Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

download Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

of 37

Transcript of Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    1/37

    http://irc.nrc-cnrc.gc.ca

    Applications and Uncertainties

    Associated with Measurements UsingFourier Transform Infrared Spectrometry,

    Gas Chromatography-Mass Spectrometry

    and Selective Ion Electrode Method in

    Fire Suppression Tests

    I RC - I R - 7 4 5

    Liu, Z.G.; S u, J .Z.; Kim, A.K .; Kanabus-Kaminska, J .M.; Lusztyk, E .

    J une 1997

    http://irc.nrc-cnrc.gc.ca/http://irc.nrc-cnrc.gc.ca/
  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    2/37

    I

    ABSTRACT

    During full-scale fire suppression tests conducted at the National Fire Laboratory

    of the National Research Council Canada, Fourier Transform Infrared (FTIR)

    Spectroscopy, Gas Chromatography-Mass Spectrometry (GC/MS) and Specific Ion

    Electrode (SIE) method were employed to monitor the concentrations of the gaseoussuppression agents in the test compartment. Also, these methods were used to determine

    the halogen acid gases generated during the fire suppression process. This report

    describes the application of these three analytical methods for a series of tests conducted

    with HCFC Blend A.. The results for the measurements of gaseous agent and halogen

    acid gas concentrations using these three methods are provided. The uncertainties

    associated with measurements using these three techniques are analyzed in detail.

    It is shown that FTIR Spectroscopy provides an instantaneous and continuous

    method for the measurement of multiple chemical species in the fire tests with reasonable

    accuracy. The GC/MS and the SIE methods involve periodical sampling and off-line

    measurements. Uncertainties associated with measurements using these analyticaltechniques are strongly dependent on the characteristics of the analytical methods

    themselves, the set-up of the sampling system, calibration procedure, sampling procedure,

    quantitative analysis and testing conditions. Considerable care is required to obtain

    accurate measurements in the presence of halogen acid gases.

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    3/37

    II

    TABLE OF CONTENTS

    ABSTRACT I

    TABLE OF CONTENTS II

    NOMENCLATURE IV

    1.0 INTRODUCTION 1

    2.0 TEST FACILITIES AND PROCEDURES 2

    2.1 Fire Suppression Agents 2

    2.2 Test Room 2

    2.3 Fire Scenarios 2

    2.4 Experimental Set-Up 2

    2.5 Test Instruments 3

    2.5.1 Fourier Transform Infrared (FTIR) Spectroscopy 3

    2.5.1.1 Calibration of the FTIR Spectroscopy 3

    2.5.1.2 Sampling and Analyzing Procedure of FTIR Spectroscopy 4

    2.5.2 Gas Chromatography-Mass Spectrometry (GC/MS) 42.5.2.1 Calibration of GC/MS 4

    2.5.2.2 Sampling and Analyzing Procedure of GC/MS 4

    2.5.3 Selective Ion Electrode (SIE) System 5

    2.5.3.1 Calibration of SIE 5

    2.5.3.2 Sampling and Analyzing Procedure of SIE 5

    2.6 Test Procedure 6

    3.0 TEST RESULTS 7

    4.0 UNCERTAINTIES ANALYSIS 9

    4.1 Uncertainties in the FTIR Spectroscopy 9

    4.2 Uncertainties in GC/MS 10

    4.3 Uncertainties in SIE System 115.0 SUMMARY 13

    REFERENCES 14

    Appendix 1: HCFC Blend A Concentrations Measured by GC/MS

    Appendix 2: F-

    and Cl-

    anion Concentrations Measured by SIE System

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    4/37

    III

    NOMENCLATURE

    FTIR Fourier Transform Infrared

    GC/MS Gas Chromatography-Mass Spectrometry

    HAPE Halon Alternative Performance Evaluation

    HCFC hydrochlorofluorocarbon

    HCl hydrogen chlorideHF hydrogen fluoride

    TDP thermal decomposition products

    SIE Selective Ion Electrode

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    5/37

    1

    1. INTRODUCTION

    Increasing concerns about the environmental consequences of the utilization of

    chemical fire suppression agents have recently prompted worldwide efforts to develop and

    evaluate environmentally-friendly replacements for Halon 1301 and Halon 1211 [1 - 5].

    An ideal halon replacement for total flooding applications should meet these importantcriteria: high fire suppression effectiveness, low direct and by-product toxicity, zero ozone

    depletion and short atmospheric lifetime. Various analytical techniques are being used to

    evaluate whether halon replacements meet such criteria or not [6].

    In full-scale fire tests for the Halon Alternatives Performance Evaluation (HAPE)

    project conducted at the National Fire Laboratory, the National Research Council Canada,

    Fourier Transform Infrared (FTIR) Spectroscopy, Gas Chromatography-Mass

    Spectrometry (GC/MS) and Selective Ion Electrode (SIE) method were used to monitor

    the concentrations of the agents used for fire suppression and to determine the combustion

    by-products generated during fire suppression [7]. Good results were obtained using

    these analytical methods. However, the test results indicated that there were someuncertainties associated with the measurements obtained using these test methods.

    Although these methods are often used in fire tests, no literature is available that discusses

    in detail the uncertainties associated with measurements obtained using these analytical

    methods.

    Previous research [8 - 11] reported that some difficulties exist when measuring

    combustion by-products in fire suppression tests, because of the severe testing conditions

    and probable contamination and/or loss of samples during the measurements. Dierdorf et

    al. [12] discussed the advantages and disadvantages of three measuring modes using a

    FTIR to determine combustion by-products in fire suppression test. Bulien [13] described

    the application of a FTIR spectrometer for measuring toxic smoke components in fire test.He indicated that missing information and details in calibration routines could lead to

    differences in results obtained using a FTIR spectrometer.

    This work describes the application of the FTIR, GC/MS and SIE methods in full-

    scale fire tests. The results of the measurements of the agent and the halogen acid gas

    concentrations using these three methods in Phase I of the HAPE project are compared.

    The uncertainties associated with measurements using these techniques are then analyzed

    in detail. The FTIR Spectroscopy provides an instantaneous and continuous method for

    the measurement of multiple chemical species during the fire tests. The GC/MS and the

    SIE methods involve periodic sampling and off-line measurements. Uncertainties related

    to measurements obtained using these techniques are strongly dependent on the

    characteristics of the methods themselves, the set-up of the sampling system, calibration,

    sampling procedure, quantitative analysis and testing conditions.

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    6/37

    2

    2.0 TEST FACILITIES AND PROCEDURE

    2.1 Fire Suppression Agents

    In Phase 1 of the HAPE project, HCFC Blend A was evaluated as a potential

    replacement for Halon-1301 [7]. HCFC Blend A is a mixture of halogenated

    hydrocarbons with 82% HCFC-22 (CHClF2), 9.5% HCFC-124 (C2HClF2), and 4.75%HCFC-123 (C2HCl2F3). The remaining 3.75% is a de-toxifying agent, lemonene, that

    gives the agent its characteristic odour. The design agent concentration for fire

    extinguishment and thermal decomposition products (mainly HF and HCl) generated

    during fire suppression were examined in full-scale fire tests. As shown in Table 1, the

    thirteen tests conducted in Phase 1 of the HAPE project involved a wide range of agent

    concentrations (8.6 - 14.7%), nozzle design, discharge manifold and fire scenarios.

    2.2 Test Room

    A 121 m3

    compartment, similar to Radar Room No.2 on the Navys Halifax Class

    frigates, was constructed inside the Burn Hall of the National Fire Laboratory. The testroom is shown in Figure 1. It has a steel access door, three viewing windows and a

    pressure relief vent near the floor of the South wall.

