basics of ir.doc
-
Upload
vijay-pradhan -
Category
Documents
-
view
214 -
download
0
Transcript of basics of ir.doc
-
7/27/2019 basics of ir.doc
1/12
.
Basic concept ofIR spectroscopy
Presented By:
Shaikh Farida Banu
M.Sc.-I
-
7/27/2019 basics of ir.doc
2/12
Infrared spectroscopy (IR spectroscopy) is thespectroscopythat deals with the infrared region
of the electromagnetic spectrum, that is light with a longerwavelengthand
lowerfrequency than visible light. It covers a range of techniques, mostly based on absorption
spectroscopy. As with all spectroscopic techniques, it can be used to identify and
study chemicals. A common laboratory instrument that uses this technique is a Fourier transform
infrared(FTIR)spectrometer.
The infrared portion of the electromagnetic spectrum is usually divided into three regions; the
near-, mid- and far- infrared, named for their relation to the visible spectrum. The higher-energy
near-IR, approximately 140004000 cm1 (0.82.5 m wavelength) can
excite overtone orharmonicvibrations. The mid-infrared, approximately 4000400 cm1 (2.5
25 m) may be used to study the fundamental vibrations and associated rotational-
vibrational structure. The far-infrared, approximately 40010cm1(251000 m), lying adjacent to
the microwaveregion, has low energy and may be used forrotational spectroscopy. The names
and classifications of these subregions are conventions, and are only loosely based on the
relative molecular or electromagnetic properties.
Theory
Infrared spectroscopy exploits the fact that molecules absorb specific frequencies that are
characteristic of their structure. These absorptions are resonant frequencies, i.e. the frequency of
the absorbed radiation matches the transition energy of the bond or group that vibrates. The
energies are determined by the shape of the molecularpotential energy surfaces, the masses of
the atoms, and the associated vibronic coupling.
In particular, in the BornOppenheimerand harmonic approximations, i.e. when the molecular
Hamiltonian corresponding to the electronic ground state can be approximated by a harmonic
oscillatorin the neighborhood of the equilibriummolecular geometry, the resonant frequencies are
associated with the normal modes corresponding to the molecular electronic ground state
potential energy surface. The resonant frequencies are also related to the strength of the bond
and the mass of the atoms at either end of it. Thus, the frequency of the vibrations are associated
with a particular normal mode of motion and a particular bond type.
Number of vibrational modes
In order for a vibrational mode in a molecule to be "IR active," it must be associated with changes
in the dipole. A permanent dipole is not necessary, as the rule requires only a change in dipole
moment.[1]
A molecule can vibrate in many ways, and each way is called a vibrational mode. For molecules
with N atoms in them, linear molecules have 3N 5 degrees of vibrational modes, whereas
nonlinear molecules have 3N 6 degrees of vibrational modes (also called vibrational degrees of
http://en.wikipedia.org/wiki/Spectroscopyhttp://en.wikipedia.org/wiki/Spectroscopyhttp://en.wikipedia.org/wiki/Spectroscopyhttp://en.wikipedia.org/wiki/Infraredhttp://en.wikipedia.org/wiki/Electromagnetic_spectrumhttp://en.wikipedia.org/wiki/Electromagnetic_spectrumhttp://en.wikipedia.org/wiki/Wavelengthhttp://en.wikipedia.org/wiki/Wavelengthhttp://en.wikipedia.org/wiki/Frequencyhttp://en.wikipedia.org/wiki/Visible_lighthttp://en.wikipedia.org/wiki/Absorption_spectroscopyhttp://en.wikipedia.org/wiki/Absorption_spectroscopyhttp://en.wikipedia.org/wiki/Chemicalhttp://en.wikipedia.org/wiki/Fourier_transform_infrared_spectroscopyhttp://en.wikipedia.org/wiki/Fourier_transform_infrared_spectroscopyhttp://en.wikipedia