An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance David Followill IROC Houston QA Center

Transcript of An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by...

Page 1: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance

David FollowillIROC Houston QA Center

Page 2: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

Outline of Presentation

• Definition of the problem• Results of audits• Possible solutions• Resources to help

Page 3: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

Learning Objectives

• Recognize that small field dosimetry is not easy and that there are tools to help you be as accurate as possible.

Page 4: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

Sometimes we all like to believe that all across the kingdom it’s bright and merry!

Page 5: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

Ladies and LordsThere is a scourge across the Kingdom

Small field dosimetry

Page 6: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

What is the truth?

From Das et al 2000

Page 7: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

Recommendation:

The unshielded stereotactic diode and micro-chambers are the detectors of choice formeasurements in radiosurgery beams. However, werecommend comparing the measurements from atleast two different detectors listed in Table 2a, andafter applying corrections taking their average.

TG-155Small Field Dosimetry

Page 8: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

TomoTherapy results (since Aug 2010)

0

20

40

60

80

100

120

0.9 0.91 0.92 0.93 0.94 0.95 0.96 0.97 0.98 0.99 1 1.01 1.02 1.03 1.04 1.05 1.06 1.07 1.08 1.09 1.1

# re

sults

TLD/Institution

Average: 0.970 +/- 0.024552 irradiations

Page 9: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

But the manufacturer told me it would work, the manufacturer set the value or I sent my own measured data to the TPS company and they modeled my beams for me.

1. Accuray calibrates all of the tomotherapy units using the Exradin A1SL chamber and the user monitors constancy.

• Correction factor of 0.980 for A1SL based on Gago-Arias et al, Med Phys, 2012

Page 10: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

Manufacturer modeling of small field dataYou make the measurementsYou send the measurementsThey create the beam model

BUTThey assume you used the correct dosimeter

They assume you made measurements correctlyThey do not know to apply additional corr. factors

RESULTIncorrect Small Beam Model

Page 11: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

Francescon et al2011 data

Small Field Dosimetry Corrections

Situation is even worse if you consider using field sizes less then 0.5 x 0.5 cm2

Page 12: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

Field Size (cm x cm)

Detectors 0.5x0.5 0.75x0.75 1.0x1.0 1.25x1.25 1.5x1.5 3.0x3.0

PTW 60012-60017 0.968±0.003 0.984±0.002 0.995±0.001 1.001±0.001 1.006±0.001 1.013±0.001

Sun Nuclear Dedge 0.932±0.003 0.951±0.002 0.967±0.001 0.978±0.003 0.986±0.002 1.001±0.002

IBA SFD 0.972±0.003 0.996±0.002 1.007±0.001 1.011±0.001 1.015±0.001 1.017±0.001

Exradin D1V 0.980±0.003 0.994±0.002 0.999±0.001 1.000±0.001 1.005±0.001 1.012±0.001

Exradin A16 1.112±0.018 1.044±0.007 1.020±0.001 1.007±0.001 1.002±0.001 0.999±0.001

PTW PinPoint 31014 1.128±0.018 1.053±0.007 1.024±0.002 1.010±0.001 1.005±0.001 1.000± 0.001

PTW MicroLion 1.023±0.009 0.997±0.002 0.993±0.001 0.992±0.001 0.994±0.001 0.998±0.001

Exradin W1 PSD 0.998±0.003 0.995±0.002 0.996±0.002 0.996±0.002 0.994±0.002 0.994±0.002

calculated as the mean over all the values obtained by changing the linacmodel, (PrimusTM Siemens 6 MV and Synergy® Elekta 6 MV), the radial FWHM andenergy of the electron source as presented by Francescon et al.97 These values canalso be used with caution for Varian 6 MV Linacs.

clin 10x10clin 10x10

f , fQ ,Qk

Unpublished data from P. Francescon

Page 13: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

Gamma Knife results (since Sep 2007)

0

10

20

30

40

50

60

0.90 0.91 0.92 0.93 0.94 0.95 0.96 0.97 0.98 0.99 1.00 1.01 1.02 1.03 1.04 1.05 1.06 1.07 1.08 1.09 1.10

# re

sults

TLD/Inst.

Average: 0.997 +/- 0.021266 results

Page 14: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

Depending on the dosimeter used, the magnitude of the additional correction, down to a 0.5 x 0.5

cm2 field size can be as much as:

6%91%1%2%1% 1. 5%

2. 2%3. No correction needed4. 13%5. 8%

Page 15: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

The answer is:

13%

The magnitude of the additional correction, down to a 0.5 x 0.5 cm2 field size can be as much as:

Francescon et al. Medical Physics, Vol. 38, No. 12, December 2011

Page 16: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

Recommendation:

It is strongly recommended to independently verifydosimetric measurements in small fields, eitherthrough measurements carried out by a differentperson and/or though an independent external audit,such as that carried out by the RPC .

TG-155Small Field Dosimetry

Page 17: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

RPC Measurements during onsite visitsMeasurement configuration:

depth = 10 cm, 100 cm SSD, Exradin A16Secondary jaws kept at a 10 x 10

MLC @ 10 x 10MLC @ 6 x 6MLC @ 4 x 4MLC @ 3 x 3MLC @ 2 x 2

Page 18: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

The Problem is that our Dragon is very small!

