Autoclave

12

description

An autoclave is a device used to sterilize equipment and supplies by subjecting them to high pressure saturated steam at 121 °C for around 15–20 minutes depending on the size of the load and the contents. It was invented by Charles Chamberland in 1879, although a precursor known as the steam digester was created by Denis Papin in 1679. The name comes from Greek auto-, ultimately meaning self, and Latin clavis meaning key—a self-locking device.

Transcript of Autoclave

Page 1: Autoclave
Page 2: Autoclave

An autoclave is a device used to sterilize equipment and supplies by subjecting them to high pressure saturated steam at 121 °C for around 15–20 minutes depending on the size of the load and the contents. It was invented by Charles Chamberland in 1879, although a precursor known as the steam digester was created by Denis Papin in 1679.

Page 3: Autoclave

Glassware:

Cap Pyrex bottles loosely-fit whether empty or filled- to prevent explosions

Cover bottles that are not made of safety glass with aluminum foil.

Polypropylene or stainless steel tubs are typically used for secondary containment.

Cap Pyrex Bottles

Stainless steel tubs

Page 4: Autoclave

1. Plug into a

standard

electric outlet.

Page 5: Autoclave

2. Pour water

(distilled water is

best) to the

chamber to flow

over its gap on the

bottom plate.

Page 6: Autoclave

3.Turn exhaust

knob on the

handle

clockwise

tightly and

close DOOR.

Cover it tightly.

Page 7: Autoclave

4. Set timer at 35

minutes

position and

make sure

STERILIZE

light is on.

(sterilization

will start)

Page 8: Autoclave

5. Upon completion of

sterilization power will be off

automatically.

Page 9: Autoclave

6. Release EXHAUST KNOB on the handle by turning 4-5 revolution until PRESSURE GAUGE, is zero returned. Then you can open DOOR COVER. (after PRESSURE GAUGE) is zero returned, wait 5-10 minutes before opening. (DOOR COVER for better sterilization).

Page 10: Autoclave

Caution1. Do not open DOOR

COVER if PRESSURE GAUGE is not yet zero returned.

2. Do not release EXHAUST KNOB on the handle during sterilization. Place Autoclave on a flat position .

3. Use distilled water as the sterilization water. (Do not use any chemical medication)

Disadvantages1. It requires more

electricity and more time to prepare and finish.

Page 11: Autoclave

Wear personal protection equipment: Lab coat

Eye protection

Closed-toe shoes

Heat-resistant gloves to remove items, especially hot glassware

Wait for the pressure gauge to drop to zero with zero time remaining before opening the door.

Never open an autoclave set for "slow exhaust" until the cycle is complete. Superheated liquids can boil over and damage the autoclave and the operator.

Open the door cautiously. Stand behind the door and slowly open it. Allow all steam to escape before reaching inside.

Let liquids stand another 10–20 minutes after the autoclave is opened to avoid any movement that could cause them to boil. Remove items carefully.

Page 12: Autoclave