.org/wiki/Spectrometerhttp://en.wikipedia.org/wiki/Spectrometerhttp://en.wikipedia.org/wiki/Overtoneshttp://en.wikipedia.org/wiki/Harmonic_(mathematics)http://en.wikipedia.org/wiki/Harmonic_(mathematics)http://en.wikipedia.org/wiki/Rotational-vibrational_spectroscopyhttp://en.wikipedia.org/wiki/Rotational-vibrational_spectroscopyhttp://en.wikipedia.org/wiki/Wavenumberhttp://en.wikipedia.org/wiki/Wavenumberhttp://en.wikipedia.org/wiki/Wavenumberhttp://en.wikipedia.org/wiki/Wavenumberhttp://en.wikipedia.org/wiki/Microwavehttp://en.wikipedia.org/wiki/Microwavehttp://en.wikipedia.org/wiki/Rotational_spectroscopyhttp://en.wikipedia.org/wiki/Rotational_spectroscopyhttp://en.wikipedia.org/wiki/Rotational_spectroscopyhttp://en.wikipedia.org/wiki/Resonant_frequencyhttp://en.wikipedia.org/wiki/Potential_energy_surfacehttp://en.wikipedia.org/wiki/Vibronic_couplinghttp://en.wikipedia.org/wiki/Born%E2%80%93Oppenheimer_approximationhttp://en.wikipedia.org/wiki/Born%E2%80%93Oppenheimer_approximationhttp://en.wikipedia.org/wiki/Molecular_Hamiltonianhttp://en.wikipedia.org/wiki/Molecular_Hamiltonianhttp://en.wikipedia.org/wiki/Ground_statehttp://en.wikipedia.org/wiki/Quantum_harmonic_oscillatorhttp://en.wikipedia.org/wiki/Quantum_harmonic_oscillatorhttp://en.wikipedia.org/wiki/Quantum_harmonic_oscillatorhttp://en.wikipedia.org/wiki/Molecular_geometryhttp://en.wikipedia.org/wiki/Molecular_geometryhttp://en.wikipedia.org/wiki/Normal_modeshttp://en.wikipedia.org/wiki/Atomic_masshttp://en.wikipedia.org/wiki/Infrared_spectroscopy#cite_note-1http://en.wikipedia.org/wiki/Spectroscopyhttp://en.wikipedia.org/wiki/Infraredhttp://en.wikipedia.org/wiki/Electromagnetic_spectrumhttp://en.wikipedia.org/wiki/Wavelengthhttp://en.wikipedia.org/wiki/Frequencyhttp://en.wikipedia.org/wiki/Visible_lighthttp://en.wikipedia.org/wiki/Absorption_spectroscopyhttp://en.wikipedia.org/wiki/Absorption_spectroscopyhttp://en.wikipedia.org/wiki/Chemicalhttp://en.wikipedia.org/wiki/Fourier_transform_infrared_spectroscopyhttp://en.wikipedia.org/wiki/Fourier_transform_infrared_spectroscopyhttp://en.wikipedia.org/wiki/Spectrometerhttp://en.wikipedia.org/wiki/Overtoneshttp://en.wikipedia.org/wiki/Harmonic_(mathematics)http://en.wikipedia.org/wiki/Rotational-vibrational_spectroscopyhttp://en.wikipedia.org/wiki/Rotational-vibrational_spectroscopyhttp://en.wikipedia.org/wiki/Wavenumberhttp://en.wikipedia.org/wiki/Microwavehttp://en.wikipedia.org/wiki/Rotational_spectroscopyhttp://en.wikipedia.org/wiki/Resonant_frequencyhttp://en.wikipedia.org/wiki/Potential_energy_surfacehttp://en.wikipedia.org/wiki/Vibronic_couplinghttp://en.wikipedia.org/wiki/Born%E2%80%93Oppenheimer_approximationhttp://en.wikipedia.org/wiki/Molecular_Hamiltonianhttp://en.wikipedia.org/wiki/Molecular_Hamiltonianhttp://en.wikipedia.org/wiki/Ground_statehttp://en.wikipedia.org/wiki/Quantum_harmonic_oscillatorhttp://en.wikipedia.org/wiki/Quantum_harmonic_oscillatorhttp://en.wikipedia.org/wiki/Molecular_geometryhttp://en.wikipedia.org/wiki/Normal_modeshttp://en.wikipedia.org/wiki/Atomic_masshttp://en.wikipedia.org/wiki/Infrared_spectroscopy#cite_note-1 -
7/27/2019 basics of ir.doc
3/12
-
7/27/2019 basics of ir.doc
4/12
unit time) and its wavenumber k(the number of waves per unit length) [3]. The
velocity for EM-waves proves to be a constant for the medium in which the waves
are propagating, that is 3 108 m/s in vacuum. It can be easily seen that
c=
(1)
the wavelength being inversely proportional to the frequency. Using the quantumcharacter of EM radiation, there is the relation
E=h
(2)
where Eis the photon energy and h is Planks constant [4]. Each quantum of radiant
energy or photon, has an energy proportional to its frequency [5]. By this means,
a photon with a higher frequency contains more energy.
