Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello 1 , J. A. Yagüe-Fabra 2 , S. Carmignato 3 , H.N. Hansen 1 1 DTU Mechanical Engineering, Technical University of Denmark, DK2800, Kgs. Lyngby, Denmark 2 I3A, University of Zaragoza, E50018 Zaragoza, Spain 3 DTG, University of Padova, I36100, Vicenza, Italy José Yagüe-Fabra Simone Carmignato Guido Tosello Hans Nørgaard Hansen

Transcript of Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… ·...

Page 1: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

Application for Computed Tomography in Metrology for Micro Manufacturing

G. Tosello1, J. A. Yagüe-Fabra2, S. Carmignato3, H.N. Hansen1

1DTU Mechanical Engineering, Technical University of Denmark, DK‐2800, Kgs. Lyngby, Denmark2I3A, University of Zaragoza, E‐50018 Zaragoza, Spain3DTG, University of Padova, I‐36100, Vicenza, Italy

José Yagüe-Fabra Simone CarmignatoGuido Tosello

Hans Nørgaard Hansen

Page 2: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/20132 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Micro Injection Moulding (µIM)

Page 3: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/20133 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Objectives

OCMM

CT

CMM

• Dimensional verification of 2 micro‐injection moulded components (actual industrial productions) using CT metrology

• Comparison Computer Tomography vs. CMM vs. OCMM.

6

UniZarUniPD

DTU DTU

Page 4: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/20134 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Outline

1. Introduction

2. Materials and methods

3. CT results - imaging

4. CT results - measurements

5. Conclusion

6. Outlook / Work in progress

Page 5: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/20135 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Introduction

• Accuracy and time demands tighter and tighter  smaller mechanical parts are characterized by smaller tolerances to be verified

• Computed Tomography (CT) metrology techniques are more and more applied for micro‐parts geometrical verification:

• Advantages: non‐contact, dense scanning, capability of measuring both internal and external geometries simultaneously, NDT

• However: challenge to obtain high accuracy measurement results, i.e. with U/T<10%‐20%

Page 6: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/20136 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Materials and methodsMicro injection moulded parts

ToggleHearing aid applicationLiquid crystal polymer (LCP)Part mass: 35 mg

Micro Dog BoneMicro mechanical material tensile testingPolyoxymethylene (POM)Part part: 35 mg

Page 7: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/20137 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Dimensions

• Both internal and external geometries

• part thickness

• internal diameter 

• external diameter

• part length

• 3 different measuring techniques: CT, TCMM, and OCMM

Length Thickness

Page 8: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/20138 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

0.2 mm

0. 9mm 

0. 9mm. d

h1 h2

A B

C

Measuring procedure – Toggle

• Different measuring systems

• Common measurand definition

• Comparison of different measuring results

D=5.400 ± 0.030 mm

d=1.550 ± 0.020 mm

H=0.380 ± 0.030 mm

0.2 mm

Page 9: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/20139 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Measuring procedure – Dog bone

a

c

b

d

a

b

c

d

y

x

Y=9.0

Y=7.5

Y=4.5

Y=3.0

ds2

ds1

Y

A B C

D E F

Y

X

X

Y=1.0 Y=2.0

Y=4.0 Y=6.0 Y=8.0

Y=10.0 Y=11.0

X=0.5

X=2.5

X=8.5

X=10.5

X=1.0

X=2

X=9.0

X=10.0

X=0.5

X=2.5

X=8.5

X=10.5

l l

X=0.5 X=2.5 X=8.5 X=10.5

y

x

x

Ay

A B

C D

A to F = 1.000 mm

± 0.020 mm

L = 11.800 mm

± 0.030 mm

a,c = 3.000 mm

± 0.030 mmb = 1.500 mm

± 0.020 mmd = 1.350 mm

± 0.020 mm

Page 10: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201310 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

• Micro-CT Scanner: General Electric

• Model: eXplore Locus SP

• X Ray source power: 50-80 KV

• Detector 2D: 2300x3500

• Maximum resolution: 8 µm

• Maximum dimensions :

Diameter: 44 mm

Height: 56 mm

Measuring machines: CT1

UniZar

• S. Ontiveros, J.A. Yagüe-Fabra, R. Jiménez, G. Tosello, S. Gasparin, A. Pierobon, S. Carmignato, H.N. Hansen (2012) Dimensional measurement of micro-moulded parts by computed tomography, Measurement Science and Technology, 23 125401 (9pp) doi:10.1088/0957-0233/23/12/125401.

