Basic Pacing Concepts Part III. Sensing Sensing zSensing is the ability of the pacemaker to...
-
Upload
shana-barnett -
Category
Documents
-
view
229 -
download
1
Transcript of Basic Pacing Concepts Part III. Sensing Sensing zSensing is the ability of the pacemaker to...
Basic Pacing ConceptsBasic Pacing ConceptsPart IIIPart III
SensingSensing
SensingSensing
Sensing is the ability of the pacemaker to “see” when a natural (intrinsic) depolarization is occurring
– Pacemakers sense cardiac depolarization by measuring changes in electrical potential of myocardial cells between the anode and cathode
An Electrogram (EGM) is the Recording of Cardiac An Electrogram (EGM) is the Recording of Cardiac Waveforms Taken From Within the HeartWaveforms Taken From Within the Heart
Intrinsic deflection on an EGM occurs when a depolarization wave passes directly under the electrodes
Two characteristics of the EGM are:
– Signal amplitude
– Slew rate
Slew Rate of the EGM Signal Measures the Change in Slew Rate of the EGM Signal Measures the Change in Voltage with Respect to the Change in TimeVoltage with Respect to the Change in Time
The longer the signal takes to move from peak to peak:
– The lower the slew rate
– The flatter the signal
Higher slew rates (number in mV) translate to greater sensing
– Measured in volts per second
Vo
lta
ge
Time
Slope
Slew rate=Change in voltage
Time duration ofvoltage change
A Pacemaker Must Be Able to Sense A Pacemaker Must Be Able to Sense and Respond to Cardiac Rhythmsand Respond to Cardiac Rhythms
Accurate sensing enables the pacemaker to determine whether or not the heart has created a beat on its own
The pacemaker is usually programmed to respond with a pacing impulse only when the heart fails to produce an intrinsic beat
Accurate Sensing...Accurate Sensing...
Ensures that undersensing will not occur –the pacemaker will not miss P or R waves that should have been sensed
Ensures that oversensing will not occur – the pacemaker will not mistake extra-cardiac activity for intrinsic cardiac events
Provides for proper timing of the pacing pulse – an appropriately sensed event resets the timing sequence of the pacemaker
Undersensing . . . Undersensing . . .
Pacemaker does not “see” the intrinsic beat, and therefore does not respond appropriately
Intrinsic beat not sensed
Scheduled pace delivered
VVI / 60
OversensingOversensing
An electrical signal other than the intended P or R wave is detected
Marker channel shows intrinsic
activity...
...though no activity is present
VVI / 60
Sensitivity – The Greater the Number, the Sensitivity – The Greater the Number, the LessLess Sensitive the Device to Intracardiac EventsSensitive the Device to Intracardiac Events
SensitivitySensitivityA
mp
litu
de
(mV
)
Time
5.0
2.5
1.25
SensitivitySensitivityA
mp
litu
de
(mV
)
Time
5.0
2.5
1.25
SensitivitySensitivityA
mp
litu
de
(mV
)
Time
5.0
2.5
1.25
Accurate Sensing Requires That Accurate Sensing Requires That Extraneous Signals Be Filtered OutExtraneous Signals Be Filtered Out
Sensing amplifiers use filters that allow appropriate sensing of P waves and R waves and reject inappropriate signals
Unwanted signals most commonly sensed are:
– T waves
– Far-field events (R waves sensed by the atrial channel)
– Skeletal myopotentials (e.g., pectoral muscle myopotentials)
Accurate Sensing is Dependent on . . .Accurate Sensing is Dependent on . . .
