Andre´ Bu¨chter Load-related implant reaction of mini ... · porary use of the native dentition...
Transcript of Andre´ Bu¨chter Load-related implant reaction of mini ... · porary use of the native dentition...
Load-related implant reaction of mini-implants used for orthodontic anchorage
Andre BuchterDirk WiechmannStefan KoerdtHans Peter WiesmannJosef PiffkoUlrich Meyer
Authors’ affiliations:Andre Buchter, Stefan Koerdt, Hans PeterWiesmann, Josef Piffko, Ulrich Meyer, Departmentof Cranio-Maxillofacial Surgery, University ofMunster, Munster, GermanyDirk Wiechmann, Orthodontic Department,Medical School Hanover, Germany
Correspondence to:Dr Andre BuchterDepartment of Cranio-Maxillofacial SurgeryUniversity of MunsterWaldeyerstra�e 30D-48129 MunsterGermanyTel.: þ49 251 834 7013Fax: þ49 251 834 7020e-mail: [email protected]
Key words: anchorage, implant stability, orthodontic tooth movement, osseointegration
Abstract: The purpose of this study was to determine the clinical and biomechanical
outcome of two different titanium mini-implant systems activated with different load
regimens. A total of 200 mini-implants (102 Abso Anchors and 98 Dual Tops) were placed
in the mandible of eight Gottinger minipigs. Two implants each were immediately loaded
in opposite direction by various forces (100, 300 or 500 cN) through tension coils.
Additionally, three different distances between the neck of the implant and the bone rim
(1, 2 and 3 mm) were used. The different load protocols were chosen to evaluate the load-
related implant performance. The load was provided by superelastic tension coils, which are
known to develop a virtually constant force. Non-loaded implants were used as a reference.
Following an experimental loading period of 22 and 70 days half of the minipigs were
sacrificed, and implant containing bone specimens evaluated for clinical performance and
implant stability. Implant loosing was found to be statistically dependent on the tip
moment (TM) at the bone rim. Clinical implant loosing were only present when load
exceeded 900 cN mm. No movement of implants through the bone was found in the
experimental groups, for any applied loads. Over the two experimental periods the non-
loaded implants of one type of implant had a higher stability than those of the loaded
implants. Dual Tops implants revealed a slightly higher removal torque compared with
Abso Anchors implants. Based on the results of this study, immediate loading of mini-
implants can be performed without loss of stability when the load-related biomechanics do
not exceed an upper limit of TM at the bone rim.
The number of adult patients requiring
orthodontic therapy has undergone a
marked increase in recent decades (Bauer
& Diedrich 1990). However adults fre-
quently present pathologic findings like
periodontal and endodontic diseases, dys-
function of the mandibular joint, and early
loss of teeth (Diedrich 2000). As the tem-
porary use of the native dentition for ortho-
dontic anchorage is frequently limited, and
extraoral appliances are rejected for aes-
thetic reasons, an alternative approach
based on stable and cooperation-indepen-
dent anchorage is needed (Fritz et al. 2003).
Numerous studies have shown that con-
ventional osseointegrated implants are sui-
table for orthodontic loading and offer
stable anchorage for orthodontic purposes
(Gray et al. 1983; Roberts et al. 1994;
Odmann et al. 1988; Smalley et al. 1988;
Wehrbein & Diedrich 1993). Wehrbein &
Diedrich (1993) moreover reported that the
application of orthodontic forces positively
affects the peri-implant bone situation of
osseointegrated implants. In additional stu-
dies, Wehrbein (1994) and Wehrbein et al.
(1999) evaluated the quantity of minera-
lized bone in the vicinity of the implantCopyright r Blackwell Munksgaard 2005
Date:Accepted 29 September 2004
To cite this article:Buchter A, Wiechmann D, Koerdt S, Wiesmann HP,Piffko J, Meyer U. Load-related implant reaction ofmini-implants used for orthodontic anchorage.Clin. Oral Impl. Res. 16, 2005; 473–479doi: 10.1111/j.1600-0501.2005.01149.x
473
surface and found that osseous adaptation
mechanisms during orthodontic loading
resulted in increased implant stability.
Mini-implants with a small diameter
were developed to ease the surgical inser-
tion procedure and to allow anchorage at
different positions of the alveolar bone.
Both clinical and experimental studies
have demonstrated that these implants
were basically able to provide sufficient
and stable anchorage for tooth movement
during the entire time period of orthodontic
therapy. Whereas dental implants have
high long-term success rates at about 90–
95% (Adell et al. 1981; Albrektsson et al.
1981), mini-implants failed to reach these
high success rates despite the shorter usage
period. Besides infection-dependent im-
plant failure, factors that decrease implant
success are related to interrelated biome-
chanical parameters such as implant de-
sign, bone quality and extent and time of
loading (Orenstein et al. 1998; Miyawaki
et al. 2003). Whereas titanium screws have
mainly been evaluated as orthodontic an-
chorage systems in clinical trials (Park
2001), less is known about the factors
affecting implant stability in respect to
load-related biomechanics. Recent reports
on mini-implant anchors used in clinical
studies (Park 2001; Sawa et al. 2001)
demonstrated a high number of failures.
