38 OSTOMY WOUND MANAGEMENT ® FEBRUARY 2015 www.o-wm.com FEATURE An Overview of Polyurethane Foams in Higher Specification Foam Mattresses Esa Soppi, MD, PhD; Juha Lehtiö; and Hannu Saarinen, MSc (Tech), BSc Abstract Soft polyurethane foams exist in thousands of grades and constitute essential components of hospital mattresses. For pressure ulcer prevention, the ability of foams to control the immersion and envelopment of patients is essential. Higher specification foam mattresses (ie, foam mattresses that relieve pressure via optimum patient immersion and envelop- ment while enabling patient position changes) are claimed to be more effective for preventing pressure ulcers than standard mattresses. Foam grade evaluations should include resiliency, density, hardness, indentation force/load de- flection, progressive hardness, tensile strength, and elongation along with essential criteria for higher specification foam mattresses. Patient-specific requirements may include optimal control of patient immersion and envelopment. Mattress cover characteristics should include breathability, impermeability to fluids, and fire safety and not affect mattress func- tion. Additional determinations such as hardness are assessed according to the guidelines of the American Society for Testing and Materials and the International Organization for Standardization. At this time, no single foam grade provides an optimal combination of the above key requirements, but the literature suggests a combination of at least 2 foams may create an optimal higher specification foam mattress for pressure ulcer prevention. Future research and the development of product specification accuracy standards are needed to help clinicians make evidence-based decisions about mat- tress use. Keywords: review, mattresses, polyurethanes, pressure ulcer, beds Index: Ostomy Wound Management 2015;61(2):38–46 Potential Conflicts of Interest: Dr. Soppi is Medical Director and Chairman of the Board; Mr. Lehtiö is Director, Research and Development; and Mr. Saarinen is Chief Executive Officer, Carital Ltd, Helsinki, Finland. Dr. Soppi is Medical Director and Chairman of the Board; Mr. Lehtiö is Director, Research and Development; and Mr. Saarinen is Chief Executive Officer, Carital Ltd, Helsinki, Finland. Please address correspondence to: Esa T. Soppi, MD, PhD, Carital Ltd, Tilketori 2a A16, Helsinki 00700 Finland; email: esa.soppi@ mitconsulting.fi. G lobally, most mattresses used in health care institutions are made of polyurethane foam; an extensive variety of different soft polyurethane foam grades are available for mattress manufacture. 1,2 Support surface manufacturers have taken advantage of this variety, producing countless types of single-layer and multiple-layer foam mattresses for the market. In 2009, the National Pressure Ulcer Advisory Panel (NPUAP) and the European Pressure Ulcer Advisory Panel (EPUAP) 3 indicated higher specification foam mattresses (HSFMs) are more effective for preventing pressure ulcers (PU) than standard hospital foam mattresses and that higher specification mattresses should be used for all individuals at risk for PU. However, defining what a HSFM should incor- porate remains difficult, and attempts to do so have yielded conflicting results. 4-6 To the authors’ knowledge, no study in- volving foam mattresses focuses on more than 2 character- istics of polyurethane foams — usually, density and relative hardness — which relate to foam type (conventional or vis- coelastic). 4-11 A lack of robust trials involving different foam mattresses leaves scientists and manufacturers seeking agree- ment on how to categorize different foams and foam mat- tresses as well as on requirements for designation as HSFM. 12 To address an obvious need, the authors conducted a search of the literature on mattress design, mattress manu- facturing, and PU development and assimilated their find- ings with nearly 100 years of cumulative combined experi- ence in the field of polyurethane foam production and work with numerous foam manufacturers. Guideline and Literature Review To complement currently available review articles and guidelines on polyurethane-based support surfaces, 3,6,12-16 the terms polyurethane or foam or flexible or density or resil- ience or mattress or support or surface were used in database searches of the Web of Science, Scopus, Medline/Ovid, and PubMed, which yielded 1,301 unique hits of material pub- lished since 1970. Among the search results were 547 English abstracts, from which 85 full-text articles (most previously DO NOT DUPLICATE

Transcript of An Overview of Polyurethane Foams in Higher Specification ...Polyurethane foams. Soft polyurethane...

Page 1: An Overview of Polyurethane Foams in Higher Specification ...Polyurethane foams. Soft polyurethane foams are com-plex, plastic-like materials that result from chemical reactions of

38 OSTOMY WOUND MANAGEMENT® FEBRUARY 2015 www.o-wm.com

FEATURE

An Overview of Polyurethane Foams in Higher Specification Foam Mattresses Esa Soppi, MD, PhD; Juha Lehtiö; and Hannu Saarinen, MSc (Tech), BSc

AbstractSoft polyurethane foams exist in thousands of grades and constitute essential components of hospital mattresses. For pressure ulcer prevention, the ability of foams to control the immersion and envelopment of patients is essential. Higher specification foam mattresses (ie, foam mattresses that relieve pressure via optimum patient immersion and envelop-ment while enabling patient position changes) are claimed to be more effective for preventing pressure ulcers than standard mattresses. Foam grade evaluations should include resiliency, density, hardness, indentation force/load de-flection, progressive hardness, tensile strength, and elongation along with essential criteria for higher specification foam mattresses. Patient-specific requirements may include optimal control of patient immersion and envelopment. Mattress cover characteristics should include breathability, impermeability to fluids, and fire safety and not affect mattress func-tion. Additional determinations such as hardness are assessed according to the guidelines of the American Society for Testing and Materials and the International Organization for Standardization. At this time, no single foam grade provides an optimal combination of the above key requirements, but the literature suggests a combination of at least 2 foams may create an optimal higher specification foam mattress for pressure ulcer prevention. Future research and the development of product specification accuracy standards are needed to help clinicians make evidence-based decisions about mat-tress use.

Keywords: review, mattresses, polyurethanes, pressure ulcer, beds

Index: Ostomy Wound Management 2015;61(2):38–46

Potential Conflicts of Interest: Dr. Soppi is Medical Director and Chairman of the Board; Mr. Lehtiö is Director, Research and Development; and Mr. Saarinen is Chief Executive Officer, Carital Ltd, Helsinki, Finland.