The EM spectrum can be divided into several regions differing in frequency only. It
extends from gamma (1020Hz) and X-rays (31016
-
7/27/2019 basics of ir.doc
5/12
vibrational and rotationa lstates.
For a molecule to absorb IR, the vibrations or rotations within a molecule must cause a net
change in the dipole moment of the molecule. The alternating electrical field of the radiation
(remember that electromagnetic radation consists of an oscillating electrical field and an
oscillating magnetic field, perpendicular to each other) interacts with fluctuations in the dipole
moment of the molecule. If the frequency of the radiation matches the vibrational frequencyof the molecule then radiation will be absorbed, causing a change in the amplitude of
molecular vibration.
Figure 2: A schematic representation of the quantized electronic and vibrational energy levels
of a molecule. It is clear that to raise the electronic energy state of a molecule from the
ground state to the excited state will cost more energy than to raise the vibrational energy
state.
An example of an electronically excited atom comes by considering the hydrogen
atom. The ground state of the hydrogen atom corresponds to having the atom's
single electron in the lowest possible orbits. By giving the atom additional energy,
the electron is able to move into a higher orbit. Now, we deem the hydrogen atom to
be in its excited state [6].
The vibrational energy of a molecule is not determined by the orbit of an electron but
by the shape of the molecule, the masses of the atoms and, eventually by the
associated vibronic coupling1). In the study of molecular vibrations I can start with a
classical model of the molecule where the nuclei are represented by mathematical
points with mass. The forces that hold the molecule together are assumed to be
similar to those exerted by massless springs. Each mass requires three coordinates
to define its position, such asx, y, andzin a Cartesian coordinate system. As aresult it has three independent degrees of freedom of motion. If there are Natomic
nuclei in the molecule, there will be a total of3Ndegrees of freedom of motion for all
the nuclear masses in the molecule. After subtracting the translational and rotational
degrees of freedom from the total 3Ndegrees of freedom, we are left with 3N
6 internal degrees of freedom for a nonlinear molecule and 3N 5internal degrees
of freedom for a linear molecule. These are the so-called normal modes of vibrations
and they result in specific natural vibrational frequencies for different molecular
-
7/27/2019 basics of ir.doc
6/12
configurations. The typical frequencies of these vibrations match the frequency of IR
radiation (!).
For example, simple diatomic molecules have only one bond allowing only stretching
vibrations. More complex molecules may have many bonds, and vibrations can be
conjugated. The atoms in a CH2 group, commonly found in organic compounds, can
vibrate in six different ways; symmetrical and antisymmetricalstretching, scissoring,rocking, wagging and twisting.
Interaction of light and moleculesNow that I have discussed the energy level structure of molecules, I will address the
interaction of these molecules with EM waves. These interactions are at the very
heart of IR spectroscopy.
