Aqua Mon 4000 Ph Red Ox Analyser
-
Upload
arvindgupta2005 -
Category
Documents
-
view
219 -
download
1
Transcript of Aqua Mon 4000 Ph Red Ox Analyser
-
7/31/2019 Aqua Mon 4000 Ph Red Ox Analyser
1/4
Aquamon 4000
pH/RedoxTransmitter
and Controller
Your Process, Our Analytics
Specialists in Process Efficiency and Energy Conservation
-
7/31/2019 Aqua Mon 4000 Ph Red Ox Analyser
2/4
Menu-driven program that simplifies set-up.
Built-in memory backup to ensure that calibration and otherinformation are not erased if power supply fails.
Push-button for calibration and electrode offset adjustment fromthe keypad.
Automatic temperature compensation (ATC).
Manual temperature compensation setting without the ATC probe,with independent setting for calibration and process temperature.
Two galvanically isolated current outputs 0/4...20mA
0 to 1999 seconds time delay adjustment on all relays minimizes false alarms.
Software selectable, two modes - pH or ORP measurement.
Special symmetrical configuration mode to facilitate operating inenvironments with electrical interference. Custom pH or ORP sensorsare available with the potential matching pin for symmetrical modemeasurement.
In the ORP mode, the Aquamon 4000 is able to measure in mV or as% concentration, with independent calibration modes. All parameterscan be operated in symmetrical or asymmetrical mode.
Push button pH calibration facility. Calibration buffer sets include US,NIST and DIN standards. Electrode condition is updated and displayedafter each calibration.
For applications where Hydrofluoric (HF) acid is present, eitherAntimony or HF-resistant glass electrodes must be used.
When Antimony electrodes are used, select 'Antimony' electrodeoption from controller menu to download relevant Antimonyelectrode data.
Separately adjustable high and low set-point hysteresis (dead bands)prevent chattering of relays around the set points.
Asymmetrical/symmetrical input for pH/ORP mode of operation.
Three control modes: limit controller, P controller and PI controller(P/PI controller as pulse length or pulse frequency).
Large dual display LCD for easy reading with clear multipleannunciators, alarm status, operational and error messages.
Two switching contacts as set-point relays.
Separate alarm relay alerting you to set point limits exceeded for acertain time and if the Pt100/Pt1000 wires are broken ordisconnected during the ATC function.
Wash relay.
Hold function to freeze output current (0/4...20mA) andrelease control relays.
LED indicators signal control activities to visually monitorcontroller status from a distance.
Protection against electromagnetic interference.
Back lighted and UV light protected LC display
Features
The Forbes Marshall Aquamon 4000 is used for measuring pH or ORP
and temperature values. The pH or ORP values can be measured using
proportional or limit control. The controller is available in two versions,
one for panel mounting and one for wall mounting in an enclosure. The
controller can be used for applications such as water treatment and
monitoring, galvanic-decontamination, chemical processing, food
processing, clean or wastewater control and neutralization processes.
-
7/31/2019 Aqua Mon 4000 Ph Red Ox Analyser
3/4
Dimensions and Connections
66.5
[2.6
2]
21.5
[8.5
]
6[.24]
90[3.54]
90[3.5
4]
Unit: MM[Inch]
80[3.15]
Holes for postmounting (4X)
Holes for wallmounting (2X)
144[5.67]
78[3.07]
39[1.54]approx.
14[.55]
144[5
.67]
Pg13.5(3 pcs) For Pg 13.5 cable glands
57.8[2.28]
24[.94]
111.50[4.39]
27.5[1.08]
84[3.30] 29[1.14]Max. 80
1-22 [0.03 - 0.87]
Seal gasket
Threaded rod for panel mounting
Panel Mount Drawing (All measuurements in mm)
Useful 12V DC output
from the controller
powers peripheral
wash mechanism,
preamplifiers, etc.
Wash relay allows automatic,
periodical cleaning of sensors
used in conjunction with
Wash program
Detachable connectors for easier wiring.
Independent Relay A and B
allows combination of Hi
andLo settings.
HOLD terminals allow for pH/
ORP Master-slave operations
(pH analyzer controls ORP
analyzer only under controlled
pH values)
Single-Pole-Double-Throw
Alarm relay for fail-safe
operations. When power
supply to the controller is
broken, the alarm relayactivates.
Automatic Temperature
Compensation with 2 or
3 wire Pt 1000 or Pt 100
temperature sensors.
3-wire system minimizes
errors from the tem-
perature sensor and
cables.
View temperature values
in degree Celsius or
degree Fahrenheit.
Manual Temperature
Compensation offers
independent input of
process and calibration
temperatures.
