UV visible spectroscopy
1
![Page 2: Best presentation of uv](https://reader033.fdocuments.in/reader033/viewer/2022061204/546ff2e1af7959ec1e8b4736/html5/thumbnails/2.jpg)
• INTRODUCTION
• PRINCIPLE
• INSTRUMENTATION&
• APPLICATIONS
Contents
2
![Page 3: Best presentation of uv](https://reader033.fdocuments.in/reader033/viewer/2022061204/546ff2e1af7959ec1e8b4736/html5/thumbnails/3.jpg)
INTRODUCTION• UV visible spectroscopy is also known as
electronic spectroscopy in which, the amount of light absorbed at each wavelength of UV and visible regions of electromagnetic spectr- um is measured.
• This absorption of electromagnetic radiations by the molecules leads to molecular
excitations.
3
![Page 4: Best presentation of uv](https://reader033.fdocuments.in/reader033/viewer/2022061204/546ff2e1af7959ec1e8b4736/html5/thumbnails/4.jpg)
Two ranges of electromagnetic radiationsin UV visible spectroscopy
ULTRA VIOLET
VISIBLE
ULTRAVIOLET REGION(10-380)n
VISIBLE REGION (380-780)nm4
![Page 5: Best presentation of uv](https://reader033.fdocuments.in/reader033/viewer/2022061204/546ff2e1af7959ec1e8b4736/html5/thumbnails/5.jpg)
UV region There are two regions in UV (1) Vacuum ultra violet region This region comprises below 200 nm in this region oxygen absorbs strongly , but range of instrument can be extended down to 150 nm , by flushing nitrogen instead of oxygen.If the instrument is evacuated,it is possible to study the whole UV region even below 200nm.the region below 200nm is therefore usually known as vacuum ultraviolet region.(2)Ordinary ultraviolet region This region comprises 200-400nm because atmosphere is transparent in this region.
5
![Page 6: Best presentation of uv](https://reader033.fdocuments.in/reader033/viewer/2022061204/546ff2e1af7959ec1e8b4736/html5/thumbnails/6.jpg)
PRINCIPLEAs the result of absorption of electromagnetic radiation molecular transitions occur.These transition occur between the electronic energy levels.As molecule absorbs energy , an electron is promoted from occupied orbital to an unoccupied orbital of greater potential energy.Generally the most probable transition is from(HOMO) to(LUMO).
6
![Page 7: Best presentation of uv](https://reader033.fdocuments.in/reader033/viewer/2022061204/546ff2e1af7959ec1e8b4736/html5/thumbnails/7.jpg)
7
![Page 8: Best presentation of uv](https://reader033.fdocuments.in/reader033/viewer/2022061204/546ff2e1af7959ec1e8b4736/html5/thumbnails/8.jpg)
FOUR TYPES OF ELECTRONIC TRANSITIONSσ→σ* for an ordinary carbon-carbon bond.(125-150)nm
1
2
3
4
π→π* for an isolated double bond. (160-190)nm
n→π* In carbonyl compounds.Aldehydes And ketones(275-295)nm
n→σ* in oxygen,nitrogen ,sulphur and Halogen compounds.(150-250)nm
8
![Page 9: Best presentation of uv](https://reader033.fdocuments.in/reader033/viewer/2022061204/546ff2e1af7959ec1e8b4736/html5/thumbnails/9.jpg)
Electronic energy levels and transitions
9
![Page 10: Best presentation of uv](https://reader033.fdocuments.in/reader033/viewer/2022061204/546ff2e1af7959ec1e8b4736/html5/thumbnails/10.jpg)
INSTRUMENTATION Two types of instruments can be used (1)Single beam spectrophotometer
A single-beam instrument uses only single beam of radiation through a single cell.The refrence cell is used to set the absorbance scale at zero for the wavelength to be studied.It is then replaced by sample cell to determine the absorbance of the sample at that wavelength.
