Analysis of Silicone Oils by High Performance Liquid ... · Analysis of Silicone Oils by High...

7
Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection Marc Plante, Bruce Bailey, Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA

Transcript of Analysis of Silicone Oils by High Performance Liquid ... · Analysis of Silicone Oils by High...

Page 1: Analysis of Silicone Oils by High Performance Liquid ... · Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection Marc Plante, Bruce

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol DetectionMarc Plante, Bruce Bailey, Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA

Page 2: Analysis of Silicone Oils by High Performance Liquid ... · Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection Marc Plante, Bruce

2 Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol DetectionMarc Plante, Bruce Bailey, Ian N. AcworthThermo Fisher Scientific, Chelmsford, MA, USA

OverviewPurpose: To develop HPLC methods for both the detailed, characterization of silicone oils and for rapid quantitation of silicone oils in samples using an HPLC system with a charged aerosol detector.

Methods: Two methods, one qualitative for characterization and one quantitative using a solid core C18 column, are outlined.

Results: Different silicone oils were characterized using the qualitative method; silicone oils in commercial products were measured, with a quantitation limit of approximately 100 ng on column.

IntroductionSilicone oils are unique materials with multitudes of significant uses in our modern world. Silicone-based materials are thermally stable, relatively inert, and are generally non-toxic, making them useful as oils and greases, lubricants, anti-foaming agents, and coatings. Silicone oils are often used in electronics, hydraulic systems, and many consumer products. The measurement of silicone oils is complicated by their solubility properties, their variations in basic structure including a wide variety of chemical moieties that may be part of this structure, as well as their general lack of chromophores making detection by ultraviolet absorption impractical.

The charged aerosol detector is a sensitive, mass-based detector, especially well-suited for the determination of any nonvolatile analyte independent of chemical characteristics. As shown in Figure 1, the detector uses nebulization to create aerosol droplets. The mobile phase evaporates in the drying tube, leaving analyte particles, which become charged in the mixing chamber. The charge is then measured by a highly sensitive electrometer, providing reproducible, nanogram-level sensitivity. This technology has greater sensitivity and precision than ELSD and refractive index (RI), is gradient compatible and is simpler to operate than a mass spectrometer (MS). Compounds do not have to possess a chromophore (unlike UV detection) or be ionized (as with MS).

Two charged aerosol detection methods were developed: one to characterize the individual components of silicone oils, and the other to provide quantitation of a near single peak of silicone oil to 100ng (o.c.). The characterization method used the Thermo Scientific™ Dionex™ UltiMate™ 3000 RS system, and the quantitation method used an UltiMate 3000 SD system in normal phase. Both methods used a solid core C18 column and the Thermo Scientific™ Dionex™ Corona™ ultra RS™ charged aerosol detector.

The combination of these chromatographic methods with the capabilities of the charged aerosol detection provides unique opportunities for detailed characterization of a silicone oil sample, as well as quantitation of silicone oil-containing samples.

.

FIGURE 1. Schematic and Functioning of Charged Aerosol Detection

1

2

3

4

5

6

7

89

101

2

3

4

5

6

7

89

10

1. Liquid eluent enters from HPLC system2. Pneumatic nebulization occurs3. Small droplets enter drying tube4. Large droplets exit to drain5. Dried particles enter mixing chamber6. Gas stream passes over corona needle7. Charged gas collides with particles and

charge is transferred8. High mobility species are removed9. Charge is measured by a highly sensitive

electrometer10.Signal transferred to chromatographic

softwareMethodsSample Preparation - CharacterizationSamples of silicone oil were dissolved in chloroform, at a concentration of approximately 40 mg/mL.

Liquid Chromatography – CharacterizationHPLC System: Thermo Scientific™ Dionex™ UltiMate™ 3000

DGP-3600RS pump, WPS-3000RS autosampler, and TCC-3000RS column oven

HPLC Column: Thermo Scientific™ Accucore™ 2.6 µm C18,3.0 × 150 mm

Column Temp.: 40 °CMobile Phase A: MethanolMobile Phase B: n-PropanolFlow Rate: 0.5 mL/minInjection Volume: 2–10 µLSample: 38 mg/mL silicone oil in chloroformDetector: Corona ultra RSNebulizer Temp.: 10 °CFilter: 5Data Rate: 10 HzPower Function: 1.00Flow Gradient for Characterization:

Time (min) %A %B Curve

-10 100 10 5

0 100 10 5

80 15 85 3

85 0 100 5

95 0 100 5

95 100 0 5

100 100 0 5

Liquid Chromatography –QuantitationHPLC System: Thermo Scientific™ Dionex™ UltiMate™ 3000

LPG-3400SD pump, WPS-3000RS autosampler, and TCC-3000RS column oven

HPLC Column: Accucore 2.6 µm C18, 3.0 × 150 mmColumn Temp.: 40 °CMobile Phase A: 0.5% Formic acid/acetonitrile/tetrahydrofuran*

(35:35:30), helium spargeMobile Phase B: Tetrahydrofuran*, helium spargeFlow Rate: 0.3–1.0 mL/minInjection Volume: 2–10 µLDetector: Corona ultra RSNebulizer Temp.: 10 °CFilter: 6Data Rate: 10 HzPower Function: 2Flow Gradient for Quantitation:

*No BHT added.

Data AnalysisAll HPLC chromatograms were obtained and compiled using Thermo Scientific™ Dionex™ Chromeleon™ Chromatography Data Station software, 7.1 SR 1.

Time (min) Flow Rate(mL/min)

%A %B

-5.0 0.6 100 0

-0.2 0.6 100 0

-0.1 0.3 100 0

0.0 0.3 100 0

0.5 0.6 100 0

Time (min) Flow Rate(mL/min)

%A %B

4.0 0.6 100 0

6.0 0.6 0 100

8.0 0.6 0 100

10.0 1.0 0 100

12.0 1.0 50 50

14.0 0.6 100 100

ResultsSample AnalysisTo investigate relative responses of different oils, three differentsilicone oil samples were characterized consisting of two similar oilswith different viscosities (200 cP and 1000 cP) and a third oil that isused as a heating bath oil. From the overlay of chromatograms shown in Figure 2, the 1000 cP oil sample is comprised of a greater amount oflarger molecular weight polysiloxanes (later elution) than the 200 cP oil sample, as characterized by the bulk of peak area at higher retention times. The heating bath oil is comprised of more isolable polysiloxanechain groups, as identified by over 170 different peaks that were atleast partially resolved for this oil.

Retention time precision was demonstrated by repetitive analysis of the heating bath silicone oil sample (Figure 3). The UltiMate 3000 LCsystems, controlled by the Chromeleon software, enables sample injections to be synchronized with the pump delivery, allowing for exactand matching system conditions for each injection. The percent relative standard deviation for retention time at 50 minutes was 0.02%.

A silicone-based topical product used for comfort with hearing aids wascharacterized using this method, as shown by the chromatogram in Figure 4. The material appears to contain a moderate amount ofsmaller molecular weight polysiloxanes mixed with an amount of highermolecular weight polysiloxanes that elute at the end of the gradient.

The method provides a high-resolution HPLC chromatogram for the detailed characterization of polysiloxane materials. The use ofthe solid core column in this method achieves the high number oftheoretical plates and therefore, the high level of resolution shown in these figures. The charged aerosol detector provides both sensitive and consistent response for these analytes.

The near-consistent response of charged aerosol detection for non-volatile analytes enables the quantitation of total silicones in the same samples used above. The second method, specifically developed forquantitation, uses an aqueous acetonitrile mobile phase with a fasttetrahydrofuran gradient to elute the analyte from the column in a rapid manner, yielding a near-single peak. Chromatographic baseline subtraction is used to remove baseline deviations that result from the rapid gradient.

The 200 cP oil showed essentially equal response to that of the 1000 cPoil, but the heating bath oil differed in response at the two amounts used in this comparison. This may be associated with the differences of the groups or numbers of associated groups contained on the silicone. Thus,it is recommended that the silicone oil calibration standard is the same (orat least similar) to the silicone that is to be quantified. As seen in Figure 2,there is a clear difference between the 200 and 1000 cP silicones and the heating bath oil silicone.

FIGURE 5. Silicone oil calibration curves, from 78 to 10,000 ng (1000 cP)in triplicate, 2000 to 10,000 ng o.c. (200 cP and Heating Bath Oil)

0

2

4

6

8

10

12

0 2000 4000 6000 8000 10000 12000

Peak

Are

a (pA

*min)

Amount (ng o.c.)

AP1000

AP200

HeatingOil

TABLE 1. Response factors of different polysiloxanes at 5000 and2000 ng o.c. relative to 1000 cP silicone oil

Sample Amount on Column (ng)

Amount Found (ng) Relative Response

AP2005000 5031 1.0062000 1994 0.997

Heating Bath Oil5000 8681 1.7362000 4199 2.100

6.50 6.80 7.00 7.20 7.40 7.60 7.80 8.00 8.20 8.40 8.60 8.80 9.00 9.20 9.40 9.60 9.80 10.00-5.0

0.0

10.0

20.0

30.0

40.0

45.0 pA

min

21 1 - 6.980 2 - 7.4523 - 7.5224 - 7.665

5 - 7.954

6 - 8.0327 - 8.187

8 - SiliconeOil - 8.295

9 - 8.417

2 - SiliconeOil – 8.060

Two of the three oils (the two most closely related) showed a similarcalibration curve, indicating that calibration curves derived from oilsof a similar composition can be used for product concentration determinations. Differences in chain length and other structures mayaffect retention time. The heating bath oil, which characterized as a different oil, showed a greater response than the viscosity-rated oils.

FIGURE 6. HPLC chromatogram overlay of over-the-counter gasrelief product (blue), dissolved at a concentration of 500 µg/mLsimethicone in tetrahydrofuran, and 500 µg/mL of 1000 cP silicone oil standard (black)

*simethicone is a mixture of silica (5.5%) and dimethicone (94.5%)1

The measurement of silicone oils in different commercial products showed the potential for quantitative determinations without extraction of the silicone from the sample or having the exact silicone oil as a calibrant.Samples were dissolved, centrifuged, diluted to a targeted amount, and analyzed. A gas relieve product chromatogram is shown in Figure 6 with an overlay of 1000 cP silicone oil. The recovery for the gas relief product(prepared at 500 µg/mL of simethicone) was 96.5%, as shown in Table 2,demonstrating the potential for this method to measure silicone oils with unknown composition in different samples.