Thus, standard clinical protocols for a bio-
mechanical by related insertion and loading
scheme are not present.
The objective of the present study was
therefore to investigate the stability of two
mini-implant systems under immediately
applied continuous orthodontic loading.
Different loads as well as various insertion
situations, mimicking the clinical situa-
tion, were used to gain insight into an
optimised application of these anchor
systems.
Material and methods
Implant designs
Screw-shaped titanium mini-implants
with a length of 10 mm and a diameter
of 1.1 mm (Abso Anchors, Dentos. Inc.,
Taegu, Seoul, Korea) and 1.6 mm (Dual
Tops, Jeil Medical Corporation, Korea)
were used in this study (Fig. 1).
Experimental animals and mini-implants
Eight male Gottinger minipigs, 14–16
months of age and with an average body
weight of 35 kg, were used. A total of 200
implants (102 Abso Anchors and 98 Dual
Tops) were placed in the mandible of
Gottinger minipigs (Fig. 2). In accordance
with the experimental design, two treat-
ment groups were tested in each animal
(Table 1). The study was approved by the
Animal Ethics Committee of the Univer-
sity of Munster under the reference num-
ber G4/2004.
Surgical procedure
All surgery was performed under sterile
conditions in a veterinary operating thea-
tre. The animals were sedated with an
intramuscular injection of ketamine
(10 mg/kg), atropine (0.06 mg/kg) and
stresnil (0.03 mg/kg). In the areas exposed
to surgery 4 ml of local anaesthesia (2%
lidocaine with 12.5mg/ml epinephrine,
xylocain/adrenalines, Astra, Wedel, Ger-
many) was injected. The left and right
mandible was exposed by skin incisions
via fascial-periosteal flaps (Fig. 3). There-
after, the implants were placed in the
caudal part of the mandible, using spiral
drills according to the standard protocol of
the manufacturer.
Two corresponding implants were in-
serted by prefabricated distance holders in
a standard distance of 17 mm (Fig. 3) at
various insertion depths (Fig. 4), leading to
force applications at different (1, 2 and
3 mm) distances from the crestal bone
(Table 1). After insertion the screws were
immediately loaded with transverse forces.
The load was provided by Sentalloys
(GAC, Grafelfing, Germany) superelastic
tension coil springs (Fig. 5), which develop
a virtually constant force of 100, 300 or
500 cN (Table 1). The tension coil springs
were protected against soft tissue ingrowth
Fig. 1. Screw-shaped titanium mini-implants with a
length of 10 mm and a diameter of 1.1 mm (Abso
Anchors) left and 1.6 mm (Dual Tops) right.
Fig. 2. Minipig with placed mini-implants in the
mandible.
Table 1. Experimental design
Minipigs 8Implants 192þ 8 fracturedSacrifice (days) 22 70Implant type Abso Anchors/Dual Tops Dual Tops/Abso Anchors
Implants 96 960 cN, 1 mm lever arm Group: LT 0 8 � 8 � 8 � 8 �100 cN, 1 mm lever arm Group: LT 1 8 � 8 � 8 � 8 �300 cN, 1 mm lever arm Group: LT 2 8 � 8 � 8 � 8 �500 cN, 1 mm lever arm Group: LT 3 8 � 8 � 8 � 8 �300 cN, 2 mm lever arm Group: LT 4 8 � 8 � 8 � 8 �300 cN, 3 mm lever arm Group: LT 5 8 � 8 � 8 � 8 �
LT, loading type.
Buchter et al . Use of mini-implants in orthodontic anchorage
474 | Clin. Oral Impl. Res. 16, 2005 / 473–479
by small latex tubes. Screws without load-
ing served as control. Six different groups
(loading types¼LT, Table 1) were evalu-
ated: 0 cN, 1 mm lever arm (LT 0); 100 cN,
1 mm lever arm (LT 1); 300 cN, 1 mm
lever arm (LT 2); 500 cN, 1 mm lever arm
(LT 3); 300 cN, 2 mm lever arm (LT 4) and
300 cN, 3 mm lever arm (LT 5). The tip
moment (TM) at the bone rim is calculated
by the formula: TM¼ force � lever arm.
After coil application the skin and the
fascia periosteum were closed in separate
layers with single resorbable sutures (Vi-
cryls4-0, Ethicon, Norderstedt, Ger-
many). Perioperatively, an antibiotic was
administered subcutaneously (2.5 ml benzyl-
penicillin/dihydrostreptomycin, Tardomy-
cels, BayerVital, Leverkusen, Germany),
every 48 h for 7 days.
Clinical follow-up
The animals were inspected after the first
few postoperative days for signs of wound
dehiscence or infection and weekly there-
after to assess general health. Loading per-
iods of 22 and 70 days were used for half of
the implants. At days 22 and 70 animals
were sacrificed with an overdose of T61
given intravenously. Following euthanasia,
mandibular block specimens containing
the two corresponding implants and sur-
rounding tissues were dissected from all of
the animals. The samples were sectioned
by a saw to remove unnecessary portions of
bone and soft tissue. Implants were con-
trolled for signs of material failure, clinical
mobility, peri-implant infection and im-
plant distance was measured.
