Bd Lsrii User's Guide
-
Upload
chinmayamaha -
Category
Documents
-
view
220 -
download
0
Transcript of Bd Lsrii User's Guide
-
7/30/2019 Bd Lsrii User's Guide
1/164
-
7/30/2019 Bd Lsrii User's Guide
2/164
-
7/30/2019 Bd Lsrii User's Guide
3/164
-
7/30/2019 Bd Lsrii User's Guide
4/164
-
7/30/2019 Bd Lsrii User's Guide
5/164
-
7/30/2019 Bd Lsrii User's Guide
6/164
-
7/30/2019 Bd Lsrii User's Guide
7/164
-
7/30/2019 Bd Lsrii User's Guide
8/164
-
7/30/2019 Bd Lsrii User's Guide
9/164
-
7/30/2019 Bd Lsrii User's Guide
10/164
-
7/30/2019 Bd Lsrii User's Guide
11/164
-
7/30/2019 Bd Lsrii User's Guide
12/164
-
7/30/2019 Bd Lsrii User's Guide
13/164
-
7/30/2019 Bd Lsrii User's Guide
14/164
-
7/30/2019 Bd Lsrii User's Guide
15/164
-
7/30/2019 Bd Lsrii User's Guide
16/164
-
7/30/2019 Bd Lsrii User's Guide
17/164
-
7/30/2019 Bd Lsrii User's Guide
18/164
-
7/30/2019 Bd Lsrii User's Guide
19/164
-
7/30/2019 Bd Lsrii User's Guide
20/164
-
7/30/2019 Bd Lsrii User's Guide
21/164
-
7/30/2019 Bd Lsrii User's Guide
22/164
-
7/30/2019 Bd Lsrii User's Guide
23/164
-
7/30/2019 Bd Lsrii User's Guide
24/164
-
7/30/2019 Bd Lsrii User's Guide
25/164
-
7/30/2019 Bd Lsrii User's Guide
26/164
-
7/30/2019 Bd Lsrii User's Guide
27/164
-
7/30/2019 Bd Lsrii User's Guide
28/164
-
7/30/2019 Bd Lsrii User's Guide
29/164
-
7/30/2019 Bd Lsrii User's Guide
30/164
-
7/30/2019 Bd Lsrii User's Guide
31/164
-
7/30/2019 Bd Lsrii User's Guide
32/164
-
7/30/2019 Bd Lsrii User's Guide
33/164
-
7/30/2019 Bd Lsrii User's Guide
34/164
-
7/30/2019 Bd Lsrii User's Guide
35/164
-
7/30/2019 Bd Lsrii User's Guide
36/164
-
7/30/2019 Bd Lsrii User's Guide
37/164
-
7/30/2019 Bd Lsrii User's Guide
38/164
-
7/30/2019 Bd Lsrii User's Guide
39/164
-
7/30/2019 Bd Lsrii User's Guide
40/164
-
7/30/2019 Bd Lsrii User's Guide
41/164
-
7/30/2019 Bd Lsrii User's Guide
42/164
-
7/30/2019 Bd Lsrii User's Guide
43/164
-
7/30/2019 Bd Lsrii User's Guide
44/164
-
7/30/2019 Bd Lsrii User's Guide
45/164
-
7/30/2019 Bd Lsrii User's Guide
46/164
-
7/30/2019 Bd Lsrii User's Guide
47/164
-
7/30/2019 Bd Lsrii User's Guide
48/164
-
7/30/2019 Bd Lsrii User's Guide
49/164
-
7/30/2019 Bd Lsrii User's Guide
50/164
-
7/30/2019 Bd Lsrii User's Guide
51/164
-
7/30/2019 Bd Lsrii User's Guide
52/164
-
7/30/2019 Bd Lsrii User's Guide
53/164
-
7/30/2019 Bd Lsrii User's Guide
54/164
-
7/30/2019 Bd Lsrii User's Guide
55/164
-
7/30/2019 Bd Lsrii User's Guide
56/164
-
7/30/2019 Bd Lsrii User's Guide
57/164
-
7/30/2019 Bd Lsrii User's Guide
58/164
-
7/30/2019 Bd Lsrii User's Guide
59/164
-
7/30/2019 Bd Lsrii User's Guide
60/164
-
7/30/2019 Bd Lsrii User's Guide
61/164
-
7/30/2019 Bd Lsrii User's Guide
62/164
-
7/30/2019 Bd Lsrii User's Guide
63/164
-
7/30/2019 Bd Lsrii User's Guide
64/164
-
7/30/2019 Bd Lsrii User's Guide
65/164
-
7/30/2019 Bd Lsrii User's Guide
66/164
-
7/30/2019 Bd Lsrii User's Guide
67/164
-
7/30/2019 Bd Lsrii User's Guide
68/164
-
7/30/2019 Bd Lsrii User's Guide
69/164
-
7/30/2019 Bd Lsrii User's Guide
70/164
-
7/30/2019 Bd Lsrii User's Guide
71/164
-
7/30/2019 Bd Lsrii User's Guide
72/164
-
7/30/2019 Bd Lsrii User's Guide
73/164
-
7/30/2019 Bd Lsrii User's Guide
74/164
-
7/30/2019 Bd Lsrii User's Guide
75/164
-
7/30/2019 Bd Lsrii User's Guide
76/164
-
7/30/2019 Bd Lsrii User's Guide
77/164
-
7/30/2019 Bd Lsrii User's Guide
78/164
-
7/30/2019 Bd Lsrii User's Guide
79/164
-
7/30/2019 Bd Lsrii User's Guide
80/164
-
7/30/2019 Bd Lsrii User's Guide
81/164
-
7/30/2019 Bd Lsrii User's Guide
82/164
-
7/30/2019 Bd Lsrii User's Guide
83/164
-
7/30/2019 Bd Lsrii User's Guide
84/164
-
7/30/2019 Bd Lsrii User's Guide
85/164
-
7/30/2019 Bd Lsrii User's Guide
86/164
-
7/30/2019 Bd Lsrii User's Guide
87/164
-
7/30/2019 Bd Lsrii User's Guide
88/164
-
7/30/2019 Bd Lsrii User's Guide
89/164
-
7/30/2019 Bd Lsrii User's Guide
90/164
-
7/30/2019 Bd Lsrii User's Guide
91/164
-
7/30/2019 Bd Lsrii User's Guide
92/164
-
7/30/2019 Bd Lsrii User's Guide
93/164
-
7/30/2019 Bd Lsrii User's Guide
94/164
-
7/30/2019 Bd Lsrii User's Guide
95/164
-
7/30/2019 Bd Lsrii User's Guide
96/164
% t
r a n s m
i s s i o n
longpass
-
7/30/2019 Bd Lsrii User's Guide
97/164
Appendix A: Technical Overview 97
Shortpass Filters
An SP filter has the opposite properties of a longpass filter. An SP filter passeslight with a shorter wavelength than the filter rating.
Bandpass Filters
A BP filter transmits a relatively narrow range or band of light. Bandpass filtersare typically designated by two numbers. The first number indicates the centerwavelength and the second refers to the width of the band of light that is passed.For example, a 500/50 BP filter transmits light that is centered at 500 nm and has
wavelength (nm)
wavelength (nm)
% t
r a n s m
i s s i o n
shortpass
a total bandwidth of 50 nm. Therefore, this filter transmits light between 475and 525 nm.
Figure A-5 Bandpass filter
bandpass
-
7/30/2019 Bd Lsrii User's Guide
98/164
98 BD LSR II Users Guide
BP and DF filters have the same general functionthey transmit a relativelynarrow band of light. The principal difference between them is theirconstruction. DF filters have more cavities or layers of optical coatings, resultingin a steeper transmission curve than the curve for a BP filter. This steep slopemeans that a DF filter is better at blocking light outside the rated bandwidth of the filter.
wavelength (nm)
% t r
a n s m
i s s i o n
t r a n s m
i s s i o n
BP 500/50 filter
DF 500/50 filter
AF 500/50 filter
-
7/30/2019 Bd Lsrii User's Guide
99/164
Appendix A: Technical Overview 99
Dichroic Mirrors
Dichroic filters that are used to direct different color light signals to differentdetectors are called dichroic mirrors or beam splitters.