Dr. Soppi is Medical Director and Chairman of the Board; Mr. Lehtiö is Director, Research and Development; and Mr. Saarinen is Chief Executive Officer, Carital Ltd, Helsinki, Finland. Please address correspondence to: Esa T. Soppi, MD, PhD, Carital Ltd, Tilketori 2a A16, Helsinki 00700 Finland; email: [email protected].

Globally, most mattresses used in health care institutions are made of polyurethane foam; an extensive variety

of different soft polyurethane foam grades are available for mattress manufacture.1,2 Support surface manufacturers have taken advantage of this variety, producing countless types of single-layer and multiple-layer foam mattresses for the market. In 2009, the National Pressure Ulcer Advisory Panel (NPUAP) and the European Pressure Ulcer Advisory Panel (EPUAP)3 indicated higher specification foam mattresses (HSFMs) are more effective for preventing pressure ulcers (PU) than standard hospital foam mattresses and that higher specification mattresses should be used for all individuals at risk for PU. However, defining what a HSFM should incor-porate remains difficult, and attempts to do so have yielded conflicting results.4-6 To the authors’ knowledge, no study in-volving foam mattresses focuses on more than 2 character-istics of polyurethane foams — usually, density and relative hardness — which relate to foam type (conventional or vis-coelastic).4-11 A lack of robust trials involving different foam

mattresses leaves scientists and manufacturers seeking agree-ment on how to categorize different foams and foam mat-tresses as well as on requirements for designation as HSFM.12

To address an obvious need, the authors conducted a search of the literature on mattress design, mattress manu-facturing, and PU development and assimilated their find-ings with nearly 100 years of cumulative combined experi-ence in the field of polyurethane foam production and work with numerous foam manufacturers.

Guideline and Literature ReviewTo complement currently available review articles and

guidelines on polyurethane-based support surfaces,3,6,12-16 the terms polyurethane or foam or flexible or density or resil-ience or mattress or support or surface were used in database searches of the Web of Science, Scopus, Medline/Ovid, and PubMed, which yielded 1,301 unique hits of material pub-lished since 1970. Among the search results were 547 English abstracts, from which 85 full-text articles (most previously

DO NOT D

UPLICATE

Page 2: An Overview of Polyurethane Foams in Higher Specification ...Polyurethane foams. Soft polyurethane foams are com-plex, plastic-like materials that result from chemical reactions of

FEBRUARY 2015 OSTOMY WOUND MANAGEMENT® 39www.o-wm.com

HIGHER SPECIFICATION FOAM MATTRESSES

familiar) were reviewed in more detail, yielding 2 studies1,17 that contributed new data to this study. Altogether, 41 of the retrieved references are cited in this study; the remainder did not add meaningful information.

Polyurethane foams. Soft polyurethane foams are com-plex, plastic-like materials that result from chemical reactions of 2 main components, polyol(s) and isocyanate(s), with water. Polyurethane also contains a number of other com-ponents that act as catalysts in the manufacturing process, as well as surfactants, fillers, and fire retardants.1,2 Once the components are mixed, the reaction starts immediately: foam and bubbles are formed and the mixture expands. The prop-erties of polyurethanes also depend on external conditions such as temperature and humidity of the ambient air during manufacture, although no clear scientific understanding ex-ists of how the properties of foams are affected by changes in the formulation process. Therefore, and due to numerous compounds involved, experimental studies2,18 have shown it is difficult to reproduce identical polyurethane foams from batch to batch and (based on the experience of foam manu-facturers) a production tolerance of ± 10% is an extremely demanding requirement for any foam manufacturer.

Variations in these multiple ingredients generate an enor-mous number of different soft polyurethane foam grades. The soft polyurethane foams are composed of an internal cell/bubble structure that is usually “broken” or open.1 These foams are made up of a network of cellular shapes comprised of tiny struts and cell windows (see Figure 1). The size of the bubbles, the wall thickness, and the bubble structure affect the density and other properties of the foam. The open cells allow air to move slowly through the foam (to “breathe”), the amount of which is expressed in cubic feet per minute and is measured as the volume that can be drawn through a 2- x 2- x 1-inch polyurethane foam sample at a 0.5-inch water pressure differential.18 An industrial standard for adequate airflow in soft polyurethane foams is at least 2.0 cubic feet (55 L) per minute.18

Main properties of flexible polyurethane foams. The methods and measures to classify polyurethane foams dif-fer from continent to continent. Industrial standards18 and guidelines are written by the American Society for Testing and Materials (ASTM) and the International Organization for Standardization (ISO).

Resilience. Resilience is an indicator of the surface elasticity or “springiness” of foams.18 Resilience is typically measured by dropping a standardized steel ball onto a specific sized foam sample and measuring how high the ball rebounds.19,20 Foam resilience ranges from about 30% ball rebound to >60% rebound in high resilience (HR) foams.18,21 Conven-tional resilience in soft, open cell foam by industry standards corresponds to a ball rebound of 20%/30%-60%.19-21

Density. Density (kg/m3) is one of the most important properties of foams.1,2,18,22,23 Density is a function of the chemistry used to produce the foam and of the additives used

during synthesis. The density of the foam affects not only its physical properties in a complex fashion, but also the foam’s durability and support.1,2,18 Typically, the higher the polymer density, the better the foam will retain its original properties; it will last longer and provide the support it was originally designed to produce.1,2,22 The densities of HR foams, visco-elastic foams, and conventional foams are typically >40 kg/m3, 30 to 100 kg/m3, and 25 to 65 kg/m3, respectively.2,18,21

Hardness. Compression load deflection (CLD, recorded in kPa) — the technical reference to “hardness” — is measured as the force required to compress a defined piece of foam a defined percent of the original with a compression plate that is larger than the piece of foam (see Figure 2).18,23,24 The foam is compressed 3 times to 70% and after the third compression readings are taken at 60%, 40%, or 25% of compression.24 Foam hardness can be controlled during production by vary-ing the chemical formulations and processing techniques. CLD hardness is independent of density. A high-density foam can be produced in such a way that it may have a low or high CLD value.1,2,25 The CLD hardness reflects how deep the patient sinks into the foam; the higher the immersion, the lower the CLD value and vice versa. Furthermore, higher

Key Points• The authors review the characteristics of various

foam mattresses and desirable characteristics for pressure ulcer prevention.

