Atomic Emission Spectroscopy

99
1 Atomic Emission Spectroscopy Lecture 21

description

Atomic Emission Spectroscopy. Lecture 21. Qualitative analysis is accomplished by comparison of the wavelengths of some emission lines to standards while the line blackness serves as the tool for semiquantitative analysis. - PowerPoint PPT Presentation

Transcript of Atomic Emission Spectroscopy

Page 1: Atomic Emission Spectroscopy

1

Atomic Emission Spectroscopy

Lecture 21

Page 2: Atomic Emission Spectroscopy

2

Qualitative analysis is accomplished by comparison of the wavelengths of some emission lines to standards while the line blackness serves as the tool for semiquantitative analysis.

Polychromators are also available as multichannel arc and spark instruments. However, these have fixed slits at certain wavelengths in order to do certain elements and thus they are not versatile.

Page 3: Atomic Emission Spectroscopy

3

Grating

DetectorsPotential Source

Page 4: Atomic Emission Spectroscopy

4

Recently, arc and spark instruments based on charge injection and charge coupled devices became available. These have extraordinarily high efficiency and performance in terms of easier calibration, short analysis time, as well as superior quantitative results.

Page 5: Atomic Emission Spectroscopy

5

Grating

CCD or CID DetectorPotential Source

Page 6: Atomic Emission Spectroscopy

6

Page 7: Atomic Emission Spectroscopy

7

Characteristics of Arc Sources1. Typical temperatures between 4000-5000 oC are high

enough to cause atomization and excitation of sample and electrode materials.

2. Usually, cyanogens compounds are formed due to reaction of graphite electrodes with atmospheric nitrogen. Emission bands from cyanogens compounds occur in the region from 350-420 nm. Unfortunately, several elements have their most sensitive lines in this same region which limits the technique. However, use of controlled atmosphere around the arc (using CO2, Helium, or argon) very much decreases the effect of cyanogens emission.

Page 8: Atomic Emission Spectroscopy

8

3. The emission signal should be integrated over a minute or so since volatilization and excitation of atoms of different species differ widely. While some species give maximum signal, others may still be in the molecular state.

4. Arc sources are very good for qualitative analysis of elements while only semiquantitative analysis is possible. It is mandatory to compare the emission spectrum of a sample with the emission spectrum of a standard. In some cases, a few milligrams of a standard is added to the sample in order to locate the emission lines of the standard and thus identify the emission wavelengths of the different elements in the sample. A comparator densitometer can be used to exactly locate the wavelengths of the standard and the sample components.

Page 9: Atomic Emission Spectroscopy

9

Standard

Sample

The lines from the standard are projected on the lines of the combined sample/standard emission spectra in order to identify sample components. Only few lines are shown in the figure.

Page 10: Atomic Emission Spectroscopy

10

Page 11: Atomic Emission Spectroscopy

11

Page 12: Atomic Emission Spectroscopy

12

Page 13: Atomic Emission Spectroscopy

13

Why use Carbon in Atomic Spectroscopy?

We have previously seen the use of graphite in electrothermal AAS as well as arc and spark AES, even though molecular spectra are real problems in both techniques due to cyanogens compounds absorption and emission. The reasons after graphite common use in atomic spectroscopy can be summarized below:

Page 14: Atomic Emission Spectroscopy

14

1. It is conductive.2. It can be obtained in a very pure state.3. Easily available and cheap.4. Thermally stable and inert.5. Carbon has few emission lines.6. Easily shaped.

Page 15: Atomic Emission Spectroscopy

15

Spark SourcesMost of the instruments in this category are arc

based instruments. Spark based instruments are of the same idea except for a spark source substituting an arc source. The spark source is constructed as in the figure below where an AC potential in the order of 10-50 KV is discharged through a capacitor which is charged and discharged through the graphite electrodes about 120 times/s; resulting in a discharge current of about 1000 A.

Page 16: Atomic Emission Spectroscopy

16

This very high current will suffer a great deal of resistance, which increase the temperature to an estimated 40000 oC. Therefore, ionic spectra are more pronounced.

Transformer

Potential Source

Page 17: Atomic Emission Spectroscopy

17

An introduction to Ultraviolet/Visible Absorption Spectroscopy

Chapter 13

Page 18: Atomic Emission Spectroscopy

18

In this chapter, absorption by molecules, rather than atoms, is considered. Absorption in the ultraviolet and visible regions occurs due to electronic transitions from the ground state to excited state. Broad band spectra are obtained since molecules have vibrational and rotational energy levels associated with electronic energy levels. The signal is either absorbance or percent transmittance of the analyte solution where:

Page 19: Atomic Emission Spectroscopy

19

Absorption measurements based upon ultraviolet and visible radiation find widespread application for the quantitative determination of a large variety species.

