Analytical Study in Kinematic of the Knee

download Analytical Study in Kinematic of the Knee

of 8

Transcript of Analytical Study in Kinematic of the Knee

  • 7/25/2019 Analytical Study in Kinematic of the Knee

    1/8

    ELSEVIER

    Med. Erg. Phys.

    Vol. 19. No. 1. 29-36.

    p. 1997

    Copyright 0 1997 Elsetier Scienc e Ltd for IPEMB. All rights resewed

    Printed in Great Britain

    PII: S13504533(96)00031-8

    1350-4533/97 $17.00 + 0.00

    Analytical study on the kinematic and dynamic

    behaviors of a knee joint

    Zhi-Kui Ling Hu-tJrng Guo and Stacey Boersma

    Department of Mechanical Engineering and Engineering Mechanics, Michigan

    Technological University, Houghton, MI 49931, USA

    Received 1 September 1995, accepted 10 May 1996

    ABSTFUCT

    A knee model in the sag&al plane is established in this Jtudy. Specafically, the model is used to study the e&s of

    inertia, articular surfaces of the knee joint, and patella on the behaviors of a knee joint. These behaviors includt the

    joint surfnce contact point, ligament f&-es, instantaneous center and slide/roll ratio between the femur and tibia.

    Model results are compawd to experimental cadaver studie.5 available in the literature, as well as between the quasi-

    .statir and dynamic mod&. We found that inertia increases the sliding ten&ncy in the latter part of pexion,, and

    lengthens the rruciate ligaments. Decreasing the curvature of the femur surface geometry tends to reduce the ligament

    forcps and moves the contact points towards the anterior positions. The introduction of the patellar ligament in the

    model seems to stabilize the behaviors of the knee joint as reflected by the behavior

    qf

    the instant centers and the

    contact po int pattern on the tibia surface. Furthermore, we ,found that diff,

    Prent magnitudes of the external load

    applied to the tibia do not alter the qualitative behaviors of the knee joint. 0 1997 Elsevim Science for IPEfiiB. ,411

    right,\ re.seroed.

    Keywords: Knee kinematics, knee dynamics, knee modeling, geometric modeling

    Med. Eng. Phys., 1997, Vol. 19, ?9-36,January

    1. INTRODUCTION

    A well-defined analytical knee model can be an

    effective tool for understanding the functionality

    of the largest musculoskeletal joint in the human

    body. Statistics show that over two million cases of-

    knee injury occur in the United States each year.

    This model can provide a scientific explanation as

    to the causes of these injuries. Therefore, preven-

    tive measure can be taken to avoid them. Further-

    more, a well-developed analytical model could

    also be used efficiently to determine the effects of

    system variables on the performance of the knee

    joint, and to guide experimental and clinical

    investigations. However, a comprehensive knee

    model does not exist in the literature.

    Analytical knee models have generally adapted

    a four bar linkage methodology, by grounding

    either the tibia or femurlm4. In these models, the

    two cruciates are assumed rigid links with neutral

    ligament fibers staying constant lengths during

    flexion or extension. Furthermore, the articular

    surfaces o f the femur and tibia are either simpli-

    fied, or their effects are ignored completely.

    Although these models have provided initial

    understanding of the knee kinematics, they can-

    Correspondence to: Zhi-Kui Ling.

    not accurately portray the actual kinematic and

    dynamic characteristics of a knee.

    Other studies -

    8 have adapted a quasi-static

    approach towards the modeling of a knee.

    Although the quasi-static models compensate for

    the deficiencies of the four bar linkage model by

    allowing the cruciate ligaments to change their

    lengths, they still cannot take into account the

    roles of inertia and other ligaments in the

    behavior of a knee.