Although dichroic mirrors have the properties of LP or SP optical filters, you cannot necessarily use any type of LP or SP filter as a beam splitter. A beam splittermust have a surface coating that reflects certain wavelengths, but many LP or SPfilters are absorbance filters that do not have any specific reflectivecharacteristics. Also, optical filters and beam splitters are rated at a specific angleof incidence. When used in front of the fluorescence detectors, they areperpendicular to the incident light, and when used as a beam splitter, they areplaced at an angle relative to the light source. Their optical properties aretherefore designed for that angle of incidence.
Compensation Theory
Fluorochromes emit light over a range of wavelengths ( Figure A-3 on page 95 ).Optical filters are used to limit the range of frequencies measured by a givendetector. However, when two or more fluorochromes are used, the overlap inwavelength ranges often makes it impossible for optical filters to isolate lightfrom a given fluorochrome. As a result, light emitted from one fluorochrome
wavelength (nm)
% t
appears in a detector intended for another ( Figure A-6 ). This is referred to asspillover. Spillover can be corrected mathematically by using a method calledcompensation.
Figure A-6 Spillover from the FITC fluorochrome to the PE detector
-
7/30/2019 Bd Lsrii User's Guide
100/164
100 BD LSR II Users Guide
For example, FITC emission appears primarily in the FITC detector, but some of its fluorescence spills over into the PE detector. The spillover is corrected orcompensated forhence the term fluorescence compensation.
Figure A-6 shows that some of the FITC emission appears in the PE detector.This can be seen in a dot plot of FITC vs PE.
Figure A-7 Theoretical display of FITC vs PE without compensation
n o r m a l
i z e d
i n t e n s
i t y
PE
FITC
unstainedparticles
FITC-positive
population
This FITC spillover in the PE detector is to be corrected as indicated by the arrowin Figure A-7 . Using the Compensation tab of the Cytometer window inBD FACSDiva software, you can adjust the PE-%FITC spectral overlap value.Compensation is optimal when the positive and negative FITC populations have
the same means or medians in the PE parameter statistics.Figure A-8 FITC spillover optimally compensated out of the PE parameter
-
7/30/2019 Bd Lsrii User's Guide
101/164
Appendix A: Technical Overview 101
Once fluorescence compensation has been set for any sample, the compensationsetting remains valid for a subsequent dim or bright sample, because
compensation subtracts a percentage of the fluorescence intensity. Figure A-9 illustrates this principle. Although the signals differ in intensity, the percentage of the FITC spillover into the PE detector remains constant.
PE
FITC
matching means
unstainedparticles
FITC-positivepopulation
Figure A-9 Two FITC signals of different intensity
FITC PE
different intensity FITC signals
y
-
7/30/2019 Bd Lsrii User's Guide
102/164
102 BD LSR II Users Guide
Electronics
As cells or other particles pass through a focused laser beam, they scatter thelaser light and can emit fluorescence. Because the laser beam is focused on a small
spot and particles move rapidly through the flow cell, the scatter or fluorescenceemission has a very brief durationonly a few microseconds. This brief flash of light is converted into an electrical signal by the detectors. The electrical signal iscalled a pulse.
1 A pulse begins when a particle enters the laser beam. At this point, both thebeam intensity and signal intensity are low.
2 The pulse reaches a maximum intensity or height when the particle reachesthe middle of the beam, where the beam and signal intensity are thebrightest. The peak intensity, or height of the pulse, is measured at thispoint.
3 As the particle leaves the beam, the pulse trails off below the threshold.
same proportion or percentage of spectral overlap in PE channel n o
r m a
l i z e
d i n t e n s i t y
Figure A-10 Anatomy of a pulse
it y
s i g n a l
i n t e n s
i t y
time
-
7/30/2019 Bd Lsrii User's Guide
103/164
Appendix A: Technical Overview 103
time
s i g n a
l i n t e n s
i ty
s i g n a l
i n t e n s i t y
time
Pulse Measurements
The pulse processors measure pulses by three characteristics: height, area, andwidth.
Figure A-11 Pulse measurements
-
7/30/2019 Bd Lsrii User's Guide
104/164
104 BD LSR II Users Guide
Pulse height is the maximum digitized intensity measured for the pulse.
Pulse area is an integration of the digitized measures over time.
Pulse width calculates:
height
threshold
baseline
0 volts
window gate:width
voltage
time
area
areaheight x 64,000
Digital Electronics
BD LSR II flow cytometer electronics digitize the signal intensity produced by adetector. The digitized data is stored in memory and further processed by theelectronics to calculate
Pulse height, area, and width
C ti
-
7/30/2019 Bd Lsrii User's Guide
105/164
Appendix A: Technical Overview 105
Compensation
Parameter ratios
These results are transferred to your workstation computer for further processingby BD FACSDiva software. For more information about digital theory, refer toDigital Theory in the BD FACSDiva Software Reference Manual .
Threshold
The threshold is the level at which the system starts to measure signal pulses.A threshold is defined for a specific detector signal. The system continuouslysamples the digitized signal data and calculates pulse area, height, and width forall channels based on the time interval during which the threshold is exceeded.
Thresholds can also be set for more than one parameter, and pulse measures arebased on either of the following:
Intervals during which ALL signals exceed their threshold value
Intervals during which ANY signal exceeds its threshold value
Laser Controls
Controls in the Laser tab of the Cytometer window are used to manually set the(laser) delay, area scaling, and window extension values.
These parameters are set by BD Biosciences service personnel during theBD LSR II flow cytometer installation and performance check and are updatedeach time you run a performance check.
-
7/30/2019 Bd Lsrii User's Guide
106/164
106 BD LSR II Users Guide
If needed, see Optimizing Laser Delay on page 157 for instructions on manuallyadjusting laser delay settings. Do not otherwise change the settings in the Laser
tab unless instructed to do so by BD Biosciences. Changing the settings affectsyour data.
Appendix B
-
7/30/2019 Bd Lsrii User's Guide
107/164
107
Troubleshooting
The tips in this section are designed to help you troubleshoot your experiments.You can find additional troubleshooting information in the BD FACSDivaSoftware Reference Manual .
If additional assistance is required, contact your local BD Biosciences technical
support representative. See Technical Assistance on page xv.
Cytometer Troubleshooting
Observation Possible Causes Recommended Solutions
Droplet containmentvacuum not functioning
Worn O-ring in retainer Replace the O-ring. See Changingthe Sample Tube O-Ring onpage 90 .
O t l i t t d i 1 L th t i ( Fig 4 4
-
7/30/2019 Bd Lsrii User's Guide
108/164
108 BD LSR II Users Guide
Outer sleeve is not seated inthe retainer
1 Loosen the retainer ( Figure 4-4on page 88 ).
2 Push the outer sleeve up intothe retainer until seated.
3 Tighten the retainer.
Outer sleeve is not on thesample injection tube
Replace the outer sleeve.
1 Loosen the retainer.
2 Slide the outer sleeve over thesample injection tube until it isseated.
3 Tighten the retainer.
Waste line is pinched,preventing proper aspiration
Check the waste line.
Waste tank is full Empty the waste tank.
Sample tube not fittingon SIP Sample tube other thanBD Falcon tubes used Use BD Falcon 12 x 75-mmsample tubes. See Equipment onpage 120.
Worn Bal seal Replace the Bal seal. See Changingthe Bal Seal on page 88.
Rapid sample aspiration Support arm is to the side Place the support arm under thesample tube.
Droplet containment moduleis failing
Call your service representative.
No events in acquisitiondisplay and RUN buttonis green.
Threshold is not set to thecorrect parameter (usuallyFSC)
Set the threshold to the correctparameter for your application.
Threshold level is too high Lower the threshold level.
Cytometer Troubleshooting (continued)
Observation Possible Causes Recommended Solutions
-
7/30/2019 Bd Lsrii User's Guide
109/164
Appendix B: Troubleshooting 109
PMT voltage for thresholdparameter is set too low
Set the PMT voltage higher for thethreshold parameter.
Gating issue Refer to the BD FACSDivaSoftware Reference Manual forinformation on setting gates.
Air in the sheath filter Purge the filter. See Removing AirBubbles on page 49.
No sample in the tube Add sample to the tube or install anew sample tube.
Sample is not mixed properly Mix the sample to suspend cells.
Waste tank is full Empty the waste tank.
PMT voltages set too low ortoo high for displayparameter
Reset the PMT voltages.