Page 19: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

Tables of standard small field factors

Followill et al 2012 and also will be in TG-155

We even make mistakes

Page 20: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

Corrected Varian DataVarian 6 MV Varian 10 MV Varian 15 MV Varian 18 MV

Field Size

(cm x cm) RPC Institution RPC Institution RPC Institution RPC Institution

10 x 10 1.000 1.000 1.000 1.000 1.000 1.000 1.000 1.000

6 x 6 0.938 0.945 0.959 0.964 0.963 0.964 0.969 0.971(0.005) (0.008) (0.002) (0.003) (0.005) (0.007) (0.003) (0.006)

[0.8%] [0.5%] [0.4%] [0.5%](n=127) (n=22) (n=32) (n=41)

4 x 4 0.886 0.900 0.916 0.927 0.927 0.928 0.928 0.933(0.006) (0.013) (0.006) (0.010) (0.006) (0.009) (0.002) (0.009)

[1.8%] [1.2%] [0.7%] [0.9%](n=122) (n=25) (n=32) (n=37)

3 x 3 0.851 0.870 0.880 0.894 0.892 0.894 0.884 0.891(0.007) (0.012) (0.006) (0.008) (0.006) (0.012) (0.005) (0.019)

[2.4%] [1.6%] [0.9%] [1.4%](n=123) (n=25) (n=31) (n=41)

2 x 2 0.804 0.825 0.823 0.838 0.824 0.823 0.806 0.804(0.008) (0.020) (0.005) (0.015) (0.011) (0.040) (0.007) (0.020)

[2.9%] [2.0%] [2.2%] [1.6%](n=136) (n=23) (n=33) (n=40)

 

Varian 6 MV Varian 10 MV Varian 15 MV Varian 18 MV Field Size (cm x cm) RPC Institution RPC Institution RPC Institution RPC Institution

10 x 10 1.000 1.000 1.000 1.000 1.000 1.000 1.000 1.000

6 x 6 0.938

(0.937)* 0.945 0.959 0.964 0.963

(0.965) 0.964 0.969 0.971 (0.005) (0.008) (0.002) (0.003) (0.005) (0.007) (0.003) (0.006) [0.8%] [0.5%] [0.4%] [0.5%]

(n=127) (n=22) (n=32) (n=41)

4 x 4 0.886

(0.885) 0.900 0.916 0.927 0.927

(0.924) 0.928 0.928 0.933 (0.006) (0.013) (0.006) (0.010) (0.006) (0.009) (0.002) (0.009) [1.8%] [1.2%] [0.7%] [0.9%]

(n=122) (n=25) (n=32) (n=37)

3 x 3 0.851

(0.849) 0.870 0.880 0.894 0.892

(0.884) 0.894 0.884 0.891 (0.007) (0.012) (0.006) (0.008) (0.006) (0.012) (0.005) (0.019) [2.4%] [1.6%] [0.9%] [1.4%]

(n=123) (n=25) (n=31) (n=41)

2 x 2 0.804

(0.802) 0.825 0.823 0.838 0.824

(0.814) 0.823 0.806 0.804 (0.008) (0.020) (0.005) (0.015) (0.011) (0.040) (0.007) (0.020) [2.9%] [2.0%] [2.2%] [1.6%]

(n=136) (n=23) (n=33) (n=40)

Page 21: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

Discrepancies Regarding:Number of InstitutionsReceiving rec. (n = 206)

Review QA Program 152 (74%)Photon Field Size Dependence 138 (67%)

Wedge Factor (WF) 66 (32%)Off-axis Factors (OAF)/Beam symmetry 60 (29%)

Electron Calibration 35 (17%)Photon Depth Dose 33 (16%)Electron Depth Dose 25 (12%)Photon Calibration 16 (8%)

Discrepancies Discovered (Jan. ’05 – April ’13)

On-Site Dosimetry Review Audit

This is a beam measurement issue and TPS beam modeling challenge.

Page 22: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

There is guidance or possibly a fair maiden after the dragon is slain.

The bigger challenge is with Varian machines due to the tertiary collimation (MLC)• Varian/Pinnacle combination:

Followill et al, JACMP, vol. 13, No. 5, 2012

• Varian/Eclipse combination:Kron et al, Med Phys, vol 39 (2), 2012

Page 23: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

Huge Resource of Published data:

239 references

5 Tables

15 Figures

13 Conclusions and recommendations

TG-155Small Field Dosimetry

Page 24: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

New IROC Houston audit

OSLD

Page 25: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

Commissioning and Testing

Three dosimeters were used to estimate the dose for each field size: Exradin A16Exradin D1V diodeExradin D1V diodePTW CC04Exradin W1

Page 26: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

First Test Irradiation away from RPC

Page 27: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

All of the following are current methods to provide a verification of your small field

dosimetry data except:

2%4%86%8%1% 1. Use of published data

2. Your 2nd set of measurements3. Rely on manufacturer’s data4. Remote audit5. Indep. meas. from another physicist

Page 28: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

The answer is:

Reliance on the manufacturers data

All of the following are current methods to provide a verification of your small field dosimetry data except:

Ref: Das et al, AAPM TG 155 report, submitted to Med Phys

Page 29: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

• The challenge continues for the <2 x 2 cm2 field sizes

• The correct dosimeter is essential.

• There are independent audits and standard data to help.

• We will defeat thisdragon!

Summary|Conclusion

Page 30: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

Thank YouQuestions?

Page 31: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

ReferencesFrancescon et al. Medical Physics, Vol. 38, No. 12, December 2011

Das et al, AAPM TG 155 report, submitted to Med Phys

Followill et al. J Appl Clin Med Phys 13(5):282-9, 2012.

Followill DS: Erratum, J Appl Clin Med Phys 15 (2): 4757 3/2014.

Page 32: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement

AcknowledgementsWork supported by PHS grant CA10953 awarded by NCI, DHHS