TABLE 2. Quantities of silicone oils found in products.

Sample Silicone Oil TypeDeterminedSilicone Oil

(ng o.c.)Product Silicone

AmountRecovery

(%)

Shampoo Dimethicone 3890 1.95 w/w-% --Conditioner Phenyltrimethicone 2702 2.70 w/w-% --Gas ReliefProduct

Simethicone* 4559 4725 ng o.c. 96.5

ConclusionsTwo methods were developed for the characterization and quantitation of silicone oils as standards and as part of products. The long characterization method can readily be modified to address specific regionsof the chromatogram or to fully characterize a specific oil in less time. The quantification method is sensitive to 100 ng o.c. and it was shown thatsimilar oils provide similar response factors. The method also had sufficientspecificity to allow for analysis of different products and silicone oils.

The use of the solid core C18 column, combined with the detection capabilities of the charged aerosol detector, enables convenient and accurate characterization and quantification of silicone oils and silicone oil-containing products.

References1. DailyMedPlus website

http://www.dailymedplus.com/monograph/view/setid/da8eba22-33e1-4f46-b43c-9ca3d8789740 (last accessed 30 Jan 13).

© 2013 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property ofThermo Fisher Scientific Inc. and its subsidiaries. This information is not intended toencourage use of these products in any manners that might infringe the intellectual propertyrights of others.

PO738_E 03/13S

Sample and Standard Preparations - QuantitationSamples containing silicone oil were dissolved in tetrahydrofuran, at a product concentration of 10—20 mg/mL and centrifuged at 10,000 g for3 minutes. HPLC sample vials were filled with the supernatant.

Standards (Silicone oil 1000 cP) were prepared by dissolving 10 mg/mL of silicone oil in tetrahydrofuran, diluted to 1 mg/mL and then sequentially diluted to lower concentrations.

FIGURE 2. Characterization of three different silicone oils, 200 cP(blue) and 1000 cP (black) viscosities, and a heating bathsilicone oil (pink)

FIGURE 3. Overlays of three separate analyses of the heatingbath silicone oil sample. Only the 50 to 55 minute segment of the total run (inset) is shown for clarity.

FIGURE 4. HPLC chromatogram of a silicone-based topicalproduct. Product appears to be made of high-molecular weightpolysiloxane (85 minutes) mixed with moderate and low-molecular weight polysiloxanes. The inset shows the low to moderate molecular weight regions.

0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100-10.0

0.0

10.0

20.0

30.0

40.0

50.0

60.0

70.0

80.0

min

pA

0.0 10.0 20.0 30.0 40.0 50.0 60.0 70.0 75.0-2.0

0.0

2.0

4.0

6.0

8.0

10.0

12.0

14.0

16.0

18.0

min

pA

This quantitative method was evaluated using different silicone oils; using the 1000 cP silicone oil standards, a calibration curve was generated from78 to 10,000 ng o.c., as shown in Figure 5. System precision (n=3) wasacceptable, with peak area percent relative standard deviations 0.92 (10,000 ng o.c.) to 7.28 (156 ng o.c.). The quantitation limit is estimated atapproximately 100 ng o.c, based on a signal to noise ratio of 3.0.

Samples that were tested included a 200 cP silicone oil, the heating bath oil, and three consumer products, including a shampoo, a hair conditioner,and a gas relief product containing different forms of silicones. The 200 cPoil and the heating bath oil were also used to generate single-injection calibration curves to investigate their response factors, relative to the 1000 cP oil standard. These results are in Table 1.

49.48 50.00 50.50 51.00 51.50 52.00 52.50 53.00 53.50 54.00 54.50 54.815.5

6.0

7.0

8.0

9.0

10.0

11.0

12.0

13.0

14.0

15.0

16.0

17.017.4

min

pA

Page 3: Analysis of Silicone Oils by High Performance Liquid ... · Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection Marc Plante, Bruce

3Thermo Scientific Poster Note • PN70538_e 06/13S

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol DetectionMarc Plante, Bruce Bailey, Ian N. AcworthThermo Fisher Scientific, Chelmsford, MA, USA

OverviewPurpose: To develop HPLC methods for both the detailed,characterization of silicone oils and for rapid quantitation of silicone oils in samples using an HPLC system with a charged aerosoldetector.

Methods: Two methods, one qualitative for characterization and one quantitative using a solid core C18 column, are outlined.

Results: Different silicone oils were characterized using the qualitative method; silicone oils in commercial products were measured, with a quantitation limit of approximately 100 ng on column.

IntroductionSilicone oils are unique materials with multitudes of significant uses in our modern world. Silicone-based materials are thermally stable,relatively inert, and are generally non-toxic, making them useful asoils and greases, lubricants, anti-foaming agents, and coatings.Silicone oils are often used in electronics, hydraulic systems, and many consumer products. The measurement of silicone oils iscomplicated by their solubility properties, their variations in basicstructure including a wide variety of chemical moieties that may be part of this structure, as well as their general lack of chromophoresmaking detection by ultraviolet absorption impractical.

The charged aerosol detector is a sensitive, mass-based detector,especially well-suited for the determination of any nonvolatile analyte independent of chemical characteristics. As shown in Figure 1, the detector uses nebulization to create aerosol droplets. The mobile phase evaporates in the drying tube, leaving analyte particles, which become charged in the mixing chamber. The charge is then measured by a highly sensitive electrometer, providing reproducible, nanogram-level sensitivity. This technology has greater sensitivity and precision than ELSD and refractive index (RI), is gradient compatible and issimpler to operate than a mass spectrometer (MS). Compounds do not have to possess a chromophore (unlike UV detection) or be ionized (as with MS).

Two charged aerosol detection methods were developed: one to characterize the individual components of silicone oils, and the otherto provide quantitation of a near single peak of silicone oil to 100ng (o.c.). The characterization method used the Thermo Scientific™Dionex™ UltiMate™ 3000 RS system, and the quantitation method used an UltiMate 3000 SD system in normal phase. Both methodsused a solid core C18 column and the Thermo Scientific™ Dionex™Corona™ ultra RS™ charged aerosol detector.

The combination of these chromatographic methods with the capabilities of the charged aerosol detection provides unique opportunities for detailed characterization of a silicone oil sample, aswell as quantitation of silicone oil-containing samples.

.

FIGURE 1. Schematic and Functioning of Charged AerosolDetection

1

2

3

4

5

6

7

89

101

2

3

4

5

6

7

89

10

1. Liquid eluent enters from HPLC system2. Pneumatic nebulization occurs3. Small droplets enter drying tube4. Large droplets exit to drain5. Dried particles enter mixing chamber6. Gas stream passes over corona needle7. Charged gas collides with particles and

charge is transferred8. High mobility species are removed9. Charge is measured by a highly sensitive

electrometer10.Signal transferred to chromatographic

softwareMethodsSample Preparation - CharacterizationSamples of silicone oil were dissolved in chloroform, at a concentration of approximately 40 mg/mL.

Liquid Chromatography – CharacterizationHPLC System: Thermo Scientific™ Dionex™ UltiMate™ 3000

DGP-3600RS pump, WPS-3000RS autosampler, and TCC-3000RS column oven

HPLC Column: Thermo Scientific™ Accucore™ 2.6 µm C18, 3.0 × 150 mm

Column Temp.: 40 °CMobile Phase A: MethanolMobile Phase B: n-PropanolFlow Rate: 0.5 mL/minInjection Volume: 2–10 µLSample: 38 mg/mL silicone oil in chloroformDetector: Corona ultra RSNebulizer Temp.: 10 °CFilter: 5Data Rate: 10 HzPower Function: 1.00Flow Gradient for Characterization:

Time (min) %A %B Curve

-10 100 10 5

0 100 10 5

80 15 85 3

85 0 100 5

95 0 100 5

95 100 0 5

100 100 0 5

Liquid Chromatography –QuantitationHPLC System: Thermo Scientific™ Dionex™ UltiMate™ 3000

LPG-3400SD pump, WPS-3000RS autosampler, and TCC-3000RS column oven

HPLC Column: Accucore 2.6 µm C18, 3.0 × 150 mmColumn Temp.: 40 °CMobile Phase A: 0.5% Formic acid/acetonitrile/tetrahydrofuran*

(35:35:30), helium spargeMobile Phase B: Tetrahydrofuran*, helium spargeFlow Rate: 0.3–1.0 mL/minInjection Volume: 2–10 µLDetector: Corona ultra RSNebulizer Temp.: 10 °CFilter: 6Data Rate: 10 HzPower Function: 2Flow Gradient for Quantitation:

*No BHT added.

Data AnalysisAll HPLC chromatograms were obtained and compiled using Thermo Scientific™ Dionex™ Chromeleon™ Chromatography Data Station software, 7.1 SR 1.

Time (min) Flow Rate(mL/min)

%A %B

-5.0 0.6 100 0

-0.2 0.6 100 0

-0.1 0.3 100 0

0.0 0.3 100 0

0.5 0.6 100 0

Time (min) Flow Rate(mL/min)

%A %B

4.0 0.6 100 0

6.0 0.6 0 100

8.0 0.6 0 100

10.0 1.0 0 100

12.0 1.0 50 50

14.0 0.6 100 100

ResultsSample AnalysisTo investigate relative responses of different oils, three differentsilicone oil samples were characterized consisting of two similar oilswith different viscosities (200 cP and 1000 cP) and a third oil that isused as a heating bath oil. From the overlay of chromatograms shown in Figure 2, the 1000 cP oil sample is comprised of a greater amount oflarger molecular weight polysiloxanes (later elution) than the 200 cP oil sample, as characterized by the bulk of peak area at higher retention times. The heating bath oil is comprised of more isolable polysiloxanechain groups, as identified by over 170 different peaks that were atleast partially resolved for this oil.

Retention time precision was demonstrated by repetitive analysis of the heating bath silicone oil sample (Figure 3). The UltiMate 3000 LCsystems, controlled by the Chromeleon software, enables sample injections to be synchronized with the pump delivery, allowing for exactand matching system conditions for each injection. The percent relative standard deviation for retention time at 50 minutes was 0.02%.

A silicone-based topical product used for comfort with hearing aids wascharacterized using this method, as shown by the chromatogram in Figure 4. The material appears to contain a moderate amount ofsmaller molecular weight polysiloxanes mixed with an amount of highermolecular weight polysiloxanes that elute at the end of the gradient.

