AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot...

24
AQUA-LAB EVO CHEMICAL DISPENSING SYSTEM User Manual REV L 4000003 revL0717 © Hydra-Flex, Inc. 2017

Transcript of AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot...

Page 1: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2016

AQUA-LAB EVO™CHEMICAL DISPENSING SYSTEM

User ManualREV L

4000003 revL0717© Hydra-Flex, Inc. 2017

Page 2: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2017

Page 3: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2016

CONTENTS

Principles of Operation 1

Specifications 1

Aqua-Lab EVO Components 2Integrated Motor Starter Wiring Diagram 3

Estimated Timeline 5

Installation Instructions 5

Startup 7 Pump Priming Instructions 7

Optimizing the System 8 Injectors vs. Metering tips vs. Nozzles 8

Application Optimization 8 Chemical Usage Measuring 8 Recommended Maintenance 9

Troubleshooting 9 Injector Vacuum Check (for troubleshooting injectors) 9 Injector Issues 10 Air Operated Valve Replacement 11 Pump Issues 11 Pressure Regulator Issues 12 Flow / Arch Issues 12 Valve Issues 12

Injector Optimization Tool 13

Reference 13 Chem-Flex Composite Injector Part Numbers 14 Chemical Dilution Ratios 15 Pressure Loss In Run Length 16 Replacement Parts List 17Warranty 18

Appendix A (Pre-Installation Checklist) 19

TABLE OF CONTENTS

Page 4: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2017

The Aqua-Lab EVO was designed to meet the application requirements of the standard roll-over automatic car wash with an available pump room to store chemical and equipment.

The system is set-up so that all that is required from the car wash controller is an output signal for each valve. The EVO electrical control box uses that signal to open both the 3.5 watt air solenoid valves and to start the pump.

With a 40 psi or greater city water pressure feed, the standard 10 GPM pump produces up to 225 psi of pressure which is regulated down to 200 psi by the bypass regulator. (The pump should never run above 225 psi.)

The200psiwaterflowsintotheHydra-Cannonmanifold,whichisfittedwithupto12air-actuatedvalvesthatarecontrolledbyairsolenoidvalves.Thewatervalvescontroltheflowofwatertoeachinjectororflow-thrurinsefitting.

TheChem-Flexinjectorsarecolor-codedbasedonthewaterflowneededforeachindividualapplication,andcombinewithcolor-codedchemicalmeteringtipstoproportiontheprecisechemicalmixratio.Aswaterflowsinto the Chem-Flex injector, the velocity increases. As the velocity increases the pressure decreases creating a vacuum which allows the precise amount of chemical to be pulled into the chamber where it mixes with water. This precise chemical mixing, along with the proper nozzle, creates an optimal display in the wash bay.

Optional components include an inlet water selection valve, a foaming air wing, a system-protection pressure switch and a tri-foam manifold.

• The inlet water selection valve is operated using an air solenoid valve. This valve allows the option of using two different water sources.

• The foaming air wing can be used for foaming air or as a blow-out frost protection. When the temperature drops below a certain pre-set point, the car wash controller opens the air valve causing air to be forced through the lines, clearing them of water or solution that could potentially freeze and cause damage.

• The pressure switch is supplied to ensure that if the pump is not getting pressure, a signal will be sent to the controller to shut the system down.

• Thetri-foammanifoldisathreepositionquick-connectmanifoldthatfitsintoasingleporttosupporttriple foam or equivalent operations.

SPECIFICATIONS

PRINCIPLES OF OPERATION

Page | 1

SYSTEM REQUIREMENTS

Power requirements

10 GPM pump Air-actuated valves

208-230V/3PH/8A or 460V/3PH/4A or 575V/3PH/4A

24VDC, 24VDC or 120VAC 3.5 watts per port

Space requirements 38” W x 44” H with pump assembly

Water inlet line 3/4” ID

Pneumatics operating pressure 80 - 100 PSI - Do not operate valves above 100 PSI

Operating ambient temperature 40 - 120° F

Solution outlet Up to 3.25 GPM 3.25 GPM - 5.5 GPM 5.5 GPM - 8 GPM

3/8”ID(1/2”polyflow) 1/2” ID 5/8” ID

Operating water pressure 200 PSI factory set (Assuming 40 PSI city feed), 180 PSI if tank fed

Max. water source temperature Recommended 110˚F (43˚C) - 137˚F (58°C) with Booster Pump

Air inlet line 3/8”ODpolyflow

Air outlet line 3/8”ODpolyflowperapplication

Air inlet pressure 20 CFM @ 80 - 120 PSI dry air

Page 5: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2016 Page | 2

AQUA-LAB EVO™ COMPONENTS

H20 InletPressure Gauge

Air Pilot Solenoid Valves 1-12 starting at the bottomright hand side and working counterclockwise

Air Pilot ActuatedHydra-Cannon Valve –1-12 starting at the bottom right hand side and working counterclockwise

Air Pilot Solenoid Valves 9-12

Air Pilot Solenoid Valves 5-8

Foaming Air Solenoid Valves 13-15

Air Pilot Solenoid Valves 1-4

Air PilotSolenoid Valve 16For Inlet Water Selector (optional)

Pressure Switch(optional)Foaming Air or

Frost Protection AirValves + Regulators(Optional)

Three Phase Power In

Multi ConductorSignal Cable

Three PhasePower to Pump

Wye Strainer

Bypass/PressureRegulator

Inlet Water Selection Valve

(Optional)

Primary AirRegulator

10 GPM QuickChange Pump

Strain Relief

IntegratedElectrical

Control Box

Mounting Slots(2 top, 2 bottom)

1

12 11 10 9

8

7

6

5

4

3

2

1

See pg. 17 for replacement parts

Page 6: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2017 Page | 3

L1

L2

L3

A2

1816

15B

1A

1

NOT3T2T1

NOL3L2L1

A2A1

95

T3T2T1

Pump3 HP

989796

Three Phase Disconnect

Off Delay Timer

240 OR480 VAC

Diode Terminal Block

Powe

r In

Contacter

RemotePressureSwitch

(V-)

ThermalOverload

Wire / Color#1#2#3#4#5#6#7#8#9#10#11#12

#1/White#2/White#3/White#4/White#5/White#6/White#7/White#8/White#9/White

#10/White#11/White#12/White

#13#14#15#16

#13/White#14/White#15/White#16/White

#18#19

#17#17#17#17#17#17#17#17

#1 & 2/Black#3 & 4/Black#5 & 6/Black#7 & 8/Black

#9 & 10/Black#11 & 12/Black#13 & 14/Black#15 & 16/Black

Over Load Must Be Set To 115% Of Motor FLA

Ground

(Control Voltage)