    To consider the effects of inertia, three studies

    have attempted to establish the dynamic model

    of a knee. The f irst9 proposed a two-dimensional

    dynamic model of a tibiofemoral joint. The model

    was used to study the contact conditions between

    the femur and tibia as well as the characteristics

    of the ligament forces. The second study investi-

    gated the role of ligaments and muscles as control

    elements for a prespecified rolling and slidin

    pattern. Recently, Abdel-Rahman and Hefzy

    I?

    presented a dynamic model which incorporates

    additional ligament constraints between the

    femur and tibia to the model by Moeinzadeh in

    describing the tibiofemoral joint. However, the

    entire articular surface of the femur was assumed

    to be a circular arc. The contact point positions

    and forces between the femur and tibia and the

    ligament forces were studied. The existing

    dynamic models do provide further understand-

  • 7/25/2019 Analytical Study in Kinematic of the Knee

    2/8

    Analytical study on the behaviors of a knee joint: Zhi-Kui Ling et al.

    ings of a knee joint. However, deficiencies still

    exist. Specifically, the role of the patella in the

    behavior of a knee joint is usually considered sep-

    arately 12*13 Furthermore, the techniques for calcu-

    lating the loci of instant centers and the roll/slide

    ratio during knee motion are absent in the exist-

    ing models. However, these are important factors

    in studying the kinematics of a knee. Further-

    more, the effects of the joint surface geometry on

    the behaviors of the knee have not been carried

    out in the literature quantitatively. Finally, the

    role of inertia in the overall behavior of the knee

    is still not quite clear.

    The aim of this study is to address the above-

    mentioned three issues in the existing twodimen-

    sional analytical knee models. Specifically, a two

    dimensional model shown in Figure I is intro-

    duced. Formulation for both the quasi-static and

    dynamic models of a knee is carried out. The

    implicit Euler and Newton-Raphsons numerical

    schemes are used to solve the established equa-

    tions of motion and nonlinear equations.

    Methods to determine the loci of the instant cen-

    ters and roll/slide ratio are introduced. In

    addition, the effect of the different articular sur-

    face geometry of the femur on the behavior of the

    knee is also investigated.

    Results of the quasi-static model provide the

    initial values for the simulation of the dynamic

    model. Although no experiments are performed

    in this study, results from the analytical model are

    compared with the available experimental data in

    the literature. These results include the ligament

    forces, the contact conditions, the loci of the

    instant centers, and the rolling/sliding behavior

    between the femur and tibia.

    The remainder of this paper is organized into

    four sections. First of all, the analytical model and

    its numerical solution are introduced. Secondly,

    methodologies used to determine the iristant cen-

    ters and slide/roll ratio are described. Results and

    discussions are presented next, followed by a con-

    clusion.

    N = Normal Force

    I I

    Fl = LCL Force

    I -t

    Fixed Femur Fz = McL Force

    F3 = ACL (mhrior) Force

    F4 = PCL Qmaior) Force

    FS = ACUposterlar) Force

    F6= PCL (~tcrtor) Force

    F7 = htdlr L@nen t Force

    Pa:

    Fext = Externd Force

    Mext = Extend Moment

    \ i $- Moving Tibia

    2. ANALYTICAL MODELS

    Because the fibula does not make contact at the

    articulating surface of the tibiofemoral joint, its

    effects are ignored in this study. The contours of

    both the femur and the tibia in the sagittal plane

    are acquired using the radiographs of a left, unat-

    tached leg of a 6%year-old female cadaver, with

    carcinoma reported as the cause of death. The

    radiograph conditions include an unloaded leg,

    horizontally positioned with the lateral side down.

    Reference axes for the model are set up with

    the Y axes centred along the bones longitudinal

    axes, pointed towards the knee joint contact sur-

    faces, as shown in Figure 1. In this study, the pro-

    file of the femur is described with two segments,

    as shown in equations (1) and (2). This reflects

    the actual shape of the femur14. A second-order

    polynomial in equation (3) is also generated to

    describe the two-dimensional profile of the tibia

    in the sa ttal plane. The maximum fit errors of

    1.8 x10-

    P

    and 4.497 ~10~~ cm are found for the

    two contours of the femur and the tibia, respect-

    ively. The three profiles are as follows:

    ~~(x_=04b0110~~~4637x-0.13492 - 0.0332

    (1)

    fib(x) = 2.733 + &8144-(x + 2.692)2

    (2)

    f&d) = 21.34-0.2578x + 0.0477~ 2

    (3)

    Five major ligaments are represented in the

    model. They are the medial collateral (MCL) , lat-

    eral collateral (LCL) , anterior cruciate (ACL) , the

    posterior cruciate (PCL), and the patellar liga-

    ment. Both the ACL and the PCL are represented

    by their anterior and posterior bundles. Their two-

    dimensional insertion and origin points are

    obtained from the literature5, and listed in Table

    1. These numbers have been adjusted to the coor-

    dinate systems discussed before.

