Beth Schueler, Ph.D. Mayo Clinic Rochester AAPM Annual Meeting 2017 1

Transcript of Beth Schueler, Ph.D. Mayo Clinic...

Page 3: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

3

1

Page 4: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

Cause: Detector support electronics are visible due to

excessive backscatter Resolution:

Reduce backscattered radiation by ○ Placing lead behind the detector ○ Avoiding overexposure ○ Collimating to no more than the detector edges

4

TU-B-201-1 Artifacts: DR 1

Page 5: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

5

1

Page 6: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

6

2

Page 7: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

Cause: Detector is constructed of 4 butted sections. Interface

between sections is visible due to variation in acquisition signal as compared to gain calibration.

Signal variation may be caused by temperature fluctuations

Resolution: Repeat gain calibration Stabilize exam room and detector temperature Use detectors produced from monolithic panels

7

TU-B-201-1 Artifacts: DR 2

Page 8: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

8

3

Page 9: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

Cause: Extra material (pillow edge) in exposure field

during gain calibration Resolution:

Repeat gain calibration with a clear exposure field

9

TU-B-201-1 Artifacts: DR 3

Page 10: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

Clinical Image Artifacts Technologists should be trained how to:

Recognize common artifacts Resolve the problem themselves when

possible or When to contact a physicist or service

engineer for assistance

10

TU-B-201-1 Artifacts: DR

Page 11: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

Flat Field Artifact Check A flat field image should be acquired

After gain calibration to detect calibration errors After a wireless detector drop to look for damage As part of routine QC

Acquire image of full detector using gain calibration technique and a uniform phantom Apply minimal image processing Use level = mean pixel value, width = 10% of the level

○ Reference: AAPM Report No. 151 (2015) 11

TU-B-201-1 Artifacts: DR

Page 12: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

12

4

Page 13: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

Cause: Fine metal flecks near tube port Artifact is an image of the focal spot

– fleck acts as a reverse pinhole camera

Shift in fleck position between gain calibration and image can cause a dark-light artifact

Resolution: Remove debris and repeat gain

calibration 13

TU-B-201-1 Artifacts: DR 4

Page 14: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

14

4

Page 15: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

15

4

Page 16: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

16

5

Page 17: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

Cause: Detector was dropped Shift in layers in the GOS detector construction

causing misregistration due to impact Resolution:

Repeat gain calibration Replace GOS with CsI detector

17

TU-B-201-1 Artifacts: DR 5

Page 18: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

18

6

Page 19: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

Cause: Detector rotated and artifact remained unchanged

– localizes artifact to the detector itself Detector delamination – bubbles between TFT and

phosphor Resolution:

Repeat gain calibration

19

TU-B-201-1 Artifacts: DR 6

Page 20: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

Flat Field Artifact Check To determine location of artifact cause within

the imaging chain, repeat exposure with Detector rotated or shifted (wireless detector) Collimator rotated Different filter Different SID Detector outside of bucky or table (evaluates grid)

20

TU-B-201-1 Artifacts: DR

Page 21: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

Clinical Evaluation of Artifacts Some artifacts visible in a flat field

image are not visible in clinical images Eliminate the artifacts that can practically be

removed Review clinical or anthropomorphic phantom

images to determine if artifact is visible and clinically important

21

TU-B-201-1 Artifacts: DR

Page 22: Beth Schueler, Ph.D. Mayo Clinic Rochesteramos3.aapm.org/abstracts/pdf/127-35529-418554-126929-520366442.pdfAAPM Report No. 151 Ongoing Quality Control in DR, 2015 Includes large library

References: AAPM Report No. 151 Ongoing Quality Control in DR, 2015

○ Includes large library of artifact examples Honey, MacKenzie. “Artifacts found during quality assurance

testing of computed radiography and digital radiography detectors.” J Digit Imaging 2009; 22:383–392.

Marshall et al. “Quality control measurements for digital x-ray detectors.” Phys Med Biol 2011; 56:979–999.

Walz-Flannigan et al. “Artifacts in Digital Radiography.” AJR 198:156–161, 2012.

Willis et al. “Artifacts and Misadventures in Digital Radiography.” Appl. Radiol. 33:11–20, 2004. 22

TU-B-201-1 Artifacts: DR