Beer’s Law:A = -logT = logP0/P = bc

A = absorbance = molar absorptivity [M-1 cm-1]c = concentration [M]P0 = incident powerP = transmitted power (after passing through sample)

Page 20: Atomic Emission Spectroscopy

20

Page 21: Atomic Emission Spectroscopy

21

Page 22: Atomic Emission Spectroscopy

22

UV-Vis Absorption Spectroscopy

Lecture 22

Page 23: Atomic Emission Spectroscopy

23

Page 24: Atomic Emission Spectroscopy

24

Measurement of Transmittance and Absorbance:The power of the beam transmitted by the analyte solution is usually compared with the power of the beam transmitted by an identical cell containing only solvent. An experimental transmittance and absorbance are then obtained with the equations.

P0 and P refers to the power of radiation after it has passed through the solvent and the analyte.

Page 25: Atomic Emission Spectroscopy

25

Beer’s law and mixturesBeer’s law and mixtures• Each analyte present in the solution absorbs Each analyte present in the solution absorbs

light! light! • The magnitude of the absorption depends on The magnitude of the absorption depends on

its its • A A total total = A= A11+A+A22+…+A+…+An n

• A A total total = = 11bcbc11++22bcbc22+…++…+nnbcbcnn

• If If 1 1 = = 2 2 == n n then simultaneous determination is then simultaneous determination is impossibleimpossible

• Need to measure A at nNeed to measure A at n’s (get n’s (get n22 ’s’s) to solve ) to solve for the concentration of species in the mixturefor the concentration of species in the mixture

Page 26: Atomic Emission Spectroscopy

26

Page 27: Atomic Emission Spectroscopy

27

Limitations to Beer’s Law

• Real limitations• Chemical deviations• Instrumental deviations

Page 28: Atomic Emission Spectroscopy

28

1.      Real Limitations a.       Beer’s law is good for dilute analyte

solutions only. High concentrations (>0.01M) will cause a negative error since as the distance between molecules become smaller the charge distribution will be affected which alter the molecules ability to absorb a specific wavelength. The same phenomenon is also observed for solutions with high electrolyte concentration, even at low analyte concentration. The molar absorptivity is altered due to electrostatic interactions.

Page 29: Atomic Emission Spectroscopy

29

b.      In the derivation of Beer’s law we have introduced a constant (). However, is dependent on the refractive index and the refractive index is a function of concentration. Therefore, will be concentration dependent. However, the refractive index changes very slightly for dilute solutions and thus we can practically assume that is constant.

c.       In rare cases, the molar absorptivity changes widely with concentration, even at dilute solutions. Therefore, Beer’s law is never a linear relation for such compounds, like methylene blue.

Page 30: Atomic Emission Spectroscopy

30

2.      Chemical Deviations This factor is an important one which

largely affects linearity in Beer’s law. It originates when an analyte dissociates, associates, or reacts in the solvent, or one of matrix constituents. For example, an acid base indicator when dissolved in water will partially dissociate according to its acid dissociation constant:

Page 31: Atomic Emission Spectroscopy

31

HIn H+ + In-

It can be easily appreciated that the amount of HIn present in solution is less than that originally dissolved where:

CHIn = [HIn] + [In-]

Assume an analytical concentration of 2x10-5 M indicator (ka = 1.42x10-5) was used, we may write:

Page 32: Atomic Emission Spectroscopy

32

1.42x10-5 = x2/(2x10-5 – x)Solving the quadratic equation gives:X = 1.12x10-5 M which means:[In-] = 1.12x10-5 M[HIn] = 2x10-5 – 1.12x10-5 = 0.88x10-5 MTherefore, the absorbance measured will be the sum

of that for HIn and In-. If a 1.00 cm cell was used and the for both HIn and In- were 7.12x103 and 9.61x102 Lmol-1cm-1 at 570 nm, respectively, the absorbance of the solution can be calculated:

Page 33: Atomic Emission Spectroscopy

33

A = AHIn + AIn

A = 7.12x103 * 1.00* 0.88x10-5 + 9.61x102 * 1.00 *1.12x10-5 = 0.073

However, if no dissociation takes place we may have:A = AHIn

A = 7.12x103 * 1.00 * 2x10-5 = 0.142 If the two results are compared we can calculate the

% decrease in anticipated signal as:% decrease in signal = {(0.142 – 0.073)/0.142}x100%

= 49%

Page 34: Atomic Emission Spectroscopy

34

However, at 430 nm, the molar absorptivities of HIn and In- are 6.30*102 and 2.06*104, respectively.

