OPHTHALMIC SURGERY, LASERS & IMAGING · VOL. 43, NO. 6 (SUPPL), 2012 S103 CLINICAL SCIENCE From Teixeira Oftalmologia (AT, CS, FTP, NS), Brasília, Brazil; the Department of Ophthalmology (AT, CS, NA), Universidade Federal de São Paulo, São Paulo, Brazil; the Department of Ophthalmology (AT, BAS), Universidade Católica de Brasília, Brasília, Brazil; the Department of Ophthalmology (FRF), Hôpital Maisonneuve-Rosemont, Université de Montréal, Montréal, Québec, Canada; the Department of Ophthalmology (FRF), Pontifícia Universidade Católica, Rio de Janeiro, Brazil; and the Department of Ophthalmology (NA), University of Illinois at Chicago, Chicago, Illinois. Originally submitted July 14, 2012. Accepted for publication October 2, 2012. The authors have no financial or proprietary interest in the materials presented herein. Address correspondence to Anderson Teixeira, MD, SDS Bloco D no 27 sala 306, Brasília 70392-901, Brazil. E-mail: [email protected] doi: 10.3928/15428877-20121003-02 n BACKGROUND AND OBJECTIVE: To com- pare the architecture of single plane self-sealing clear corneal incision (SP-CCI) and shallow grooved self- sealing clear corneal incision (SG-CCI) after cataract surgery using Fourier-domain optical coherence to- mography (FD-OCT). n PATIENTS AND METHODS: An FD-OCT sys- tem with a corneal adaptor module was used to image the corneal incisions in radial sections in 44 eyes. A line scan pattern was used to measure the corneal inci- sion, positioning the caliper mark perpendicular to the limbus. Measurements were performed immediately after cataract surgery and at postoperative days 1, 7, and 30. Incisions were analyzed regarding length, loca- tion, angle, architecture, and anatomic imperfections. n RESULTS: All incisions were located superiorly (temporal, 24 eyes; nasal, 20 eyes). The mean SG-CCI length (20 eyes) was 1.79 ± 0.31 mm (range: 0.93–2.64 mm) and the mean incision angle was 37 ± 7 degrees (range: 24–34 degrees). The mean SP-CCI length (22 eyes) was 1.64 ± 0.22 mm (range: 1.25–2.36 mm) and the mean incision angle was 38.5 ± 5 degrees (range: 27–52 degrees) (P < .05). Anatomic imperfections were observed at postoperative day 1 in 19 eyes for SP- CCI and 14 eyes for SG-CCI. No patient presented endophthalmitis during 30 days of follow-up. n CONCLUSION: Epithelial imperfection at the cor- neal incision site was observed in more than 36% of the wounds (36% in SP-CCI and 45% in SG-CCI) at postoperative day 1 with spontaneous resolution. SG-CCI had the greatest length and lowest angle of corneal incision. Reduced incision length and inap- propriate construction may determine risk factors for wound architectural imperfections. Further studies including more patients with an architectural analysis of clear corneal incisions are needed to confirm these preliminary results. [Ophthalmic Surg Lasers Imaging 2012;43:S103- S108.] Architectural Analysis of Clear Corneal Incision Techniques in Cataract Surgery Using Fourier-Domain OCT Anderson Teixeira, MD, PhD; Camila Salaroli, MD; Flavio Rezende Filho, MD, PhD; Francisco Teixeira Pinto, MD; Nonato Souza; Benedito Antonio Sousa, MD; Norma Allemann, MD, PhD

Transcript of Architectural Analysis of Clear Corneal Incision ......Fourier-domain optical coherence tomography...

Page 1: Architectural Analysis of Clear Corneal Incision ......Fourier-domain optical coherence tomography after cata-ract extraction showing the corneal incision. Incision length was mea-sured

Ophthalmic Surgery, laSerS & imaging · VOl. 43, nO. 6 (Suppl), 2012 S103

■ C L I N I C A L S C I E N C E ■

From Teixeira Oftalmologia (AT, CS, FTP, NS), Brasília, Brazil; the Department of Ophthalmology (AT, CS, NA), Universidade Federal de São Paulo, São Paulo, Brazil; the Department of Ophthalmology (AT, BAS), Universidade Católica de Brasília, Brasília, Brazil; the Department of Ophthalmology (FRF), Hôpital Maisonneuve-Rosemont, Université de Montréal, Montréal, Québec, Canada; the Department of Ophthalmology (FRF), Pontifícia Universidade Católica, Rio de Janeiro, Brazil; and the Department of Ophthalmology (NA), University of Illinois at Chicago, Chicago, Illinois.