The electrophysiological properties of the myocardium
The characteristics of the electrode and its placement within the heart
The sensing amplifiers of the pacemaker
Factors That May Affect Sensing Are:Factors That May Affect Sensing Are:
Lead polarity (unipolar vs. bipolar)
Lead integrity
– Insulation break
– Wire fracture
EMI – Electromagnetic Interference
Unipolar SensingUnipolar Sensing
Produces a large potential difference due to:
– A cathode and anode that are farther apart than in a bipolar system
_
Bipolar SensingBipolar Sensing
Produces a smaller potential difference due to the short interelectrode distance
– Electrical signals from outside the heart such as myopotentials are less likely to be sensed
An Insulation Break May Cause Both An Insulation Break May Cause Both Undersensing or OversensingUndersensing or Oversensing
Undersensing occurs when inner and outer conductor coils are in continuos contact
– Signals from intrinsic beats are reduced at the sense amplifier and amplitude no longer meets the programmed sensing value
Oversensing occurs when inner and outer conductor coils make intermittent contact
– Signals are incorrectly interpreted as P or R waves
A Wire Fracture Can Cause Both A Wire Fracture Can Cause Both Undersensing and OversensingUndersensing and Oversensing
Undersensing occurs when the cardiac signal is unable to get back to the pacemaker – intrinsic signals cannot cross the wire fracture
Oversensing occurs when the severed ends of the wire intermittently make contact, which creates potentials interpreted by the pacemaker as P or R waves
Electromagnetic InterferenceElectromagnetic Interference
Electromagnetic Interference (EMI)Electromagnetic Interference (EMI)
Interference is caused by electromagnetic energy with a source that is outside the body
Electromagnetic fields that may affect pacemakers are radio-frequency waves
– 50-60 Hz are most frequently associated with pacemaker interference
Few sources of EMI are found in the home or office but several exist in hospitals
EMI May Result in the Following Problems:EMI May Result in the Following Problems:
Oversensing
Transient mode change (noise reversion)
Reprogramming (Power on Reset or “POR”)
Oversensing May Occur When EMI Signals Are Oversensing May Occur When EMI Signals Are Incorrectly Interpreted as P Waves or R WavesIncorrectly Interpreted as P Waves or R Waves
Pacing rates will vary as a result of EMI:
– Rates will accelerate if sensed as P waves in dual-chamber systems (P waves are “tracked”)
– Rates will be low or inhibited if sensed in single-chamber systems, or on ventricular lead in dual-chamber systems
EMIEMI
“Noise” sensed by the pacemaker
Should have paced
Noise ReversionNoise Reversion
VPVPSRSR SR SR
Noise Sensed
Lower Rate Interval
VVI/60
Continuous refractory sensing will cause pacing at the lower or sensor driven rate
EMI May Lead to Inadvertent Reprogramming EMI May Lead to Inadvertent Reprogramming of the Pacing Parametersof the Pacing Parameters
Device will revert to Power on Reset (POR or “backup” mode)
– Power on Reset may exhibit a mode and rate change which are often the same as ERI
– In some cases, reprogrammed parameters may be permanent
New technologies will continue to create new, unanticipated sources of EMI:
Cellular phones (digital)
Sources of EMI Are Found Most Sources of EMI Are Found Most Commonly in Hospital EnvironmentsCommonly in Hospital Environments
Sources of EMI that interfere with pacemaker operation include surgical/therapeutic equipment such as:
– Electrocautery
– Transthoracic defibrillation
– Extracorporeal shock-wave lithotripsy
– Therapeutic radiation
– RF ablation
– TENS units
– MRI
Sources of EMI Are Found More Rarely in:Sources of EMI Are Found More Rarely in:
Home, office, and shopping environments
Industrial environments with very high electrical outputs
Transportation systems with high electrical energy exposure or with high-powered radar and radio transmission
– Engines or subway braking systems
– Airport radar
– Airplane engines
TV and radio transmission sites
Electrocautery is the Most Common Electrocautery is the Most Common Hospital Source of Pacemaker EMIHospital Source of Pacemaker EMI
Outcomes
– Oversensing–inhibition
– Undersensing (noise reversion)
– Power on Reset
– Permanent loss of pacemaker output (if battery voltage is low)
Precautions
– Reprogram mode to VOO/DOO, or place a magnet over device
– Strategically place the grounding plate
– Limit electrocautery bursts to 1-second burst every 10 seconds
– Use bipolar electrocautery forceps
Transthoracic Defibrillation Transthoracic Defibrillation
Outcome
– Inappropriate reprogramming of the pulse generator (POR)
– Damage to pacemaker circuitry
Precautions
– Position defibrillation paddles apex-posterior (AP) and as far from the pacemaker and leads as possible
Magnetic Resonance Imaging (MRI) is Generally Magnetic Resonance Imaging (MRI) is Generally Contraindicated in Patients with PacemakersContraindicated in Patients with Pacemakers
Outcomes
– Extremely high pacing rate
– Reversion to asynchronous pacing
Precautions
– Program pacemaker output low enough to create persistentnon-capture, ODO or OVO mode
Lithotripsy Shock Waves May Have an Lithotripsy Shock Waves May Have an Effect on Pacemaker SystemsEffect on Pacemaker Systems
Outcomes indual-chamber modes:
– Inhibition of ventricular pacing
Outcomes in rate adaptive pacemakers
– High pacing rates
– Piezoelectric crystal damage
Precautions:
– Program pacemaker to VVI or VOO mode
– Lithotriptor focal point should be greater than 6 inches from the pacemaker
– Carefully monitor heart function throughout procedure
Radiation Energy May Cause Radiation Energy May Cause Permanent Damage Permanent Damage
Certain kinds of radiation energy may cause damage to the semi-conductor circuitry
– Ionizing radiation used for breast or lung cancer therapy
Damage can be permanent and requires replacement of the pacemaker
Therapeutic Radiation May Therapeutic Radiation May Cause Severe DamageCause Severe Damage
Outcomes:
– Pacemaker circuit damage
