Aquaread Catalog V1.00.pdf

download Aquaread Catalog V1.00.pdf

If you can't read please download the document

description

Aquaread Catalog V1.00.pdf

Transcript of Aquaread Catalog V1.00.pdf

AQUAREADMultiparameter Water Quality Instruments

www.aquaread.com / [email protected] / @aquaread / +44 (0) 1843 609 695

AP-2000 AQUAPROBE quality testing instrument Portable multiparameter waterStandard Parameters: pH, ORP, Optical DO, EC, Salinity, Resistivity, TDS, SSG, & Temperature.ISE Electrodes: Ammonium & Ammonia, Calcium, Fluoride, Chloride, Nitrate.

3 Easy Options: 1. With or without Depth? 2. Which ISE electrode? 3. Which Optical electrode?

Optical Electrodes: Turbidity, Rhodamine, Chlorophyll, Fluorescein, Blue Green Algae, Refined Oil.

Mechanical SpecificationProtection Class Immersion Depth Operating Temperature Dimensions (L x Dia) Weight IP68 (permanent immersion) Min 75mm. MAX 30M -5C - +50C 290mm x 42mm 700g

The AP-2000 Sockets, pH/ORP, 1 x ISE and 1 x Optical

Fully loaded AP-2000

AP-2000 web page

www.aquaread.com / [email protected] / @aquaread / +44 (0) 1843 609 695

PROBE FEATURESOnly probe to measure Turbidity, Optical DO, EC, pH, ORP, Depth, Temperature and ISE, at the same time, in a diameter under 2 Probe size makes it ideal for portable use within our rugged carry case Made from marine grade anodised aluminium to withstand all environments Probe now features the rugged, metal AquaConn connectors for greater strength and protection AP-2000 can be used with a range of extension cable lengths that feature our AquaConn connectors. The cables are made to last on the outside and on the inside. All extension cables contain our new KevlarCore; internal strands of Kevlar that run the length of the cable to provide extra tensile strength, meaning there is no need for messy external support cables

AP-2000 AquaConn probe connection The AP-2000 allows you to add 1 ISE electrode and 1 optical electrode. You can choose from the large range of sensors available, swap and change as you please. There is also the option to add 2 ISEs. Email the team for more details

AP-2000 mated AquaConn connector and Kevlarcore extension cable

Large range of exchangeable electrodes available for use in the AP-2000

www.aquaread.com / [email protected] / @aquaread / +44 (0) 1843 609 695

AM-200 AQUAMETERdata and its GPS coordinates Recording your water quality

Mechanical SpecificationDimensions (W x H x D) Weight (inc. batteries) Display Data Memory GPS Receiver 90mm x 180mm x 39mm 450g 80 character with backlight Over 1000 full data sets12 channel, internal antenna

Atmospheric Pressure PC interface Power Supply Operating Temperatures Protection Class

150mb - 1150mb +/-1mb

USB Cable (provided) 5 x AA battery -5C - +50C IP67

Record your dataset and your precise location with built in GPSGoogle Earth and Google Map images showing data locations

AM-200 web page

www.aquaread.com / [email protected] / @aquaread / +44 (0) 1843 609 695

METER FEATURESRugged metal AquaConn connectors that wont let you down in the field

Built in GPS receiver enables location tagging with every dataset; view data in Google Earth Display of all the parameters being measured; scroll left and right to view Record your dataset at the touch of a single button, including all parameters, GPS data and calibration data Built in air pressure sensor for atmospheric compensation Ergonomic curved design sits nicely in the hand, with rubber gasket for added grip in wet conditions Rugged, metal AquaConn connectors provide extreme durability where its needed most Supplied with AquaLink PC utility and USB cable for data management, report creation and Google Earth file generation

AM-200 Aquameter can be used with both the AP-2000 and the AP-7000 Aquaprobes

www.aquaread.com / [email protected] / @aquaread / +44 (0) 1843 609 695

AP-7000 AQUAPROBE with novel self cleaning Long term deployment systemStandard Parameters: pH, ORP, Optical DO, EC, Salinity, Resistivity, TDS, SSG, Depth & Temperature. Full of Options: The top of the range AP-7000 has 6 user configurable sensor ports in addition to the complete set of standard parameters Choose from a wide range of ISE and Optical sensors to populate the probe with no restrictions; each sensor port will take ISE or Optical electrodes. See the sensor specifications for full details

Mechanical SpecificationProtection Class Immersion Depth Operating Temperature Dimensions (L x Dia) Weight IP68 (permanent immersion) Min 75mm. MAX 100M -5C - +50C 450mm x 77mm 1350g

Novel self cleaning mechanism is easy to maintain

AP-7000 web page

Remove the pin to free the cleaning arm for easy maintenance such as brush cleaning /replacement. www.aquaread.com / [email protected] / @aquaread / +44 (0) 1843 609 695

PROBE FEATURESNovel self cleaning mechanism allowing accurate data capture for longer; scan the QR code for a video demonstrationAP-7000 on YouTube