Roughly, there are five possible effects of interaction between radiation and
molecules. These are scattering,absorption, and emission. [8]. Absorption is the
process by which the energy of a photon is taken up by the matter, and this process
plays a key role in IR spectroscopy. There are several types of physical processes
that could lie behind absorption, depending on the quantum energy of the particular
frequency of EM radiation. For example, high energetic ultraviolet (UV) radiation cancause ionization and visible light usually causes electron transitions. As told before,
the energy levels for all physical processes at the atomic and molecular levels are
quantized, and if there are no available quantized energy levels with spacings which
match the quantum energy of the incident radiation, then the material will be
transparent to that radiation. In view of IR spectroscopy, I will now focus on the
absorption of IR radiation by matter.
IR radiation does not have enough energy to induce electronic transitions as seen
with UV and visible light. Absorption of IR is restricted to excite vibrational and
rotational states of a molecule.
Even though the total charge on a molecule is zero, the nature of chemical bonds is
such that the positive and negative charges do not necessarily overlap in this case.Such molecules are said to be polar because they possess a permanent dipole
moment. For a molecule to absorb IR, the vibrations or rotations within a molecule
must cause a net change in the dipole moment of the molecule. The alternating
electrical field of the radiation interacts with fluctuations in the dipole moment of the
molecule. If the frequency of the radiation matches the vibrational frequency of the
molecule then radiation will be absorbed, causing a change in the amplitude of
molecular vibration. The result of IR absorption is heating of the matter since it
increases molecular vibrational energy. Molecular vibrations give rise to absorption
bands throughout most of the IR region of the spectrum. The far IR, lying adjacent
to the microwave region, has low energy and may be used for rotational
spectroscopy.
http://en.wikipedia.org/wiki/Image:Symmetrical_stretching.gifhttp://en.wikipedia.org/wiki/Image:Asymmetrical_stretching.gifhttp://en.wikipedia.org/wiki/Image:Asymmetrical_stretching.gifhttp://en.wikipedia.org/wiki/Image:Scissoring.gifhttp://en.wikipedia.org/wiki/Image:Rocking.gifhttp://en.wikipedia.org/wiki/Image:Wagging.gifhttp://en.wikipedia.org/wiki/Image:Twisting.gifhttp://en.wikipedia.org/wiki/Scattering_(optics)http://fuse.pha.jhu.edu/~wpb/spectroscopy/basics.htmlhttp://en.wikipedia.org/wiki/Ionizationhttp://start.student.utwente.nl/quantumoptics/rabi.phphttp://hyperphysics.phy-astr.gsu.edu/hbase/electric/diph2o.htmlhttp://hyperphysics.phy-astr.gsu.edu/hbase/electric/diph2o.htmlhttp://en.wikipedia.org/wiki/Image:Symmetrical_stretching.gifhttp://en.wikipedia.org/wiki/Image:Asymmetrical_stretching.gifhttp://en.wikipedia.org/wiki/Image:Asymmetrical_stretching.gifhttp://en.wikipedia.org/wiki/Image:Scissoring.gifhttp://en.wikipedia.org/wiki/Image:Rocking.gifhttp://en.wikipedia.org/wiki/Image:Wagging.gifhttp://en.wikipedia.org/wiki/Image:Twisting.gifhttp://en.wikipedia.org/wiki/Scattering_(optics)http://fuse.pha.jhu.edu/~wpb/spectroscopy/basics.htmlhttp://en.wikipedia.org/wiki/Ionizationhttp://start.student.utwente.nl/quantumoptics/rabi.phphttp://hyperphysics.phy-astr.gsu.edu/hbase/electric/diph2o.htmlhttp://hyperphysics.phy-astr.gsu.edu/hbase/electric/diph2o.html -
7/27/2019 basics of ir.doc
7/12
Figure 3: Energy levels of a molecule during the absorption of a photon. Photons with specific
energies will be absorbed by the molecule if this energy is equal to the difference between the
energy levels; this is when the frequency of the IR radiation matches the vibrational frequency
of the molecule. Before absorption of the incident photon the molecule is in the ground state.
After absorption the molecule is in an excited vibrational state but still in the ground state of
the electronic energy level. Of course the molecule does not remain in this excited state
forever. The energy absorbed by a molecule is rapidly dissipated: it will be transformed into
kinetic energy as result of collisions or released again as photon.