A = AHIn + AIn

A = 6.30*102 * 1.00* 0.88x10-5 + 2.06*104 * 1.00 *1.12x10-5 = 0.236

Again, if no dissociation takes place we may have:A = AHIn

A = 6.30*102 * 1.00 * 2x10-5 = 0.013 If the two results are compared we can calculate the %

increase in anticipated signal as:% decrease in signal = {(0.236 – 0.013)/0.013}x100% = V.

large

Page 35: Atomic Emission Spectroscopy

35

Comparison between results obtained at 570 nm and 430 nm show large dependence on the values of the molar absorptivities of HIn and In- at these wavelength.

At 570 nm:A = AHIn + AIn

A = 7.12x103 * 1.00* 0.88x10-5 + 9.61x102 * 1.00 *1.12x10-5 = 0.073

And at 430 nm:A = 6.30*102 * 1.00* 0.88x10-5 + 2.06*104 * 1.00

*1.12x10-5 = 0.236

Page 36: Atomic Emission Spectroscopy

36

Chemical deviations from Beer’s law for unbuffered solutions of the indicator Hln.

Note that there are positive deviations at 430 nm and negative deviations at 570 nm. At 430 nm, the absorbance is primarily due to the ionized In- form of the indicator and is proportional to the fraction ionized, which varies nonlinearly with the total indicator concentration. At 570 nm, the absorbance is due principally to the undissociated acid Hln, which increases nonlinearly with the total concentration.

Page 37: Atomic Emission Spectroscopy

37

Page 38: Atomic Emission Spectroscopy

38

Calculated Absorbance Data for Various Indicator Concentrations

Page 39: Atomic Emission Spectroscopy

39

An example of association equilibria Association of chromate in acidic solution to form

the dichromate according to the equation below: 

2 CrO42- + 2 H+ Cr2O7

2- + H2O 

The absorbance of the chromate ions will change according to the mentioned equilibrium and will thus be nonlinearly related to concentration.

A = CrO4 *b*CCrO4 + Cr2O7 *b*CCr2O7

Page 40: Atomic Emission Spectroscopy

40

3.      Instrumental Deviations a.       Beer’s law is good for monochromatic

light only since is wavelength dependent. It is enough to assume a dichromatic beam passing through a sample to appreciate the need for a monochromatic light. Assume that the radiant power of incident radiation is Po and Po’ while transmitted power is P and P’. The absorbance of solution can be written as:

Page 41: Atomic Emission Spectroscopy

41

A = log (Po + Po’)/(P + P’)

P = Po10-bc, substituting in the above equation:

A = log (Po + Po’)/(Po10-bc Po’10-’bc)

Assume = ’ = A = log (Po + Po’)/(Po + Po’) 10-bc

A = bc

However, since ’ # , since is wavelength dependent, then A # bc

Page 42: Atomic Emission Spectroscopy

42

Page 43: Atomic Emission Spectroscopy

43

The effect of polychromatic radiation on Beer’s law.

In the spectrum at the top, the molar absorptivity of the analyte is nearly constant over band A. Note that in Beer’s law plot at the bottom, using band A gives a linear relationship. In the spectrum, band B corresponds to a region where the absorptivity shows substantial changes. In the lower plot, note the dramatic deviation from Beer’s law that results.

Page 44: Atomic Emission Spectroscopy

44

Page 45: Atomic Emission Spectroscopy

45

Therefore, the linearity between absorbance and concentration breaks down if incident radiation was polychromatic. In most cases with UV-Vis spectroscopy, the effect is small especially at the wavelength maximum. The small changes in signal is insignificant since differs only slightly.

Page 46: Atomic Emission Spectroscopy

46

UV-Vis Absorption Spectroscopy

Lecture 23

Page 47: Atomic Emission Spectroscopy

47

b.      Stray Radiation Stray radiation resulting from scattering or

various reflections in the instrument will reach the detector without passing through the sample. The problem can be severe in cases of high absorbance or when the wavelengths of stray radiation is in such a range where the detector is highly sensitive as well as at wavelengths extremes of an instrument. The absorbance recorded can be represented by the relation:

A = log (Po + Ps)/(P + Ps)Where; Ps is the radiant power of stray radiation.