Originally submitted July 14, 2012. Accepted for publication October 2, 2012.The authors have no financial or proprietary interest in the materials presented herein.Address correspondence to Anderson Teixeira, MD, SDS Bloco D no 27 sala 306, Brasília 70392-901, Brazil. E-mail: [email protected]: 10.3928/15428877-20121003-02

n BACKGROUND AND OBJECTIVE: To com-pare the architecture of single plane self-sealing clear corneal incision (SP-CCI) and shallow grooved self-sealing clear corneal incision (SG-CCI) after cataract surgery using Fourier-domain optical coherence to-mography (FD-OCT).

n PATIENTS AND METHODS: An FD-OCT sys-tem with a corneal adaptor module was used to image the corneal incisions in radial sections in 44 eyes. A line scan pattern was used to measure the corneal inci-sion, positioning the caliper mark perpendicular to the limbus. Measurements were performed immediately after cataract surgery and at postoperative days 1, 7, and 30. Incisions were analyzed regarding length, loca-tion, angle, architecture, and anatomic imperfections.

n RESULTS: All incisions were located superiorly (temporal, 24 eyes; nasal, 20 eyes). The mean SG-CCI length (20 eyes) was 1.79 ± 0.31 mm (range: 0.93–2.64 mm) and the mean incision angle was 37 ± 7 degrees (range: 24–34 degrees). The mean SP-CCI length (22

eyes) was 1.64 ± 0.22 mm (range: 1.25–2.36 mm) and the mean incision angle was 38.5 ± 5 degrees (range: 27–52 degrees) (P < .05). Anatomic imperfections were observed at postoperative day 1 in 19 eyes for SP-CCI and 14 eyes for SG-CCI. No patient presented endophthalmitis during 30 days of follow-up.

n CONCLUSION: Epithelial imperfection at the cor-neal incision site was observed in more than 36% of the wounds (36% in SP-CCI and 45% in SG-CCI) at postoperative day 1 with spontaneous resolution. SG-CCI had the greatest length and lowest angle of corneal incision. Reduced incision length and inap-propriate construction may determine risk factors for wound architectural imperfections. Further studies including more patients with an architectural analysis of clear corneal incisions are needed to confirm these preliminary results.

[Ophthalmic Surg Lasers Imaging 2012;43:S103-S108.]

Architectural Analysis of Clear Corneal Incision Techniques in Cataract Surgery Using

Fourier-Domain OCT

Anderson Teixeira, MD, PhD; Camila Salaroli, MD; Flavio Rezende Filho, MD, PhD; Francisco Teixeira Pinto, MD; Nonato Souza; Benedito Antonio Sousa, MD; Norma Allemann, MD, PhD

Page 2: Architectural Analysis of Clear Corneal Incision ......Fourier-domain optical coherence tomography after cata-ract extraction showing the corneal incision. Incision length was mea-sured

S104 cOpyright © SlacK incOrpOrated

INTRODUCTION

Clear corneal incision (CCI) in cataract surgery is accepted worldwide. It allows rapid visual rehabilita-tion and involves less induced corneal astigmatism, less bleeding during surgery, higher structural stability of the anterior chamber, and ease in construction.1-5 On the other hand, some studies suggest that CCI may be associated with an increased risk of postoperative endo-phthalmitis4-7 and self-sealing behavior of the wound is an important barrier against intraocular infection. CCI can be performed by either single plane self-seal-ing construction or two plane self-sealing construction (shallow grooved self-sealing construction). The differ-ence between the techniques depends on the corneal construction tunnel to generate a self-sealing valve and safety postoperative workout.8

Anterior segment optical coherence tomography (AS-OCT) is a non-contact method used to examine the architecture features of CCI in vivo after cataract surgery such as epithelial and endothelial gaping, en-dothelial misalignment, Descemet’s membrane detach-ment, distance from the limbus, and length, shape, angle, and thickness of the CCI.3,9-14 Spectral-domain OCT (SD-OCT) has good quality resolution and ac-quisition is faster, allowing adequate screening for post-operative follow-up to analyze the corneal incision.

The objective of this study was to demonstrate the architectural features and the dimensions of the corneal wound after cataract surgery using SD-OCT and to correlate them to surgical outcomes. This study provid-ed accurate measurement of the wound in vivo using high-resolution, clearly visible SD-OCT images of the cornea right after the phacoemulsification procedure without significant motion error, enabling analysis of the differences of the incision techniques.