The method provides a high-resolution HPLC chromatogram for the detailed characterization of polysiloxane materials. The use ofthe solid core column in this method achieves the high number oftheoretical plates and therefore, the high level of resolution shown in these figures. The charged aerosol detector provides both sensitive and consistent response for these analytes.

The near-consistent response of charged aerosol detection for non-volatile analytes enables the quantitation of total silicones in the same samples used above. The second method, specifically developed forquantitation, uses an aqueous acetonitrile mobile phase with a fasttetrahydrofuran gradient to elute the analyte from the column in a rapid manner, yielding a near-single peak. Chromatographic baseline subtraction is used to remove baseline deviations that result from the rapid gradient.

The 200 cP oil showed essentially equal response to that of the 1000 cPoil, but the heating bath oil differed in response at the two amounts used in this comparison. This may be associated with the differences of the groups or numbers of associated groups contained on the silicone. Thus,it is recommended that the silicone oil calibration standard is the same (orat least similar) to the silicone that is to be quantified. As seen in Figure 2,there is a clear difference between the 200 and 1000 cP silicones and the heating bath oil silicone.

FIGURE 5. Silicone oil calibration curves, from 78 to 10,000 ng (1000 cP)in triplicate, 2000 to 10,000 ng o.c. (200 cP and Heating Bath Oil)

0

2

4

6

8

10

12

0 2000 4000 6000 8000 10000 12000

Peak

Are

a (pA

*min)

Amount (ng o.c.)

AP1000

AP200

HeatingOil

TABLE 1. Response factors of different polysiloxanes at 5000 and2000 ng o.c. relative to 1000 cP silicone oil

Sample Amount on Column (ng)

Amount Found (ng) Relative Response

AP2005000 5031 1.0062000 1994 0.997

Heating Bath Oil5000 8681 1.7362000 4199 2.100

6.50 6.80 7.00 7.20 7.40 7.60 7.80 8.00 8.20 8.40 8.60 8.80 9.00 9.20 9.40 9.60 9.80 10.00-5.0

0.0

10.0

20.0

30.0

40.0

45.0 pA

min

21 1 - 6.980 2 - 7.4523 - 7.5224 - 7.665

5 - 7.954

6 - 8.0327 - 8.187

8 - SiliconeOil - 8.295

9 - 8.417

2 - SiliconeOil – 8.060

Two of the three oils (the two most closely related) showed a similarcalibration curve, indicating that calibration curves derived from oilsof a similar composition can be used for product concentration determinations. Differences in chain length and other structures mayaffect retention time. The heating bath oil, which characterized as a different oil, showed a greater response than the viscosity-rated oils.

FIGURE 6. HPLC chromatogram overlay of over-the-counter gasrelief product (blue), dissolved at a concentration of 500 µg/mLsimethicone in tetrahydrofuran, and 500 µg/mL of 1000 cP silicone oil standard (black)

*simethicone is a mixture of silica (5.5%) and dimethicone (94.5%)1

The measurement of silicone oils in different commercial products showed the potential for quantitative determinations without extraction of the silicone from the sample or having the exact silicone oil as a calibrant.Samples were dissolved, centrifuged, diluted to a targeted amount, and analyzed. A gas relieve product chromatogram is shown in Figure 6 with an overlay of 1000 cP silicone oil. The recovery for the gas relief product(prepared at 500 µg/mL of simethicone) was 96.5%, as shown in Table 2,demonstrating the potential for this method to measure silicone oils with unknown composition in different samples.

TABLE 2. Quantities of silicone oils found in products.

Sample Silicone Oil TypeDeterminedSilicone Oil

(ng o.c.)Product Silicone

AmountRecovery

(%)

Shampoo Dimethicone 3890 1.95 w/w-% --Conditioner Phenyltrimethicone 2702 2.70 w/w-% --Gas ReliefProduct

Simethicone* 4559 4725 ng o.c. 96.5

ConclusionsTwo methods were developed for the characterization and quantitation of silicone oils as standards and as part of products. The long characterization method can readily be modified to address specific regionsof the chromatogram or to fully characterize a specific oil in less time. The quantification method is sensitive to 100 ng o.c. and it was shown thatsimilar oils provide similar response factors. The method also had sufficientspecificity to allow for analysis of different products and silicone oils.

The use of the solid core C18 column, combined with the detection capabilities of the charged aerosol detector, enables convenient and accurate characterization and quantification of silicone oils and silicone oil-containing products.

References1. DailyMedPlus website

http://www.dailymedplus.com/monograph/view/setid/da8eba22-33e1-4f46-b43c-9ca3d8789740 (last accessed 30 Jan 13).

© 2013 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property ofThermo Fisher Scientific Inc. and its subsidiaries. This information is not intended toencourage use of these products in any manners that might infringe the intellectual propertyrights of others.

PO738_E 03/13S

Sample and Standard Preparations - QuantitationSamples containing silicone oil were dissolved in tetrahydrofuran, at a product concentration of 10—20 mg/mL and centrifuged at 10,000 g for 3 minutes. HPLC sample vials were filled with the supernatant.

Standards (Silicone oil 1000 cP) were prepared by dissolving 10 mg/mL of silicone oil in tetrahydrofuran, diluted to 1 mg/mL and then sequentially diluted to lower concentrations.

FIGURE 2. Characterization of three different silicone oils, 200 cP(blue) and 1000 cP (black) viscosities, and a heating bathsilicone oil (pink)

FIGURE 3. Overlays of three separate analyses of the heatingbath silicone oil sample. Only the 50 to 55 minute segment of the total run (inset) is shown for clarity.

FIGURE 4. HPLC chromatogram of a silicone-based topicalproduct. Product appears to be made of high-molecular weightpolysiloxane (85 minutes) mixed with moderate and low-molecular weight polysiloxanes. The inset shows the low to moderate molecular weight regions.

0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100-10.0

0.0

10.0

20.0

30.0

40.0

50.0

60.0

70.0

80.0

min

pA

0.0 10.0 20.0 30.0 40.0 50.0 60.0 70.0 75.0-2.0

0.0

2.0

4.0

6.0

8.0

10.0

12.0

14.0

16.0

18.0

min

pA

This quantitative method was evaluated using different silicone oils; using the 1000 cP silicone oil standards, a calibration curve was generated from78 to 10,000 ng o.c., as shown in Figure 5. System precision (n=3) wasacceptable, with peak area percent relative standard deviations 0.92 (10,000 ng o.c.) to 7.28 (156 ng o.c.). The quantitation limit is estimated atapproximately 100 ng o.c, based on a signal to noise ratio of 3.0.

Samples that were tested included a 200 cP silicone oil, the heating bath oil, and three consumer products, including a shampoo, a hair conditioner,and a gas relief product containing different forms of silicones. The 200 cPoil and the heating bath oil were also used to generate single-injection calibration curves to investigate their response factors, relative to the 1000 cP oil standard. These results are in Table 1.

49.48 50.00 50.50 51.00 51.50 52.00 52.50 53.00 53.50 54.00 54.50 54.815.5

6.0

7.0

8.0

9.0

10.0

11.0

12.0

13.0

14.0

15.0

16.0

17.017.4

min

pA

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol DetectionMarc Plante, Bruce Bailey, Ian N. AcworthThermo Fisher Scientific, Chelmsford, MA, USA

OverviewPurpose: To develop HPLC methods for both the detailed,characterization of silicone oils and for rapid quantitation of silicone oils in samples using an HPLC system with a charged aerosoldetector.

Methods: Two methods, one qualitative for characterization and one quantitative using a solid core C18 column, are outlined.

Results: Different silicone oils were characterized using the qualitative method; silicone oils in commercial products were measured, with a quantitation limit of approximately 100 ng on column.

IntroductionSilicone oils are unique materials with multitudes of significant uses in our modern world. Silicone-based materials are thermally stable,relatively inert, and are generally non-toxic, making them useful asoils and greases, lubricants, anti-foaming agents, and coatings.Silicone oils are often used in electronics, hydraulic systems, and many consumer products. The measurement of silicone oils iscomplicated by their solubility properties, their variations in basicstructure including a wide variety of chemical moieties that may be part of this structure, as well as their general lack of chromophoresmaking detection by ultraviolet absorption impractical.

The charged aerosol detector is a sensitive, mass-based detector,especially well-suited for the determination of any nonvolatile analyte independent of chemical characteristics. As shown in Figure 1, the detector uses nebulization to create aerosol droplets. The mobile phase evaporates in the drying tube, leaving analyte particles, which become charged in the mixing chamber. The charge is then measured by a highly sensitive electrometer, providing reproducible, nanogram-level sensitivity. This technology has greater sensitivity and precision than ELSD and refractive index (RI), is gradient compatible and issimpler to operate than a mass spectrometer (MS). Compounds do not have to possess a chromophore (unlike UV detection) or be ionized (as with MS).

Two charged aerosol detection methods were developed: one to characterize the individual components of silicone oils, and the otherto provide quantitation of a near single peak of silicone oil to 100ng (o.c.). The characterization method used the Thermo Scientific™Dionex™ UltiMate™ 3000 RS system, and the quantitation method used an UltiMate 3000 SD system in normal phase. Both methodsused a solid core C18 column and the Thermo Scientific™ Dionex™Corona™ ultra RS™ charged aerosol detector.

The combination of these chromatographic methods with the capabilities of the charged aerosol detection provides unique opportunities for detailed characterization of a silicone oil sample, aswell as quantitation of silicone oil-containing samples.

.

FIGURE 1. Schematic and Functioning of Charged AerosolDetection

1

2

3

4

5

6

7

89

101

2

3

4

5

6

7

89

10

1. Liquid eluent enters from HPLC system2. Pneumatic nebulization occurs3. Small droplets enter drying tube4. Large droplets exit to drain5. Dried particles enter mixing chamber6. Gas stream passes over corona needle7. Charged gas collides with particles and

charge is transferred8. High mobility species are removed9. Charge is measured by a highly sensitive

electrometer10.Signal transferred to chromatographic

softwareMethodsSample Preparation - CharacterizationSamples of silicone oil were dissolved in chloroform, at a concentration of approximately 40 mg/mL.

