BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

51
The Leader in Measurement & Control Basis Weight Sensor BASIS WEIGHT SENSOR SERVICE MANUAL V5.0 October, 2010

Transcript of BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

Page 1: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control

Basis Weight Sensor

BASIS WEIGHT SENSOR

SERVICE MANUAL V5.0 October, 2010

Page 2: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 2 of 51

TABLE OF CONTENTS CHAPTER 1 ................................................................................................................................................................................. 4

1.0 BASIS WEIGHT SENSOR SOURCE ............................................................................................................................. 4 1.1 PRINCIPLE OF OPERATION ...................................................................................................................................... 4 1.1.1 RADIOACTIVE MATERIAL ................................................................................................................................... 5 1.1.2 MEASUREMENT OF PAPER ................................................................................................................................ 5 1.2 SOURCE ASSEMBLY ................................................................................................................................................. 5 1.3 SHUTTER OPERATION ............................................................................................................................................. 6 1.4 SAFETY FEATURES ................................................................................................................................................... 6 1.4.1 LOCKING MECHANISM ....................................................................................................................................... 6 1.4.2 FIRE SAFETY ....................................................................................................................................................... 7 1.4.3 RADIATION SHIELDING ...................................................................................................................................... 7 1.4.4 ENCAPSULATION ................................................................................................................................................ 7 1.4.5 RADIATION WARNING ....................................................................................................................................... 7 1.4.6 MAGNETIC SWITCH ........................................................................................................................................... 7 1.5 FLAG OPERATION ................................................................................................................................................... 8 1.6 AIR CYLINDER OPERATION & ADJUSTMENT ........................................................................................................... 8 1.7 WINDOW REPLACEMENT & CLEANING .................................................................................................................. 8 1.8 TROUBLE SHOOTING ............................................................................................................................................ 10

CHAPTER 2 ............................................................................................................................................................................... 15 2.0 BASIS WEIGHT SENSOR RECEIVER ........................................................................................................................ 15

2.1 PRINCIPLE OF OPERATION .................................................................................................................................... 15 2.2 RECEIVER ASSEMBLY DESIGN DETAIL ................................................................................................................... 16 2.2.1 ION CHAMBER ................................................................................................................................................. 16 2.2.2 SPECIAL OPTICS ................................................................................................................................................ 16 2.2.3 AMPLIFIER ........................................................................................................................................................ 17 2.2.4 ALUMINIZED MYLAR WINDOW ....................................................................................................................... 17 2.3 WINDOW REPLACEMENT AND CLEANING ............................................................................................................ 17 2.3.1 REPLACEMENT OF THE BASIS WEIGHT WINDOW ............................................................................................ 18 2.4 ION CHAMBER REPLACEMENT ............................................................................................................................. 19 2.5 ELECTROMETER AMPLIFIER DETAILED DESCRIPTION ........................................................................................... 19 2.6 POWER SUPPLY & SIGNALS .................................................................................................................................. 20 2.7 TROUBLESHOOTING ............................................................................................................................................. 21

CHAPTER 3 ............................................................................................................................................................................... 29 3.0 BASIS WEIGHT SENSOR CALIBRATION .................................................................................................................. 29

3.1 CALIBRATION OVERVIEW ..................................................................................................................................... 29 3.2 CALIBRATION ALGORITHMS ................................................................................................................................. 29 3.3 STATIC CALIBRATION ............................................................................................................................................ 31 3.3.1 SENSOR STABILITY CHECK ................................................................................................................................ 31 3.3.2 SAMPLING ........................................................................................................................................................ 33 3.4 DYNAMIC CALIBRATION ....................................................................................................................................... 34

CHAPTER 4 ............................................................................................................................................................................... 39 4.0 RADIATION SAFETY ............................................................................................................................................... 39

4.1 ABOUT NUCLEAR RADIATION ............................................................................................................................... 39 4.2 BIOLOGICAL EFFECTS OF RADIATION ................................................................................................................... 40 4.3 RADIOACTIVE MATERIALS USED IN SSS SYSTEMS ................................................................................................. 40 4.4 LIMITS FOR EXPOSURE .......................................................................................................................................... 41 4.5 EMERGENCY PROCEDURES ................................................................................................................................... 42 4.6 SAFETY PRECAUTIONS .......................................................................................................................................... 42 4.7 WARNING LIGHTS ................................................................................................................................................. 43 4.8 TRANSPORTATION OF SOURCE ............................................................................................................................. 43 4.9 CLEANING AND REPLACING SENSOR WINDOWS .................................................................................................. 43 4.10 SOURCE CAPSULE INFORMATION .................................................................................................................... 44 4.11 INSTALLING OR REMOVING THE SOURCE CAPSULE ......................................................................................... 45

CHAPTER 5 ............................................................................................................................................................................... 51 5.0 KEY COMPONENTS ..................................................................................................................................................... 51

Page 3: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 3 of 51

CHAPTER 1:

BASIS WEIGHT SENSOR SOURCE

Page 4: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 4 of 51

CHAPTER 1

1.0 BASIS WEIGHT SENSOR SOURCE

1.1 PRINCIPLE OF OPERATION The basis Weight Sensor uses the principle of radiation absorption by the paper in order to

measure the paper weight per unit area. The sensor is divided into a Source Assembly containing the Radioactive Source, and a Receiver Assembly containing the radiation sensing Ion Chamber and Amplifier. During measurement, radiation is emitted from the source and some passes through the paper to the Ion Chamber. The heavier the paper, the more radiation is absorbed by the paper, and the less is measured by the Ion Chamber.

Calibration formulas convert this measured signal into an absolute measurement of Basis

Weight in engineering units.

Kr-85 or Pm-147

Page 5: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 5 of 51

1.1.1 RADIOACTIVE MATERIAL The S‐tec Basis Weight Sensor for Tissue and light weight applications uses a 500mCi source

of the isotope Promethium 147 (Pm‐147) as the emitter material. The promethium material is solid and is contained in a 22mm diameter x 6mm deep .

For S‐tec Basis Weight Sensor for heavier weight applications uses 300mCi source of the

isotope Kr‐85. Krypton is a noble gas and is contained in a small capsule. Further information about the nature of radioactivity, radioactive isotopes, kinds of radiation

and radiation safety, and capsule specifications can be found later in this manual in the Radiation Safety Section.