    In the following discussion, formulation of the

    analytical model is divided into two cases, the

    quasi-static model and the dynamic model. Two

    constraint equations exist for both models. First,

    the tibia surface must be in contact with the femur

    surface at one point.

    Where, x, y0 is the origin location of the tibia

    with respect to the femur, fmxc and fmyc, tibxc

    and tibyc are the femur and tibia contact points,

    respectively.

    The second constraint requires colinearity of

    Table 1 Ligament insertion and origin coordinates (cm)

    Ligament

    Tibia X Tibia Y Femur X Femur Y

    LCL 3.849 17.579 -2.5 1.9

    MCL 2.149 16.079 -2.3 1.4

    ACL (anterior) 0.849 21.079 -2.3 1.9

    ACL (posterior) 1.149 21.079 -1.9 1.9

    PCL (posterior) 3.849 20.579 -3.2 2.4

    PCL (anterior) 3.849 20.579 -1.2 2.4

    Fii 1 A knee model in its sagittal plane

    30

  • 7/25/2019 Analytical Study in Kinematic of the Knee

    3/8

    Analytical study on t/w behaviors of a knee joint: Zhi-Kui Ling et al.

    the knee capsular provides negligible resistance,

    a zero coefficient of friction is assumed, resulting

    in only a normal force component.

    The tibia is governed by three equations of

    motion, as shown below. In this study, the tibia

    mass is estimated at 3.45 kg, the mass for a 50

    percentile malel, and the mass moment of inertia

    is set to be 392.8 kg cm2.

    Table 2 Ligament Stiffnesses

    Ligament

    k, (kg cm- SK*)

    LCL

    150 000

    MCL 150 000

    ACL (anterior) 200 000

    ACL (posterior) 100 000

    PCL (anterior) 175 000

    PCL (posterior) 175 000

    the unit normals at the respective contact points

    on both the femur and tibia. This can be rep-

    resented by the zero cross product of the two nor-

    mals. The simplified form of this condition is

    shown below.

    (5)

    In this study, the nonlinearity of ligament forces

    for the cruciate and collateral ligaments is mod-

    eled with the following expression.

    muss

    = (F,.), + W + A(nm) (%)y

    R

    kj( how,- Istart,) *

    smagi= o

    if howi &art,)

    if< lnOWj5 hart,)

    (6)

    X[fmq$+ fma&l +

    Here, j is an index number representing differ-

    ent ligament. The stiffness of the collateral and

    cruciate ligaments, kj, are taken from the litera-

    ture.16, and are listed in Table 2, and lnow is the

    current length of a ligament.

    In equation (6), hart, the taut length of the

    ligament j, is calculated by multiplying the initial

    length at full extension of the ligament by its

    strain ratio5. This is shown in equation ( { ), and

    the strain ratios of the corresponding ligaments

    are shown in Table 3.

    Istart, = linit ial?q

    (7)

    In this study, the patellar ligament is assumed

    to be in the sagittal plane during flexionl. The

    magnitude of the pate110 ligament force is

    obtained through the ratio between patellar liga-

    ment force and quadriceps force versus flexion

    angle . The insertion point of the patellar liga-

    ment on the tibia is specified at (-0.251, 17.779)

    with respect to the tibia coordinate system. The

    angular orientation of the patellar ligament is ref-

    erenced from the literature*.

    Besides ligament forces, a force at the contact

    point exists. Because the synovial fluid present in

    Table 3 Ligament strain ratio

    Ligament

    Strain ratio

    LCL 1.02

    MCL

    1.02

    ACL (anterior) 1.05

    ACL (posterior)

    1.035

    PCL (anterior) 1.05

    PCL (posterior) 1.05

    (9)

    where Norm is the magnitude of the normal con-

    tact force, and A can be either positive or negative

    depending on the curvature.