    2.3 Fire Scenarios

    Test fires in Phase 1 of the HAPE project included eight tell-tale (TT) fires, three

    50 kW square-pan (SP) fires and one 400 kW round-pan (RP) fire. The fire scenarios

    used for tests are given in Table 1. Heptane was used as the fuel for the tell-tale, square-

    pan and round-pan fires. Their locations in the compartment are indicated in Figure 1.

    2.4 Experimental Set-Up

    The mass loss rate of the agent as it discharged from the cylinder was measured by

    a weight scale with a digital readout. Thermocouples were placed at each fire location to

    determine the extinguishing time. Three thermocouple trees were set up in different

    locations of the compartment to measure room temperatures (see Figure 1).

    Three quartz ports connected to stainless steel sampling lines were placed at the

    same locations as the thermocouple trees. The sampling ports were arranged at different

    elevations to measure the distribution of the agent in the compartment. The elevation of

    these ports is indicated in Figure 1.

    As shown in Figure 2, the sample gases in the compartment were sampled

    continuously using three sampling quartz ports. The samples were filtered with microfiber

    filters in the sampling lines. A switching system, using solenoid valves, connected the

    FTIR to the gas-sampling lines from each of the three locations at 30 second intervals.

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    7/37

    3

    The sampling port for the collection of samples for measurement using GC/MS

    was located after the FTIR cell. These samples were used to determine the concentration

    of volatile organic compositions.

    Secondary, samples were obtained for analysis using SIE method. These samples

    were obtained using a 6 meter length of sampling line connected to the GC/MS samplingport (see Figure 2). The SIE method was used to determine acid gas concentration.

    All the sampling lines from the sampling ports in the room to the gas cell of the

    FTIR and the sampling port of the GC/MS were heated and maintained at the temperature

    of 155oC. This prevented the condensation of sample gases in the sampling line. The

    sampling line from the GC/MS sampling port to the SIE sampling port was not heated.

    All sampling lines were pumped continuously prior to the test and after the test to ensure

    clean gas samples.

    2.5 Test Instruments

    2.5.1 Fourier Transform Infrared (FTIR) Spectroscopy

    The FTIR spectroscopic method is based on the combination of an interferometer,

    usually of the Michelson type, with a sensitive infrared detector and a computer [14 - 17].

    FTIR spectroscopy provides sensitive and very fast measurements with high resolution

    over a wide frequency range for the identification of a broad range of chemical

    compounds. In the HAPE project, a BioRad FTS 175 FTIR spectrometer was used on-

    line to measure the concentrations of the agent and its thermal decomposition products

    during full-scale fire tests.

    2.5.1.1 Calibration of the FTIR Spectroscopy

    The calibration of the FTIR spectrometer for the agent, HCFC Blend A, was

    carried out by using commercial calibration gases for the three major components of the

    agent. The concentrations of HCFC-22, HCFC-123 and HCFC-124 used in the

    calibration mixture were 7.50%, 0.49% and 0.4%, respectively, which was the same as

    used in HCFC Blend A. Each HCFC component has a characteristic infrared absorption

    peaks. For HCFC-22, three infrared absorption peaks centered at frequencies of 3,025,

    2,213 and 596 cm-1

    , were selected as the reference. For components HCFC-123 and

    HCFC-124, the peaks centered at the frequencies of 697 and 672 cm-1

    were used as

    references, respectively.

    Commercial HF samples, with concentrations of 970, 5339, 11,400 and 75,000

    ppm, were used for calibration. HF has a unique absorption band in the frequency region

    of 3,600 to 4,300 cm-1

    with approximately 10 to 18 sharp absorption peaks. The eight HF

    peaks at 3,788.9, 3,878.4, 3,921.0, 4,001.7, 4,039.7, 4,076.0 and 4,110.7 cm-1

    were

    selected as references for HF. During each fire test, the HF concentration at a given time

    was determined by averaging the values corresponding to the eight peaks.

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    8/37

    4

    The calibration of the FTIR spectrometer for HCl was carried out using 500 and

    75,000 ppm commercial samples. More than 16 HCl absorption peaks were found in its

    characteristic frequency region of 2,600 to 3,100 cm-1

    . The peaks in the P-branch at

    2,776, 2,799, 2,822 and 2,844 cm-1

    were selected as the HCl references. In each fire test,

    the HCl concentration at a given time was determined by averaging the values

    corresponding to the four peaks.

    2.5.1.2 Sampling and Analyzing Procedures of FTIR Spectroscopy

    As shown in Figure 2, gas samples were drawn continuously from the

    compartment through three gas-sampling lines to the FTIR spectrometer. The gas sample

    from each sampling location was analyzed for 30 second using switching valve system.

    An infrared (FTIR) spectrum over a wide frequency range of 400 to 4,500 cm-1

    was

    collected from the sample gas. Measurement using the FTIR lasted more than 12 minutes

    for each test.

    Quantitative analysis of the agent and thermal decomposition products wereobtained by comparing infrared absorption peaks from the fire tests with the

    corresponding reference peaks obtained using the calibration gases.

    2.5.2 Gas Chromatography-Mass Spectrometry (GC/MS)

    Gas Chromatography-Mass Spectrometry (GC/MS) is the combination of the

    chromatographic technique with mass spectrometry: full use is made of the separating

    power of GC together with the structural information derivable from MS [18 - 20]. In the

    HAPE project, a Model 5995 Hewlett/Packard Gas Chromatography/Mass Spectrometer

    was used to measure the concentration of the agents used and other components present

    in the test room.

    2.5.2.1 Calibration of GC/MS

    Several commercial single component gases were used to calibrate the GC/MS.

    The concentrations of HCFC-22, HCFC-123 and HCFC-124 components used for the

    GC/MS calibration were the same as those used in the FTIR calibration (7.50%, 0.49%

    and 0.4%, respectively). The single component standards were sampled using an air-tight

    syringe from a decompressed free-flowing stream. Various volumes of calibration gases

    were then loaded on the sorption tubes.

    An Envirochem Model 810 A with two stage thermal desorber/concentrator was

    used to thermally desorb and concentrate the volatile component prior to analysis in the

    GC/MS. The volatile constituents were then separated in the capillary columns and

    analyzed by the MS detector. The corresponding ion spectrum for each component was

    compared with the NBS75 spectral library with the aid of a computer. Their peaks were

    identified and quantified as MS counts. Volume normalized MS area for each component

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    9/37

    5

    was then plotted against the tube load. A linear equation was fitted through the

    experimental points for each component:

    Ci = Ai + Bi X (1)

    where Ci is the concentration of ith component [ppm], A i and Bi are response factors ofith component respectively and X is the normalized MS response [cts/mL].

    2.5.2.2 Sampling and Analyzing Procedures of GC/MS

    The GC/MS was not conducted on-line. The measurement using the GC/MS

    involved: preparing sorption tubes, sampling, transporting and storing samples, and

    analyzing samples in the laboratory.

    In Phase 1 of the HAPE project, three types of sorption tube (Carbotrap and

    Carbotrap tubes and Standard Three Layer Glass Beads-Teenax-Ambersorb tubes)

    were used. Prior to the test, the sampling tubes were cleaned by purging with ultra purehelium for 10 to 15 minutes while they were heated to a temperature of 250

    0C.