PATIENTS AND METHODS

Study participants were recruited from clinical practices (AT, FTP, and CS) according to a prospec-tive study protocol approved by the institutional review board. Informed consent was obtained and document-ed for every participant. The study adhered to the te-nets of the Declaration of Helsinki. Forty-four eyes of 30 patients with cataract and no other ocular diseases were enrolled in the study.

Randomly selected eyes underwent cataract sur-

gery with the clear corneal incision performed using either technique. Patients were divided in two groups: 24 eyes received single plane self-sealing clear corneal incision (SP-CCI) construction and 20 eyes received shallow grooved clear corneal incision (SG-CCI) con-struction.

All phacoemulsification surgeries were performed by the same surgeon (AT) and CCI was performed su-peronasally in the left eye or superotemporally in the right eye. The incisions were made with a 2.75-mm metal keratome (Angiotech Inc., Reading, PA) and phacoemulsification surgery was performed using the Alcon Legacy system (Alcon Laboratories, Inc., Fort Worth, TX) under topical anesthesia with less than 0.5 minute of ultrasound (burst mode). Lens cortex was removed by automated irrigation and aspiration using a bimanual set-up. A foldable acrylic intraocu-lar lens (Akreos AO; Bausch & Lomb, St. Louis, MO) was injected into the capsular bag using a disposable implantation system, without any incision tunnel enlargement. Phacoemulsification and side-port inci-sions were sealed with stromal hydration using bal-anced salt solution. The eyes were filled with balanced salt solution to normal pressure on palpation without leaking.

A Fourier-domain OCT system (RTVue; Optovue Inc., Fremont, CA) with a corneal adaptor module was used to image the CCI immediately after the surgery and at postoperative days 1, 7, and 30 to evaluate the incision morphology. A line scan pattern was used to measure the corneal incisions, positioning the caliper mark perpendicular to the limbus. On OCT, radial scans were performed at the corneal incision site to an-alyze the following parameters: curvilinear length (to-tal length measured by six points between the internal and external wound openings, Figure 1B), linear length (line between the internal and external wound open-ings, Figure 1A), angle between the corneal surface tan-gent, architectural deformation, and external depth of the incision at each of the three stages of the wound.

SAS version 9.1 programming language (SAS In-stitute, Inc., Cary, NC) was used for all analyses. The data were analyzed by linear and curve length, shallow groove, incision angle, and epithelial and endothe-lial gaping. The paired t test was used to compare the different incision techniques and analysis of variance tested across days. A P value of less than .05 was con-sidered significant for all tests.

Page 3: Architectural Analysis of Clear Corneal Incision ......Fourier-domain optical coherence tomography after cata-ract extraction showing the corneal incision. Incision length was mea-sured

Ophthalmic Surgery, laSerS & imaging · VOl. 43, nO. 6 (Suppl), 2012 S105

RESULTS

Thirty participants (44 eyes) were enrolled and completed the study: 18 men and 12 women. Mean age was 68 ± 12 years (range: 52 to 82 years). No intra-operative complication or incision leakage was seen in any of the participants. During 30 days of follow-up, there were no reports of hypotony, shallow chamber, or endophthalmitis.

The average curvilinear length was 1.64 ± 0.22 mm (range: 1.25 to 2.36 mm) for SP-CCI and 1.79 ± 0.31 mm (range: 0.93 to 2.64 mm) for SG-CCI, which was statistically significant (P = .04). The average linear length of the incision from external to internal open-ing was 1.61 ± 0.24 mm (range: 1.25 to 2.28 mm) for SP-CCI and 1.62 ± 0.27 mm (range: 0.89 to 2.40 mm) for SG-CCI, which was not statistically signifi-cant (P = .92). The average angle of the incision rela-tive to the tangent plane to the corneal surface was 38 ± 5 degrees (range: 27 to 52 degrees) for SP-CCI and 37 ± 7 degrees (range: 21 to 52 degrees) for SP-CCI;

there were no significant differences between groups (P = .12). The shallow grooved average was 0.24 ± 0.61 mm (range: 0.14 to 0.40 mm). Tables 1 and 2 list all postoperative average measurements for both groups with the analysis of variance between days.

Changes on the Epithelial Side of the WoundEpithelial gaping was observed in both groups. It

was noted in 21% of the participants on the day of sur-gery and 8% on postoperative day 1 for SP-CCI and 47% and 10%, respectively, for SG-CCI. No epithe-lial changes were observed after postoperative day 1 in both groups. The average of epithelial changes was 0.11 ± 0.04 mm (range: 0.05 to 0.15 mm) for SP-CCI and 0.13 ± 0.1 mm (range: 0.02 to 0.33 mm) for SG-CCI, with no significant differences between groups (P = .66).