Removal torque testing
The removal torque test was performed by
applying a counter-clockwise rotation of
the implant about its axis at a rate of
0.11/s according to the experimental set
up of Li et al. (2002) (Fig. 6). For each
implant the torque rotation curve was
recorded. The removal torque was defined
as the maximum torque (N mm) on the
curve.
Statistical analysis
Mean values and standard deviations were
calculated for removal torque testing. Be-
tween both groups descriptive statistic was
used, because of the different implant geo-
metry. Multiple comparisons between
groups were performed using two-way ana-
lysis of variance and t-tests. Difference was
considered significant when Po0.05. All
calculations were performed through the
use of SPSS for Windows (SPSS Inc.,
Chicago, IL, USA).
Results
Clinical observation
During the insertion process six Abso An-
chors and two Dual Tops implants frac-
tured the caudal of the head and were not
included in the study; new implants were
inserted. All other implants were clinically
immobile after insertion. Implants were
monocortically inserted in the mandibular
bone. The animals recovered well after
surgery and no signs of infection were
noted at any time during the observation
period. All together five tension coils were
lost (two Abso Anchors LT 4 (one after 22
and one after 70 days); two Abso Anchors
LT 5 (22 days); one LT 3 (22 days) (Tables 2
and 3) and one Dual Tops LT 5 (22 days).
Three of the Abso Anchors implants
showed a little tip in the direction of the
applied force, so that the tension coils
slipped from the implants head. The reason
for the loss of the two tension coils is
unknown. All other implants, without
the five loosening implants, showed a
good stability at a clinical level. None of
the implants was found to have moved
through the bone. One Dual Tops, one
Abso Anchors LT 5 (22 days) and three
Abso Anchors LT 5 (70 days) showed an
implant bending in the upper part of the
screws accompanied by peri-implant
bone loss and slight signs of inflammation
(Table 4).
Removal torque
During the removal torque test one Abso
Anchors and one Dual Tops implants (22
days) fractured the caudal of the head. Over
the experimental periods the Dual Tops
(22 (without LT 3) and 70 days) and Abso
Anchors (22 days) reference implants of
both groups revealed a significantly higher
implant/bone stability than those of the
loaded implants (Tables 5 and 6). Dual
Tops implants revealed a higher mean
removal torque value compared with
Fig. 3. Prefabricated distance holder for two corre-
sponding implants.
Fig. 4. Various insertion depths by prefabricated
distance holder.
Fig. 5. Loaded implants with transverse force by
superelastic tension coil springs.
Fig. 6. Removal torque test.
Buchter et al . Use of mini-implants in orthodontic anchorage
475 | Clin. Oral Impl. Res. 16, 2005 / 473–479
Abso Anchors implants. As the implant
geometry differed between both implants,
removal torque testing was not statistically
compared between these two groups. The
results of the torque testing increased for
Abso Anchors implants stability over
time (significantly at LT 3). Whereas
Dual Tops implants showed no increase
in implant stability during the experiment.
Implant stability in terms of resistance
towards torsion forces were not statistically
dependent from the applied TMs.
Discussion
Whereas conventional or modified oral
implants have been shown to successfully
serve as anchorage for orthodontic appli-
ances (Wehrbein et al. 1996) mini-implants
failed to reach these high success rates.
When the high failure rates of mini-
implants are under evaluation two main
factors have to be considered. The biome-
chanical loading of peri-implant bone as
well as the time schedule of loading, have
been shown to have a major impact on the
peri-implant bone healing (Meyer et al.
2004) and can be assumed to determine
the clinical fate of mini-implants. A mono-
cortical mandibular anchorage system as
well as a load application in magnitudes of
orthodontic practice was therefore used in
the present study to gain insight into an
optimised clinical loading protocol.
Two main findings can be demonstrated
through our experimental approach. First,
implant failure is directly related to the
tipping moment (TM) at the bone rim.
Second, by reducing the main TM under
a threshold of 900 cN mm (300 cN and
3 mm lever arm) mini-implants can be
loaded immediately without impairment
of implant stability and implant success
rates. In most of the long-term clinical
studies implant failures have been attribu-
ted to overloading or excessive loading
when no peri-implantitis phenomena
were present. Most of the implants losses
were considered to be the result of exces-
sive strains and stresses at the bone/
implant interface (Adell et al. 1981). The
present study shows that all the implants
installed into mandibular bone were suc-
cessfully loaded and remained stable
throughout the entire duration of the study,
when tipping forces were not higher then
900 cN mm. This finding is in agreement
with earlier studies (Roberts et al. 1989,
1994) demonstrating that implants re-
mained stable when subjected to tip forces
ranging from 100 to 300 cN. Studies by
Isidor (1996, 1997) revealed that excessive
lateral loading of conventional osseointe-
grated implants, a loading regimen that
was also used in our experimental set up
(900 cN mm), resulted in a high risk for the
loss of osseointegration. The fact that the
degree of loading has a direct influence on
the stability of implants confirms the
superior influence of loads generated by
function. A number of in vitro studies,
including finite-element analysis (FEA),
have reported stress concentrations to oc-
cur in the marginal peri-implant bone after
lateral or oblique load application (Borchers
& Reichart 1983; Soltesz & Siegele 1984;
Mihalko et al. 1992). FEA clearly demon-
strate that the local strain distribution had
a significant impact on the biological activ-
ity of the adjacent bone tissue (Meyer et al.