    In this study, the quasi-static model constitutes

    equations (l-7), and equations (8-10) with the

    left hand sides equal to zero. The dynamic model

    consists of equations (l-10). For the quasi-static

    model, Newton-Raphsons method is used to solve

    for the six nonlinear equations, i.e. equations (S-

    10) with zero accelerations, and equations (4) and

    (5). The independent variable in these six equa-

    tions is the flexion angle, and the six unknowns

    are the tibia mass center (3~0, y,J, contact point

    with respect to the femur and tibia (femxc, tibxc),

    the normal force (nomn), and the required exter-

    nal moment (K,,). The solutions starting from

    0 with 10 increment up to 90 are calculated.

    The solution at each of these positions is found,

    if the tolerance is smaller than 0.0001.

    For the dynamic model, there are also six equa-

    tions: the three equations of motion, equations

    (S-10)) and three algebraic constraints, equations

    (4) and (5). Smce the six equations are a mix of

    nonlinear and differential equations, this study

    uses both the implicit Euler and Newton-

    Raphsons in simulating the dynamic model. The

    advantage of implicit Euler is its stability, however,

    the method only provides a first order accuracy.

    Iterations are performed at each time step until

    convergence occurs. In this study, the time

    increment for the implicit Euler method is set at

    0.0001 s. The solution for each time step is achi-

    eved if the tolerance of 0.0001 is achieved.

    31

  • 7/25/2019 Analytical Study in Kinematic of the Knee

    4/8

    Analytical study on the behaviors of a knee joint: Zki-Kui Ling et al.

    2.1. Instant centers and slide/roll ratio

    The instantaneous centers of rotation for both the

    quasi-static and dynamic models are found by

    using the instantaneous velocities of two points on

    the moving tibia. They are the tibia contact point

    and the tibia mass center. For the quasi-static

    model, the line representing the velocity at the

    tibia contact point is approximated by the line

    connection between the contact point, and the

    same contact point at its next position. The same

    technique is used to determine the velocity line

    at the mass center of the tibia. The instant center

    is then determined by constructing two perpen-

    dicular lines to the two velocity lines at the contact

    point and the mass center. The intersection point

    of the two perpendicular lines is the instant center

    location at that particular instance.

    For the dynamic model, the velocity at the mass

    center of the tibia is known from model simul-

    ation. However, the velocity at the contact point

    is unknown. In this study, the x and y components

    of the contact point velocity,

    (dXtitml/&

    df2( tibxc,)/dt), are found by using the contact

    point and the same point at the next time step.

    A first order forward difference approximation

    scheme is used in the determination of the con-

    tact point velocity. The same two perpendicular

    lines to the velocities at the mass center of the

    tibia and the contact point are constructed to

    determine the corresponding instant centers.

    To calculate the slide/roll ratio, the arc lengths

    travelled on the surfaces of the tibia and femur

    between two consecutive simulation times are

    determined with the following numerical inte-

    gration.

    (11)

    For either contour surface (fi or f2), the lower and

    upper limits are the X components of two adjac-

    ent contact points. The slide/roll ratio is defined

    as the difference. between the larger distance (D)

    and the smaller distance (d) travelled on the

    femur and tibia over the smaller of the two arc

    lengths travelled (d) .

    3. RESULTS AND DISCUSSIONS

    The already-established analytical models are used

    to provide a comparison between the kinematic

    and dynamic results in terms of the following

    characteristics: the contact points on the femur

    and tibia; the ligament forces; the instant center

    locations; the slide/roll ratio, all with respect to

    the flexion angle.

    The effects of the articular surface geometry on

    the dynamic behavior of a knee are also investi-

    gated. This is accomplished with the reduction of

    the curvatures of the femur surface. Specifically,

    the coefficient for the linear term, 0.4637, of the

    first femur profile in equation (1) is reduced to

    0.4137 and 0.3637, respectively. The second pro-

    file of the femur is also changed. In this case, the

    Profun --

    Pnnnoz.---

    P lww3 - . .

    KiMllUllO.0

    I

    x

    -3.5 -

    -4-

    Figure 2 Femur contact points

    radius of the circular arc in equation (2) is

    reduced by 0.1 cm and 0.22 cm, respectively. The

    original and the two new profiles of the femur are

    identified as profile 1, 2 and 3 hereafter. Finally,

    the effect of the patella on the knee behavior is

    also studied with the model.

    A constant impulse force with a magnitude of

    20 N is applied along the x axis of the tibia with

    a duration of 0.1 ..s in the analytical model.