    As shown in Figure 2, the sampling port for GC/MS was located after the FTIR

    cell and consisted of a SS-Tee that was equipped with a teflonized septum under a Swage-

    Lok Fitting. During the test, the gases from the test room were sampled manually with a

    gas-tight high precision syringe (Hamilton 1 mL). The contents of the syringe were then

    injected in a stream of laboratory air into a sorption tube. The tube was later stored in a

    screw capped glass sheath. At the end of the test, the exposed sorption tubes were

    transported to the GC/MS facility where they were thermally desorbed and analyzed.

    The components of the sample gas were identified with the aid of a HP supplied

    75,000 compound library with a probability matching algorithm and three other libraries

    [21]. Quantitative analysis of the agent concentrations was obtained by using equation (1)

    and comparing the peaks recorded from the fire tests with the corresponding reference

    peaks produced using the calibration gases.

    2.5.3 Selective Ion Electrode System

    Thermal decomposition products, including hydrogen fluoride (HF) and hydrogen

    chloride (HCl), were produced in the fire suppression process due to the interactions

    between the agent and the flame. In Phase 1 of the HAPE project, the thermal

    decomposition products were collected by Wet Chemistry and then analyzed using a

    Selective Ion Electrodes (SIE) method.

    2.5.3.1 Calibration of SIE System

    Orion solid state Selective Ion Electrodes were used to determine the

    concentrations of HF and HCl acid gases produced during the fire suppression process by

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    10/37

    6

    analyzing the levels of F-

    and Cl-

    anions in a solution prepared using the gas samples [22,

    23]. Each Selective Ion Electrode system was calibrated using calibration standards

    prepared from 0.1 N Acculute solutions of NaCl and NaF using the Incremental

    Technique method. A semi-logarithmic plot of each electrode potential versus halide ion

    concentration was constructed. A calibration curve was fitted to more than 30 calibration

    points to provide the appropriate response factor equation. For calibration, theconcentrations for F

    -and Cl

    -anions were varied from 0.0 to 8.0 and 0.0 to 40.0 mmol/L,

    respectively.

    2.5.3.2 Sampling and Analyzing Procedures of SIE System

    As shown in Figure 2, the sampling port for Wet Chemistry using a SS-Tee was

    located at the exit of the cold exhaust line. An all-glass threaded Supelco Midget bubbler

    equipped with a gas frit diffuser to increase the gas/liquid contact surfaces. 22 mL screw-

    cap vials were used as collection vessels. The collection vessels were prefilled with 10

    mL of 0.1 N NaOH solution prepared using an Anachemia Acculute Standard Volumetric

    Solution.

    The gases were sampled through the vials at a volumetric flow rate of 100-200

    mL/min for 90-120 seconds. The mass transfer between the gas stream and the solution

    occurred in irreversible chemisorption-neutralization reactions. The gaseous hydrogen

    fluoride (HF) and chloride (HCl) were then trapped from the sampling gases and dissolved

    in the hydroxide solution while most other components in the sample remained

    undissolved.

    In the laboratory, the pH of the solution in each vial was adjusted with a small

    volume of concentrated sulphuric acid. An aliquot of the exposed solution was thenadjusted using TISAB II solution. A Fluoride Selective Electrode was used to determine

    the concentration of F-

    anion. Another aliquot of the exposed solution was adjusted using

    an ISA buffer and a Chloride Selective Electrode was used to determine the concentration

    of Cl-

    anion. The quantitative analysis of the F-

    and Cl-

    anions concentrations were

    obtained by comparing the electrode potentials of the fire gases with the corresponding

    calibration curves produced using the standard solutions.

    2.6 Test Procedure

    Computers controlling the data acquisition system and the FTIR were started

    simultaneously at time zero. Test fires were allowed at least a 30 s pre-burn. During thepre-burn period, the door of the compartment was kept open. Agent discharge was

    activated manually at 60 s and lasted about 10 s. The pressure relief vent and the access

    door were deliberately held closed during the tests. The tests were terminated after 10

    minutes. The pressure relief vent was then opened and a fan connected through an

    exhaust duct was activated to ventilate the room.

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    11/37

    7

    3.0 TEST RESULTS

    The FTIR and GC/MS were used to measure the agent concentrations and the

    FTIR and SIE were used to measure the acid gases generated during the fire suppression

    process.

    Agent Concentrations

    As shown in Table 1, Test H-1 was a dump test, without any fire, to determine the

    agent discharge time and its concentration in the room. Test H-12 was a fire test without

    any extinguishing agent used. These two tests are not discussed in the present work. In

    Tests H-2 to H-6, the door was opened for a moment to relieve some of the pressure

    during the discharge. For all subsequent tests, both the louvers of the pressure relief vent

    opening and the door were kept closed during the discharge.

    The filters in the sample lines are designed to remove most of the particles from

    the sampling gases. The filters must be changed prior to every test. However, the filters

    in the sampling lines were not changed until Test H-7 in Phase 1 of the HAPE project. Asa result, the filters in the sampling lines were clogged during Tests H-2 to H-6. The

    clogged filters affected the measuring accuracy of FTIR, GC/MS and SIE. After Test H-

    7, the filters were changed prior to every test.

    The agent concentrations measured by GC/MS in Phase 1 of the HAPE project are

    listed in Appendix 1. The agent concentrations measured by FTIR are shown in Figure 3

    (for Tests H-2, H-3 and H-7 to H-13). The agent concentrations measured by GC/MS in

    the corresponding tests were plotted in Figure 4.

    The FTIR spectroscopy method continuously records the changes of agent

    concentrations in the compartment from the beginning of the discharge. The agentconcentration in the compartment quickly reached its design value and was distributed

    uniformly in the compartment shortly after the discharge. The agent concentrations

    measured by the FTIR from different sampling ports did not display any obvious

    fluctuation.

    For Test H-2, the target agent concentration in the room was 8.6%. The FTIR

    results showed that the maximum agent concentration actually reached 9% at the

    beginning of the test, then continuously dropped. For Test H-3, the agent concentrations

    as measured by FTIR were lower than the design value (8.6%), because the filter in the

    sampling lines was clogged during the test. After Test H-7, the agent leakage from the

    room was minimized during the discharge by closing both the door and the louvers of the

    pressure relief vent. The filters in the sample lines were also changed for every test. As a

    result, the agent concentration in the compartment did not drop after the discharge and the

    agent concentrations measured by the FTIR were very close to the target concentration.

    The target concentration in Test H-7 was 9.3% and the average agent concentration

    measured by the FTIR was 9.2%.

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    12/37

    8

    The number of points measured by GC/MS method for in full scale fire tests was

    limited. In Test H-8, only four GC/MS data points were obtained in a testing period of

    nearly 400 seconds, in comparison with 180 data points obtained using the FTIR method.

    Also with the switching system, the gases from some sampling lines were missed during

    the sampling process for GC/MS.

    The agent concentrations measured by GC/MS showed significant fluctuations

    from one testing point to another, as shown in Figure 4. The measured values obtained

    using the GC/MS method were lower than the target concentrations as well as the values

    measured by FTIR. In Test H-11, the average agent concentration measured by GC/MS

    for a testing period of nearly 400 seconds was 5.58% compared with the design

    concentration of 8.6%.

    Thermal Decomposition Products (TDP)

    Thermal decomposition products generated during the fire tests are strongly

    dependent on the size of the fire and the agent concentration used for suppression. The

    HF and HCl concentrations measured by SIE in Phase 1 of the HAPE project are listed inAppendix 2. HF and HCl concentrations measured by FTIR and SIE methods are shown

    in Figures 5, 6 and 7.