2001a). FEA calculations of Isidor (1997)
revealed that not all forces may be tolerated
by the crestal bone on a long-term basis.
High strain values above 6700 mstrain re-
sulted in peri-implant bone resorption and
a negative balance between bone apposition
Table 2. Clinical results after 22 days
Implants types Dual Tops Abso Anchors
Tensioncoil lost
Implantloosening
Implant fracture/removal torque
Tensioncoil lost
Implantloosening
Implant fracture/removal torque
0 cN, 1 mm neck/bone distance100 cN, 1 mm distance to the bone300 cN, 1 mm neck/bone distance500 cN, 1 mm neck/bone distance 1300 cN, 2 mm neck/bone distance 1 1300 cN, 3 mm neck/bone distance 1 1 2 1 1
Table 3. Clinical results after 70 days
Implants types Dual Tops Abso Anchors
Tensioncoil lost
Implantloosening
Implant fracture/removal torque
Tensioncoil lost
Implantloosening
Implant fracture/removal torque
0 cN, 1 mm neck/bone distance100 cN, 1 mm distance to the bone300 cN, 1 mm neck/bone distance500 cN, 1 mm neck/bone distance300 cN, 2 mm neck/bone distance 1300 cN, 3 mm neck/bone distance 3
Table 4. Implants loosening in relation totip moment (TM) [TM (N mm)¼ force(N) � lever arm (mm)]
0
1
2
3
4
Imp
lan
t lo
ose
nin
g (
n)
100cNmm
500cNmm
900cNmm
Abso Anchor
Dual Top
Buchter et al . Use of mini-implants in orthodontic anchorage
476 | Clin. Oral Impl. Res. 16, 2005 / 473–479
and bone resorption (Isidor 1997), confirm-
ing the hypothesis posted by Frost (1998) as
well as the findings of several authors, who
reported adaptation mechanisms to a win-
dow of physiological loading (Meyer et al.
1999a). This is also in agreement with
observations made by Hoshaw et al.
(1994), who reported that high stress con-
centration may cause marginal bone re-
sorption around implants in a load model
(Meyer et al. 2001b; Buchter et al. 2005).
Some studies using static load models con-
firm our results that low mechanical load is
not accompanied by marginal bone defects
and impaired bone mineralization (Meyer
et al. 2001a). This, in turn, means that
mini-implants could serve as anchorage for
orthodontic force systems when loads do
not exceed a tolerable strain level. It is
important to note that the amount of
stresses and strains are dependent on the
geometry of the screw as well as on the
mechanical properties of the implant and
bone. As we have not performed an FEA
analysis in our investigation, strain and
stress values in peri-implant bone cannot
directly be correlated to implant instabil-
ities and loss rate in our study. The finding
of differences in implant stability between
implants systems may be based in the
differences in implant geometry influen-
cing the strain and stress distribution in
peri-implant bone (Buchter et al. 2004).
The results of the present study confirm
that implants can be immediately loaded
by continuous forces. While it has been
stated in some studies that immediate or
early loading of oral implants impedes
successful osseointegration (Sagara et al.
1993), recent studies suggest that immedi-
ate loading can be performed under defined
conditions (Meyer et al. 1999a, 1999b,
2004). Different investigations have indi-
cated that although premature loading has
been interpreted as inducing fibrous tissue
interposition, immediate loading per se is
not responsible for fibrous encapsulation. It
is the excess of micromotions during the
healing phase that interferes with bone
repair (Szmukler-Moncler et al. 1998). Ex-
perimental observations suggest that mi-
cromotions do not systematically lead to
fibrous tissue interposition and that toler-
ance to micromotion is design and load
dependent. The effect of a continuous lat-
eral or oblique load on the peri-implant
bone has been described in in vitro experi-
ments, also using FEA (Soltesz & Siegele
1984; Mihalko et al. 1992). As these mod-
els reveal that under conditions of a direct
bone/implant contact and moderate load
applications, marginal strains are lower
than the upper limit of tolerable bone
strains, immediate loading seems to be
possible without impairment of implant
stability. Furthermore, experimental stu-
dies have not generally been able to de-
monstrate implant loosening induced by
orthodontic load (Wehrbein et al. 1997;
Akin-Nergiz et al. 1998) even when these
loads were applied immediately. Addition-
ally, a number of clinical studies have
confirmed a positive effect of orthodontic
loading on the stability of titanium screw
implants, as well as the effects on peri-
implant bone (Roberts et al. 1994; Wehr-
bein et al. 1993, 1997; Hurzeler et al.
1998). The clinical view that loaded mini-
implants showed no movement through
the bone is confirmed by the present study.