(V+)(Control Voltage)

INTEGRATED MOTOR STARTER WIRING DIAGRAM

Page 7: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2016 Page | 4

5 EASY STEPS FOR INSTALLATIONCOMPLETE PRE-INSTALLATION CHECKLIST BEFORE THESE STEPS (Appendix A pg. 19)1. Unpack

2. Hang the equipment

3. Make connections

4. Start-up

5. Optimizing the system (see pg. 7)

INSTALLATION

HFI TERMINAL POSITION WIRE NUMBER

AIR ACTUATED HYDRA-CANNON VALVE-1 WIRE #1

AIR ACTUATED HYDRA-CANNON VALVE-2 WIRE #2

AIR ACTUATED HYDRA-CANNON VALVE-3 WIRE #3

AIR ACTUATED HYDRA-CANNON VALVE-4 WIRE #4

AIR ACTUATED HYDRA-CANNON VALVE-5 WIRE #5

AIR ACTUATED HYDRA-CANNON VALVE-6 WIRE #6

AIR ACTUATED HYDRA-CANNON VALVE-7 WIRE #7

AIR ACTUATED HYDRA-CANNON VALVE-8 WIRE #8

AIR ACTUATED HYDRA-CANNON VALVE-9 WIRE #9

AIR ACTUATED HYDRA-CANNON VALVE-10 WIRE #10

AIR ACTUATED HYDRA-CANNON VALVE-11 WIRE #11

AIR ACTUATED HYDRA-CANNON VALVE-12 WIRE #12

FOAMING AIR VALVE-13 WIRE #13

FOAMING AIR VALVE-14 WIRE #14

FOAMING AIR VALVE-15 WIRE #15

INLET SELECTOR-16 WIRE #16

COMMON-17 WIRE #17

+ CONTROL VOLTAGE-18 WIRE #18

LOW WATER FAULT-19 WIRE #19

Page 8: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2017 Page | 5

General Skill Level• Mechanical: Basic - mounting equipment.

• Electrical: Advanced - three phase power and controls knowledge (local codes knowledge required).

• Plumbing: Moderate – principal supply line required.

INSTALLATION INSTRUCTIONS

WHO TASK EST. TIME

DISTRIBUTOR / TECHNICIAN HANG EQUIPMENT 1 HR

TECHNICIAN CONNECT WATER, AIR AND SOLUTION LINES 1/2 HR

TECHNICIAN CONNECT CONTROL LEADS TO MAIN CONTROLLER OR JUNCTION BOX 1 HR

DISTRIBUTOR/TECHNICIAN STARTUP (INJECTOR, METERING TIP AND NOZZLE SELECTION) 1/2 HR

DISTRIBUTOR/TECHNICIAN DOCUMENT CONFIGURATION 1/2 HR

TOTAL LABOR HOURS 3 1/2 HRS

WHO TASK EST. TIME

DISTRIBUTOR MONITOR & RECORD PERFORMANCE 1 HR / WK

DISTRIBUTOR MAINTENANCE PER SCHEDULE OR AS NEEDED

POST INSTALLATION

TOTAL HOURS SPENT

INSTALLATION

TOTAL CUSTOMER 1/4 HR

TOTAL DISTRIBUTOR/TECHNICIAN 6 HR

TOTAL ELECTRICIAN 1 HR

TOTAL PLUMBER 1 HR

Installation takes approximately ONE day. An electrician and a plumber are needed for half a day.

PRE-INSTALLATION

ESTIMATED TIMELINE

WHO TASK EST. TIME

DISTRIBUTOR & CUSTOMER DETERMINE LOCATION TO INSTALL EQUIPMENT 1/4 HR

PLUMBER INSTALL WATER SUPPLY LINE 1 HR

ELECTRICIAN INSTALL ELECTRICAL SUPPLY LINE 1 HR

TECHNICIAN LABEL ALL CONTROLLER RELAYS AT CONTROLLER 1/4 HR

TECHNICIAN RUN SOLUTION AND AIR LINES (IF NECESSARY) 1 HR

TECHNICIAN INSTALL AIR SUPPLY LINE 1 HR

TOTAL LABOR HOURS 4 1/2 HRS

Page 9: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2016

PNEUMATIC CONNECTIONS• Connect pre-run 3/8” OD poly air feed line to push connect

fittingontheinletsideoftheprimaryregulator.

• Connect 3/8” OD poly lines from EVO to foam generator(s).

**If there are unused air ports, back out the individual line regulator until airnolongerflows.

Page | 6

UNPACKINGThe AQUA-LAB EVO is shipped in a wooden crate for protection.

1. Cut straps holding boxes onto pallet.

2. Lift the pump from the pallet. Use assistance if necessary.

3. ***Be sure not to discard the manuals and accessories box.

LOCATION & MOUNTING**IflocationwasnotidentifiedduringthePre-InstallationProcess,makesuretoconsidertheproximityto feed water, power supply, and the control cabinet as well as space near the system to store chemical containers.

***See drawing in reference for general layout (in appendix A)

Pump Assembly - Need To Be Mounted On Left Side Of Aqua-Lab Evo Panel1. Drill (4x) 3/8” holes in wall for pump mounting bracket.

2. Insert concrete anchors, set pump mounting bracket on bolts & tighten down.

3. Slide pump onto stand.

AQUA-LAB EVO1. Drill (4) 5/32” holes on the wall. 19 1/4” wide, 33” tall.

2. Press EVO board tight against the wall.

3. Screw in 3/16” concrete screws with a washer (not included).

FEED WATER CONNECTION**Prior To Connection, Ensure That The Feed Lines Are Free Of Debris By Flushing Out The Lines For 15 Minutes.

• Connect pre-run main water supply line to pump inlet 3/4” NPT female connection.

• Hammer

• Tape measure

• Level

• Utility knife

• Wire stripper

• Socket set

• Adjustable wrench

• Screw driver set

• Teflontape

Tools & Equipment Needed• Drill with Phillips head

• Concrete drill bit 3/8”

• Concrete drill bit 5/32”

• Pneumatic: Basic- pneumatic utility connection required.

• Chemical Knowledge: Moderate - chemical titrations required.

