Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... ·...

35
Anesthesia Machine Workshop James H. Philip, ME(E), MD, CCE

Transcript of Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... ·...

Page 1: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Anesthesia Machine Workshop

James H. Philip, ME(E), MD, CCE

Page 2: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Anesthesia Machine Workshop

James H. Philip, ME(E), MD, CCE

Anesthesiologist and Director Bioengineering

Department of Anesthesiology, Pain, and Peri-operative Medicine,

Brigham and Women’s Hospital

Medical Liaison for Anesthesia, Department of Biomedical Engineering

Partners HealthCare System

Associate Professor of Anaesthesia

Harvard Medical School

©

Copyright 2011, James H. Philip, all rights reserved.

Page 3: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Anesthesia Intranet Site contains all

www.bwhanesthesia.orgEducation

Anesthesia TechnologyEverything

Today (2011-08-18) :Anesthesia Machine Workshop 2011Aisys

Pre-Use Check –

Text and slideshowFabius

Pre-use Check –

Text and slideshow

Page 4: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Breathing Circuit Review

Page 5: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

The Circle-Absorber System and Fresh Gas Flow

Sampled & measured I, E

ExhaustExpired

CO2

Ab-

sorb-ant

200 mL/min

InspiredFresh Gas

Flow(FGF)

Ventilation Flow

Page 6: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

OR Ventilators

Page 7: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Old fashioned Uncompensated Bellows Movement

Gas Compression and and tubing expansion

decreased actual tidal volumeFresh gas flow (during inspiration) added to actual tidal volumeOlder Machines

Ohmeda Modulus 2

Draeger Narkomed

Page 8: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Fresh Gas Flow Independent of Ventilation

1998 - WYSIWYG What You Set

Is What You Get

Set Tidal Volume - patient receives what you set

Two ways to achieve this

GE- Fresh Gas Compensation

Active, feedback-control of inspired tidal volume

Draeger - Fresh Gas Decoupling

Passive separation of FGF during inspiration

Page 9: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Aestiva Aisys

GE – white with green trim to touch

Page 10: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Apollo GSFabius

GS

Draeger - white and blue

Page 11: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Exhaled tidal volume feedbackBellows moves approximately set tidal volumeExhaled volume is measuredNext and subsequent inspired bellows movement changesBellows volume changes based on previous exhaled tidal volumesSmartVent

in GE Aisys

and Aestiva

Page 12: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

GE-Datex- Ohmeda

Aestiva

GE Aestiva

example of all modern GE models

Pneumatic (oxygen)-driven bellows

Inspiratory and Expiratory Flow Sensor

Feedback from flow sensors to bellows drive

Bellows stops pushing when Measured Inspired Tidal Volume =

Set Inspired Tidal Volume

If you press the Flush button Bellows stops

Next breaths are confused

GE has many alarms that stop ventilation

Fresh Gas Compensation

Active, feedback control of inspired tidal volume

Page 13: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Aestiva VOSAll GE models have

similar differential-pressure flow sensor

Page 14: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

BTW, You only need to slide it out to here to zero the flow sensors

Page 15: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Flow sensors in manifold

Page 16: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Flow sensors in manifold

Reversed

Page 17: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Variable-Orifice flow Sensor (VOS)

Pressure Transducer

P1 P2

Flow

FAccurate over a wide range of flowEach one is calibrated in GE factoryMemory chip stores the calibration curve

P

Variable orifice

GE

Page 18: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Variable-Orifice flow Sensor (VOS)

Pressure Transducer

P1 P2

Flow

F

P

Variable orifice

This is what makes the SmartVent™

smart

This is GE’s core technology

Accurate over a wide range of flowEach one is calibrated in GE factoryMemory chip stores the calibration curve

Page 19: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Variable-Orifice flow Sensor (VOS)

Pressure Transducer

P1 P2

Flow

F

P

Variable orifice

One drop of water will make this fail

Accurate over a wide range of flowEach one is calibrated in GE factoryMemory chip stores the calibration curve

Page 20: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Water Management is important Use an HME or HMEFilter

Heat and Moisture Exchangerbetween patient and circuit

HMEWet Dry

Page 21: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Empty Aisys water reservoir

Between casesIf full during a caseUnexpected tidal volume readings

Page 22: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Draeger Ventilation

Virtual Fabius

www.simanest.org

Page 23: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Draeger Ventilation

Fresh Gas Flow DecouplingPassive separation of FGF during inspiration

Tiro, Fabius, Apollo

Page 24: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Draeger E-VentDraeger

Fabius

GS, Tiro, or Apollo

Piston

Page 25: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Fresh Gas Flow decouplingPiston pushes set volume into circuit and lungsFresh gas is stored in reservoir bag during inspirationPiston takes first from reservoir bag Piston takes second from exhaled gasReservoir bag moves with Mech

Vent (V or P mode)

Several phasesDraeger

Fabius, Tiro,

Apollo

Page 26: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Draeger Circuit

Page 27: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Draeger

Hot

wire

Anemometer

is cooled

by

flow

Calibrated for heat conductivity. User sets Desflurane Y/N

Fabius –

Used to Display, Expired tidal volume

Apollo –

Sensor #2 Used to measure

and control

inspired tidal volume

Creates “Auto-Flow VV”Same as GE PCVG

Page 28: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Implications of “Decoupling”

Collateral damageReservoir bag is in the circuit during ventilationAdds a second sequential mixing chamberOld gas stays in-play even when FGF is highThis does not matter much during inductionThis matters a lot during emergenceIn emergence we desire Inspired agent concentration = 0Even with bag ventilation, expired gas contaminates inspired

Page 29: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Example Fabius, End of case, VCV, FGF = 15 LPM

Page 30: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Inspired does not = setting even with high FGF (12 LPM)

Set

Set

I

IE

E

Page 31: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Draeger Circuit

Page 32: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Draeger has difficulty clearing the last bit of agent

Malignant Hyperthermia PreparationTarget is FI Agent

< 5 ppm Most ADSs require 10 minutesFabius and Apollo require > 1 hour

Inspired Charcoal agent absorber will overcome this

They are on order

Gunter JB, Ball, J Than-Win S. Preparation of the Draeger Fabius Anesthesia Machine for the Malignant–Hyperthermia Susceptible Patient. Anesth Analg 2008;107:1936 –45

Page 33: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

Aisys Inspired = 0 by second breath

3 minutes3 minutes

Page 34: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

GE Circuit

Pneumatic bellows ventilator

Page 35: Anesthesia Machine Workshopetherweb.bwh.harvard.edu/education/PHILIP/adsworkshop... · 2015-05-06 · Medical Liaison for Anesthesia, Department of Biomedical Engineering ... Text

End of Breathing Circuit and Ventilators

Next – Qustions

and Answers

Finally – review of Pre-Use Check