    Although the effect of different external loads on

    the behavior of the knee is not the focus of this

    study, a qualitative study is performed. In the

    remainder of this section, the results of the afore-

    mentioned studies are presented.

    The contact point with respect to the femur and

    tibia travels posteriorly with flexion as shown in

    Figures 2 and 3. This is in aFeement with the

    results reported by othersgs, . In Figure 2 it can

    be observed that as the curvature of the femur

    surface profile becomes smaller, contact points

    shift towards the anterior direction. The inertia

    has a greater impact on the contact point

    behavior towards the latter part of the knee

    flexion. From Figure 2, it is also shown that the

    transition from the firs t to the second profile of

    the femur is not perfectly smooth. This is due to

    the fact that the slope at the connection point of

    0

    0 10 M 30 40 50 60 70 30 so

    Flaxion Dqrn

    Figure 3 Tibia contact points

  • 7/25/2019 Analytical Study in Kinematic of the Knee

    5/8

    the two profiles is not completely continuous in

    the modeling of the femur contour. The rest of

    the results are also affected by this phenomenon.

    In Figure 3, the contact point with respect to the

    tibia travels posteriorly as well with flexion. How-

    ever, the contact point moves in a faster rate

    towards the posterior direction for the dynamic

    model than for the kinematic model as flexion

    exceeds 40. The effect of the patellar ligament

    on the contact point pattern of both the femur

    and tibia is barely noticeable.

    Ligament forces in both LCL and MCL exhibit

    maximum magnitudes at the full extension pos-

    ition. As flexion starts, the ligament forces start to

    decrease and are faded to zero before the full

    flexion of 90 is reached. It is found that the effect

    of inertia on the collateral ligaments is not appar-

    ent. Furthermore, the ligament forces abate as the

    curvature of the femur surface decreases. Intro-

    duction of the patella in this study does not make

    any difference in the behavior of the MCL, but

    stretches the LCL ligaments. This is shown in Fig-

    ure 4.

    It may not be appropriate to compare the

    results of the ligament forces from the analytical

    models with those of experiments quantitatively,

    since different boundary conditions were used in

    those experiments and the analytical studies. Fur-

    thermore, different opinions exist with regard to

    the functions of ligaments20-24. Finally, most of the

    experimental studies are quasi-static in nature.

    Therefore, the discussion here on the ligament

    forces is confined to qualitative comparison. For

    the collateral ligaments, the results from this study

    using both the quasi-static and dynamic models

    qualitatively agree with those found in the exyeri-

    mental studies. The experimental results21.2 ,24.L5

    indicated that from full extension, the collateral

    ligament forces have the maximum values, and

    decrease as the knee flexes.

    While the ligament force of the anterior PCL

    increases and then decreases with respect to the

    flexion angle, the force in the posterior PCL

    decreases very rapidly, as shown in Figure 5. Fur-

    thermore, while the force in the anterior fiber of

    the PCL from the dynamic model is larger than

    300

    I

    h

    250 \

    \

    MCL with Pateliar l igament = -

    I

    I\

    MCL without PaWar l igament =.-.-.-

    200

    LCL wkh Paldlar liint =

    (

    \

    LCL withoul Patek tiit = +

    f 1

    T+

    0 +++&

    0

    10 20 30 40 50 M) 70 80 90

    Ftexion Degree

    Figure4 h~fluence of the patellar ligament over the forces in the

    collateral ligaments

    Analytical study on the behaviors of a knee joint: %hi-Kui Ling et al.

    Figure 5 The behaviours of the PCI. ligament force iu terms of in

    antrrior and posterior bundles

    that from the quasi-static model during the early

    part of flexion, the trend reverses after flexion

    angle passes 42.

    As the curvature of the femur declines, the liga-

    ment forces in both fibers of the PCL decrease.

    While the patellar ligament has no effect on the

    posterior fiber of the PCL, it stretches the anterior

    fiber of the PCL in the early part of the flexion,

    and provides relief for the fiber in the later part

    of the flexion. From the modeling standpoint, this

    can be explained with the fact that the posterior

    fiber is used to provide the moment to balance

    that produced by the patellar ligament in the early

    part of flexion.