Page 10: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2017

PUMP PRIMING INSTRUCTIONS (ASSUMES > 40 PSI CITY FEED TO PUMP)1. Pull the pump outlet line at the manifold quick-connect. Open ball valve and direct toward a drain or

container to remove the majority of the air from the pump until a steady streamofwaterisflowing(approx.1min).

2. Start the pump for a short period of time. ENSURE THAT PUMP ROTATION IS CORRECT as indicated by the arrow on the bottom of the pump and that 200 psi can be reached.

• If pump cannot regulate to 200 psi, look at flat spot on the shaft or remove pump motor cover and look at shaft to confirm correct rotation.

3. Start the pump and slowly open ball valve until it is wide open. Allow to runfor60secondstoflushlinesandthenclosevalve.

4. Reconnect the high pressure line to the Hydra-Cannon Manifold and open valve.

5. Confirmthatthepumpcanobtain200psi(180psifortankfed)whilefiringsolenoidsandthat the pump housing (stainless steel tube) is cool to the touch after a minute in operation.

• If housing is hot or noisy, pump did not prime correctly.

• It is normal for external housing of the motor to exceed 150° F.

6. If pump does not prime, repeat steps 3-5.

7. Verify pump prime 24 hours after operation to ensure prime held. Pay close attention to the temperature of the pump housing (stainless steel area).If it starts getting hotter than the supply water or greater than 140° F, then it is likely that the pump did not prime correctly which WILL CAUSE DAMAGE to pumps. The motor housing (painted portion) will be hot during operation.

START UP

!WARNING!PUMP MUST BE PRIMED BEFORE OPERATION.

Page | 7

CONSISTENTLY ACHIEVE THE DESIRED CLEANING AND PRESENTATION/PERFORMANCE USING THE LEAST AMOUNT OF CHEMICAL AND WATERINJECTORS VS. METERING TIPS VS. NOZZLESTHE KEY TO OPTIMIZING THE SYSTEM IS THROUGH TRIAL AND ERROR. DON’T BE AFRAID TO TRY THESE STEPS TO ACHIEVE YOUR IDEAL PERFORMANCE.

OPTIMIZING THE SYSTEM

Page 11: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2016 Page | 8

What do injectors do?• Injectors increases or decreases the amount of water in the solution

What do metering tips do?• Metering tips increases or decreases the amount of chemical in the solution

What do nozzles do?• Nozzles determine the pattern and back pressure of the solution

APPLICATION OPTIMIZATION(REPEAT FOR EACH APPLICATION)• Application too wet

• Increase air• Reduce injector size (decreases water)• Increase metering tip (increases chemical)

• Application too dry• Decrease air• Increase injector size (increases water)• Decrease metering tip (decreases chemical)

• Nozzle sputters• Decrease air• Decrease nozzle(s) and/or size used on arch• Increase injector size (increases water)

• Too much chemical used• Decrease metering tip• Decrease metering tip and injector size (to maintain desired ratio)

• No chemical • Check vacuum of injector (see instructions on page 9 or page 13 if your using HFI’s Injector

Optimization Tool.)• Check foot valve• Check metering tip

• Nozzle fan pattern not filled• Reduce nozzle size• Increase injector size (increases water)

• Water not present at all nozzles on arch• Verify check valves are functioning• Verify nozzles are not plugged• Reduce number of nozzles• Reduce nozzle size• Increase injector size (increases water)

CHEMICAL USAGE MEASURINGVERIFY TITRATION OF CHEMICALS BEFORE PROCEEDING1. Set up lab scale with small bucket of chemical to be measured.

2. Put the suction line into the bucket.

3. Run the application being tested to “prime” the line. (All air bubbles must be removed for accuracy.)

4. Record the Initial Weight from the scale. (Tarring the scale with weight on the scale can affect accuracy.)

5. Run the application for 6 vehicles (or manually for the same amount of time it would be on for 6 vehicles).

6. Record the Final Weight from the scale.

7. Subtract the Initial Weight from the Final Weight to determine the weight of used product.

8. Divide this Used Weight by 6 to get a per car weight.

9. Divide the Per Car Weightingramsbythespecificgravityofthechemicaltodeterminethemillilitersof chemical used per vehicle.

10. Repeat for each chemical application.

0

8

16

24

Page 12: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2017 Page | 9

RECOMMENDED MAINTENANCETHE RECOMMENDED SERVICE AND MAINTENANCE ON THE AQUA-LAB SYSTEM ARE AS FOLLOWS. Monthly

• Check air regulator.• Checkwaterfilterandreplaceasneeded(ifinstalled).

Semi-Annually• Check and replace injector metering tips.• Check and clean wye strainer.• Inspect and replace chemical lines as needed.• Ensure chemical draw lines are tightly secured to injector hose barbs. Chemical draw lines loosen

overtime which can cause deviation in chemical usage; clip 1” off old hose as needed that was stretched by hose barb.

Annually• Clean water regulator.• Inspectmotorstarterforcorrosion,ifidentifiedorderreplacement/spareparts.

1-3 Years• Inspect and replace injectors.• Replace water valves.• Replace main water pressure regulator.

TROUBLESHOOTING

C

B

INJECTOR VACUUM CHECK (FOR TROUBLESHOOTING INJECTORS) 1. At the Chem-Flex injector, remove the chemical feed line from

the injector hose barb.

2. Attach the tubing of the vacuum gauge to the Chem-Flex hose barb (Image A).

3. With the pump(s) on, manually activate the chemical that is to be tested at the main car wash control cabinet. An injector that is working properly will have a reading greater than or equalto(≥)20inHg

4. If vacuum reads <20 in/Hg (image B), remove solution metering tip (image C) and retest. a. If retest vacuum reads >20 in/Hg (image D), The solution metering tip is clogged. Replace the metering tip.

b. If Retest vacuum reads <20 in/Hg, continue to STEP 5

5. Remove a nozzle on the arch or the chemical feed line from the foam generator and retest vacuum. a. If retest vacuum reads >20 in/Hg, back pressure is being created. Continue to STEP 6.

b. If back pressure is not still not being created try these steps and retest after each:

1. Clean nozzle tips.

2. Loosely replace media in foam generator. Do no over pack.

A

Page 13: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2016

INJECTOR ISSUESPROBLEM POTENTIAL CAUSES SOLUTIONSInjector is not drawing chemical - PASSES vacuum pressure check

Clogged chemical feed Check chemical hose, foot valve, metering tip and hose barb for debris or clogs.

