Page 1: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 1

An Introduction to Mass Spectrometry

Guy Wilson UPC2 & Purification Business Development,

Northern Europe, Waters Ltd

Page 2: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 2

What is a Mass Spectrometer?

Sample Introduction

Ion Source Mass Analyser

Detector

Data System

LC, GC etc. Mass Spectrometer

Page 3: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 3

Atmospheric Pressure Ionisation

Atmospheric Pressure Ionisation (API)

– A term for any ionisation process that is carried out at

atmospheric pressure

The most common modes of API are…

– Electrospray Ionisation (ESI)

o Most commonly used

– Atmospheric Pressure Chemical Ionisation (APCI)

o Less commonly used than ESI but still significant usage

– Atmospheric Pressure Photo Ionisation (APPI)

o A more niche ionisation mode for certain compounds that will

not ionise by either ESI or APCI.

Page 4: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 4

Electrospray Ionisation (ESI)

Page 5: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 5

Electrospray Ionisation Overview

Page 6: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 6

Electrospray Ionisation Producing Charged Droplets

Liquid Flow

Electrospray Probe Tip at high voltage

Taylor Cone

More Positive Ions than Negative Ions

Positively Charged Droplets

More Negative Ions than Positive Ions

Page 7: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 7

Electrospray Ionisation Fission of Charged Droplets

Droplets produced from the spray have a surface charge

Surface charged droplets undergo solvent evaporation and

droplet fission to produce smaller droplets

Like charge repulsion becomes greater than droplet surface

tension and fission occurs to produce smaller charged

droplets

+

+

+

+

+

+

+

+

+

+

+

+

+ +

+

+

+

+

+ + + +

+

+

+

+

+ Coulombic Fission

+

+

+

Solvent Evaporation

Page 8: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 8

Electrospray Ionisation Formation of Gas Phase Ions

Ion Desorption Mechanism

Charged Residue Mechanism

+

+

+

+

+

+

+

+

+

+

+

+

+

+

+

+

+

+

+

+ +

+

+ +

+

+

+

+

+

+

+ + +

+ + +

+ + + +

+

+

+ +

+ +

+ +

+

+

+

+

+

+

+

+

+

+ +

+ +

Page 9: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 9

Electrospray Ionisation Z-spray Ion Source

Sample

Nebulizer Gas

Desolvation Gas

Source Exhaust

Cone Gas

Roughing Pump Turbomolecular Pump Extraction Cone

ESI Probe

Sample Cone

Isolation Valve

RF Lens (Hexapole) Quadrupole

Atmospheric Pressure

Moderate Vacuum

Full Vacuum

Page 10: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 10

Atmospheric Pressure Chemical Ionisation (APCI)

Page 11: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 11

APCI – Overview

Sample Solution

Nebulizer Gas

APCI Probe Heater (400-650°C)

Flash Vaporisation

Corona Discharge Needle

Sample Cone

Corona Discharge

Liquid flow is forced through a narrow capillary to give it a

high linear velocity

The APCI Probe heater combined with nebulizer gas then

vaporises the liquid flow

The solvent and analyte vapour passes though the corona

discharge region to produce gas phase ions

Solvent and Analyte Vapour

Page 12: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 12

Atmospheric Pressure Photo Ionisation (APPI)

Page 13: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 13

APPI – Overview

Sample Solution

Probe Heater

Flash Vaporisation

Vacuum UV Source

Sample Cone

APPI uses the same probe as APCI with a Vacuum UV light

source instead of the corona discharge needle

As with APCI:

– Liquid flow is forced through a capillary to give it a high linear

velocity

– The probe heater and nebulizer gas vaporise the liquid flow

Ionisation occurs by either direct or chemical ionisation type

processes in the region of the light source

Nebulizer Gas

Solvent and Analyte Vapour

Repeller Electrode

h

Page 14: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 14

ESI / APCI / APPI Comparison

ESI

APCI

APPI

Page 15: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 15

Universal Source platform

Also compatible with 3rd party source options:

Phytronix LDTD — IonSense DART Prosolia DESI — Advion NanoMate — Protea LAESI

ESI APCI ESCi™

APPI APCI

APGC TRIZAIC UPLC®

nanoFlow™ ESI

ASAP

Page 16: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 16

Quadrupole Theory

Pre-filter Quadrupole Mass Filter Post-filter

Rejected Ions Stable (Resonant) Ions

Page 17: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 17

Single Quadrupole MS

Page 18: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 18

Tandem Quadrupole MS

Page 19: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 19

Ele

ctr

ic F

ield

Diffuse Ion

Cloud

Maximising signal

Maximising robustness

Transfer Region Ion Optics

Page 20: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 20

Traveling Wave Ion Transport

RF

RF

Page 21: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 21

Traveling Wave Ion Transport

Time

Ions

Traveling Wave Pulse

Page 22: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 22

Different Types of Experiments

MS2

Collision Cell MS1

Page 23: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 23

MS2 Scan

Works in the same way as MS Scan except:

– MS1 and Collision Cell act only as Ion guides

– MS2 is the scanning quadrupole

Minimises distance and amount of ion optics to interact with

between mass resolved ions and the detector

Better performance for fast scanning (> 2,000 Da/sec)

MS1 Ion Guide (RF only)

MS2 Scanning

T-wave Collision Cell

Page 24: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 24

MS Scan - Example

TIC

Ex

tracte

d M

ass C

hro

m

Mass S

pectr

um

Page 25: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 25

SIR – Single Ion Recording

Only selected ions are transmitted through the instrument

and are monitored constantly

Due to the instrument not acquiring other ions considerably

more time is spent on each selected ion making this a much

more sensitive acquisition mode than scanning acquisition

modes

No mass spectra are produced for this type of acquisition

MS1 Static – Single Ion

MS2 Ion Guide (RF only)

T-wave Collision Cell No Gas

Page 26: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 26

SIR - Example

S/N = 109

S/N = 38

S/N = 101

S/N = 129

10ng/m

L o

f each c

om

ponent

Page 27: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 27

MRM – Multiple Reaction Monitoring

A selected ion is transmitted through the first quadrupole

(precursor ion), fragmented in the collision cell, and a

specified fragment ion is then transmitted through the

second quadrupole (product ion).

More selective and sensitive than SIR due to

– Specific transition needed for response to be seen

– Less interference by background ions of the same mass

No mass spectra are generated by MRM experiments

MS1 Static – Single Ion

MS2 Static – Single Fragment Ion

T-wave Collision Cell Gas on

Page 28: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 28

MRM - Example

S/N = 357

S/N = 772

S/N = 6600

S/N = 1234

10ng/m

L o

f each c

om

ponent

Page 29: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 29

Selected ions are transmitted through the first quadrupole

(precursor ions) and fragmented in the collision cell

MS2 is then scanned over a user defined mass range

A mass spectrum of the product ions generated by

fragmentation is acquired at each time point throughout the

acquisition.

MS1 Static – Single Ion

MS2 Scanning

T-wave Collision Cell Gas on

Product Ion Scanning

Page 30: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 30

Product Ion Scanning - Example

Page 31: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 31

MS1 is scanned over a specified mass range and all ions are

sequentially passed through to the collision cell where they

are fragmented

MS2 is set to transmit only the mass of a specific fragment

ion and does not scan

Any ions that fragment to give the specified product ion will

generate a result.