IR SpectrometerIn an IR spectrometer molecules are irradiated with a whole range of IR frequencies
but are only capable of absorbing radiation energy at certain specific frequencies
which match the vibration frequencies of the molecule. The energy absorbed by a
molecule is rapidly dissipated. The excited molecule loses its energy in less than 10 -6seconds. The energy is either transformed into kinetic energy as result of collisions
or released again as photon. As the direction of the emitted photon is random in
space, and because the absorption process can be repeated for such a photon on its
way trough the medium, it can be seen that for a once-absorbed photon the
probability of re-emerging from the medium in the direction of the transmitted beam
is negligibly small [1]. A spectrometer usually measures the radiation intensity as a
function of the wavelength of the light behind a sample. At the vibrational
frequencies of the molecules an intensity decrease is obtained and a transmittance or
absorbance spectrum is plotted. In this way a sample can be characterized and
allows to determine which chemical bonds are present.
http://en.wikipedia.org/wiki/Infrared_Spectroscopy_Correlation_Tablehttp://en.wikipedia.org/wiki/Infrared_Spectroscopy_Correlation_Table -
7/27/2019 basics of ir.doc
8/12
Figure 4: FTIR spectrum of the Benzoic acid dimer recorded at room temperature 1 M CDCl3
solution [9]. The wavelength is often expressed as some unit directly proportional to the
photon energy, such as wavenumber which has a reciprocal relationship to wavelength.
Practical IR spectroscopy
The infrared spectrum of a sample is recorded by passing a beam of infrared light through the
sample. When the frequency of the IR is the same as the vibrational frequency of a bond,
absorption occurs. Examination of the transmitted light reveals how much energy was absorbed
at each frequency (or wavelength). This can be achieved by scanning the wavelength range
using a monochromator. Alternatively, the whole wavelength range is measured at once using
aFourier transforminstrument and then atransmittance orabsorbancespectrum is generated
using a dedicated procedure. Analysis of the position, shape and intensity of peaks in this
spectrum reveals details about the molecular structure of the sample.
This technique works almost exclusively on samples withcovalent bonds. Simple spectra are
obtained from samples with few IR active bonds and high levels of purity. More complex
molecular structures lead to more absorption bands and more complex spectra. The technique
has been used for the characterization of very complex mixtures. Spectra issues with infrared
fluorescence are rare.
Sample preparation
Gaseous samples require a sample cell with a longpathlength to compensate for the diluteness.
The pathlength of the sample cell depends on the concentration of the compound of interest. A
simple glass tube with length of 5 to 10 cm equipped with infrared windows at the both ends of
the tube can be used for concentrations down to several hundred ppm. Sample gas
concentrations well below ppm can be measured with a White's cell in which the infrared light is
guided with mirrors to travel through the gas. White's cells are available with optical pathlength
starting from 0.5 m up to hundred meters.
Liquid samples can be sandwiched between two plates of a salt (commonly sodium chloride, or
common salt, although a number of other salts such aspotassium bromide orcalcium fluoride are
also used).[2]The plates are transparent to the infrared light and do not introduce any lines onto
the spectra.