Too much back pressure on injector Clean or replace downstream check valves, increase nozzle size or quantity, use larger tubing, or use smallerflowinjectors.

Clogged injector check valve Blow compressed air through the chemical hose barb on the injector to remove debris.

Injector is not drawing chemical - Clogged injector nozzle Remove injector and blow out any debris with compressed air.FAILS vacuum pressure check

Defective injector Replace injector.

Productspecific-Sonny’sRainBar Remove elbow at inlet to foam generator and remove nozzle.

Manifold inlet clogged (rare) Removeendfittingsandretentionrod.Cleanoutinletholestoallowfullflow.

Valve malfunction, valve not opening Ensure minimum 60 psi on primary air regulator, ensure valve receiving signal.

No flow from injector Clogged injector Remove injector and blow out debris with compressed air.

No water supply Check that the system has a supply of water.

Injector stainless steel disintegrating

Strong Acid Call Hydra-Flex and order composite version of injectors.

Page | 10

3. Decrease air pressure for foaming.

4. Try smaller injector (this will produce less flow and thus less back pressure).

c. If retest vacuum reads <20 in/Hg, replace injector and retest. If vacuum continues to read <20 in/Hg, call your service provider.

6. Repeat steps 2-5 for each chemical lane that a vacuum reading is needed for.

7. Once testing is complete, turn off the AQUA-LAB pump from the main car wash control cabinet.

There is a variation of performance in the injectors that comes from slight variations in the dimensions of the parts and in assembly that are unavoidable. It is common to see the resultant vacuum range from 20 in Hg all the way up to 28. There is also variation in the through hole size on the meter tips from Dema (within their manufacturing tolerances). Using the same tip color from site to site is a good starting point. However with the potential for variation from part to part it is reasonable to still need to do some adjustments from there.

D

Page 14: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2017

AIR OPERATED VALVE REPLACEMENT(PN: 3000931)1. Shut off the ball valve to Hydra-Cannon manifold.

2. Disconnect air line from front of valve.

3. Unscrewquickconnectfittingbyhand(DON’T LOSE BLACK WASHER).4. Unscrew valve assembly from the Hydra-Cannon manifold.

5. Screw new valve into manifold until hand tight and threaded pilot port is facing forward.

6. Removethecapfrompilotportandthreadinquickconnectfittingtofront of valve – Hand Tight ONLY.

7. Pushairlinebackintofitting.

8. Open the ball valve to the Hydra-Cannon manifold.

PUMP ISSUES

Unscrew from manifold using this portion of valve

PROBLEM POTENTIAL CAUSES SOLUTIONSPump Operates, But Only Delivering 100-150 Psi Incorrect motor rotation

Reverse rotation by interchanging two power leads in motor starter box.

Pump Operates, But Delivers Little Or No Water

Pump not priming See priming instructions.

Missing 1 of 3 phases Wire according to diagram.

Inadequate water supplyCheck pressure on inlet side of pump to be sure positive pressure is maintained.

Undersized piping Replace with larger piping.

Leak on the inlet side Make sure connections are tight.

Worn or defective pump partsReplace worn parts or entire pump, clean parts if required.

Pump Will Not Start Or Run AtFull Speed

Constant hot not connectedMake sure constant control voltage is supplied in car wash controller.

Blown fuse or circuit breaker

Could be due to blown pump motor. Try to turn breaker back on or replace fuse. If breaker trips after trying to firethemotorismostlikelyburnedout.Replacewithnew motor and pump.

Defective motor starter contactor Replace motor starter contactor.

Thermal overload set too low Adjust setting on thermal overload to match voltage.

Incorrect motor voltage Voltage must be within 10% of motor rated voltage.

Defective motor Replace motor.

E-Stop is depressed Unlock E-stop.

Pump components damaged Replace worn part or entire pump.

Not primed Re-prime pump.

Page | 11

Page 15: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2016

PROBLEM POTENTIAL CAUSES SOLUTIONS

Excessive Noise From Pump

Pumpnotsecuredfirmly Secure properly.

Restricted inlet Clean or correct restriction.

Waterregulatorfluttering/chattering

Adjust regulator down and then back up or replace regulator.

Cavitation (sounds like marbles in pump)

Increase inlet size.

Worn mechanical seal Replace pump.

Pump Leaks Loosefittings,andornotenoughthread tape

Tightenfittings,andortakepartoffandputnewthread tape on.

PRESSURE REGULATOR ISSUES

FLOW / ARCH ISSUES

VALVE ISSUES

PROBLEM POTENTIAL CAUSES SOLUTIONS

System Won’t Regulate Up To 200 Psi

Pump not primed Follow priming instructions.

Debris in regulator Remove regulator and clean out debris.

Motor rotation incorrect Verify rotation.

Opening too many valves at onceSystem is limited by size of pump and size of injectors, increaseflowbyaddingsecondarypumpsorreducesize / number of injectors open.

Defective check valve (if applicable) Replace check valve.

Defective Regulator Replace regulator.

Defective Pump Replace Pump.

PROBLEM POTENTIAL CAUSES SOLUTIONS

Flow At Arch Is Too Low

Incorrectinjectorflowrateselection

Replace with larger injector.

System pressure too low Ensure system pressure is set at 200psi.

Foam generator plugged Ensure cleaned and clear.

Downstream plumbing restrictiveIncrease size of plumbing / tubing, ensure check valves are cleaned or new, reduce elbows in line or other turns that would restrict.

PROBLEM POTENTIAL CAUSES SOLUTIONS

Valve Will Not Open

Air pressure too lowEnsure primary air regulator reading at least 60 psi, turn up to 80-90psi if possible and check again.

System pressure too low

Remove valve from manifold, Carefully remove top of valve (caution – under high spring pressure) push white piston up with small allen wrench from opposite end and check o-ring condition. Replace and lubricate if needed.

Valve Leaks Air Or Water Out Top

Internal o-ring seal damaged / worn

Remove valve from manifold, Carefully remove top of valve (caution – under high spring pressure) push white piston up with small screwdriver from opposite end and check o-ring condition. Replace with 018 Viton O-ring and lubricate with Dow 111 valve lube.

Valve Remains Open After Signal Is Off Manifold pressure is above 230 psi

Reduce pressure to manifold to 200 psi operating pressure.