Page 11: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201311 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Measuring machines: CT2

• Micro-CT Scanner: Tomolab(developed by the ELETTRA Laboratory in Trieste)

• cone-beam microCT• X Ray source power: 40-130 KV• Spot size: 5 µm• Maximum dimensions:

Diameter: 45 mm

UniPD

• S. Ontiveros, J.A. Yagüe-Fabra, R. Jiménez, G. Tosello, S. Gasparin, A. Pierobon, S. Carmignato, H.N. Hansen (2012) Dimensional measurement of micro-moulded parts by computed tomography, Measurement Science and Technology, 23 125401 (9pp) doi:10.1088/0957-0233/23/12/125401.

Page 12: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201312 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

• Optical CMM: DeMeet 220 (2½ D)• Measuring volume 220 mm x 150

mm x 100 mm• MPE

X-Y= 4 + L/150 µm, L in mm

• MPEZ

= 3.5 µm• Fast measurements and in-line

quality • Validation instrument

Measuring machines: OCMM and TCMM

• Tactile CMM: measuring volume 850 mm x 1150 mm x 600 mm• MPE = 0.4 + L/900 µm, L in mm• Toggle parts measured OCMM compensation

• G. Tosello, H.N. Hansen, S. Gasparin ”Applications of dimensional micro metrology to the product and process quality control in manufacturing of precision polymer micro components” CIRP Annals - Manufacturing Technology 58 (2009) 467–472.

DTU

Page 13: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201313 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Outline

1. Introduction

2. Materials and methods

3. CT results - imaging

4. CT results - measurements

5. Conclusion

6. Outlook / Work in progress

Page 14: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201314 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

6

Work Piece Scanning

ReconstructionEvaluation

Slices

Correction factors

Surface Extraction

3. CT Metrology: Process

Page 15: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201315 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Part and support materials cannot be distinguished

Support

PartArtifacts

CT Image quality

Waves

Streaking artifact

Page 16: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201316 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Streaking artifact

Shadingartifacts

Ring Artifact

CT Image quality

Waves

Extra material and bubbles

Extra material

Page 17: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201317 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Work partdefects

CT Evaluation

Workpartdefects

Software

Correction Factors

Post-Process

6

Threshold determination

Strategy

Page 18: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201318 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Outline

1. Introduction

2. Materials and methods

3. CT results - imaging

4. CT results - measurements

5. Conclusion

6. Outlook / Work in progress

Page 19: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201319 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Correction Process 1Features with external and internal dimensions

1. Determine the ratio between Inner Diameter (ID) and Outer Diameter (OD) of the reference measurement. This ratio does not depend on the scale factor.

2. Adjust the threshold in the CT measurements and calculate the ratio, repeat this process until value calculated in step 1 is obtained.

3. Once obtained the ratio in the step 2, calculate the scale factor.

4. Apply the scale factor to the CT measurements.

Page 20: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201320 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Results – Toggle

Maximum Error: -93 µm

Minimum Error: -66 µm

Average Error(abs): 75 µm

Maximum Error: -25 µm

Minimum Error: -21 µm

Average Error(abs): 17 µm

Before Correction After Correction

Page 21: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201321 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Results – Toggles

Maximum Error: -114 µm

Minimum Error: -59 µm

Average Error(abs): 77 µm

Maximum Error: -58 µm

Minimum Error: 0.4 µm

Average Error(abs): 19 µm

Before Correction After Correction

Page 22: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201322 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Correction Process 2Features with external dimensions

1. Make the surface extraction at ISO 50% and measure the workpiece2. Calculate the deviations of the CT measurement results from the referencecalibrated values

3. Find the slope of the line that passes through the error points

4. Apply the equation of the slope to the measurements made at step 1

Page 23: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201323 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Results – Dog Bones

[µm] Left Side Right Side Thickness

MaximumError ‐43 ‐37 19,3

MinimumError 4,8 11,3 2,8

Average Error(abs) 12,2 11,6 10,6

[µm] Left Side Right Side Thickness

MaximumError 7,4 ‐13.3 ‐13.7

MinimumError ‐6,9 1,6 2,8

Average Error(abs) 5,1 3,5 5,9

Before Correction After Correction

Page 24: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201324 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Measurements results comparisonCT vs. OCMM

• H=0.380 ± 0.030 mm

Page 25: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201325 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Measurements results comparisonCT vs. OCMM