Whereas this study was of short duration
and, hence, possible long-term effects on
implant stability were not elucidated, it
must be stressed that most of the load
Table 5. Removal torque values (N mm), Abso Anchors
Treatment group Loadingtype (LT)
Days 22 AbsoAnchors (N mm)
Days 70 AbsoAnchors (N mm)
Changewithin group
Change days22 vs. 70
Reference, 1 mm distance neck/bone 0 31 � 1.41 29.85 � 1.48 � 1.15 � 1.53 P¼ 0.482100 cN, 1 mm distance neck/bone 1 11.15 � 1.33 38.25 � 11.99 27.1 � 11.99 P¼ 0.109300 cN, 1 mm distance neck/bone 2 15.35 � 6.54 41.75 � 6.53 26.4 � 6.83 P¼ 0.23500 cN, 1 mm distance neck/bone 3 13.95 � 4.08 33 � 141 19.05 � 4.1 P¼ 0.018300 cN, 2 mm distance neck/bone 4 15.5 � 4.18 30.15 � 10.86 14.65 � 11.06 P¼ 0.233300 cN, 3 mm distance neck/bone 5 15.05 � 4.24 15.25 � 1.25 0.2 � 2.46 P¼ 0.938Change reference vs. 100 cN, 1 mm distance P¼ 0.001 P¼ 0.535Change reference vs. 300 cN, 1 mm distance P¼ 0.003 P¼ 0.165Change reference vs. 500 cN, 1 mm distance P¼ 0.024 P¼ 0.12Change reference vs. 300 cN, 2 mm distance P¼ 0.004 P¼ 0.98Change reference vs. 300 cN, 3 mm distance P¼ 0.0043 P¼ 0.001
Table 6. Removal torque values (N mm), Dual Tops
Treatment group Loadingtype (LT)
Days 22 DualTops (N mm)
Days 70 DualTops (N mm)
Change withingroup
Change days22 vs. 70
Reference, 1 mm distance neck/bone 0 109 � 2.08 111.05 � 3.05 2.05 � 2.08 P¼ 0.363100 cN, 1 mm distance neck/bone 1 76.25 � 3.8 73.9 � 6.49 � 2.35 � 7.53 P¼ 0.765300 cN, 1 mm distance neck/bone 2 72.75 � 5.28 55.6 � 5.32 � 17.15 � 7.5 P¼ 0.062500 cN, 1 mm distance neck/bone 3 82.3 � 15.72 52.05 � 6.89 � 30.25 � 17.17 P¼ 1.29300 cN, 2 mm distance neck/bone 4 42.95 � 5.09 55.15 � 7.41 12.2 � 6.3 P¼ 0.101300 cN, 3 mm distance neck/bone 5 60.15 � 5.87 59.95 � 18.1 � 0.2 � 19.59 P¼ 0.992Change reference vs. 100 cN, 1 mm distance P¼ 0.001 P¼ 0.002Change reference vs. 300 cN, 1 mm distance P¼ 0.003 P¼ 0.004Change reference vs. 500 cN, 1 mm distance P¼ 0.143 P¼ 0.01Change reference vs. 300 cN, 2 mm distance P¼ 0.001 P¼ 0.001Change reference vs. 300 cN, 3 mm distance P¼ 0.002 P¼ 0.0797
Buchter et al . Use of mini-implants in orthodontic anchorage
477 | Clin. Oral Impl. Res. 16, 2005 / 473–479
application are time limited in clinical
orthodontic practice.
We used in the surgical procedures a
closed-flap technique differs from the clin-
ical procedure in which no closed flap is
used. Depending on the localization the
miniscrews are either through attached
gingiva or the screw head is covered again
with mucosa letting only the wire or oil
spring penetrate the mucosa. In any case
there is an increased infection risk com-
pared with a closed-flap technique thus in
turn increasing the risk for screw failure.
To eliminate this infection risk, we used in
our study the closed-flap technique. Clin-
ical randomised prospective studies will be
necessary to investigate the implant loss
rate.
In conclusion the present study indicate
that a moderate static, lateral loading of
mini-implants with forces typically used
for orthodontic movement, does not impair
implant stability. It may be assumed that
above a defined threshold marginal bone
strains cannot be tolerated by the bone
tissue, leading to implant loosening.
Resume
Le but de cette etude a ete de determiner ce qui se
passait cliniquement et biomecaniquement au ni-
veau de deux systemes de mini-implants en titane
actives par differents regimes de mise en charge.
Deux cents mini-implants (102 Abso Anchors et 98
Dual Tops) ont ete places dans la mandibule de huit
mini-porcs de Gottinger. Deux implants chacun ont
ete immediatement mis en charge dans une direc-
tion opposee avec des forces variables (100 cN,
300 cN ou 500 cN) par des ressorts de tension. De
plus trois distances differentes entre l’epaulement de
l’implant et le rebord osseux (1, 2, 3 mm) ont ete
utilises. Les differents protocoles de mise en charge
ont ete choisis pour evaluer la performance implan-
taire vis-a-vis de la mise en charge. Cette charge
etait produite par des ressorts de tension super-
elastiques connues pour developper une force
quasi-constante. Des implants sans charge ont servi
de controle. Suivant une periode de mise en charge
de 22 et 70 jours la moitie des mini-porcs ont ete
euthanasies et les specimens comprenant l’os et
l’implant evalues pour leur performance clinique et
leur stabilite implantaire. La mobilite implantaire a
ete reconnue comme statisquement dependante du
moment de l’extremite au niveau du bord osseux. La
mobilite implantaire clinique a seulement ete re-
ncontree lorsque la charge depassait 900 cN mm.