Liquid Chromatography – CharacterizationHPLC System: Thermo Scientific™ Dionex™ UltiMate™ 3000

DGP-3600RS pump, WPS-3000RS autosampler,and TCC-3000RS column oven

HPLC Column: Thermo Scientific™ Accucore™ 2.6 µm C18,3.0 × 150 mm

Column Temp.: 40 °CMobile Phase A: MethanolMobile Phase B: n-PropanolFlow Rate: 0.5 mL/minInjection Volume: 2–10 µLSample: 38 mg/mL silicone oil in chloroformDetector: Corona ultra RSNebulizer Temp.: 10 °CFilter: 5Data Rate: 10 HzPower Function: 1.00Flow Gradient for Characterization:

Time (min) %A %B Curve

-10 100 10 5

0 100 10 5

80 15 85 3

85 0 100 5

95 0 100 5

95 100 0 5

100 100 0 5

Liquid Chromatography –Quantitation HPLC System: Thermo Scientific™ Dionex™ UltiMate™ 3000

LPG-3400SD pump, WPS-3000RS autosampler, andTCC-3000RS column ovenHPLC Column: Accucore 2.6 µm C18, 3.0 × 150 mmColumn Temp.: 40 °CMobile Phase A: 0.5% Formic acid/acetonitrile/tetrahydrofuran*

(35:35:30), helium spargeMobile Phase B: Tetrahydrofuran*, helium spargeFlow Rate: 0.3–1.0 mL/minInjection Volume: 2–10 µLDetector: Corona ultra RSNebulizer Temp.: 10 °CFilter: 6Data Rate: 10 HzPower Function: 2Flow Gradient for Quantitation:

*No BHT added.

Data AnalysisAll HPLC chromatograms were obtained and compiled using Thermo Scientific™ Dionex™ Chromeleon™ Chromatography Data Station software, 7.1 SR 1.

Time (min) Flow Rate (mL/min)

%A %B

-5.0 0.6 100 0

-0.2 0.6 100 0

-0.1 0.3 100 0

0.0 0.3 100 0

0.5 0.6 100 0

Time (min) Flow Rate (mL/min)

%A %B

4.0 0.6 100 0

6.0 0.6 0 100

8.0 0.6 0 100

10.0 1.0 0 100

12.0 1.0 50 50

14.0 0.6 100 100

ResultsSample AnalysisTo investigate relative responses of different oils, three differentsilicone oil samples were characterized consisting of two similar oilswith different viscosities (200 cP and 1000 cP) and a third oil that isused as a heating bath oil. From the overlay of chromatograms shown in Figure 2, the 1000 cP oil sample is comprised of a greater amount oflarger molecular weight polysiloxanes (later elution) than the 200 cP oil sample, as characterized by the bulk of peak area at higher retention times. The heating bath oil is comprised of more isolable polysiloxanechain groups, as identified by over 170 different peaks that were atleast partially resolved for this oil.

Retention time precision was demonstrated by repetitive analysis of the heating bath silicone oil sample (Figure 3). The UltiMate 3000 LCsystems, controlled by the Chromeleon software, enables sample injections to be synchronized with the pump delivery, allowing for exactand matching system conditions for each injection. The percent relative standard deviation for retention time at 50 minutes was 0.02%.

A silicone-based topical product used for comfort with hearing aids wascharacterized using this method, as shown by the chromatogram in Figure 4. The material appears to contain a moderate amount ofsmaller molecular weight polysiloxanes mixed with an amount of highermolecular weight polysiloxanes that elute at the end of the gradient.

The method provides a high-resolution HPLC chromatogram for the detailed characterization of polysiloxane materials. The use of the solid core column in this method achieves the high number oftheoretical plates and therefore, the high level of resolution shown in these figures. The charged aerosol detector provides both sensitive and consistent response for these analytes.

The near-consistent response of charged aerosol detection for non-volatile analytes enables the quantitation of total silicones in the same samples used above. The second method, specifically developed forquantitation, uses an aqueous acetonitrile mobile phase with a fasttetrahydrofuran gradient to elute the analyte from the column in a rapid manner, yielding a near-single peak. Chromatographic baseline subtraction is used to remove baseline deviations that result from the rapid gradient.

The 200 cP oil showed essentially equal response to that of the 1000 cPoil, but the heating bath oil differed in response at the two amounts used in this comparison. This may be associated with the differences of the groups or numbers of associated groups contained on the silicone. Thus,it is recommended that the silicone oil calibration standard is the same (orat least similar) to the silicone that is to be quantified. As seen in Figure 2,there is a clear difference between the 200 and 1000 cP silicones and the heating bath oil silicone.

FIGURE 5. Silicone oil calibration curves, from 78 to 10,000 ng (1000 cP)in triplicate, 2000 to 10,000 ng o.c. (200 cP and Heating Bath Oil)

0

2

4

6

8

10

12

0 2000 4000 6000 8000 10000 12000

Peak

Are

a (pA

*min)

Amount (ng o.c.)

AP1000

AP200

HeatingOil

TABLE 1. Response factors of different polysiloxanes at 5000 and2000 ng o.c. relative to 1000 cP silicone oil

Sample Amount on Column (ng)

Amount Found (ng) Relative Response

AP2005000 5031 1.0062000 1994 0.997

Heating Bath Oil5000 8681 1.7362000 4199 2.100

6.50 6.80 7.00 7.20 7.40 7.60 7.80 8.00 8.20 8.40 8.60 8.80 9.00 9.20 9.40 9.60 9.80 10.00-5.0

0.0

10.0

20.0

30.0

40.0

45.0 pA

min

21 1 - 6.980 2 - 7.4523 - 7.5224 - 7.665

5 - 7.954

6 - 8.0327 - 8.187

8 - SiliconeOil - 8.295

9 - 8.417

2 - SiliconeOil – 8.060

Two of the three oils (the two most closely related) showed a similarcalibration curve, indicating that calibration curves derived from oils of a similar composition can be used for product concentration determinations. Differences in chain length and other structures mayaffect retention time. The heating bath oil, which characterized as a different oil, showed a greater response than the viscosity-rated oils.

FIGURE 6. HPLC chromatogram overlay of over-the-counter gasrelief product (blue), dissolved at a concentration of 500 µg/mLsimethicone in tetrahydrofuran, and 500 µg/mL of 1000 cP silicone oil standard (black)

*simethicone is a mixture of silica (5.5%) and dimethicone (94.5%)1

The measurement of silicone oils in different commercial products showed the potential for quantitative determinations without extraction of the silicone from the sample or having the exact silicone oil as a calibrant.Samples were dissolved, centrifuged, diluted to a targeted amount, and analyzed. A gas relieve product chromatogram is shown in Figure 6 with an overlay of 1000 cP silicone oil. The recovery for the gas relief product(prepared at 500 µg/mL of simethicone) was 96.5%, as shown in Table 2,demonstrating the potential for this method to measure silicone oils with unknown composition in different samples.

TABLE 2. Quantities of silicone oils found in products.

Sample Silicone Oil TypeDeterminedSilicone Oil

(ng o.c.)Product Silicone

AmountRecovery

(%)

Shampoo Dimethicone 3890 1.95 w/w-% --Conditioner Phenyltrimethicone 2702 2.70 w/w-% --Gas ReliefProduct

Simethicone* 4559 4725 ng o.c. 96.5

ConclusionsTwo methods were developed for the characterization and quantitation ofsilicone oils as standards and as part of products. The long characterization method can readily be modified to address specific regionsof the chromatogram or to fully characterize a specific oil in less time. The quantification method is sensitive to 100 ng o.c. and it was shown thatsimilar oils provide similar response factors. The method also had sufficientspecificity to allow for analysis of different products and silicone oils.

The use of the solid core C18 column, combined with the detection capabilities of the charged aerosol detector, enables convenient and accurate characterization and quantification of silicone oils and silicone oil-containing products.

References1http://www.dailymedplus.com/monograph/view/setid/da8eba22-33e1-4f46-b43c-9ca3d8789740 (last accessed 30 Jan 13).

© 2013 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property ofThermo Fisher Scientific Inc. and its subsidiaries. This information is not intended toencourage use of these products in any manners that might infringe the intellectual propertyrights of others.

PO738_E 03/13S

Sample and Standard Preparations - QuantitationSamples containing silicone oil were dissolved in tetrahydrofuran, at a product concentration of 10—20 mg/mL and centrifuged at 10,000 g for3 minutes. HPLC sample vials were filled with the supernatant.

Standards (Silicone oil 1000 cP) were prepared by dissolving 10 mg/mL of silicone oil in tetrahydrofuran, diluted to 1 mg/mL and then sequentially diluted to lower concentrations.

FIGURE 2. Characterization of three different silicone oils, 200 cP(blue) and 1000 cP (black) viscosities, and a heating bathsilicone oil (pink)

49.48 50.00 50.50 51.00 51.50 52.00 52.50 53.00 53.50 54.00 54.50 54.815.5

6.0

7.0

8.0

9.0

10.0

11.0

12.0

13.0

14.0

15.0

16.0

17.017.4

min

pA

FIGURE 3. Overlays of three separate analyses of the heatingbath silicone oil sample. Only the 50 to 55 minute segment of the total run (inset) is shown for clarity.

FIGURE 4. HPLC chromatogram of a silicone-based topicalproduct. Product appears to be made of high-molecular weightpolysiloxane (85 minutes) mixed with moderate and low-molecular weight polysiloxanes. The inset shows the low to moderate molecular weight regions.

0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100-10.0

0.0

10.0

20.0

30.0

40.0

50.0

60.0

70.0

80.0

min

pA

0.0 10.0 20.0 30.0 40.0 50.0 60.0 70.0 75.0-2.0

0.0

2.0

4.0

6.0

8.0

10.0

12.0

14.0

16.0

18.0

min

pA

This quantitative method was evaluated using different silicone oils; using the 1000 cP silicone oil standards, a calibration curve was generated from78 to 10,000 ng o.c., as shown in Figure 5. System precision (n=3) wasacceptable, with peak area percent relative standard deviations 0.92 (10,000 ng o.c.) to 7.28 (156 ng o.c.). The quantitation limit is estimated atapproximately 100 ng o.c, based on a signal to noise ratio of 3.0.

Samples that were tested included a 200 cP silicone oil, the heating bath oil, and three consumer products, including a shampoo, a hair conditioner,and a gas relief product containing different forms of silicones. The 200 cPoil and the heating bath oil were also used to generate single-injection calibration curves to investigate their response factors, relative to the 1000 cP oil standard. These results are in Table 1.

Page 4: Analysis of Silicone Oils by High Performance Liquid ... · Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection Marc Plante, Bruce

4 Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol DetectionMarc Plante, Bruce Bailey, Ian N. AcworthThermo Fisher Scientific, Chelmsford, MA, USA

OverviewPurpose: To develop HPLC methods for both the detailed, characterization of silicone oils and for rapid quantitation of silicone oils in samples using an HPLC system with a charged aerosol detector.

