463A Read this manual thoroughly before use Users Manual Analog Multimeter

Transcript of Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The...

Page 1: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used

463A

Read this manual thoroughly before use

Users Manual

Analog Multimeter

Page 2: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used
Page 3: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used

1

1

2

3

4

5

9

8

76

INTRODUCTIONThis meter is a rectifier type, permanent-magnet moving coilinstrument designed for DC and AC voltage, DC and AC current,resistance and temperature measurements.

STRUCTURE

Figure 1

Page 4: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used

2

1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement.

2. Trigger Used to open and close the jaws.

3. " V mA " Terminal Plug-in connector for the red test lead for all measurements except ac current measurements.

4. Pointer Zero Corrector Used for zeroing.

5. Pointer

6. 0 Adjuster Knob 1. Used for zeroing for resistance measurements. 2. Used for calibration for temperature measurements.

7. " COM " Terminal Plug-in connector for the black test lead for all measurements except ac current measurements.

8. Function/Range Switch Used to select the desired function or range.

9. Tactile Barrier Used to prevent finger from touching the conductor under test. Do not hold the meter anywhere beyond the tactile barrier.

Page 5: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used

3

To avoid possible electric shock or personal injury, follow theseguidelines:

Do not use the meter if it is damaged. Before you use the meter, inspect the case. Pay particular attention to the insulation surrounding the connectors. Inspect the test leads for damaged insulation or exposed metal.

Check the test leads for continuity. Replace damaged test leads before you use the meter. Do not use the meter if it operates abnormally. Protection may be

impaired. When in doubt, have the meter serviced. Do not operate the meter around explosive gas, vapor, or dust. Do not apply a voltage or current higher than the selected range's

upper limit between terminals. Before use, verify the meter's operation by measuring a known

voltage. When measuring dc current, turn off power to the circuit before

connecting the meter to the circuit. Remember to place the meter in series with the circuit. When servicing the meter, use only specified replacement parts. Use caution when working with voltage above 30V ac rms, 42V

Warning

SAFETY INFORMATIONThe meter has been designed according to IEC-61010 concerningelectronic measuring instruments with a measurement category III(CAT III 500V) and Pollution degree 2.

Page 6: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used

4

peak, or 60V dc. Such voltages pose a shock hazard. When using the probes, keep your fingers behind the finger

guards on the probes. Connect the black test lead before you connect the red test lead.

When you disconnect test leads, disconnect the red test lead first. Remove the test leads from the meter and remove the clamp jaws

from any clamped object before you open the battery cover or the case. Do not operate the meter with the battery cover or portion of the

case removed or loosened. To avoid electric shock, do not touch any naked conductor with

hand or skin and do not ground yourself. Adhere to local and national safety codes. Individual protective

equipment must be used to prevent shock and arc blast injury where hazardous live conductors are exposed. Do not hold the meter anywhere beyond the tactile barrier. Do not use the test leads with other equipment. Remaining endangerment:

When an input terminal is connected to dangerous live potential it is to be noted that this potential can occur at all other terminals! CAT III - Measurement Category III is for measurements

performed in the building installation. Examples are measurements on distribution boards, circuit breakers, wiring, including cables, bus-bars, junction boxes, switches, socket outlets in the fixed installation, and equipment for industrial use and some other equipment, for example, stationary motors with permanent connection to the fixed installation. Do not use the meter for measurements within Measurement Categories IV.

Page 7: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used

5

CautionTo avoid possible damage to the meter or to the equipment undertest, follow these guidelines:

Disconnect circuit power and discharge all capacitors before measuring resistance or temperature. Use the proper function and range for your measurements. Before measuring dc current, turn off power to the circuit before

connecting the meter to the circuit. Before rotating the rotary switch to change functions, disconnect

test leads from the circuit under test and remove the clamp jaws from any clamped object.

SymbolsAlternating CurrentDirect CurrentCaution, risk of danger, refer to the operating manual before use.Caution, risk of electric shock.Earth (ground) TerminalFuseConforms to European Union directivesThe equipment is protected throughout by double insulation orreinforced insulation.Application around and removal from HAZARDOUSLIVE conductors is permitted.