– Loss of output
– “Runaway”
Precautions:
– Keep cumulative radiation absorbed by the pacemaker to less than 500 rads; shielding may be required
– Check pacemaker after radiation sessions for changes in pacemaker function (can be done transtelephonically)
Pacemaker Features That Pacemaker Features That Address Interference Address Interference
InputBandpass
filterAbsolute
valueReversion
circuitLevel
detector
Sensitivity adjustment
Pacemakerlogic
Pacemaker sensing circuits amplify, filter and either process or reject incoming signals
Rate Responsive PacingRate Responsive Pacing
Rate ResponseRate Response
Rate responsive (also called rate modulated) pacemakers provide patients with the ability to vary heart rate when the sinus node cannot provide the appropriate rate
Rate responsive pacing is indicated for:
– Patients who are chronotropically incompetent (heart rate cannot reach appropriate levels during exercise or to meet other metabolic demands)
– Patients in chronic atrial fibrillation with slow ventricular response
Rate Responsive PacingRate Responsive Pacing
Cardiac output (CO) is determined by the combination of stroke volume (SV) and heart rate (HR)
SV X HR = CO
Changes in cardiac output depend on the ability of the HR and SV to respond to metabolic requirements
Rate Responsive PacingRate Responsive Pacing
SV reserves can account for increases in cardiac output of up to 50%
HR reserves can nearly triple total cardiac output in response to metabolic demands
Rate Responsive PacingRate Responsive Pacing
When the need for oxygenated blood increases, the pacemaker ensures that the heart rate increases to provide additional cardiac output
Adjusting Heart Rate to Activity
Normal Heart Rate
Rate Responsive PacingFixed-Rate Pacing
Daily Activities
A Variety of Rate Response Sensors ExistA Variety of Rate Response Sensors Exist
Those most accepted in the market place are:
– Activity sensors that detect physical movement and increase the rate according to the level of activity
– Minute ventilation sensors that measure the change in respiration rate and tidal volume via transthoracic impedance readings
Rate Responsive PacingRate Responsive Pacing
Activity sensors employ a piezoelectric crystal that detects mechanical signals produced by movement
The crystal translates the mechanical signals into electrical signals that in turn increase the rate of the pacemaker
Piezoelectric crystal
Rate Responsive PacingRate Responsive Pacing
Minute Ventilation (MV) is the volume of air introduced into the lungs per unit of time
MV has two components:
– Tidal volume–the volume of air introduced into the lungs in a single respiration cycle
– Respiration rate–the number of respiration cycles per minute
Rate Responsive PacingRate Responsive Pacing
Minute ventilation can be measured by measuring the changes in electrical impedance across the chest cavity to calculate changes in lung volume over time
Pacing systems
Electrical concepts
Stimulation thresholds
Sensing
Electromagnetic Interference (EMI)
Rate response
Summary of Basic Pacing Concepts Summary of Basic Pacing Concepts ModuleModule
General Medtronic Pacemaker DisclaimerINDICATIONS
Medtronic pacemakers are indicated for rate adaptive pacing in patients who may benefit from increased pacing rates concurrent with increases in activity (Thera, Thera-i, Prodigy, Preva and Medtronic.Kappa 700 Series) or increases in activity and/or minute ventilation (Medtronic.Kappa 400 Series).
Medtronic pacemakers are also indicated for dual chamber and atrial tracking modes in patients who may benefit from maintenance of AV synchrony. Dual chamber modes are specifically indicated for treatment of conduction disorders that require restoration of both rate and AV synchrony, which include various degrees of AV block to maintain the atrial contribution to cardiac output and VVI intolerance (e.g., pacemaker syndrome) in the presence of persistent sinus rhythm.
9790 Programmer
The Medtronic 9790 Programmers are portable, microprocessor based instruments used to program Medtronic implantable devices.
9462
The Model 9462 Remote Assistant™ is intended for use in combination with a Medtronic implantable pacemaker with Remote Assistant diagnostic capabilities.
CONTRAINDICATIONS
Medtronic pacemakers are contraindicated for the following applications:
Dual chamber atrial pacing in patients with chronic refractory atrial tachyarrhythmias.
Asynchronous pacing in the presence (or likelihood) of competitive paced and intrinsic rhythms.
Unipolar pacing for patients with an implanted cardioverter-defibrillator because it may cause unwanted delivery or inhibition of ICD therapy.
Medtronic.Kappa 400 Series pacemakers are contraindicated for use with epicardial leads and with abdominal implantation.
WARNINGS/PRECAUTIONS
Pacemaker patients should avoid sources of magnetic resonance imaging, diathermy, high sources of radiation, electrosurgical cautery, external defibrillation, lithotripsy, and radiofrequency ablation to avoid electrical reset of the device, inappropriate sensing and/or therapy.
9462
Operation of the Model 9462 Remote Assistant™ Cardiac Monitor near sources of electromagnetic interference, such as cellular phones, computer monitors, etc. may adversely affect the performance of this device.
See the appropriate technical manual for detailed information regarding indications, contraindications, warnings, and precautions.
Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician.
Medtronic Leads
For Indications, Contraindications, Warnings, and Precautions for Medtronic Leads, please refer to the appropriate Leads Technical Manual or call your local Medtronic Representative.
Caution: Federal law restricts this device to sale by or on the order of a Physician.
Note:
This presentation is provided for general educational purposes only and should not be considered the exclusive source for this type of information. At all times, it is the professional responsibility of the practitioner to exercise independent clinical judgment in a particular situation.