    The FTIR spectroscopy continuously recorded the formation of halogen acid gases

    in fire suppression from the beginning of the discharge. In Test H-2, eight tell-tale fires

    were employed but not extinguished. The FTIR results show that only a small amount of

    halogen acid gases were produced in the test with relatively small fires. With the increase

    in the fire sizes (e.g. 200 kW in Tests H-3 and H-8), very high concentrations of halogen

    acid gases were produced, when the fires were not extinguished. However, when the

    agent concentration was increased to 14.7% in Test H-10, Figure 7 shows that only a

    small amount of HF and HCl acid gases were produced as the fire was quicklyextinguished. When the fires were reignited at 680 s in Test H-10, after an agent

    concentration had decreased due to the operation of the exhaust fan, a significant amount

    of halogen acid gases were produced.

    The FTIR data in Figures 5 and 6 shows that HF and HCl concentrations measured

    in Test H-3 were lower than those measured in Test H-8 under the same fire scenarios.

    This indicates that the clogged filters in Test H-3 resulted in an underestimation of the

    production of halogen acid gases during the fire suppression process.

    In some unextinguished tests (e.g. H-8, H-9 and H-13), HF concentrations

    measured by FTIR displayed significant fluctuation from one sampling port to another,

    although the agent was distributed uniformly in the test room. This is because the

    formation of halogen acid gases varied depending on the local temperature. In Test H-8,

    sampling port 3 was close to the ceiling of the room where the local gas temperature was

    high. Sampling ports 1 and 2 were located at low elevations close to the floor, where the

    local gas temperature was relatively low. Hence, halogen acid gases measured at sample

    port 3 were higher than those measured at the other two ports.

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    13/37

    9

    In comparison with the measurements using the FTIR, the acid gases measured by

    the Wet Chemistry method were integrative samples from three sample points and

    represented a time and space-averaging of the concentrations in the test room. The results

    shown in Figure 6 for Tests H-8, H-9 and H-13 indicated that HF and HCl concentrations

    measured by the Selective Ion Electrode System were much lower than those measured bythe FTIR spectroscopy. This suggests that some of the acid gases were lost in the

    sampling systems.

    In Test H-10, HF and HCl concentrations measured by the Selective Ion Electrode

    System were much higher than the concentration measured by the FTIR and had almost

    the same levels as measured in Tests H-9 and H-13 in which the fires were not

    extinguished. This was not true for the case where the fire was quickly extinguished. In

    this test, the sampling system for the Wet Chemistry method were probably contaminated

    when the sample gases flowed through a long and cold stainless steel sampling line.

    4.0 UNCERTAINTY ANALYSIS

    Uncertainties generated from measuring agent concentrations and thermal

    decomposition products are strongly dependent on the characteristics of the analytical

    methods themselves, the set-up of the sampling system, calibration and sampling

    procedures and testing conditions.

    4.1 Uncertainties in FTIR Spectroscopy Measurement

    The FTIR was used to continuously measure the components of the gases sampled

    from the test room. The uncertainties related to the measurements with the FTIR are due

    to the calibration, sampling and quantification of experimental signals of the FTIR.

    Uncertainties in the Calibration of the FTIR

    The uncertainties associated with the FTIR calibration can be attributed mainly to

    the impurity and concentration range of the calibration gas, and calibrated data points

    required to form a proper calibrating curve.

    For the calibration of the agent, the concentration of the agent was about 8.5%

    with a single-point calibration. This could cause some uncertainties in some tests as the

    agent concentrations were higher than the calibrated concentration.

    For the calibration of HF, commercial samples with four HF acid concentrations

    were used to construct a proper calibration curve. For the calibration of HCl, however,

    commercial samples with only two HCl acid concentrations were employed to produce a

    calibration curve. During the full-scale fire tests, a large amount of acid gases were

    produced in the unextinguished fires. The calibration ranges for the halogen acid gases

    covered the concentrations of HF and HCl acid gases produced in the tests. However, the

    increase in the acid gas concentration is not linearly [2]. As a result, the limited numbers

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    14/37

    10

    of calibration points used for HCl could result in some uncertainties in the measurement of

    acid gases.

    Uncertainties in the Test of the FTIR

    The measurement using FTIR spectroscopy in the HAPE project was the external

    absorption mode. The gas samples were withdrawn through the metal sampling line andmeasured at the outside of the test room. The filters in the sample lines could result in

    some uncertainties in the measurements as the filters were clogged in some tests. Also,

    the particles retained by the filter, which were highly absorbent and reactive, could also

    prevent a fraction of the halogen acid gases reaching the FTIR for analysis. As shown in

    Figures 3 and 5, the concentrations of the agent and halogen acid gases measured by the

    FTIR in Test H-3 were significantly lower than those without the clogged filters in the

    later test.

    Also, the acid gases are very reactive and, even though stainless steel sampling

    lines were heated, the interaction between the reactive acid components and metal tubes

    occurred, resulting in losses of sampling acid gases and increased the uncertainties in themeasurements.

    Uncertainties in the Quantitative Analysis of the FTIR

    Since multiple chemical compounds were produced in the fire tests, there were

    overlaps of the FTIR signals in some analysis regions. The signal noise could lead to

    ambiguous determination of the exact height and area of some absorption peaks for the

    analysis of the agent and acid gases. Hence, this could result in an underestimated or

    overestimated concentrations of the agent and halogen acid gases.

    It was estimated that for the cold discharge test, the uncertainties of the FTIR in

    measuring the agent concentration was less than 7%. For the fire tests, the uncertaintieswere within 13%, because of the increasing uncertainties due to interference from the

    decomposition products. The uncertainties of the FTIR in the measurement of HF and

    HCl concentrations were estimated about 20% higher than those for the measurement of

    the agent concentration [7].

    4.2 Uncertainties in GC/MS Measurement

    The GC/MS analysis for the measurement of the agent concentrations was

    conducted off-line. The gases from the test room were sampled manually, put in storage,

    transported to the laboratory and then analyzed by GC/MS. Thus, the uncertainties

    produced in the measurement by GC/MS include not only those in the FTIR analysis, such

    as calibration and signal quantification uncertainty, but also those uncertainties related to

    the collection of the sample gas, and the storage and transport of the sample gas to the

    laboratory.

    Uncertainties in the Calibration of GC/MS

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    15/37

    11

    As the same as discussed in the calibration of FTIR, the uncertainties related to the

    calibration of GC/MS were due to impurities of the calibration chemicals, to the

    calibration concentration and to the number of calibrated data points. The purity of

    chemicals used for calibration was usually very high (99%). The number of data points for

    the GC/MS calibration was also increased by diluting the calibration standards. However,

    the concentration of the calibration standard was still lower than some values used in thetests. This could result in some uncertainties in determining the agent concentration, when

    the agent concentration exceeded the calibrated range.

    Uncertainties in the Sampling Procedure of GC/MS

    There were two uncertainties arising from the sampling process for GC/MS

    analysis. The first uncertainty resulted from the sampling period as the sampling gas was

    quickly and manually drawn with a gas-tight syringe at the sample port. The second

    uncertainty in the GC/MS sampling process was associated with the retaining capabilities

    of the sorption tube.

    The gas samples were affected by the clogged filters, the non-uniform pumpingcycles, the switch from one sample port to the other, and the fluctuation of the gas

    composition in the compartment during the short manual sampling period. The sampling

    volume of the gas-tight syringe also influences the sampling accuracy. In Test 6, the size

    of the gas-tight syringe was changed from 1 mL to 0.5 mL. As shown in Appendix 1, the

    agent concentrations measured by GC/MS in Test H-6 were significantly lower than those

    measured in the test in which samples were obtained using a 1 mL syringe.

