BD FACSAria Options Manual

http://www.bdbiosciences.com/ Part No. 335910 Rev. A March 2003

BD Biosciences 2350 Qume Drive San Jose, CA 95131-1807 USA Tel (877) 232-8995 Fax (408) 954-2347

Asia Pacific Tel (65) 6-861-0633 Fax (65) 6-860-1590

Brazil Tel (55) 11-5185-9995 Fax (55) 11-5185-9895

Canada Tel (888) 259-0187 (905) 542-8028 Fax (905) 542-9391 [email protected] Mexico Tel (52) 55 5999 8296 Fax (52) 55 5999 8288

Europe Tel (32) 53-720211 Fax (32) 53-720450

Japan Nippon Becton Dickinson Company, Ltd. Tel 0120-8555-90

2003, Becton, Dickinson and Company. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into any language or computer language, in any form or by any means: electronic, mechanical, magnetic, optical, chemical, manual, or otherwise, without prior written permission from BD Biosciences. The information in this guide is subject to change without notice. BD Biosciences reserves the right to change its products and services at any time to incorporate the latest technological developments. Although this guide has been prepared with every precaution to ensure accuracy, BD Biosciences assumes no liability for any errors or omissions, nor for any damages resulting from the application or use of this information. BD Biosciences welcomes customer input on corrections and suggestions for improvement. BD, the BD logo, and BD FACSAria are trademarks of Becton, Dickinson and Company. Kimwipes is a registered trademark of the Kimberly-Clark Corporation. Lauda is a registered trademark of Lauda Thermostate GmbH & Co. KG. All other company and product names may be trademarks of the respective companies with which they are associated. Guide written by Kim Gautho; edited by Joan Miller; produced by Pushpa MacFarlane.

HistoryRevision 335910 Rev. A Date 03/03 Change Made Initial release

ContentsBD Aerosol Management Option Option Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Evacuator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ULPA Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Operating the BD Aerosol Management Option . . . . . . . . . . . . . . . . . . . . . . Starting Up the Evacuator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting Up for Sorting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Opening the Sort Collection Chamber Door . . . . . . . . . . . . . . . . . . . . . Turning Off the Evacuator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Replacing the ULPA Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Replacing the Air Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Control Panel Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Filter Flow Gauge Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . BD Temperature Control Option Option Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Using the BD Temperature Control Option . . . . . . . . . . . . . . . . . . . . . . . . . Setting Up the Water Bath . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting Up the Tube Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting Up the ACDU Stage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Starting Up the Water Bath . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 6 6 7 8 8 11 11 12 13 14 18 19 19 21 22 23 24 25 25 26 28 30

iii

Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

31 32

iv

BD FACSAria Options Manual

BD Aerosol Management OptionThe BD aerosol management option (AMO) is a device that promotes the containment of aerosols by evacuating the sort collection chamber in the BD FACSAria ow cytometer. The option uses an attached vacuum source to rapidly evacuate aerosolized particles through an ultra-low penetrating air lter during routine sorting or analysis. The following topics are described in this guide: Option Components on page 6 Operating the BD Aerosol Management Option on page 8 Maintenance on page 13 Troubleshooting on page 19 Specications on page 22

5

Option ComponentsThe BD aerosol management option (AMO) includes the following: an evacuator to generate negative pressure an ultra-low penetrating air (ULPA) lter to trap particles, with attached tubing that connects the evacuator to the instrument an air lter for the sort collection chamber door

The AMO also requires the splash shield in the BD FACSAria accessory kit.

B CAUTION

The BD aerosol management option does not eliminate the health risks of working with biohazardous material and must be used in conjunction with good laboratory practice. The BD aerosol management option is For Research Use Only. It is not for use in diagnostic or therapeutic procedures.

EvacuatorThe evacuator holds the ULPA lter and attached tubing. Air ow is controlled using pushbuttons within a membrane panel on the front of the unit. The evacuator sits on castor wheels for easy maneuverability. It can be moved using the handle attached to the unit. See Figure 1 on page 7.

6

BD FACSAria Options Manual

handle tubing membrane panel filter hold-down ULPA filter flow gauge

Figure 1 AMO evacuator

ULPA FilterThe ULPA lter used in the BD aerosol management option captures and retains 99.9999% of all particles down to and including particles 0.01 microns in size. For particles of this size, only 1 out of 1,000,000 passes through the ULPA lter (on average).

Figure 2 ULPA lter

BD Aerosol Management Option

7

Operating the BD Aerosol Management OptionThis section describes the following: Starting Up the Evacuator on page 8 Setting Up for Sorting on page 11 Opening the Sort Collection Chamber Door on page 11 Turning Off the Evacuator on page 12

Starting Up the EvacuatorNOTICE Before starting up the evacuator, make sure that the ULPA lter is completely seated against the bottom of the evacuator lter well, and that the tubing is securely attached to the lter and instrument manifold.

B WARNING

Any instrument surface that comes in contact with biological specimens can transmit potentially fatal disease. Use universal precautions when handling instrument hardware. Wear suitable protective clothing and gloves.

1 Install the splash shield or the tube holder below the aspirator drawer(Figure 3 on page 9). The splash shield is required for sorting into a multiwell plate or onto a slide. NOTICE If you are using the tube holder, make sure an O-ring is installed in the groove on top of the tube holder. The O-ring minimizes the chance of aerosols escaping. To install the splash shield, close the sort block door and open the sort collection chamber door, if needed. (The sort block door must be closed in order to open the collection chamber door.)

8

BD FACSAria Options Manual

Remove the tube holder, if one is installed, and tilt the splash shield to insert it into the slotted ttings below the sort aspirator drawer. (See Figure 3.) Push the splash shield all the way in.

splash shield

Figure 3 Installing the splash shield

2 Ensure that an air lter is installed in the sort collection chamber door.The lter traps airborne dust that could clog the ULPA lter. Note that the lter should be changed on a monthly basis. See Replacing the Air Filter on page 18.

air filter

3 Close the sort collection chamber door.NOTICE The sort collection chamber door must be closed for the evacuator to generate negative pressure in the chamber.

BD Aerosol Management Option

9

4 Switch on the main power on the back of the evacuator (Figure 4).main power

Figure 4 Turning on main power

5 Press the POWER button on the membrane panel of the evacuator.

power button

6 Press the up or down arrow button to set the suction control rate to 20%.Each time either arrow button is pressed, the suction will increase or decrease by 10%. When two lights are lit on the suction control indicator, the actual air ow is the value between the two illuminated percentages. NOTICE Do not set the suction control rate above 20%. Higher rates could affect the stability of the side streams.

7 Verify that the lter ow gauge reads less than 2.4 inches of H2O (Figure 5on page 11). For a new lter, the gauge should read 1.11.4 inches of H2O. As the lter is used, the reading will increase. If the gauge reads 2.4 inches of H2O or greater, replace the lter. See Replacing the ULPA Filter on page 14.

10

BD FACSAria Options Manual

new filter

used filter

Figure 5 Reading the lter ow gauge

Setting Up for SortingRefer to your instrument manual for instructions.

1 Start up the ow cytometer system. 2 Start up the evacuator as described in the previous section. 3 Follow the standard sort setup procedure.NOTICE Always start up the evacuator before setti