1.1.2 MEASUREMENT OF PAPER The signal measured by the Ion Chamber approximates to Beer’s Law as follows: I = Io * exp(‐μ*BW) Where: I = Intensity of the signal measuring paper Io = Intensity of the signal with no paper

μ = Absorption Coefficient BW = Basis Weight of the Paper The SSS Basis Weight Sensor uses this principle but adds corrections for Bremsstrahlang

effects (gamma), automatic correction of dirt build‐up, source decay and air density changes. Most of these are corrected during the sensor standardisation procedure.

1.2 SOURCE ASSEMBLY Drawings of the Source Assembly are in the appendix to this section. The Source Assembly

consists of a source body inside which can rotate the source holder. The source holder houses the capsule of radioactive material. When the source is rotated at 90° to the vertical, no radiation can exit the source so the “shutter” is said to be closed. When the source holder is rotated so that the source points downwards radiation can exit the source assembly through a 10mm diameter aperture in the Source Body. The shutter is then said to be “open”.

During standardise the flag solenoid moves a 20mm diameter piece of Mylar plastic into the

beam to simulate a constant Basis Weight. This is used to correct for any dirt accumulation since calibration, source activity decay and for window wear or air density effects.

Page 6: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 6 of 51

1.3 SHUTTER OPERATION Movement of the shutter is controlled by the air pressure in the pneumatic cylinder.

Operation of the cylinder causes the source holder to rotate 90° to the open position by pushing on the coupling leaver. When air pressure is taken off the pneumatic cylinder the source holder rotates back to the closed position under the action of the cylinder return spring.

The source holder rotates on two bearings fitted into the bearing holders. These bearing

should never need attention. As the source holder rotates from the closed position, the coupling lever releases the micro

switch so that the Green Radiation Warning Lights go off and the Red Lights come on. Unless the Green lights are on, the shutter is not fully closed.

The shutter can be locked into the closed position by loosening the lock nut and screwing the

M4 screw in the centre of the Source Body Lid so that it enters the hole for this purpose in the source holder. For shipping purposes ensure that the screw is tight. In normal operation the screw should be locked off with the lock nut so that the shutter is free to move.

1.4 SAFETY FEATURES The design of the Source Assembly and associated equipment includes the many safety

features below:

1.4.1 LOCKING MECHANISM When working on the source assembly or in the area of the source assembly, it is advisable

to LOCK the source in the CLOSED position. Locking the source closed is also necessary during shipping or movement of the source.

The source includes a simple locking mechanism. On the top of the Source Assembly there

is an M4 screw hole in the top of the Source Body Lid. Normally an M4 stainless steel screw is located in this hole with a locking nut. The nut locks the screw in the raised position where it does not obstruct the operation of the shutter. To lock the source closed simply loosen the locking nut and screw the M4 screw as far as it will go with the shutter in the fully closed position. The screw must extend into the Source Holder Body where a hole is provided. Locking the nut again will ensure that it cannot come loose. With the screw fully in place, the source is prevented from rotating and cannot open.

Page 7: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 7 of 51

1.4.2 FIRE SAFETY In the event of a fire either within the heads or outside, the plastic compressed air tubes

feeding compressed air to the heads and distributing it within the heads, will melt or rupture. When the compressed air pressure fails, for any reason, the shutter will return to the closed position automatically. The shutter mechanism is therefore fail‐safe under fire conditions.

The Source Holder is made from stainless steel, as are the bearing holders. The Source

Body is made from a solid piece of brass. All fasteners are stainless steel. Therefore in the event of a fire, or even mechanical damage, the source capsule is well protected.

1.4.3 RADIATION SHIELDING More than adequate radiation shielding in the closed position is provided by the stainless

steel Source Holder and solid Brass Source Body. In the open position the small gap size (nominally 5‐6mm on tissue applications) and the

structure of the opposing sensor heads, provide excellent shielding. Non‐the‐less it is advisable that Operators and Engineers ensure that the shutter is properly closed before working on the heads or cleaning the sensor windows.

1.4.4 ENCAPSULATION The Pm‐147 material is housed in a Brass capsule with a thin Gold window. The capsule is

leak proof and is designed to have a working life of up to 5 years.

1.4.5 RADIATION WARNING On the Source Assembly is located a micro switch to control the radiation warning lights.

Only when the shutter is fully in the closed position will the Red lights go out and the Green lights come on. Radiation warning lights are located on each side of both the upper and lower heads. The lights are LEDs so should never need to be replaced. External warning lights on the end of the scanner are also be installed.

It is required that radiation warning labels be placed on the scanner outlining the type of

radioactive material and activity. 1.4.6 MAGNETIC SWITCH

Page 8: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 8 of 51

For additional Operator and Engineer safety, a magnetic reed switch is installed in the circuit

of the shutter solenoid. The reed switch is in the lower head and closes when the magnet on the corresponding upper head is close. This simple device ensures that the shutter cannot open unless the heads are very close together. Should he heads separate, for any reason, then a potential radiation hazard would exist if the shutter should open. This is because the lower head may not be present to act as a shield. When the heads are displaced by more than about 30mm, the reed switch opens and the solenoid valve releases, closing the shutter.

1.5 FLAG OPERATION The Flag is inserted by energising the 24V dc solenoid mounted on the Source Mounting

Plate. The solenoid rotates the flag assembly so that the Mylar sample is centred in the radiation beam when the shutter is open.

The flag should normally never need adjustment. The clearance between the flag and the

Source Body and to the Source Front Ring is very tight. Should it be necessary to adjust the flag, care should be taken to ensure the flag does not bind during the movement. Also ensure that the flag Mylar sample is exactly centred in the beam when the solenoid is fully energised.

1.6 AIR CYLINDER OPERATION & ADJUSTMENT A miniature stainless steel air cylinder operates the shutter mechanism. Compressed air to

the air cylinder is switched from the solenoid valve. The solenoid valve is a 3‐port valve which either supplies compressed air to the cylinder at the full line pressure or vents the cylinder to atmosphere. In either case the air flows to the cylinder to operate it or from the cylinder when releasing. Flow is restricted by the miniature needle valve. It is important to adjust the needle valves for a slope operation and release of the cylinder.

The purpose of the miniature needle valve, located on the cylinder entry, is to control the

rate of operation and release. The valve is adjusted so that the cylinder operates slowly and releases slowly, therefore extending the life of the cylinder and source components. If the cylinder releases too quickly the shutter warning lights can be seen to flicker as the shutter “bounces”.

1.7 WINDOW REPLACEMENT & CLEANING The Basis Weight Sensor window is an aluminized Mylar film (Pm‐147) or Antistatic Heavy

Film (Kr‐85). The window provides a barrier to dirt, dust and hot or humid air into the sensor. At the same time it is sufficiently low density to allow most of the beta particles from the source to pass through. The Source window and the Receiver window are identical.