    In Phase 1 of the HAPE project, three types of sorption tubes (Carbotrap ,

    Carbotrap and the Standard Three Layer Glass Beads-Teenax-Ambersorb tubes) were

    used. As shown in Figure 8, the concentrations of HCFC-123 and HCFC-124

    components in the agent measured by GC/MS did not fluctuate with the changes in

    sorption tubes used. But, the concentration of the HCFC-22 components changed from

    one sampling point to another. The test results also showed that the retaining capabilities

    of the Standard Glass Beads-Teenax-Ambersorb tubes was better than those of the other

    two types of sorption tube, but still resulted in an underestimation of the concentrations of

    HCFC Blend A in the tests.

    The retaining capabilities of sorption tubes were also affected by their age and the

    frequency of their utilization. Those sorption tubes with high utilization frequencies had

    lower retaining capabilities due to the deactivation of the sorption sites. In Phase 1 of the

    HAPE project, the age and utilization frequency of the sorption tubes were not identified.

    This lead to differences in the retaining capabilities of sorption tubes for samples, because

    different sorption tubes were used. The existence of halogen acid components in the

    sample gases can further reduce the retaining capabilities of the sorption tubes by

    damaging the beads of the sorption material.

    Uncertainties in Storage and Transport for GC/MS Analysis

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    16/37

    12

    The components of the agent HCFC Blend A are relatively stable. However,

    during storage and transport, the chemical interaction between the halogen acid

    components in the sample gases and the sorption material could damage the beads in the

    sorption tubes. This resulted in some uncertainties for the GC/MS analysis.

    Uncertainties in the Quantitative Analysis of GC/MSThe uncertainties produced in the quantitative analysis using the GC/MS method

    are relatively small in comparison with those observed for the FTIR method. During

    quantitative analysis of the sample gases, the interference from other chemical compounds

    could be reduced to a low level by the proper separation of Gas Chromatography.

    Hence, the uncertainties in the measurements obtained using GC/MS are mainly

    due to the manual sampling procedure and the retaining capabilities of the sorption tubes.

    The existence of the halogen acid components in the sampling gases further increases the

    uncertainties in the GC/MS measurement. As a result, the agent concentrations measured

    by GC/MS are significantly lower than the target concentrations. The uncertainties related

    to the measurements with GC/MS are higher than those related to the measurements usingFTIR spectroscopy.

    4.3 Uncertainties in SIE Measurement

    The Selective Ion Electrode (SIE) method for the measurement of halogen acid

    gases was also performed off-line. The sample gases were collected using Wet Chemistry

    and then analyzed by the Selective Ion Electrode System. The uncertainties associated

    with this analysis procedure include uncertainties in the collection and storage of the

    samples using the Wet Chemistry method, and uncertainties in the calibration and

    quantitative analysis of acid gases using the Selective Ion Electrode method.

    Uncertainties in the Collection and Storage of Sampling Gases Using Wet Chemistry

    For the HAPE project, heated stainless steel was used for the primary sampling

    lines. However, the sample line that connected the GC/MS sample port to the SIE sample

    port was not heated. Hence, the water vapour in the sampling gases could condense in the

    tube, and halogen acid components in the sampling gases could react with the metal tube

    in the presence of air and moisture during the test.

    The acid gases are very reactive with the stainless steel [24]. Sheinson et al. [8, 9]

    found that HF sampled using a short stainless steel tube showed significantly lower

    concentrations than when sampled using a Teflon tube. As shown in Figure 6, the results

    in the present tests also indicated that the concentrations of HF and HCl acid gases

    measured by SIE were significantly lower than those measured using the FTIR, indicating

    loses in the sampling procedure due to the interaction of the sample gases with the cold

    metal walls. Also, cold metal sample lines can contaminate samples if the same sampling

    system was repeatedly used in different tests. As shown in Figure 7, HF and HCl

    concentrations measured by Selective Ion Electrode System in Test H-10 were much

    higher than those values measured by FTIR, when the fire was quickly extinguished.

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    17/37

    13

    The effectiveness of the sampling procedure for the Wet Chemistry method was

    also affected by the collective efficiency of the bubbler vials and the flow rate of the

    sampling gases through the bubbler vials. The collective efficiency of bubbler vials was

    designed for a certain range of hydrogen halides concentrations. In the full-scale tests, a

    large amount of halogen acid gases was produced, when the fire was not extinguished bythe suppression agent. The high concentrations of hydrogen halides produced during the

    fire suppression process in some cases exceeded the upper range for the vial and lead to

    premature saturation of the ion trapping solutions in the bubbler vial. Hence, halogen acid

    components in the sampling gases were not all collected by Wet Chemistry in some tests.

    The high volumetric flow rate of the sampling gases could also reduce the stripping of acid

    components from the gas stream if the contact time for the gas bubbles within the solution

    was too short.

    Great care was also needed to store and transport the acid gases to prevent

    unwanted reactions and loss, because of the reactivity of the acid gases.

    Uncertainties in the Calibration and Quantitative Analysis Using SIE

    For the calibration of the Selective Ion Electrode System, calibration standards

    prepared from Acculute solutions were used. The calibration curves were plotted based

    on more than 30 calibration points with a wide range of F-

    and Cl-

    anion concentrations.

    The uncertainties related to the calibration of the SIE, therefore, are relatively small.

    Other halogen components, such as COF2 and COFCl, were also produced during

    the fire suppression process and hydrolyzed rapidly in the presence of water. Hence, in

    quantitative analysis of the acid concentrations, F-

    and Cl-

    anions measured by Selective

    Ion Electrode System include F

    -

    and Cl

    -

    anions from both halogen acid gases and othercomponents in the samples. As a result, the concentrations of the halogen acid

    components in the samples could be overestimated by the SIE method. The FTIR results

    showed that the amount of COF2 and other components generated during the fire

    suppression process was relatively small. These results indicated a 5 - 10% uncertainty in

    the HF and HCl measurement due to the presence of COF2 and other components in the

    sample gases.

    5.0 SUMMARY

    Fourier Transform Infrared (FTIR) Spectroscopy can provide an instantaneous and

    continuous method for the measurement of multiple chemical species in fire tests withreasonable accuracy. The uncertainties related to the measurements by FTIR are

    influenced by the calibration concentrations, the number of calibrated data points, the

    filters in the sampling lines and the signal interference in the quantitative analysis. The

    chemical interaction between the sample gases and metal sampling lines could further

    increase the FTIR uncertainties in the measurement of multiple chemical species.

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    18/37

    14

    Gas Chromatography-Mass Spectrometry (GC/MS) for the measurement of the

    agent concentration was conducted off-line. GC/MS cannot continuously record the

    changes in the agent concentrations in the test room. The number of measuring points is

    limited. The GC/MS method requires more elaborate calibration, and more time-

    consuming and labour-intensive sampling procedure and quantitative analysis. For the

    GC/MS, the uncertainties produced from the calibration procedure and quantitativeanalysis are relatively small, in comparison with those in the FTIR method. However, the

    sampling procedures used to obtain samples for GC/MS analysis has a significant effect on

    the measuring accuracy. Also, the retaining capabilities of the sorption tubes have a

    significant effect. The existence of the halogen acid components in the sample gases can

    further increase the uncertainties of the measurements by GC/MS. The overall

    uncertainties in the measurement of the agent concentration by GC/MS are higher than

    those with the FTIR spectroscopy. For the GC/MS method, the collection efficiency for

    the sample gases must be systematically verified prior to the test.

