Made in France ARTIS ® ARTIS ® INTRAOCULAR LENS The only hydrophobic IOL which is injected in a 1.8 mm diameter cartridge ARTIS ® , two innovations are better than one Ophthalmologic surgery Ophthalmologic surgery Cristalens 12 allée Rosa Luxemburg - BP 50240 Eragny 95615 Cergy Pontoise cedex - France Tél : + 33 (0) 1 34 32 30 50 - Fax : + 33 (0) 1 34 32 30 60 Cristalens Industrie 4 rue Louis de Broglie - 22300 Lannion - France Tél : + 33 (0) 2 96 48 92 92 - Fax : + 33 (0) 2 96 48 97 87 www.cristalens.fr 14/06/2012 DOC_4_PHOBIC

Transcript of ARTIS The only hydrophobic IOL which is injected in a 1.8 ... · ARTIS® New design Anatomic...

Made in France

ARTIS®ARTIS®

INTRAOCULAR LENS

The only hydrophobic IOL which is injected

in a 1.8 mm diameter cartridge

ARTIS®, two innovations are better than one

Ophthalmologicsurgery

Ophthalmologicsurgery

Cristalens12 allée Rosa Luxemburg - BP 50240 Eragny

95615 Cergy Pontoise cedex - France

Tél : + 33 (0) 1 34 32 30 50 - Fax : + 33 (0) 1 34 32 30 60

Cristalens Industrie4 rue Louis de Broglie - 22300 Lannion - France

Tél : + 33 (0) 2 96 48 92 92 - Fax : + 33 (0) 2 96 48 97 87

www.cristalens.fr

14/0

6/20

12 D

OC

_4_P

HO

BIC

4 reasons to prefer the new material Cristalens

Faster unfolding, better control

Optimised refractive index

Transparent material

Positioning in the capsular bag easier to control

ARTIS® The only hydrophobic iol which is injected in a 1.8 mm diameter cartridge

Reflectance and index contrast

The gradient of the curve shows the unfolding time against temperature.In practice, above 30 °C in a cohesive OVD, com-

plete unfolding (optical and haptic) is achieved in:

> 15 seconds for a hydrophilic iol

> 22 seconds for the hydrophobic Cristalens

> 62 seconds for a hydrophobic lens manufactured

by the competition

Due to the presence of significant vacuoles in a lens, light diffusion increases as the angle of

incidence of the light approaches the normal.

If the pressure required for injection is low, the entire implant positioning process is simplified:

> regular effortless sliding > controlled unfolding of the implant

> positioning of the haptics at the required position

In a 2.2 mm diameter cartridge, the injection pressure

of our material is 1.8 kg. Other hydrophobic materials

require a pressure of 2.6 to 3.8 kg.

In a 1.8 mm cartridge, we are at 2.2 kg, when no other

hydrophobic material can be injected.

1

2

3

4

10 000

1000

100

10

0

70

60

50

40

30

20

10

0

10 12 14 16 18 20 22 24 26 28 30 32 34

1,35 1,4 1,45 1,5 1,55 1,6

ARTIS®

Normalangle of

observation

Lens transparent in water

Lens with glistening

Maximumangle ofobservation

Lens 1 Lens 2 Lens 3 Lens 4

ARTIS® reflection coefficient

Coefficient of another referencehydrophobic material

Lens index

Relative coefficient

Deployment time

Temperature

1.8 mm cartridge 2.2 kg pressure

33° 23,5° 14°

To reduce both the iol optic thickness

and the visual discomfort related to

reflectance of the material.

ARTIS®

New design

Anatomic conditions for precise, stable refraction

ARTIS®: design better adapted to variationsin bag diameter

• A single iol diameter does not take into

account the high variability in lens diameters:

8.25 to 10.80 mm.

• This diameter does not always vary with

the axial length, the implant must therefore

adapt to the bag diameter and not the

opposite. This condition is required to

guarantee the precision and stability of

the refractive result.

• An implant which adapts to variations in

capsular bag diameter will be better positioned in the ciliary diaphragm and

on the optical axis.

>> The implant adapts to the bag

diameter to ensure centration and

correct position on the optical axis.

>> With better centration and no

tilting of the haptics or the optics.

The experiment conducted in wells and on bank eyes (views in Miyake Apple,Liliana Werner laboratory, Salt Lake City) has demonstrated the ADAPTABILITYof the ARTIS® design.

• In 9 to 10 mm wells, corresponding to the largest number of diameters encountered

in clinics, the loops deform with no anterior or posterior angulation, keeping

the same contact arc.

• An implant which adapts to the bag (and not the opposite)

will be better centered in the bag and better positioned on

the optical axis at the required position.

>> Diameter variation induced by the implant.

View in Miyake Apple (Liliana Werner, Salt Lake City).

+ 0.3 mm

+ 1.7 mm

2

Ø 10.0 mm Ø 9.5 mm Ø 9.0 mm

Diameter of the empty lens bag Ø 9.3 mm

ARTIS® Ø 9.6 mm

Other Ø 10.99 mm

1

The only hydrophobic iol which is injected in a 1.8 mm diameter cartridge

ARTIS® TECHNICAL SPECIFICATIONS

Lens type For implantation in the capsular bag

Optical diameter6.0 mm 25.0 D

5.8 mm 25.5 D

Overall diameter10.79 mm 25.0 D

10.50 mm 25.5 D

Design One piece with square edges on 360°

OpticalAspherical on the posterior surface,

partial compensation of corneal aberrations, biconvex

Angulation 5°

Material Hydrophobic Acrylic for micro-incision

Dioptric powers

From + 10.0 D to + 30.0 D

- from + 10.0 D to + 14.0 D by + 1.0 D

- from + 14.5 D to + 25.5 D by + 0.5 D

- from + 26.0 D to + 30.0 D by + 1.0 D

Estimated A-constant119.3 ultrasound biometry

119.7 interference laser biometry

Refractive index 1.54

Sterilization Ethylene oxide

Recommended incision size 1.8 mm – 2.0 mm

DESIGNATION TECHNICAL SPECIFICATIONS