Solid samples can be prepared in a variety of ways. One common method is to crush the sample
with an oily mulling agent (usually Nujol) in amarble oragatemortar, with a pestle. A thin film of
the mull is smeared onto salt plates and measured. The second method is to grind a quantity of
http://en.wikipedia.org/wiki/Monochromatorhttp://en.wikipedia.org/wiki/Fourier_transformhttp://en.wikipedia.org/wiki/Fourier_transformhttp://en.wikipedia.org/wiki/Transmittancehttp://en.wikipedia.org/wiki/Transmittancehttp://en.wikipedia.org/wiki/Absorbancehttp://en.wikipedia.org/wiki/Absorbancehttp://en.wikipedia.org/wiki/Covalent_bondhttp://en.wikipedia.org/wiki/Covalent_bondhttp://en.wikipedia.org/wiki/Pathlengthhttp://en.wikipedia.org/wiki/Pathlengthhttp://en.wikipedia.org/wiki/White_cell_(spectroscopy)http://en.wikipedia.org/wiki/Sodium_chloridehttp://en.wikipedia.org/wiki/Potassium_bromidehttp://en.wikipedia.org/wiki/Potassium_bromidehttp://en.wikipedia.org/wiki/Calcium_fluoridehttp://en.wikipedia.org/wiki/Calcium_fluoridehttp://en.wikipedia.org/wiki/Infrared_spectroscopy#cite_note-Har-2http://en.wikipedia.org/wiki/Infrared_spectroscopy#cite_note-Har-2http://en.wikipedia.org/wiki/Nujolhttp://en.wikipedia.org/wiki/Marblehttp://en.wikipedia.org/wiki/Marblehttp://en.wikipedia.org/wiki/Agatehttp://en.wikipedia.org/wiki/Mortar_and_pestlehttp://en.wikipedia.org/wiki/Mortar_and_pestlehttp://en.wikipedia.org/wiki/Mortar_and_pestlehttp://en.wikipedia.org/wiki/Monochromatorhttp://en.wikipedia.org/wiki/Fourier_transformhttp://en.wikipedia.org/wiki/Transmittancehttp://en.wikipedia.org/wiki/Absorbancehttp://en.wikipedia.org/wiki/Covalent_bondhttp://en.wikipedia.org/wiki/Pathlengthhttp://en.wikipedia.org/wiki/White_cell_(spectroscopy)http://en.wikipedia.org/wiki/Sodium_chloridehttp://en.wikipedia.org/wiki/Potassium_bromidehttp://en.wikipedia.org/wiki/Calcium_fluoridehttp://en.wikipedia.org/wiki/Infrared_spectroscopy#cite_note-Har-2http://en.wikipedia.org/wiki/Nujolhttp://en.wikipedia.org/wiki/Marblehttp://en.wikipedia.org/wiki/Agatehttp://en.wikipedia.org/wiki/Mortar_and_pestle -
7/27/2019 basics of ir.doc
9/12
the sample with a specially purified salt (usually potassium bromide) finely (to remove scattering
effects from large crystals). This powder mixture is then pressed in a mechanicalpress to form a
translucent pellet through which the beam of the spectrometer can pass. [2] A third technique is the
"cast film" technique, which is used mainly for polymeric materials. The sample is first dissolved
in a suitable, non hygroscopic solvent. A drop of this solution is deposited on surface of KBr or
NaCl cell. The solution is then evaporated to dryness and the film formed on the cell is analyseddirectly. Care is important to ensure that the film is not too thick otherwise light cannot pass
through. This technique is suitable for qualitative analysis. The final method is to
use microtomy to cut a thin (20100 m) film from a solid sample. This is one of the most
important ways of analyzing failed plastic products for example because the integrity of the solid
is preserved.
In photoacoustic spectroscopythe need for sample treatment is minimal. The sample, liquid or
solid, is placed into the sample cup which is inserted into the photoacoustic cell which is then
sealed for the measurement. The sample may be one solid piece, powder or basically in any form
for the measurement. For example, a piece of rock can be inserted into the sample cup and the
spectrum measured from it.
It is important to note that spectra obtained from different sample preparation methods will look
slightly different from each other due to differences in the samples' physical states.
Comparing to a reference
Schematics of a two-beam absorption spectrometer. A beam of infrared light is produced, passed through
an interferometer(not shown), and then split into two separate beams. One is passed through the sample, the
other passed through a reference. The beams are both reflected back towards a detector, however first they pass
through a splitter, which quickly alternates which of the two beams enters the detector. The two signals are then
compared and a printout is obtained. This "two-beam" setup gives accurate spectra even if the intensity of the light
source drifts over time.