• At this time, the literature suggests no single foam grade can meet these characteristics.

• The authors conclude research on foam combina-tions is promising and product specification accuracy for all mattresses is needed.

Ostomy Wound Management 2015;61(2):38–46

Figure 1. Scanning electron micrograph of the open cell structure of a soft polyurethane foam (Courtesy of Herman-Joseph Christen, Kabelwerk Eupen AG).

DO NOT D

UPLICATE

Page 3: An Overview of Polyurethane Foams in Higher Specification ...Polyurethane foams. Soft polyurethane foams are com-plex, plastic-like materials that result from chemical reactions of

40 OSTOMY WOUND MANAGEMENT® FEBRUARY 2015 www.o-wm.com

FEATURE

CLD hardness values reflect the foam’s ability to support the body and to prevent the bottoming out.

Indentation force/load deflection and progressive hardness. Indentation force/load deflection (IFD or ILD, recorded in Newtons) — referred to as “firmness” or “stiffness” — is de-fined as the force required to compress a defined piece of foam with a compression plate that is smaller than the piece of foam.18,26,27 IFD/ILD is independent of foam density, al-though it is often thought higher density foams are firmer than low density foams,18,26 and it is possible to produce high-density foams that are less firm and lower density foams that are firm.1,2,27,28 IFD reflects the surface feel of the foam and is a measure of the immersion and envelopment characteristics of the foam (see Figure 3).

Progressive hardness (support factor, or the SAG-index [from the verb sag]) is measured by dividing the force needed to indent the foam 65% of its original height by the force needed to indent the foam 25% of its original height (see Figure 4).20,27,29 In high-resilience foams, the SAG-index is >2.4.18,21 The SAG-index reflects how well the foam envelops the body contours; the better envelopment, the lower the SAG index and vice versa. Furthermore, higher SAG index values reflect the foam’s ability to support the body and to prevent the bottoming out.

Tensile strength. Tensile strength describes the durability of the foam and is defined as the maximum force at the point of rupture when a piece of foam is subjected to the pull of op-posite forces. The result is usually expressed in PSI, kg/cm2, or kPa.20,28,39 German reimbursement standards require a tensile strength of all types of mattress foams of >80 kPa.18,31

Elongation. Elongation describes the durability of the foam and is defined as the length of elongation at the break point recorded as a percent of the original length.18,20,30 The

elongation at the breaking point of all types of medical use mattress foams must be >115%.18,31

For the clinician, these characteristics reflect the high quality and durability of the foam in mattress.

Viscoelastic foam (ie, differentiating viscoelastic foam from other foams). When the human body is positioned on viscoelastic foam, the foam progressively conforms to the shape of the body, and after the weight is removed, the foam usually slowly reassumes its initial shape; this is a type of “slow recovery” foam.26 The physical properties of visco-elastic foam can be greatly influenced by temperature. In testing viscoelastic foam performance characteristics, it is also important to make notation of the ambient tempera-ture and be certain all comparison testing is performed under like conditions. The sample must be conditioned at specified temperature and humidity before testing.18 Even slight changes in room temperature can affect the measured

Figure 2. Compression load deflection (CLD). CLD is used to measure hardness by compressing a defined piece of foam with a compression plate that is larger than the piece of foam.18,24,26

Figure 3. Schematic representation of immersion:a) no immersion of the body with high compression load deflection (CLD) hardness mattress; b): immersion and envelopment with typical of medium density foam and low CLD hardness mattress; c) limited envelopment ca-pability typical of a standard/conventional foam mat-tress; d) more optimized envelopment, immersion, and support capability typical of a 2-layer higher specifica-tion foam mattress.

DO NOT D

UPLICATE

Page 4: An Overview of Polyurethane Foams in Higher Specification ...Polyurethane foams. Soft polyurethane foams are com-plex, plastic-like materials that result from chemical reactions of

FEBRUARY 2015 OSTOMY WOUND MANAGEMENT® 41www.o-wm.com

HIGHER SPECIFICATION FOAM MATTRESSES

firmness and recovery rates. Depending on the formulation, some viscoelastic foams maintain their “memory” feature at temperatures as low as 30˚ F (0˚ C), but the optimum range for the best “memory” action is between 55˚ F (13˚ C) and 85˚ F (30˚ C).18

Viscoelastic foam is characterized by a ball rebound of <20% (ie, this type of foam exhibits low resilience [LR]). In fact, certain viscoelastic foams can absorb approximately 90% of the ball impact.

The optimal combination is defined by the application, but for high-resiliency and viscoelastic foams the densities vary from >40 kg/m3 to >50 kg/m3 and CLD hardness from 4–5 kPa to 1–3 kPa, respectively.

Due to its conforming aspect and low resilience, viscoelas-tic material provides a comfortable yet supportive mattress. People with impaired mobility can benefit from the foam’s capacity to redistribute weight and surface pressure, poten-tially reducing PU development risk.4-6,13,14,16 Unfortunately, production economics often restrict end-use applications. Unlike traditional foam processing, viscoelastic formulation flexibility is more restrictive. Raw materials must be carefully altered to generate foam with varying properties. Cutting, profiling, and other fabrication techniques also may require more care due to the slow recovery aspect of the foam.2,22 These considerations affect product economics.