    The Specific Ion Electrode (SIE) method for the measurement of halogen acid

    gases was also conducted off-line in the tests. The concentrations obtained with the SIEmethod are time and space-averaged. The uncertainties for this method are mainly due to

    chemical interactions between the acid gases and metal sampling lines. In addition, the

    improper design of the bubbler vials and flow rates in the sampling process also increases

    the uncertainties of the SIE measurement. The concentrations of F-

    and Cl-

    anions

    measured by the Specific Ion Electrode method are the sum of both halogen acid gases

    and other components (e.g. COF2) in sampling gases.

    REFERENCES

    1. Grant, C.C., From the Halons of Yesterday to the Challenges of Tomorrow,

    Proceedings of Fire Safety -- Without Halon?, Zurich, Switzerland, September

    7-9, 1994, pp. 7-39

    2. Su, J.Z., Kim, A.K. and Mawhinney, J.R., Review of Total Flooding Gaseous

    Agents, J. of Fire Protection Engineering, vol.8, No.2, pp.45-64, 1996

    3. Tapscott, R.E., Summary of Current Agents, Presented at 1994 International

    CFC and Halon Alternatives Conference, Washington, DC, U.S.A., 1994

    4. Metchis, K., The Regulation of Halon and Halon Substitutes, Proceedings of

    Halon Options Technical Working Conference, Albuquerque, NM, USA., 1994

    pp. 7-30

    5. Kim, A.K., Su, J.Z., Mawhinney, J.R. and Kanabus-Kaminska, M., Full-Scale Fire

    Testing of HFC-227ea and HCFC Blend A, Proceedings of Halon OptionsTechnical Working Conference, Albuquerque, NM, USA., 1996

    6. Miller, T.L. and Green, A.E.S., A Survey of Techniques for Fire Suppressant

    Studies, Proceedings: Halon Alternatives Technical Working Conference 1994,

    PP.453-463.

    7. Mawhinney, J.R., Kim, A.K., Su., J.Z., Kanabus-Kaminska, M., Crampton, G.,

    and Lusztyk, E., Report on Full-Scale Fire Testing of HCFC Blend A - ND

    Research Project HAPE 7665-1 (CFFM3-2), Client Report of National Fire

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    19/37

    15

    Laboratory of the National Research Council Canada, 1995

    8. Sheinson, R.S., Musick, J.K., and Carhart, H.W., HF and Hbr Production from

    Full-Scale CF3Br (Halon 1301) Fire Suppression Tests, J. of Fire and

    Flammability, Vol.12, P.229, 1981

    9. Sheinson, R.S. and Alexander, J.I., HF and Hbr from Halon 1301 Extinguished

    Pan Fires, Fall Meeting, Eastern States Section Meeting/the Combustion Institute,Pittsburgh PA, 1982

    10. Di Nenno, P.J., Forssell, E.W., Peatross, M.J., Wong, J.T., and Maynard, M.,

    Evaluation of Alternative Agents for Halon 1301 in Total Flooding Fire

    Suppression Systems: Thermal Decomposition Product Testing, Proceeding:

    Halon Alternatives Technical Working Conference, Albuquerque, P. 162, 1993

    11. Linteris, G.T., King, M.D., Liu, A., Womeldorf, C. and Hsin, Y.E., Acid Gas

    Production in Inhibited Diffusion Flames, Proceeding: Halon Alternatives

    Technical Working Conference, Albuquerque, P. 177, 1994

    12. Dierdorf, D.S., Moore, T.A. and Skaggs, S.R., Decomposition Product Analysis

    During Intermediate-Scale (645 FT3) and Laboratory Scale (6.18 FT3) Testing of

    NFPA 2001 Agents, Proceeding: Halon Alternatives Technical WorkingConference, Albuquerque, P. 177, 1993

    13. Bulien, O. K., FTIR Spectrometer for Measuring Toxic Smoke Components in

    Fire Testing - Review of Equipment and Calibration Routines in NT FIRE 047,

    Fire and Materials, Vol. 20, PP.225-333, 1996

    14. Smyth, M.R., Chemical Analysis in Complex Matrices, Ellis Horwood and

    Prentice Hall, England, 1992

    15. Zyka, J., Instrumentation in Analytical Chemistry, Ellis Horwood and Prentice

    Hall, England, 1994

    16. Ferraro, J.R. and Baslle, L.J., Fourier Transform Infrared Spectroscopy -

    Techniques Using Fourier Transform Infrared, Vol. 3, Academic Press, New

    York, 198417. Ferraro, J.R. and Baslle, L.J., Fourier Transform Infrared Spectroscopy -

    Applications to Chemical Systems, Vol. 1, Academic Press, New York,

    1984

    18. Pecsok, R.L. and Shields, L.D., Modern Methods of Chemical Analysis John

    Wiley & Sons, Inc., New York, 1968

    19. Middleditch, B. S., Practical Mass Spectrometry, Plenum Press, New York,

    1979

    20. Ravindranath, B., Principles and Practice of Chromatography, Fllis Horwood

    Limited, New York, 1989

    21. Bruno, T.J., Spectroscopic Library for Alternative Refrigeram Analysis, NIST

    Special Publication:794, 1991

    22. Freiser, H., Ion-Selective Electrodes in Analytical Chemistry, Plenum Press,

    New York, 1978

    23. Kenkel J., Analytical Chemistry for Technicians, Lewis Publishers, Inc.,

    Michigan, 1989

    24. Lodge, J.P., Methods of Air Sampling and Analysis, 3rd edition, Lewis Publi.

    Inc., Michigan, 1991

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    20/37

    16

    25. LaQue, F.L. and Copson, H.R., Corrosion Resistance of Metals and Alloys, 2nd

    Ed., American Chemical Society Monograph No. 158, New York, 1963

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    21/37

    17

    Appendix 1: HCFC Blend A Concentrations Measured by GC/MS

    HAPE 2-13March-April 1995

    SAMPLE TIME of F22 F124 F123 AGENTSampling % % % as sum %min

    HAPE 2, 17-Feb-95 8.6%NAFS IIIT2-1 0.5 0.00 0.00 0.00 0.00T2-2 2 2.48 0.18 0.23 2.88T2-3 3.5 2.36 0.18 0.24 2.79T2-5 6.5 2.04 0.20 0.29 2.52T2-6 8 3.10 0.21 0.31 3.62T2-7 9.5 2.82 0.21 0.30 3.33T2-8 11 2.34 0.19 0.28 2.82T2-10 15.5 2.70 0.18 0.26 3.13

    T2-11 18.5 2.04 0.15 0.23 2.42T2-12 21.5 1.33 0.11 0.18 1.63

    HAPE 3, 22-Feb-95 8.6%NAFS IIIT3-1 0.5 0.00 0.00 0.00 0.00T3-2 1 2.44 0.16 0.24 2.85T3-3 2 2.91 0.18 0.24 3.33T3-5 5 2.63 0.16 0.23 3.02T3-6 6.5 2.09 0.15 0.22 2.45T3-8 12.5 2.19 0.14 0.21 2.54T3-9 14 1.02 0.11 0.16 1.29T3-10 17 0.58 0.08 0.12 0.79T3-11 23 0.00 0.06 0.08 0.13

    HAPE 4, 24-Feb-95 9.3%NAFS IIIT4-1 0 0.00 0.05 0.00 0.05T4-2 0.5 0.26 0.05 0.00 0.31T4-3 1 2.91 0.15 0.21 3.27T4-4 2 2.22 0.12 0.18 2.52T4-5 3.5 2.80 0.14 0.19 3.13T4-6 5 2.90 0.14 0.18 3.23T4-8 8 2.77 0.13 0.20 3.09T4-9 9.5 2.14 0.11 0.18 2.43T4-10 11 2.21 0.11 0.16 2.49T4-11 14 1.20 0.08 0.13 1.41T4-12 20 0.63 0.05 0.00 0.69