Methods: Two methods, one qualitative for characterization and one quantitative using a solid core C18 column, are outlined.

Results: Different silicone oils were characterized using the qualitative method; silicone oils in commercial products were measured, with a quantitation limit of approximately 100 ng on column.

IntroductionSilicone oils are unique materials with multitudes of significant uses in our modern world. Silicone-based materials are thermally stable, relatively inert, and are generally non-toxic, making them useful as oils and greases, lubricants, anti-foaming agents, and coatings. Silicone oils are often used in electronics, hydraulic systems, and many consumer products. The measurement of silicone oils is complicated by their solubility properties, their variations in basic structure including a wide variety of chemical moieties that may be part of this structure, as well as their general lack of chromophores making detection by ultraviolet absorption impractical.

The charged aerosol detector is a sensitive, mass-based detector, especially well-suited for the determination of any nonvolatile analyte independent of chemical characteristics. As shown in Figure 1, the detector uses nebulization to create aerosol droplets. The mobile phase evaporates in the drying tube, leaving analyte particles, which become charged in the mixing chamber. The charge is then measured by a highly sensitive electrometer, providing reproducible, nanogram-level sensitivity. This technology has greater sensitivity and precision than ELSD and refractive index (RI), is gradient compatible and is simpler to operate than a mass spectrometer (MS). Compounds do not have to possess a chromophore (unlike UV detection) or be ionized (as with MS).

Two charged aerosol detection methods were developed: one to characterize the individual components of silicone oils, and the other to provide quantitation of a near single peak of silicone oil to 100ng (o.c.). The characterization method used the Thermo Scientific™ Dionex™ UltiMate™ 3000 RS system, and the quantitation method used an UltiMate 3000 SD system in normal phase. Both methods used a solid core C18 column and the Thermo Scientific™ Dionex™ Corona™ ultra RS™ charged aerosol detector.

The combination of these chromatographic methods with the capabilities of the charged aerosol detection provides unique opportunities for detailed characterization of a silicone oil sample, as well as quantitation of silicone oil-containing samples.

.

FIGURE 1. Schematic and Functioning of Charged Aerosol Detection

1

2

3

4

5

6

7

89

101

2

3

4

5

6

7

89

10

1. Liquid eluent enters from HPLC system2. Pneumatic nebulization occurs3. Small droplets enter drying tube4. Large droplets exit to drain5. Dried particles enter mixing chamber6. Gas stream passes over corona needle7. Charged gas collides with particles and

charge is transferred8. High mobility species are removed9. Charge is measured by a highly sensitive

electrometer10.Signal transferred to chromatographic

softwareMethodsSample Preparation - CharacterizationSamples of silicone oil were dissolved in chloroform, at a concentration of approximately 40 mg/mL.

Liquid Chromatography – CharacterizationHPLC System: Thermo Scientific™ Dionex™ UltiMate™ 3000

DGP-3600RS pump, WPS-3000RS autosampler, and TCC-3000RS column oven

HPLC Column: Thermo Scientific™ Accucore™ 2.6 µm C18, 3.0 × 150 mm

Column Temp.: 40 °CMobile Phase A: MethanolMobile Phase B: n-PropanolFlow Rate: 0.5 mL/minInjection Volume: 2–10 µLSample: 38 mg/mL silicone oil in chloroformDetector: Corona ultra RSNebulizer Temp.: 10 °CFilter: 5Data Rate: 10 HzPower Function: 1.00Flow Gradient for Characterization:

Time (min) %A %B Curve

-10 100 10 5

0 100 10 5

80 15 85 3

85 0 100 5

95 0 100 5

95 100 0 5

100 100 0 5

Liquid Chromatography –QuantitationHPLC System: Thermo Scientific™ Dionex™ UltiMate™ 3000

LPG-3400SD pump, WPS-3000RS autosampler, and TCC-3000RS column oven

HPLC Column: Accucore 2.6 µm C18, 3.0 × 150 mmColumn Temp.: 40 °CMobile Phase A: 0.5% Formic acid/acetonitrile/tetrahydrofuran*

(35:35:30), helium spargeMobile Phase B: Tetrahydrofuran*, helium spargeFlow Rate: 0.3–1.0 mL/minInjection Volume: 2–10 µLDetector: Corona ultra RSNebulizer Temp.: 10 °CFilter: 6Data Rate: 10 HzPower Function: 2Flow Gradient for Quantitation:

*No BHT added.

Data AnalysisAll HPLC chromatograms were obtained and compiled using Thermo Scientific™ Dionex™ Chromeleon™ Chromatography Data Station software, 7.1 SR 1.

Time (min) Flow Rate (mL/min)

%A %B

-5.0 0.6 100 0

-0.2 0.6 100 0

-0.1 0.3 100 0

0.0 0.3 100 0

0.5 0.6 100 0

Time (min) Flow Rate (mL/min)

%A %B

4.0 0.6 100 0

6.0 0.6 0 100

8.0 0.6 0 100

10.0 1.0 0 100

12.0 1.0 50 50

14.0 0.6 100 100

ResultsSample AnalysisTo investigate relative responses of different oils, three different silicone oil samples were characterized consisting of two similar oils with different viscosities (200 cP and 1000 cP) and a third oil that is used as a heating bath oil. From the overlay of chromatograms shown in Figure 2, the 1000 cP oil sample is comprised of a greater amount of larger molecular weight polysiloxanes (later elution) than the 200 cP oil sample, as characterized by the bulk of peak area at higher retention times. The heating bath oil is comprised of more isolable polysiloxanechain groups, as identified by over 170 different peaks that were at least partially resolved for this oil.

Retention time precision was demonstrated by repetitive analysis of the heating bath silicone oil sample (Figure 3). The UltiMate 3000 LC systems, controlled by the Chromeleon software, enables sample injections to be synchronized with the pump delivery, allowing for exact and matching system conditions for each injection. The percent relative standard deviation for retention time at 50 minutes was 0.02%.

A silicone-based topical product used for comfort with hearing aids was characterized using this method, as shown by the chromatogram in Figure 4. The material appears to contain a moderate amount of smaller molecular weight polysiloxanes mixed with an amount of higher molecular weight polysiloxanes that elute at the end of the gradient.

The method provides a high-resolution HPLC chromatogram for the detailed characterization of polysiloxane materials. The use of the solid core column in this method achieves the high number of theoretical plates and therefore, the high level of resolution shown in these figures. The charged aerosol detector provides both sensitive and consistent response for these analytes.

The near-consistent response of charged aerosol detection for non-volatile analytes enables the quantitation of total silicones in the same samples used above. The second method, specifically developed for quantitation, uses an aqueous acetonitrile mobile phase with a fast tetrahydrofuran gradient to elute the analyte from the column in a rapid manner, yielding a near-single peak. Chromatographic baseline subtraction is used to remove baseline deviations that result from the rapid gradient.

The 200 cP oil showed essentially equal response to that of the 1000 cP oil, but the heating bath oil differed in response at the two amounts used in this comparison. This may be associated with the differences of the groups or numbers of associated groups contained on the silicone. Thus, it is recommended that the silicone oil calibration standard is the same (or at least similar) to the silicone that is to be quantified. As seen in Figure 2, there is a clear difference between the 200 and 1000 cP silicones and the heating bath oil silicone.

FIGURE 5. Silicone oil calibration curves, from 78 to 10,000 ng (1000 cP) in triplicate, 2000 to 10,000 ng o.c. (200 cP and Heating Bath Oil)

0

2

4

6

8

10

12

0 2000 4000 6000 8000 10000 12000

Peak

Are

a (pA

*min)

Amount (ng o.c.)

AP1000

AP200

HeatingOil

TABLE 1. Response factors of different polysiloxanes at 5000 and 2000 ng o.c. relative to 1000 cP silicone oil

Sample Amount on

Column (ng) Amount Found

(ng) Relative Response

AP200 5000 5031 1.006

2000 1994 0.997

Heating Bath Oil 5000 8681 1.736

2000 4199 2.100

6.50 6.80 7.00 7.20 7.40 7.60 7.80 8.00 8.20 8.40 8.60 8.80 9.00 9.20 9.40 9.60 9.80 10.00-5.0

0.0

10.0

20.0

30.0

40.0

45.0 pA

min

21 1 - 6.980 2 - 7.4523 - 7.5224 - 7.665

5 - 7.954

6 - 8.0327 - 8.187

8 - SiliconeOil - 8.295

9 - 8.417

2 - SiliconeOil – 8.060

Two of the three oils (the two most closely related) showed a similar calibration curve, indicating that calibration curves derived from oils of a similar composition can be used for product concentration determinations. Differences in chain length and other structures may affect retention time. The heating bath oil, which characterized as a different oil, showed a greater response than the viscosity-rated oils.

FIGURE 6. HPLC chromatogram overlay of over-the-counter gas relief product (blue), dissolved at a concentration of 500 µg/mL simethicone in tetrahydrofuran, and 500 µg/mL of 1000 cP silicone oil standard (black)

*simethicone is a mixture of silica (5.5%) and dimethicone (94.5%)1

The measurement of silicone oils in different commercial products showed the potential for quantitative determinations without extraction of the silicone from the sample or having the exact silicone oil as a calibrant. Samples were dissolved, centrifuged, diluted to a targeted amount, and analyzed. A gas relieve product chromatogram is shown in Figure 6 with an overlay of 1000 cP silicone oil. The recovery for the gas relief product (prepared at 500 µg/mL of simethicone) was 96.5%, as shown in Table 2, demonstrating the potential for this method to measure silicone oils with unknown composition in different samples.

TABLE 2. Quantities of silicone oils found in products.

Sample Silicone Oil Type DeterminedSilicone Oil

(ng o.c.) Product Silicone

Amount Recovery

(%)

Shampoo Dimethicone 3890 1.95 w/w-% -- Conditioner Phenyltrimethicone 2702 2.70 w/w-% -- Gas ReliefProduct

Simethicone* 4559 4725 ng o.c. 96.5

ConclusionsTwo methods were developed for the characterization and quantitation of silicone oils as standards and as part of products. The long characterization method can readily be modified to address specific regions of the chromatogram or to fully characterize a specific oil in less time. The quantification method is sensitive to 100 ng o.c. and it was shown that similar oils provide similar response factors. The method also had sufficient specificity to allow for analysis of different products and silicone oils.