Patient issues with pressure and physiological change. Pressure causes stress in the tissues and deforms the skin, fat, and muscle. It affects blood flow and metabolism of the tissues subjected to pressure32,33; the latter have different elastic modu-lus characteristics: resistance of tissues to become deformed elastically (ie, nonpermanently when a force is applied to it), and pressure causes shear between and within the tissues. The magnitude of the shear depends on the person’s characteristics,

such as the anatomical configuration of the bony prominences and the elasticity of the skin and tissues.3,34 These concerns have been examined in experimental and clinical studies.3,32-34

Support surfaces should be designed to manage tissue loads and stress reactions to pressure, shear force, temper-ature, and microclimate by redistributing the pressure and managing the interactions between the tissues and support surface combined immersion and envelopment proper-ties. When a person lays or sits on a support surface, his/her weight deforms both the support surface (typical property of viscoelastic foam) and his/her soft tissues.32 As a patient sinks (immerses) into the support surface, equalization of pressure off bony prominences should occur. Bottoming out needs to be prevented; this can be achieved by increasing the CLD hardness, but equalization properties of the foam will be lost (see Figure 3a). Increasing the density and height may help to some extent, but also increases the weight of the mattress. Furthermore, the height of the mattress can be increased or multiple layers of different types of support surface need to be introduced. However, as the patient sinks deeply into a very deep mattress, the property of deep immersion of the mattress (a disadvantage of viscoelastic foam) may make it difficult to change positions (see Figure 3). Furthermore, the “memory” function (ie, the slow recovery of original confor-mation of viscoelastic foams) is hypothesized to cause shear forces in the tissues when the patient is repositioned by the nursing staff because the original immersion configuration in the mattress will remain long after repositioning.

The support surface’s envelopment capability (ie, capabil-ity to conform to the shape of the person) describes its ability to deform around irregularities of the body without causing any substantial pressure increase (see Figure 5). Furthermore, the pressure of the individual’s body will be more evenly dis-tributed and the strain on bony prominences, where PUs typically develop, alleviated.3,12,22,35 This may reduce internal tissue pressure, strain, and stress to a great extent.33 The ex-tent to which small areas of the body are subjected to high pressure will determine the degree of damaging deformation and physiological changes.16,32,35

Circulatory and metabolic changes result from the pres-sure exerted on the tissues.32,33,36 As shown in clinical experi-mental studies,36,37 pressure-induced stress in the tissues will change the blood flow in the microvasculature and decrease tissue oxygen tension, irrespective of the type of mattress the patient is using. Interindividual variation in this change in ox-ygen supply is enormous.33,36,38 The counterbalancing physio-logical effects aim to maintain tissue oxygen levels, and these effects are markedly influenced by the support surface (see Figures 5 and 6).37,38 In one randomized controlled study,37 10 healthy volunteers were placed on a conventional, nonvis-coelastic foam (35 kg/m3) or dynamic (the microchip-guided control unit with pumps, functioning for approximately 1 hour/day. This nonalternating air mattress [Carital® Optima, Carital Ltd, Helsinki, Finland] adjusts automatically to the

Figure 4. Indentation force/load deflection (IFD/ILD) firm-ness is measured by compressing a defined piece of foam with a compression plate that is smaller than the piece of foam. SAG-index = IFD at 65% divided by IFD at 25%.18,20,27

DO NOT D

UPLICATE

Page 5: An Overview of Polyurethane Foams in Higher Specification ...Polyurethane foams. Soft polyurethane foams are com-plex, plastic-like materials that result from chemical reactions of

42 OSTOMY WOUND MANAGEMENT® FEBRUARY 2015 www.o-wm.com

FEATURE

pressure values of the air within it to predefined levels irre-spective of patient’s weight, body shape, or position). Mean tissue oxygen tension39 (see Figure 6a), capillary flow using Doppler37 (see Figure 6b), and skin temperature (see Figure 6c) were measured. Although the mean tissue oxygen tensions were similar on both mattresses, the standard mattress caused the capillary flow to increase throughout the whole 2 hours of exposure to pressure. This increase was due to the following mechanism: As the oxygen consumption and temperature rose in the tissue exposed to pressure, the tissue reacted by increas-ing blood flow in an effort to uphold a stable oxygen pressure in the tissue. Naturally, this compensatory effect cannot con-tinue indefinitely on a standard mattress, because the capillary flow must even out, at some point in time, and this will result in a drop in tissue oxygen tension. The dynamic, nonalternat-ing air mattress has shown optimized immersion and envelop-ment capabilities40 and to be highly effective for prevention of pressure ulcers among extremely sick intensive care patients in a randomized, controlled trial.6,41

These findings suggest tissues exposed to sufficiently high local pressures exhaust their compensatory blood flow re-serves and this may lead to ischemic tissue injury and PU formation. However, this did not occur in healthy volun-teers (the ethical committee had limited the test period for 2 hours).37,38 This effect will take place even in healthy indi-viduals if the pressure or exposure time is sufficiently long.12 These concerns become even more crucial in severely ill peo-ple who have fever and compromised blood flow.37,41 Taken together, the results suggest that optimal control of patient immersion and envelopment on any mattresses are crucial for optimal tissue perfusion and oxygenation.

Figure 5. Sacral area tissue oxygen tension in healthy volunteer placed on conventional, non-viscoelastic, polyurethane foam (35 kg/m3) mattress.38 Footnote: Change in position after 120 minutes. Reprinted with permission of the Finnish Medical Journal Duodecim.