It is important to regularly inspect and clean the window. Cleaning of the window should be done on a daily basis by the Operator.

Page 9: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 9 of 51

1.7.1 CLEANING THE BASIS WEIGHT WINDOWS The procedure for cleaning the Basis Weight Mylar windows is as follows:

1. Send the scanner off sheet and ensure the green “shutter closed” lights are on. This ensures there is no radiation hazard to the operator.

2. Command the scanner to “Head Separate”. The scanner heads will then split

allowing the Operator access to the sensor windows. 3. Using water or alcohol, very gently wipe the surface of the Basis Weight window to

remove dirt and dust. Both the upper, Receiver side, window and the lower, Source side, window should be cleaned. At the same time clean the Moisture Sensor windows and the edge sensor sapphire windows.

4. Inspect the basis weight window for any holes or cuts. Any damage to the window

might result in dirt entering the sensor. In this case the window should be replaced immediately.

5. After all cleaning has been completed make sure there are no tools or materials left

on the heads, then press the “Off Sheet” button to close and align the heads.

In the event of any window damage the window should replaced be immediately. 1.7.2 REPLACEMENT OF THE BASIS WEIGHT WINDOW When it is necessary to replace the Basis Weight window, use the following procedure:

1. Send the scanner off sheet and ensure the green “shutter closed” lights are on. This ensures there is no radiation hazard to the operator.

2. Command the scanner to “Head Separate”. The scanner heads will then split

allowing the Operator access to the sensor windows. 3. Remove the lower head cover and lock the basis weight source in the closed

position. Check again to ensure the Green “shutter closed” lights are on. 4 Carefully remove each of the 16 countersunk screws holding the Ring, Window

Holder. 5. Remove the Ring, Window Holder, by inserting a small screwdriver into one of the

screw holes and levering the ring loose very gently. DO NOT INSERT ANY OBJECT INTO THE BW WINDOW because it could puncture the Ion Chamber Window or damage the source capsule. Wear safety glasses while performing this procedure.

6. Remove the old Basis Weight Mylar Window and install a new window, if necessary

cut the window material to the correct size. 7. Fit the new window.

Page 10: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 10 of 51

8. Replace the Ring, Window Holder, and align the screw holes. 9. Insert and progressively tighten the 16 countersunk screws which hold the Ring,

Window Holder in place. It may be necessary to prick holes in the window to allow the screws to easily penetrate.

10. When all the screws are tight the window should be as flat as possible. 11. After ensuring that all tools and materials are clear of the sensor heads, command

the scanner “off sheet” to re‐align the heads.

1.8 TROUBLE SHOOTING The source assembly is very simple and fault finding is normally a case of simple observation.

Possible problems might be: 1. The shutter will not operate.

In this case first check that the solenoid valve has operated and there is compressed air supplied to the solenoid valve. Check there is air supplied to the cylinder and that the miniature needle valve is not fully closed. Also check to make sure the shutter is not locked in the closed position.

After checking all of the above, operate the shutter mechanism by hand. If it is clearly binding then investigate the source mechanism itself. If it seems necessary to dismantle the source mechanism, call a qualified person. Do not attempt any dismantling unless qualified, trained and knowledgeable.

ONLY QUALIFIED PERSONNEL WITH A LICENSE TO WORK ON RADIOACTIVE

DEVICES MAY DISMANTLE THE SOURCE MECHANISM. 2. Radiation Lights do not operate or operate incorrectly

In this case check the Micro Switch on the source mechanism. Operate the shutter by hand ‐ just a little. As soon as the shutter moves towards the open position the Micro Switch should operate to turn the Green Lights off and the Red Lights on.

3. Flag will not insert

First check that a full 24V is applied to the flag solenoid, if not look elsewhere for the problem. If the solenoid has 24V applied but will not insert, move the solenoid by hand by gently rotating the solenoid. If the solenoid operates in this way but not electrically then the solenoid coil is likely to be open circuit. If the solenoid will not rotate or if the solenoid binds, then either the flag itself is binding as it moves, or the solenoid itself is faulty or is contaminated with dirt.

Changing the solenoid or adjusting the flag involves substantial dismantling of the

source assembly. THIS MUST ONLY BE ATTEMPTED BY QUALIFIED PERSONNEL WITH PROPER TRAINING. RADIATION HAZARD!

Page 11: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 11 of 51

Page 12: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 12 of 51

SOURCE ASSEMBLY

Page 13: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 13 of 51

PICTURE SHOWING BASIS WEIGHT SOURCE ASSEMBLY INSTALLED IN HEAD

Page 14: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 14 of 51

CHAPTER 2:

BASIS WEIGHT SENSOR RECEIVER

Page 15: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 15 of 51

CHAPTER 2

2.0 BASIS WEIGHT SENSOR RECEIVER

2.1 PRINCIPLE OF OPERATION The Basis Weight Sensor Receiver is the active part of the Basis Weight Sensor. It measures

the amount of radiation that has passed through the paper from the source. This signal is then used to compute the Basis Weight of the paper. The receiver assembly is normally located in the lower enclosure head, the source is normally in the upper head.

FIG 2.1 BASIS WEIGHT RECEIVER SCHEMATIC

Beta particles from the source pass through the source Mylar window, where some are absorbed. The remainder pass through the paper where they are absorbed at a rate depending primarily on the sheet mass, then through the receiver Mylar window, optics and to the ion chamber. Particles are also absorbed by the air column mass that the particles must travel through. Once reaching the Ion Chamber the particles must then penetrate the Ion Chamber titanium steel window. The particles that succeed in entering the Ion Chamber cause the argon gas inside to ionise. Ionisation of the gas will result in a minute current flow to the operational amplifier. This current flow is proportional to the radiation energy received.

The very small current from the Ion Chamber is amplified by the Operational Amplifier circuit

on the PCB and converted to a 0 to 2mA current output to the ADC.

Amplifier Shield

Stainless Steel Window

Amplifier PCB

Page 16: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 16 of 51

2.2 RECEIVER ASSEMBLY DESIGN DETAIL The Basis Weight Sensor is designed to measure the basis weight of the paper very

accurately. To do this it needs to be very stable, rugged, insensitive to flutter & ash, and be able to be re‐calibrated on‐line automatically (standardisation), to correct for source decay and environmental changes. A detailed description of the Basis Weight Sensor Receiver would describe the following components:

2.2.1 ION CHAMBER The Ion Chamber is a sealed aluminium cylinder with an insulated electrode inside. Inside

the cylinder is Argon gas at a pressure of 1520 Torr. A window of 25 micron (0.001 inch) stainless steel allows the radiation to enter the chamber while retaining the pressurised Argon gas. Beta particles entering the ion chamber collide with the Argon gas molecules causing ionisation.