Page | 12

Page 16: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2017

BACKGROUND:This tool is for initial setup and troubleshooting of Chem-Flex™ Injectors and an Aqua-Lab™ Chemical Dispensing System. In order for the injector to work properly and draw chemical this gauge must be in the “GREEN” section when installed immediately after an injector that is running. If the gauge is in the red you will either see: intermittent chemical, no chemical draw, or chemical being applied at a very low pressure.

Back pressure refers to the pressure in the solution output line. Excessive back pressure is the main reason that injectors will not draw. If there is ever any concern to why an injector is not drawing chemical, the best and easiest way to diagnose the problem is to check the back pressure. See instructions below:

INJECTOR OPTIMIZATION TOOL

STEPS:1. Plug the optimization tool into the

outlet line of injector and connect solution output line.

2. Turn on function from car wash controller to actuate Hydra-Cannon valvesuchthatfluidisflowingthrough both the injector and injector optimization tool and out to the applicator.

3. Read injector optimization tool.

4. If the gauge is in the “RED ZONE” the back pressure of the outlet line is either too low or too high. See steps below to correct.

BACK PRESSURE TOO HIGH (UPPER RED SECTION):

(Back Pressure May Be Affected By One Or Several Of These Things)1. Foam generators are clogged/degraded. Clean or replace media in generator.

2. Injectorflowsizeistoolarge.Godownaninjectorsize(lessGPM).

3. Nozzle size on the arch is too small. Go up in nozzle size.

4. Check valves are dirty and or failing. Clean or replace check valves.

5. Thereisakinkinthelineorexcessfittings(elbowsandreducersincreasethebackpressure).Checklineandreplaceanykinkedsections.Trytoreducefittings.

6. ID of tubing going out to the tunnel is too small. Go up a size in inside diameter.

7. Check valves have too high of cracking pressure. Replace check valve with lower pressure check valve.

8. Clean foot valve.

BACK PRESSURE TOO LOW (LOWER RED SECTION):

(Back Pressure May Be Affected By One Or Several Of These Things)1. Injectorflowtoolow.Increaseinjectorsize.

2. Nozzle size too large. Reduce nozzle size.

SolutionOutput Line

Hydra-Cannon Valve

Injector Optimization Tool

Injector

Chemical Draw Line

Chemical Bucket

Page | 13

Page 17: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2016 Page | 14

Max Pressure InletNPT up to 1000 PSI (69 bar)

PC2 up to 500 PSI (34 bar)

Max Pressure Outlet 333 PSI (23 bar)

Temperature Range 33°F - 175°F (.5°C - 79°C)

Maximum Wrench Torque 30 ft-lbs (41 N-m)

SPECIFICATIONS

STANDARD SST CHEM-FLEX INJECTOR PART NUMBERSINLET CONNECTION PC2* (QUICK CONNECT)

OUTLET CONNECTION 3/8” MNPT

FOR USE WITH AQUA-LAB MODEL(S) MD GEN2, XD GEN2, HD

COLOR FLOW ORIFICE FLOW RATE @ 200 PSI (14 bar) SINGLE BARB DUAL BARB

White 0.029” (0.74 mm) 0.25 GPM (1.0 LPM) 618029 -

Yellow 0.040” (1.02 mm) 0.50 GPM (1.9 LPM) 618040 629040

Tan 0.051” (1.39 mm) 0.75 GPM (2.8 LPM) 618051 629051

Red 0.057” (1.45 mm) 1.00 GPM (3.8 LPM) 618057 629057

Orange 0.070” (1.78 mm) 1.50 GPM (5.7 LPM) 618070 629070

Gray 0.083” (2.11 mm) 2.00 GPM (7.6 LPM) 618083 629083

Dark Blue 0.086” (2.18 mm) 2.25 GPM (8.5 LPM) 618086 629086

Light Blue 0.095” (2.41 mm) 3.00 GPM (11.3 LPM) 618095 629095

Light Green 0.098” (2.49 mm) 3.25 GPM (12.3 LPM) 618098 629098

Pink 0.106” (2.69 mm) 3.75 GPM (14.2 LPM) 618106 629106

Purple 0.117” (2.98 mm) 4.50 GPM (17.0 LPM) 618117 629117

Dark Green 0.125” (3.18 mm) 5.50 GPM (20.8 LPM) 618125 629125

*For exclusive use with Aqua-Lab Chemical Dispensing Systems.

Page 18: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2017 Page | 15

REFERENCE - CHEMICAL DILUTION RATIOS

3.00

”1:

251

1: 5

031:

754

1: 1

006

1: 1

509

1: 2

012

1: 2

263

1: 2

416

1: 2

460

1: 2

628

1: 4

526

1: 5

532

1: 8

047

1: 1

0059

1: 1

2070

1: 1

5088

2.00

”1:

181

1: 3

631:

544

1: 7

261:

108

91:

145

11:

163

31:

140

21:

133

71:

151

01:

326

61:

399

11:

580

61:

725

71:

870

81:

108

85

1.00

”1:

104

1: 2

081:

311

1: 4

151:

623

1: 8

311:

934

1: 8

351:

778

1: 8

981:

186

91:

228

41:

332

21:

415

31:

498

31:

622

9

0.75

”1:

82

1: 1

651:

247

1: 3

291:

494

1: 6

591:

741

1: 6

421:

668

1: 7

551:

148

31:

181

21:

263

61:

329

51:

395

41:

494

2

0.50

”1:

59

1: 1

191:

178

1: 2

381:

357

1: 4

751:

535

1: 4

871:

496

1: 5

581:

106

91:

130

71:

190

11:

237

61:

285

21:

356

4

0.25

”1:

34

1: 6

81:

102

1: 1

361:

204

1: 2

721:

306

1: 2

821:

304

1: 3

441:

612

1: 7

481:

108

81:

136

01:

163

21:

204

0

Spiral Plug Length

Flow

Rat

e(G

PM) a

t 200

PSI

0.25

0.50

0.75

1.00

1.50

2.00

2.25

3.00

3.25

3.75

4.50

5.50

8.0

10.0

12.0

15.0

Inje

ctor

Col

orW

hite

Yello

wTa

nRe

dO

rang

eG

ray

Blue

Ligh

t Bl

ueLi

ght

Gre

enPi

nkPu

rple

Dar

k G

reen

Blac

kBl

ack

Blac

kBl

ack

Noz

zle

Size

0.02

9”

(0.7

mm

)0.

040”

(1.0

mm

)0.

051”

(1.3

mm

)0.

057”

(1.4

mm

)0.