Too few events are displayed Increase the number of events todisplay.
Sample injection tube isclogged
Remove the sample tube to allowbackflushing.
If the event rate is still erratic,clean the sample injection tube.See Daily Cleaning and Shutdownon page 78.
Bal seal is worn Replace the Bal seal. See Changingthe Bal Seal on page 88.
No events in acquisitiondisplay and RUN buttonis green (continued)
Laser is not warmed up Wait the recommended amount of time for the laser to warm up.
30 min for the 488-nm (blue)
30 min for the 355-nm (UV)
Cytometer Troubleshooting (continued)
Observation Possible Causes Recommended Solutions
-
7/30/2019 Bd Lsrii User's Guide
110/164
110 BD LSR II Users Guide
15 min for the 405-nm (violet)
20 min for the 633-nm (red)
Laser delay is set incorrectly Adjust the laser delay settings.
See Setting Laser Delay onpage 155 .
Laser is not functioning Verify the malfunction bychanging the threshold to analternative laser while running theappropriate sample. If unsuccessful, contactBD Biosciences.
No events in acquisitiondisplay and RUN buttonis orange
RUN is not activated Press the RUN button.
Sample tube is not installedor is not properly seated
Install the sample tube correctlyon the SIP.
Sample tube is cracked Replace the sample tube
Cytometer Troubleshooting (continued)
Observation Possible Causes Recommended Solutions
-
7/30/2019 Bd Lsrii User's Guide
111/164
Appendix B: Troubleshooting 111
Sample tube is cracked Replace the sample tube.
Sheath container is not
pressurized
Ensure that the sheath
container lid and all connectorsare securely seated.
Inspect the O-ring and replaceit if necessary. See Changing theSample Tube O-Ring onpage 90.
Bal seal is worn Replace the Bal seal. See Changing
the Bal Seal on page 88.Air leak at sheath container Ensure that the sheath container
lid and all connectors are securelyseated.
No events in acquisitiondisplay and RUN buttonis orange (continued)
Sheath container is empty Fill the sheath container.
Air in sheath filter Purge the filter. See Removing Air
Bubbles on page 49.
No fluorescent signal Incorrect fluorochromeassignment
Make sure the cytometerconfiguration in the softwarematches the optical filters in thecytometer.
Wrong filter is installed Make sure the appropriate filter is
Cytometer Troubleshooting (continued)
Observation Possible Causes Recommended Solutions
-
7/30/2019 Bd Lsrii User's Guide
112/164
112 BD LSR II Users Guide
Wrong filter is installed Make sure the appropriate filter isinstalled for each fluorochrome.See Changing Optical Filters orMirrors on page 43.
Laser is not functioning Verify the laser malfunction bychanging the threshold to analternative laser while running theappropriate sample. If unsuccessful, contactBD Biosciences.
High event rate Air bubble in the sheath filteror flow cell
Remove the air bubble. SeeRemoving Air Bubbles onpage 49.
Threshold level is too low Increase the threshold level. Referto the BD FACSDiva SoftwareReference Manual forinstructions.
PMT voltage for thethreshold parameter set toohigh
Set the PMT voltage lower for thethreshold parameter. Refer to theBD FACSDiva SoftwareReference Manual forinstructions.
Sample is too concentrated Dilute the sample.
Sample flow rate is set on HI Set the sample flow rate to MEDor LO.
Low event rate Threshold level is too high Lower the threshold level. Refer tothe BD FACSDiva SoftwareReference Manual forinstructions.
PMT voltage for the Set the PMT voltage higher for the
Cytometer Troubleshooting (continued)
Observation Possible Causes Recommended Solutions
-
7/30/2019 Bd Lsrii User's Guide
113/164
Appendix B: Troubleshooting 113
PMT voltage for thethreshold parameter is set toolow
Set the PMT voltage higher for thethreshold parameter. Refer to theBD FACSDiva SoftwareReference Manual forinstructions.
Sample is not adequatelymixed
Mix the sample to suspend thecells.
Sample is too diluted Concentrate the sample. If theflow rate setting is not critical tothe application, set the flow rateswitch to MED or HI.
Sample injection tube isclogged
Remove the sample tube to allowbackflushing.
If the event rate is still erratic,clean the sample injection tube.See Daily Cleaning and Shutdown
on page 78.Erratic event rate Sample tube is cracked Replace the sample tube.
Bal seal is worn Replace the Bal seal. See Changingthe Bal Seal on page 88.
Sample injection tube isclogged
Remove the sample tube to allowbackflushing.
If the event rate is still erratic,clean the sample injection tube.See Daily Cleaning and Shutdownon page 78.
Erratic event rate(continued)
Sample injection tube isclogged
Remove the sample tube to allowbackflushing.
If the event rate is still erratic,clean the sample injection tube.See Daily Cleaning and Shutdown
Cytometer Troubleshooting (continued)
Observation Possible Causes Recommended Solutions
http://-/?-http://-/?- -
7/30/2019 Bd Lsrii User's Guide
114/164
114 BD LSR II Users Guide
on page 78.
Contaminated sample Prepare the specimen again.Ensure that the tube is clean.
Sheath filter is dirty Replace the fil ter. See Changingthe Shea th Filter on page 85.
Distorted scatterparameters
Cytometer settings areimproperly adjusted
Optimize the scatter parameters.Refer to the BD FACSDivaSoftware Reference Manual forinstructions.
Air bubble in sheath filter orflow cell
Purge the air from the filter. SeeRemoving Air Bubbles onpage 49.
Flow cell is dirty Perform the system flushprocedure. See System Flush onpage 80.
Air leak at sheath container Ensure that the sheath containerlid is tight and all connectors aresecure.
Hypertonic buffers or fixative Replace the buffers and fixative.
Excessive amount of debris in display
Threshold level is too low Increase the threshold level.
Sheath filter is dirty Replace the filter. See Changingthe Sheath Filter on page 85.
Flow cell is dirty Flush the system. See System Flush
Cytometer Troubleshooting (continued)
Observation Possible Causes Recommended Solutions
http://-/?-http://-/?-http://-/?-http://-/?- -
7/30/2019 Bd Lsrii User's Guide
115/164
Appendix B: Troubleshooting 115
on page 80.
Dead cells or debris in sample Examine the sample under amicroscope.
Sample is contaminated Re-stain the sample, ensure tube isclean.
Stock sheath fluid iscontaminated
Rinse the sheath container with DIwater, then fill with sheath fluidfrom another (or new lot) bulk
container.High CV Air bubble in sheath filter or
flow cellPurge the filter. See Removing AirBubbles on page 49.
Sample flow rate is set toohigh
Set the sample flow rate lower.
Air leak at sheath container Ensure that the sheath containerlid is tight and all connectors aresecure.
Flow cell is dirty Flush the system. See System Flushon page 80.
Poor sample preparation Repeat sample preparation.
Sample not diluted in samefluid as sheath fluid
Dilute the sample in the same fluidas you are using for sheath.
Poor QC results Air bubble or debris in flowcell
Prime the fluidics system. SeePriming the Fluidics on page 50.
Old or contaminated QCparticles
Make new QC samples andperform the quality controlprocedure again.
Cytometer Troubleshooting (continued)
Observation Possible Causes Recommended Solutions
-
7/30/2019 Bd Lsrii User's Guide
116/164
116 BD LSR II Users Guide
Sample not diluted in same
fluid as sheath fluid
Dilute the sample in the same fluid
as you are using for sheath.Laser not warmed up Wait the recommended amount of
time for the laser to warm up.
30 min for the 488-nm (blue)
30 min for the 355-nm (UV)
15 min for the 405-nm (violet)
20 min for the 633-nm (red)
Laser not functioning Contact BD Biosciences.
Optical alignment problem Contact BD Biosciences.
Appendix C
Supplies and Consumables
-
7/30/2019 Bd Lsrii User's Guide
117/164
117
Supplies and Consumables
To order spare parts and consumables, such as bulk fluids, from BD Biosciences:
Within the US, call (877) 232-8995.
Outside the US, contact your local BD Biosciences customer support
representative.
Worldwide contact information can be found at bdbiosciences.com.
Use the following part numbers to order supplies for your BD LSR II system:
QC Particles on page 118
QC Cytometer Setup and Tracking Particles on page 118
Reagents on page 119
Equipment on page 120
QC Particles
Particle Laser Supplier Catalog No.