    HAPE 5, 28-Feb-95 9.3%NAFS IIIT5-1 0 0.00 0.00 0.00 0.00T5-2 0.5 0.06 0.00 0.02 0.08T5-3 1 3.79 0.36 0.24 4.39T5-4 2 2.18 0.19 0.17 2.54T5-5 3.5 3.84 0.27 0.19 4.30T5-6 5 4.79 0.32 0.20 5.31T5-7 8 3.06 0.30 0.19 3.56T5-8 11 4.03 0.29 0.20 4.52

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    22/37

    18

    T5-9 12.5 2.28 0.17 0.12 2.57T5-11 20.5 0.43 0.01 0.00 0.44

    HAPE 6, 02-Mar-95 9.3%NAFS IIIT6-1 0 -0.00 0.00 0.00 -0.00T6-2 0.5 -0.00 0.00 0.00 -0.00T6-3 1 0.49 0.06 0.06 0.61T6-4 2 1.04 0.05 0.04 1.13T6-5 3.5 1.77 0.09 0.08 1.94T6-6 5 1.69 0.08 0.06 1.83T6-7 6.5 1.34 0.10 0.08 1.53T6-8 9.5 1.38 0.08 0.06 1.52T6-9 12.5 0.79 0.07 0.05 0.91T6-10 15.5 0.59 0.03 0.02 0.65T6-11 20.5 0.34 0.02 0.00 0.36

    HAPE 7, 07-Mar-95 9.3%NAFS IIIT7-1 0 -0.00 0.00 0.00 -0.00T7-2 0.5 -0.00 0.00 0.03 0.03T7-3 1 3.40 0.24 0.18 3.82

    T7-4 2 4.72 0.47 0.41 5.59T7-5 3.5 3.68 0.40 0.34 4.42T7-6 5 4.52 0.43 0.31 5.26T7-7 6.5 4.29 0.43 0.32 5.04T7-8 9.5 7.52 0.56 0.38 8.46T7-9 12.5 3.92 0.34 0.24 4.50T7-10 18.5 1.79 0.08 0.08 1.96T7-11 24 0.40 0.02 0.03 0.45

    HAPE 8, 09-Mar-95 9.3%NAFS IIIT8-1 0 n/dT8-2 0.5 n/dT8-3 1 n/d

    T8-4 2 3.75 0.34 0.30 4.39T8-5 3.5 3.48 0.35 0.29 4.11T8-6 5 2.43 0.36 0.28 3.06T8-7 6.5 2.80 0.38 0.28 3.46T8-8 9.5 5.23 0.45 0.30 5.98T8-9 12.5 2.39 0.20 0.15 2.74T8-10 17 1.84 0.11 0.08 2.03

    HAPE 9, 16-Mar-95 8.6%NAFS IIIT9-1 0 -0.00 0.00 0.01 0.01T9-2 0.5 -0.00 0.00 0.01 0.01T9-3 1 0.28 0.02 0.03 0.32T9-4 2 5.37 0.45 0.33 6.15

    T9-5 3.5 4.60 0.44 0.32 5.36T9-6 5 3.07 0.43 0.34 3.84T9-7 8 5.03 0.49 0.37 5.88T9-8 11 5.76 0.49 0.42 6.68T9-9 12.5 8.46 0.50 0.37 9.34T9-10 15.5 1.53 0.12 0.10 1.76

    HAPE 10, 21-Mar-95 2x8.6%NAFS IIIT10-1 0 -0.00 0.00 0.00 -0.00T10-2 0.5 -0.00 0.00 0.00 -0.00

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    23/37

    19

    T10-3 1 -0.00 0.18 0.11 0.29T10-4 2 4.48 0.55 0.52 5.54T10-5 3.5 5.90 0.61 0.59 7.10T10-6 5 12.66 1.09 0.69 14.44T10-7 6.5 8.45 0.85 0.65 9.95T10-8 9.5 12.79 0.97 0.58 14.34T10-9 11 4.78 0.45 0.31 5.54T10-10 12.5 3.54 0.36 0.28 4.18T10-11 17 2.27 0.19 0.15 2.61T10-12 27.5 0.14 0.00 0.00 0.14

    HAPE 11, 22-Mar-95 8.6%NAFS IIIT11-1 0T11-2 0.5 -0.00 0.00 0.00 -0.00T11-3 1 5.40 0.39 0.29 6.09T11-4 2.5 4.78 0.38 0.29 5.45T11-6 5.5 4.74 0.38 0.29 5.41T11-7 7 4.67 0.38 0.33 5.38T11-8 8.5 6.01 0.44 0.35 6.80T11-9 11.5 5.06 0.43 0.36 5.85

    T11-10 13 4.64 0.28 0.25 5.17T11-11 17.5 1.22 0.05 0.08 1.35

    HAPE 13, 24-Mar-95 10%NAFS IIIT13-1 0 -0.00 0.00 0.00 -0.00T13-2 0.5 -0.00 0.00 0.00 -0.00T13-3 1 4.38 0.27 0.22 4.86T13-4 2.5 5.32 0.46 0.34 6.13T13-5 4 4.11 0.34 0.27 4.73T13-6 5.3 4.58 0.37 0.29 5.24T13-7 8.5 6.65 0.40 0.33 7.39

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    24/37

    20

    Appendix 2: F- and Cl- Concentration Measured by Ion Selective Electrode Method

    HAPE 3 - 13

    March - April

    1995

    Sample middle time Cl- conc in gas F- conc in gass ppm ppm

    HAPE 3, 22-Feb-95 8.6%NAFS III0 0 0

    3-01 30 28 263-02 95.5 2598 59783-03 172 1852 84253-04 242 3887 124293-07 525 16765 132503-08 705 14501 10955

    HAPE 4, 24-Feb-95 9.3%NAFS III0 0 0

    4-01 120 2633 33974-02 285 19383 118214-03 375 23519 113794-04 465 21351 122794-05 573 20343 127554-07 1140 4780 5834

    HAPE 5, 28-Feb-95 9.3%NAFS III0 0 0

    5-01 60 3942 74095-02 150 5619 127275-04 270 11058 132205-05 330 12181 137335-06 390 14781 137335-09 705 14508 106605-10 855 8310 43985-11 1095 4882 4076

    HAPE 6, 02-Mar-95 9.3%NAFS III6-01 0 14 1166-02 60 1817 56166-03 180 4555 95666-04 300 9407 92096-05 420 11980 8535

    6-06 540 14537 85356-07 660 11980 79096-08 900 8540 48246-09 1140 3754 2527

    HAPE 7, 07-Mar-95 9.3%NAFS III

    7-01 0 15 477-02 60 1394 3475

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    25/37

    21

    7-03 180 1955 66357-04 300 5398 80257-05 420 9189 71607-06 540 11149 66357-07 840 9644 52817-08 1140 3022 2202

    HAPE 8, 09-Mar-95 9.3%NAFS III

    0 0 08-01 55 1058 40768-02 235 1615 48578-03 303 5516 91558-04 415 4513 72118-05 535 6032 74918-06 655 4300 66838-07 900 1732 28638-08 1140 387 1065

    HAPE 9, 16-Mar-95 8.6%NAFS III

    0 0 09-01 55 1307 42959-02 175 1511 85179-03 295 1665 81999-04 415 2225 85179-05 535 2120 75999-06 660 2141 69669-07 900 897 25919-08 1140 255 1258

    HAPE 10, 21-Mar-95 2x8.6%NAFS III

    0 0 010-01 55 1983 435110-02 175 2184 469510-03 297 2436 547910-04 474 2572 536010-05 535 2406 487710-06 660 627 262510-07 895 465 227910-08 1014 337 152810-09 1140 170

    HAPE 11, 22-Mar-95 8.6%NAFS III

    0 0 0

    11-01 55 1457 348311-02 175 1684 390411-03 295 1388 571011-04 415 1855 835411-05 535 1388 717511-06 670 1769 631911-07 797.5 944 4189

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    26/37

    22

    11-08 895 610 287911-09 1020 345 1737

    HAPE 13, 24-Mar-95 10%NAFS III

    0 0 0

    13-01 65 1218 324213-02 182.5 2229 847013-04 415 2834 884813-05 535 3438 884813-06 655 2974 704213-07 782.5 1614 377913-08 912 854 231113-09 1095 422 1538

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    27/37

    23

    Table 1 Sequence of Tests in Phase 1 of HAPE Project

    Test Date

    1995

    Target Conc.