MS1 Scanning

MS2 Static – Single Fragment Ion

T-wave Collision Cell Gas on

Precursor Ion Scanning

Page 32: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 32

Precursors of 156 Da

Precursors of 85 Da

Precursor Ion Scanning - Example

TIC

Precurso

r Ion

Mass

Spectra

Page 33: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 33

MS1 is scanned over a specified mass range and all ions are

sequentially passed through to the collision cell where they

are fragmented

MS2 is scanned in sync with MS1 over the same mass range

minus an offset

A response in only seen if a precursor ion loses a neutral

fragment in the collision cell of the same mass to charge

ratio as the offset between MS1 and MS2

MS1 Scanning

MS2 Scanning with mass offset

T-wave Collision Cell Gas on

Constant Neutral Loss

Page 34: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 34

Xevo TQ-S micro

Xevo TQD

SQD 2

Waters Quadrupole Mass Spectrometers

Xevo TQ-S

Page 35: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 35

Time Of Flight MS

Time-of-flight mass spectrometry (Tof-MS) is probably the

simplest method of mass measurement to conceptualise,

although there are hidden complexities when it comes to higher

resolution instruments.

The inherent characteristics of Tof MS are extreme sensitivity

(all ions are detected), almost unlimited mass range, speed of

analysis (modern instruments can obtain >10 full spectra per

second) and sub 5ppm mass accuracy.

35

Page 36: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 36

Time Of Flight Theory

A voltage pulse applied to a pusher plate (typically, 800-1000V)

initiates an orthogonal acceleration TOF (oa-TOF) m/z

measurement

The energy imparted by this pulse in combination with an

electric field is converted into kinetic energy (KE)

The time it takes an ion to travel (TOF) a given distance (d) is

proportional to its mass-to-charge ratio or m/z

Page 37: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 37

Time Of Flight Theory

Pusher Detector

Time from Push

Page 38: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 38

Maximising Resolution

Page 39: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 39

Accelerating the Ions

P u sher

Ideal Case: perfectly straight line no orthogonal velocity component

P u sher

Spatial distribution: some ions are closer to the plate than others

P u sher

Energy distribution: small orthogonal velocity component

Page 40: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 40

Waters TOF Mass Spectrometers

Xevo G2-XS QTof

Synapt G2-Si HDMS

Page 41: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 41

Xevo G2-XS

Page 42: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 42

LC/MSE Data Acquisition

Collision cell cycles between low (MS of peptides, 4V)

and elevated (MSE of fragments, 20 45 V) energy No selection of peptide ions

MS

MSE

UPLC

• Global analysis: acquire data on all the ions • Maximize the LC/MS duty cycle • Minimize bias/selection of ions ( Reproducibility) • Obtain qualitative and quantitative information from the same analysis run

Page 43: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 43

SYNAPT G2-Si HDMS

Page 44: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 44

Ion mobility

Page 45: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 45

Ion Mobility Separation (IMS)

+ +

+

+

+

+

+ +

+

Size Shape Charge

The mobility of an ionised molecule is dependant on its…

Page 46: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 46

“Classic” Ion Mobility Spectrometry

The Bowers Group Website [Ion Mobility Theory]:

http://bowers.chem.ucsb.edu/theory_analysis/ion-mobility/index.shtml

Page 47: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 47

TriWave Device

Trap

He Cell

IMS Transfer

Page 48: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 48

iNFORMATION Selectivity (and Specificity)

Retention CCS Mass Peak

capacity

QuanTof UPLC/HDMS

Page 49: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 49

Mass Detection At Its Most Advanced

Page 50: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 50

Know your peaks

Page 51: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 51

Complementary & Compatible

Information-rich mass spectral data

Targeted for small molecule applications

Complementary to and compatible with PDA

– UPLC, UPC2, purification

– Optional ISM module to split and dilute

Intuitive system for chromatographers

• 30-1250 DA • PRE-OPTIMIZED ES± • 20 HZ FS • 100 HZ SIR • 4 ORDERS • INTUITIVE •

Page 52: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 52

6.5 mins

22 mins

Page 53: An Introduction to Mass Spectrometry - Waters Corporation accounts_an introduction to mass...Time-of-flight mass spectrometry (Tof-MS) is probably the simplest method of mass measurement

©2013 Waters Corporation 53

Thank You