Changes on the Endothelium Side of the WoundEndothelium gaping was observed in both groups.

For SP-CCI, it was noted in 79% of the participants on the day of surgery, in 42% on postoperative day

Figure 1. Fourier-domain optical coherence tomography after cata-ract extraction showing the corneal incision. Incision length was mea-sured by a 6-point (dots) line traced between the internal and external wound openings. (A) Single plane self-sealing clear corneal incision (SP-CCI group) and (B) shallow grooved self-sealing clear corneal incision (SG-CCI group).

Page 4: Architectural Analysis of Clear Corneal Incision ......Fourier-domain optical coherence tomography after cata-ract extraction showing the corneal incision. Incision length was mea-sured

S106 cOpyright © SlacK incOrpOrated

1, and in 37% on postoperative day 7. For SG-CCI, it was observed in 63%, 31%, and 21%, respectively. Three patients had endothelial gaping at postoperative day 30 (2 for SP-CCI and 1 for SG-CCI). The average

of endothelial gaping was 0.13 ± 0.06 mm (range: 0.04 to 0.28 mm) for SP-CCI and 0.13 ± 0.05 mm (range: 0.05 to 0.22 mm) for SG-CCI, with no significant dif-ferences between groups (P = .82). Descemet’s detach-ment and endothelial misalignment were observed in both groups (Table 3) with a complete resolution at 30 days postoperatively.

DISCUSSION

CCIs are classified by their architecture as being ei-ther single plane when there is no groove at the external edge of the incision, shallow grooved when the initial groove is less than 400 microns, and deeply grooved when it is deeper than 400 microns.15 There are articles about CCI that have studied the integrity and heal-ing of the architecture at the incision site, and wounds greater than 2.0 mm in length and 3.0 to 3.5 mm in width demonstrated sufficient self-sealing.13,14,16,17 In our series, there was a discrepancy comparing linear and curve length measurements, and our results dem-

TABLE 3

No. of Eyes With Architectural Changes Observed Using FD-OCT in Self-Sealing Clear Corneal Incisions After Cataract

Extraction in Both GroupsParameter SP-CCI SG-CCI

Descemet’s detachment 7 6

Endothelial misalignment 0 1

Loss of coaptation 1 3

Endothelium gaping 19 12

Epithelium gaping 5 9

FD-OCT = Fourier-domain optical coherence tomography; SP-CCI = single plane self-sealing clear corneal incision; SG-CCI = shallow grooved self-sealing clear corneal incision.

TABLE 1

Mean and Standard Deviation of FD-OCT Measurements Performed in the Group With SG-CCI After Cataract Extraction (n = 20 Eyes)

Postoperative Period

SG-CCI Linear

Length (mm)

SG-CCI Length

Curve (mm)

SG-CCI Shallow

Groove (mm)

SG-CCI Angle

(degrees)Epithelial

Gaping (mm)Endothelial

Gaping (mm)

Day 0 1.68 ± 0.32 1.80 ± 0.31 0.25 ± 0.06 41 ± 7 0.12 ± 0.1 0.15 ± 0.07

1 day 1.63 ± 0.44 1.75 ± 0.30 0.22 ± 0.07 39 ± 8 0.05 ± 0.05 0.11 ± 0.04

7 days 1.55 ± 0.27 1.66 ± 0.31 0.23 ± 0.06 38 ± 6 0.0 0.10 ± 0.03

30 days 1.50 ± 0.27 1.65 ± 0.35 0.24 ± 0.06 29 ± 7 0.0 0.22

ANOVA 0.253 0.426 0.494 < .0001 < .0001 < .0001

FD-OCT = Fourier-domain optical coherence tomography; SG-CCI = shallow grooved self-sealing clear corneal incision; ANOVA = analysis of variance.