Figure 6. Physiological changes due to pressure exert-ed on the tissues.37,38 Skin temperature from 30 minutes onwards, P <0.01 to <0.001 (Figure 6c).Footnote: Reprinted with permission

DO NOT D

UPLICATE

Page 6: An Overview of Polyurethane Foams in Higher Specification ...Polyurethane foams. Soft polyurethane foams are com-plex, plastic-like materials that result from chemical reactions of

FEBRUARY 2015 OSTOMY WOUND MANAGEMENT® 43www.o-wm.com

HIGHER SPECIFICATION FOAM MATTRESSES

Higher Specification Foam MattressesCurrently, the distinguishing characteristics of HSFMs are

not well described. Outcomes of patients placed on standard hospital foam mattresses and other foam mattresses are compared in a Cochrane review6 covering 7 studies,3,4,42-46 5 of which are randomized trials.4,42-46 The studies included in the Cochrane review are detailed in Table 1. The studies that compared standard and alternative foam mattresses varied in quality and all failed to adequately define standard hospital

and higher specification foam mattresses. Furthermore, in some of the studies higher specification mattresses were used as standard mattresses in other studies (see Table 1). Still, the meta-analysis3,6,12 concluded HSFMs reduced the incidence of PU-related events for individuals at risk to less than half of the incidence among standard mattress users (risk ratio = 0.41, P = 0.021).3,6

The key randomized, controlled study by Russell et al4 in-volved elderly patients requiring acute care in orthopedic and

Table 1. Summary of trials investigating the performance of standard and higher specification foam mattresses (HSFM)

Patient characteristics

Risk scale (risk score)

Standard foam mattress

properties (N)

Higher specification foam mattress, HSFM (N)

Primary outcome P value

Hofman et al 199442

Patients with femoral neck

fracture

Dutch consensus scale (score >8)

Standard SG40 hospital mat-tressa (N=23)

Comfortex DeCube mattressa (N=21)

Pressure ulcer incidence. Standard: 68% (13/19)

versus HSFM 24% (4/17)

Not significant

Gray and Campbell 199443

Orthopedic trauma, vascular and medical on-cology patients

without breaks in the skin

Waterlow (score >15)

Standard NHS foam mattressa (thickness 130 mm) (N=80)

Soft foam mattressa (N=90)

Pressure ulcer incidence. Standard 33.8% versus

HSFM 6.7%

<0.001

Santy et al 199444

Patients >55 years of age with hip fracture, with or without pres-

sure ulcers

Standard foama (NHS contract

surface) (N= 64)

Clinifloata, Trans-foama, Theraresta,

Vaperma mattresses (N=441)

Pressure ulcer incidence. Standard 26.5% versus

HSFM 9.5%

P =0.002

Collier 199645

Patients attend-ing at a general medical ward

Standard hospi-tal mattress (Re-lyon)a (thickness 130 mm) (N=9)

Clinifloata, Omni-foama, Softfoama, STM5a, Theraresta,

Transfoama, Vapour-luxa mattresses

(N = 81)

Pressure ulcer incidence. No pressure ulcers in any

of the mattresses

Gray and Smith 200046

Surgical, ortho-pedic, and medi-

cal patients

Waterlow (mean score 14 for stan-dard; score for 13

HSFM)

Transfoama (N=50)

Transfoamwavea (N=50)

Pressure ulcer incidence. 2% in both groups. In

addition, nonblanching erythema 2% in both

groups

Not significant

Russell et al 20034

Elderly acute care, rehabilita-tion, and ortho-pedic patients

Waterlow(score 15-20)

Numerous stan-dard foam mat-tresses (King’s

Funda, Soft-foama, Trans-

foama, Linknur-sea, or a King’s Fund mattress

with a Spenco or Propad mattress overlay) (N=604)

CONFOR-Med: 3 inch (7.5cm) vis-

coelastic foam on the top and 3 inch (7.5cm) standard

polyurethane foam on the bottoma

(N=562)

Incidence of nonblanching erythema or worse Intact skin with blanching ery-thema: standard (N=161)

26.6% versus HSFM (N=110/19.6%. Intact skin

with nonblanching ery-thema: standard (N=66)

19.6% versus HSFM (N=48) 9.5%

P =0.004 for all patients and P =0.005 for those with-

out blanching erythema on trial

admission

Berthe et al 20075

Medical and surgical patients

Modified Ek’s scale (on Kliniplot 14.6% of patients the mobility score

1 or 2, on stan-dard mattress the

corresponding group included

11.0% of patients)

Numerous stan-dard mattresses.

No specifica-tions available

(N=1072)

Kliniplot: Foam (Bultex) mattress with block struc-

turea (height 18 cm) (N=657)

Pressure ulcer incidence. Standard 1.9% versus Kliniplot 3.2%. Time to

pressure ulcer 31 days for Kliniplot versus 18 days for conventional foam

mattress

P=0.154 for pressure ulcer

incidence. P<0.001 for the time to pressure ulcer develop-

ment

a Not known to be in production any more. No other specifications, as shown in the table, are available

DO NOT D

UPLICATE

Page 7: An Overview of Polyurethane Foams in Higher Specification ...Polyurethane foams. Soft polyurethane foams are com-plex, plastic-like materials that result from chemical reactions of

44 OSTOMY WOUND MANAGEMENT® FEBRUARY 2015 www.o-wm.com

FEATURE

rehabilitation wards. The support surfaces included standard foam hospital mattresses (see Table 1) and a HSFM com-posed of a 3-inch (7.5 cm) layer of standard foam on the bot-tom and a 3-inch (7.5 cm) layer of viscoelastic foam on the top. Use of the HSFM was associated with a significant reduc-tion (P = 0.004) in the incidence of blanching erythema (pri-mary outcome measure) and a nonsignificant decrease in the incidence of nonblanching erythema (10.9% versus 8.5%, P = 0.17). A statistically significant decrease was noted in all nonblanching erythema groups in favor of HSFM at day 7 (P = 0.0015 and P = 0.042, respectively).4

Thus, currently no consensus based on published studies and reviews has been reached on the requirements of HS-FMs.12 A HSFM needs to fulfill 3 essential functional speci-fications: 1) it must possess optimum immersion and en-velopment properties; 2) the patient must be able to change position; and 3) it must be easy for the patient to change po-sition or for the nursing staff to move the patient. At present, there is no evidence a single foam grade provides an optimal combination of these key specifications. The specifications may be achieved only by combining at least 2 layers of dif-ferent types of foams, as suggested by the results reported by Russell et al.4 Table 2 summarizes the main features of higher

specification foam mattress requirements as based on this review. The minimum requirements of polyurethane foam qualifications rely on the properties of different grades of polyurethane foams and on expert experience on the behav-ior of different grades of foams used in support surfaces. A thin layer of viscoelastic foam on the top utilizes its beneficial properties and simultaneously eliminates its disadvantages (deep immersion and slow recovery) when thick support surfaces are used alone. The essential properties of high resil-ience (on the bottom) and viscoelastic foams are character-ized by ball rebound, density, CLD hardness, and SAG-index.