070”

(1.8

mm

)0.

083”

(2.1

mm

)0.

086”

(2.2

mm

)0.

095”

(2.4

mm

)0.

098”

(2.5

mm

)0.

106”

(2.7

mm

)0.

117”

(2.9

mm

)0.

125”

(3.2

mm

)0.

161”

(4.1

mm

)0.

117”

(4.5

mm

)0.

186”

(4.7

mm

)0.

207”

(5.3

mm

)

CO

PPER

1: 5

71:

104

1: 1

551:

195

1: 2

811:

406

1: 4

681:

598

1: 6

291:

739

1: 8

811:

107

41:

201

91:

231

81:

332

41:

425

9

PUM

PKIN

1: 4

31:

82

1: 1

191:

126

1: 2

381:

348

1: 3

981:

519

1: 5

541:

647

1: 7

741:

946

1: 1

465

1: 1

832

1: 2

310

1: 3

042

BURG

UN

DY

1: 3

41:

67

1: 9

71:

111

1: 2

071:

304

1: 3

471:

461

1: 4

951:

575

1: 6

901:

845

1: 1

244

1: 1

670

1: 1

975

1: 2

469

LIM

E1:

28

1: 5

71:

81

1: 1

001:

183

1: 2

701:

307

1: 4

141:

447

1: 5

181:

622

1: 7

641:

988

1: 1

262

1: 1

481

1: 1

958

TAN

1: 2

81:

57

1: 8

11:

100

1: 1

831:

270

1: 3

071:

414

1: 4

471:

518

1: 6

221:

764

1: 8

261:

104

21:

128

61:

155

7

ORA

NG

E1:

23

1: 4

41:

64

1: 7

81:

137

1: 1

961:

215

1: 2

931:

314

1: 3

651:

437

1: 5

361:

751

1: 8

741:

113

61:

126

2

TURQ

UO

ISE

1: 1

71:

31

1: 4

51:

55

1: 9

11:

126

1: 1

341:

185

1: 1

971:

230

1: 2

741:

336

1: 5

611:

752

1: 8

851:

109

9

PIN

K1:

14

1: 2

41:

35

1: 4

21:

68

1: 9

31:

98

1: 1

301:

143

1: 1

591:

188

1: 2

241:

372

1: 4

731:

570

1: 7

28

LIG

HT

BLU

E1:

11

1: 1

71:

24

1: 3

11:

47

1: 6

41:

66

1: 9

31:

98

1: 1

141:

136

1: 1

661:

318

1: 4

061:

483

1: 5

72

BRO

WN

1: 1

01:

15

1: 2

21:

28

1: 4

31:

58

1: 5

91:

84

1: 8

81:

103

1: 1

221:

150

1: 2

701:

338

1: 4

081:

507

RED

1: 1

21:

17

1: 2

31:

34

1: 4

51:

46

1: 6

51:

69

1: 8

01:

95

1: 1

161:

202

1: 2

541:

306

1: 3

75

WH

ITE

1: 1

21:

16

1: 2

21:

31

1: 4

21:

43

1: 6

11:

64

1: 7

41:

88

1: 1

081:

180

1: 2

261:

272

1: 3

43

GRE

EN1:

11

1: 1

41:

20

1: 2

81:

37

1: 3

81:

53

1: 5

51:

65

1: 7

61:

94

1: 1

601:

201

1: 2

411:

300

BLU

E1:

10

1: 1

21:

17

1: 2

31:

30

1: 3

11:

44

1: 4

61:

53

1: 6

31:

77

1: 1

221:

152

1: 1

851:

224

YELL

OW

1: 9

1: 1

21:

16

1: 2

01:

22

1: 3

01:

31

1: 3

61:

42

1: 5

21:

80

1: 1

001:

120

1: 1

48

BLA

CK

1: 1

01:

13

1: 1

61:

17

1: 2

31:

24

1: 2

81:

33

1: 4

01:

61

1: 7

71:

92

1: 1

15

PURP

LE1:

6.6

1: 8

.31:

91:

10

1: 1

31:

13

1: 1

51:

17

1: 2

11:

31

1: 3

91:

46

1: 5

7

GRA

Y1:

5.3

1: 6

.71:

6.9

1: 7

.61:

9.9

1: 1

01:

11.

61:

13

1: 1

61:

24

1: 3

01:

35

1: 4

4

OPE

N1:

4.9

1: 5

.31:

5.2

1: 6

.01:

6.1

1: 6

.11:

7.6

1: 8

.31:

10

1: 1

21:

15

1: 1

81:

23

Metering TipNO

TE: D

iluti

on

rati

os

giv

en a

re b

ased

on

dra

win

g w

ater

thr

oug

h th

e m

eter

ing

tip

s an

d a

re m

eant

as

a st

arti

ng p

oin

t fo

r sy

stem

co

nfig

urat

ion.

R

esul

ts a

re e

xpec

ted

to

var

y w

hen

dra

win

g c

hem

ical

s d

ue t

o d

iffer

ence

s in

vis

cosi

ty a

nd t

emp

erat

ure.

The

re m

ay b

e sl

ight

var

iati

ons

of

per

form

ance

in

inje

cto

rs a

nd m

eter

ing

tip

s th

at a

re u

navo

idab

le d

ue t

o m

anuf

actu

re t

ole

ranc

es. U

sing

the

sam

e ti

p c

olo

r fr

om

sit

e to

sit

e is

a g

oo

d s

tart

ing

po

int.

H

ow

ever

wit

h th

e p

ote

ntia

l fo

r va

riat

ion

fro

m p

art

to p

art

it is

rea

sona

ble

to

sti

ll ne

ed t

o d

o s

om

e ad

just

men

ts f

rom

the

re.

3/7”

po

lyflo

w (L

LDP

E) t

ubin

g is

req

uire

d t

o e

nsur

e a

seal

bet

wee

n th

e tu

be

wal

l and

the

sp

iral

plu

g.

(Ass

umes

feed

pre

ssur

e of

200

PSI

(14

bar)

)

Page 19: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2016 Page | 16

PRESSURE LOSS IN RUN LENGTH

All solution line tubing should be selected for 10 PSI or less of pressure loss.

*20 PSI pressure loss may be acceptable depending upon nozzle sizing, foamers, check valves and other line restrictions present in application.

All numbers represent pressure loss in PSI for selected solution line tubing.