To take the infrared spectrum of a sample, it is necessary to measure both the sample and a
"reference" (or "control"). This is because each measurement is affected by not only the light-
absorption properties of the sample, but also the properties of the instrument (for example, what
light source is used, what infrared detectoris used, etc.). The reference measurement makes it
possible to eliminate the instrument influence. Mathematically, the sample transmission spectrum
is divided by the reference transmission spectrum.
http://en.wikipedia.org/wiki/Potassium_bromidehttp://en.wikipedia.org/wiki/Machine_presshttp://en.wikipedia.org/wiki/Machine_presshttp://en.wikipedia.org/wiki/Infrared_spectroscopy#cite_note-Har-2http://en.wikipedia.org/wiki/Microtomyhttp://en.wikipedia.org/wiki/Photoacoustic_spectroscopyhttp://en.wikipedia.org/wiki/Photoacoustic_spectroscopyhttp://en.wikipedia.org/wiki/Interferometerhttp://en.wikipedia.org/wiki/Interferometerhttp://en.wikipedia.org/wiki/Infrared_detectorhttp://en.wikipedia.org/wiki/File:IR_spectroscopy_apparatus.svghttp://en.wikipedia.org/wiki/File:IR_spectroscopy_apparatus.svghttp://en.wikipedia.org/wiki/Potassium_bromidehttp://en.wikipedia.org/wiki/Machine_presshttp://en.wikipedia.org/wiki/Infrared_spectroscopy#cite_note-Har-2http://en.wikipedia.org/wiki/Microtomyhttp://en.wikipedia.org/wiki/Photoacoustic_spectroscopyhttp://en.wikipedia.org/wiki/Interferometerhttp://en.wikipedia.org/wiki/Infrared_detector -
7/27/2019 basics of ir.doc
10/12
The appropriate "reference" depends on the measurement and its goal. The simplest reference
measurement is to simply remove the sample (replacing it by air). However, sometimes a
different reference is more useful. For example, if the sample is a dilute solute dissolved in water
in a beaker, then a good reference measurement might be to measure pure water in the same
beaker. Then the reference measurement would cancel out not only all the instrumental
properties (like what light source is used), but also the light-absorbing and light-reflectingproperties of the water and beaker, and the final result would just show the properties of the
solute (at least approximately).
A common way to compare to a reference is sequentially: first measure the reference, then
replace the reference by the sample and measure the sample. This technique is not perfectly
reliable; if the infrared lamp is a bit brighter during the reference measurement, then a bit dimmer
during the sample measurement, the measurement will be distorted. More elaborate methods,
such as a "two-beam" setup (see figure), can correct for these types of effects to give very
accurate results. The Standard additionmethod can be used to statistically cancel these errors.
Absorption bands
IR spectroscopy is often used to identify structures because functional groups give rise to
characteristic bands both in terms of intensity and position (frequency). The positions of these
bands is summarized in correlation tables as shown below.
Main article: Infrared Spectroscopy Correlation Table
Wavenumbers listed in cm1 .
Badger's Rule
For many kinds of samples, the assignments are known, i.e. which bond deformation(s) are
associated with which frequency. In such cases further information can be gleaned about the
strength on a bond, relying on the empirical guideline called Badger's Rule. This rule states that
the strength of a bond correlates with the frequency of its vibrational mode.
Uses and applications
Infrared spectroscopy is a simple and reliable technique widely used in both organic and
inorganic chemistry, in research and industry. It is used in quality control, dynamic measurement,
and monitoring applications such as the long-term unattended measurement of
CO2 concentrations in greenhouses and growth chambers by infrared gas analyzers.
http://en.wikipedia.org/wiki/Standard_additionhttp://en.wikipedia.org/wiki/Standard_additionhttp://en.wikipedia.org/wiki/Functional_grouphttp://en.wikipedia.org/wiki/Infrared_Spectroscopy_Correlation_Tablehttp://en.wikipedia.org/wiki/Wavenumberhttp://en.wikipedia.org/wiki/Wavenumberhttp://en.wikipedia.org/wiki/Wavenumberhttp://en.wikipedia.org/wiki/File:IR_summary_version_2.gifhttp://en.wikipedia.org/wiki/Standard_additionhttp://en.wikipedia.org/wiki/Functional_grouphttp://en.wikipedia.org/wiki/Infrared_Spectroscopy_Correlation_Tablehttp://en.wikipedia.org/wiki/Wavenumber -
7/27/2019 basics of ir.doc
11/12
It is also used inforensic analysisin both criminal and civil cases, for example in
identifying polymer degradation. It can be used in detecting how much alcohol is in the blood of a
suspected drunk driver measured as 1/10,000 g/mL = 100 g/mL. [citation needed]
A useful way of analysing solid samples without the need for cutting samples uses ATR
orattenuated total reflectancespectroscopy. Using this approach, samples are pressed against
the face of a single crystal. The infrared radiation passes through the crystal and only interacts
with the sample at the interface between the two materials.