SPHERORainbowCalibration Particles(8 peak)
SPHERO UltraR i b
all
all
BD Biosciences
Spherotech, Inc.
559123
URFP-30 2
-
7/30/2019 Bd Lsrii User's Guide
118/164
118 BD LSR II Users Guide
QC Cytometer Setup and Tracking Particles
RainbowFluorescent Particles
(single peak)
30-2
DNA QC Particles kit blue 488 nm BD Biosciences 349523
Particle Laser Supplier
BD Cytometer Setupand Tracking beads
UV (355 nm and 375 nm)
violet (405 nm and 407 nm)
blue (488 nm)
green (532 nm)
red (633 nm and 645 nm)
BD Biosciences(contact BDBiosciences for moreinformation)
Reagents
Reagent Supplier Catalog No.
BD FACSFlow sheath fluid BD Biosciences 340398(US and Latin
America)
342003(Europe)
-
7/30/2019 Bd Lsrii User's Guide
119/164
Appendix C: Supplies and Consumables 119
Monoclonal antibodies BD Biosciences a
a. Refer to the BD Biosciences Product Catalog or the BD Biosciences website (bdbiosciences.com).
BD FACS lysing solution BD Biosciences 349202
BD FACSRinse solution BD Biosciences 340346
BD FACSClean solution BD Biosciences 340345
Dyes and fluorochromes BD Biosciences
Molecular Probes
Sigma
Chlorine bleach (5% sodiumhypochlorite)
Clorox or other majorsupplier (to ensure thatthe bleach is at thecorrect concentrationand free of particulatematter)
Equipment
Equipment Item Supplier Catalog No.
Bal seal BD Biosciences 343509
O-ring, sample tube 343615
Sheath filter assembly 344678
BD Falcon TM polystyrene test 352052
-
7/30/2019 Bd Lsrii User's Guide
120/164
120 BD LSR II Users Guide
p y ytubes, 12 x 75-mm 352054
352058
Appendix D
Standard Base Configuration
-
7/30/2019 Bd Lsrii User's Guide
121/164
121
g
The standard base configuration for a BD LSR II cytometer supports detectors,filters, and mirrors for one to four lasers. This appendix describes how to set upthe cytometer optics using standard default configuration components.
4-Blue 2-Violet 2-355 UV 2-Red Configuration on page 122
Additional Optics on page 128
The BD LSR II cytometer can also be ordered with one of several optionalconfigurations, which are described in Appendix E .
4-Blue 2-Violet 2-355 UV 2-Red Configuration
The standard configuration supports a blue octagon, and violet, UV, and redtrigons. Table D-1 shows the detectors, filters, and mirrors used in the standarddefault configuration, and recommended fluorochromes for each detector. Theword blank indicates that a blank optical holder should be used instead of amirror or filter. A dash () indicates that no slot exists for a mirror in that PMTposition.
Table D-1 Default filters and fluorochromes
-
7/30/2019 Bd Lsrii User's Guide
122/164
122 BD LSR II Users Guide
Detector Array(Laser)
PMT(Detector)
LongpassDichroicMirror
BandpassFilter
Fluorochrome orScatter Parameter
blue octagon(488-nm laser)
A 735 LP 780/60 PE-Cy7
B 685 LP 695/40 PerCP-Cy5.5
C 550 LP 575/26 PE, PID 505 LP 530/30 FITC, GFP
E blank 488/10 SSC
F blank blank none
G blank blank none
H blank none
violet trigon(405-nm laser)
A 505 LP 525/50 AmCyan
B blank 440/40 Pacific Blue
C blank none
UV trigon(355-nm laser)
A 505 LP 530/30 Indo-1 (Blue)
B blank 450/50 Indo-1 (Violet), DAPI
C blank none
red trigon(633-nm laser)
A 735 LP 780/60 APC-Cy7
B blank 660/20 APC
C blank none
Table D-1 Default filters and fluorochromes (continued)
Detector Array(Laser)
PMT(Detector)
LongpassDichroicMirror
BandpassFilter
Fluorochrome orScatter Parameter
-
7/30/2019 Bd Lsrii User's Guide
123/164
Appendix D: Standard Base Configuration 123
Base ConfigurationFigure D-1 shows a default base cytometer configuration.
Figure D-1 Base configuration
-
7/30/2019 Bd Lsrii User's Guide
124/164
124 BD LSR II Users Guide
Octagon and Trigon MapsThis section shows how to install mirrors and filters in your octagon and trigonsfor the standard default configuration.
If a slot is filled with a filter or mirror, an identifying number appears in thatposition on the configuration map. If a slot is filled with a blank optical holder,
that position on the configuration map is unlabeled.
Figure D-2 Standard default configuration: blue octagon
F I TC
6 9 5 / 4 0
P e r C P -C y 5.5
488-nm blue laser
-
7/30/2019 Bd Lsrii User's Guide
125/164
Appendix D: Standard Base Configuration 125
4 8 8 /
1 0
5 3 0 / 3 0
5 0 5 L P
T C
S S C
6 8 5 LP
7 8 0 / 6 0
7 3 5 L P
5 7 5 / 2 6
5 5 0 L P
P E - C y 7 P E
Figure D-3 Standard default configuration: red and violet trigons
6 6 0 / 2 0
AP C
4 4 0 / 4 0
P ac i f i c Bl ue
633-nm red laser 405-nm violet laser
-
7/30/2019 Bd Lsrii User's Guide
126/164
126 BD LSR II Users Guide
7 8 0 / 6 0
7 3 5 L P
A P C - C y 7
5 2 5 / 5 0
5 0 5 L P
A m C y a n
Figure D-4 Standard default configuration: UV trigon
4 5 0 / 5 0
D AP I
355-nm UV laser
-
7/30/2019 Bd Lsrii User's Guide
127/164
Appendix D: Standard Base Configuration 127
5 3 0 / 3 0
5 0 5 L P
I n d o - 1 ( B l u e )
Additional OpticsThis section describes some common custom filter and mirror configurations.Table D-2 shows the detector arrays, mirrors, and filters used in the customconfigurations, and recommended fluorochromes for each detector. The mirrorsand filters used in these custom configurations are contained in the BD LSR IIcytometer spares kit.
Table D-2 Additional filters and mirrors
( ) l l h
-
7/30/2019 Bd Lsrii User's Guide
128/164
128 BD LSR II Users Guide
Maps on the following pages show how to install mirrors and filters in youroctagon and trigons for common custom configurations.
Detector Array (Laser) Mirror Filter Fluorochrome
blue octagonstandard 488-nm blue laser
600 LP 610/20 PE-Texas Red
635 LP 670/14 PerCPBD Cy-Chrome reagent
585/42 DsRed
violet trigonoptional 405-nm violet laser
none none none
UV trigonoptional 355-nm UV laser
450 LP Indo-1 (Blue)
405/20 Indo-1 (Violet)
red trigonoptional 633-nm red laser
none none none
PE-Texas Red
To use PE-Texas Red, replace the mirror and filter for the B PMT of the blueoctagon as shown below.
488-nm blue laser
-
7/30/2019 Bd Lsrii User's Guide
129/164
Appendix D: Standard Base Configuration 129
4 8 8 /
1 0
5 3 0 / 3 0 5 0 5 L P
F I T C
S S C
6 1 0 / 2 0
6 0 0 LP
7 8 0 / 6 0
7 3 5 L P
5 7 5 / 2 6
5 5 0 L P
P E - C y 7 P E
P E -T e x as Re d
Indo-1
If you have a violet laser, replace the mirror for the A PMT of the UV trigon asshown on the left below.
If you do not have a violet laser, replace both the mirror for the A PMT and thefilter for the B PMT of the UV trigon as shown on the right below.
355-nm UV laser(with violet laser)
355-nm UV laser(no violet laser)
-
7/30/2019 Bd Lsrii User's Guide
130/164
130 BD LSR II Users Guide
4 5 0 / 5 0
5 3 0 / 3 0
4 5 0 L P
I n d o - 1 ( B l u e )
Ind o -1 ( V i o l e t )
4 0 5 / 2 0
5 3 0 / 3 0
4 5 0 L P
I n d o - 1 ( B l u e )
Ind o -1 ( V i o l e t )
DsRed
To use DsRed, replace the filter for the C PMT of the blue octagon as shownbelow.