Mattress CoversMattress covers are also important and may modify the

functionality of the mattress, especially microclimate manage-ment.12 A cover can ruin the functionality of a HSFM, but a good cover, as defined below, cannot transform a conventional foam mattress into a HSFM. Thus, a HSFM needs a cover with clearly specified properties. Studies on mattress cov-ers are lacking, but generally used minimum requirements include the following: The cover needs to be breathable (ie, the microclimate control and enough elasticity that it does not impair the functionality of mattress or cause a

Table 2. Key properties of higher specification foam mattresses (HSFM)

Specification Foam layer, bottom Foam layer, top Middle layer(s) Comment

Main functionalities of foams

Immersion control; supports body and prevents bottoming

out

Envelopment; ca-pability to deform

around and encom-pass the contour of

the body

Improves immersion control and prevents

bottoming out in heavy patients. Re-

duces point pressures

Middle layer is needed for patients

weighing >100 kg

Type of foam High resiliency (HR) foam

Viscoelastic foam HR/ Viscoelastic/ Other

A huge number of various soft foams are

available

Ball rebound (%) ≥60 <20 According to the desired function

A measure of resil-ience

Density (kg/m³) ≥40 >50 ≥40 The dense, the more durable

CLD hardness at 40% (kPa)

4–5 1–2.5 HR: 4–5 Viscoelastic: 2–3

Main measure of im-mersion control

Progressive hardness (SAG-index)

2.7 –3.1 <2.4 According to desired function

SAG-index = IFD at 65% / IFD at 25%

Tensile strength (kPa) >80 >80 >80 Describes durability of foam

Elongation at break (%)

>115 >115 >115 Describes durability of foam

Layer thickness (cm) Minimum 8 (at least 3.2 inches)

Minimum 4–5 (at least 1.6–2.0 inches)

According to desired functiona

Layers must be glued together since it en-hances the durability

Production tolerance (%)

±10 ±10 ±10 ±10% is today an extremely demanding quality requirement

IFD=indentation force deflection; a can reduce the thickness of top and bottom layers

DO NOT D

UPLICATE

Page 8: An Overview of Polyurethane Foams in Higher Specification ...Polyurethane foams. Soft polyurethane foams are com-plex, plastic-like materials that result from chemical reactions of

FEBRUARY 2015 OSTOMY WOUND MANAGEMENT® 45www.o-wm.com

HIGHER SPECIFICATION FOAM MATTRESSES

hammocking effect). For hygiene reasons, the cover also must be impermeable to biological fluids; the cover ma-terials should be resistant to urea in urine and sweat and organic acids as well as disinfectant antimicrobial agents. Both the cover and the mattress need to be fire safe ac-cording to the local regulations.

DiscussionA large number of soft polyurethane foam grades are suit-

able for mattress use.1,2,18,29 In the health care setting, mat-tresses are used continuously, and the durability of the foams and mattresses needs to better than in home use. Measures of durability (eg, density, tensile strength, and elongation capability) and the construction of the mattress layers are important. The cells within the foams resemble microscopic mechanical springs and gradually lose their functionality and thickness because of fatigue.20,22,47,48 Foam mattress manufac-turers claim a 3-year to 5-year lifespan for conventional mat-tresses in continuous use.48 Over time, the comfort and capa-bility of standard foam mattresses to prevent PUs becomes impaired, although the mattress may look intact.47,48 Profiling or scoring the top layer of the foam may improve its immer-sion and envelopment properties, but profiling and use by patients weighing >100 kg causes extra stress to the foam and mattress structures, and this may reduce the mattress’ life ex-pectancy.22,34,47,48 The foams used in HSFM should have a life span of at least 5 years. To prevent bottoming out of patients weighing >100 kg, a third, middle layer is needed if the aim is to maintain the mattress thickness at a level that fulfills the distance requirement from the mattress to the top of the side rail49 to reduce the possibility of a patient accidentally falling from the bed.

When the mattress contacts the body surface, pressure is generated and causes physiological changes in the tissues.35

One of the main changes is reduced tissue oxygen tension (to approximately 30 mm Hg), with marked individual varia-tions. The values can be extremely low, but it is not known whether this affects the individual’s susceptibility to PU de-velopment.3,32,33,36-38 The tissue oxygen tensions are similar in low-end and high-end mattresses and are a function of the internal tissue pressures created, temperature and capillary flow profiles, and other circumstances. The skin-mattress contact raises the local temperature and modifies capillary flow; this phenomenon is exacerbated by any general infec-tion the patient may have, because proinflammatory cyto-kines and free radicals are released. A rise in body tempera-ture of 0.5˚ C increases oxygen consumption by the tissues by 6% to 7% and increases the susceptibility of the skin and subcutaneous tissues to PU.3,34,50 This disadvantageous bio-logical process can be modified by improved control of the microclimate of the mattress and its cover.12

The density, resilience, and progressive hardness (SAG-index) of a foam are the most important determinants of the immersion and envelopment properties of the foam.1,18

Immersion and envelopment, on the other hand, are essen-tial user specifications and crucial for the functionality of any mattress.12 Optimal control of immersion and envelopment is still extremely difficult to achieve with only 1 foam type without profiling5 (cutting blocks, waves, or holes into the foam) or using a thick layer of a single foam (>6 inches/15 cm). However, to ensure patient safety, the top of the side rail should be more than 8.7 inches (22 cm) above the mattress49; this limits the height of the mattress. Even if the required immersion and envelopment characteristics probably were fulfilled, the 2 other essential user specifications (the patient must be able to change his/her position without help and it must be easy for the nursing staff to change the patient’s posi-tion and to move the patient) cannot be achieved.