Page 48: Atomic Emission Spectroscopy

48

Page 49: Atomic Emission Spectroscopy

49

Instrumental Noise as a Function in Transmittance

The uncertainty in concentration as a function of the uncertainty in transmittance can be statistically represented as:

 sc

2 = (dc/dT)2 sT2

 A = -log T = bc = -0.434 ln Tc = -(1/b)*0.434 ln T (1)dc/dt = - 0.434/bT sc

2 = (-0.434/bT)2 sT2 (2)

Page 50: Atomic Emission Spectroscopy

50

Dividing equation 2 by the square of equation 1 (sc

/c)2 = (-0.434/bT)2 sT2/{(-0.434 ln T)2/(b)2}

sc/c = (sT / T ln T)

 Therefore, it is clear that the uncertainty in

concentration of a sample is nonlinearly related to the magnitude of the transmittance. Substitution for different values of transmittance and assuming sT is constant, we get:

Page 51: Atomic Emission Spectroscopy

51

Page 52: Atomic Emission Spectroscopy

52

Page 53: Atomic Emission Spectroscopy

53

Therefore, an absorbance between 0.2-0.7 may be advantageous in terms of a lower uncertainty in concentration measurements. At higher or lower absorbances, an increase in uncertainty is encountered. It is therefore advised that the test solution be in the concentration range which gives an absorbance value in the range from 0.2-0.7 for best precision.

However, it should also be remembered that we ended up with this conclusion provided that sT is constant. Unfortunately, sT is not always constant which complicates the conclusions above.

Page 54: Atomic Emission Spectroscopy

54

Effect of bandwidth on spectral detail for a sample of benzene vapor. Note that as the spectral bandwidth increases, the fine structure in the spectrum is lost. At a bandwidth of 10 nm, only a broad absorption band is observed.

EFFECT OF bandwidth

Page 55: Atomic Emission Spectroscopy

55

Effect of slit width (spectral bandwidth) on peak heights. Here, the sample was s solution of praseodymium chloride. Note that as the spectral bandwidth decreases by decreasing the slit width from 1.0 mm to 0.1 mm, the peak heights increase.

Page 56: Atomic Emission Spectroscopy

56

Effect of Scattered Radiation at Wavelength Extremes of an Instrument

Wavelength extremes of an instrument are dependent on type of source, detector and optical components used in the manufacture of the instrument. Outside the working range of the instrument, it is not possible to use it for accurate determinations. However, the extremes of the instrument are very close to the region of invalid instrumental performance and would thus be not very accurate. An example may be a visible photometer which, in principle, can be used in the range from 340-780 nm. It may be obvious that glass windows, cells and prism will start to absorb significantly below 380 nm and thus a decrease in the incident radiant power is significant.

Page 57: Atomic Emission Spectroscopy

57

What defines the instrumental wavelength extremes?

Three main Factors:1. Source2. Detector3. Optical components (lenses, windows, etc)Measurements at wavelength extremes should

be avoided since errors are very possible due to:

1. Source limitations2. Detector limitations3. Sample cell limitations4. Scattered radiation

Page 58: Atomic Emission Spectroscopy

58

Spectrum of cerium (IV) obtained with a spectrophotometer having glass optics (A) and quartz optics (B). The false peak in A arises from transmission of stray radiation of longer wavelengths.

B: UV-VIS spectrophotometerA: VIS spectrophotometer EFFECT OF SCATTERED

RADIATION

Page 59: Atomic Emission Spectroscopy

59

The output from the source at the low wavelength range is minimal. Also, the detector has best sensitivities around 550 nm which means that away up and down this value, the sensitivity significantly decrease. However, scattered radiation, and stray radiation in general, will reach the detector without passing through these surfaces as well as these radiation are constituted from wavelengths for which the detector is highly sensitive. In some cases, stray and scattered radiation reaching the detector can be far more intense than the monochromatic beam from the source. False peaks may appear in such cases and one should be aware of this cause of such peaks.