Aucun mouvement des implants n’a ete trouvee
dans le groupe experimental pour aucune des forces
appliquees. Durant les deux periodes experimentales
les implants non-charges d’un type d’implant avai-
ent une stabilite superieure a celle des implants
charges. Les implants Dual Tops avaient une force
de torsion d’enlevement un peu plus importante que
les implants Abso Anchors. La mise en charge
immediate de mini-implants peut etre effectuee sans
perte de stabilite lorsque la charge en relation avec la
biomecanique n’excede pas une limite superieure du
moment de l’extremite au niveau du bord osseux.
Zusammenfassung
Belastungsbedingte Reaktionen von Mini-Implan-
taten, welche fur orthodontische Verankerungen
verwendet werden
Es war das Ziel dieser Studie, die klinische und
biomechanische Reaktion bei zwei verschiedenen
Titan Mini-Implantat-Systemen, welche verschie-
denen Belastungen ausgesetzt wurden, zu bestim-
men. Insgesamt wurden 200 Mini-Implantate (102
Abso Anchors und 98 Dual Tops) in die Unterkie-
fer von 8 Gottinger Minipigs eingesetzt. Immer zwei
Implantate wurden jeweils sofort in entgegen gesetz-
ter Richtung mit unterschiedlichen Kraften
(100 cN, 300 cN oder 500 cN) durch Zugfedern
belastet. Zusatzlich wurden drei verschiedene Ab-
stande zwischen dem Hals des Implantats und der
Knochenkante (1 mm, 2 mm, 3 mm) verwendet.
Die unterschiedlichen Belastungsprotokolle wurden
ausgewahlt, um die belastungsbedingte Implanta-
treaktion auszuwerten. Die Belastung wurde mit
superelastischen Zugfedern durchgefuhrt, welche
dafur bekannt sind, dass sie eine relativ konstante
Kraft entwickeln. Unbelastete Implantate dienten
als Kontrolle. Nach einer experimentellen Belas-
tungsdauer von 22 und 70 Tagen wurde je die Halfte
der Minipigs geopfert. An Knochenpraparaten mit
Implantaten wurde die klinische Reaktion und die
Stabilitat der Implantate ausgewertet. Es wurde
entdeckt, dass die Lockerung von Implantaten sta-
tistisch signifikant abhangig vom Kippmoment auf
Hohe der Knochenkante war. Eine klinische Lock-
erung von Implantaten trat nur auf, wenn die Belas-
tung 900 cN mm uberstieg. Bei den experimentellen
Gruppen konnte mit den applizierten Kraften keine
Bewegung der Implantate durch den Knochen aus-
gelost werden. Ueber die zwei Abschnitte des Ex-
periments wiesen die nicht belasteten Implantate
eines Typs eine grossere Stabilitat als die belasteten
Implantate auf. Die Dual Tops Implantate zeigten
leicht hohere Ausdrehmomente als die Abso An-
chors Implantate. Aufgrund der Resultate dieser
Studie kann eine Sofortbelastung von Mini-Implan-
taten ohne Stabilitatsverlust durchgefuhrt werden,
sofern durch die Belastung die obere Grenze eines
Kippmoments auf Hohe der Knochenkante nicht
uberschreitet.
Resumen
El proposito de este estudio fue determinar los
resultados clınicos y biomecanicos de dos sistemas
diferentes de mini-implantes de titanio activados
con diferentes regimenes de carga. Se colocaron un
total de 200 mini-implantes (102 Abso Anchors y
98 Dual Tops) en la mandıbula de 8 minicerdos
Gottinger. Se cargaron dos implantes inmediata-
mente en direcciones opuestas con varias fuerzas
(100 cN, 300 cN o 500 cN) a traves de resortes de
tension. Se usaron, adicionalmente, tres distancias
diferentes entre el cuello del implante y el reborde
oseo (1 mm, 2 mm, 3 mm). Los diferentes protocolos
de carga se eligieron para evaluar el rendimiento
relacionado con la carga. La carga se suministro por
medio de resortes de tension superelasticos, conoci-
dos por desarrollar una fuerza virtualmente con-
stante. Los implantes no cargados se usaron como
referencia. Tras un periodo experimental de carga de
22 y 70 dıas la mitad de los minicerdos se sacrifi-
caron, y se evaluaron los implantes conteniendo
especımenes oseos para el rendimiento clınico y
estabilidad de los implantes. El aflojamiento del
implante se encontro que era estadısticamente de-
pendiente de la punta del momento en el reborde
oseo. El aflojamiento clınico del implante solo se
presento cuando la carga excedio los 900 cN mm.