The Ion Chamber is mounted inside the aluminium sensor enclosure and insulated from it

with Delron rings. The body of the Ion Chamber is connected to the ‐400V dc supply. Positively charged ions created by the beta particle collisions with the Argon gas, migrate to the body attracted by the ‐400V potential. The Negatively charged ions are attracted to the Ion Chamber electrode, which is at virtual ground potential (+400V relative to the body), and create the small current to the amplifier. The current is proportional to the beta particle energy entering the ion chamber.

The Ion Chamber window surface should be tight due to the pressure of Argon inside. If

gas leaks from the chamber, the window may become slack and crinkled looking. The Ion Chamber cannot be repaired and must be replaced if damaged or if the gas leaks. Avoid touching the Ion Chamber window because a rupturing of the Ion Chamber Window could be a hazard to the eyes due to the pressure of Argon inside.

2.2.2 SPECIAL OPTICS The purpose of the optics is to eliminate the sensitivity of the sensor to sheet flutter and to

ash content in the paper. Flutter sensitivity means the sensitivity to the position of the sheet in the gap. The optic parts are of a proprietary nature and depend on the application.

The optics is located beneath the sensor Mylar window and immediately before the Ion

Chamber. The optics includes the source window material. All beta particles must pass through the optics before they can enter the Ion Chamber.

Page 17: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 17 of 51

2.2.3 AMPLIFIER The Amplifier PCB contains a high gain electrometer amplifier. The amplifier circuit has a

very high impedance and 10mS time constant. Gain is selected, with appropriate choice of feedback resistor, to achieve an output signal to the ADC of between 6 ‐ 8V at the ADC, with the shutter open and no paper. Care should be taken if changing the amplifier gain to not exceed 9V output for a clean sensor with no paper in the gap. More than this may overload the input of the ADC and result in a non‐linear output signal. A more detailed description of the amplifier circuit is given later in this section.

The printed circuit board is made of 2 separate printed circuit boards. The smaller, inner,

board is mounted directly on to the Ion Chamber electrode. The larger, outer, board is mounted on the top of the receiver assembly body. A stainless plate protects the amplifier board and provides a shield against electrostatic noise.

2.2.4 ALUMINIZED MYLAR WINDOW The Basis Weight Sensor window is a conductive aluminized mylar film. The window

provides a barrier to dirt, dust and hot or humid air into the sensor. At the same time it is sufficiently low density to allow most of the beta particles from the source to pass through. The Source window and the Receiver window are identical. Cleaning and replacement of the window is described below.

2.3 WINDOW REPLACEMENT AND CLEANING It is important to regularly inspect and clean the window. Cleaning of the window should

be done on a daily basis by the Operator. The procedure for cleaning the Basis Weight Mylar windows is as follows:

1. Send the scanner off sheet and ensure the green “shutter closed” lights are on.

This ensures there is no radiation hazard to the operator. 2. Command the scanner to “Head Separate”. The scanner heads will then split

allowing the Operator access to the sensor windows. 3. Using water or alcohol, very gently wipe the surface of the Basis Weight window to

remove dirt and dust. Both the upper, Source, window and the lower, Receiver, window should be cleaned. At the same time clean the Moisture Sensor windows and the edge sensor sapphire windows.

4. Inspect the basis weight window for any holes or cuts. Any damage to the window

might result in dirt entering the sensor. In this case the window should be replaced as soon as possible.

5. After all cleaning has been completed make sure there are no tools or materials left

on the heads, then press the “Off Sheet” button to close and align the heads. In the event of the window being damaged, it should be immediately replaced.

Page 18: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 18 of 51

2.3.1 REPLACEMENT OF THE BASIS WEIGHT WINDOW When it is necessary to replace the Basis Weight window, use the following procedure:

1. Send the scanner off sheet and ensure the green “shutter closed” lights are on. This ensures there is no radiation hazard to the operator.

2. Command the scanner to “Head Separate”. The scanner heads will then split

allowing the Operator access to the sensor windows. 3. Remove the upper head cover and lock the basis weight source in the closed

position. Check again to ensure the Green “shutter closed” lights are on. 4 Carefully remove each of the 16 countersunk screws holding the Ring, Window

Holder. 5. Remove the Ring, Window Holder, by inserting a small screwdriver into one of the

screw holes and levering the ring loose very gently. DO NOT INSERT ANY OBJECT INTO THE BW WINDOW because it could puncture the Ion Chamber Window. Wear eye protection during this procedure.

6. Remove the old Basis Weight Mylar Window and cut the window material to the

same size. 7. Fit the new window. 8. Replace the Ring, Window Holder, and align the screw holes. 9. Insert and progressively tighten the 16 countersunk screws which hold the Ring,

Window Holder in place. It may be necessary to prick holes in the window to allow the screws to penetrate.

10. When all the screws are tight the window should be as flat as possible. 11. After ensuring that all tools and materials are clear of the sensor heads, command

the scanner “off sheet” to re‐align the heads.

Page 19: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 19 of 51

2.4 ION CHAMBER REPLACEMENT Changing the Ion Chamber is straight forward. Proceed as follows:

1. Remove the Receiver Assembly from the heads and take to a work bench in a clean area.

2. Remove the PCB assembly, taking great care when removing the centre screw

connection to the ion chamber electrode. 3. Remove the “Plate, Retainer, Ion Chamber” cover plate. 4. Lift the Ion Chamber out of the assembly and replace with the new Ion Chamber. 5. Reassemble in the reverse order the above. Take care not to over tighten the

screw on the Ion Chamber electrode. Care should be used when handling the Ion Chamber to ensure that the thin titanium

window is not damaged. Avoid touching the window and never allow any tools or other sharp objects to go near to the window. Rupture of the window with the pressure of the Argon gas, could present a danger to the eyes.

2.5 ELECTROMETER AMPLIFIER DETAILED DESCRIPTION The Electrometer Amplifier consists of 2 printed circuit board, one inside the other. The

smaller board is connected directly on to the centre electrode of the Ion Chamber. R2 resistor (10MΩ) on this board along with C1 (100PF) provides the 10ms time constant smoothing for the signal.

The main board contains all the active elements of the amplifier. U1 is the electrometer

amplifier, AD549, which amplifies the very small signal from the Ion Chamber. The gain of this amplifier is determined by the feedback resistor R1. R1 is selected at installation in order to make the output current to the ADC about 1.6mA (approx. 8V across the ADC input). The signal will fall over time as the source decays, however adjustment of the feedback resistor is not usually necessary.