    HCFC Blend A

    (%)

    Pipe System Nozzle/

    Orifice

    Diam.

    Fire Comments

    Details

    H-1 Jan-27 8.6 Navy standard

    65 mm manifold

    Type A

    8.8 mm

    None

    H-2 Feb-17 8.6 Navy standard

    65 mm manifold

    Type A

    11.5 mm

    8 TTs

    (see Note)

    H-3 Feb-22 8.6 Navy standard

    65 mm manifold

    Type A

    11.5 mm

    8 TTs

    3 SPs

    200 kW

    SPs 50 mm

    from each

    wall

    H-4 Feb-24 8.6 Navy standard

    65 mm manifold

    Type A

    11.5 mm

    8 TTs

    1 RPs

    450 kW

    RP between

    nozzles

    H-5 Feb-28 9.3 Navy standard

    65 mm manifold

    Type A

    11.5 mm

    8 TTs

    3 SPs

    200 kW

    SPs 50 mm

    from each

    wall

    H-6 Mar-2 9.3 Navy standard65 mm manifold

    Type B6 @ 8 mm

    8 TTs3 SPs

    200 kW

    Gas samplinglines adjusted

    vertically

    H-7 Mar-7 9.3 Navy standard

    65 mm manifold

    Type B

    6 @ 8 mm

    8 TTs

    3 SPs

    200 kW

    SPs 50 mm

    from each

    wall

    H-8 Mar-9 9.3 Modified

    Removed 65 mm

    manifold

    Type B

    6 @ 8 mm

    8 TTs

    3 SPs

    200 kW

    SPs 50 mm

    from each

    wall

    H-9 Mar-16 9.3 Modified

    Removed 65 mm

    manifold

    Type B

    6 @ 8 mm

    8 TTs

    2 SPs

    1 RP in centre

    550 kW

    1 SP moved

    out to 300 mm

    from each

    wall

    H-10 Mar-21 2 @ 8.6

    Target 15 with

    controlled

    venting

    Replaced 65 mm

    manifold

    two cylinders

    discharge

    Type B

    6 @ 8 mm

    8 TTs

    3 SPs

    200 kW

    SPs 50 mm

    from each

    wall

    H-11 Mar-22 8.6

    Post-discharge

    ignition

    Navy standard

    65 mm manifold

    Type B

    6 @ 8 mm

    3 SPs

    150 kW

    Pans in

    corners

    SPs 50 mm

    from each

    wall

    H-12 Mar-23 No discharge N/A N/A 3 SPs

    150 kW

    Pans in

    corners

    Non-

    suppression

    test

    H-13 Mar-24 10 Navy standard65 mm manifold

    Type B6 @ 8 mm

    8 TTs3 SPs

    200 kW

    SPs 50 mmfrom each

    wall

    Note: Total heat output from eight TTs is estimated to be 50 kW; three SPs, 150 kW; one RP, 400 kW

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    28/37

    24

    Time (s)

    60 90 120 150 180

    VolumetricConcentration(%)

    0

    2

    4

    6

    8

    10

    12

    H-2

    H-3

    H-7

    H-8

    Figure 3a. HCFC Blend A concentration-time profiles measured by FTIR

    Spectroscopy for Tests H-2, H-3, H-7 and H-8.

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    29/37

    25

    Time (s)

    60 90 120 150 180

    VolumetricConcentration(%)

    0

    2

    4

    6

    8

    10

    12

    14

    16

    H-9

    H-10

    H-11

    H-13

    Figure 3b. HCFC Blend A concentration-time profiles measured by FTIR

    Spectroscopy for Tests H-9, H-10, H-11 and H-13.

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    30/37

    26

    Time (S)

    0 200 400 600 800 1000 1200 1400

    VolumetricConcentration(%)

    0

    2

    4

    6

    8

    10

    12

    H-2

    H-3

    H-7

    H-8

    Figure 4a. HCFC Blend A concentration-time profiles measured by GC/MS for

    Tests H-2, H-3, H-7 and H-8.

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    31/37

    27

    Time (S)

    0 200 400 600 800 1000 1200 1400

    VolumetricConcentration(%)

    0

    2

    4

    6

    8

    10

    12

    14

    16

    H-9

    H-10

    H-11H-13

    Figure 4b. HCFC Blend A concentration-time profiles measured by GC/MS for

    Tests H-9, H-10, H-11 and H-13.

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    32/37

    28

    Time (s)

    0 60 120 180 240 300 360 420 480 540 600 660 720 780 840 900 960

    VolumetricConcentration(ppm)

    0

    2000

    4000

    6000

    8000

    10000

    12000

    14000

    16000

    18000

    H-2 (FTIR)

    H-3 (FTIR)

    H-3 (SIE)

    Figure 5a. HF concentration-time profiles measured by FTIR and SIE for Tests

    H-2 and H-3.

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    33/37

    29

    Time (s)

    0 60 120 180 240 300 360 420 480 540 600 660 720 780 840 900 960

    VolumetricConcentration(ppm)

    0

    2000

    4000

    6000

    8000

    10000

    12000

    14000

    16000

    18000

    H-2 (FTIR)

    H-3 (FTIR)

    H-3 (SIE)

    Figure 5b. HCI concentration-time profiles measured by FTIR and SIE for Tests

    H-2 and H-3.

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    34/37

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    35/37

    31

    Time (s)

    60 120 180 240 300 360 420 480 540 600 660 720

    VolumetricConcentration(ppm)

    0

    5000

    10000

    15000

    20000

    25000

    H-8 (FTIR)

    H-9 (FTIR)

    H-13 (FTIR)

    H-8 (SIE)

    H-9 (SIE)H-13 (SIE)

    Figure 6b. HCI concentration-time profiles measured by FTIR and SIE for Tests

    H-8, H-9 and H-13.

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    36/37

    32

    Time (s)

    60 120 180 240 300 360 420 480 540 600 660 720

    H

    FandHCIConcentration(ppm)

    0

    1000

    2000

    3000

    4000

    5000

    6000

    HF (FTIR)

    HCl (FTIR)

    HF (SIE)

    HCI (SIE)

    Figure 7. HF and HCI concentration-time profiles measured by FTIR and SIE

    for Test H-10.

  • 7/29/2019 Applications and Uncertainties Associated With Measurements Using FTIR Spectrometry

    37/37

    Time (s)

    0 200 400 600 800 1000 1200

    VolumetricConcentration(%)

    0

    1

    2

    3

    4

    5

    6

    HCFC-22

    HCFC-124

    HCFC-123

    Figure 8. The components of HCFC Blend A-time profiles measured by GC/MS

    for Tests H-8.