488-nm blue laser
-
7/30/2019 Bd Lsrii User's Guide
131/164
Appendix D: Standard Base Configuration 131
4 8 8 /
1 0
5 3 0 / 3 0 5 0 5 L P
F I T C
S S C
6 9 5 / 4 0
6 8 5 LP
7 8 0 / 6 0
7 3 5 L P
5 8 5 / 4 2
5 5 0 L P
P E - C y 7 D s R e d
P e r C P -C y 5.5
PerCP or BD Cy-Chrome Reagent
To use PerCP or BD Cy-Chrome reagent, replace the mirror and filter for the BPMT of the blue octagon as shown below.
488-nm blue laser
-
7/30/2019 Bd Lsrii User's Guide
132/164
132 BD LSR II Users Guide
4 8 8 /
1 0
5 3 0 / 3 0 5 0 5 L P
F I T C
S S C
6 7 0 / 1 4
6 3 5 LP
7 8 0 / 6 0
7 3 5 L P
5 7 5 / 2 6
5 5 0 L P
P E - C y 7 P E
P e r C P
Appendix E
Special Order Configurations
-
7/30/2019 Bd Lsrii User's Guide
133/164
133
The BD LSR II cytometer can be ordered with, or upgraded to, one of severallaser and detector array options. You must use the Cytometer Setup and Trackingapplication to set up new configurations. See the Cytometer Setup and Tracking Applications Guide for more information.
This appendix contains the following information:
Common Special Order Configurations on page 134
Special Order Configuration Trigon and Octagon Maps on page 148
Common Special Order Configurations
The following are commonly used configurations.
6-Blue 0-Violet 0-UV 3-Red Configuration on page 135
6-Blue 2-Violet 0-UV 3-Red Configuration on page 136
6-Blue 0-Violet 2-UV 3-Red Configuration on page 137
6-Blue 2-Violet 2-UV 3-Red Configuration on page 139
-
7/30/2019 Bd Lsrii User's Guide
134/164
134 BD LSR II Users Guide
6-Blue 6-Violet 0-UV 3-Red Configuration on page 140
6-Blue 6-Violet 0-UV 4-Red Configuration on page 142
6-Blue 6-Violet 2-UV 3-Red Configuration on page 144
6-Blue 6-Violet 2-UV 4-Red Configuration on page 146
6-Blue 0-Violet 0-UV 3-Red Configuration
6-Blue 0-Violet 0-UV 3-Red supports a blue octagon and a red trigon.Table E-1 shows the detectors, filters, and mirrors used in the defaultconfiguration. The word blank indicates that a blank optical holder shouldbe used instead of a mirror or filter. A dash () indicates that no slot exists fora mirror in that PMT position.
The 6-Blue 0-Violet 0-UV 3-Red maps are:
6-Color Blue Octagon Default Configuration Map on page 149
-
7/30/2019 Bd Lsrii User's Guide
135/164
Appendix E: Special Order Configurations 135
3-Color Red Trigon Default Configuration Map on page 153
Table E-1 6-Blue 0-Violet 0-UV 3-Red default mirror and filter configuration
Detector Array(Laser)
PMT(Detector)
Longpass
DichroicMirror
BandpassFilter
Fluorochrome orScatter Parameter
blue octagon(488-nm laser)
A 755 LP 780/60 PE-Cy7
B 685 LP 695/40 PerCP-Cy5.5
C 655 LP 660/20 PE-Cy5
D 600 LP 610/20 PE-Texas RedE 550 LP 575/26 PE
F 505 LP 530/30 FITC, Alexa Fluor 488
G blank 488/10 SSC
H blank none
red trigon(633-nm laser)
A 755 LP 780/60 APC-Cy7
B 710 LP 730/45 Alexa Fluor 700
C 660/20 APC
6-Blue 2-Violet 0-UV 3-Red Configuration
6-Blue 2-Violet 0-UV 3-Red supports a blue octagon, and violet and red trigons.Table E-2 shows the detectors, filters, and mirrors used in the defaultconfiguration. The word blank indicates that a blank optical holder should beused instead of a mirror or filter. A dash () indicates that no slot exists for amirror in that PMT position.
The 6-Blue 2-Violet 0-UV 3-Red maps are:
6-Color Blue Octagon Default Configuration Map on page 149
-
7/30/2019 Bd Lsrii User's Guide
136/164
136 BD LSR II Users Guide
2-Color Violet Trigon Default Configuration Map on page 150
3-Color Red Trigon Default Configuration Map on page 153
Table E-2 6-Blue 2-Violet 0-UV 3-Red default mirror and filter configuration
Detector Array(Laser)
PMT(Detector)
LongpassDichroicMirror
BandpassFilter
Fluorochrome orScatter Parameter
blue octagon(488-nm laser)
A 755 LP 780/60 PE-Cy7
B 685 LP 695/40 PerCP-Cy5.5
C 655 LP 660/20 PE-Cy5
D 600 LP 610/20 PE-Texas Red
E 550 LP 575/26 PE
F 505 LP 530/30 FITC, Alexa Fluor 488
G blank 488/10 SSC
H blank none
violet trigon(405-nm laser)
A 505 LP 525/50 AmCyan
B blank 450/50 Pacific Blue
C blank none
red trigon(633-nm laser)
A 755 LP 780/60 APC-Cy7
B 710 LP 730/45 Al Fl 700
Table E-2 6-Blue 2-Violet 0-UV 3-Red default mirror and filter configuration (continued)
Detector Array(Laser)
PMT(Detector)
LongpassDichroicMirror
BandpassFilter
Fluorochrome orScatter Parameter
-
7/30/2019 Bd Lsrii User's Guide
137/164
Appendix E: Special Order Configurations 137
6-Blue 0-Violet 2-UV 3-Red Configuration
6-Blue 0-Violet 2-UV 3-Red supports a blue octagon, and UV and red trigons.Table E-3 on page 138 shows the detectors, filters, and mirrors used in thedefault configuration. The word blank indicates that a blank optical holdershould be used instead of a mirror or filter. A dash () indicates that no slotexists for a mirror in that PMT position.
The 6-Blue 0-Violet 2-UV 3-Red maps are:
6-Color Blue Octagon Default Configuration Map on page 149
2-Color UV Trigon Default Configuration Map on page 152
3-Color Red Trigon Default Configuration Map on page 153
B 710 LP 730/45 Alexa Fluor 700
C 660/20 APC
Table E-3 6-Blue 0-Violet 2-UV 3-Red default mirror and filter configuration
Detector Array(Laser)
PMT(Detector)
Longpass
DichroicMirror
BandpassFilter
Fluorochrome orScatter Parameter
blue octagon(488-nm laser)
A 755 LP 780/60 PE-Cy7
B 685 LP 695/40 PerCP-Cy5.5
C 655 LP 660/20 PE-Cy5
D 600 LP 610/20 PE T R d
-
7/30/2019 Bd Lsrii User's Guide
138/164
138 BD LSR II Users Guide
D 600 LP 610/20 PE-Texas RedE 550 LP 575/26 PE
F 505 LP 530/30 FITC, Alexa Fluor 488
G blank 488/10 SSC
H blank none
UV trigon(355-nm laser)
A 505 LP 530/30 Indo-1 (Blue)
B blank 450/50 Indo-1 (Violet), DAPI
C blank none
red trigon(633-nm laser)
A 755 LP 780/60 APC-Cy7
B 710 LP 730/45 Alexa Fluor 700
C 660/20 APC
6-Blue 2-Violet 2-UV 3-Red Configuration
6-Blue 2-Violet 2-UV 3-Red supports a blue octagon, and violet, UV, and redtrigons. Table E-4 shows the detectors, filters, and mirrors used in the defaultconfiguration. The word blank indicates that a blank optical holder should beused instead of a mirror or filter. A dash () indicates that no slot exists for amirror in that PMT position.