Page 8: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used

6

SPECIFICATION

Accuracy is specified for a period of one year after calibration and at18°C to 28°C, with relative humidity < 75%.

DC VoltageRanges: 10V - 50V - 500VAccuracy: ± 3% of Full Scale

AC VoltageRanges: 50V - 250V - 500VAccuracy: ± 3% of Full ScaleFrequency Response: 50Hz to 400Hz

DC CurrentRanges: 0.5mA - 10mA - 250mAAccuracy: ± 3% of Full Scale

AC CurrentRanges: 25A - 100A - 500A - 1000A - 2000AFrequency Response: 50Hz to 60HzAccuracy: ± 5% of Full Scale

ResistanceRanges: ×1 range --- 0 to 3k , scale center at 50

×100 range --- 0 to 300k , scale center at 5kAccuracy: 3% of arc

Page 9: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used

7

Temperature

Range: -20°C to 150°C (using the temperature probe)

Accuracy: 3% of arc

Jaw Opening Capability: 53mm (approx.)

Max. Measurable Conductor: Ø49mm (approx.)

Battery: 1.5V battery, AAA or equivalent, 3 pieces

Standard Applied: IEC61010-1 CAT III 500V

Operating Environment: temperature: 0°C to 40°C

relative humidity: 75%

Storage Environment: temperature: -10°C to 50°C

relative humidity: 85%

Size: 275 × 100 × 47mm

Weight: about 452g (including battery)

Page 10: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used

8

Test Range Position Scale To Read Multiplier

10V50V500V

B 10B 50B 50

× 1× 1× 10

DC voltage

AC voltage

DC current

50V250V500V

B 50B 250B 50

× 1× 1× 10

0.5mA10mA250mA

B 50B 10B 250

× 0.01× 1× 1

ABCD

REFERENCE TABLE FOR READING

Figure 2

Page 11: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used

9

Explanation for multiplier in the table:If you use " X100 " range in resistance measurement, you shouldmultiply the reading on scale C by 100, the result is the resistancevalue of the resistor under test.

CC

×1×100

× 1× 100

Temperature -20°C to 150°C D × 1

Resistance

Test Range Position Scale To Read Multiplier

25A100A500A1000A2000A

A 25A 100A 500A 100A 100

× 1× 1× 1× 10× 20

AC current

( continued )

OPERATION INSTRUCTION

Preparation for Measurements

Use a small flat tip screwdriver to adjust the Pointer Zero Correctoruntil the pointer is directly over the zero at the left end of the scale A( or scale B ).

Page 12: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used

10

1. Connect the black test lead to the " COM " terminal and the red test lead to the " V mA " terminal.2. Set the range switch to ×1 or ×100 position.3. Turn the 0 adjuster knob to set the pointer to the " 0 " position of the scale C while shorting the test leads. Note: If the pointer can't be adjusted to the " 0 " position of the the scale C, replace the batteries.4. Make sure that the object to be tested is without power.5. Connect the test leads across the object and read the scale C referring the reference table.

Measuring Resistance

Figure 3

Page 13: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used

1. Connect the black test lead to the " COM " terminal and the red test lead to the " V mA " terminal.2. Set the range switch to the desired range position. If the magnitude of the voltage to be measured is not known beforehand, set the range switch to the highest range position first and then reduce it range by range if necessary.3. Connect the red test lead to the positive terminal of the circuit to be tested and the black one to the negative terminal of the circuit. Note: If the pointer deflects to the left of the scale B, reverse the test leads.4. Read the scale B referring the reference table.

11

Measuring DC Voltage

Figure 4

Page 14: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used

12

1. Connect the black test lead to the " COM " terminal and the red test lead to the " V mA " terminal.2. Set the range switch to the desired range position. If the magnitude of the voltage to be measured is not known beforehand, set the range switch to the highest range position first and then reduce it range by range if necessary.3. Connect the test leads across the circuit to be tested.4. Read the scale B referring the reference table.