GREEN = GOOD YELLOW= USE CAUTION RED = NOT RECOMMENDED

INJECTORFlow Rate (GPM) at 200 PSI 0.25 0.50 0.75 1.00 1.50 2.00 2.25 3.00 3.25 3.75 4.50 5.50

Injector Color White Yellow Tan Red Orange Gray DarkBlue

LightBlue

Light Green Pink Purple Dark

Green

Nozzle Size 0.029”(0.7 mm)

0.040”(1.0 mm)

0.051”(1.3 mm)

0.057”(1.4 mm)

0.070”(1.8 mm)

0.083”(2.1 mm)

0.086”(2.2 mm)

0.095” (2.41 mm)

0.098”(2.5 mm)

0.106” (2.69 mm)

0.117”(2.9 mm)

0.125”(3.2 mm)

150’ 1 PSI 1 PSI 1 PSI 1 PSI 2 PSI 3 PSI 4 PSI 6 PSI 7 PSI 9 PSI 13 PSI* 18 PSI*125’ 1 PSI 1 PSI 1 PSI 1 PSI 2 PSI 3 PSI 4 PSI 5 PSI 6 PSI 8 PSI 11 PSI* 16 PSI*100’ 1 PSI 1 PSI 1 PSI 1 PSI 2 PSI 2 PSI 3 PSI 4 PSI 5 PSI 6 PSI 9 PSI 13 PSI*75’ 1 PSI 1 PSI 1 PSI 1 PSI 1 PSI 2 PSI 3 PSI 3 PSI 4 PSI 5 PSI 7 PSI 9 PSI50’ 1 PSI 1 PSI 1 PSI 1 PSI 1 PSI 1 PSI 2 PSI 2 PSI 3 PSI 3 PSI 5 PSI 6 PSI25’ 1 PSI 1 PSI 1 PSI 1 PSI 1 PSI 1 PSI 1 PSI 1 PSI 2 PSI 2 PSI 3 PSI 4 PSI

Run

Leng

th

5/8” ID BRAIDED

INJECTORFlow Rate (GPM) at 200 PSI 0.25 0.50 0.75 1.00 1.50 2.00 2.25 3.00 3.25 3.75 4.50 5.50

Injector Color White Yellow Tan Red Orange Gray Blue LightBlue

LightGreen Pink Purple Dark

Green

Nozzle Size 0.029”(0.7 mm)

0.040”(1.0 mm)

0.051”(1.3 mm)

0.057”(1.4 mm)

0.070”(1.8 mm)

0.083”(2.1 mm)

0.086”(2.2 mm)

0.095” (2.41 mm)

0.098”(2.5 mm)

0.106” (2.69 mm)

0.117”(2.9 mm)

0.125”(3.2 mm)

150’ 1 PSI 1 PSI 2 PSI 3 PSI 6 PSI 9 PSI 11 PSI* 18 PSI* 21 PSI 27 PSI 36 PSI 56 PSI125’ 1 PSI 1 PSI 2 PSI 3 PSI 5 PSI 8 PSI 9 PSI 15 PSI* 18 PSI* 23 PSI 30 PSI 47 PSI100’ 1 PSI 1 PSI 2 PSI 2 PSI 4 PSI 7 PSI 8 PSI 12 PSI* 14 PSI* 18 PSI* 24 PSI 36 PSI75’ 1 PSI 1 PSI 1 PSI 2 PSI 3 PSI 5 PSI 6 PSI 9 PSI 11 PSI* 14 PSI* 18 PSI* 26 PSI50’ 1 PSI 1 PSI 1 PSI 1 PSI 2 PSI 4 PSI 4 PSI 6 PSI 8 PSI 9 PSI 12 PSI* 18 PSI*25’ 1 PSI 1 PSI 1 PSI 1 PSI 1 PSI 2 PSI 2 PSI 3 PSI 4 PSI 5 PSI 6 PSI 9 PSI

Run

Leng

th

1/2” ID BRAIDED

INJECTORFlow Rate (GPM) at 200 PSI 0.25 0.50 0.75 1.00 1.50 2.00 2.25 3.00 3.25 3.75 4.50 5.50

Injector Color White Yellow Tan Red Orange Gray Blue LightBlue

LightGreen Pink Purple Dark

Green

Nozzle Size 0.029”(0.7 mm)

0.040”(1.0 mm)

0.051”(1.3 mm)

0.057”(1.4 mm)

0.070”(1.8 mm)

0.083”(2.1 mm)

0.086”(2.2 mm)

0.095” (2.41 mm)

0.098”(2.5 mm)

0.106” (2.69 mm)

0.117”(2.9 mm)

0.125”(3.2 mm)

150’ 1 PSI 5 PSI 7 PSI 11 PSI* 22 PSI 36 PSI 44 PSI 76 PSI 88 PSI 115 PSI 154 PSI 239 PSI125’ 1 PSI 4 PSI 6 PSI 9 PSI 18 PSI* 30 PSI 37 PSI 63 PSI 73 PSI 101 PSI 129 PSI 200 PSI100’ 1 PSI 3 PSI 5 PSI 7 PSI 14 PSI* 23 PSI 28 PSI 50 PSI 54 PSI 77 PSI 103 PSI 142 PSI75’ 1 PSI 2 PSI 4 PSI 6 PSI 12 PSI* 18 PSI* 22 PSI 37 PSI 42 PSI 53 PSI 77 PSI 106 PSI50’ 1 PSI 1 PSI 3 PSI 4 PSI 8 PSI 12 PSI* 15 PSI* 25 PSI 28 PSI 39 PSI 52 PSI 73 PSI25’ 1 PSI 1 PSI 1 PSI 2 PSI 3 PSI 6 PSI 7 PSI 12 PSI* 13 PSI* 19 PSI* 26 PSI 34 PSI