A null adjustment is provided so that when the shutter is closed the amplifier signal is

close to zero. Because it is better to avoid the possibility of a negative voltage to the ADC the null potentiometer should be adjusted to provide a slight positive signal with the shutter closed. When measured by the computer, the offset voltage should be between 20 and 100mV. The system measures these “background counts” during a periodic “background count check” procedure and subtracts them from all open shutter measurements.

U2 is a dual OP‐270Z Operational Amplifier. The first half of the IC, U2‐A, acts as a buffer

amplifier of unity gain. This feeds the second half of the IC which is a Voltage to Current output driver. The input voltage to the driver, U2‐B, is a 0 to 4V dc signal. Output is a 0 to

Page 20: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 20 of 51

2mA current signal to the ADC channel. The ADC input has a 5KΩ resistor which converts

the 0 to 2mA signal into 0 to 10V. Because the computer calculates the Basis Weight of the paper from the ratio of the open and the on‐sheet signals, the absolute value of the signal is unimportant provided the signals never saturate and are in reasonable range.

2.6 POWER SUPPLY & SIGNALS The Receiver Assembly requires +12V, ‐12V and ‐400V power and provides a 0 to 2mA

output signal. Facility exists on the board for a high voltage DC/DC converter to provide the ‐400V power, this is only used for a fixed point, stand alone, Basis Weight Sensor application. The pin connections for the power and signals are:

J1 ‐ 1 N/C J1 ‐ 2 N/C J1 ‐ 3 GND J1 ‐ 4 +12VDC J1 ‐ 5 ‐12VDC J1 ‐ 6 GND J1 ‐ 7 GND J1 ‐ 8 N/C J1 ‐ 9 GND J1 ‐ 10 ‐400VDC J2 ‐ 1 Signal Output, 0 to 2mA J2 ‐ 2 Signal Return to Ground

Page 21: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 21 of 51

2.7 TROUBLESHOOTING Most of the problems with Basis Weight Receiver would be:

1. No Signal If there is no Basis Weight signal then check the following: a) Is the shutter open? The signal will only be present if there is radiation to be

measured. Physically check the source assembly to make sure the shutter is really in the fully open position. Do not rely just on the shutter lights.

b) Are the supply voltages correct for the Basis Weight Sensor? The sensor needs +12V, ‐12V, ‐400V & Ground.

c) Are the connectors properly on? d) Is there a signal to the ADC? If yes, then the problem is elsewhere.

The above are the most likely causes. Check them before doing anything else. After this the only possible causes are with the Ion Chamber itself or with the amplifier PCB.

2. Too Small a Signal If the signal is too small and the problem happened suddenly (i.e. not due to normal

source decay over years), then check the following: a) Is the shutter opening fully? Any obstruction in the radiation path, including the

shutter not opening fully, will cause a reduced signal. b) Is there anything in the gap ‐ such as a piece of paper or excessive dirt? Are the

windows damaged allowing dirt or water to get into the sensor? c) Check the supply voltages, especially the ‐400V supply. Are they correct? d) Is the Flag stuck in or is it loose? Is the flag operating correctly? e) Check the signal at the ADC. If OK at the ADC, (up to 10V with no paper) then the

problem is elsewhere. f) Faulty Ion Chamber. g) Faulty amplifier 3. Noise A noisy signal could be: a) Some object in the paper path that is moving e.g. paper, loose flag, broken window,

shutter movement, low compressed air pressure (shutter moving). b) Water entering the sensor or on the window. c) An Ion Chamber that has leaked gas (is the window still "tight"). d) Loose connections ‐ check all, including the connections to the Ion Chamber e) Faulty amplifier. Check if the noise is coming from the sensor or elsewhere, such as ADC or Computer.

Page 22: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 22 of 51

Page 23: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 23 of 51

PICTURE SHOWING BASIS WEIGHT RECEIVER INSTALLED

Page 24: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 24 of 51

ION CHAMBER

Page 25: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 25 of 51

BW AMPLIFIER SCHEMATIC

Page 26: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 26 of 51

VOLTAGE DIVIDER RESISTORS - GAIN MULTIPLIER

R7 R7-1 R7-2 R7-3 R7-1/2 R7-1/3 R7-2/3 R7-1/2/3

R5 1 2 4 0.67 0.80 1.33 0.57

R5-1 10.00 10.00 5.00 2.50 15.00 12.50 7.50 17.50

R5-2 15.00 15.00 7.50 3.75 22.50 18.75 11.25 26.25

R5-3 20.00 20.00 10.00 5.00 30.00 25.00 15.00 35.00

R5-1/2 6.00 6.00 3.00 1.50 9.00 7.50 4.50 10.50

R5-1/3 6.67 6.67 3.33 1.67 10.00 8.33 5.00 11.67

R5-2/3 8.57 8.57 4.29 2.14 12.86 10.71 6.43 15.00

R5-1/2/3 4.62 4.62 2.31 1.15 6.92 5.77 3.46 8.08

MAIN F/B RESISTORS - GAIN MULTIPLIER

RES KΩ GAIN

1 100 3.0

2 100 3.0

3 100 3.0

1+2 50 1.5

1+3 50 1.5

2+3 50 1.5

1+2+3 33 1.0

Page 27: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 27 of 51

BASIS WEIGHT SENSOR AMPLIFIER PCB ‐ LAYOUT

Page 28: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 28 of 51

CHAPTER 3:

BASIS WEIGHT SENSOR CALIBRATION

Page 29: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 29 of 51

CHAPTER 3

3.0 BASIS WEIGHT SENSOR CALIBRATION

3.1 CALIBRATION OVERVIEW The Basis Weight Sensor is capable of very accurate measurement of the basis weight of

paper as it is being made on a paper machine. The Sensor itself, together with the ADC, provides an input to the computer. The voltage is a measurement of the radiation being received. From this signal the computer must calculate the basis weight of the paper, fully compensating for dirt build‐up, air density changes, radioactive source decay, electronic drift etc.. To achieve this, the computer is programmed with proprietary calibration algorithms.

3.2 CALIBRATION ALGORITHMS The following algorithms are proprietary and must not be disclosed to others. These

Algorithms are programmed into the computer.