The 6-Blue 2-Violet 2-UV 3-Red maps are:
6-Color Blue Octagon Default Configuration Map on page 149
2 Color Violet Trigon Default Configuration Map on page 150
-
7/30/2019 Bd Lsrii User's Guide
139/164
Appendix E: Special Order Configurations 139
2-Color Violet Trigon Default Configuration Map on page 150
2-Color UV Trigon Default Configuration Map on page 152
3-Color Red Trigon Default Configuration Map on page 153
Table E-4 6-Blue 2-Violet 2-UV 3-Red default mirror and filter configuration
Detector Array(Laser)
PMT(Detector)
LongpassDichroicMirror
BandpassFilter
Fluorochrome orScatter Parameter
blue octagon
(488-nm laser)
A 755 LP 780/60 PE-Cy7
B 685 LP 695/40 PerCP-Cy5.5
C 655 LP 660/20 PE-Cy5
D 600 LP 610/20 PE-Texas Red
E 550 LP 575/26 PE
F 505 LP 530/30 FITC, Alexa Fluor 488
G blank 488/10 SSC
H blank none
violet trigon(405-nm laser)
A 505 LP 525/50 AmCyan
B blank 450/50 Pacific Blue
C blank none
UV trigon(355-nm laser)
A 505 LP 530/30 Indo-1 (Blue)
B blank 450/50 Indo-1 (Violet) DAPI
Table E-4 6-Blue 2-Violet 2-UV 3-Red default mirror and filter configuration (continued)
Detector Array(Laser)
PMT(Detector)
LongpassDichroicMirror
BandpassFilter
Fluorochrome orScatter Parameter
-
7/30/2019 Bd Lsrii User's Guide
140/164
140 BD LSR II Users Guide
6-Blue 6-Violet 0-UV 3-Red Configuration
6-Blue 6-Violet 0-UV 3-Red supports blue violet octagons, and a red trigon.Table E-5 on page 141 shows the detectors, filters, and mirrors used in thedefault configuration. The word blank indicates that a blank optical holder
should be used instead of a mirror or filter. A dash () indicates that no slotexists for a mirror in that PMT position.
The 6-Blue 6-Violet 0-UV 3-Red maps are:
6-Color Blue Octagon Default Configuration Map on page 149
6-Color Violet Octagon Default Configuration Map on page 151
3-Color Red Trigon Default Configuration Map on page 153
B blank 450/50 Indo 1 (Violet), DAPI
C blank none
red trigon(633-nm laser)
A 755 LP 780/60 APC-Cy7
B 710 LP 730/45 Alexa Fluor 700
C 660/20 APC
Table E-5 6-Blue 6-Violet 0-UV 3-Red default mirror and filter configuration
Detector Array(Laser)
PMT(Detector)
LongpassDichroicMirror
BandpassFilter
Fluorochrome orScatter Parameter
blue octagon(488-nm laser)
A 755 LP 780/60 PE-Cy7
B 685 LP 695/40 PerCP-Cy5.5
C 655 LP 660/20 PE-Cy5
D 600 LP 610/20 PE-Texas Red
E 550 LP 575/26 PE
-
7/30/2019 Bd Lsrii User's Guide
141/164
Appendix E: Special Order Configurations 141
E 550 LP 575/26 PE
F 505 LP 530/30 FITC, Alexa Fluor 488
G blank 488/10 SSC
H blank none
violet octagon
(405-nm laser)
A 630 LP 655/8 Qdot 655
B 595 LP 605/12 Qdot 605
C 575 LP 585/15 Qdot 585
D 545 LP 560/20 Qdot 565
E 475 LP 525/50 AmCyan, Qdot 525
F blank 450/50 Pacific Blue
G blank blank none
H blank none
red trigon(633-nm laser)
A 755 LP 780/60 APC-Cy7
B 710 LP 730/45 Alexa Fluor 700
C 660/20 APC
6-Blue 6-Violet 0-UV 4-Red Configuration
6-Blue 6-Violet 0-UV 4-Red supports blue, violet, and red octagons.Table E-6 shows the detectors, filters, and mirrors used in the defaultconfiguration. The word blank indicates that a blank optical holder shouldbe used instead of a mirror or filter. A dash () indicates that no slot exists fora mirror in that PMT position.
The 6-Blue 6-Violet 0-UV 4-Red maps are:
6-Color Blue Octagon Default Configuration Map on page 149
6-Color Violet Octagon Default Configuration Map on page 151
-
7/30/2019 Bd Lsrii User's Guide
142/164
142 BD LSR II Users Guide
g g p p g
4-Color Red Octagon Default Configuration Map on page 154
Table E-6 6-Blue 6-Violet 0-UV 4-Red default mirror and filter configuration
Detector Array(Laser)
PMT(Detector)
LongpassDichroicMirror
BandpassFilter
Fluorochrome orScatter Parameter
blue octagon(488-nm laser)
A 755 LP 780/60 PE-Cy7
B 685 LP 695/40 PerCP-Cy5.5
C 655 LP 660/20 PE-Cy5D 600 LP 610/20 PE-Texas Red
E 550 LP 575/26 PE
F 505 LP 530/30 FITC, Alexa Fluor 488
G blank 488/10 SSC
H blank none
violet octagon(405-nm laser)
A 630 LP 655/8 Qdot 655
B 595 LP 605/12 Qdot 605
C 575 LP 585/15 Qdot 585
D 545 LP 560/20 Qdot 565
E 475 LP 525/50 AmCyan, Qdot 525
Table E-6 6-Blue 6-Violet 0-UV 4-Red default mirror and filter configuration (continued)
Detector Array(Laser)
PMT(Detector)
LongpassDichroicMirror
BandpassFilter
Fluorochrome orScatter Parameter
-
7/30/2019 Bd Lsrii User's Guide
143/164
Appendix E: Special Order Configurations 143
F blank 450/50 Pacific Blue
G blank blank none
H blank none
red octagon
(633-nm laser)
A 755 LP 780/60 APC-Cy7
B 710 LP 730/45 Alexa Fluor 700
C 675 LP 685/35 Alexa Fluor 680
D 660/20 APC
E blank blank none
F blank blank none
G blank blank none
H blank none
6-Blue 6-Violet 2-UV 3-Red Configuration
6-Blue 6-Violet 2-UV 3-Red supports blue and violet octagons, and UV and redtrigons. Table E-7 shows the detectors, filters, and mirrors used in the defaultconfiguration. The word blank indicates that a blank optical holder should beused instead of a mirror or filter. A dash () indicates that no slot exists for amirror in that PMT position.
The 6-Blue 6-Violet 2-UV 3-Red maps are:
6-Color Blue Octagon Default Configuration Map on page 149
6-Color Violet Octagon Default Configuration Map on page 151
-
7/30/2019 Bd Lsrii User's Guide
144/164
144 BD LSR II Users Guide
2-Color UV Trigon Default Configuration Map on page 152
3-Color Red Trigon Default Configuration Map on page 153
Table E-7 6-Blue 6-Violet 2-UV 3-Red default mirror and filter configuration
Detector Array(Laser)
PMT(Detector)
LongpassDichroicMirror
BandpassFilter
Fluorochrome orScatter Parameter
blue octagon
(488-nm laser)
A 755 LP 780/60 PE-Cy7
B 685 LP 695/40 PerCP-Cy5.5
C 655 LP 660/20 PE-Cy5
D 600 LP 610/20 PE-Texas Red
E 550 LP 575/26 PE
F 505 LP 530/30 FITC, Alexa Fluor 488
G blank 488/10 SSC
H blank none
violet octagon(405-nm laser)
A 630 LP 655/8 Qdot 655
B 595 LP 605/12 Qdot 605
C 575 LP 585/15 Qdot 585
D 545 LP 560/20 Qdot 565
E 475 LP 525/50 AmCyan, Qdot 525
F blank 450/50 Pacific Blue
G blank blank none
H blank none
Table E-7 6-Blue 6-Violet 2-UV 3-Red default mirror and filter configuration (continued)
-
7/30/2019 Bd Lsrii User's Guide
145/164
Appendix E: Special Order Configurations 145
H blank none
UV trigon(355-nm laser)
A 505 LP 530/30 Indo-1 (Blue)
B blank 450/50 Indo-1 (Violet), DAPI
C blank none
red trigon(633-nm laser) A 755 LP 780/60 APC-Cy7B 710 LP 730/45 Alexa Fluor 700
C 660/20 APC
6-Blue 6-Violet 2-UV 4-Red Configuration
6-Blue 6-Violet 2-UV 4-Red supports blue, violet, and red octagons, and a UVtrigon. Table E-8 shows the detectors, filters, and mirrors used in the defaultconfiguration. The word blank indicates that a blank optical holder should beused instead of a mirror or filter. A dash () indicates that no slot exists for amirror in that PMT position.