Page 60: Atomic Emission Spectroscopy

60

Instrumentation• Light source - selector• Sample container • Detector• Signal processing• Light Sources (commercial

instruments)– D2 lamp (UV: 160 – 375 nm)– W lamp (vis: 350 – 2500 nm)

Page 61: Atomic Emission Spectroscopy

61

SourcesDeuterium and hydrogen lamps (160 – 375 nm)

DD22 + E + Eee → D → D22* → D’ + D’’ + * → D’ + D’’ + hh

Page 62: Atomic Emission Spectroscopy

62

(a) A deuterium lamp of the type used in spectrophotometers and (b) its spectrum. The plot is of irradiance Eλ (proportional to radiant power) versus wavelength. Note that the maximum intensity occurs at ~225 m.Typically, instruments switch from deuterium to tungsten at ~350 nm.

Deuterium lampUV region

Page 63: Atomic Emission Spectroscopy

63

(a) A tungsten lamp of the type used in spectroscopy and its spectrum

(b). Intensity of the

tungsten source is usually quite low at wavelengths shorter than about 350 nm. Note that the intensity reaches a maximum in the near-IR region of the spectrum.

Visible and near-IR region

Page 64: Atomic Emission Spectroscopy

64

The tungsten lamp is by far the most common source in the visible and near IR region with a continuum output wavelength in the range from 350-2500 nm. The lamp is formed from a tungsten filament heated to about 3000 oC housed in a glass envelope. The output of the lamp approaches a black body radiation where it is observed that the energy of a tungsten lamp varies as the fourth power of the operating voltage.

Page 65: Atomic Emission Spectroscopy

65

Tungsten halogen lamps are currently more popular than just tungsten lamps since they have longer lifetime. Tungsten halogen lamps contain small quantities of iodine in a quartz envelope. The quartz envelope is necessary due to the higher temperature of the tungsten halogen lamps (3500 oC). The longer lifetime of tungsten halogen lamps stems from the fact that sublimed tungsten forms volatile WI2 which redeposits on the filament thus increasing its lifetime. The output of tungsten halogen lamps are more efficient and extend well into the UV.

Page 66: Atomic Emission Spectroscopy

66

Tungsten lamps (350-2500 nm)

Why add IWhy add I22 in the lamps? in the lamps?W + IW + I22 → WI → WI22

Low limit: 330 nm

1)Low intensity2)Glass or quartz envelope

Page 67: Atomic Emission Spectroscopy

67

3. Xenon Arc Lamps Passage of current through an atmosphere of

high pressured xenon excites xenon and produces a continuum in the range from 200-1000 nm with maximum output at about 500 nm. Although the output of the xenon arc lamp covers the whole UV and visible regions, it is seldom used as a conventional source in the UV-Vis. The radiant power of the lamp is very high as to preclude the use of the lamp in UV-Vis instruments. However, an important application of this source will be discussed in luminescence spectroscopy which will be discussed later.

Page 68: Atomic Emission Spectroscopy

68

Page 69: Atomic Emission Spectroscopy

69

UV-Vis Absorption Spectroscopy

Lecture 24

Page 70: Atomic Emission Spectroscopy

70

Instrumental Components• Source - selector (monochromators)• Sample holders • Cuvettes (b = 1 cm, typically)

1. Glass (Vis)2. Fused silica (UV+Vis)

• Detectors– Photodiodes– PMTs

Page 71: Atomic Emission Spectroscopy

71

Sample ContainersSample containers are called cells or cuvettes and are

made of either glass or quartz depending on the region of the electromagnetic spectrum. The path length of the cell varies between 0.1 and 10 cm but the most common path length is 1.0 cm. Rectangular cells or cylindrical cells are routinely used. In addition, disposable polypropylene cells are used in the visible region. The quality of the absorbance signal is dependent on the quality of the cells used in terms of matching, cleaning as well as freedom from scratches.

Page 72: Atomic Emission Spectroscopy

72

Instrumental designs for UV-visible photometers or spectrophotometers. In (a), a single-beam instrument

is shown. Radiation from the filter or monochromator passes through either the reference cell or the sample cell before striking the photodetector.

Types of Instruments

Page 73: Atomic Emission Spectroscopy

73

1. Single beam–Place cuvette with blank in place

in the instrument and take a reading 100% T

–Replace blank with sample and take reading % T for analyte (from which absorbance is calc’d)

Page 74: Atomic Emission Spectroscopy

74

Most common spectrophotometer Spectronic 20

1. On/Off switch and zero transmission adjustment knob

2. Wavelength selector/Readout

3. Sample chamber

4. Blank adjustment knob

5. Absorbance/Transmittance scale

Page 75: Atomic Emission Spectroscopy

75

Page 76: Atomic Emission Spectroscopy

76

End view of the exit slit of the Spectronic 20 spectrophotometer pictured earlier

Page 77: Atomic Emission Spectroscopy

77

• Single-Beam Instruments for the Ultraviolet/Visible Region

Page 78: Atomic Emission Spectroscopy

78

• Single-Beam Computerized Spectrophotometers Inside of a

single-beam spectrophotometer connected to a computer.