The probe features a full set of standard parameters plus 6 user configurable auxillary ports which can take any combination of our 12 ISE or optical screw-in sensors, which are swappable in the field Test at depths of up to 100m in both fresh and marine watersThe AP-7000 sensor ports give you maximun Use the probe the perfect eectrode layout freedom to create with the GPS Aquameter

to view and capture data in real-time and

to perform calibration Alternatively use the probe with the AquaLogger-7000 for extended periods of unmanned monitoring Available with SDI-12 + Modbus (RS-485) outputs allowing connection to third-party data loggers or telemetry devices - hook it straight up to existing recording networks for easy integration Utilises our rugged metal connectors and KevlarCore cables for high tensile cable strength and a solid connection

For use with the GPS Aquameter and AquaLogger-7000

The AP-7000 is designed for use with the AquaLogger-7000. When combined you can utilise the full list of features available such as customisable AP-7000 self cleaning and event triggers

LoggerLink PC application provided with every AquaLoggerAquaLogger-7000 web page

Standard Parameter Specification

SPECIFICATION

Optical Dissolved Oxygen Depth

Range Resolution Accuracy Range Resolution Accuracy Range Resolution Accuracy Range Resolution Accuracy Range

0 500.0% / 0 50.00 mg/L 0.1% / 0.01mg/L 0 - 200%: 1% of reading. 200% - 500%: 10% 0 100m (0 330F) 0.01m/F 1% 0 200 mS/cm (0 - 200,000 S/cm) 3 Auto-range scales: 0 9999 S/cm, 10.00 99.99 mS/cm, 100.0 200.0mS/cm 1% of reading or 1 S/cm if greater 0 100,000 mg/L (ppm) 2 Auto-range scales: 0 9999mg/L, 10.00 100.00g/L 1% of reading or 1mg/L if greater 5 cm 1 M cm 2 Auto-range scales: 5 9999 cm, 10.0 1000.0 K cm 1% of reading or 1 cm if greater 0 70 PSU / 0 70.00 ppt (g/Kg) 0.01 PSU / 0.01 ppt 1% of reading or 0.1 unit if greater 0 50 t 0.1 t 1.0 t 0 14 pH / 625mV 0.01 pH / 0.1mV 0.01 pH / 5mV 2000mV 0.1mV 5mV -5C +50C (23F 122F) 0.1 C/F 0.5 C

Conductivity (EC) TDS*

Resistivity*

Resolution Accuracy Range Resolution Accuracy Range Resolution Accuracy Range Resolution Accuracy [email protected] Range

Salinity* Seawater Specific Gravity* pH

www.aquaread.com / [email protected] / @aquaread / +44 (0) 1843 609 695

ORP

Resolution Accuracy Range Resolution Accuracy

Temperature

* Readings calculated from EC and temperature electrode values

ISE Electrode SpecificationRange Ammonium Resolution Accuracy Range Resolution Accuracy Range Resolution Accuracy Range Resolution Accuracy Range Resolution Accuracy Range Resolution Accuracy 0 1,000mg/L (ppm) 2 Auto-range scales: 0.00 - 99.99 mg/L, 100.0 999.9 mg/L 10% of reading 0 1,000mg/L (ppm) 2 Auto-range scales: 0.00 - 99.99 mg/L, 100.0 999.9 mg/L 10% of reading 0 1,000mg/L (ppm) Ammonia

Promotional price

Chloride

2 Auto-range scales: 0.00 - 99.99 mg/L, 100.0 999.9 mg/L 10% of reading 0 1,000mg/L (ppm) 2 Auto-range scales: 0.00 - 99.99 mg/L, 100.0 999.9 mg/L 10% of reading 0 1,000mg/L (ppm) 2 Auto-range scales: 0.00 - 99.99 mg/L, 100.0 999.9 mg/L 10% of reading 0 1,000mg/L (ppm) 2 Auto-range scales: 0.00 - 99.99 mg/L, 100.0 999.9 mg/L 10% of reading

Fluoride

Nitrate

Calcium

Ammonium electrode required. Readings calculated from ammonium, pH and temperature values.

Optical Electrode SpecificationRange Turbidity Resolution Repeatability Range Chlorophyll Phycocyanin(Freshwater Blue -Green Algae)

0 3000 NTU 2 Auto-range scales: 0.0 99.9 NTU, 100 3000 NTU 2% of reading 0 500 g/L (ppb) 0.1 g/L 2% of reading 0 300,000 cells/mL 1 cell/mL 2% of reading 0 200,000 cells/mL 1 cell/mL 2% of reading 0 500 g/L (ppb) 0.1 g/L 5% of reading 0 500 g/L (ppb) 0.1 g/L 5% of reading 0 10,000 g/L (ppb) 0.1 g/L 2% of reading

Resolution Repeatability Range Resolution Repeatability Range Resolution Repeatability Range Resolution Accuracy Range

Phycoerythrin(Marine BlueGreen Algae)

Rhodamine WT Dye

Fluorescein Dye

Resolution Accuracy Range Resolution Repeatability

Refined Oil

The accuracy figures quoted above represent the equipments capability at the calibration points at 25C. These figures do not take into account errors introduced by variations in the accuracy of calibration solutions and errors beyond the control of the manufacturer that may be introduced by environmental conditions in the field.