The 6-Blue 6-Violet 2-UV 4-Red maps are:
6-Color Blue Octagon Default Configuration Map on page 149
6-Color Violet Octagon Default Configuration Map on page 151
-
7/30/2019 Bd Lsrii User's Guide
146/164
146 BD LSR II Users Guide
2-Color UV Trigon Default Configuration Map on page 152
4-Color Red Octagon Default Configuration Map on page 154
Table E-8 6-Blue 6-Violet 2-UV 4-Red default mirror and filter configuration
Detector Array(Laser)
PMT(Detector)
LongpassDichroicMirror
BandpassFilter
Fluorochrome orScatter Parameter
blue octagon
(488-nm laser)
A 755 LP 780/60 PE-Cy7
B 685 LP 695/40 PerCP-Cy5.5
C 655 LP 660/20 PE-Cy5
D 600 LP 610/20 PE-Texas Red
E 550 LP 575/26 PE
F 505 LP 530/30 FITC, Alexa Fluor 488
G blank 488/10 SSC
H blank none
violet octagon(405-nm laser)
A 630 LP 655/8 Qdot 655
B 595 LP 605/12 Qdot 605
C 575 LP 585/15 Qdot 585
D 545 LP 560/20 Qdot 565
E 475 LP 525/50 AmCyan, Qdot 525F bl k 450/50 P ifi Bl
Table E-8 6-Blue 6-Violet 2-UV 4-Red default mirror and filter configuration (continued)
Detector Array(Laser)
PMT(Detector)
LongpassDichroicMirror
BandpassFilter
Fluorochrome orScatter Parameter
-
7/30/2019 Bd Lsrii User's Guide
147/164
Appendix E: Special Order Configurations 147
F blank 450/50 Pacific Blue
G blank blank none
H blank none
UV trigon
(355-nm laser)
A 505 LP 530/30 Indo-1 (Blue)
B blank 450/50 Indo-1 (Violet), DAPI
C blank none
red octagon(633-nm laser)
A 755 LP 780/60 APC-Cy7
B 710 LP 730/45 Alexa Fluor 700
C 675 LP 685/35 Alexa Fluor 680
D 660/20 APC
E blank blank none
F blank blank none
G blank blank none
H blank none
Special Order Configuration Trigon andOctagon Maps
If a slot contains a filter or mirror, a number appears in the correspondingposition on the configuration map. If a slot contains a blank optical holder,nothing is written in the corresponding position on the configuration map.
The default configuration maps are:
6-Color Blue Octagon Default Configuration Map on page 149
2-Color Violet Trigon Default Configuration Map on page 150
-
7/30/2019 Bd Lsrii User's Guide
148/164
148 BD LSR II Users Guide
6-Color Violet Octagon Default Configuration Map on page 151
2-Color UV Trigon Default Configuration Map on page 152
3-Color Red Trigon Default Configuration Map on page 153
4-Color Red Octagon Default Configuration Map on page 154
6-Color Blue Octagon Default Configuration Map
6 1 0 / 2 0
P E - T e x a s R e d
6 9 5 / 4 0
6 8 5 LP
P e r C P -C y 5.5
488-nm blue laser
-
7/30/2019 Bd Lsrii User's Guide
149/164
Appendix E: Special Order Configurations 149
5 7 5 /
2 6
0 6 0 0 L P
P E
5 3 0 / 3 0
5 0 5 L
P
5 5 0
L P
4 8 8 / 1 0
7 8 0 / 6 0
7 5 5 L P
6 6 0 / 2 0
6 5 5 L P
P E - C y 7 P E - C y
5
S S C
F I T C
2-Color Violet Trigon Default Configuration Map
4 5 0 / 5 0
P ac i f i c Bl ue
405-nm violet laser
-
7/30/2019 Bd Lsrii User's Guide
150/164
150 BD LSR II Users Guide
5 2 5 / 5 0
5 0 5 L P
A m C y a n
6-Color Violet Octagon Default Configuration Map
5 6 0 / 2 0
Q d o t 5 6 5
6 0 5 / 1 2
5 9 5 LP
Q d o t 60 5
405-nm violet laser
-
7/30/2019 Bd Lsrii User's Guide
151/164
Appendix E: Special Order Configurations 151
5 2 5 /
5 0
5 4 5 L P
A m C y a n
4 5 0 / 5 0
4 7 5
L P
P a c i f i c B l u e
6 5 5 / 8
6 3 0 L P
5 8 5 / 1 5
5 7 5 L P
Q d o t 6 5 5 Q d o t 5
8 5
2-Color UV Trigon Default Configuration Map
4 5 0 / 5 0
Ind o -1 ( V i o l e t )
355-nm UV laser
-
7/30/2019 Bd Lsrii User's Guide
152/164
152 BD LSR II Users Guide
5 3 0 / 3 0
5 0 5 L P
I n d o - 1 ( B l u e )
3-Color Red Trigon Default Configuration Map
7 3 0 / 4 5
7 10 LP
Al e x a F l uo r 7 0 0
633-nm red laser
-
7/30/2019 Bd Lsrii User's Guide
153/164
Appendix E: Special Order Configurations 153
7 8 0 / 6 0
7 5 5 L P
6 6 0 / 2 0
A P C - C y 7 A P C
4-Color Red Octagon Default Configuration Map
6 6 0 / 2 0
A P C
7 3 0 / 4 5
7 10 LP
Al e x a F l uo r 7 0 0
633-nm red laser
-
7/30/2019 Bd Lsrii User's Guide
154/164
154 BD LSR II Users Guide
7 8 0 / 6 0
7 5 5 L P
6 8 5 / 3 5
6 7 5 L P
A P C - C y 7 A l e x a F l u
o r 6 8 0
Appendix FSetting Laser Delay
This appendix describes how to optimize laser delay settings in a multiple laser
-
7/30/2019 Bd Lsrii User's Guide
155/164
155
system.
About Laser Delay on page 156
Optimizing Laser Delay on page 157
About Laser Delay
Sample interrogation takes place within the cuvette flow cell. Laser light isdirected through a series of prisms that focus multiple lasers on the event stream
at different positions. This allows optimal detection of fluorescent signal fromeach laser with minimal cross-contamination from the other beams.
In the BD LSR II four-laser system, the blue laser intercepts the stream first,followed by the violet, UV, and red lasers. Because the laser signals are spatiallyseparated, there is a slight delay between the detection of each lasers signal(Figure F-1 ).
Figure F-1 Signal separation over time
event intercepts red laser
-
7/30/2019 Bd Lsrii User's Guide
156/164
156 BD LSR II Users Guide
The laser delay setting in BD FACSDiva software is used to realign the signals sothey can be measured and displayed on the same time scale. Signals are alignedwith respect to the blue laser, so the blue laser will have a 0 delay value, and thered laser will have the longest delay.
time
red
blue
UV
violet
event intercepts blue laser
event intercepts violet laser
event intercepts UV laser
event intercepts red laser
Optimizing Laser Delay
Laser delay is set using BD FACSDiva software. To optimize the delay for a givenlaser, you acquire events from a sample with a fluorescence signal excited by that
laser. Follow the procedures in Running Samples on page 53 , for sampleoptimization and acquiring data.
To optimize laser delay:
1 While acquiring data from your sample, create a histogram to show thefluorescence signal excited by the laser in which the delay is to beoptimized.
2 In the Acquisition Dashboard, set the Events to Display to 500 evt.
-
7/30/2019 Bd Lsrii User's Guide
157/164
Appendix F: Setting Laser Delay 157
3 Select the Laser tab in the Cytometer window.
Window extension and laser delay values are displayed in microseconds(sec).
Figure F-2 Laser tab of the Cytometer window
4 Set the window extension value to 0 sec.
5 Set an initial laser delay value ONLY for the laser you are optimizing.
If you are optimizing the violet laser, set its delay to 20 sec.
If you are optimizing the UV laser, set its delay to 40 sec.
If you are optimizing the red laser, set its delay to 60 sec.
6 While observing the positive events on the histogram, adjust the laser delayin 1 sec increments within a range of 10 sec of the initial setting.
Choose the setting that moves the events farthest to the right (highestfluorescence intensity).