No se encontro movimiento de los implantes a
traves del hueso en los grupos experimentales, en
ninguna de las cargas aplicadas. A lo largo de los dos
periodos experimentales los implantes sin carga de
un tipo de implante tuvieron una mayor estabilidad
que aquellos de los implantes cargados. Los im-
plantes Dual Tops necesitaron un mayor torque
de remocion en comparacion con los implantes Abso
Anchors. Basandose en los resultados de este estu-
dio, la carga inmediata se puede llevar a cabo sin
perdida de estabilidad cuando la biomecanica rela-
cionada con la carga no supere un lımite superior de
la punta del momento en el reborde oseo.
Buchter et al . Use of mini-implants in orthodontic anchorage
478 | Clin. Oral Impl. Res. 16, 2005 / 473–479
References
Adell, R., Lekholm, U., Rockler, B. & Branemark,
P.I. (1981) A 15-year study of osseointegratd
implants in the treatment of the edentulous jaw.
International Journal of Oral Surgery 10:
387–416.
Akin-Nergiz, N., Nergiz, I., Schulz, A., Arpak, N.
& Niedermeier, W. (1998) Reactions of peri-im-
plant tissues to continuous loading of osseointe-
grated implants. American Journal of Ortho-
dontics Dentofacial Orthopedics 114: 292–298.
Albrektsson, T., Branemark, P.-I., Hasson, H.A. &
Lindstrom, J. (1981) Titanium implant. Require-
ments for ensuring a long-lasting direct bone
anchorage in man. Acta Orthopaedica Scandina-
vica 52: 155–170.
Bauer, W. & Diedrich, P. (1990) Motivation und
Erfolgsbeurteilung erwachsener Patienten zur
kieferorthopadischen Behandlung – Interpretation
einer Befragung. Fortschritte der Kieferorthopadie
51: 180–188.
Borchers, L. & Reichart, P. (1983) Three-dimen-
sional stress distribution around a dental implant
at different stages of interface development. Jour-
nal of Dental Research 62: 155–159.
Buchter, A., Kleinheinz, J., Wiesmann, H.-P., Jayar-
anan, M., Joos, U. & Meyer, U. (2005) Inter-
face reaction at dental implants inserted in con-
densed bone. Clinical Oral Implants Research
16: 1–8.
Buchter, A., Kleinheinz, J., Wiesmann, H.-P., Ker-
sken, J., Weytother, H., Nienkemper, M., Joos, U.
& Meyer, U. (2004) Biological and biomechanical
evaluation of bone remodelling and implant stabi-
lity after using an osteotome technique. Clinical
Oral Implants Research 12, in press.
Diedrich, P. (2000) Kieferorthopadische Behandlung
Erwachsener. In: Diedrich, P., ed. Hrsg. Praxis
der Zahnheilkunde. Kieferorthopadie III, 174–
208. Munchen-Jena: Urbarn & Fischer.
Fritz, U., Diedrich, P., Kinzlnger, G. & Al-Said, M.
(2003) The anchorage quality of mini-implants
towards. Journal of Orofacial Orthopedics 64:
293–304.
Frost, H.M. (1998) A brief review for orthopaedic
surgeons: fatigue damage (microdamage in bone
(its determinants and clinical implications)). Jour-
nal of Orthopedics Science 3: 272–281.
Gray, B., Steen, M. & King, G. & Clark, A.E. (1983)
Studies on the efficacy of implants as orthodontic
anchorage. American Journal of Orthodontics
Dentofacial Orthopedics 83: 311–317.
Hoshaw, S.J., Brunski, J.B. & Cochran, G.V.B.
(1994) Mechanical loading of Branemark implants
affects interfacial bone modeling and remodeling.
International Journal of Oral & Maxillofacial
Implants 9: 345–360.
Hurzeler, M.B., Qumones, C.R., Kohal, R.J.,
Rohde, M., Slurb, J.R. & Teuscher, U., et al.
(1998) Changes in peri-implant tissues sub-
jected to orthodontic forces and ligature break-
down in monkeys. Journal of Periodontology 69:
396–404.
Isidor, F. (1996) Loss of osseointegration caused
by occlusal load of oral implants. A clinical and
radiographic study in monkeys. Clinical Oral
Implants Research 7: 143–152.
Isidor, F. (1997) Histological evaluation of peri-im-
plant bone at implants subjected to occlusal over-
load or plaque accumulation. Clinical Oral
Implants Research 8: 1–9.
Li, D., Ferguson, S., Beutler, T., Cochran, D., Sittig,
C., Hirt, P. & Buser, D. (2002) Biomechanical
comparison of sandblasted and acid-etched and
the machined and acid-etched titanium surface
for dental implants. Journal of Biomedical Mate-
rials Research 60: 345–332.
Meyer, T., Meyer, U., Stratmann, U., Wiesmann,
H.-P. & Joos, U. (1999a) Identification of apopto-
tic cell death in distraction osteogenesis. Cell
Biology International 23: 439–446.
Meyer, U., Joos, U., Mythili, J., Stamm, T., Hohoff,
A., Stratmann, U. & Wiesmann, H.-P. (2004)
Ultrastructural characterization of the implant/
bone interface of immediately loaded dental im-
plants. Biometerials 25: 1959–1967.