I on-sheet signal (profile corrected) Io clean gauge open signal (calculated) I1 open gauge signal (actual) I2 flag signal I3 background signal aIo bremsstrahlang correction BWo dirt weight BW basis weight FW flag weight - dirt and flag readings can be modelled over narrow ranges: (1) BWo = 1 ln Io μ1 I1 (2) BWo + FW = 1 ln Io - W μ2 I2 where μ1, μ2 and W are parameters

Page 30: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 30 of 51

Standardise (Stdz): BWo & Io can be calculated from (1) and (2) at Stdz: (3) BWo = 1 [ln (I1) - μ2(FW +W)] -μ2-μ1 (I2) (4) Io = I1eμ1BWo On Sheet: (5) X = Io - aIo - I3 I - aIo - I3 where a is a calibration parameter (6) BW = AlnX - BWo 1 + (B) (X) where A and B are calibration parameters Dynamic Standardisation: Dynamic standardisation corrects for dirt build-up between full standardisations. This is done by assuming that Io stay unchanged between standardisation and by measuring I1 (profile corrected) while the scanner turns around outside the sheet. (7) BWo = 1 ln(Io) μ1 (I1)

Page 31: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 31 of 51

3.3 STATIC CALIBRATION Static Calibration of the sensor is normally carried out using Mylar or Aluminised Mylar

samples of know basis weight. A range of Mylar samples covering the total basis range of paper is required.

The procedure for static calibration should be broken down into the following steps: 1. Sensor stability check. 2. Sampling. 3. Calculation of the calibration constants. 4. Recheck of calibration A description of each of these steps is below.

3.3.1 SENSOR STABILITY CHECK Before starting the calibration procedure it is important to ensure that the sensor is

stable. Any problems with stability should be resolved before proceeding. Stability of the sensor means that the sensor gives repeatable and consistent readings for the same conditions.

The stability check is performed by doing a total of 30 successive standardise checks.

The standardization is commanded from the video screen. The results should be tabulated in the form below:

The variation of the counts above should not change by more than about 10mV

Standard Deviation (Sigma) for the Open and Flag signals. For the Closed signal this should not have varied by more than about 5mV.

Page 32: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 32 of 51

SENSOR STABILITY WORKSHEET

STZ # OPEN VOLTAGE FLAG VOLTAGE FLAG/AIR RATIO CLOSED VOLTAGE

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30

Page 33: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 33 of 51

3.3.2 SAMPLING Insert each of the Mylar samples into the gap using the sample holder to ensure exact

registration. For each sample do 5 sample checks and note the voltage for each sample check. The sample check is commanded from the video screen.

Record the results of the sampling as shown in the table below:

SAMPLE # SAMPLE 1 VOLTAGE

SAMPLE 2 VOLTAGE

SAMPLE 3 VOLTAGE

SAMPLE 4 VOLTAGE

SAMPLE 5 VOLTAGE

AVERAGE VOLTAGE

Sample1 GSM

Sample 2 GSM:

Sample 3 GSM

Sample 4 GSM

Sample 5 GSM

Next tape a fixed piece of very light Mylar or other film material of accurately known

basis weight over the sensor lower window. This is to simulate an accumulated amount of dirt of known weight. The sensor Open Voltage and Flag Voltage are measured with the “dirt” in place. These voltages will therefore be lower than those without the dirt.

STZ with Dirt

STZ 1

STZ 2

STZ 3

STZ 4

STDZ 5

AVERAGE VOLTAGE

Open Volts

Flag Volts Actual Dirt Wt

Sample 4 GSM

Sample 5 GSM

When all the results of the static sampling have been tabulated, fax these and the

stability check results to SSS. SSS engineers will use a computer program to fit the best calibration curve through the data and calculate the optimum calibration constants. This same procedure is used in‐house to calibrate the Basis Weight Sensors before shipment.

Page 34: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 34 of 51

3.4 DYNAMIC CALIBRATION Dynamic calibration is needed to make the sensor read the correct value of real paper

on a real paper machine. Although the sensor may very accurate on static (non‐moving) samples, machine conditions can change the calibration depending on such things as: sheet furnish; sheet temperature; moisture gain/loss; sheet stretch under tension, environmental effects etc. Whatever the source of the error when measuring moving paper on the machine, a method is needed to check the calibration of the sensor and make suitable corrections to the calibration.

There are 4 major ways to calibrate the Basis Weight Sensor on the paper machine.

These are: Calibrate to Mill Laboratory This method simply compares the sensor end of reel Basis Weight measurement with

that of the Mill laboratory result. To do this well, it is essential that a large number of comparisons are made before adjusting the sensor calibration.

The advantage of this method, which is useful for getting the sensor to correct “ball

park” reading, is that it is simple and the sensor reads similar values to that of the laboratory.

The disadvantage of this method is that the laboratory result is often noisy, can have a

systematic error, often changes as the laboratory person changes at each shift, and most importantly, proper precautions are rarely taken by laboratory people to protect the samples from moisture gain or loss.

This method, then, is recommended only to get the sensor reading close to the correct

value or where it is a requirement that the sensor measure the same as the laboratory. Roll Weight Check The SSS system measures the length of paper made during a reel and also the width.

From this data and from the measured average Basis Weight, the computer calculates the weight of the paper on the reel.

It is possible to accurately weigh the roll of paper after the change of reel. Subtracting

the weight of the steel shell from this will give the weight of the paper. This can be compared to the measured weight to calculate the difference.

The advantage of this method is that it is relatively easy to do if accurate scales are

available. The disadvantage of this method is that the accuracy of the scales is very important and

the paper weight is the result of two measurement ‐ paper + steel shell, and steel shell without paper. The overall accuracy also depends on an accurate width measurement and an accurate tachometer measurement of the reel speed. Another disadvantage is that when measuring paper grades that contract when no longer under tension (such as crepe tissue), the weight calculated is often significantly different from the weight measured by the laboratory of a relaxed, and so contracted, sheet of paper.

Page 35: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 35 of 51

Single Point Measurement(s) In this method the scanner heads are put in single point (fixed position on sheet) to

continuously measure the Basis Weight of the paper in that position. Towards the end of the reel, a card is placed in the paper to indicate the start of measurement. At the same time the single point readings are integrated and averaged until reel turn‐up occurs.

When the reel is out, many samples are cut from the paper in the position of the sensor during single point measurement. The average Basis Weight reading calculated from the samples, is compared to the average sensor measurement. If 3 or more such samples (on different reels) show a consistent error, then the sensor calibration is adjusted by this amount.

The advantage of this method is that it is reasonably accurate and not too difficult to do.

It does use laboratory sampling methods but precautions should be taken to avoid moisture gain or loss ‐ such as using polythene bags to hold the samples.

