Automatic Quality Assessment of Whole Slide Images

download Automatic Quality Assessment of Whole Slide Images

of 14

Transcript of Automatic Quality Assessment of Whole Slide Images

  • 8/11/2019 Automatic Quality Assessment of Whole Slide Images

    1/14

    Towards Better Digital Pathology Workflows:

    Programming libraries for high-speedsharpness assessment of Whole Slide Images

    David Ameisen - [email protected]

    mailto:[email protected]:[email protected]
  • 8/11/2019 Automatic Quality Assessment of Whole Slide Images

    2/14

    Quality assessment in WSI:

    a key step towards systematic use WSI scanners - FDA Class III Medical Devices :

    13% of Whole Slide Images need to be scanned at least twice.Campbell WS et al., Human Path, 2012;

    Concordance between whole-slide imaging and light microscopy for routine surgical pathology.

    A posteriori visual quality verification is insufficient: partial verification.

    Automatic Quality Assurance tools are not adapted to high dimensions (>10 9 pixels).

    Our perspective: Make WSI Quality Assessment tools fast enough for any pathology workflow.

  • 8/11/2019 Automatic Quality Assessment of Whole Slide Images

    3/14

    7 years ofresearch in

    WSIcollaborations

    Quality

    Assurance?

    Major problem:

    Blur

    Research

    Teaching

    Clinical

    Manual / Visual

    Non-systematic

    Non-quantified

    Towards awidespread use

    of WSI inpathology

    Background

  • 8/11/2019 Automatic Quality Assessment of Whole Slide Images

    4/14

    State of the artCurrent methods and limits of automated image quality assurance

    No Fast Blur Detection and Quantification

    Solution for Whole Slide Images

    Auto-focus One shot

    Comparison Gold standard

    No reference Too slow

    Z-stack Too big

  • 8/11/2019 Automatic Quality Assessment of Whole Slide Images

    5/14

    Sharp; partially blurred; out of focus Absolute

    Blur/SharpnessQuantification

    Highly scalable, multi-core, multi-node

    Compatibility: Windows, Mac OSX, Linux

    No gold standard required

    A Fast Blur Detection and QuantificationSolution for Whole Slide Images

    3 billion pixels analyzed per minute on a recent computer(10 to 100 times faster than other methods)

  • 8/11/2019 Automatic Quality Assessment of Whole Slide Images

    6/14

    6

    Whole Slide Imagechecked at all magnifications

  • 8/11/2019 Automatic Quality Assessment of Whole Slide Images

    7/14

    20x : 1.7 Billion pixels6589 tiles of 512x512px24% relevant tiles (non-blank)amon which 25% are

  • 8/11/2019 Automatic Quality Assessment of Whole Slide Images

    8/14

    Video : real-time

    analysis

    https://www.youtube.com/watch?v=eIunSx-a-ug

    https://www.youtube.com/watch?v=eIunSx-a-ughttps://www.youtube.com/watch?v=eIunSx-a-ug
  • 8/11/2019 Automatic Quality Assessment of Whole Slide Images

    9/14

    5000 images analyzed in our labInternet survey:

    100 images, 7500+ answers

  • 8/11/2019 Automatic Quality Assessment of Whole Slide Images

    10/14

    Correlation algorithm/user votesImage classification into three groups:Out of focus, Partially blurred, Sharp

  • 8/11/2019 Automatic Quality Assessment of Whole Slide Images

    11/14

    DuringWSI

    digitization

    Applications :Saving time, increasing the quality

    Identification and reacquisition of blurred tiles during thescanCertification of the WSIs general image quality

    Just afterWSI

    digitization

    Reacquisition of the WSI with more suitable profiles

    Notification sent to reacquire the WSI

    ContinuousQuality

    Assusrance

    Notification sent to recalibrate the WSI acquisition system

    WSI quality scoring integrated in the files metadata

  • 8/11/2019 Automatic Quality Assessment of Whole Slide Images

    12/14

    C++Library

    ImplementationsScaling

    Optimized, multi-core, multi-node

    Highly scalable to match any throughput

    APIsC++Java

    Python

    Embeddable in Whole Slide Image acquisition systems

    Embeddable in Whole Slide Image management platforms

    Embeddable in Whole Slide Image viewers

  • 8/11/2019 Automatic Quality Assessment of Whole Slide Images

    13/14

    Our references

    Talk, 11th European Congress on Telepathology5th International Congress on Virtual Microscopy6-9 June 2012, Venice, Italy

    Poster, Pathology Visions Conference ,October, 28-31, 2012, Baltimore, MA, USA.

    Talk, 12th European Congress on Digital Pathology18-21 June 2014, Paris, France

    FlexMIm - Projet R&D - 4 989 K Fonds Unique Interministriel n 142013-2015, Paris, France.

    Talk, Pathology Visions Conference ,October, 20-21, 2014, San Francisco, CA, USA.

    pre-seed funding

  • 8/11/2019 Automatic Quality Assessment of Whole Slide Images

    14/14

    Contact :

    Laboratory : David Ameisen - [email protected] Idfinnov : Robert Marino [email protected]

    mailto:[email protected]:[email protected]:[email protected]:[email protected]