    The analytical results of the PCL match with

    those from experiments22. The difference exists in

    the anterior fiber, where the maximum ligament

    force occurs during the early part of flexion in the

    analytical modeling. However, the overall trend of

    the ligament forces follows the experimental

    results. The effect of inertia on the posterior fiber

    of PCL is difficult to observe, since the ligament

    force becomes zero before 10 of flexion. The

    small value of posterior PCL force was also indi-

    cated by others 26,2. The anterior portion of PCL

    exhibits a decrease in the ligament force when the

    inertia force is considered in the later part of

    flexion. This is probably due to the fact that the

    inertia force acts along the same direction as the

    ligament force in the anterior PCL during flexion

    of the knee.

    While the ligament force of the posterior ACL

    increases with respect to the flexion angle, force

    in the anterior ACL decreases and then increases

    for the dynamic model. Yet, while the posterior

    ACL increases and then decreases with respect to

    the flexion angle, the anterior ACL decreases in

    the kinematic model. They are shown in Figure 6.

    The difference between the kinematic and

    dynamic model is due to the presence of the iner-

    tia, which changes the contact pattern on both

    the femur and tibia as discussed in the previous

    section. Consequently, the insertion point of the

    ACL on tibia changes its pattern of motion at the

    later stage of flexion, which causes itself to be

    stretched in the process. As the curvature of the

    33

  • 7/25/2019 Analytical Study in Kinematic of the Knee

    6/8

    Analytical

    on the behaviors

    o?a

    nee joint: Zhi-Kui Ling et al.

    350,

    250

    ACL(a) Profile 1

    ACL(a) P rofile 2

    ACL(a) P rofile 3

    ACL(a) Kinematic

    ACL(p) Profile 1

    ACL(p) Profile 2

    ACL(p) Profile 3

    .ACL(p) Kinematic

    it

    1 20 30 40 50 60 70 60 90

    Flexion Degree

    Figure 6 The behaviours of the ACL ligament force in terms of its anterior and posterior bundle

    femur decreases, the ligament forces in both fib-

    ers of the ACL decrease.

    While the patellar ligament has little effect on

    the pattern of the ligament forces in either fiber

    of the ACL, the forces in ACL increase as flexion

    increases. From the modeling standpoint, this can

    be explained with the fact that the ACL is used

    to provide the moment to balance that produced

    by the patellar ligament in the later part of

    flexion. For the anterior cruciate ligament, results

    from the analytical model without the patella do

    match with those from experiments, such as

    France et a~ . The ligament force in the anterior

    portion of the ACL decreases from 0 to 90 of

    flexion, while the posterior portion increases from

    0 to 50 and then decreases towards the 90

    flexion.

    The instantaneous centers obtained from the

    quasi-static model follow a circular path, begin-

    ning at around 20 of flexion anteriorly on the

    proximal femoral condyle, and ending at 90 pos-

    teriorly closer to the joint surface also on the

    proximal femoral condyle. When the inertia is

    considered, the instantaneous centers have the

    same pattern as demonstrated by the quasi-static

    model, nevertheless, they are located in the pos-

    terior side of the instant centers from the kinem-

    atic model. The effect of changing the surface

    geometry of the femur on the instant centers is

    minimal. However, the instant centers shift

    anteriorly as the curvature of the femur decreases.

    Patellar ligament constraint makes a big differ-

    ence in the locus of the instant centers for the

    dynamic model. Figure 7 demonstrates this differ-

    ence where the model without the patella locates

    the instant centers in the tibia side of the joint at

    the higher degrees of flexion, while the instant

    centers of the model with the patella are located

    on the femur side.

    The loci of the instant centers from experi-

    mental studies are only available with the quasi-

    static approach. These instant centers were found

    using X-rays at incremental degrees of rotation**.

    Therefore, comparison between the results from

    34

    x am (a)

    Figure 7 Influence of the patellar ligament over the loci of the

    instant centres

    this study and the experimental results may not

    be appropriate, as the methods used to determine

    each individual instant center are different. How-

    ever, for the quasi-static model, the analytical

    results display the trend of the instant center locus

    which is similar to the experimental results.