• d=1.550 ± 0.020 mm

Page 26: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201326 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Measurements results comparisonCT vs. OCMM

• D=5.400 ± 0.030 mm

Page 27: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201327 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Outline

1. Introduction

2. Materials and methods

3. CT results - imaging

4. CT results - measurements

5. Conclusion

6. Outlook / Work in progress

Page 28: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201328 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Conclusion

• CT measuring techniques are feasible for a complete quality control of 3D micro moulded parts

• Capability to provide morphological information such as sink marks on the surface and voids inside the mouldings

• Ability to collect comprehensive point clouds from internal and external geometries and simultaneously gathering information on material properties

• Correction strategies effective to improve measurement accuracy

Page 29: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201329 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

29

Department of Mechanical EngineeringSection of Manufacturing EngineeringMicro/Nano and Precision Manufacturing (MPP)

• Department of Mechanical Engineering– 6 sections, administration and workshops– Scientific personnel = 107 (2012)– Technical & Administrative = 85 (2012)– PhD students graduates = 23 (2011)

• Section of Manufacturing Engineering– Micro/Nano and Precision Manufacturing (MPP)– Scientific personnel = 19 + Technical personnel = 23 (2013)– PhD students = 18 (2013)– MSc students graduates = 15‐20 / year– BSc students graduates = 5‐10 / year– MSc Programme on Materials and Manufacturing Engineering– Micro Mechanical Systems Design and Manufacture (MSc course)– PhD Summer School on Multi‐Material Micro Manufacture(since 2006)

Page 30: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201330 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

• Group Leader → Prof. Hans N. Hansen• Research Group →Micro/Nano and Precision Manufacturing (MPP) (established on 2002)

• Research projects focus (European, national, industrial projects)– Mechanical, thermal and chemical manufacturing PROCESSES and their combinations

– MATERIALS →metal, polymers and ceramics

– Development, analysis and SIMULATION of processes, process machines, tooling systems

– MASS PRODUCTION processes → µ‐injection moulding and µ‐forming

– GEOMETRIC METROLOGY as the basis for decisions on control of modern constructions, manufacturing and function

Si

Ni

PP

Section of Manufacturing EngineeringMicro/Nano and Precision Manufacturing (MPP)

Page 31: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201331 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Si

Ni

PP

DesignMaterial

developm.Process

developm.

Tooling technologies

Joining

Manufacturing systems

Characteri-sation

• Metals

• Polymers

• Ceramics

• Metal forming

• Inj. moulding

• µMachining

• µAM

• Laser welding

• Soldering

• Resistance weld.

• Mech. assembly

• Electroforming

• µEDM

• µMilling

• µECM

• Dimesion and geometry

• Materials charact.

• Product driven

• Tooling

• Tolerancing

Section of Manufacturing EngineeringMicro/Nano and Precision Manufacturing (MPP)

Page 32: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201332 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Injection Moulding (IM) Micro Injection Moulding (µIM)Injection Compression Moulding (ICM)

• Process development

• Process simulation

• Design & Tooling

Page 33: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201333 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Section of Manufacturing EngineeringLaboratory of Geometrical Metrology

Page 34: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201334 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Precision & Micro MachiningMicro milling / Micro electrical discharge machining

Page 35: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201335 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Research Projects (National / European)Micro/Nano/MultimaterialManufacturing

• POLYMETAL DK (2005‐2008) Metallization of polymers for micro manufacturing• MASMICRO EU FP6 (2004‐2008) Micro‐assembly techniques for micro products• 4M EU FP6 (2004‐2008) Multi‐Material Micro Manufacture• NANOCMM EU FP6 (2006‐2012) Flexible Coordinate Metrology for Micro and Nano Components Production

• COTECH EU FP7 (2008‐2012) Converging Technologies for Micro Manufacture• PolyNano DK (2011‐2015) for Micro/Nano Fluidic Manufacturing Platform• Hi‐Micro EU FP7 (2012‐2015) High Precision Technology for Micro Manufacture• HINMICO EU FP7 (2013‐2016) High Throughput Integrated Technologies for MultimaterialFunctional Micro Components

Page 36: Application for Computed Tomography in Metrology for Micro ... for Computed Tomogr… · Application for Computed Tomography in Metrology for Micro Manufacturing G. Tosello1, J. A.

24/10/201336 G. Tosello, Technical University of DenmarkGerman-Austrian-Danish Workshop on Industrial CT Scanning

Thank you for your kind attention