitate the selection of the abutment height.
Use a shoUlder depth gaUge to determine abUtment height3
Uncover implantsplace gUide pins4
Uncover implants2
Uncover each implant using a small
crestal incision and use the healing plugremoval instrument to facilitate theremoval of the black healing plug.
Uncover implants2
Place guide pins into the implants todetermine their axial inclinations.
-
7/29/2019 Bicon Brevis Technique Manual
9/16
6
Indirect Transfer Technique
Rotate a combination of 0 and/or 15 angled abutments to achieve parallelism prior to their being
seated. Brevis abutments are available in heights of 2.0, 4.0 and 6.0mm.
abUtments5a parallel abUtments5b
4.0 15 6.0 04.0 02.0 152.0 0 6.0 15
2.0 0 6.0 156.0 04.0 154.0 02.0 153.0mmP
ost
2.0mmP
ost
Seat black plastic impression caps ontothe Brevis abutments.
seat impression caps6
Make impression using Brevisimpression caps. d nt ae a retressn the ttanu autents.
make impression Using caps7
Block out undercuts on the model.
block oUt UndercUts9
Insert the aluminum transfer dies into theimpression caps prior to the pouring of amaster stone model.
insert transfer dies8
-
7/29/2019 Bicon Brevis Technique Manual
10/16
7
Removing/Inserting Rubber O-Ringinto Brevis Housing
Remove rubber o-rings with a scaling
instrument or explorer.
remove hoUsings1
Insert the explorer in between the o-ring
and Brevis housing and loosen theo-ring from the housing.
insert explorer2
Using cotton pliers, squeeze the o-ringinto a figure eight and insert it intothe housing.
sqUeeze o-ring3
Place the entire o-ring into the retentivegroove within the Brevis housing.
place o-ring4
Uncover implants
If the o-ring is fully inserted, yet not securein the retention lip of the housing, use anexplorer to move the o-ring into place.
move o-ring into place5
-
7/29/2019 Bicon Brevis Technique Manual
11/16
8
The design of the Bicon abutment system is such that the hemispherical base of the abutment
does not sit ush with the neck of the implant. By design, there is a space below the abutmentpost when the abutment is fully seated. When viewing this on a radiograph, it can be seen as aradiolucency. please nte that the st any autent ay nt e fe. lterng theautent st n any anner r usng a eentng eu wll aet the rtnal ft the lng taer resultng n a tentally nn-retentve autent. The following diagramdepicts the nal seating of a Brevis abutment as well as the method for measuring each abutment.
3Measurement Guide
The height of the Brevis abutment is measured from thetop of the implant to the shoulder of the Brevis abutment.
The available heights are 2.0, 4.0 or 6.0mm. The total heightfrom the top of the implant to the top of the abutment isapproximately 5.0, 7.0 or 9.0mm, respectively.
brevs utent
Rubber O-Ring Maintenance
Each patient maintains and inserts his denture dierently. The o-ring should be changed every 6 to 24
months depending upon its wear.
The o-rings should not be soaked in a solution with high alcohol content. Some cleaning solutions
may dry out the rubber o-rings and cause them to lose retention faster than normal.
An inappropriately aligned housing will result in excessive wear of the rubber o-ring.
If excessive wear of o-ring is noted, remove and re-align the housing in denture with a suciently
owable acrylic injected onto the housing and into the relieved denture.
3keys t uess
Height
-
7/29/2019 Bicon Brevis Technique Manual
12/16
9
Chairside Brevis Technique
Place soft wax to record relative position
of abutments.
1
Record position of abutments in soft wax.
Alternatively, abutments could be markedwith a wet ink prior to inserting denturefor recording an ink transfer mark.
2
Mark position of abutments on denture.
3
Liberally relieve denture to accommodateBrevis housings and confirm positionintra-orally.
4
Place rubber dam apron over abutments. Apply Vaseline under rubber dam apron.
65
-
7/29/2019 Bicon Brevis Technique Manual
13/16
10
Snap Brevis housing onto abutments. Sufficiently apply fluid acrylic to Brevis
housing to prevent displacement ofhousing upon insertion of denture.
Place minimal acrylic into relieved denture.
9
Patient clenches bilaterally on cotton rollswhile acrylic sets.
10
Polish excess acrylic from denture andremove palatal coverage for maxillarydentures.
11
Chairside Brevis Technique
7 8
-
7/29/2019 Bicon Brevis Technique Manual
14/16
11
Place plastic alignment caps on
abutments and record impression.
1
Insert aluminum transfer dies into
alignment caps in the impression andpour the master model.
2
Snap Brevis housing on to the aluminumtransfer die and block out undercut withstone.
3
Indirect Transfer Technique
Brevis Abutment
2.0mm
3.0mm
3.0mm
4.0mm
3.0mm
6.0mm
Guida alla misurazione Example
The height of the Brevis
abutment is measured from
the top of the implant to
the shoulder of the Brevisabutment. The available
heights are 2.0, 4.0 or 6.0mm.
The total height from the top
of the implant to the top of the
abutment is approximately 5.0,
7.0 or 9.0mm, respectively.
-
7/29/2019 Bicon Brevis Technique Manual
15/16
12
Brevis Abutments*
2.0mm PostDescription Height Angle Post Part Number 0 15
2.0mm 0 Brevis Abutment 2.0mm Post 2.0mm 0 2.0mm 260-100-404
2.0mm 15 Brevis Abutment 2.0mm Post 2.0mm 15 2.0mm 260-100-405
4.0mm 0 Brevis Abutment 2.0mm Post 4.0mm 0 2.0mm 260-100-406
4.0mm 15 Brevis Abutment 2.0mm Post 4.0mm 15 2.0mm 260-100-407
6.0mm 0 Brevis Abutment 2.0mm Post 6.0mm 0 2.0mm 260-100-408
6.0mm 15 Brevis Abutment 2.0mm Post 6.0mm 15 2.0mm 260-100-409
Brevis Abutments*3.0mm Post
Description Height Angle Post Part Number 0 15
2.0mm 0 Brevis Abutment 3.0mm Post 2.0mm 0 3.0mm 260-300-434
2.0mm 15 Brevis Abutment 3.0mm Post 2.0mm 15 3.0mm 260-300-435
4.0mm 0 Brevis Abutment 3.0mm Post 4.0mm 0 3.0mm 260-300-436
4.0mm 15 Brevis Abutment 3.0mm Post 4.0mm 15 3.0mm 260-300-437
6.0mm 0 Brevis Abutment 3.0mm Post 6.0mm 0 3.0mm 260-300-438
6.0mm 15 Brevis Abutment 3.0mm Post 6.0mm 15 3.0mm 260-300-439
*Each abutment is packaged with one titanium Brevis housing.
Restorative Components
Description Part Number
Brevis Abutment Chairside Kit 260-100-212 Rubber O-Ring Brevis Housing
Brevis Impression Kit without Housing 260-100-218
Impression Cap Aluminum Transfer Die
Brevis Abutment Impression Kit 260-100-217
Rubber O-Ring Brevis Housing Impression Cap Aluminum Transfer Die
Brevis Rubber O-Rings (10) 260-100-014 Rubber O-Ring
Brevis Abutment System
-
7/29/2019 Bicon Brevis Technique Manual
16/16
Copyright 2008 Bicon LIT-255 R1208
WORLD HEADQUARTERS
501 rrway
bstn, m 02130 U
tel: 800.88.bico u 617.524.4443
ax: 800.28.bico u 617.524.0096
we: www.n.