TABLE 2

Mean and Standard Deviation of FD-OCT Measurements Performed in the Group With SP-CCI After Cataract Extraction (n = 24 Eyes)

Postoperative Period

SP-CCI Linear Length (mm)

SP-CCI Length Curve (mm)

SP-CCI Angle (degrees)

Epithelial Defect (mm)

Endothelial Defect (mm)

Day 0 1.73 ± 0.22 1.75 ± 0.24 41 ± 6 0.11 ± 0.07 0.15 ± 0.08

1 day 1.62 ± 0.24 1.67 ± 0.26 42 ± 5 0.95 ± 0.02 0.10 ± 0.04

7 days 1.56 ± 0.23 1.59 ± 0.24 37 ± 5 0.0 0.12 ± 0.08

30 days 1.52 ± 0.39 1.54 ± 0.25 32 ± 6 0.0 0.10 ± 0.03

ANOVA 0.058 0.023 < .0001 < .0001 0.019

FD-OCT = Fourier-domain optical coherence tomography; SP-CCI = single plane self-sealing clear corneal incision; ANOVA = analysis of variance.

Page 5: Architectural Analysis of Clear Corneal Incision ......Fourier-domain optical coherence tomography after cata-ract extraction showing the corneal incision. Incision length was mea-sured

Ophthalmic Surgery, laSerS & imaging · VOl. 43, nO. 6 (Suppl), 2012 S107

onstrate that average linear measurements were smaller than 2.00 mm in both groups and that the SG-CCI group had a longer length. The shortfall might be ex-plained by the stretching of the cornea under the force of the advancing metal blade.3 SG-CCI had a lower incidence of endothelial gaping comparing to SP-CCI, but both groups had patients with endothelial gaping still present after postoperative day 30 (1 eye for SG-CCI and 2 eyes for SP-CCI). Epithelial defects were more frequently observed for SG-CCI, with resolution between postoperative days 1 and 7. Independent of the technique used for the corneal incision, all eyes enrolled in the current study had a proper self-sealing CCI. In the literature, some authors advocate the use of SG-CCI and others suggest that SP-CCI is safe in most cases.3,18-20

The angle of the incision is an important factor to maintain the integrity and sealing properties of the wound. Small angles have been shown to be more ef-fective at creating a self-sealing condition.3,21 In the current study, the angle appeared to be shallow enough to create sealing (average angle = 39 degrees for SP-

CCI and 37 degrees for SG-CCI, P = .48) for all eyes in both groups.

We observed that there was external wound gap-ing in 21% of the OCT images on the day of surgery for SP-CCI, which decreased to 8% on postoperative day 1 and was not observed at days 7 and 30. For SG-CCI, external wound gaping was observed in 47% on the day of surgery, 10% on postoperative day 1, and was not observed during the following examinations. External gaps can be found in CCI with a fast reso-lution4,10,21,22 and can be associated with intraocular pressure fluctuation and stromal hydration.4,10,14

Endothelial wound gaping was observed in both groups, with a higher incidence for SP-CCI (P = .82). In the literature, the incidence of endothelial wound gaping varies between 25% and 70%, depending on the postoperative day.4,10,14,21,22 Some authors suggest-ed possible explanations for endothelial wound gaping, such as phacoemulsification instrument sizing or ex-cessive ultrasound power, corneal stromal edema, and endothelial damage.13,14,23 Other authors added intra-ocular pressure fluctuation as a factor.4,21

Figure 2. Complications of corneal incision after cataract surgery observed with Fourier-domain optical coherence tomography imag-ing. (A) Postoperative day 7: shallow grooved self-sealing clear corneal incision with endothelium misalignment. (B) Postoperative day 7: shallow grooved self-sealing clear corneal incision with loss of coaptation. (C) Postoperative day 7: single plane self-sealing clear corneal incision with loss of coaptation. (D) Postoperative day 1: single plane self-sealing clear corneal incision with endothelium gaping and Descemet’s detachment.

Page 6: Architectural Analysis of Clear Corneal Incision ......Fourier-domain optical coherence tomography after cata-ract extraction showing the corneal incision. Incision length was mea-sured

S108 cOpyright © SlacK incOrpOrated

Corneal incisions related to cataract surgery in both groups imaged with SD-OCT presented internal and external gaps in the immediate postoperative pe-riod with a spontaneous resolution. Both techniques seemed to be safe regarding self-sealing, but SG-CCI demonstrated a longer incision length. SD-OCT im-aging allowed immediate postoperative non-contact evaluation of patients undergoing cataract surgery and provided high-resolution measurements for evaluating the incision’s morphology. This information may be used to select corneal incision instruments and tech-niques to diminish the risk of gaping, which can be related to visual rehabilitation and to an increased risk of postoperative complications.

REFERENCES

1. Long DA, Monica ML. A prospective evaluation of corneal curvature changes with 3.0- to 3.5-mm corneal tunnel phacoemulsification. Ophthalmology. 1996;103:226-232.