The use of the solid core C18 column, combined with the detection capabilities of the charged aerosol detector, enables convenient and accurate characterization and quantification of silicone oils and silicone oil-containing products.

References1. DailyMedPlus website

http://www.dailymedplus.com/monograph/view/setid/da8eba22-33e1-4f46-b43c-9ca3d8789740 (last accessed 30 Jan 13).

© 2013 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property ofThermo Fisher Scientific Inc. and its subsidiaries. This information is not intended toencourage use of these products in any manners that might infringe the intellectual propertyrights of others.

PO738_E 03/13S

Sample and Standard Preparations - QuantitationSamples containing silicone oil were dissolved in tetrahydrofuran, at a product concentration of 10—20 mg/mL and centrifuged at 10,000 g for 3 minutes. HPLC sample vials were filled with the supernatant.

Standards (Silicone oil 1000 cP) were prepared by dissolving 10 mg/mL of silicone oil in tetrahydrofuran, diluted to 1 mg/mL and then sequentially diluted to lower concentrations.

FIGURE 2. Characterization of three different silicone oils, 200 cP(blue) and 1000 cP (black) viscosities, and a heating bath silicone oil (pink)

FIGURE 3. Overlays of three separate analyses of the heating bath silicone oil sample. Only the 50 to 55 minute segment of the total run (inset) is shown for clarity.

FIGURE 4. HPLC chromatogram of a silicone-based topical product. Product appears to be made of high-molecular weight polysiloxane (85 minutes) mixed with moderate and low-molecular weight polysiloxanes. The inset shows the low to moderate molecular weight regions.

0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100-10.0

0.0

10.0

20.0

30.0

40.0

50.0

60.0

70.0

80.0

min

pA

0.0 10.0 20.0 30.0 40.0 50.0 60.0 70.0 75.0-2.0

0.0

2.0

4.0

6.0

8.0

10.0

12.0

14.0

16.0

18.0

min

pA

This quantitative method was evaluated using different silicone oils; using the 1000 cP silicone oil standards, a calibration curve was generated from 78 to 10,000 ng o.c., as shown in Figure 5. System precision (n=3) was acceptable, with peak area percent relative standard deviations 0.92 (10,000 ng o.c.) to 7.28 (156 ng o.c.). The quantitation limit is estimated at approximately 100 ng o.c, based on a signal to noise ratio of 3.0.

Samples that were tested included a 200 cP silicone oil, the heating bath oil, and three consumer products, including a shampoo, a hair conditioner, and a gas relief product containing different forms of silicones. The 200 cP oil and the heating bath oil were also used to generate single-injection calibration curves to investigate their response factors, relative to the 1000 cP oil standard. These results are in Table 1.

49.48 50.00 50.50 51.00 51.50 52.00 52.50 53.00 53.50 54.00 54.50 54.81 5.5 6.0

7.0

8.0

9.0

10.0

11.0

12.0

13.0

14.0

15.0

16.0

17.0 17.4

min

pA

Page 5: Analysis of Silicone Oils by High Performance Liquid ... · Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection Marc Plante, Bruce

5Thermo Scientific Poster Note • PN70538_e 06/13S

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol DetectionMarc Plante, Bruce Bailey, Ian N. AcworthThermo Fisher Scientific, Chelmsford, MA, USA

OverviewPurpose: To develop HPLC methods for both the detailed,characterization of silicone oils and for rapid quantitation of silicone oils in samples using an HPLC system with a charged aerosoldetector.

Methods: Two methods, one qualitative for characterization and one quantitative using a solid core C18 column, are outlined.

Results: Different silicone oils were characterized using the qualitative method; silicone oils in commercial products were measured, with a quantitation limit of approximately 100 ng on column.

IntroductionSilicone oils are unique materials with multitudes of significant uses in our modern world. Silicone-based materials are thermally stable,relatively inert, and are generally non-toxic, making them useful asoils and greases, lubricants, anti-foaming agents, and coatings.Silicone oils are often used in electronics, hydraulic systems, and many consumer products. The measurement of silicone oils iscomplicated by their solubility properties, their variations in basicstructure including a wide variety of chemical moieties that may be part of this structure, as well as their general lack of chromophoresmaking detection by ultraviolet absorption impractical.

The charged aerosol detector is a sensitive, mass-based detector,especially well-suited for the determination of any nonvolatile analyte independent of chemical characteristics. As shown in Figure 1, the detector uses nebulization to create aerosol droplets. The mobile phase evaporates in the drying tube, leaving analyte particles, which become charged in the mixing chamber. The charge is then measured by a highly sensitive electrometer, providing reproducible, nanogram-level sensitivity. This technology has greater sensitivity and precision than ELSD and refractive index (RI), is gradient compatible and issimpler to operate than a mass spectrometer (MS). Compounds do not have to possess a chromophore (unlike UV detection) or be ionized (as with MS).

Two charged aerosol detection methods were developed: one to characterize the individual components of silicone oils, and the otherto provide quantitation of a near single peak of silicone oil to 100ng (o.c.). The characterization method used the Thermo Scientific™Dionex™ UltiMate™ 3000 RS system, and the quantitation method used an UltiMate 3000 SD system in normal phase. Both methodsused a solid core C18 column and the Thermo Scientific™ Dionex™Corona™ ultra RS™ charged aerosol detector.

The combination of these chromatographic methods with the capabilities of the charged aerosol detection provides unique opportunities for detailed characterization of a silicone oil sample, aswell as quantitation of silicone oil-containing samples.

.

FIGURE 1. Schematic and Functioning of Charged AerosolDetection

1

2

3

4

5

6

7

89

101

2

3

4

5

6

7

89

10

1. Liquid eluent enters from HPLC system2. Pneumatic nebulization occurs3. Small droplets enter drying tube4. Large droplets exit to drain5. Dried particles enter mixing chamber6. Gas stream passes over corona needle7. Charged gas collides with particles and

charge is transferred8. High mobility species are removed9. Charge is measured by a highly sensitive

electrometer10.Signal transferred to chromatographic

softwareMethodsSample Preparation - CharacterizationSamples of silicone oil were dissolved in chloroform, at a concentration of approximately 40 mg/mL.

Liquid Chromatography – CharacterizationHPLC System: Thermo Scientific™ Dionex™ UltiMate™ 3000

DGP-3600RS pump, WPS-3000RS autosampler, and TCC-3000RS column oven

HPLC Column: Thermo Scientific™ Accucore™ 2.6 µm C18,3.0 × 150 mm

Column Temp.: 40 °CMobile Phase A: MethanolMobile Phase B: n-PropanolFlow Rate: 0.5 mL/minInjection Volume: 2–10 µLSample: 38 mg/mL silicone oil in chloroformDetector: Corona ultra RSNebulizer Temp.: 10 °CFilter: 5Data Rate: 10 HzPower Function: 1.00Flow Gradient for Characterization:

Time (min) %A %B Curve

-10 100 10 5

0 100 10 5

80 15 85 3

85 0 100 5

95 0 100 5

95 100 0 5

100 100 0 5

Liquid Chromatography –QuantitationHPLC System: Thermo Scientific™ Dionex™ UltiMate™ 3000

LPG-3400SD pump, WPS-3000RS autosampler, and TCC-3000RS column oven

HPLC Column: Accucore 2.6 µm C18, 3.0 × 150 mmColumn Temp.: 40 °CMobile Phase A: 0.5% Formic acid/acetonitrile/tetrahydrofuran*

(35:35:30), helium spargeMobile Phase B: Tetrahydrofuran*, helium spargeFlow Rate: 0.3–1.0 mL/minInjection Volume: 2–10 µLDetector: Corona ultra RSNebulizer Temp.: 10 °CFilter: 6Data Rate: 10 HzPower Function: 2Flow Gradient for Quantitation:

*No BHT added.

Data AnalysisAll HPLC chromatograms were obtained and compiled using Thermo Scientific™ Dionex™ Chromeleon™ Chromatography Data Station software, 7.1 SR 1.

Time (min) Flow Rate(mL/min)

%A %B

-5.0 0.6 100 0

-0.2 0.6 100 0

-0.1 0.3 100 0

0.0 0.3 100 0

0.5 0.6 100 0

Time (min) Flow Rate(mL/min)

%A %B

4.0 0.6 100 0

6.0 0.6 0 100

8.0 0.6 0 100

10.0 1.0 0 100

12.0 1.0 50 50

14.0 0.6 100 100

ResultsSample AnalysisTo investigate relative responses of different oils, three differentsilicone oil samples were characterized consisting of two similar oilswith different viscosities (200 cP and 1000 cP) and a third oil that isused as a heating bath oil. From the overlay of chromatograms shown in Figure 2, the 1000 cP oil sample is comprised of a greater amount oflarger molecular weight polysiloxanes (later elution) than the 200 cP oil sample, as characterized by the bulk of peak area at higher retention times. The heating bath oil is comprised of more isolable polysiloxanechain groups, as identified by over 170 different peaks that were atleast partially resolved for this oil.

Retention time precision was demonstrated by repetitive analysis of the heating bath silicone oil sample (Figure 3). The UltiMate 3000 LCsystems, controlled by the Chromeleon software, enables sample injections to be synchronized with the pump delivery, allowing for exactand matching system conditions for each injection. The percent relative standard deviation for retention time at 50 minutes was 0.02%.

A silicone-based topical product used for comfort with hearing aids wascharacterized using this method, as shown by the chromatogram in Figure 4. The material appears to contain a moderate amount ofsmaller molecular weight polysiloxanes mixed with an amount of highermolecular weight polysiloxanes that elute at the end of the gradient.

The method provides a high-resolution HPLC chromatogram for the detailed characterization of polysiloxane materials. The use ofthe solid core column in this method achieves the high number oftheoretical plates and therefore, the high level of resolution shown in these figures. The charged aerosol detector provides both sensitive and consistent response for these analytes.

The near-consistent response of charged aerosol detection for non-volatile analytes enables the quantitation of total silicones in the same samples used above. The second method, specifically developed forquantitation, uses an aqueous acetonitrile mobile phase with a fasttetrahydrofuran gradient to elute the analyte from the column in a rapid manner, yielding a near-single peak. Chromatographic baseline subtraction is used to remove baseline deviations that result from the rapid gradient.