Page 79: Atomic Emission Spectroscopy

79

2. Double beam (most commercial instruments)– Light is split and directed towards both reference

cell (blank) and sample cell– One or two detectors; electronics measure ratio

(i.e., measure/calculate absorbance)– Advantages:

• Compensates for fluctuations in source intensity and drift in detector

• Better design for continuous recording of spectra

• Faster

Page 80: Atomic Emission Spectroscopy

80

General Instrument DesignsDouble Beam: In - Space

Needs two detectors

Page 81: Atomic Emission Spectroscopy

81

General Instrument DesignsDouble Beam: In - Time

Page 82: Atomic Emission Spectroscopy

82

Page 83: Atomic Emission Spectroscopy

83

Page 84: Atomic Emission Spectroscopy

84

Merits of Double Beam Instruments

1.Compensate for all but the most short term fluctuation in radiant output of the source

2.Compensate drift in transducer and amplifier

3.Compensate for wide variations in source intensity with wavelength

Page 85: Atomic Emission Spectroscopy

85

3 .Dual Beam Instruments

Page 86: Atomic Emission Spectroscopy

86

4. Multichannel Instruments• Photodiode array detectors used

(multichannel detector, can measure all wavelengths dispersed by grating simultaneously).

• Advantage: scan spectrum very quickly “snapshot” < 1 sec.

• Powerful tool for studies of transient intermediates in moderately fast reactions.

• Useful for kinetic studies.• Useful for qualitative and quantitative

determination of the components exiting from a liquid chromatographic column.

Page 87: Atomic Emission Spectroscopy

87

A multichannel diode-array spectrophotometer

Page 88: Atomic Emission Spectroscopy

88

Page 89: Atomic Emission Spectroscopy

89

Location of Sample cellIn all photometers and scanning spectrophotpmeters

described above, the cell has been positioned after the monochromators. This is important to decrease the possibility of sample photodecomposition due to prolonged exposure to all frequencies coming from the source. However, the sample is positioned before the monochromator in multichannel instruments like a photodiode array spectrophotometer. This can be done without fear of photodecomposition since the sample exposure time is usually less than 1 s. Therefore, it is now clear that in UV-Vis where photodecomposition of samples can take place, the sample is placed after the monochromators in scanning instruments while positioning of the sample before the monochromators is advised in multichannel instruments.

Page 90: Atomic Emission Spectroscopy

90

5 .Probe Type Instruments

These are the same as conventional single beam instruments but the beam from the monochromators is guided through a bifurcated optical fiber to the sample container where absorption takes place. The attenuation in reflected beam at the specified wavelength is thus measured and related to concentration of analyte in the sample.

Page 91: Atomic Emission Spectroscopy

91

A fiber optic cable can be referred to as a light pipe where light can be transmitted by the fiber without loss in intensity (when light hits the internal surface of the fiber at an angle larger than a critical angle). Therefore, fiber optics can be used to transmit light for very long distances without losses. A group of fibers can be combined together to form a fiber optic cable or bundle. A bifurcated fiber optic cable has three terminals where fibers from two separate cables are combined at one end to form the new configuration.

Page 92: Atomic Emission Spectroscopy

92

Page 93: Atomic Emission Spectroscopy

93

Page 94: Atomic Emission Spectroscopy

94

Page 95: Atomic Emission Spectroscopy

95

Fiber optic probeinstrument

Page 96: Atomic Emission Spectroscopy

96

Page 97: Atomic Emission Spectroscopy

97

6 .Double Dispersing Instruments

The instrument in this case has two gratings where the light beam leaving the first monochromators at a specified wavelength is directed to the second grating. This procedure results in better spectral resolution as well as decreased scattered radiation. However, double dispersing instruments are expensive and seem to offer limited advantages as compared to cost; especially in the UV-Vis region where exact wavelength may not be crucial.

Page 98: Atomic Emission Spectroscopy

98

Page 99: Atomic Emission Spectroscopy

99

Optical diagram of the Varian Cary 300 double-dispersing spectrophotometer. A second monochromator is added immediately after the source.