Analyzing the Image Density. Density Overall blackening of the image.
-
Upload
amara-martell -
Category
Documents
-
view
217 -
download
0
Transcript of Analyzing the Image Density. Density Overall blackening of the image.
![Page 1: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/1.jpg)
Analyzing the ImageAnalyzing the Image
DensityDensity
![Page 2: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/2.jpg)
DensityDensity
Overall blackening of the image
![Page 3: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/3.jpg)
Assessing DensityAssessing Density
Anatomical area of interest– OD 0.25-2.50
Darker image preferable to light image
![Page 4: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/4.jpg)
Factors Affecting DensityFactors Affecting DensitymAsmAs
Controlling Factor
In the straight line portion of the sensitometric curve, density proportional to log relative exposure.
mAs can be adjusted to compensate for changes made in other technical factors.
![Page 5: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/5.jpg)
DensityDensitymAs as Controlling FactormAs as Controlling Factor
Reciprocity Law– Density should remain the same as long as the
same mAs is used regardless of the mA and time combination
Reciprocity Law failure– Occurs at short exposure times (less than .01)
or long exposure times (several seconds)
![Page 6: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/6.jpg)
DensityDensitymAs as Controlling FactormAs as Controlling Factor
Minimum change 30% in exposure– Either 30% change in mAs or other factors that equal 30%
change in exposure
For mAs changes, adjust in increments of 2 X or ½ mAs
![Page 7: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/7.jpg)
BEWAREBEWARE
The “Step in Time”
![Page 8: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/8.jpg)
DensityDensityKilovoltageKilovoltage
15% increase in kVp doubles the exposure to the film
15% decrease in kVp halves the exposure to the film– In the lower kVp range (30-50), a change of 4-5 % in kVp may be detectable; in the middle
range (50-90), a change of 8-9% may be necessary; in the higher range (90-130), 10-12% is necessary to detect change in density.
![Page 9: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/9.jpg)
Density Density kVp – 15% RulekVp – 15% Rule
May vary up to 25% rule at high kVp
15% change in kVp will always change
Contrast!
![Page 10: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/10.jpg)
DensityDensityInfluencing FactorsInfluencing Factors
Focal Spot Size – Properly calibrated equipment should not
exhibit change in density with a change in focal spot size.
![Page 11: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/11.jpg)
DensityDensityInfluencing FactorsInfluencing Factors
Anode Heel Effect– The x-ray beam intensity may vary up to 45% from the
cathode to anode end of the beam– More noticeable
With small angle anodes With the collimator open wide
– Advantage Place thicker or denser body part under cathode end of tube
![Page 12: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/12.jpg)
DensityDensityAnode Heel EffectAnode Heel Effect
Examples– Femur– Lower leg– Humerus– Forearm– Thoracic Spine– Lumbar spine
![Page 13: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/13.jpg)
DensityDensityDistanceDistance
SID – Inverse Square Law– Intensity of radiation is inversely proportional
to the square of the distance from the source– Increase SID, decrease Density
I1 D 22
______ = ______
I 2 D 1 2
![Page 14: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/14.jpg)
Density MaintenanceDensity MaintenanceFormula Formula
mAs1 D12
_____ = _____
mAs2 D22
![Page 15: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/15.jpg)
DensityDensity
Filtration – Increase in added filtration decreases density
![Page 16: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/16.jpg)
DensityDensityBeam RestrictionBeam Restriction
– Reduction of beam size reduces the production of scattered radiation, therefore reducing the amount of radiation reaching the film and reducing density.
– Dependent upon: Amount of scatter produced Efficiency of grid Amount of reduction
![Page 17: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/17.jpg)
DensityDensityBeam RestrictionBeam Restriction
Usually not necessary to compensate for a restriction in beam size because the change in density is
less than 30% and therefore not noticeable.Compensation may be required:
Large patientHigh Kilovoltage techniquelow grid efficiency (or no grid)
![Page 18: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/18.jpg)
DensityDensityAnatomical PartAnatomical Part
Greater tissue thickness, the less density. Radiopaque contrast media decreases density. Radiolucent contrast media increases density. Destructive pathology increases density. Additive pathology decreases density. Angled techniques decrease density
– Variable (depends on degree of angulation)
![Page 19: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/19.jpg)
DensityDensityAnatomical PartAnatomical Part
Angled techniques decrease density– Variable (depends on degree of angulation)
![Page 20: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/20.jpg)
DensityDensityGridsGrids
Grids absorb scattered radiation, therefore decreasing density.
Loss of density compensated for by increasing mAs.
Primary purpose of grid is to increase contrast
![Page 21: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/21.jpg)
DensityDensityGridsGrids
Grid ConversionsNo grid to:GFC
– 5:1 2 MAS1 GCF1
– 6:1 3 -------- = -------
– 8:1 4 MAS2 GCF2
– 12:1 5– 16:1 6
![Page 22: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/22.jpg)
DensityDensityFilm/Screen CombinationFilm/Screen Combination
As intensifying screen speed increases, density increases; to compensate mAs is decreased.
MAS1 RS2
----------- = -----------
MAS2 RS1
MAS2
![Page 23: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/23.jpg)
DensityDensity
Film Processing should not affect density as long as processing parameters are maintained.
![Page 24: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/24.jpg)
DensityDensity
Increase mAs + Decrease mAs -
Increase kVp + Decrease kVp -
Increase generator + Decrease generator –
Focal Spot Size 0 Anode Heel Effect 0
Increase SID - Decrease SID +
Increase filtration - Decrease filtration +
Inc. Beam restrict - Dec beam restrict +
Inc. part thickness - Dec. part thickness-
![Page 25: Analyzing the Image Density. Density Overall blackening of the image.](https://reader030.fdocuments.in/reader030/viewer/2022032517/56649cbf5503460f949844b8/html5/thumbnails/25.jpg)
DensityDensity
Radiolucent CM + Radiopaque CM -
Additive pathology -Destructive pathology +
Increase grid ratio - Decrease grid ratio +
Increase F/S RS + Decrease F/S RD -
Increase Processing + Decrease Processing -
Increase CR angle - Decrease CR angle +