7 Draw an interval gate on the histogram for the positive events.
8 Create a statistics view to display the mean fluorescence intensity of thegated population.
-
7/30/2019 Bd Lsrii User's Guide
158/164
158 BD LSR II Users Guide
g p p
9 While observing the mean fluorescence intensity for the gated population,adjust the laser delay in 0.1 sec increments within a range of 2.0 sec of the setting obtained in step 6 .
You should also stay within a range of 10 sec of the initial setting(see step 5 ).
Preserve the setting that maximizes the fluorescence intensity.
10 Reset the window extension to 10 sec.
IndexAacridine orange (AO) 81adjusting
threshold 63voltages 63air in filter, removing 50l i 31
BD LSR IIcomponents 26cytometer optics 55online Help xiiworkstation 38
beam splitters See dichroic mirrors.biohazards xx, 31
-
7/30/2019 Bd Lsrii User's Guide
159/164
159
alarm, waste container 31ALPHA filters (AFs) 35analysis
data 68immunophenotyping 72reusing 77saving 77
antibodies 106arm, tube support 30assistance, technical xv
BBal seal, replacing 90bandpass (BP) filters 35, 109
holder 42specifications 45
battery test 85Battery Test switch, shown 85battery, changing 86BD FACSClean solution 81BD FACSDiva software See softwareBD FACSFlow solution 47BD FACSRinse solution 81BD Falcon tubes 122
blank optical holders 42bleach solution 81blue laser 34bubbles, removing air 50
buttonsfluid control 28sample flow rate control 28
Ccalculating compensation 66, 67capacity, waste container 31changing battery 86Class I laser product xviiiCoherent lasers 34compensation
calculating 66, 67theory 111tubes, creating 61
componentscytometer shown 26optical bench 32
computer system, about 38configuration maps 126
containers, sheath and waste 31, 46, 49control panel, cytometer 27controls
compensation 61single-stained 55, 68
cord, damaged xixcovers, cytometer 26creating
analysis objects 72compensation tubes 61global worksheets 69statistics view 73
custom configurations 130cuvette flow cell 158CV, troubleshooting 100Cy-Chrome, filter configuration for 134
detectorsphotodiode 37photomultiplier tubes (PMTs) 37sensitivity 37
dichroicfilters 35mirrors 35, 111
digital data 26discriminating filters (DFs) 110DNA
flow rate for analysis 104droplet containment system 30
troubleshooting 94DsRed, filter configuration for 133
E
http://-/?-http://-/?- -
7/30/2019 Bd Lsrii User's Guide
160/164
160 BD LSR II Users Guide
cytometer 26components shown 26configuration 43, 55covers and door 26, 44
dimensions xiiihandles 27, 32setup 54spares kit 130symbols and labels xxi
D
DAPIcleaning after using 81
dataanalyzing 68, 72gating 72recording 68, 70
debris
excessive 100removing 63delay, laser 158
Eelectrical
requirements xiiisafety xix
electronics 114emission duration 114environmental requirements xiiievent rate
erratic 98, 99high 97low 98zero 95, 96, 97
excessive debris 100excitation wavelength 106experiments
immunophenotyping 69sample optimization 58
extra filters and mirrors 130
FFalcon tubes 94
filtersbandpass 35optical
bandpass 35, 109changing 44dichroic 35discriminating 110extra 130longpass 35, 108shortpass 35, 109specifications 45theory 107
sheathreplacing 88
FITC and Stokes shift 106fixed-alignment lasers 26
Ggating data 72global worksheets 72
creating 69previewing data 68, 77
Hhandles, cytometer 27, 32hazards
biological xx, 31electrical xix
laser xviiHoechst 81hydrodynamic focusing 104
http://-/?-http://-/?- -
7/30/2019 Bd Lsrii User's Guide
161/164
161
flow cell 104draining 51
flow rate control buttons 28fluid control buttons 28
fluidics 104cleaning 80description 28, 104flow rate control 28flushing system 83priming 51removing air bubbles 50run mode 28sample injection port (SIP) 29, 30
fluidics interconnect 50fluorescence 106
emission 114fluorochromes 111
emission spectra 107filter configurations and 124, 130
role in light emission 106which measured 43
forward scatter (FSC) 33, 34, 105FSC and SSC voltages, adjusting 63FSC threshold, adjusting 63
Iimmunophenotyping 104
analysis 72
data 69experiment 69
indo-1, filter configuration for 132
J JDS Uniphase HeNe 34
Kknob
SAMPLE FINE ADJ 28sheath clamp, shown 46
Llabels on cytometer xxilaser delay 157
lasersabout 34classification xviiipower 34quality control (QC) 52, 120safety xviiwarming up 40warmup times 34
longpass (LP) filters 35, 108longpass dichroic mirrors
holder 42specifications 45
LSR II See cytometer, BD LSR II.
Mmaintenance
Ooctagon
description 32detector 26location 44shown 36
optical bench components 32optics 55
components 32cover 33custom filter configurations
DsRed 133
indo-1 132PerCP or BD Cy-Chrome 134PE-Texas Red 131
dichroic mirrors 35 111
http://-/?-http://-/?-http://-/?-http://-/?- -
7/30/2019 Bd Lsrii User's Guide
162/164
162 BD LSR II Users Guide
maintenancefluidics
daily cleaning 80flushing system 83
periodicBal seal 90replacing sheath filter 88sample tube O-ring 92
scheduledbattery change 86battery test 85system flush 83waste management system 85
Microsoft Windows operatingsystem 38
mirrorsdichroic longpass 35extra 130specifications 45
NNCCLS documents xx
dichroic mirrors 35, 111filters 35, 44, 107location 26steering 32, 36
optimizationsample 54
ordering spare parts 119O-ring, sample tube, replacing 92
PPerCP
filter configuration for 134Stokes shift 106PE-Texas Red 131photodiode 37photomultiplier tubes (PMTs) 32, 37population hierarchy 72power
requirements xixswitch 26
priming fluidics system 51propidium iodide (PI)
cleaning after using 81
pulsedefinition 114measurements 116processors 116
Qquality control (QC)
about 52particles 120troubleshooting 101
RRadius laser 34recording
compensation Tubes 65, 66
sampleinjection tube 30optimization 54
experiment 58single-stained controls 55, 68
tube requirements 94SAMPLE FINE ADJ knob 28sample injection port (SIP) 104
cleaning 80components 29, 30problems with 94replacing Bal seal 90replacing sample tube O-ring 92
samples, running 70Sapphire laser 34saving analyses 77scatter
http://-/?-http://-/?- -
7/30/2019 Bd Lsrii User's Guide
163/164
163
co pe sat o ubes 65, 66data 68, 70
red laser 34removing air bubbles, filter 50replacing
Bal seal 90battery 86optical filters 44sample tube O-ring 92sheath filter 88
requirements specifications xiiireusing analyses 77
Ssafety
biological xxelectrical xixgeneral xxilaser xviisymbols and labels xxi
scatterlight 105parameter distortion 99
setting laser delay 157
setup, cytometer 54sheathcontainer 31, 46filter
replacing 88flow 104fluid 104
backflush 30pressure 104
sheath clamp knob 46sheath container
cap 47closing 47depressurize 47
shortpass (SP) filters 35, 109
Show Population Hierarchy 73shutdown, cytometer 80side scatter (SSC) 34, 105signals, amplifying 37SIP See sample injection port.
softwareadjusting detector voltages 37cytometer control xi
sparesfilters and mirrors 130kit 130parts, ordering 119
specifications, filter and mirror 45spillover 112statistics views 72steering optics 32, 36Stokes shift 106switch, Battery Test 85
symbols on cytometer xxi
T
Vviolet laser 34virus protection software xxiiivoltages, PMT, adjusting 63
Wwaste container 31, 46
alarm 31capacity 31emptying 48
window extension 159workstation, about 38
http://-/?-http://-/?- -
7/30/2019 Bd Lsrii User's Guide
164/164
164 BD LSR II Users Guide
technical assistance xvTechnical Assistance, assistance
technical xvtemperature requirements xiiitesting battery 85text conventions xithiazole orange (TO) 81threshold
adjusting 63defined 117
trigon 37description 32detector 26location 44
tubescompensation 61requirements 94
Uungrounded receptacles xixuser preferences 56UV laser 34