The disadvantage of this method is that it can sample only one spot each time and so

can give only a required intercept change to the calibration. More sampling in different CD positions is needed to determine if there is a slope error. If the paper is uneven with narrow basis weight streaks, then it is difficult to align the measured paper with the sampled paper. This is especially so given the narrow measurement beam (10mm at the source window) of the sensor. The manual samples taken will always be larger than 10mm, so sensor measured paper and laboratory sampled paper will be different.

CD Sampling Method With this method the trend average profile at the reel turn‐up is compared to multiple

samples of paper taken across the full width of the reel after turn‐up. One way to do this is to slab off the top 25mm of paper onto the floor. With a template of know area and sharp knife, samples are taken across the full width of paper. By alternating the position of the sample in the MD direction, all the paper can be sampled. By cutting into the paper layers, many pieces of paper can be averaged in each sample. Proper polythene bagging procedures need to be adhered to in order to avoid the possibility of moisture gain or loss in the long time elapsed from the reel turn‐up to the measuring the sample weights in the laboratory.

After weighting each sample and calculating the Basis Weight from the sample weight

(less bag) and the paper area, the results of each sample are graphed. The graph of sample basis weight can be compared to the graph of sensor profile to show any differences.

The advantage of this method is that the full CD width of the paper is measured at one

time and compared with the CD profile. From this result both intercept (offset) and slope (scale) errors can be determined and corrected.

The disadvantage of this method is that it is very time consuming (allow one hour). It

can only be performed during stable running of the machine when the end of reel profile can be taken as truly representing the measurement of the slabbed‐off paper. A significant error can be introduced if the sample cutting is not done perfect, or is done inconsistently.

Page 36: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 36 of 51

Example Calibration Curve for Kr‐85

Kr-85 BW Calibration

0

1

2

3

4

5

6

7

8

9

10

0 20 40 60 80 100 120 140 160

g/m2

Sign

al V

olta

ge

Page 37: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 37 of 51

Example Calibration Curve for Pm‐147

BW / VOLTS

0

1

2

3

4

5

6

0 10 20 30 40 50 60

BW GSM

Vol

ts

After Linearization

BW / LN X

0

0.2

0.4

0.6

0.8

1

1.2

0 10 20 30 40 50 60

BW GSM

LN X

Page 38: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 38 of 51

CHAPTER 4:

RADIATION SAFETY

Page 39: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 39 of 51

CHAPTER 4

4.0 RADIATION SAFETY

4.1 ABOUT NUCLEAR RADIATION Radioactive atoms give off ionising radiation in

the form of fast‐flying particles or as rays. Some of these particles are called Alpha

Particles. They travel only a few centimetres in air and have little or no penetrating power. Alpha particles are fast moving Protons.

Other particles are called Beta Particles. They travel about one meter or so in air but

will not penetrate much beyond the depth of a person’s skin. Beta particles are fast moving electrons.

Still stronger radiation, in the form of electromagnetic rays, is Gamma Rays. They

travel at the speed of light. An only dense material with a high atomic number such as lead is effective in stopping them. Gamma rays are the same as X‐rays except that they are generated in the atom rather than in an X‐ray tube.

Note: The SSS Basis Weight Sensor uses Beta radiation.

Page 40: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 40 of 51

4.2 BIOLOGICAL EFFECTS OF RADIATION All radiation exposure is of concern. When working with or near radiation sources of

any kind, you should be aware of the possible risks. Allowable levels of personal exposure will depend on the type of radioactive material and the situation.

Danger to the human body of Alpha and Beta Particles is mostly to the skin and the

eyes. This is because these particles cannot penetrate very far into the body. Danger from Gamma Rays is to the whole body.

Over exposure of radiation may damage or kill the tiny living cells that make up our

body. Large amounts of radiation or repeated high exposures may increase the risk of serious disease such as cancer or leukaemia later in life.

For these reasons it is essential that proper safety precautions are adhered to and that

only properly trained personnel work on or near radioactive materials.

4.3 RADIOACTIVE MATERIALS USED IN SSS SYSTEMS SSS Basis Weight sensors are used to measure paper and paper board. The principle of

measurement is that radiation is absorbed by the paper or board. Therefore the type of radiation that is used in SSS systems MUST be easily absorbed by paper but not be totally absorbed. Only Beta particles satisfy this need.

The Basis Weight sensor in this system uses Promethium 147 isotope (Pm‐147). This

material is one of the safest source materials available since the beta radiation is of low energy.

Pm‐147 has a half life of 2.6 years. This means that the source activity will fall to half

after this time. Given this relatively short half life, the source will need to be replacing at least every 5 years.

For comparison purposes below are the source materials often found in Paper Sensors

are: SOURCE RADIATION MATERIAL TYPE HALF LIFE Krypton 85 Beta, some Gamma Gas 10.7 years Promethium 147 Beta, some Gamma Solid 2.6 years Strontium 90 Beta , some Gamma Solid 28 years Iron 55 x‐ray (Gamma) Solid 2.7 years

Page 41: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 41 of 51

4.4 LIMITS FOR EXPOSURE The quantity of ionising radiation exposure is measured in terms of ionisation called

roentgens but may also be measured in terms of energy absorbed called Rads or Rems.

R = Roentgen This the special unit of exposure. 2.58 x 10‐4 Coulomb per kilogram of air under standard conditions.

Rad Unit of absorbed dose is the energy imparted to matter (such as human

tissue) by ionising radiation per unit mass of irradiated material at the place of interest. For X‐rays or Gamma Rays, 1 R (Roentgen) of exposure will produce about 1 Rad of absorbed dose in soft tissue.

Gy = (Gray) The SI unit of absorbed dose. 1 Rad = 0.01 Gy Rem Stands for “Roentgen Equivalent Man”. A special unit of dose

equivalent. The dose equivalent in Rems is numerically equal to the absorbed dose in Rads multiplied by a factor weighted for relative biological effect on the tissue being exposed. For routine protection calculations of exposure to x‐rays, 1R = 1 Rad = 1 Rem.