The 200 cP oil showed essentially equal response to that of the 1000 cP oil, but the heating bath oil differed in response at the two amounts used in this comparison. This may be associated with the differences of the groups or numbers of associated groups contained on the silicone. Thus, it is recommended that the silicone oil calibration standard is the same (or at least similar) to the silicone that is to be quantified. As seen in Figure 2, there is a clear difference between the 200 and 1000 cP silicones and the heating bath oil silicone.

FIGURE 5. Silicone oil calibration curves, from 78 to 10,000 ng (1000 cP) in triplicate, 2000 to 10,000 ng o.c. (200 cP and Heating Bath Oil)

0

2

4

6

8

10

12

0 2000 4000 6000 8000 10000 12000

Peak

Are

a (pA

*min)

Amount (ng o.c.)

AP1000

AP200

HeatingOil

TABLE 1. Response factors of different polysiloxanes at 5000 and 2000 ng o.c. relative to 1000 cP silicone oil

Sample Amount on

Column (ng)Amount Found

(ng) Relative Response

AP200 5000 5031 1.006

2000 1994 0.997

Heating Bath Oil5000 8681 1.736

2000 4199 2.100

6.50 6.80 7.00 7.20 7.40 7.60 7.80 8.00 8.20 8.40 8.60 8.80 9.00 9.20 9.40 9.60 9.80 10.00-5.0

0.0

10.0

20.0

30.0

40.0

45.0 pA

min

21 1 - 6.980 2 - 7.4523 - 7.5224 - 7.665

5 - 7.954

6 - 8.0327 - 8.187

8 - SiliconeOil - 8.295

9 - 8.417

2 - SiliconeOil – 8.060

Two of the three oils (the two most closely related) showed a similar calibration curve, indicating that calibration curves derived from oils of a similar composition can be used for product concentration determinations. Differences in chain length and other structures may affect retention time. The heating bath oil, which characterized as a different oil, showed a greater response than the viscosity-rated oils.

FIGURE 6. HPLC chromatogram overlay of over-the-counter gas relief product (blue), dissolved at a concentration of 500 µg/mL simethicone in tetrahydrofuran, and 500 µg/mL of 1000 cP silicone oil standard (black)

*simethicone is a mixture of silica (5.5%) and dimethicone (94.5%)1

The measurement of silicone oils in different commercial products showed the potential for quantitative determinations without extraction of the silicone from the sample or having the exact silicone oil as a calibrant.Samples were dissolved, centrifuged, diluted to a targeted amount, and analyzed. A gas relieve product chromatogram is shown in Figure 6 with an overlay of 1000 cP silicone oil. The recovery for the gas relief product(prepared at 500 µg/mL of simethicone) was 96.5%, as shown in Table 2,demonstrating the potential for this method to measure silicone oils with unknown composition in different samples.

TABLE 2. Quantities of silicone oils found in products.

Sample Silicone Oil Type DeterminedSilicone Oil

(ng o.c.)Product Silicone

AmountRecovery

(%)

Shampoo Dimethicone 3890 1.95 w/w-% --Conditioner Phenyltrimethicone 2702 2.70 w/w-% --Gas ReliefProduct

Simethicone* 4559 4725 ng o.c. 96.5

ConclusionsTwo methods were developed for the characterization and quantitation of silicone oils as standards and as part of products. The long characterization method can readily be modified to address specific regionsof the chromatogram or to fully characterize a specific oil in less time. The quantification method is sensitive to 100 ng o.c. and it was shown thatsimilar oils provide similar response factors. The method also had sufficientspecificity to allow for analysis of different products and silicone oils.

The use of the solid core C18 column, combined with the detection capabilities of the charged aerosol detector, enables convenient and accurate characterization and quantification of silicone oils and silicone oil-containing products.

References1. DailyMedPlus website

http://www.dailymedplus.com/monograph/view/setid/da8eba22-33e1-4f46-b43c-9ca3d8789740 (last accessed 30 Jan 13).

© 2013 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property ofThermo Fisher Scientific Inc. and its subsidiaries. This information is not intended toencourage use of these products in any manners that might infringe the intellectual propertyrights of others.

PO738_E 03/13S

Sample and Standard Preparations - QuantitationSamples containing silicone oil were dissolved in tetrahydrofuran, at a product concentration of 10—20 mg/mL and centrifuged at 10,000 g for3 minutes. HPLC sample vials were filled with the supernatant.

Standards (Silicone oil 1000 cP) were prepared by dissolving 10 mg/mL of silicone oil in tetrahydrofuran, diluted to 1 mg/mL and then sequentially diluted to lower concentrations.

FIGURE 2. Characterization of three different silicone oils, 200 cP(blue) and 1000 cP (black) viscosities, and a heating bathsilicone oil (pink)

FIGURE 3. Overlays of three separate analyses of the heatingbath silicone oil sample. Only the 50 to 55 minute segment of the total run (inset) is shown for clarity.

FIGURE 4. HPLC chromatogram of a silicone-based topicalproduct. Product appears to be made of high-molecular weightpolysiloxane (85 minutes) mixed with moderate and low-molecular weight polysiloxanes. The inset shows the low to moderate molecular weight regions.

0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100-10.0

0.0

10.0

20.0

30.0

40.0

50.0

60.0

70.0

80.0

min

pA

0.0 10.0 20.0 30.0 40.0 50.0 60.0 70.0 75.0-2.0

0.0

2.0

4.0

6.0

8.0

10.0

12.0

14.0

16.0

18.0

min

pA

This quantitative method was evaluated using different silicone oils; using the 1000 cP silicone oil standards, a calibration curve was generated from 78 to 10,000 ng o.c., as shown in Figure 5. System precision (n=3) was acceptable, with peak area percent relative standard deviations 0.92 (10,000 ng o.c.) to 7.28 (156 ng o.c.). The quantitation limit is estimated at approximately 100 ng o.c, based on a signal to noise ratio of 3.0.

Samples that were tested included a 200 cP silicone oil, the heating bath oil, and three consumer products, including a shampoo, a hair conditioner, and a gas relief product containing different forms of silicones. The 200 cP oil and the heating bath oil were also used to generate single-injection calibration curves to investigate their response factors, relative to the 1000 cP oil standard. These results are in Table 1.

49.48 50.00 50.50 51.00 51.50 52.00 52.50 53.00 53.50 54.00 54.50 54.815.5

6.0

7.0

8.0

9.0

10.0

11.0

12.0

13.0

14.0

15.0

16.0

17.017.4

min

pA

Page 6: Analysis of Silicone Oils by High Performance Liquid ... · Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection Marc Plante, Bruce

6 Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol DetectionMarc Plante, Bruce Bailey, Ian N. AcworthThermo Fisher Scientific, Chelmsford, MA, USA

OverviewPurpose: To develop HPLC methods for both the detailed,characterization of silicone oils and for rapid quantitation of silicone oils in samples using an HPLC system with a charged aerosoldetector.

Methods: Two methods, one qualitative for characterization and one quantitative using a solid core C18 column, are outlined.

Results: Different silicone oils were characterized using the qualitative method; silicone oils in commercial products were measured, with a quantitation limit of approximately 100 ng on column.

IntroductionSilicone oils are unique materials with multitudes of significant uses in our modern world. Silicone-based materials are thermally stable,relatively inert, and are generally non-toxic, making them useful asoils and greases, lubricants, anti-foaming agents, and coatings.Silicone oils are often used in electronics, hydraulic systems, and many consumer products. The measurement of silicone oils iscomplicated by their solubility properties, their variations in basicstructure including a wide variety of chemical moieties that may be part of this structure, as well as their general lack of chromophoresmaking detection by ultraviolet absorption impractical.

The charged aerosol detector is a sensitive, mass-based detector,especially well-suited for the determination of any nonvolatile analyte independent of chemical characteristics. As shown in Figure 1, the detector uses nebulization to create aerosol droplets. The mobile phase evaporates in the drying tube, leaving analyte particles, which become charged in the mixing chamber. The charge is then measured by a highly sensitive electrometer, providing reproducible, nanogram-level sensitivity. This technology has greater sensitivity and precision than ELSD and refractive index (RI), is gradient compatible and issimpler to operate than a mass spectrometer (MS). Compounds do not have to possess a chromophore (unlike UV detection) or be ionized (as with MS).

Two charged aerosol detection methods were developed: one to characterize the individual components of silicone oils, and the otherto provide quantitation of a near single peak of silicone oil to 100ng (o.c.). The characterization method used the Thermo Scientific™Dionex™ UltiMate™ 3000 RS system, and the quantitation method used an UltiMate 3000 SD system in normal phase. Both methodsused a solid core C18 column and the Thermo Scientific™ Dionex™Corona™ ultra RS™ charged aerosol detector.

The combination of these chromatographic methods with the capabilities of the charged aerosol detection provides unique opportunities for detailed characterization of a silicone oil sample, aswell as quantitation of silicone oil-containing samples.

.

FIGURE 1. Schematic and Functioning of Charged AerosolDetection

1

2

3

4

5

6

7

89

101

2

3

4

5

6

7

89

10

1. Liquid eluent enters from HPLC system2. Pneumatic nebulization occurs3. Small droplets enter drying tube4. Large droplets exit to drain5. Dried particles enter mixing chamber6. Gas stream passes over corona needle7. Charged gas collides with particles and

charge is transferred8. High mobility species are removed9. Charge is measured by a highly sensitive

electrometer10.Signal transferred to chromatographic

softwareMethodsSample Preparation - CharacterizationSamples of silicone oil were dissolved in chloroform, at a concentration of approximately 40 mg/mL.

Liquid Chromatography – CharacterizationHPLC System: Thermo Scientific™ Dionex™ UltiMate™ 3000

DGP-3600RS pump, WPS-3000RS autosampler, and TCC-3000RS column oven

HPLC Column: Thermo Scientific™ Accucore™ 2.6 µm C18,3.0 × 150 mm

Column Temp.: 40 °CMobile Phase A: MethanolMobile Phase B: n-PropanolFlow Rate: 0.5 mL/minInjection Volume: 2–10 µLSample: 38 mg/mL silicone oil in chloroformDetector: Corona ultra RSNebulizer Temp.: 10 °CFilter: 5Data Rate: 10 HzPower Function: 1.00Flow Gradient for Characterization:

Time (min) %A %B Curve

-10 100 10 5

0 100 10 5

80 15 85 3

85 0 100 5

95 0 100 5

95 100 0 5

100 100 0 5

Liquid Chromatography –QuantitationHPLC System: Thermo Scientific™ Dionex™ UltiMate™ 3000

LPG-3400SD pump, WPS-3000RS autosampler, and TCC-3000RS column oven

HPLC Column: Accucore 2.6 µm C18, 3.0 × 150 mmColumn Temp.: 40 °CMobile Phase A: 0.5% Formic acid/acetonitrile/tetrahydrofuran*

(35:35:30), helium spargeMobile Phase B: Tetrahydrofuran*, helium spargeFlow Rate: 0.3–1.0 mL/minInjection Volume: 2–10 µLDetector: Corona ultra RSNebulizer Temp.: 10 °CFilter: 6Data Rate: 10 HzPower Function: 2Flow Gradient for Quantitation:

*No BHT added.