Meyer, U., Vollmer, D., Runte, C., Bourauel, C. &
Joos, U. (2001a) Bone loading pattern around
implants in average and atrophic edentulous max-
illae: a finite-element analysis. Jounal of Cranio-
Maxillofacial Surgery 29: 100–105.
Meyer, U., Wiesmann, H.P., Kruse-Losler, B.,
Handschel, J., Stratmann, U. & Joos, U. (1999b)
Strainrelated bone remodeling in distraction os-
teogenesis of the mandible. Plastic and Recon-
structive Surgery 103: 800–807.
Meyer, U., Wismann, H.-P., Meyer, T., Schulze-
Osthoff, D., Jasche, J., Kruse-Losler, B. & Joos, U.
(2001b) Microstructural investigations of strain-
related collagen mineralization. British Journal of
Oral and Maxillofacial Surgery 39: 381–289.
Mihalko, W.M., May, T.C., Kay, J.F. & Krause,
W.R. (1992) Finite element analysis of inter-
face geometry effects on the crestal bone sur-
rounding a dental implant. Implant Dentistry
Fall 1: 212–217.
Miyawaki, S., Koyama, I., Masahide, I., Katsuaki,
M., Sugahara, T. & Tankano-Yamamoto, T.
(2003) Factors associated with the stability of
titanium screw placed in the posterior region
for orthodontic anchorage. Orthodontic Waves
11: 373–378.
Odman, J., Lekholm, U. & Lemt, T., et al. (1988)
Osseointegrated titanium implants a new ap-
proach in orthodontic treatment. European Jour-
nal of Orthodontics 10: 98–105.
Orenstein, I.H, Tarnow, D.P., Morris, H.F. & Ochi,
S. (1998) Factors affecting implant mobility at
placement and integration of mobile implants
at uncovering. Journal of Periodontology 69:
1404–1412.
Park, H.S. (2001) The orthodontic treatment using
micro-implant: the clinical application of MIA
(micro-implant anchorage). American Journal of
Orthodontics and Dentofacial Orthopedics 86:
95–111.
Roberts, E.W., Helm, F.R., Marshall, K.J. & Gongl-
off, R.K. (1989) Rigid endosseous implants for
orthodontic and orthopedic anchorage. The Angle
Orthodontist 59: 247–256.
Roberts, W.E., Nelson, C.L. & Goodacre, C.J.
(1994) Rigid implant anchorage to close a man-
dibular first molar extraction site. Journal of
Clinical Orthodontics 27: 693–704.
Sagara, M., Akagawa, Y., Nikai, H. & Tsuru, H.
(1993) The effects of early occlusal loading on one-
stage titanium alloy implants in beagle dogs: a
pilot study. Journal of Prosthetic Dentistry 69:
281–288.
Sawa, Y., Goto, N., Suzuki, K. & Kamo, K. (2001)
The new method for maxillary retraction of the
anterior teeth using a titanium microscrew as
anchorage. Orthodontic Waves 60: 328–331.
Smalley, W., Shapiro, P., Hohl, T., Kokich, V.G. &
Branemark, P.I. (1988) Osseointegrated titanium
implants for maxillofacial protraction in mon-
keys. American Journal of Orthodontics and
Dentofacial Orthopedics 94: 285–295.
Soltesz, U. & Siegele, D. (1984) Einfluss der Stei-
figkeit des Implantmaterials auf die in Knochen
erzeugten spannungen. Deutsche Zahnarztliche
Zeitung 39: 183–186.
Szmukler-Moncler, S., Salama, S., Reingewirtz, Y.
& Dubruille, J.-H. (1998) Timing of loading
and effect of micro-motion on bone–implant
interface: a review of experimental literature.
Journal of Biomedical Materials Research 43:
193–203.
Wehrbein, H. (1994) Enossale Titanimplantate als
orthodontische Verankerungselemente: Experi-
mentelle Untersuchungen und klinische Anwen-
dungen. Fortschritte der Kieferorthopadie 55:
236–250.
Wehrbein, H. & Diedrich, P. (1993) Endosseous
titanium implants during and after orthodontic
load an experimental study in the dog. Clinical
Oral Implants Research 4: 76–82.
Wehrbein, H., Glatzmaier, J., Mundwiller, U. &
Diedrich, P. (1996) The orthosystem – a new
implant system for orthodontic anchorage in
the palate. Journal of Orofacial Orthopedics 57:
142–153.
Wehrbein, H., Glatzmaier, J. & Yildirim, M. (1997)
Orthodontic anchorage-capacity of short titanium
screw implants in the maxillas. An experimental
study in the dog. Clinical Oral Implants Re-
search 8: 131–141.
Wehrbein, H., Yildirim, M. & Diedrith, P. (1999)
Osteodynamics around orthodontically loaded
short maxillary implants. Journal of Orofacial
Orthopedics 60: 409–415.
Buchter et al . Use of mini-implants in orthodontic anchorage
479 | Clin. Oral Impl. Res. 16, 2005 / 473–479