Measuring AC Voltage

Figure 5

Page 15: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used

Measuring DC Current

1. Connect the black test lead to the " COM " terminal and the red test lead to the " V mA " terminal.2. Set the range switch to the desired range position.3. Turn off power to the circuit you want to measure. Then discharge all capacitors.4. Break the circuit path to be measured, then connect the red test lead to the more positive side of the break and the black test lead to the more negative side of the break.5. Turn on power to the circuit, then read the scale B referring the reference table. ( If the pointer deflects to the left of the scale, reverse the test leads. )

13

Figure 6

Page 16: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used

Note :If the magnitude of the current to be measured is not knownbeforehand, set the range switch to the highest range position firstand then reduce it range by range if necessary.

Measuring AC Current

14

1. Remove all test leads from the meter. Then set the range switch to the desired range position.2. Press the trigger and clamp the jaws around the wire to be measured. Make sure that the jaws are perfectly closed. Note: a. Each time only one wire should be clamped.

Figure 7

Page 17: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used

15

b. The wire should be in the center of the jaws in order to get an accurate reading.3. Read the scale A referring the reference table.

Note:1. For measurements > 400A, measurement duration must not exceed 30 secs and interval between measurements should be more than 5 minutes.2. When you measure large current, the jaws may emit a faint sound at their joint. This is normal and does not affect measurement accuracy.

Measuring Temperature

Object

Temperature Probe

Figure 8

Page 18: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used

16

1. Set the range switch to the ×100 ( Temp. ) position.2. Connect the black test lead to the " COM " terminal and the red test lead to the " V mA " terminal. Short the two test leads together and adjust the 0 adjuster knob until the pointer is directly over the " CAL " scale mark on the scale D.3. Remove the test leads. Then insert the black plug of the Temperature Probe into the " COM " terminal and the red plug of the Temperature Probe into the " V mA " terminal.4. Touch the probe tip of the Temperature Probe to the object to be measured, read scale D when they reach thermal equilibrium.

Note:1. Do not make measurement on any live conductor.2. To avoid heat injury, do not touch hot probe with hand or skin.2. Do not measure any object whose temperature exceeds 150°C.

BATTERY AND FUSE REPLACEMENTConnect the black test lead to the " COM " terminal and the red testlead to the " V mA " terminal. If the pointer can not be set to " 0 "position of the scale C by turning the 0 adjuster knob while therange switch is in ×1 or ×100 position and the test leads are beingshorted, the batteries are low and should be replaced.

To replace batteries, remove the screw on the battery cover andremove the battery cover, replace the exhausted batteries with newbatteries of the same type ( 1.5V battery, AAA or equivalent ), makesure that the polarity connections are correct. Reinstall the batterycover and the screw.

Page 19: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used

17

If the meter can not work, you should replace the fuse.To replace fuse, remove the screw on the battery cover and removethe battery cover, then remove the screws on the back cover andremove the back cover, replace the blown fuse with a new one of thesame ratings. Reinstall the back cover, the battery cover and all thescrews.

The meter uses one fuse: 250mA/500V, FAST, Ø6.35X32mm

ACCESSORIESManual: 1 pieceTest Lead: 1 pair

PRESENTTemperature Probe: 1 piece

NOTE1. This manual is subject to change without notice.2. Our company will not take the other responsibilities for any loss.3. The contents of this manual can not be used as the reason to use the meter for any special application.

Page 20: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used

V120425

DISPOSAL OF THIS ARTICLE

Dear Customer,If you at some point intend to dispose of this article,then please keep in mind that many of its componentsconsist of valuable materials, which can be recycled.

Please do not discharge it in the garbage bin, but checkwith your local council for recycling facilities in your area.

18

Page 21: Analog Multimeter...2 1. Jaws Used to clamp the conductor whose ac current is to be measured. The conductor should be in the center of the jaws during measurement. 2. Trigger Used