From the available data, the best combination to reach all essential user specifications appears to be a 2-layer construc-tion (as a minimum): a viscoelastic foam on the top and a high resilience foam on the bottom as discussed by Russell et al.4 The foams also should have a low production variation tolerance; otherwise, the mattress specifications will vary too much from one foam batch to another, and the quality speci-fications will no longer be reached.

Conclusion Based on the literature, definitive meanings and charac-

teristics/qualifications designating standard/conventional foam mattresses and HSFMs are yet to be determined. The data reviewed provide a basis for further research. In the fu-ture, product specification accuracy will be a factor of the key properties of each foam type used in mattresses. The authors hope their findings improve the essential requirements of mattresses in the implementation of patient care globally. n

AcknowledgmentThe language of the article was reviewed by Robert Paul,

MD, PhD, certified translator.

References1. King DA. Guide to flexible polyurethane foams. J Tissue Viability.

1998;8(1):19–25. 2. Randall D, Lee S. The Polyurethanes Book. New York, NY: Wiley;2003. 3. National Pressure Ulcer Advisory Panel and European Pressure Ulcer Ad-

visory Panel. Pressure ulcer prevention and treatment: clinical practice guideline. Washington, DC: National Pressure Ulcer Advisory Panel;2009.

4. Russell LJ, Reynolds TM, Park C, Rithalia S, Gonsalkorale M, Birch J, et al. Randomized clinical trial comparing 2 support surfaces: results of the preven-tion of pressure ulcers study. Adv Skin Wound Care. 2003;16(6):317–327.

5. Berthe JV, Bustillo A. Melot C, de Fontaine S. Does a foamy-block mat-tress system prevent pressure sores ? A prospective randomised clinical trial in 1729 patients. Acta Chir Belg. 2007;107(2):155–161.

6. McInnes H, Jammali-Blasi A, Bell-Syer SEM, Dumville JC, Cullum N. Sup-port surfaces for pressure ulcer prevention. Cochrane Database System Rev. 2011;4:CD001735.

7. Tweed M, Tweed C, Perkins GD. The effect of differing support surfaces on the efficacy of chest compressions using a resuscitation manikin model. Resuscitation. 2001;51(2):179–183.

8. Perkins GD, Benny R, Giles S, Gao F, Tweed MJ. Do different mattresses affect the quality of cardiopulmonary resuscitation? Intensive Care Med. 2003;29(12):2330–2335.

9. Vanderwee K, Grypdonck MH, Defloor T. Effectiveness of an alternating pressure air mattress for the prevention of pressure ulcers. Age Aging

DO NOT D

UPLICATE

Page 9: An Overview of Polyurethane Foams in Higher Specification ...Polyurethane foams. Soft polyurethane foams are com-plex, plastic-like materials that result from chemical reactions of

46 OSTOMY WOUND MANAGEMENT® FEBRUARY 2015 www.o-wm.com

FEATURE

2005;34(3):261–267.10. Noordergraaf GJ, Paulussen IW, Venema A, van Berkom PFJ, Woerlee

PH, Scheffer GJ, Noordergraaf A. The impact of compliant surfaces on in-hospital chest compressions: Effects of common mattresses and a back-board. Resuscitation. 2009;80(5):546–552.

11. Cloette G, Dellimore KH, Scheffer C, Smuts MS, Wallis LA. The impact of backboard size and orientation on sternum-to-spine compression depth and compression stiffness in a manikin study of CPR using two mattress types. Resuscitation. 2011;82(8):1064–1070.

12. National Pressure Ulcer Advisory Panel, European Pressure Ulcer Advi-sory Panel and Pan Pacific Pressure Injury Alliance. Prevention and Treat-ment of Pressure Ulcers; Clinical Practice Guideline. Emily Haesler (Ed.) Osborne Park, Western Australia: Cambridge Media; 2014.

13. Stechmiller JK, Cowan, L, Whitney JD, Phillips, L, Aslam, R, Barbul, A, et al. Guidelines for the prevention of pressure ulcers. Wound Repair Regen. 2008;16(2):151–168.

14. Australian Wound Management Association. Pan Pacific Clinical Practice Guideline for the Prevention and Management of Pressure Injury. Osborne Park, WA: Cambridge Media:2012.

15. Chou R, Dana T, Bougatsos C, Blazina I, Starmer AJ, Reitel K, Buckley DI. Pressure ulcer risk assessment and prevention. A systematic comparative effectiveness review. Ann Intern Med. 2013;159(1):28–38.

16. Levine SM, Sinno S, Levine JP, Saadeh PB. Current thoughts for the pre-vention and treatment of pressure ulcers. Using the evidence to determine fact or fiction. Ann Surg. 2013;257(4):603-608.

17. Zaretsky E, Asaf Z, Ran E, Aizik F. Impact response of high density flexible polyurethane foam. Int J Impact Engineer. 2012;39(1):1–7.

18. Polyurethane Foam Association. Joint Industry Foam Standards and Guidelines, Sections 1, 2 and 4. 1994. Available at: www.pfa.org/jifsg. Ac-cessed March 13, 2014

19. Flexible cellular polymeric materials - Determination of resilience by ball rebound. 2007; ISO 8307. Available at: www.iso.org/iso/catalogue_detail.htm?csnumber=41057. Accessed January 17, 2015.

20. Standard test methods for flexible cellular materials—Slab, bonded, and molded urethane foams. 2011; ASTM D3574. Available at: www.astm.org/Standards/D3574.htm. Accessed January 17, 2015.

21. Polyurethane Foam Association. Flexible polyurethane foam in mattress construction. In Touch. 1995;5(1):1–7.

22. Kaushiva BD. Structure-property relationships of flexible polyurethane foams. PhD Thesis. Virginia Polytechnic Institute 1999. Available at: www.google.fi/url?sa=t&rct=j&q=&esrc=s&source=web&cd=1&ved=0CB4QFjAA&url=http%3A%2F%2Fscholar.lib.vt.edu%2Ftheses%2Favailable%2Fetd-083199-185156%2Funrestricted%2FKAUSHIVA1.PDF&ei=9aC6VJaaC-fVygPCyIHwAw&usg=AFQjCNGnamR7aXah_R3dN12zwCdgroMnSg&bvm=bv.83829542,d.bGQ. Accessed January 17, 2015.