Dual 4-20 mA scaleable
outputs for measured
parameter and temperature.Unique Out-of-range
current setting at 22 mA
differentiates normal
readings from Out-of-
range values.
Liquid ground (potential matching
pin) for symmetrical operations
(pH / ORP controllers only)
Unit: MM[Inch]Panel cut out
92 +0.5 [3.62 +.02]
92+0.5
[3.6
2+.0
2]
Flat gasket 1mm [0.4](to be inserted by customer)
96[3.7
8]
92[3.6
2]
56[2.2
0]
32[1.2
6]
Max. 180 [7.09]
Max. 45 [1.77]
Wall Mount Drawing (All measuurements in mm)
Wall Mount Opening
-
7/31/2019 Aqua Mon 4000 Ph Red Ox Analyser
4/4
Forbes Marshall Pvt. Ltd.A-34/35, MIDC, Industrial Estate, 'H' Block, Pimpri, Pune - 411 018. India.Tel.: 91(0)20 - 27442020 Fax: 91(0)20 - 27442040E-mail: [email protected]
Branches : Ahmedabad, Alibag, Bangalore, Bhopal, Chandigarh, Chennai, Coimbatore, Delhi,Hyderabad, Indore, Jamshedpur, Kolkata, Lucknow, Madurai, Mumbai, Nagpur, Navi Mumbai,Pune, Surat, Vadodara, Visakhapatnam
Website : www.forbesmarshall.com
Subject to change without notice FM CIG | Analytical | Aquamon4K | Rev1 | 0708 | Solutions
R
pH Resolution 0.01 pH
Accuracy 0.01 pH
Range 0 to 100% or -1000 to 1000 mV
mV Resolution 0.1% or 1 mV
Accuracy 1 mV
Range -9.9 to 125 C (15.0 to 257.0 F)
Resolution 0.1
Temperature Relative Accuracy 0.5 C ( 1.0 F)
Sensor Pt100 / Pt1000 (jumper selectable); 2 or 3 wire
Compensation Automatic ( 10 C / 18 F offset adjustment) / Manual
Function Limit / P / PI Control (Pulse Length or Pulse Frequency)
Integral time 0 to 999.9 minutes
Adjustable period with pulse length controller 0.5 to 20 sec.
Adjustable period with pulse frequency controller 60 to 120 pulses / min
Set-Point and Pickup / Dropout delay 0 to 1999 seconds
Controller Functions Wash cycle 0.1 to 199.9 hours
Wash duration 1 to 1999 seconds
Switching pH Hysteresis 0.1 to 1 pH
Switching ORP Hysteresis 1 to 10.0 % / 10 to 100 mV
Contact Outputs, Controller 1 SPDT; 3 SPST relays
Switching Voltage/Current/Power Max 250 VAC / Max 3A / Max 600 VA
Function (switchable) Latching / pulse
Alarm Functions Pickup delay 0 to 1999 seconds
Switching Voltage/Current/Power Max 250 VAC / Max 3A / Max 600 VA
Display LCD UV coat, backlit 14 segments display with symbols for status information
Backlight On/Off selection with four level of brightness control
Electromagnetic Compliance Emitted Interference EN 61 326
(EMC) Specifications Immunity to Interference EN 61 326
Environmental Conditions Ambient Temp. Operating Range -10 to 50 oC (14 to 122 oF)
Rel. Humidity 10 to 95% (non-condensing)Input 80 to 250 VAC/DC 48/62 Hz
Power Supply Main Fuse 250 mA anti-surge
Pollution Degree 2
Transient Overvoltage category II
Signal Output Two 0/4 to 20 mA outputs for pH/mv and temperature, galvanically isolated
Electrical Data Load Max.600
and Connections pH / ORP Input Screw Terminals
Connection terminal Wall mount (2 x 3-pins; 8-pin, 9-pin & 13-pin terminal blocks)
Panel mount (1 x 3-pin; 1 x 9-pin & 1 x 19-pin terminal blocks)
Dimensions (W x H x D) Wall Mount 144 x 144 x 111.5 mm (5.67 x 5.67 x 4.39 inch)
Panel Mount 175 x 96 x 96 mm ( 6.89 x 3.78 x3.78 inch)
Mechanical Weight Wall Mount 745g (unit); 1100g (boxed)Specifications Panel Mount 550g (unit); 950g (boxed)
Material Wall Mount PBT
Panel Mount ABS
Insulation Wall Mount NEMA 4X, IP 65
Panel Mount IP 54 (front) , IP 40 (housing)
Ordering Information Wall mounted pH transmitter AQUAMON 4000 W
Panel mounted pH transmitter AQUAMON 4000 P
ange -2.00 to 16.00 pH
Sr. No. Description Code
1.
2.
Specifications