Sv = (Sievent) The SI unit dose equivalent to 1 Rm = 0.01 Sv DOSE LIMITS

All radiation is of concern. When working around sources of radiation you should be aware of the possible risks. Allowable levels of personal exposure will vary depending on the situation. Limit of exposure are established in each country. Typical dose limits are:

Maximum Yearly dose: REM mSv Occupationally exposed ‐ whole body 5 50 Skin or any extremities 50 500 Eyes 15 150 Dose to Embryo / Foetus 0.5 5 Public or Non‐occupationally Exposed 0.1 1

Page 42: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 42 of 51

4.5 EMERGENCY PROCEDURES In the event of an emergency such as a serious fire affecting the source enclosure, or if

the red shutter light stays on and cannot be made to go off, do the following:

1. Send the scanner off sheet. 2. Keep unauthorized personnel away from the scanner heads. Keep a distance

of about 3 Metres for safety. 3. Immediately inform the site radiation safety officer or SSS trained engineer. The trained engineer or safety officer will investigate the problem and determine if any risk exists. In the case of a shutter not closing properly he may be able to rectify the problem without any exposure risk to himself or others. If there is evidence of possible damage to the source capsule then the action taken will depend on the type of source material. For Promethium 147 a wipe test must be done and the resulting swab tested for any sign of radioactive material leakage using a radiation survey meter.

4.6 SAFETY PRECAUTIONS It is essential that all personnel who work on or near a radiation source take sensible

safety precautions to minimise any exposure to radiation. Because the source material in SSS sensors emits mostly Beta radiation which cannot

penetrate more than a little over one meter of air, avoiding risk of exposure can be done in 3 ways:

Shielding ‐ ensure that the shutter is closed and the green light is on, before working on

the heads or the sensor. Distance ‐ keep a distance from the source. Beta radiation cannot penetrate far in air

(about 1.5M) so maximise the distance between any part of the body and the source. Time ‐ Minimise the time near to the source. Any exposure is halved if only half the

time is spent in an exposed condition. Other safety precautions are: Operators and Engineers should avoid going close to the scanner heads when the red

radiation warning lights are on. Press “Off Sheet” to send the scanner to the off sheet position and to close the shutter. The shutter should be closed whenever cleaning the sensor windows or otherwise working near to the sensor heads. The shutter is confirmed closed when the green shutter close lights are on. If working on the Basis Weight sensor, it is suggested that the shutter be locked in the closed position. Details

Page 43: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 43 of 51

of the source assembly and the method of locking the source in the closed position are explained in an earlier section of this manual.

4.7 WARNING LIGHTS Radiation warning lights (large LEDs) are located on both sides of the top and the

bottom scanner heads. The Red lights indicate that the shutter is open or not closed. The Green lights indicate that the shutter is in the fully closed position. The lights are operated by a micro switch on the source assembly.

When working on or near to the scanner heads ensure that the shutter is fully closed

and observe the warning lights.

4.8 TRANSPORTATION OF SOURCE Should the source have to be moved or transported, it is very important that it is made

safe for the journey. If transporting the source within the source assembly, then lock the source assembly and ensure that it cannot come loose.

If the source capsule is transported outside of the source assembly (not advisable except

by specially trained personnel), then it MUST be shipped inside the original lead pouch and container. The source package must then be enclosed in the original box with radiation label on the outside identifying the source material and activity.

Whenever a radiation source is moved, the proper authorities must be informed and

permits obtained. Because regulations vary from country to country so it is not possible to outline the procedures in this manual.

4.9 CLEANING AND REPLACING SENSOR WINDOWS Cleaning and replacing sensor windows risks exposure to radiation if not performed

correctly or with safety in mind. Please refer to earlier sections of this manual (Basis Weight Sensor Source & Basis Weight Sensor Receiver) for full details about how to change or replace sensor windows in a safe manner.

Page 44: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 44 of 51

4.10 SOURCE CAPSULE INFORMATION

PROMETHIUM SOURCE CAPSULE Radioisotope: Pm‐147 Half Life: 2.6234 years Activity (typical): 500mCi (22 GBq) Radiation Type: Beta, some Gamma Capsule Type: VZ‐2824 Window Daimeter: 15.6mm Window: Titanium, 5 micron Source Body: Brass Rec.Working Life: 5 Years Kr‐85 SOURCE CAPSULE Radioisotope: Kr‐85 Half Life: 10.756 years Activity (typical): 300mCi (11.1GBq) Radiation Type: Beta, some Gamma Capsule Type: X.1088 Window Diameter: 6.3mm Window Material: Titanium Source Body: Titanium Rec.Working Life: 10 Years

Page 45: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 45 of 51

4.11 INSTALLING OR REMOVING THE SOURCE CAPSULE

ONLY QUALIFIED PERSONS ARE ALLOWED TO INSTALL, REMOVE OR DO ANY WORK ON THE RADIOACTIVE SOURCE. Should it be necessary to remove, install or work with the radioactive source, then the following provide a guide regarding the safety procedures and what to expect.

Safe Working Conditions Prepare safe working conditions. Necessary items to prepare are: 1. Safety glasses to protect the eyes. 2. Rubber gloves to protect the fingers and hands 3. A thick desk located against a solid wall to absorb any radiation. 4. A sheet of glass to protect the body.

The picture below shows a typical set‐up. The glass (or thick Perspex) shield is placed against the wall to provide a protective screed to work behind. Gloves and Safety glasses are also used. The work is done by keeping the source material behind the glass and always pointing the source away from any part of the body.

Page 46: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 46 of 51

TYPICAL SETUP TO REDUCE THE RISK OF RADIATION EXPOSURE.

THICK BRICK WALL

GLASS SCREEN

HEAVY TABLE

RUBBER GLOVES

Page 47: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 47 of 51

REMOVING THE SOURCE AND PLACING IN THE SOURCE HOLDER REMOVE THE BRASS CAP ONLY AT THE LAST MOMENT

KR‐85 SOURCE PICTURES

SOURCE PACKAGING WITH LABEL

SOURCE BODY BRASS SOURCE CAP

SOURCE HOLDER

HOLD THE SHUTTER OPEN AND INSERT THE SOUCE HOLDER IN HERE.

Page 48: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 48 of 51

KR‐85 SOURCE CAPUSULE

Page 49: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 49 of 51

Page 50: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 50 of 51

CHAPTER 5:

KEY SENSOR COMPONENTS

Page 51: BASIS WEIGHT SENSOR SERVICE MANUAL - VremSoft

The Leader in Measurement & Control Page 51 of 51

CHAPTER 5

5.0 KEY COMPONENTS

COMPONENT MODEL MANUFACTURER COUNTRY OF ORIGIN

Ion Chamber 52024 (Kr‐85), 520121 (Pm‐147) LND Inc. USA

Source KAC10883 (Kr-85) or PHCB11998 (Pm-147) QCS Global Germany

Pneumatic Cylinder USR-08-1 Cylinder, 1" stroke Clippard USA

Source Shielding Brass S‐tec China

Metal Parts Anodized (or Alodyne) Aluminium S‐tec China

Amplifier IC AD549 / OP‐270 Analog Devices USA

MicroSwitch Standard Any quality supplier China

Flag Solenoid