With increasing technology in computer filtering and manipulation of the results, samples in
solution can now be measured accurately (water produces a broad absorbance across the range
of interest, and thus renders the spectra unreadable without this computer treatment).
Some instruments will also automatically tell you what substance is being measured from a store
of thousands of reference spectra held in storage.
Infrared spectroscopy is also useful in measuring the degree of polymerization
inpolymermanufacture. Changes in the character or quantity of a particular bond are assessed
by measuring at a specific frequency over time. Modern research instruments can take infraredmeasurements across the range of interest as frequently as 32 times a second. This can be done
whilst simultaneous measurements are made using other techniques. This makes the
observations of chemical reactions and processes quicker and more accurate.
Infrared spectroscopy has also been successfully utilized in the field of semiconductor
microelectronics:[4]for example, infrared spectroscopy can be applied to semiconductors
like silicon, gallium arsenide,gallium nitride,zinc selenide, amorphous silicon,silicon nitride, etc.
The instruments are now small, and can be transported, even for use in field trials.
References
Applications of absorption spectroscopy of organic compounds by JOHN R.DYER
Principles of instrumental analysis by SKOOG &WEST 3Rd EDITION
Google. COM
http://en.wikipedia.org/wiki/Forensic_analysishttp://en.wikipedia.org/wiki/Forensic_analysishttp://en.wikipedia.org/wiki/Forensic_analysishttp://en.wikipedia.org/wiki/Polymer_degradationhttp://en.wikipedia.org/wiki/Wikipedia:Citation_neededhttp://en.wikipedia.org/wiki/Wikipedia:Citation_neededhttp://en.wikipedia.org/wiki/Wikipedia:Citation_neededhttp://en.wikipedia.org/wiki/Attenuated_total_reflectancehttp://en.wikipedia.org/wiki/Attenuated_total_reflectancehttp://en.wikipedia.org/wiki/Polymerhttp://en.wikipedia.org/wiki/Polymerhttp://en.wikipedia.org/wiki/Infrared_spectroscopy#cite_note-4http://en.wikipedia.org/wiki/Infrared_spectroscopy#cite_note-4http://en.wikipedia.org/wiki/Siliconhttp://en.wikipedia.org/wiki/Gallium_arsenidehttp://en.wikipedia.org/wiki/Gallium_nitridehttp://en.wikipedia.org/wiki/Gallium_nitridehttp://en.wikipedia.org/wiki/Gallium_nitridehttp://en.wikipedia.org/wiki/Zinc_selenidehttp://en.wikipedia.org/wiki/Silicon_nitridehttp://en.wikipedia.org/wiki/Silicon_nitridehttp://en.wikipedia.org/wiki/Silicon_nitridehttp://en.wikipedia.org/wiki/Forensic_analysishttp://en.wikipedia.org/wiki/Polymer_degradationhttp://en.wikipedia.org/wiki/Wikipedia:Citation_neededhttp://en.wikipedia.org/wiki/Attenuated_total_reflectancehttp://en.wikipedia.org/wiki/Polymerhttp://en.wikipedia.org/wiki/Infrared_spectroscopy#cite_note-4http://en.wikipedia.org/wiki/Siliconhttp://en.wikipedia.org/wiki/Gallium_arsenidehttp://en.wikipedia.org/wiki/Gallium_nitridehttp://en.wikipedia.org/wiki/Zinc_selenidehttp://en.wikipedia.org/wiki/Silicon_nitride -
7/27/2019 basics of ir.doc
12/12