23. Polyurethane Foam Association. Flexible polyurethane foam. The impor-tance of density. In Touch. 1991;1(2):1–4.

24. Polymeric materials, cellular flexible - Determination of stress-strain char-acteristics in compression - Part 1: Low-density materials (ISO 3386-1:1986 + Amd 1:2010); German version EN ISO 3386-1:1997 + A1:2010 (Foreign Standard). 2010; DIN EN ISO 3386-1. Available at: http://shop.bsigroup.com/ProductDetail/?pid=000000000030180438. Accessed Jan-uary 17, 2015.

25. Polyurethane Foam Association. Flexible polyurethane foam: a primer. In Touch. 1994;4(3):1–5.

26. Polyurethane Foam Association. Information on flexible polyurethane foam. In Touch 2003;11(1):1–7

27. Flexible cellular polymeric materials. Determination of hardness (indentation technique). 2008; BS EN ISO 2439. Available at: http://shop.bsigroup.com/ProductDetail/?pid=000000000030148754. AccessedJanuary 17, 2015.

28. Polyurethane Foam Association. Flexible polyurethane foam: a primer. In Touch. 1993;3(1):1–4.

29. Polyurethane Foam Association. Flexible polyurethane foam: a primer. In Touch. 1991;1(1):1–8.

30. Tensile strength and elongation at break of flexible cellular polymeric materials. 1999; ISO 1798. Available at: ww.iso.org/iso/catalogue_detail.htm?csnumber=41059. Accessed January 17, 2015.

31. Hilfsmittelverzeichnis. (requirements or reimbursement of medical devices in Germany). Available from: www.gkv-spitzenverband.de/krankenver-sicherung/hilfsmittel/hilfsmittelverzeichnis/hilfsmittelverzeichnis.jsp, Ac-cessed March 3, 2014.

32. Gefen A. Bioengineering models of deep tissue injury. Adv Skin Wound Care. 2008;21(1):30–36.

33. Levy A, Kopplin K, Gefen A. An air-cell-based cushion for pressure ulcer protection remarkably reduces tissue stresses in the seated buttocks with respect to foams: finite element studies. J Tissue Viability. 2014;23(1):13–23.

34. Brienza D, Geyer M. Understanding support surface technologies. Adv Skin Wound Care. 2000;13(5):237–244.

35. Jonsson A, Linden M, Lindgren M, Malmqvist LÅ, Bäcklund Y. Evaluation of antidecubitus mattresses. Med Biol Eng Comput. 2005;43(5):541–547.

36. Bader D, Oomens C. Recent advances in pressure ulcer research. In: Ro-manelli M, Clark M, Cherry G, Colin D, Defloor T, eds. Science and Practice of Pressure Ulcer Management. London, UK: Springer;2006:11–26

37. Takala J, Soini H, Soppi E, Kataja M, Olkkonen K. Voidaanko painehaavau-man riskitekijöitä vähentää erikoispatjalla? (Can the risk factors for the development of pressure ulcers be reduced?). Finnish Med J Duodecim. 1994;110(4):407–414.

38. Soppi E. Painehaava – esiintyminen, patofysiologia ja ehkäisy (Pressure ul-cers, occurence, pathophysiology and prevention. A review). Finnish Med J Duodecim. 2010;126(3):261–268.

39. Soini HO, Takala J. Measurement of tissue oxygen tension: comparison be-tween two subcutaneous oxygen tonometers. J Clin Monit.1991;7(3):227–231.

40. Soppi E. Functionality of different type of mattresses in the prevention of pressure ulcers. Presentation at the 16th European Pressure Ulcer Advi-sory Panel Meeting. Wien, Austria August 28–30, 2013.

41. Takala J, Varmavuo S, Soppi E. Prevention of pressure sores in acute respira-tory failure: a randomized, controlled trial. Clin Int Care. 1996;7(5):228–235.

42. Hofman A, Geelkerken RH, Wille J, Hamming JJ, Hermans J, Breslau PJ. Pressure sores and pressure-decreasing mattresses: controlled clinical trial. Lancet. 1994; 343(8897):568–571.

43. Gray DG, Campbell M. A randomized clinical trial of two types of foam mattresses. J Tissue Viability. 1994;4(4):128–132.

44. Santy JE, Butler MK, Whyman JD. A comparison study of 6 types of hos-pital mattress to determine which most effectively reduces the incidence of pressure sores in elderly patients with hip fractures in a District Gen-eral Hospital. Report to Northern and Yorkshire Regional Health Authority. 1994. Available through special libraries.

45. Collier ME. Pressure-reducing mattresses. J Wound Care. 1996;5(5):207–211.46. Gray DG, Smith M. Comparison of a new foam mattress with the standard

hospital mattress. J Wound Care. 2000;9(1):29–31.47. Heule EJ, Goossens RH, Mugge R, Dietz E, Heule F. Using an indenta-

tion measurement device to assess foam mattress quality. Ostomy Wound Manage. 2007;53(11):56–62.

48. Gray DG, Cooper PJ, Campbell M. A study of the performance of a pres-sure reducing foam mattress after three years of use. J Tissue Viability. 1998;8(3):9–13.

49. Medical electrical equipment — particular requirements for the basic safe-ty and essential performance of medical beds. 2009; ISO EN 60601-2-52. Available at: www.iso.org/iso/catalogue_detail.htm?csnumber=36067. Accessed January 17, 2015.

50. Soppi E, Iivanainen A, Korhonen P. Concordance of Shape Risk Scale, a new pressure ulcer risk tool, with the Braden scale. Int Wound J. 2012;DOI: 10.1111/iwj.12015.