2. Lyle WA, Jin GJ. Prospective evaluation of early visual and refractive effects with small clear corneal incision for cataract surgery. J Cataract Refract Surg. 1996;22:1456-1460.

3. Schallhorn JM, Tang M, Li Y, Song JC, Huang D. Optical coherence tomography of clear corneal incisions for cataract surgery. J Cataract Refract Surg. 2008;34:1561-1565.

4. McDonnell PJ, Taban M, Sarayba M, et al. Dynamic morphology of clear corneal cataract incisions. Ophthalmology. 2003;110:2342-2348.

5. Nagaki Y, Hayasaka S, Kadoi C, et al. Bacterial endophthalmitis after small-incision cataract surgery: effect of incision placement and intra-ocular lens type. J Cataract Refract Surg. 2003;29:20-26.

6. Cooper BA, Holekamp NM, Bohigian G, Thompson PA. Case-con-trol study of endophthalmitis after cataract surgery comparing scleral tunnel and clear corneal wounds. Am J Ophthalmol. 2003;136:300-305.

7. Taban M, Behrens A, Newcomb RL, et al. Acute endophthalmitis following cataract surgery: a systematic review of the literature. Arch Ophthalmol. 2005;123:613-620.

8. Crema A. Incisões para Faco. In: Dias FR, ed. Cirurgia de Catarata. Rio de Janeiro: Cultura medica; 2000:203-210.

9. Torres LF, Saez-Espinola F, Colina JM, et al. In vivo architectural anal-ysis of 3.2 mm clear corneal incisions for phacoemulsification using optical coherence tomography. J Cataract Refract Surg. 2006;32:1820-1826.

10. Calladine D, Packard R. Clear corneal incision architecture in the immediate postoperative period evaluated using optical coherence to-mography. J Cataract Refract Surg. 2007;33:1429-1435.

11. Can I, Bayhan HA, Celik H, Bostanci Ceran B. Anterior segment optical coherence tomography evaluation and comparison of main clear corneal incisions in microcoaxial and biaxial cataract surgery. J Cataract Refract Surg. 2011;37:490-500.

12. Lyles GW, Cohen KL, Lam D. OCT-documented incision features and natural history of clear corneal incisions used for bimanual mi-croincision cataract surgery. Cornea. 2011;30:681-686.

13. Fukuda S, Kawana K, Yasuno Y, Oshika T. Wound architecture of clear corneal incision with or without stromal hydration observed with 3-dimensional optical coherence tomography. Am J Ophthalmol. 2011;151:413-419.

14. Xia Y, Liu X, Luo L, et al. Early changes in clear cornea incision after phacoemulsification: an anterior segment optical coherence tomogra-phy study. Acta Ophthalmol. 2009;87:764-768.

15. Hooffman RSFI, Packer M. Phacoemulsification principles and tech-niques. In: Buratto LWL, Zanini M, Apple D, eds. Incision Construc-tion, 2nd ed. Thorofare, NJ: SLACK Incorporated; 2003:265-277.

16. Mackool RJ, Russell RS. Strength of clear corneal incisions in cadaver eyes. J Cataract Refract Surg. 1996;22:721-725.

17. Fine IH. Architecture and construction of a self-sealing incision for cataract surgery. J Cataract Refract Surg. 1991;17(suppl):672-676.

18. Masket S, Belani S. Proper wound construction to prevent short-term ocular hypotony after clear corneal incision cataract surgery. J Cata-ract Refract Surg. 2007;33:383-386.

19. Ernest PH, Lavery KT, Kiessling LA. Relative strength of scleral cor-neal and clear corneal incisions constructed in cadaver eyes. J Cataract Refract Surg. 1994;20:626-629.

20. Ernest PH. Cataract incision architecture. Int Ophthalmol Clin. 1994;34:31-57.

21. Taban M, Rao B, Reznik J, Zhang J, Chen Z, McDonnell PJ. Dy-namic morphology of sutureless cataract wounds: effect of incision angle and location. Surv Ophthalmol. 2004;49(suppl 2):S62-S72.

22. Fine IH, Hoffman RS, Packer M. Profile of clear corneal cataract incisions demonstrated by ocular coherence tomography. J Cataract Refract Surg. 2007;33:94-97.

23. Fukuda S, Kawana K, Yasuno Y, Oshika T. Repeatability and repro-ducibility of anterior chamber volume measurements using 3-dimen-sional corneal and anterior segment optical coherence tomography. J Cataract Refract Surg. 2011;37:461-468.