Data AnalysisAll HPLC chromatograms were obtained and compiled using Thermo Scientific™ Dionex™ Chromeleon™ Chromatography Data Station software, 7.1 SR 1.

Time (min) Flow Rate(mL/min)

%A %B

-5.0 0.6 100 0

-0.2 0.6 100 0

-0.1 0.3 100 0

0.0 0.3 100 0

0.5 0.6 100 0

Time (min) Flow Rate(mL/min)

%A %B

4.0 0.6 100 0

6.0 0.6 0 100

8.0 0.6 0 100

10.0 1.0 0 100

12.0 1.0 50 50

14.0 0.6 100 100

ResultsSample AnalysisTo investigate relative responses of different oils, three differentsilicone oil samples were characterized consisting of two similar oilswith different viscosities (200 cP and 1000 cP) and a third oil that isused as a heating bath oil. From the overlay of chromatograms shown in Figure 2, the 1000 cP oil sample is comprised of a greater amount oflarger molecular weight polysiloxanes (later elution) than the 200 cP oil sample, as characterized by the bulk of peak area at higher retention times. The heating bath oil is comprised of more isolable polysiloxanechain groups, as identified by over 170 different peaks that were atleast partially resolved for this oil.

Retention time precision was demonstrated by repetitive analysis of the heating bath silicone oil sample (Figure 3). The UltiMate 3000 LCsystems, controlled by the Chromeleon software, enables sample injections to be synchronized with the pump delivery, allowing for exactand matching system conditions for each injection. The percent relative standard deviation for retention time at 50 minutes was 0.02%.

A silicone-based topical product used for comfort with hearing aids wascharacterized using this method, as shown by the chromatogram in Figure 4. The material appears to contain a moderate amount ofsmaller molecular weight polysiloxanes mixed with an amount of highermolecular weight polysiloxanes that elute at the end of the gradient.

The method provides a high-resolution HPLC chromatogram for the detailed characterization of polysiloxane materials. The use ofthe solid core column in this method achieves the high number oftheoretical plates and therefore, the high level of resolution shown in these figures. The charged aerosol detector provides both sensitive and consistent response for these analytes.

The near-consistent response of charged aerosol detection for non-volatile analytes enables the quantitation of total silicones in the same samples used above. The second method, specifically developed forquantitation, uses an aqueous acetonitrile mobile phase with a fasttetrahydrofuran gradient to elute the analyte from the column in a rapid manner, yielding a near-single peak. Chromatographic baseline subtraction is used to remove baseline deviations that result from the rapid gradient.

The 200 cP oil showed essentially equal response to that of the 1000 cPoil, but the heating bath oil differed in response at the two amounts used in this comparison. This may be associated with the differences of the groups or numbers of associated groups contained on the silicone. Thus,it is recommended that the silicone oil calibration standard is the same (orat least similar) to the silicone that is to be quantified. As seen in Figure 2,there is a clear difference between the 200 and 1000 cP silicones and the heating bath oil silicone.

FIGURE 5. Silicone oil calibration curves, from 78 to 10,000 ng (1000 cP)in triplicate, 2000 to 10,000 ng o.c. (200 cP and Heating Bath Oil)

0

2

4

6

8

10

12

0 2000 4000 6000 8000 10000 12000

Peak

Are

a (pA

*min)

Amount (ng o.c.)

AP1000

AP200

HeatingOil

TABLE 1. Response factors of different polysiloxanes at 5000 and2000 ng o.c. relative to 1000 cP silicone oil

Sample Amount on Column (ng)

Amount Found (ng) Relative Response

AP2005000 5031 1.0062000 1994 0.997

Heating Bath Oil5000 8681 1.7362000 4199 2.100

6.50 6.80 7.00 7.20 7.40 7.60 7.80 8.00 8.20 8.40 8.60 8.80 9.00 9.20 9.40 9.60 9.80 10.00-5.0

0.0

10.0

20.0

30.0

40.0

45.0 pA

min

21 1 - 6.980 2 - 7.4523 - 7.5224 - 7.665

5 - 7.954

6 - 8.0327 - 8.187

8 - SiliconeOil - 8.295

9 - 8.417

2 - SiliconeOil – 8.060

Two of the three oils (the two most closely related) showed a similarcalibration curve, indicating that calibration curves derived from oilsof a similar composition can be used for product concentration determinations. Differences in chain length and other structures mayaffect retention time. The heating bath oil, which characterized as a different oil, showed a greater response than the viscosity-rated oils.

FIGURE 6. HPLC chromatogram overlay of over-the-counter gasrelief product (blue), dissolved at a concentration of 500 µg/mLsimethicone in tetrahydrofuran, and 500 µg/mL of 1000 cP silicone oil standard (black)

*simethicone is a mixture of silica (5.5%) and dimethicone (94.5%)1

The measurement of silicone oils in different commercial products showed the potential for quantitative determinations without extraction of the silicone from the sample or having the exact silicone oil as a calibrant. Samples were dissolved, centrifuged, diluted to a targeted amount, and analyzed. A gas relieve product chromatogram is shown in Figure 6 with an overlay of 1000 cP silicone oil. The recovery for the gas relief product (prepared at 500 µg/mL of simethicone) was 96.5%, as shown in Table 2, demonstrating the potential for this method to measure silicone oils with unknown composition in different samples.

TABLE 2. Quantities of silicone oils found in products.

Sample Silicone Oil TypeDeterminedSilicone Oil

(ng o.c.)Product Silicone

AmountRecovery

(%)

Shampoo Dimethicone 3890 1.95 w/w-% --Conditioner Phenyltrimethicone 2702 2.70 w/w-% --Gas ReliefProduct

Simethicone* 4559 4725 ng o.c. 96.5

ConclusionsTwo methods were developed for the characterization and quantitation of silicone oils as standards and as part of products. The long characterization method can readily be modified to address specific regions of the chromatogram or to fully characterize a specific oil in less time. The quantification method is sensitive to 100 ng o.c. and it was shown that similar oils provide similar response factors. The method also had sufficient specificity to allow for analysis of different products and silicone oils.

The use of the solid core C18 column, combined with the detection capabilities of the charged aerosol detector, enables convenient and accurate characterization and quantification of silicone oils and silicone oil-containing products.

References1. DailyMedPlus website

http://www.dailymedplus.com/monograph/view/setid/da8eba22-33e1-4f46-b43c-9ca3d8789740 (last accessed 30 Jan 13).

© 2013 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property ofThermo Fisher Scientific Inc. and its subsidiaries. This information is not intended toencourage use of these products in any manners that might infringe the intellectual propertyrights of others.

PO738_E 03/13S

Sample and Standard Preparations - QuantitationSamples containing silicone oil were dissolved in tetrahydrofuran, at a product concentration of 10—20 mg/mL and centrifuged at 10,000 g for3 minutes. HPLC sample vials were filled with the supernatant.

Standards (Silicone oil 1000 cP) were prepared by dissolving 10 mg/mL of silicone oil in tetrahydrofuran, diluted to 1 mg/mL and then sequentially diluted to lower concentrations.

FIGURE 2. Characterization of three different silicone oils, 200 cP(blue) and 1000 cP (black) viscosities, and a heating bathsilicone oil (pink)

FIGURE 3. Overlays of three separate analyses of the heatingbath silicone oil sample. Only the 50 to 55 minute segment of the total run (inset) is shown for clarity.

FIGURE 4. HPLC chromatogram of a silicone-based topicalproduct. Product appears to be made of high-molecular weightpolysiloxane (85 minutes) mixed with moderate and low-molecular weight polysiloxanes. The inset shows the low to moderate molecular weight regions.

0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100-10.0

0.0

10.0

20.0

30.0

40.0

50.0

60.0

70.0

80.0

min

pA

0.0 10.0 20.0 30.0 40.0 50.0 60.0 70.0 75.0-2.0

0.0

2.0

4.0

6.0

8.0

10.0

12.0

14.0

16.0

18.0

min

pA

This quantitative method was evaluated using different silicone oils; using the 1000 cP silicone oil standards, a calibration curve was generated from78 to 10,000 ng o.c., as shown in Figure 5. System precision (n=3) wasacceptable, with peak area percent relative standard deviations 0.92 (10,000 ng o.c.) to 7.28 (156 ng o.c.). The quantitation limit is estimated atapproximately 100 ng o.c, based on a signal to noise ratio of 3.0.

Samples that were tested included a 200 cP silicone oil, the heating bath oil, and three consumer products, including a shampoo, a hair conditioner,and a gas relief product containing different forms of silicones. The 200 cPoil and the heating bath oil were also used to generate single-injection calibration curves to investigate their response factors, relative to the 1000 cP oil standard. These results are in Table 1.

49.48 50.00 50.50 51.00 51.50 52.00 52.50 53.00 53.50 54.00 54.50 54.815.5

6.0

7.0

8.0

9.0

10.0

11.0

12.0

13.0

14.0

15.0

16.0

17.017.4

min

pA

Page 7: Analysis of Silicone Oils by High Performance Liquid ... · Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection Marc Plante, Bruce

www.thermofisher.com©2016 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.

Australia +61 3 9757 4486 Austria +43 1 333 50 34 0 Belgium +32 53 73 42 41 Brazil +55 11 3731 5140 China +852 2428 3282

PN70538_E 07/16S

Denmark +45 70 23 62 60France +33 1 60 92 48 00Germany +49 6126 991 0India +91 22 6742 9494Italy +39 02 51 62 1267

Japan +81 6 6885 1213Korea +82 2 3420 8600Netherlands +31 76 579 55 55Singapore +65 6289 1190 Sweden +46 8 473 3380

Switzerland +41 62 205 9966Taiwan +886 2 8751 6655UK/Ireland +44 1442 233555USA and Canada +847 295 7500