Run

Leng

th

3/8” ID, 1/2” OD POLY TUBE

Page 20: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2017

T EF

U

NP

O

G

A,B,C,D

K

UH

I,J

M

VZY

L

QRS

1

12 11 10 9

8

7

6

5

4

3

2

1

W or X

G

REPLACEMENT PARTS LIST - AQUA-LAB EVO™

Page | 17

AE

C

DB

Page 21: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2016

PART NAME PART NUMBER

A Contactor

24 VAC - 3001336

24 VDC - 3001337

120 VAC - 3001338

B Thermal Overload 3001335

C Diode Terminal Block 3001200

D Terminal Block 3000338

E Timer Relay 3000664

F Din I x Field Wireable Cable - With LED 24 VAC/VDC - 3001157

120 VAC 3001156

G Air Actuated Hydra-Cannon Valve Replacement Kit - Composite 3000931

H 20 GPM Bypass Pressure Regulator - Brass 3000915

I 3/4” Wye Strainer - 20 Mesh 3000668

J3/4” NPT T16 Wye Strainer Screen & Gasket Kit 1001936

3/4” NPT T15 Wye Strainer Screen & Gasket Kit 1001937

K 1/2” Inlet Ball Valve 3000565

L Solenoid Actuated Air Valve - Non Foaming

24VAC - 3001152

24 VDC - 3001153

120 VAC - 3001154

M Solenoid Actuated Air Valve - Foaming

24 VAC - 1001428

24 VDC - 1001429

120 VAC - 1001430

N Foaming Small Air Regulator & 0-60 PSI Gauge 3000808

O Primary Air Regulator 3001160

P Thermal Relief Valve - 1/2” NPT 3000323

Not Shown 1/4” Hose Barb Foaming Air Check Valve 3000819

Q 3/4” Air Actuated 3-Way Ball Valve Inlet Selector 3001165

R 3/4” Male Quick Connect - Brass 3000196

S 3/4” Female Quick Connect - Brass 3000197

T Flow-Thru Adapter 3000173

Not Shown Duo-Foam Manifold (Injectors Not Included) 1001232

U Manifold Inlet Hose 1/2” ID hose X 1/2” NPT Ends 72” (1.8 m) Long 3000579

Not Shown Hydra-Cannon Valve O-Ring Replacement Kit 1001155

V Low Water Fault Kit 3001205

W 10 GPM Pump 208-230V/3PH/8A or 460V/3PH/4A 3001159

X 10 GPM Pump 575V/3PH/4A 3001434

Y 1/2” Female Quick Connect - Brass 3000502

Z 1/2” Male Quick Connect - Brass 3000500

Page 22: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2017

AQUA-LAB ACCESSORIES

PART NAME PART NUMBERDuo-Foam Kit 1001869

Triple-Foam Kit 1001948

Quad-Foam Kit 1001949

High-Flow Rinse Kit 1001230

Anti-Siphon Kit (PC2) 1001436

Injector Optimization Tool 1002018

Vacuum Tester Kit 1001104

Water Filtration System - Cold 1001431

Water Filtration System - Cold Replacement Filter 3000855

PC2 Plug - Stainless Steel 3000022

Flow-Through Adapter 3000173

Vinyl Hose Barb Cap 3000745

Remote Metering Device/Check Hub 1001756

3/8” Coupler 3000522

3/8” Elbow 3000523

3/8” x 1/2” Push-To-Connect 3000526

3/8” Check Valve (PC2) 3000970

Page 23: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2016

FACTORY LIMITEDHydra-Flex, Inc warrants its equipment to be free from defect in material or workmanship under proper

normal use for a period of one (1) year beginning the date of purchase.

Hydra-Flex,Inc’sliabilityshallbelimitedtorepairorreplacementofpartsfoundtobedefectivewithinthe

warrantyperiodandfollowingHydra-Flex,Inc’sinspection.Hydra-Flex,Incshallhavetheoptionrequiring

thereturnofdefectivematerialtoestablishthepurchaser’sclaim.Intheeventofrepairorreplacement

this limited warranty is non-cumulative. Neither labor nor transportation charges are included in this

warranty.

This warranty is based upon the proper care and maintenance of the warranted equipment. Warranty

doesnotapplyifthemerchandiseisalteredormodifiedinanyway.Warrantydoesnotapplytoany

equipment which has been subject to misuse, inappropriate use of tools, including exposure to harsh

chemicals,neglect,lackofmaintenance,freezing,fluidhammer,accident,thirdpartydamage,fluid

impurities such as sand or minerals, acts of God or acts of war. Nor does it apply to any equipment which

has been repaired or altered by anyone not so authorized by Hydra-Flex, Inc. All equipment must be

properlyinstalledinaccordancewithspecifiedplumbing,electrical,andmechanicalrequirements.The

warranty does not apply to normal wear and tear or routine maintenance components as described in the

equipment manual.

Except as expressly stated herein, Hydra-Flex, Inc shall not be liable for damages of any kind in

connectionwiththepurchase,maintenance,oruseofthisequipmentincludinglossofprofitsandall

claims for consequential damages. This limited warranty is in lieu of all other warranties expressed or

implied. Hydra-Flex, Inc neither assumes nor authorizes any person to assume for it any other obligation

or liability in connection herewith. This warranty is neither assignable nor transferable.

Transportation damage claims are to be submitted to the carrier of the damaged material.

AQUA-LAB EVO™ WARRANTY

Page | 18

Page 24: AQUA-LAB EVO - Hydra-Flex, Inc....AQUA-LAB EVO™ COMPONENTS H20 Inlet Pressure Gauge Air Pilot Solenoid Valves 1-12 starting at the bottom right hand side and working counterclockwise

© Hydra-Flex, Inc. 2017

Distributor / Technician:

Chem. Supplier:

Car wash Owner:

Site Manager:

Electrician:

Plumber:

Other:

AQUA-LAB EVO PRE-INSTALLATION CHECKLISTSTEP 1: Determine who will be involved in installation

STEP 2: Determine Aqua-Lab location• EVO – 38” wide x 44” high • Pump–8”widex4’high

STEP 3: Electrician to provide breaker and line run to the location where the motor starter will be mounted per local code.

• 15 amp Breaker per Operating PumpSTEP 4: Run water supply line to location

• Single Pump – 3/4” Supply LineSTEP 5: Locate main car wash control panel. Identify and label all relays.STEP 6: Locate main compressed air line and drop a 3/8” feed to the location of the AQUA LAB.STEP 7: Run new ½” poly lines for solution from the gantry to the location of the AQUA LAB EVO.

• Carefully label each line for the application and chemical. STEP 8: Run new 3/8” poly lines for air from the gantry to the location of the AQUA LAB EVO. STEP 9: Confirm delivery of AQUA LAB EVO with HFI

Delivery Date: ________________________STEP 10: Coordinate Install date with all parties involved

Install Date: ________________________

APPENDIX A

Page | 19

34"

3612"

2114"

5014"

1914"

2014"

4518"

GENERAL LAYOUT

T: 952-808-3640 • F: 952-808-3650 • www.hydraflexinc.com•[email protected]

4000003 revL0717