10
![Page 11: Best presentation of uv](https://reader033.fdocuments.in/reader033/viewer/2022061204/546ff2e1af7959ec1e8b4736/html5/thumbnails/11.jpg)
Single beam instrument
11
![Page 12: Best presentation of uv](https://reader033.fdocuments.in/reader033/viewer/2022061204/546ff2e1af7959ec1e8b4736/html5/thumbnails/12.jpg)
(2)Double beam instrument A double beam instrument divides the radiation into two beams of equal intensity which are passed through two separate cells. one of the two cells contain the sample solution, while other, called the reference cell, contains either the pure solvent or a blank solution.Since the absorption by the sample is automatically corrected for absorption occurring in the solvent, the readout from the instrument is the difference between amounts of the radiations absorbed in the two cells.
12
![Page 13: Best presentation of uv](https://reader033.fdocuments.in/reader033/viewer/2022061204/546ff2e1af7959ec1e8b4736/html5/thumbnails/13.jpg)
Working of double beam instrument
13
![Page 14: Best presentation of uv](https://reader033.fdocuments.in/reader033/viewer/2022061204/546ff2e1af7959ec1e8b4736/html5/thumbnails/14.jpg)
14
Effects of solventsBlue shift (n- p*) (Hypsocromic shift)
Increasing polarity of solvent better solvation of electron pairs (n level has lower Energy)
peak shifts to the blue (more energetic)
30 nm (hydrogen bond energy)
Red shift (n- p* and p –p*) (Bathochromic shift)
Increasing polarity of solvent, then increase the attractive polarization forces between solvent and absorber, thus decreases the energy of the unexcited and excited states with the later greater
peaks shift to the red
5 nm
![Page 15: Best presentation of uv](https://reader033.fdocuments.in/reader033/viewer/2022061204/546ff2e1af7959ec1e8b4736/html5/thumbnails/15.jpg)
15
![Page 16: Best presentation of uv](https://reader033.fdocuments.in/reader033/viewer/2022061204/546ff2e1af7959ec1e8b4736/html5/thumbnails/16.jpg)
Wavelength absorbed by functional groups
16
![Page 17: Best presentation of uv](https://reader033.fdocuments.in/reader033/viewer/2022061204/546ff2e1af7959ec1e8b4736/html5/thumbnails/17.jpg)
17
EXAMPLE
![Page 18: Best presentation of uv](https://reader033.fdocuments.in/reader033/viewer/2022061204/546ff2e1af7959ec1e8b4736/html5/thumbnails/18.jpg)
Example of a Method to Determine the Absorption Spectra of an Organic Compound
Woodward’s Rules For Conjugated Carbonyl Compounds
Aldehyde: 208 nm
Extended conjugation: 30 nm
Homodyne component: 39 nm
a-Alkyl groups or ring residues: 10 nm
d-Alkyl groups or ring residues: 18 nm
Calculated: 304 nm
18
![Page 19: Best presentation of uv](https://reader033.fdocuments.in/reader033/viewer/2022061204/546ff2e1af7959ec1e8b4736/html5/thumbnails/19.jpg)
APPLICATIONS• Structure determination As different chromosphores absorbs at different wavelength, it is some times possible
to identify the absorbing chromophore in an unknown compound by comparing the wave- length of its maximum absorption with the a- bsorption characteristics of known compounds.
19
![Page 20: Best presentation of uv](https://reader033.fdocuments.in/reader033/viewer/2022061204/546ff2e1af7959ec1e8b4736/html5/thumbnails/20.jpg)
• Analytical application Since UV/visible absorption bands are usually
intense, small quantities of an observing substance can be easily detected .
• Stereo chemical studies The absorption by a conjugated system is due
to the π→π* transition. In the conjugated system the conjugation is decreased from preventing coplanarity, which affects the absorption spectrum.The UV/visible, therefore, provides useful tool for studying steriochemical details in certain molecules.