    It can be concluded that rolling is dominant at

    the beginning of flexion, and sliding becomes the

    dominant factor as the flexion increases. There is

    very little difference between the slide/roll ratio

    of the quasi-static and the dynamic model during

    flexion from 0 to 60. However, the ratio dips

    lower for the quasi-static model when the flexion

    angle exceeds 60. The change of femur curvature

    has very little effect on the slide/roll ratio. Fur-

    thermore, the patellar ligament facilitates the

    increase of rolling in the latter part of knee

    flexion, as illustrated in Figure 8. Although there

    are no experimental results available, the

    slide/roll ratio obtained from this study matches

    with the consensus as related to the

    ing versus rolling in the literature

    r

    attern of slid-

    .

    Although no graphs are presented in this paper

    to illustrate the behaviors of the knee joint under

  • 7/25/2019 Analytical Study in Kinematic of the Knee

    7/8

    5

    with patekr l igament = -

    4.5 -

    without pstellar lQa"Mt = - - -

    /

    4-

    , -

    /

    3.5

    -

    , -

    1-

    Slii

  • 7/25/2019 Analytical Study in Kinematic of the Knee

    8/8

    Analytical study on the behaviors of a knee joint: Zhi-Kui Ling et al.

    ments. Journal of Bone and Joint Surgery, 1976, 58(A),

    350-355.

    17. Reilly, D. T. and Martens, M., Experimental analysis of

    the quadriceps muscle force and patello-femoral joint

    reaction force for various activities. Acta &hop. &and.,

    1972, 43, 126-137.

    18. Webb Associates: Anthropometric Source Book, ed. Anthro-

    pology Research Project, NASA, Scientific and Technical

    Information Offi ce, Washington, 1978.

    19. Draganich, L. F., Andriacche, T. P. and Anderson, G. B.

    J., Interaction between intrinsic knee mechanics and the

    knee extensor mechanism. Jarnal of Orthopaedic Research,

    1987, 5, 539-547.

    20. Edwards, R. G., Laf fer ty, J. F. and Lange, K O., Ligament

    strain in the human knee joint. ASME Journal of Basic

    En@ne&ng, 1970, 92, 131-136.

    21. Wang, Ching-Jen and Walker, P.S., The ef fec ts of flexion

    and rotation on the length patterns of the ligaments and

    the knee. Journal of Biomechanics, 1973, 6 , 587-596.

    22. France, P. E., Daniels, A. U., Goble, E. M. and Dunn, H.

    K, Simultaneous quantization of knee ligament forces.

    Journal of Biomechanics, 1983, 16, 553-564.

    23. Ahmed, A. M., Hyder, A., Burke, D. L. and Chan, K. H.,

    In-vitro ligament tension pattern in the flexed knee in

    passive loading. J.

    Orthop. Res.,

    1987, 5, 217-230.

    24. Blankevoort, L., Huiskes, R. and De Lange, A., Recruit-

    ment of knee joint ligaments. Transactions of the

    ASME,

    1991, 113, 94-103.

    25. Huiskes, R., Blankevoort, L., van Dijk, R. , de Lange, A.

    and van Rens, Th. J. G., Ligament deformation patterns

    in passive knee-joint motions. In Proceedings of ASME Win-

    ter Annual Meeting: Bioenginee ring Symposium, 1984.

    26. Girgis, F. G., Marshall, J. L. and Monajem, A. R. S. A l.,

    The cruciate ligaments of the knee joint. Clinical Ortho-

    paedics and Related Research,

    1975, 106, 216-231.

    27. Markolf, K L., Mensch, J. S. and Amstutz, H. C., Stiffness

    and laxity of the knee - the contributions of supporting

    structures. J. Bone Joint Surgery, 1976, 58(A), 583-593.

    28. Soudan, K. and Van Audekercke, R., Methods, difficu lties

    and inaccuracies in the study of human joint kinematics

    and pathokinematics by the instant axis concept.

    Example, the knee joint. Journal of Biomechanics, 1979, 12,

    27-33.

    29. Steindler, A,,

    Kinesiology

    of

    the Human Body Under Normal

    and Patholog ical Conditions. Charles C. Thomas, Spring-

    field, Illinois, 1973.

    30. Attf ield, S. F., Warren-Forward, M., Wilton, T. and Sam-

    batakakis, A., Measurement of sof t tissue imbalance in

    total knee arthroplasty using electronic instrumentation.

    Medical Engineering & Physics,

    1994, 16, 501-505.

    36